performance of commercial reverse line blot assays for hpv
TRANSCRIPT
1
Performance of Commercial Reverse 1
Line Blot Assays for HPV Genotyping 2
3
Running Title: Performance of HPV line blot assays 4
5
Martin Steinau1*, Juanita M Onyekwuluje1, Mariela Z Scarbrough 1, Elizabeth R Unger1, Joakim 6
Dillner2, Tiequn Zhou3‡ 7
8
1 WHO Human Papillomavirus Laboratory Network Global Reference Laboratory, Chronic Viral 9
Diseases Branch, Division of High-Consequence Pathogens and, National Center for Emerging 10
and Zoonotic Infectious Diseases Pathology, Centers for Disease Control and Prevention, 11
Atlanta, GA 30333, U.S.A. 2WHO Human Papillomavirus Laboratory Network Global Reference 12
Laboratory, Region Skåne, Malmö and Departments of Laboratory Medicine, Medical 13
Epidemiology & Biostatistics, Karolinska Institute, Stockholm, Sweden. 14
3Quality, Safety and Standards Team, Immunizations, Vaccines and Biologicals Department, 15
World Health Organization, Geneva, Switzerland 16
17
18
Key Words: Human papillomavirus, HPV typing, reverse line-blot assay 19
Copyright © 2012, American Society for Microbiology. All Rights Reserved.J. Clin. Microbiol. doi:10.1128/JCM.06576-11 JCM Accepts, published online ahead of print on 22 February 2012
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
2
* Corresponding Author: 20
Martin Steinau, PhD 21
Centers for Disease Control and Prevention 22
1600 Clifton Rd, Atlanta, GA 30333 23
Ph: +01 (404) 639-0561 24
Fax: +01 (404) 639-3540 25
Email: [email protected] 26
27
Word Count: 28
Abstract: 217 29
Text: 3202 30
Tables: 5 31
Figures: 2 32
Financial Disclosures: This work was supported by the WHO via a project funded by the Bill and 33
Melinda Gates Foundation 34
Disclaimers: The findings and conclusions in this report are those of the authors and do not 35
necessarily represent the views of the supporting agencies. ‡The author is a staff member of 36
the World Health Organization. The author alone is responsible for the views expressed in this 37
publication and they do not necessarily represent the decisions or policies of the World Health 38
Organization. 39
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
3
Abstract 40
The performance of three line blot assays (LBA) - Linear Array HPV genotyping assay (LA; Roche 41
Diagnostics), INNO-LiPA HPV Genotyping Extra (LiPA; Innogenetics) and the Reverse 42
Hybridization assay (RH; Qiagen) was evaluated using quantitated whole genomic HPV plasmids 43
(types 6, 11, 16, 18, 31, 33, 35, 39, 51, 52, 56, 58, 59, 68b) as well as epidemiologic samples. In a 44
plasmid titration series LiPA and RH did not detect 50 international units (IU) of HPV 18 in the 45
presence of 5 x 104 or more IU HPV16. HPV DNA (1 - 6 types) in the plasmid challenges at 50 46
(IU) or genome equivalents (GE) were identified with an accuracy of 99.9% by LA, 97.3% by LiPA 47
and 95.4% by RH with positive reproducibility of 99.8% (Kappa = 0.992), 88.2% (Kappa = 0.928) 48
and 88.1% (Kappa = 0.926) respectively. Two instances of mis-typing occured with LiPA. Of the 49
120 epidemiologic samples 76 were positive for high-risk types by LA, 90 by LiPA and 69 by RH 50
with a positive reproducibility of 87.3% (Kappa = 0.925), 83.9% (Kappa = 0.899) and 90.2% 51
(Kappa = 0.942) respectively. Although all the assays had good concordance in the clinical 52
samples, the greater accuracy and specificity in the plasmid panel suggest that the LA assay has 53
an advantage for internationally comparable genotyping studies. 54
55
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
4
Introduction 56
The introduction of HPV vaccines has highlighted the need for standardized accurate HPV 57
genotyping assays to assure comparability of results in laboratories world-wide, both before 58
and after vaccine introduction. A large variety of assays, using both in-house methods and 59
commercial kits, are available for HPV detection and typing (6), and more can be anticipated. 60
International proficiency testing organized by the WHO HPV LabNet has documented significant 61
