panoura 18s with ceph calibration and alignment quick guide

38
Panoura 18S Panoramic X-Ray with Ceph Calibration and Alignment

Upload: don-vibbert

Post on 20-Jan-2017

176 views

Category:

Health & Medicine


5 download

TRANSCRIPT

Page 1: Panoura 18S with Ceph Calibration and Alignment Quick Guide

Panoura 18SPanoramic X-Ray with Ceph Calibration and

Alignment

Page 2: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

2ImageWorks Proprietary – Confidential Information

These instructions serve as supplemental assembly guidance of the Panoura. They are not meant to

replace the installation manual, but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

Summary

Page 3: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

3ImageWorks Proprietary – Confidential Information

1. Computer set up2. Tube head aging3. Panoramic alignment4. Panoramic sensor calibration5. Ceph alignment6. Ceph sensor calibration7. Secondary collimater alignment8. Ear rod alignment9. Final Test Exposure

Agenda

Page 4: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

4ImageWorks Proprietary – Confidential Information

This procedure assumes PC was either provided or staged by ImageWorks

Start PC and open Task Manager (Ctl-Alt-Del and choose Task Manager)Verify that these processes are running:

1. P18_CM.exe2. P18_Launcher.exe3. P18S_Setup.exe4. Panoura18S.exe

1. Computer Set Up

Page 5: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

5ImageWorks Proprietary – Confidential Information

Confirm that key programs are running by looking for the following icons in the system tray (usually bottom right of screen):

1. Control Manager

2. Image Creator

1. Computer Set Up

Page 6: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

6ImageWorks Proprietary – Confidential Information

On the computer in the lower right hand corner of the screen, Double click the Control Manager icon (white tooth on light blue background).

Hold down the shift key and click on the setup button to open the CS service menus.

It will ask for password, which is p18p18

Select CS95 to perform Tube Head Aging Routing

2. Tube Head Aging

Page 7: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

7ImageWorks Proprietary – Confidential Information

Follow directions on screen. To initiate exposure, press hand switch and hold until beeping stops (x-ray cycle will be about 4 seconds).

Screen will show countdown during cool-down period and will say when it’s ready for next exposure(about 80 seconds).

Repeat this process 13 times. At the completion of the 13th cycle, the screen should show “Pass”

Click on “Main” tab to exit CS mode

2. Tube Head AgingNote:

If exposure switch is pressed before cooling cycle is complete, an

FL03 warning is displayed.

If switch is released before exposure is complete, an ER08

warning is displayed.

These error codes can be cleared by pressing

RESET on panel.

Page 8: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

8ImageWorks Proprietary – Confidential Information

Repeat this process 13 times. At the completion of the 13th cycle, the screen should show “Pass”

Click on “Main” tab to exit CS mode

2. Tube Head Aging

Page 9: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

9ImageWorks Proprietary – Confidential Information

3. Panoramic Mechanical Alignment

If inside cover on tube head is in place, remove this to reveal collimater assembly

Collimator

assembly

Page 10: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

10ImageWorks Proprietary – Confidential Information

For mechanical alignment and calibration, the FL07 exposure lock must be switched off. Do this by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.Make sure sensor is installed in Pan position

3. Panoramic Mechanical Alignment

Open Ajat Dental Imaging Software

Select Tools Maintenance Mechanical Alignment

The mechanical alignment tools screen is displayed

Page 11: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

11ImageWorks Proprietary – Confidential Information

On the computer in the lower right hand corner of the screen, Double click the Control Manager icon (white tooth on light blue background).

Hold down the shift key and click on the setup button to open the CS service menus.

It will ask for password, which is p18p18

Select CS75 to set parameters for the scan.

