pacemaker basics frijo jose a. cathode: negative – electrode on the tip of a pacing lead anode:...
TRANSCRIPT
• Cathode: negative– Electrode on the tip of a
pacing lead• Anode: positive– The “ring” electrode on
a bipolar lead– The PG case on a
unipolar systemAnode
Cathode
Implantable Pacemaker Circuit
• Pulse generator (PG): – Battery– Circuitry– Connector(s)
• Leads– Cathode– Anode
• Body tissueIPG
Lead
Anode
Cathode
The Pulse Generator
• Battery - provide energy• Circuitry - controls
pacemaker operations• Connector- join the PG to
the lead(s)Circuitry
Battery
Connector Block
Lead Characterization
• Position within the heart– Endocardial/transvenous leads– Epicardial leads
• Fixation mechanism– Active/Screw-in– Passive/Tined
• Shape– Straight– J-shaped used in the atrium
• Polarity– Unipolar– Bipolar
• Insulator– Silicone– Polyurethane
Endocardial Passive Fixation Leads• The tines become lodged
in the trabeculae,a fibrous meshwork, of the heart
Tines
Transvenous Active Fixation Leads• The helix, or screw,
extends into the endocardial tissue– Allows for lead positioning
anywhere in the heart’s chamber
– The helix is extended using an included tool
Epicardial Leads
• Leads applied directly to the surface of the heart– Fixation mechanisms include:
• Epicardial stab-in• Myocardial screw-in• Suture-on
– Applied via sternotomy or laproscopy
Lead Polarity
• Unipolar leads – Smaller diameter lead body
than bipolar leads– Usually larger pacing artifacts
on ECG
• Bipolar leads – Usually less susceptible to
oversensing of non-cardiac signals
Unipolar lead
Bipolar coaxial lead
To tip (cathode)
Unipolar Pacing System• Lead has only one electrode –
cathode – at the tip• PG can - anode• When pacing, the impulse:
– Flows through the tip electrode (cathode)
– Stimulates the heart– Returns through body fluid and
tissue to PG can (anode)
Cathode -
Anode +
•more interference (myopotentials)
• Big spike on ECG
• Pectoral (pocket) stimulation possible
Anode
Bipolar Pacing System
• The lead has both an anode and cathode
• The pacing impulse: – Flows through the tip
electrode located at the end of the lead wire
– Stimulates the heart– Returns to the ring electrode,
the anode, above the lead tip
Cathode
Anode +
Cathode -•less interference
• Spike difficult to see on ECG
• No pectoral (pocket) stimulation
STIMULATION THRESHOLD
• The minimum stimulus intensity & duration necessary to reliably initiate a propagated depolarising wavefront from an electrode
• For a pacing stimulus to “capture”, the stimulus must exceed a critical amplitude & must be applied for a sufficient duration
• Stimulus amplitude & duration interact- minimal amplitude required to capture depends on the pulse duration
Strength–Duration Relation
Stimulus amplitude for endocardial stimulation has an exponential relation to duration of the pulse- rapidly rising strength–duration curve at pulse durations <0.25 ms and a relatively flat curve at pulse durations > 1.0 ms
• Chronaxie pulse duration ~ point of min threshold energy on strength–duration curve
• With pulse durations > chronaxie- relatively little ↓ in threshold V
• Wider pulse dur→wasting of energy without providing a substantial ↑ in safety margin
• When threshold determined by ↓ amplitude, stimulus V set twice the threshold value
• PG that determine threshold by automatically ↓pulse duration-at least 3times threshold
• Wedensky effect
Hyperacute phase of threshold evolution- active fixation electrodes may produce,immediately following implantation,an ↑ stimulation threshold that ↓over the next 20-30 min- transient high threshold – a/c injury at myocardial–electrode interface
Proper programming of the stimulus amplitude
• The strength–duration relation must be appreciated• The safety margin that is chosen for a particular pt
must be based on the degree of pacemaker dependency
• An appreciation of the effect of stimulus amplitude & duration on battery longevity
• The overall metabolic & pharmacologic history of the pt must be considered
• Pacing threshold varies inversely with surface area of the stimulating electrode
Strength–Interval Relation
• The stimulation threshold is influenced by the coupling interval of electrical stimuli and frequency of stimulation
• ICD- greater stimulus intensity during ATP than during antibradycardia pacing
(Ztotal) = Zc Ze ZpZp - related to movmnt
of charged ionmyo toward the s in cathode
Zp -directly related to pulse duration & can be ↓ by use of relatively ↓ pulse durations
Polarization is inversely related to SA of electrode- to ↓ Zp but ↑ Ze, SA of electrode can be made large but radius small by use of a porous coating
Afterpotential of opposite chargeis induced in the myocardium at the interface of
the stimulating electrode
The slope of intrinsic deflectn (dV/dt) expressed in V/s -referred to as slew rate
For an EKG to be sensed, the amplitude & slew rate must exceed the sensing threshold
• For a battery,the decay characteristics should be predictable. The ideal battery should have a predictable fall in V near the end of life, yet provide sufficient service life after initial voltage decay to allow time for the elective replacement indicator to be detected and for replacement to be performed
• P- Native atrial depolarization• A- Atrial paced event• R- Native ventricular depolarization• V- Ventricular paced event• AV- Sequential pacing in the atrium and ventricle• AVI- Programmed AV pacing interval• AR- Atrial paced event foll by intrinsic ventricular• ARP- Atrial refractory period• PV- Native atrial foll by paced ventricular, P-
synchronous• AEI- Interval from a ventricular sensed/paced to atrial
paced event, the VA interval
NASPE/ BPEG Generic (NBG) Pacemaker Code
I. Chamber II. Chamber III. Response to IV. Programmability V. Antitachy Paced Sensed Sensing Rate Modulation arrhythmia funct.