differences in sensitivity, specificity and reproducibility between different laboratories using a 62
variety of testing platforms (2). Variation in performance for laboratories using the same assay 63
implicates laboratory practice; however the results also indicated differences between assays. 64
The WHO HPV LabNet has further recognized the need for HPV typing assays that can easily be 65
standardized, require minimal equipment for assay performance and interpretation, and which 66
can be used in a variety of laboratory settings. Following its meeting on the standardization of 67
HPV assays and the role of WHO HPV LabNet in supporting vaccine introduction (WHO/HQ, 68
Geneva, Switzerland, 23-25 January 2008, Ref: 69
http://www.who.int/biologicals/publications/meetings/areas/vaccines/human_ papillomavirus 70
/HPV%20Jan%20meeting%20report_20080909%20_Clean_.pdf) it was agreed that commercial 71
assay kits should be evaluated in collaborative studies for proficiency. 72
Reverse line blot assays (LBA), based on consensus amplification of conserved regions of HPV 73
followed by hybridization to type specific probes on line blot strips, have been widely used. 74
Available LBAs differ in the primer set (1,3,4) used in the amplification phase and in number and 75
sequence of the detection probes. Beyond thermocyclers for target amplifications, only 76
temperature control water baths and visual inspection are required to carry out these assays. 77
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
5
Therefore the LBA platforms have the potential to meet the needs of the WHO for high 78
performance genotyping that does not involve expensive equipment. 79
Studies have been conducted previously to compare and evaluate performance of these HPV 80
typing assays. Typically, these were restricted to DNA extracts from patient-collected anogenital 81
specimens, which cannot be validated independently and limit the analysis to relative 82
comparisons and assessments of type prevalence found by the individual assays (5,7,10). The 83
use of plasmid standards offers clear advantages for more objective evaluations beyond that 84
level. Wittingly, cloned, full-length HPV genomic DNA provides complete control over genotype 85
identity as well as copy number input and eliminates the need for a gold standard test. 86
We selected three commercial LBAs using different primer sets and subjected them to a 87
number of different test samples. These included plasmid standards as well as patient extracts 88
from different sources. The objective was a comparison of these kits’ performance with the 89
potential needs of the WHO in mind. Specifically, assay sensitivity, specificity and 90
reproducibility were assessed. 91
92
Materials and Methods 93
Study Design: HPV plasmids were used to prepare test samples assessing the assays’ 94
performance parameters. Human genomic DNA was added as carrier diluents to a final 95
concentration of 1 ng/µL in each plasmid preparation to simulate the situation in actual 96
samples. For the evaluation of competition between HPV types with unequal copy numbers a 97
titration series containing a constant 50 IU of HPV18 DNA and different numbers of HPV16 IU in 98
10-fold increments from 5 x 100 to 5 x 106 was prepared. To evaluate detection of HPV at the 99
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
6
desired level of sensitivity – 50 IU or GE per assay, the final concentration of each type was 100
adjusted so that 50 IU (or GE) would be assayed. Samples containing multiple types [from two 101
to six of the 15 types in different combinations] were prepared in order to evaluate impact of 102
multiple types on assay performance (see Table 1). Two additional control samples containing 103
no template DNA and one with only Human genomic DNA (Roche Diagnostics) were also 104
prepared. Since sample source and collection media can influence the performance of 105
molecular assays, particularly those based on PCR technology, we included extracts of cervical 106
