OTS80PB & OTS60PBOil Test Set
USER GUIDE
2
G SAFETY WARNINGS
n SafetywarningsandprecautionsmustbereadandunderstoodbeforetheOTS80PB/OTS60PBisused.Theymust beobservedduringuse.
n TheOTS80PB/OTS60PBmustonlybeusedinthewayintendedandforthestatedpurposesdescribedbyMegger.
n TheOTS80PB/OTS60PBmust not beusedifanypartofitisdamaged.
n TheOTS80PB/OTS60PBmustbeproperlyearthed.Asupplementaryearthterminalisprovided.
n PositiontheOTS80PB/OTS60PBsothatpowercanbeeasilyremoved.
n TheOTS80PB/OTS60PBmust notbeusedinwetlocationsorwithrainfallingontheinstrument.
n TheOTS80PB/OTS60PBmustonlybeusedfortestingelectricalgradesofinsulatingliquids.Nootherobjectsor substancesshouldbeplacedinthetestchamber.
n TheOTS80PB/OTS60PBmustonlybeusedwithoneofMegger’sprecisiontestvesselsorVCM100Dcorrectlyfitted onthesupporthornsinthetestchamber.
n Replacementfusesmustbeofthecorrecttypeandrating.
n Ifbatteriesarefitted,the12VDCchargermustonlybepoweredfromanegativeearthedvehicle12VDCsource usingtheMeggersuppliedlead.
n CalibrationorrepairmustonlybecarriedoutbyaMeggerqualifiedrepairfacility.
n Thetestchambermustbekeptcleanandfreefromlintorotherfibres.
n TheOTS80PB/OTS60PBweighsbetween16kgand23kg.Careshouldbetakenwhenliftingtheinstrument.
n Itisprohibitedtoinsertforeignobjectsintoanygapontheinstrumentbeforeorduringatest.
n Unitmustbeusedinawellventilatedlocation.SmallquantitiesofOzonemaybecreatedafterprolongedtesting.
NOTETHE OTS80PB/60PB MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.
IF THESE INSTRUMENTS ARE NOT USED IN THE MANNER SPECIFIED PROTECTION MAY BE IMPAIRED
3
Symbols used on this instrument
G Caution:refertoaccompanyingnotes
F Dangerhighvoltage,riskofelectricalshock
DCvoltage
a ACvoltage
c EquipmentcomplieswithcurrentEUDirectives
u Earthterminal
f Fuse
USBterminal
Equipmentcomplieswith‘Ctick’requirements
Donotdisposeofinthewastestream
4
ContentsSafetywarnings 2Symbolsusedonthisinstrument 3Generaldescription 5Featuresandbenefits 6Applications 6Instrumentcontrolsandindicators 7 Keypadandprinterpanel 7 Rearpanel 8Preparationsforuse 9 Gettingstarted 9 Supplyvoltage 9 Assembly 9 Instrumentsetupprocedure 9 Calibration 9 Testvesselmortisedstirrer 9 Transportationandstorage 9Operatinginstructions 10 Topleveltabs 10 Navigationandcharacterentry 11 Setfavouriteteststandards 12 Performingabreakdowntest 12 Testinprogress 12 Testresults 12Customtests 13 Basicmemoryfunctions 14 TransferresultstoUSBflashdrive 14 Printqualityandchangingpaper 15 Vesselandelectrodes 15 VCM100D/VCM80D 15 Testvesselmotorisedstirrer 15Preventivemaintenance 16 General 16 Cleaning 16 CareforOTSelectrodes 16 Newelectrodes 16 Electrodestorage 16 Cleaningoxidising/dirtyelectrodes 16Technicalspecifications 17 OTS80PB/60PBtechnicalspecification 17 VCM100D/VCM80Dvoltagemeterstechnicalspecification 18Storageandtransportation 18Accessoriesandequipment 19 Optional 19 Spares 19
Repairandwarranty 20
5
General description
Megger’sOTS80PB/60PBOilTestSetsareportableliquiddielectricbreakdowntesters.Theyarefullyauto-matic,poweredbybattery/mainsandfacilitateaccuratebreakdownvoltagetestsonmineral,esterandsiliconinsulatingoils.Usedonsiteaswellasinthelaboratory,theprecisiontestvesselsetstheaccuracystandardfortheliquidinsulationtestindustry.Foreaseofusetheinstrumentscomespre-programmedwitha‘library’ofinternationalbreakdownteststandardsandtheabilitytocreatecustomtests.