O= none O= none O= none O= none O= noneA=atrium A= atrium T= triggered P= simple P= pacingV= ventricle V= ventricle I= inhibited M= multi S= shockD= dual D= dual D= dual C= communication D= dual(A+V) (A+V) (T+I) R= Rate Modulation
Manufacturers’ Designation only:
S= single S= single(A or V) (A or V)
AOO & VOO
• By application of magnet• Useful in diagnosing pacemaker dysfunction• During surgery to prevent interference from
electrocautery
Ventricular asynchronous (VOO) pacing
• Neither sensing nor mode of response• Irrespective of any other events, V pacing
artifacts occur at programmed rate. Timing cycle cannot be reset by any intrinsic event
Atrial asynchronous (AOO)AV sequential asynchronous (DOO)
• AVI & VAI or AEI are both fixed- intervals never change, as is insensitive to any atrial or ventricular activity, and timers never reset
Ventricular demand inhibited (VVI)
• Sensing on V channel- output inhibited by sensed V event
• Refractory after V/R- VRP- any V event in VRP-not sensed & doesn’t reset V timer
• LRL
VVI MODE
Automatic Interval
• Automatic interval starts from a paced complex (to the next paced complex)
• Escape interval starts from a sensed complex (to the next paced complex)
Escape Interval
If the intervals are equal:
• No hysteresis
If the escape interval > automatic interval:
• Hysteresis
VVI MODE (with hysteresis)
1000 ms
850 ms
Escape interval = 1000 ms (60 ppm)
Automatic interval = 850 ms (70 ppm)
AAI• Useful for SSS with N- AV conduction• Should be capable of 1:1 AV to rates 120-140 b/m • Atrial tachyarrhythmias should not be present• Atria should not be “silent”• If no A activity, atria paced at LOWER RATE limit (LR)• If A activity occurs before LR,- “resetting”• An A activity too early may not cause depolarisation & better
not sensed- ARP • Caution- far-field sensing of V activity
Single-Chamber Triggered-Mode
• Output pulse every time a native event sensed• ↑current drain• Deforms native signal• Prevent inappropriate inhibition from
oversensing when pt does not have a stable native escape rhythm
• Can be used for noninvasive EPS,with already implanted PPI tracking chest wall stimuli created by a programmable stimulator
AV sequential-V Inhibited Pacing (DVI)
• PPI is inhibited & reset by sensed V activity but ignores all intrinsic A complexes
• Native R during AVI sensed – V output inhibited & AEI reset
• For both A&V stimuli to be inhibited, sensed R must occur during AEI
AV Sequential, Non–P-Synchronous Pacing with D-Chamber Sensing (DDI)
• AAI + VVI• Difference btw DVI & DDI- DDI incorporates A
sensing as well as V sensing- prevents competitive A pacing
• Mode of response is inhibition only- no tracking of P waves - paced V rate cannot be > programmed LRL
• Goes by LR for each• Timing cycles- LRL, AVI, PVARP & VRP
A Synchronous (P-Tracking) Pacing (VDD)
• VAT + VVI • Timing cycle- LRL,AVI,PVARP,VRP,& URL• Concept of AV interval (AVI)• A sensed atrial event initiates AVI • Goes by LR• If no A event occurs, PPI escapes with a V pace
at LRL- VVI • AV block with intact sinus node function (esp
useful in congenital AV block)
DDD• VAT + AAI + VVI• AV interval & VA interval• VA interval replaces the LR• Concept of PVARP (to prevent endless loop or PMT)
• Indications• 1. The combination of AV block and SSS• 2. Patients with LV dysfunction & LV hypertrophy who
need coordination of atrial & ventricular contractions to maintain adequate CO
DDD ExamplesThe Four Faces of DDD
• Atrial and ventricular pacing
– Atrial pace re-starts the lower rate timer and triggers an AV delay timer (PAV)• The PAV expires without being inhibited by a ventricular sense, resulting in
a ventricular pace
AP
AP
VP
VP
DDD ExamplesThe Four Faces of DDD
• Atrial pacing and ventricular sensing