cells in specimen transport media (STM) and archived formalin fixed paraffin embedded (FFPE) 107
tissue in the test panel. These samples are described below. In particular, FFPE extracts are 108
known to be challenging for assays with large amplicons, and the LBAs could perform 109
differently 110
The final testing panel included 66 plasmid challenge samples (7 titration, 15 individual, 18 111
multiple, , each in duplicate), 60 epidemiologic samples and 2 negative controls. Six replicate 112
aliquots of the testing panel were prepared to provide duplicate testing on each of the three 113
LBA platforms. All DNA samples were randomly coded by a member of the laboratory not 114
involved in testing disguising the origin of the sample, expected HPV status and types. Two 115
technologists independently tested a panel on each of the platforms and interpreted the 116
results. Prior to initiating testing, each technologist successfully completed proficiency testing 117
on each assay platform; correctly identifying one to five HPV types in 10 unknown samples with 118
at least 80% accuracy. 119
Each assay used a 10 µL DNA aliquot per PCR reaction. Results for all possible HPV types were 120
entered into MS Access database tables. A sample was termed HPV positive when at least one 121
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
7
type was detected, HPV negative if none of the 15 types but the genomic control was positive 122
and inadequate if neither HPV nor control were detected. Testing was completed over a period 123
of two weeks. The order in which each technologist used each platform varied. 124
125
Plasmid samples: Full length clones of HPV types 6, 11, 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 126
59, 68 in plasmid vectors (100 ng in TE buffer) were provided by the WHO HPV Global 127
Reference Laboratory (Region Skåne, Malmö, Sweden). Each plasmid was diluted to 5 x 104 128
genome equivalents per mL in 200 μg/mL yeast tRNA solution (Invitrogen, Carlsbad, CA). 129
Number of genomic copies for HPV16 and 18 had been previously standardized by WHO and 130
was measured in international units (IU) accordingly. Copy numbers of other types were 131
calculated from the molecular mass and concentration as Genome Equivalents (GE). Human 132
genomic DNA used as background for HPV plasmid samples was obtained from Roche 133
Diagnostics, (Indianapolis, IN and diluted to 10 ng/μl in 0.1 mM TE. 134
135
Epidemiological Samples: Residual DNA extracts were retrieved from anonymous archived 136
epidemiologic samples from studies of HPV, either populations with high HPV prevalence or 137
samples of HPV-associated cancers. These were randomly selected without knowledge of prior 138
HPV results to include each 30 DNA extracts from cervical cells in Digene STM (Qiagen, Valencia, 139
CA, USA) and 30 from archived FFPE cancer tissues. 140
141
HPV Genotyping: The selection of kits to be evaluated was based on the following criteria: (a) 142
ability to detect and individually identify all or the majority of the thirteen high risk HPV types 143
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
8
16, 18, 31, 33, 35, 39, 51, 52, 56, 58, 59, 68b and the two low risk types 6 and 11, (b) no 144
requirment for expensive instrumentation to perform and interpret assay, and (c) commercial 145
availability in the US. Meeting most of these criteria, the Linear Array (LA) HPV Genotyping 146
Assay (Roche Diagnostics, Indianapolis, IN), INNO-LiPA HPV Genotyping Extra (Innogenetics, 147
Gent, Belgium) and the RH Probe Assay (Qiagen, Valencia, CA) were chosen. All three tests are 148
based on general HPV amplification via a low stringency PCR amplification of the L1 region and 149
subsequent detection via reverse line blot assay with type specific probes. Technical details are 150
compared in Table 2. 151
The LA uses PGMY 09/11 consensus primers and identifies a total of 37 individual types. The 152
test was performed according to the manufacturer’s specification, except that 40 μl DNase free 153