TheOTS60PBweighslessthan17kgandtheOTS80PB21.5kgto23.5kgdependingonbatterychoice.Itcomescompletewithoptionalcarrybagandoptionalimpactresistanttransportcase.Thetransportcaseprovidesspacefortheunit,vesselandelectrodeset.
Theinstrumentishousedinametalcasewithmouldedendcapsthataccommodateconvenientcarryhandles.Thetestchamberisdesignedwithaneasy-cleansurfaceandprovidesadrainforspiltoilthroughaninternalpiperoutedtotherearoftheinstrument.Abeakerorsimilarcanbeusedforspillagecollection.Thetestareaiscoveredbyaplasticlidwithanelectricallyprotectiveinnerconductivelayerwhich,withthemetalcase,providesasafeequipotentialenclosureforhighvoltagetesting.
Thelidallowsinplentyofambientlightandusersareabletoviewthetestandwitnessstirringandthebreakdown.Thechamberlidisdualredundantsafetyprotectedwithmicro-switchestopreventaccidentalelectrification.
Theadvancedtestvesseldesignyieldshighlyaccurateandrepeatableresultspreviouslyonlyproducedbylaboratoryinstruments.Twostirringoptionsareavailable;aninternalmagneticbeanstirrerwhichisdroppedintotheoilandanoptionallidmountedimpellerthatconformstotheASTMD1816standard.Stirringduringbreakdowntestingisemployedtoproduceanhomogeneousdistributionofcontaminantswithintheoilandtofreeanycarbongeneratedbetweenelectrodesduringtesting.
Electrodegapissetbyprecisionadjustmentwheelsthatlockinpositionensuringthattheelectrodesdonotmoveduringabusydayinthefield.Itis,however,advisabletochecktheelectrodeconditionandgapdailybecauseitisacriticalparameterinbreakdownvoltagetesting.FlatanodisedAluminiumgapsettinggaugesaresuppliedinpopularsizestofacilitateelectrodegapsetting.Electrodesaredetachableandareavailableinshapesthatmeetallinternationalstandards,i.e.disk,sphericalandmushroom(VDE)electrodes.
TheOTS80PBislinepoweredwiththeoptionofinternalbatteries(eitherLeadAcid,NiMHornobattery)buttheOTS60PBonlyoffersNiMHbatteryornobatteryoption.Inadditiontolinepowercharging,a12VDCchargerfacilityisfactoryfittedwitheitherbatteryoption.Avehicle12VDCchargercableisprovidedtochargetheinstrument’sbatteries.Customerswhoorderoneofthebatterypackoptionsshouldbemindfuloftheneedtomaintainbatterycapacity.Asaminimum,leadacidbatteriesshouldbefullychargedevery3months.CustomerswhochosetheNiMHbatteryoptionareadvisedtooperatefrombatterywithoutlinepowerperiodicallytoensuremaximumbatterylife.
AselectionofinternationalbreakdowntestsarechosenfromalibraryofstandardsequencesandstartedfromtheuserfriendlymenuontheLCDdisplay.Thetopleveltestscreendisplaysallpertinenttestinformationtoensurethecorrectelectrodesareusedwiththespecifiedelectrodegapsetting.
ResultsareviewedonscreenandprintedoneitheranexternalUSBprinterortheinternalprinterfollowingatestresult.Intheeventofabreakdownnotoccurringbeforereachingthemaximumvoltage,theinstrumentwillreturnmaximumvoltage.Thisisasafetymeasuretoensuretheinstrumentisnotdrivenoveritsdesignedmaximumvoltage.
Theinstrumentautomaticallyturnsoffthedisplay15minutesaftercompletionofatestsequenceorlastuse.
Safetyfeaturesincorporatedinthedesignincludedualsafetymicro-switchesandzerovolttouchbar,thescreenedchamberlidandconvienentpowersupplyOn/Offswitch.Pressinganybuttonduringatestwillimmediatelyremovethehighvoltageandterminatethetest.Itisimportanttoearththisinstrumentusingthegroundconnectionattherearoftheinstrumenttoensuresafeoperation.
