polyfocal intraocular lens: 6 months results from

23
Polyfocal intraocular lens: 6 months results from observational e-registry Stodulka P. 1 , Urminsky J. 2 , Hlozanek M. 3 ,Krizova D 3 1. GEMINI Eye Clinic, Zlin, Czech Republic 2. Department of Ophthalmology, T.Bata Regional Hospital Zlin, Czech Republic 3. Department of Ophthalmology, University Hospital Kralovske Vinohrady and 3rd Medical Faculty Prague, Czech Republic Presenting author is a consultant to Bausch+Lomb

Upload: others

Post on 19-May-2022

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Polyfocal intraocular lens: 6 months results from

Polyfocal intraocular lens:

6 months results from

observational e-registry

Stodulka P.1,

Urminsky J.2, Hlozanek M.3 ,Krizova D 3

1. GEMINI Eye Clinic, Zlin, Czech Republic 2. Department of Ophthalmology, T.Bata Regional Hospital Zlin, Czech Republic 3. Department of Ophthalmology, University Hospital Kralovske Vinohrady and

3rd Medical Faculty Prague, Czech Republic

Presenting author is a consultant to Bausch+Lomb

Page 2: Polyfocal intraocular lens: 6 months results from

Which way in IOL development?

„Technical“ or „Bioanalogic“?

Technical Bioanalogic

Page 3: Polyfocal intraocular lens: 6 months results from

WIOL-CF: the next step in IOL evolution

original Ridley lens

Page 4: Polyfocal intraocular lens: 6 months results from

WIOL-CF Intraocular lens

• Bioanalogic intraocular lens full size

no haptics design

Polyfocality leads to the range of foci

coresponding to a range of refractive powers

diameter 8.6 – 9.0 mm

• Polyfocal hyperbolic optics: refractive power is maximal in

the center and continuously

decreases without steps to

periphery

• Material: WIGEL® biocompatible hydrogel

42% water

refractive index 1.43

thickness 0.8–1.7 mm

Page 5: Polyfocal intraocular lens: 6 months results from

WIOL-CF intraocular lens pedigree

Prof. Otto Wichterle, Czech scientist,

inventor of the soft contact lenses

Technology sold to US in 1960, laid foundation of B&L contact lens production

More at: http://en.wikipedia.org/wiki/Otto_Wichterle

Page 6: Polyfocal intraocular lens: 6 months results from

WIOL-CF in the eye

Page 7: Polyfocal intraocular lens: 6 months results from

Implantation of WIOL-CF

Page 8: Polyfocal intraocular lens: 6 months results from

Young crystalline lens = the ideal IOL

WIOL-CF

Design Inspired by Nature

WIOL-CF Bioanalogic IOL

Hydrogel-like tissue with negative charge (carboxylate and sulphate groups)

High water content (66%), refractive index 1.42

Full size (10.5 mm) lens, smooth, glare-free optics

Hyperbolic surface featuring polyfocal optics enabling large depth of focus and accommodation

Accommodation range of up to 10 D decreasing with age

Hydrogel with negative charge (carboxylate groups) High water content (42%), refractive index 1.43

Full-optics (9 mm) lens, smooth, glare-free optics

Hyperbolic surface featuring polyfocal optics enabling large depth of focus and pseudoaccommodation

Depth of focus exceeding 2 D (data about stability for over 12 years)

Atchison DA.: Accommodation and presbyopia.,Ophthalmic Physiol Opt., Jul 1995;15(4):255-72 Pasta J. et al, Abstract p. 102, ESCRS, Munich, Germany,September, 2003 Fisher RF et al.: J Physiol. (1973),234, 443-7 Kasthurirangan S.:Investigative Ophthalmology & Visual Science, June 2008, Vol. 49, No. 6, 2531-40 Kasthurirangan S. Journal of Vision (2011) 11(3)19, 1–16

Dubbelman M. et al.: Vision Research 41 (2001) 1867–1877 Dubbelman M. et al: Vision Research 43,2003; 2363–2375 Dubbelman M. et al: Vision Research 45 ,2005; 117-132 Manns F. et al: Experimental Eye Research 78, 2004; 39–51 Pasta J. et al: Abstract No 320, ASCRS, San Francisco, CA, April 2006

Page 9: Polyfocal intraocular lens: 6 months results from

Study design:

• Prospective

• Non randomized

• Multicenter - 12 centers in the Czech Republic

• Bilateral WIOL-CF implantation

• Central data registry

Method

Page 10: Polyfocal intraocular lens: 6 months results from

Preoperative

medical history

12 month follow-up with 3-monthly

assessments (3-6-12 m)

Bilateral implantation of WIOL-CF

within 2 weeks

Retrospective Prospective

-2 weeks

Patient characteristics

Biometry

Implanted IOL

UDVA, UIVA, UNVA, CDVA

Subjective refraction

Adverse optical phenomena (glare, halo)

Contrast Sensitivity

Spectacles independence, subjective satisfaction

Patient enters registry 2 weeks after bilateral WIOL-CF implantation

2 weeks 3 months 6 months 12 months

Central data registry

Page 11: Polyfocal intraocular lens: 6 months results from

• 104 eyes of 52 patients (23 male, 29 female)

• Mean age 64.1 years, Median 65 (48-82)

• Cataract without any other intraocular pathology

• Corneal astigmatism < 1.25 cylD

• 6 months follow up

Patient data

Page 12: Polyfocal intraocular lens: 6 months results from

CDVA monocular @ 6 months

N Median Mean (D) SD

CDVA (decimal) 104 1.0 0.90 0,19

0%

50%

100%

1.0 and better 0.9 and better 0.8 and better

66% 72% 82%

0.8 and better 82%

0.9 and better 72%

1.0 and better 66%

Page 13: Polyfocal intraocular lens: 6 months results from

CNVA monocular @ 6 months

N Median Mean (D) SD

CNVA (J) 104 J1 J1.33 1.07

J3 and better 98%

J2 and better 95%

J1 51%

0%

50%

100%

J1 J2 and better J3 and better

51%

95% 98%

Page 14: Polyfocal intraocular lens: 6 months results from

UDVA binocular @ 6 months

0%

50%

100%

1.0 and better 0.9 and better 0.8 and better

72% 83% 94%

N Mean SD

52 0.95 0.14

0.8 and better 94%

0.9 and better 83%

1.0 and better 72%

Page 15: Polyfocal intraocular lens: 6 months results from

UIVA (70cm) binocular @ 6 months

N Median Mean SD

52 J1 J1.36 1.53

0%

20%

40%

60%

80%

100%

J1 J2 and better J5 and better

70% 85%

100%

J5 and better 100%

J2 and better 85%

J1 70%

Page 16: Polyfocal intraocular lens: 6 months results from

UNVA (40cm) binocular @ 6 months

N Median Mean SD

52 J2 J3.2 2.03

J5 and better 83%

J3 and better 65%

J2 and better 55%

0%

20%

40%

60%

80%

100%

J2 and better J3 and better J5 and better

55%

65%

83%

Page 17: Polyfocal intraocular lens: 6 months results from

SAFETY

CDVA & CNVA

binocular N Median Mean (D) SD

CDVA (decimal) 52 1.0 0.98 0.11

CNVA (J) 52 J1 J 1.1 1.02

monocular N Median Mean (D) SD

CDVA (decimal) 104 1.0 0.90 0.19

CNVA (J) 104 J1 J 1.33 1.08

Page 18: Polyfocal intraocular lens: 6 months results from

N Mean (D) SD

Distance sferical equivalent

104 + 0.024 0.51

Near sferical equivalent

104 + 1.56 0.93

Subjective refraction @ 6 months

Page 19: Polyfocal intraocular lens: 6 months results from

Spectacle independence

& subjective satisfaction

85% of patiens

spectacle independent

85%

15%

Not use spectacles

Use spectacles on near

85%

15%

Satisfied Not Satisfied

Not satisfied patients expected better near vision

85% of patients express subjective satisfaction

Page 20: Polyfocal intraocular lens: 6 months results from

Glare & Halo

91% of WIOL-CF patients without

disturbing optical phenomena

No explantation due to severe optical phenomena

N=52

Page 21: Polyfocal intraocular lens: 6 months results from

Contrast Sensitivity - monocular

Mesopic Conditions Photopic Conditions

N=104

Page 22: Polyfocal intraocular lens: 6 months results from

Conclusions

Excellent far and intermediate visual acuity results

Near visual acuity in the range of „social reading“ for

majority of patients

Polyfocal optics safety confirmed by BCVA results for far and

near distance

High subjective satisfaction rate

Exceptional contrast sensitivity

Low level of serious/disturbing optical phenomena

glare/halo

Page 23: Polyfocal intraocular lens: 6 months results from

Thank you for your attention!