water was added with the 10 μl template DNA to fill to the required volume. Hybridization and 154
washing steps of the reverse line blot assay were done automatically with Beeblot instruments 155
(Bee Robotics, Caernarton Gwynedd, UK). HPV52 is only detected by an “XR” probe which 156
cross-hybridizes with HPV33, 35 and 58. In the presence of any of these three types, HPV52 157
cannot be identified unequivocally. 158
LiPA applies SPF10 primers and includes probes to detect 29 HPV types. The assay was 159
performed in accordance with the manufacturer’s protocol using an AutoBlot 3000H (MedTec, 160
Buffalo, IL) for hybridization to the genotyping strips. Some types are defined by a single 161
positive probe on the genotyping strip (i.e. HPV6, 11, 16), but others are interpreted as a 162
combination of two to four probes (i.e. HPV18, 33, 58). While the manufacturer provides 163
interpretation of type detection, including “possible” types, only those that were detected 164
unequivocally were included in the analysis. 165
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
9
The RH assay was not officially available in the US at the time this study was conducted and the 166
kits were kindly provided by Qiagen. The test utilizes the GP5+/6+ primers and line probes to 167
distinguish 18 high risk types; it does not detect HPV6 and 11. PCR amplification and line blot 168
detection were carried out as specified by Qiagen. Hybridization and washing of HPV 169
genotyping strips was achieved manually using a Gemini Twin shaking water bath (SciGene, 170
Sunnyvale, CA) and an aspiration system with 8-needle Stream Splitter (Art Robbins Instrument, 171
Sunnyvale, CA). 172
A water blank and SiHa DNA were included as negative and positive controls in every run with 173
each assay to monitor validity. The results from these controls were not included in the 174
analysis. 175
176
Analysis: The 15 HPV types 6, 11, 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 68 were 177
considered for the analysis. Some calculations (positive samples, accuracy, reproducibility) for 178
RH results were restricted to the 13 high-risk types as indicated. The two technologists’ results 179
for each sample and platform were counted as independent events. For plasmid samples, 180
accuracy was calculated as the percentage of correctly detected types (true positive and true 181
negative) from the total number of types assessed in all samples. [Total types assessed = 15 182
types per assay x 66 samples x 2 assays = 990; 858 if restricted to 13 hr types] 183
Results from STM and FFPE extracts were compared descriptively. Positive reproducibility (PR) 184
was calculated as the percentage of HPV types identified in both replicates among the total 185
number of types detected in either of the duplicate samples. Unweighted Cohen’s Kappa 186
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
10
coefficients were calculated with SPSS 15.0 for Windows for all types assessed (positive and 187
negative) in the relevant sample set. 188
189
Results 190
Titration of HPV16 against 50 IU HPV18: Detection of HPV16, HPV18 and genomic control are 191
listed in Table 3 at all input concentrations of HPV16. When HPV 16 is present below 5 X 104 IU, 192
all three targets are detected by all assays. While LA results were robust to increasing IU of HPV 193
16, LiPA and RH failed to detect 50 IU HPV18 and at the highest input of HPV 16 IU failed to 194
detect the genomic control. The negative control containing only human gDNA gave expected 195
results with all assays as did the no template control (data not shown). 196
197
Detection of HPV plasmid standards: Table 4 lists false positive, false negative and inadequate 198
results by LBA among plasmid samples. Both technologist’s results are reported so number of 199
results are double the number of samples. In the single HPV plasmid challenges, accuracy of LA 200
was 29/30 (96.7%), LiPA was 27/30 (90.0%), and RH was 23/30 (76.7%). [RH detected 23/26 201