– Atrial pace restarts the lower rate timer and triggers an AV delay timer (PAV)• Before the PAV can expire, it is inhibited by an intrinsic ventricular event
(R-wave)
AP
AP
VS
VS
DDD ExamplesThe Four Faces of DDD
• Atrial sensing, ventricular pacing
– The intrinsic atrial event (P-wave) inhibits the lower rate timer and triggers an AV delay timer (SAV)• The SAV expires without being inhibited by an intrinsic ventricular event,
resulting in a ventricular pace
AS
AS
VP
VP
DDD ExamplesThe Four Faces of DDD
• Atrial and ventricular sensing
– The intrinsic atrial event (P-wave) inhibits the lower rate timer and triggers an AV delay timer (SAV)• Before the SAV can expire, it is inhibited by an intrinsic ventricular event (R-
wave)
AS
AS
VS
VS
Dual Response to SensingDDD
• The pacemaker can:– Inhibit and trigger– A P-wave inhibits atrial pacing and triggers an SAV
interval– An atrial pace triggers a PAV interval– An R-wave inhibits ventricular pacing
Lower Rate Interval
• Defines the lowest rate the pacemaker will pace
Lower Rate Interval
VP VP VVI / 60
Refractory Period
• Interval initiated by a paced or sensed event
• Designed to prevent inhibition by cardiac or non-cardiac events
Lower Rate Interval
VP VP VVI / 60Refractory Period
Blanking Period
• The first portion of the refractory period• Pacemaker is “blind” to any activity• Designed to prevent oversensing pacing stimulus
Lower Rate Interval
VP VP VVI / 60Blanking PeriodRefractory Period
Upper Sensor Rate Interval
• Defines the shortest interval (highest rate) the pacemaker can pace as dictated by the sensor (AAIR, VVIR modes)
Lower Rate Interval
VP VP VVIR / 60 / 120Blanking PeriodRefractory Period
Upper Sensor Rate Interval
VP VP VS VP
Lower Rate Interval-60 ppm
Hysteresis
• Allows the rate to fall below the programmed lower rate following an intrinsic beat
Hysteresis Rate-50 ppm
Dual Chamber Timing Parameters
• Lower rate• AV and VA intervals• Upper rate intervals• Refractory periods• Blanking periods
Lower Rate Interval
APVP
APVP
Lower Rate
• The lowest rate the pacemaker will pace the atrium in the absence of intrinsic atrial events
DDD 60 / 120
APVP
ASVP
PAV SAV
200 ms 170 ms
Lower Rate Interval
AV Intervals
• Initiated by a paced or non-refractory sensed atrial event– Separately programmable AV intervals – SAV /PAV
DDD 60 / 120
Lower Rate Interval
APVP
APVP
AV Interval VA Interval
Atrial Escape Interval (V-A Interval)Atrial Escape Interval (V-A Interval)
• The interval initiated by a paced or sensed ventricular event to the next atrial event
DDD 60 / 120PAV 200 ms; V-A 800 ms
200 ms 800 ms
DDDR 60 / 120A-A = 500 ms
APVP
APVP
Upper Activity Rate Limit
Lower Rate Limit
V-APAV V-APAV
Upper Activity (Sensor) Rate
• In rate responsive modes, the Upper Activity Rate provides the limit for sensor-indicated pacing
ASVP
ASVP
DDDR 60 / 100 (upper tracking rate) Sinus rate: 100 bpm
Lower Rate Interval {
Upper Tracking Rate Limit
Upper Tracking Rate
• The maximum rate the ventricle can be paced in response to sensed atrial events
SAV SAVVA VA
Post Ventricular Atrial Refractory Period (PVARP)
Refractory Periods• VRP and PVARP are initiated by sensed or paced
ventricular events– The VRP is intended to prevent self-inhibition such as
sensing of T-waves – The PVARP is intended primarily to prevent sensing of
retrograde P waves
AP
VPVentricular Refractory Period (VRP)
A-V Interval(Atrial Refractory)
Blanking Periods
• First portion of the refractory period-sensing is disabled
AP
VP
AP
Post Ventricular Atrial Blanking (PVAB)
Post Atrial Ventricular Blanking
Ventricular Blanking (Nonprogrammable)
Atrial Blanking (Nonprogrammable)
Dual Chamber Timing
• Atrial Pace (AP) - Ventricular Pace (VP) example
DDD 60
A-A interval A-A interval
VRP
ARP PVARP
PVAB
VRP
ARP PVARP
PVAB
PAVV-A interval V-A interval
PAV