6
Features and benefitsn Fullyautomaticbreakdowntesterwithtestvoltagesupto80kV/60kVdependingonmodeln Advanced400mloilvesselwithaccurateelectrodegapsettingandadjustmentlockingmechanismn Flatelectrodegapgaugesthatwillnotdamageelectrodesn Easycleantestchamberwithoildrainn Magneticbeadstirreroroptionallidmountedimpeller(conformingtoASTMD1816)n Directtripcontrolemployingvoltageandcurrentmeasurementn Rapidresponsetotrip(<10μs)n Automaticoiltemperaturemeasurementn IntuitiveuserinterfacewithbacklitQVGAcolourdisplay(visibleoutdoors)n SimpletransferoftestresultsviaUSBflashdriven OptionalVCM100D/VCM80Dvoltagecheckunitsmeasurevoltagedirectlyn Optionalcarrycasewithshoulderstrapn Optionaltransportcase
ApplicationsTheOTS80PB/60PBisusedfordeterminingthedielectricstrengthofhighvoltageinsulatingliquids,suchastheinsulatingoilsusedintransformers,switchgearandotherelectricalapparatus.Thedielectricbreakdownvoltageisameasureoftheinsulatingabilityofaliquidtowithstandelectricalstress.Contaminantssuchascellulosicfibres,conductingparticles,dirt,andwaterreducebreakdownvoltageofaninsulatingliquid.Alowresultindicatesthepresenceofoneormoreofthesecontaminants.
Thetestsetisfullyautomatic.Theoperatorhassimplytopreparethetestvessel,loaditwithsampleoil,placeitinthetestchamber,selecttheappropriatespecificationforthetestsandthenstartthetestsequence.Thetestsetautomaticallycarriesoutthesequenceoftestsasdefinedbythepre-selectednationalspecifications.Oiltestingspecifications,forwhichthesetsarepre-programmed,areasfollows:
n ASTMD1816-04
n ASTMD877A&B
n BSEN60156
n CEIEN60156
n IEC601561995-05
n IRAM2341-72
n JISC2101-99(M)&(S)
n PASEVEN60156
n UNEEN60156
n NFEN60156
n SABSEN60156
n VDE0370Part5
n AS1767.2.1
Acriticalfactorwithrespecttofieldtestingofoilisitstemperature,becausesamplesfromatransformertakenshortlyafterswitchoffcanbemanydegreeshotterthanequivalentroomtemperaturelaboratorytestsperformedat20ºCto23ºC.TheOTS80PB/60PBinstrumentsmeasuretheoiltemperatureduringthetestsothattemperaturecorrectioncanbeappliedtoresults.
NB:Certainstandardsrequirefieldteststobeperformedbetween20ºCand30ºC.
7
Instrument controls and indicatorsKeypad and internal printer panelThecontrolpanel,illustratedbelow,islocatedtothefrontrightsideoftheinstrumentandcontainsaQVGAcolourscreen(1),navigationkeypadwith‘OK’(4)buttonanda‘TEST’(5)button.Theoperationisuserfriendlyanddrivenfromsixtoplevelwindows.Astart-upLED(2)isusedduringinstrumentinitialisation.Thelower,verticalsectionofthefrontpanelhousesanON/OFFpowerswitch(6)andaUSBTypeAinterfacesocket
Item Description
1 3.5inch,backlight,brightcolourQVGA display
2 On/offLED
3 ExternalpowerLED/icon
4 TEST’buttonwithassociatedredwarning LED
5 USBsocket 6 Navigationkeypadwith‘OK’selectbutton
Theinstrumentprinterpanel,shownbelow,housestheoptionalinternalprinter.Aflatpanelreplacestheprinterifnotordered.Theprinterisadotmatriximpactprinterwith57.5mmwidepaperroll.Paperresultsprintedonthisprinterwillnotblackeninhightemperatures.Thepanelhastwolockingstudsandacontrolprintadvancebutton.
8
Rear panelTherearpanelillustratedbelowaccommodatesallconnectionstotheinstrumentwiththeexceptionofafrontfacingUSBinterface.