(88.5%), if HPV6 and 11 are excluded.] The challenges with multiple HPV plasmids included a 202
total of 61 HPVs. Of these, LA assaysidentified all 122 types correctly (100%), LiPA 100/122 203
(82.0%) and RH 83/122 (68.0%). If HPV6 and 11 were excluded, 83/108 (76.9%) types were 204
correctly found by RH. 205
Accuracy for all 15 types in the 66 plasmid samples was 99.9, 97.3 and 95.4% for LA, LiPA and 206
RH respectively. Restricted to the 13 high-risk types, RH’s accuracy was 96.7%. 207
208
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
11
Epidemiologic samples: Results for STM and FFPE samples are combined as differences 209
between the two sample types were negligible for all LBAs. From the 120 test results derived 210
from each assay, the genomic control probe was positive in 116, 90 and 87 tests by LA, LiPA and 211
RH respectively. However results were adequate in 118, 118 and 111 by LA, LiPA and RH 212
respectively because some samples were positive for HPV but negative for the genomic control. 213
One or more HPV types were detected in 76, 90, and 69 samples by LA, LiPA and RH 214
respectively. Complete type-specific agreement amoung all three LBAs was noted for 81 215
samples (67.5%). Of these 42 were HPV positive and 39 HPV negative. An additional 24 216
samples (20%) had at least one type detected concurrently by all assays and one sample was 217
HPV positive in all assays but without type-agreement. 218
The overall detection of individual types was very similar for the three LBAs with the exception 219
of HPV52, which was only detected by LiPA in 9 cases (Figure 1). LA detected a total of 111 220
types (108 without 6 and 11) in 76 samples, LiPA detected 121 types (111 without 6 and 11) in 221
90 samples and RH found 97 types in 69 samples. Instances of type concordances and 222
discordances between the assays are illustrated in Figure 2. 223
224
Reproducibility: All test samples with the exception of the seven titration samples were 225
included in the reproducibility assessment. In the 33 test samples prepared with single or 226
multiple plasmids, expected results were for a total of 76 instances of HPV type detection (67 227
without HPV6 and 11). Differences in reproducibility between STM and FFPE samples, were not 228
significant and they were combined in Table 5. 229
230
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
12
Discussion 231
Results from the plasmid standards revealed differences between the HPV typing assays. The 232
titration of HPV16 in different amounts against constant HPV18 template numbers highlighted 233
the robustness of the assays to competing types with different concentrations. Copy numbers 234
of more than 5 x 103 or a 100-fold higher template amount of HPV16 suppressed amplification 235
of HPV 18 in LiPA and RH PCRs. These observations are in line with assessments by van Doorn at 236
al. (9), who found that 100 copies of HPV18 were outcompeted by 1000-fold higher 237
concentration of HPV16. Only the LA was able to detect both types even at 100,000 times 238
higher HPV16 concentrations. The larger reaction volume (100 μl in LA vs 50 μl in the others) 239
might be responsible for the superior tolerance to competition. The extreme differences in 240
copy number are unlikely to occur in actual biologic samples, however pushing the technical 241
limits of the assays allowed for evaluation of robustness to a variety of challenges. All three 242
LBAs generally could detect 50 IU/GE of the single high risk HPV types. Only HPV68b was not 243
detected by either RH duplicate, which is not surprising since the LOD is stated as 105 viral 244
copies in the RH Detection Kit Handbook. The sporadic lack of reproducibility may indicate that 245
this input amount is in the range of the lowest limit of detection (LOD) for at least some types 246
(Table 3). 247
Significant deficiencies were seen in samples that included more than one type. LiPA failed to 248
detect 18% of the types included in the 36 multi-plasmid challenges. The 22 missed types 249