3. Panoramic Mechanical Alignment

Page 12: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

12ImageWorks Proprietary – Confidential Information

Adjust parameters (kv, ma and time)click “Set” …

…to match those listed in the mechanical alignment tool

3. Panoramic Mechanical Alignment

Page 13: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

13ImageWorks Proprietary – Confidential Information

Place 20 mm aluminum block over the pan collimator slit

(a magnetized block does notrequire any tape)

3. Panoramic Mechanical Alignment

pan slit

ceph slit

Page 14: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

14ImageWorks Proprietary – Confidential Information

Press handswitch to initiate x-ray. White column will appear on the screen with arrows above and below showing how far each end is out of alignment.

Number on top and bottom represent distance each side is out of alignment. If number is less than 0,20 mm, it will be in green, which is within the tolerance for alignment. The objective is to adjust collimater position such that both top and bottom are green (< 0,20 mm).

If either number is red (greater than 0,20), then the collimater must be adjusted.

Note: these distances do NOT correlate with how far collimater is out of alignment

3. Panoramic Mechanical Alignment

Page 15: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

15ImageWorks Proprietary – Confidential Information

The direction of the arrow represents the side to which the beam has moved too far when facing the sensor. Therefore, the beam must be moved in the opposite direction when facing the sensor.

In the example shown, the red arrows pointing to the left means that the beam is too far to the left (when facing the sensor), and must be moved to the right. Therefore, in this situation, the collimator slit must be moved to the left when facing the collimator.

Tip: The mechanical alignment process is most efficient if each adjustment of the collimator is done with deliberation and care. Small adjustments to collimator can have large impact on alignment, so it’s best to start with very small adjustments so that larger misalignments don’t result.

3. Panoramic Mechanical Alignment

Page 16: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

16ImageWorks Proprietary – Confidential Information

To adjust collimator left or right, loosen the two adjusting screws (these simply hold the assembly in place).Use pushing screw to move the assembly.

Note: The pushing screw only pushes, but does not pull. So turning the push screw clockwise by one turn, will move the collimater approximately 1 mm to the left (when facing the collimater). However, to move in the opposite direction, turning the screw counterclockwise one turn will create space, and the collimater needs to be physically “tapped” to move it.

Note: vertical or rotational adjustments may be required. Please refer to manual

3. Panoramic Mechanical Alignment

Adjusting Screws

Pushing Screw

Page 17: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

17ImageWorks Proprietary – Confidential Information

Tip: Loosening Adjusting screws too much will make the assembly too loose to make adjustments

Tip: For each scan, handswitch can be released when white area appears on alignment screen (as shown in picture) instead of doing entire exposure. This may help avoid delays caused by tubehead overheating, which require longer cooling periods between test exposures. This approach will create an Er08 code with beeps, but these can be cleared by pressing RESET.

3. Panoramic Mechanical Alignment

LevelNote: after each adjustment, confirm that surface of bracket is approximately level

Page 18: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

18ImageWorks Proprietary – Confidential Information

After adjusting the position of the collimator, turn the unit off and then on. This will cycle the motor above the collimator.

Turn off FL07 by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.

Repeat steps to open Control Manager • Double click the Control Manager icon• Hold down the shift key and click on the setup button to open the CS

service menus.• Enter password: p18p18• Select CS75 to set parameters for the scan.

Initiate scan.

If arrows are red, repeat collimator position adjustment steps

If arrows are green, tighten adjusting screws, and repeat steps for power cycling the unit, reopening Control Manager and scanning.Confirm that arrows remain green after tightening. If all is green, then click Close.

3. Panoramic Mechanical Alignment

Page 19: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

19ImageWorks Proprietary – Confidential Information

4. Panoramic Sensor CalibrationRemove all adapters such as chin rest block from irradiation path of the x-ray. Push temple supports up.

Open Ajat Dental Imaging Software

Select Tools Maintenance Calibration

Calibration screen is displayed. Click Calibrate

Click Yes

Note: unit needs to be

powered on for about 5 min to

perform calibration

Page 20: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

20ImageWorks Proprietary – Confidential Information

4. Panoramic Sensor CalibrationThe calibration includes a series of 3 x-ray exposures.

The display details the parameters needed for each exposure. Set the indicator panel to reflect these parameters.