Item Description 1 Groundconnectionwithcablelug
2 IEClinepowersocket
3 Fuses
4 USBTypeAsocket
5 USBTypeBsocket
6 12VDCchargersocket
Itismandatoryforsafeoperationthatanearthgroundcableisattachedtothisinstrumentbeforeperformingtests.Thegroundconnectionisidentifiedasitem1intheillustration.Attentionmustbepaidtoproductwarningsandmarkings.
LinepowerisappliedviaanIECsocket(2)andisfused(3).Theinternalbatteries,ifsupplied,canbechargedbyajacktypesocket(6)fromavehicleorcorrectlyratedwallplugtransformer.Ifbatteriesareorderedwiththisproductachargerandvehiclechargingcableissupplied.Notethattheinstrumentisnotdesignedtooperatefrom12VDC,onlytochargethebatteries.
TwoUSBinterfaceconnectors,onetypeA(4)andonetypeB(5),offerUSBhostanddeviceinterfaces.AsecondUSBtypeA(host)interfaceislocatedonthefrontoftheinstrumentforuserconvenience.TheUSBtypeAinterface(4)isusedtodriveaprinterorflash/harddiskdriveandtheUSBtypeBinterface(5)canbeusedtoconnecttheinstrumenttoalaptopPCorothercomputinghost.
9
Preparations for useGetting startedItisimportanttoreadtheuserguidethoroughlybeforeoperatingtheOTS80PB/60PBforthefirsttime.Thesafetywarningsareparticularlyimportant.
Supply voltageBeforeconnectinglinepowertotheinstrumentthegroundconnectorshouldbeconnectedtoasuitableearthpoint.
TheOTS80PB/60PBisdesignedtooperateon50/60HzAClinepowerintherange90to265VAC.A12VDCpowercircuitforbatterychargingisfittedwheneitherNiMHorleadacidbatterypacksareordered.Thebatterycanbeeitherchargedusingthesupply12Vvehiclechargercableorfrommainspower.
Assembly
n Unpackinstrumentandcontentsofthepackingbox.
n Ensurethattheinstrumentisplacedonasolidsurfaceortablewithsufficientworkingspace.
n Connectgroundtoasuitableearthconnector.
n Fitselectedelectrodestothethreadedshaftsinthetestvesseltakingcarenottoscratchelectrodesurfaces. Electrodesshouldbehandtightenedontotheshaftstopreventanymovementduringuse.
n Adjustmentoftheelectrodegapshouldbedonewiththefeelergaugesprovided,adjustmentwheelslockedand thegaprechecked.
n Fittestvesselintoinstrumentchamber.
n Ifbatteriesarefitted,beforeswitchingontheinstrumentforthefirsttime,connectpowerandleaveforaperiodof atleastfourhours.Thiswillensureafullchargeofthebattery.
n Readthesectioninthisdocumententitled,“InstrumentOperation”beforepoweringuptheOTS80PB/60PBforthe firsttime.
n Connectlinepower.
n Turnontheinstrument.
n Followtheinstrumentsetupproceduredetailedbelow.
Instrument setup procedureTherequiredlanguageoptionisselectedfromthetopleveltabbedwindow.Usetherightarrowkeytonavigateacrossthetabstothelanguagescreen.Theflagsymbolisusedasaniconrepresentinglanguageoption.Simplyusetheup/downarrowbuttonstoselecttherequiredlanguageandpress‘OK’.Thefollowingactions,accessedfromtheSettingstabwindow,shouldbetakenbeforerunningatest:1. Setdateandtime–Time/Dateoption.2. Adjustthedisplaybacklightforoptimumlegibility–DisplaySetupoption.3. Selectunits-metricorimperial.4. Setprinteroptions.
CalibrationTheOTS80PB/60PBarecalibratedinthefactorypriortodeliverysothereisnoneedtocalibrateoninitialsetup.
TheVCM100D/VCM80Dvoltagecheckdevicescanbeusedtoconfirmtheoutputvoltageoftheinstrumentbyreadingthevoltagedirectlyattheelectrodebeforerunningtests.PeriodiccheckingofactualtestvoltageusingtheVCM100D/VCM80Disrecommended.ThevoltagechecktestisaccessedfromtheToolstabwindow.
Test vessel motorised stirrer Unpacktheoptionalmotorisedstirrerassemblyfromitsboxandfitthestirrershafttotheundersideofthestirrerassembly.Theunitisreadyforuse.Theoiltestvessellidisremovedandthestirrerassemblyusedasalidforteststhatrequirestirring.