consisted exclusively of HPV39, 58, 59 and 69. In some instances types 39 and 68b were 250
identified as “possible” types due to the LiPA’s multi probe set (see Table 4). Nevertheless, both 251
HPV39 and 68b were also missed in other samples that were free of this ambiguity suggesting 252
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
13
unequal amplification efficiency or competition. Disregarding HPV 6 and 11 the RH assay failed 253
to detect 20.5% of the types in this subset. Besides HPV68b, types 52, 59 and 39 were missed in 254
several instances.. A critical limitation of the RH assay might result from the large discrepancy in 255
detection sensitivities for different HPV types. According to the handbook LODs differ 25,000-256
fold ranging from four copies for HPV16 to 100,000 copies for HPV53 and 68. 257
Reproducibility of results for the plasmid challenges was greatest among the LA test, but 258
generally good in all assays. Among the results derived from the patient samples RH had the 259
highest positive reproducibility (90.2%), but the denominator was also lowered to 51 types 260
since HPV6 and 11 are not detected for this test. It was rather surprising that no significant 261
differences were found between STM and FFPE extracts as fractured and poor quality DNA from 262
archived tissues should favor shorter amplicon lengths as targeted by the SPF primers (8). The 263
sample size may not have been sufficient to allow differences in assay performance to be 264
identified. Generally, results from the patient samples were comparable between the assays 265
with at least partial agreement in 87.5%. Performance differences found with the plasmid 266
samples might reflect rather extreme situations which are not relevant for the majority of real 267
clinical specimens. 268
False positive results were a particular concern. Two instances were observed among LiPA 269
results in plasmid samples. In both cases the algorithm for identifying the expected type 270
(HPV18, 58) requires that more than probe hybridizes with the amplicon. As only one of the 271
required probes hybridized, another type (HPV39, 52) was falsely indicated. It is likely that 272
individual LiPA probes differ in their affinity to the same amplicon and generate an incomplete 273
band pattern at low target amounts which would consequently lead to attributing the result to 274
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
14
an incorrect type. Detection of HPV52 by LiPA might be particularly vulnerable to this problem. 275
In the LiPA assay the HPV 52 amplicon hybridizes to a single probe. However that probe also 276
hybridizes to amplicons of five additional HPV types that are distinguished by combinations 277
with other probes. In this regard it is noteworthy that HPV52 was detected in 9 of the 15 types 278
exclusively found by LiPA among the patient samples (Figure 1) and two of three patient 279
samples exclusively HPV positive by LiPA had only type 52. It seems likely that at least some of 280
these cases are false positive. 281
Conversely, some HPV52 may have been missed by the LA as it has a similar shortcoming and 282
detects HPV52 only through the cross-binding XR probe. While this ambiguity occurred in 28 283
cases among the LA results only two of them tested in fact positive for HPV52 by LiPA. 284
Every laboratory test is also influenced by the accuracy of human handling. The samples 285
prepared from plasmid DNA yielded one inadequate result each, by LA and LiPA and the (single) 286
HPV types not detected in these results were also counted as “missed”. Although performed 287
with utmost care in a clinically certified laboratory, operator errors cannot be ruled out as a 288
cause and may have lowered the real tests’ performances. However this possible distortion 289
should be minimal. 290
Analysis for this comparison study was restricted to the most relevant high risk types as well 291