Note: when 20 mm Al is displayed, then the aluminum block must be attached over the panoramic slit. When “No Filter” is displayed, then remove the block

Repeat cycles updating parameters requested for each exposures

When the calibration is finished, press Ok

Page 21: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

21ImageWorks Proprietary – Confidential Information

Make sure that secondary collimator and ear rods are removed

Move sensor to ceph position

Press MODE button on the panel

Switch the display to CEPH and LA.

Press RESET

Press BEAM

5. Ceph Mechanical Alignment

Page 22: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

22ImageWorks Proprietary – Confidential Information

For mechanical alignment, the FL07 exposure lock must be switched off. Do think by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.

Open Ajat Dental Imaging Software

Select Tools Maintenance Mechanical Alignment

The mechanical alignment tools screen is displayed

5. Ceph Mechanical Alignment

Page 23: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

23ImageWorks Proprietary – Confidential Information

Press handswitch to initiate x-ray. White column will appear on the screen with arrows above and below showing how far each end is out of alignment.

Tip: handswitch can be released when white area appears on alignment screen (as shown in picture) instead of doing entire exposure. This may help avoid delays caused by tubehead overheating, which require longer cooling periods between test exposures. This approach will create an Er08 code with beeps, but these can be cleared by pressing RESET.

Number on top and bottom represent distance each side is out of alignment. If number is less than 0,20 mm, it will be in green, which is within the tolerance for alignment. The objective is to adjust collimater position such that both top and bottom are green (< 0,20 mm).

If either number is red (greater than 0,20), then the collimater must be adjusted.

5. Ceph Mechanical Alignment

Page 24: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

24ImageWorks Proprietary – Confidential Information

The direction of the arrow represents the side to which the beam has moved too far when facing the sensor. Therefore, the beam must be moved in the opposite direction when facing the sensor.

In the example shown, the red arrows pointing to the left means that the beam is too far to the left (when facing the sensor), and must be moved to the right. Therefore, in this situation, the collimator slit must be moved to the left when facing the collimator.

Tip: The mechanical alignment process is most efficient if each adjustment of the collimator is done with deliberation and care. Small adjustments to collimator can have large impact on alignment, so it’s best to start with very small adjustments so that larger misalignments don’t result.

5. Ceph Mechanical Alignment

Page 25: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

25ImageWorks Proprietary – Confidential Information

Note: The push screw only pushes, but does not pull. So turning the push screw clockwise by one turn, will move the collimator approximately 1 mm to the right (when facing the collimater). However, to move in the opposite direction, turning the screw counterclockwise one turn will create space, and the collimater needs to be physically “tapped” to move it.

Note: vertical or rotational adjustments may be required. Please refer to manual

5. Ceph Mechanical AlignmentTo adjust collimater left or right, loosen the two Horizontal Direction Adjusting screws slightly (these simply hold the assembly in place).Note: loosening Horiz Dir Adj screws too much will make the assembly too loose to make adjustmentsLoosen the nut on the Push Screw, and then rotate the push screw to adjust the position.

Push Screws

and Nuts

Horizontal Direction Adjusting Screws

Page 26: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

26ImageWorks Proprietary – Confidential Information

After adjusting the position of the collimator, turn the unit off and then on. This will cycle the motor above the collimator.

Remove the FL07 exposure lock by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2-3 seconds.

Initiate scan.

If arrows are red, repeat collimator position adjustment steps

If arrows are green, tighten adjusting screws and locking nuts, and repeat steps for power cycling the unit, reopening Control Manager and scanning.Confirm that arrows remain green after tightening. If all is green, then click Close.