Whennotinusecareshouldbetakentostorethestirrershaftandstirrerlidassemblyinthecarrybagprovidedwiththeinstrument.Transportation and storageTheOTS80PB/60PBareprecisioninstrumentsthatgeneratesHVto80kVandneedstobetransportedandstoredcarefully.AtransportationcaseisavailableasanoptionandrequiredbycustomerswantingtotransporttheOTS80PB/60PBbyroad,railorair.
Theunitshouldbestoredinaroomorareawheretheenvironmentiswithinthetemperaturerangeof-30ºCto65ºCandmaximumhumidityof95%at45ºC.
10
Operating instructionsTop level tabsTheuserispresentedwithasetofsixtabbed,toplevelwindowsasdepictedbelowwhichfacilitateinstrumentcontrol.Leftandrightarrowkeysnavigatebetweentabs.
Home Setup File
Information Tools Language
Thehometabcontainssummaryinformationforeachshowingtheselectedautomaticteststandard,IDoftheassettheinsulatingliquidcomesfrom,thetypeofinsulator,electrodegapandthestirrerifrequired.Thisisthemainscreenfromwhichtestsarerun.
Setupwindowoffersinstrumentsettingsoflanguage(canalsobesetfromlanguagewindow–overcomesdifficultywheninstrumentissettounknownlanguages),time/date,display,units,printerandasettingtoselectfastdetectionofbreakdown.
Userscanprint,save,recall,deleteanddownloadresultstoPCorUSBdrivefromthefilewindow.
TheInformationwindowoffersdetailsabouttheinstrumentsoftwareconfiguration/versionaswellasahelpfunctionthatservesasanindexeduserguide.
Thetoolswindowofferstheabilitytomanageteststandardsincludingtheabilitytodefinecustomtests,favouritetestsandaddandremovenewstandardsandstandardupgradesviaUSBdiskorcomputerconnection.Anautomaticelectrodecleanfunctionisincludedasameansofremovingdepositsfromelectrodesurfaces.ThiswindowalsorunsatestusingtheVCM100D/VCM80Dvoltagetesterstocheckoutputvoltageattheelectrodebarrier.
Theflagsymbolisusedasaniconrepresentinglanguageoption.Simplyusetheup/downarrowbuttonstoselecttherequiredlanguageandpress‘OK’.
11
Navigation and Character EntryThekeypadconsistsofarrowkeyssurroundingan‘OK’buttonusedfornavigatingthescreensandselectingfunctionsandoptions.Thesesixbuttonsfullycontroltheinstrument.
Withineachuserinterfacewindowtheupanddownarrowbuttonsareusedtonavigateupanddownthefunctionslistedontheleftsideofthescreen.The‘OK’keyeitherscrollsthroughoptionsassociatedwitheachfunctionorentersalowerlevelwindowtoperformafunction.
Leftandrightarrowbuttonsnavigatebetweentabsonthetoplevelwindowforoptimumspeed.Individualtabwindowssavethehighlightfunctionandreturntothesamepositiononeachscreen.
Userscannavigateintolowerlevelwindowsusingtheup/downarrowbuttonsandthe‘OK’buttontoreach,forexample,the‘DisplaySetup’function.
Theusernavigatesto‘DisplaySetup’andpresses‘OK’togettothedisplaysetupwindow.Theup/downortheleft/rightbuttonsadjustbrightnesswithabarindicatorshowingbrightnesspercentage.The‘OK’buttonisusedtosetbrightnessafteradjustmentandsubsequentreturntothesettingswindow.
Thefollowingimagesillustratethisfunction
AvirtualQUERTYkeypadthatfacilitatesalphanumericentrytospecifyassetID,filenameandcustomtestname.Acharacterisselectedbynavigatingtoit,currentcharactershowswithabluebackground,andpressingthe‘OK’button.Aftertypingtherequiredcharactersthevirtualkeyboard‘OK’isselectedtoreturntopreviouswindow.
Numericentryisviaavirtualkeypadasshowntotheright.Digitentryisthesameasthefullvirtualkeypadwithakeypad‘OK’activatingareturntopreviouswindow.