asHPV6 and 11. It should be considered however, that Linear array detected 50, LiPA 16 and RH 292
4 additional HPV types in the 60 patient samples. Depending on the number of types covered 293
by each assay, the scope of HPV detection will always be limited and does not directly allow an 294
“HPV negative” interpretation. 295
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
15
For the majority of samples, HPV typing results will be very similar and comparable for the 296
three LBAs evaluated. In difficult situations such as multiple infections, low copy numbers or 297
large difference in viral copies, LA has an advantage. Some limitations of LiPA are due to its’ 298
multiple probe detection system. RH is disadvantaged by low sensitivity for some types, 299
particularly HPV53, 68 and 82. LA performed nearly perfect and is only hampered by the cross-300
reacting XR probe for HPV52 detection. 301
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
16
References 302
303
1. de Roda Husmann AM, Walboomvers JM, van den Brule AJ, Maijer CJ, and Snijders PJ. 304
1995. The use of general primers GP5 and GP6 elongated at their 3' ends with adjacent 305
highly conservative sequences improves human papillomavirus detection by PCR. J Gen 306
Virol 76:1057-1062. 307
2. Eklund C, Zhou T, Dillner J, and WHO Human Papillomavirus Laboratory Network. 2010. 308
Global proficiency study of human papillomavirus genotyping. J Clin Microbiol 48:4147-309
4155. 310
3. Gravitt PE, Peyton CL, Alessi TQ, Wheeler CM, Coutlee F, Hildesheim A, Schiffman MH, 311
Scott DR, and Apple RJ. 2000. Improved Amplification of Genital Human Papillomaviruses. 312
J Clin Microbiol 38:357-361. 313
4. Kleter B, van Doorn LJ, ter Schegget J, Schrauwen L, van Krimpen K, Burger M, ter 314
Harmsel B, and Quint W. 1998. Novel short-fragment PCR assay for highly sensitive 315
broad-spectrum detection of anogenital Human papillomaviruses. Am J Pathol 153:1731-316
1739. 317
5. Klug SJ, Molijn A, Shopp B, Holz B, Iftner A, Quint W, Snijders PJ, Petry KU, Kruger Kajaer 318
S, Munk C, and Iftner T. 2008. Comparison of the performance of different HPV 319
genotyping methods for detecting genital HPV types. J Med Viol 80:1264-1274. 320
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
17
6. Molijn A, Kleter B, Quint W, and van Doorn LJ. 2005. Molecular diagnosis of human 321
papillomavirus (HPV) infections. J Clin Virol 32S:S43-S51. 322
7. Seme K, Lepej SZ, Lunar MM, Iscic-Bes J, Planinic A, Kojan BJ, Vince A, and Polja M. 2009. 323
Digene HPV Genotyping RH Test RUO: comparative evaluation with INNO-LiPA HPV 324
Genotyping Extra Test for detection of 18 high-risk and probable high-risk human 325
papillomavirus genotypes. J Clin Virol 46:176-179. 326
8. Tan SE, Garland SM, Rumbold AR, and Tabrizi SN. 2010. Human papillomavirus 327
genotyping using archival vulva dysplastic or neoplastic biopsy tissues: comparison 328
between the INNO-LiPA and linear array assays. J Clin Microbiol 48:1458-1460. 329
9. van Doorn LJ, Molijn A, Kleter B, Quint W, and Colau B. 2006. Highly effective detection 330
of human papillomavirus 16 and 18 DNA by a testing algorithm combining broad-331
spectrum type-specific PCR. J Clin Microbiol 44:3292-3298. 332
10. van Hamont D, van Ham MA, Bakkers JM, Massunger LF, and Melchers WJ. 2006. 333
Evaluation of the SPF10-INNO LiPA human papillomavirus (HPV) genotyping test and the 334
roche linear array HPV genotyping test. J Clin Microbiol 44:3122-3129. 335
336 337
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
18
Tables and Figures: 338 339 Table 1. HPV types in the plasmid test samples. 340
Plasmid Titration HPV type (Copy number)
Plasmid Individual*
Plasmid Mix*
HPV 18 (5E01), HPV 16 (5E00) HPV 6 HPV16, HPV68b
HPV 18 (5E01), HPV 16 (5E01) HPV 11 HPV18, HPV59
HPV 18 (5E01), HPV 16 (5E02) HPV 16 HPV6, HPV56
HPV 18 (5E01), HPV 16 (5E03) HPV 18 HPV35, HPV11
HPV 18 (5E01), HPV 16 (5E04) HPV 31 HPV39, HPV52
HPV 18 (5E01), HPV 16 (5E05) HPV 33 HPV31, HPV51
HPV 18 (5E01), HPV 16 (5E06) HPV 35 HPV33, HPV58
No HPV HPV 39 HPV45, HPV16, HPV52