5. Ceph Mechanical Alignment

Page 27: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

27ImageWorks Proprietary – Confidential Information

6. Ceph Sensor CalibrationRemove all adapters such as chin rest block from irradiation path of the x-ray

Open Ajat Dental Imaging Software

Select Tools Maintenance Calibration

Calibration screen is displayed. Click Calibrate

Click Yes

Note: unit needs to be

powered on for about 5 min to

perform calibration

Page 28: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

28ImageWorks Proprietary – Confidential Information

6. Ceph Sensor CalibrationThe calibration includes a series of 4 exposures.

The display details the parameters of each exposure. Set the indicator panel to reflect these parameters.

Note: when 20 mm Al is displayed, then the aluminum block must be attached over the panoramic slit. When “No Filter” is displayed, then remove the block

When the calibration is finished, press Ok

Page 29: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

29ImageWorks Proprietary – Confidential Information

Attach secondary collimater

7. Secondary Collimator Alignment

Page 30: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

30ImageWorks Proprietary – Confidential Information

Make sure sensor is in ceph position

Press MODE button on the panel

Switch the display to CEPH and LA.

Press RESET

Press BEAM

7. Secondary Collimator Alignment

Page 31: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

31ImageWorks Proprietary – Confidential Information

Make sure FL07 exposure lock is off by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2-3 seconds.

Open Ajat Dental Imaging Software

Select Tools Maintenance Mechanical Alignment

The mechanical alignment tools screen is displayed

7. Secondary Collimator Alignment

Page 32: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

32ImageWorks Proprietary – Confidential Information

Watch the screen during the exposure to see white column in alignment screen.

Press handswitch to initiate x-ray.

During exposure, check to see that the green rectangle does not significantly come inside the center 4 squares of the grid on the mechanical alignment screen (denoted by dotted rectangle in picture)

7. Secondary Collimator Alignment

For exposure cycle, check that left and right sides of

green rectangle…

…remain outside of center 4

squares of grid

If sides of white column fall inside the center 4 squares, than secondary collimator must be adjusted

Page 33: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

33ImageWorks Proprietary – Confidential Information

Adjust secondary collimater by loosening the 4 screws.

When facing the secondary collimator, it needs to be moved in the direction that the green rectangle needs to be shifted to satisfy the alignment test above (i.e. to stay outside of the center 4 grid squares).

Initiate exposure and check alignment again.

Repeat until green rectangle is outside the center 4 squares.

7. Secondary Collimator Alignment

Page 34: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

34ImageWorks Proprietary – Confidential Information

Attach the ear rods and nose support

8. Ear rod Alignment

Tip: when installing ear rods, before tightening bolts, first allow gravity to get them to hang plum (vertical), and then tighten.

Page 35: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

35ImageWorks Proprietary – Confidential Information

Set positions of the ear rods and nose support to Ceph LA positionsOpen ear rods to maximum widthPress MODE on the control panelSwitch display to LAPress RESETTake exposure and inspect the alignment of the ear rod circles in the image

8. Ear rod Alignment

Note: the side with the larger ear ring is the side farthest from the sensor

Adjust ear rods until circles are closely aligned.

Tip: after first exposure, the handswitch for each following exposure can be released as soon as ear rings are revealed in the image. This can help avoid cool down wait times.

Page 36: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

36ImageWorks Proprietary – Confidential Information

Remove chinrest block

Mount installation phantom

9. Final Test Exposure

Page 37: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

37ImageWorks Proprietary – Confidential Information

Run exposure and evaluate image

Seven ball bearings should be checked to see that:1. Each ball appears circular2. Distance between balls is symmetrical to left and right of

center ball

9. Final Test Exposure

Note: Because the test scan does

not have patient anatomy, it is possible that lines or artifacts may be seen in other parts of the test image. This will not affect quality of actual dental

images taken.

Page 38: Panoura 18S with Ceph Calibration and Alignment Quick Guide

05/0

1/20

23

38ImageWorks Proprietary – Confidential Information

These instructions serve as supplemental assembly guidance of the Panoura. They are not meant to

replace the installation manual, but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

Summary