12
Favourite test standardsItisimportanttosetupfavouriteteststandardsfromthestandardslibrarytoavoidnavigatingthoroughtheentirelisteachtimeyouwanttochangetoadifferenttest.Todothisnavigatetothe‘Tools’toplevelwindowandselect‘Manageteststandards’.Thenmovedownto‘Selectfavouritetests’andchooseonlythosetestswhichyourequire,makingsureyouhavethemostrecentversionofeachtest,unlessyouspecificallywanttouseanoldteststandard.Testsaregreyedoutwhenselected.Simplyreturntothepreviousscreenstosaveyourselection.Thescreensbelowillustratethisprocess.
Performing a breakdown testSelectthe‘Home’tabwindowusingthearrowkeysand‘OK’button.Thiswindowoffersasummaryofinformationrequiredforameasurement.Thefollowingimageshowsatypicalsetupwiththeselectedstandardtest,thetestidentificationorassetIDcomprisedofanassetnumberandtwodigitserialnumber,thetypeofoil,requiredelectrodegap,electrodeshapeandfinallythestirrer.
Whentheuserselectsaparticularstandardonlyvalidoptionsarepresentedwithrespecttogapsetting,electrodesandstirreroptions.
Navigationissimple;upanddownarrowbuttonsmoveupanddowntheleftsideofthewindowwhilstoptionsforeachlinecanbescrolledthroughusingthe‘OK’key,orinformationenteredinthecaseofTestID.
Test in progressDuringatestthescreenflashesredwarningHVsymbolstoremindtheuserthatahighvoltagetestisinprogress.Asasafetymeasure,anykeypresswillimmediatelyhaltthetest.
Theelectrodepotentialdifferenceisdisplayedasitrampsupinover-sizeddigitstoemphasisethepresenceofhighvoltage.
Betweenteststhereisadwelltimeandpossiblyastirringaction.Acountdowntimerindicatesperiodofdwellorstirringremaining.
13
Test resultsAttheendofanytestresultsareshownonthescreen.AtypicalexampleistheIEC60156test.Thisimageshowsthecompleteinformationpresentedtotheuserhoweverthescreensizeisnaturallysmallersoasliderfunctionenablestheusertonavigatethroughtheresultandprintitifrequired.
ThetestresultforIEC60156containsresultsofsixsequentialbreakdowntestswithteststatisticsofaveragebreakdownvoltage,datadispersionandstandarddeviation.
Theresultswindowfacilitatesadirectprintfunctionatthebottomoftheresultwindow.
Custom testsCustomtestsareteststhatcanbedefinedbyauserbygivingtheusercontrolovertherangeofparameterssetinatest.Threecustomtestshavebeenpre-programmedforuserdefinition;n Custom(5)–thistestcontainsfivesuccessivebreakdowntestsn Custom(6)–thistestcontainssixsuccessivebreakdowntestsn Custom(10)–thistestcontainstensuccessivebreakdowntests
NavigatetotheToolstoplevelscreenasillustratedbelow.Select,“ManageTestStandards”bypressingtheOKbutton.Godownto,“EditCustomTestParameters”andpresstheOKbutton.Thecustomparametersscreenshowsasetofvariablesthatcontrolabreakdowntest.Eachvariableisusersettable,simplynavigatedowntotherequiredsetting,presstheOKbuttonandproceedtochangethevalueasrequired.FornumericentryontheOTSAFmodelsthedecimalpointcanbeenteredbyquicklypressingthedigit‘1’twice.Sotoenter2.5mmgapsettingtheuserentersthefollowingsequence:n 2n 1,1n 5n OK
Inthecustomstestparametersscreen‘setmax.kV’mayappearas,‘setkV(max.xxx)’incertainversionsoffirmware.Allentryfieldsinthecustomtestparametersscreenarerangecheckedtoensureitisvalid.WhenallvaluesaresettheusernavigatesbacktotheToolsscreenviatheRETURNfunction.TheusershouldensurethatCustom(5)andCustom(6)areselectedasfavouritetestsbeforeleavingthemanageteststandardsscreen.
14
Basic memory functionsTheinternalmemoryislimitedtofiftytestrecordswherearecordrepresentsafullresultofoneoftheprogrammedstandardtests.BestpracticeistobackupinternalresultsonadailybasisbyeitherdownloadingdataviaUSBcabletoPowerDBLiteorsavingontoaUSBmemorydevice.WhensavingtoUSBmemorydevice,afilenameintheformatofamaximumeightcharactersisrequiredandthefilenameenteredmustonlycontaincharactersA-Z,a-zand0-9.