Water blank HPV 45 HPV18, HPV39, HPV59
HPV 51 HPV6, HPV35, HPV51
HPV 52 HPV33, HPV18, HPV56
HPV 56 HPV58, HPV16, HPV11, HPV45
HPV 58 HPV6, HPV59, HPV52, HPV39
HPV 59 HPV31, HPV33, HPV35, HPV58, HPV56
HPV 68b HPV18, HPV31, HPV11, HPV45, HPV68b
No HPV HPV52, HPV39, HPV45, HPV68b, HPV51
HPV6, HPV31, HPV35, HPV16, HPV56, HPV59
HPV18, HPV33, HPV39, HPV51, HPV58, HPV68b
*All plasmids in the individual and mixed samples were prepared to represent the equivalent of 341
50 IU or GE per test sample; sensitivity suggested by the WHO HPV LabNet. 342
343
344
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
19
Table 2. Technical specification of commercial LBAs. 345
LA LiPA RH
Primer set PGMY 09/11 SPF10 GP5+/6+
HPV amplicon size 450 bp 65 bp 150 bp
PCR volume 100 μL 50 μL 50 μL
Hybridization Temp. 53°C 49°C 50°C
Detection signal Horseradish Peroxidase
AlkalinePhosphatase
Alkaline Phosphatase
Genomic control
Control amplicon size
β-Globin
268 bp
HLA-DPB1
280 bp
β-Globin
258 bp
LA = Linear Array HPV Genotyping Test 346
LiPA = Line Probe Assay 347
RH = Reverse Hybridization 348
349
350 on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
20
Table 3. Impact of Increasing HPV 16 IU on Detection of 50 IU HPV 18* 351
Input HPV 18 = 50 IU
H.s. DNA = 10 ng LA LiPA RH
Test 16 IU HPV16 HPV18 β-Glob HPV16 HPV18 hDNA HPV16 HPV18 gDNA
1st 5E+00 + + + + + + + + +
2nd 5E+00 + + + + + + + + +
1st 5E+01 + + + + + + + + +
2nd 5E+01 + + + + + + + + +
1st 5E+02 + + + + + + + + +
2nd 5E+02 + + + + + + + + +
1st 5E+03 + + + + + + + + +
2nd 5E+03 + + + + + + + + +
1st 5E+04 + + + + fp + + + +
2nd 5E+04 + + + + + + +
1st 5E+05 + + + + + +
2nd 5E+05 + + + + + + +
1st 5E+06 + + + + +
2nd 5E+06 + + + + +
1st 0 + + +
2nd 0 + + +
* Detected = +, Not detected = blank; fp= false positive result – HPV 39 detected instead of HPV 18 352
353
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
21
Table 4. Incorrect Line Blot Assay results in HPV plasmid challenges¥. 354
No. of types (No. of samples)
LA LiPA RH
1 (n = 30) 59 (ia) 39 (ia), 58 (fp 52), 59 6‡, 6‡, 11‡, 11‡, 52, 68, 68 2 (n = 14) 58, 59, 59 6‡, 6‡, 11‡, 11‡, 56, 52
3 (n = 8) 39*, 39*, 68, 59 52, 52, 68, 68, 59, 6‡, 6‡,
39 4 (n = 4) 58 (fp 52), 59, 59 11‡, 11‡, 6‡, 6‡, 39, 52, 59
5 (n = 6) 39, 39, 68, 68, 68*, 68* 58, 58, 39, 39, 52, 52,
11‡, 11‡, 68, 68, 68, 68 6 (n = 4) 39*, 39*, 58, 68, 59, 59 68, 68, 6‡, 6‡, 59, 59
355
ia = inadequate, fp = false positive and falsely detected type 356
* detected as “possible types” 357
‡ probe not included in assay 358
¥Each entry reflects failue to detect a type by one of the technologists. 359
360
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
22
Table 5. Positive reproducibility of Line Blot Assays for plasmid standards and extracts from 361
epidemiologic samples. 362
363
LA LiPA RH *
Plas
mid
DN
A
(n =
33)
Types to be detected 76 76 67
Discrepant types 1 9 8
Samples affected 1 6 8
Positive Reproducibility 98.7% 88.2% 88.1%
Kappa 0.992 0.928 0.926
Epid
emio
logi
c (n
= 6
0)
Pos. types (by either test) 63 66 51
Discrepant types 8 11 5
Samples affected 6 9 5
Positive Reproducibility 87.3% 83.3% 90.2%
Kappa 0.925 0.899 0.942 364
* for RH analysis types 6 and 11 were omitted and the total number of types adjusted. 365
366
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
23
367
368
Figure 1. Type-specific detection by the three LBAs in the 120 epidemiologic samples. White 369
bars indicate results found by by LA, gray by LiPA and black by RH LBA. 370
371
0
5
10
15
20
25
30
35
40
45
Num
ber o
f tim
es d
etec
ted
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from
24
372
373
374
Figure 2: Venn diagram illustrating instances of type specific concordance and discordance between 375
assays (restricted to 13 high risk types- 18, 18, 31,33,35,39,45,51,52,56,58,59 and 68). Numbers in 376
parentheses indicate total number of HPV types detected by each LBA. 377
378
379
380
381
382
383
LA (108)
RH (97)
LiPA (111)
82
4
4 10
12 7
15
on February 15, 2018 by guest
http://jcm.asm
.org/D
ownloaded from