Memoryfunctionsareaccessiblefromthefilesmenuandincludethefollowing:• Save• Recall• Delete• Printlastresult
TheusercanselectbetweeninternalmemoryandaUSBmemorydeviceforthesave,recallanddeletefunctions.Afterselectionofamemorylocationtherelevantfilesaredisplayedonscreen.Navigationisviathearrowkeysasfollows:
Upanddownarrowmoveupanddownoneitematatime.Leftandrightarrowsimplementapageupandpagedownfunctiontoassistnavigationwherelargenumbersoffilearepresent.
Inthedeletefunctionwindowtheusercanselectall,deselectallorselectonefileatatime.Selectasinglefilebynavigatingtothefile(bluehighlight)andpressingOKtoselect(redhighlightchangingtogreywhennavigatingawayfromtheselectedfile.)
Therecallfunctionoperatesononefileatatime.
Transfer results to USB flash driveUsersmaywanttocopyresultsfrominternalmemorytoaUSBmemorydeviceaseitheranend-of-dayprocedureorwheneverthe50resultfilelimitisreached.ThefollowingprocedurewillbackupfilestoaUSBmemorydevice:
1. Fromthefilemenu,select‘Save’2. Then‘SelectUSB’3. EnterafilenameusingonlycharactersA-Z,a-zand0-9viathevirtualkeypadandreturntosavefunctionscreen4. Navigatedownto‘SelectResults’–afulllistofinternallysavedresultsarepresentedbyTestID.5. Selectrequiredresultsandthen‘SAVE’totransfertheresultstofilename(asentered)onthememorydevice.6. Navigatebacktotoplevelfunctionsandremovethememorydevice(ifrequired)
15
Print quality and changing paperTheprintercoverisremovedbyturningthetwofastenersthroughaquarterturnandpullingthepaneloff.
Theprinterribbonmayneedtobemovedontoimproveprintquality.Todothis,removetheprintercoverandturntherighthandrotarywinderindirectionindicatedontheribboncase.Tochangetheprinterribbon,simplypresstheribbonassemblywhereitsays,“PUSHEJECT,”toremovetheoldribbonandreplacewithanewone.ThepartisanEpsonERC-09,Meggerpartnumber25995-002.
Paperisfedintothehousingwiththelosepaperfacingtherearoftheinstrument.Ensurethatthepaperisneatlycutandnotcurledupattheend.
Holdingtheendofthepaperroll,insertthepaperunderthecrossbarandintotheprinterslotasshown(bluearrows)inthedrawingopposite.Presstheprintfeedkeyrepeatedlyuntilthepaperisjustvisiblefeedingthroughtheprintmechanism.
Replacetheprinterfrontcoverbyfirstlocatingitatthebaseofthepanelandsecureitwiththefastenerswhichareturnedbackaquarterturntosecurethepanelinplace.Finallypressthepaperfeedbuttontoadvancethepaperthroughthefrontcoverpaperslot.
Vessel and ElectrodesTheoiltestvesselandelectrodesrequirenomaintenanceotherthancleaning.Electrodescanbeunscrewedandreplacedwithanothershapeelectrode.Electrodesshouldbeexaminedregularlytoensurenopittinghasoccurredandtheelectrodesrenewedwherepittingisevident.
Theinstrumentincludesabuiltinelectrodecleaningfunctionwhichcanbeactivatedfromthetoolsmenu.Thisfunctionwillperform24sequentialbreakdownstoremovedepositsonelectrodesurfaces.
VCM100D/VCM80DTheVCM100D/VCM80DarenotuserserviceableitemsbutmayrequireperiodiccleaningusingacleanclothdampenedwithIsopropylAlcohol(IPA).
Test vessel motorised stirrerMegger’smotorisedstirrerisintegratedintoamodifiedlidassemblyandsimplyreplacesthevessellid.Powerisconnectedviaaflushmountedrearinsertandisdisconnectedasthestirrerassemblyislifted.Thisunitisnotuserserviceable.
16
Preventive maintenance
GeneralTheOTS80PB/60PBisaHVinstrumentgeneratingupto80kVandisnotuserserviceable.Theusershouldensurethatthelidisnotcrackedordistortedbeforeuse.Theonlyinternalpartoftheinstrumentthatisaccessibletotheuseristheprinterpaperhousingallowingrenewalofpaperandribbonwhenrequired.OpeningtheOTS80PB/60PBisstrictlyforbiddenandwillconstituteabreachofwarrantee.Thereisnoneedforanytoolstobeusedwiththisinstrumentotherthanthesuppliedgaugesandmagneticbeanretriever.
CleaningTocleantheoutersurfacesoftheOTS80PB/60PB,firstdisconnectfromlinepowerandthenwipetheinstrumentusingacleanclothdampenedwithIsopropylAlcohol(IPA).
Theusermustensurethatthetestchamberisalwayskeptclean,particularlypriortoatest.Wipeawayanyspiltoilinthechamberorontheoutsideofthetestvesselwithalintfreecloth.Whenalotofoilisspiltinthetestchambertheunitisdesignedwithadrainfacilityattherear.Simplyunclipthecleartubeanddrainoilintoabeakerorothersuitablecontainer.
Tocleantheinsideofthetestvesselfollowtheinstructionsgivenintherelevanttestspecificationor,inthecaseofnoinstructions,cleanedwithasmallvolumeofthenextsampleofoiltobemeasured.
Care for OTS electrodesBrasselectrodessuppliedwiththeMeggeroiltestsetswillshowsignsofdepositsbuiltupfrominsulatingoilbreakdowntesting.Iflefttheywilloxidiseandappeardull.
New electrodesMeggerelectrodesaresuppliedinakitwithgapgaugesandmagneticstirrerbeansandretrievalstick.theelectrodeshavedifferentshapesbutarecaredforinthesameway.Ifnewelectrodesrequirecleaning,isopropylalcohol(IPA)maybeusedtocleanthemoff.Itisadvisabletoimmersetheelectrodesincleaninsulatingoilforacoupleofhourspriortouse.
Electrode storageElectrodesmaybestoredinasuitablecontainerorimmersedincleanmineralinsulatingoil.Electrodesinatestvessellefttostandovernightmaybeleftwiththelastoilsampletestedinthevessel.
Cleaning oxidised/dirty electrodesDirtyelectrodesmaybecleanedwithaclean,softclothandbrasscleanerbutcareshouldbetakentouseminimalpressuretoavoidremovalofexcessiveelectrodematerial.AfterremovingthedirtcleantheelectrodeswithacleanclothandIPA.Itisadvisabletoimmersetheelectrodesincleaninsulatingoilforacoupleofhourspriortouse.
Ifelectrodesarepittedorscratchedtheyshouldbediscardedandnewelectrodesfitted.
17
OTS80PB/OTS60PBTechnical SpecificationThefollowingtechnicalspecificationsapplytotheOTS80PB/60PB:
Test voltage -40to0to40kV,[email protected]
Voltage resolution 0.1kV
Voltage accuracy ±1%±2digits
Vessels400ml(standard)150ml(option) Thermoplasticpolyamidechamberprovidesprecisionelectrode alignment,adjustmentwheelslockelectrodesinposition
Temperature sensor resolution 1ºC
Power supply Linevoltage/power 90to264VAC,300VA Linefrequency 48/63Hz
Fuses 2x4ACFSTceramic,20mmx5mm,for250V
Batteries (optional) LeadAcid2x12V4Ah,orNiMH24V2Ah
DC power supply 12VDC,60VA–fittedwithbatteryoption,usedtochargebatteryonly
Internal printer (option) Dotmatriximpactprinterwithgraphicaloutput Paper57.5mmwide,weight52gmperroll EpsonblackribboncassetteERC-09B
Protection Safetyinterlockontestchamber Dischargebarrieronpolycarbonatetestchambercover
Display 320x240QVGAcolourdisplaywithbacklight
Dimensions (W x H x D) OTS80PB 520mmx380mmx250mm
OTS60PB 520mmx330mmx250mm
Weight OTS80PB 20.5kg(nobattery) 21kg(NiMHbattery) 23.5kg(withLeadacidbatteries)
OTS60PB 16.3kg(nobattery) 16.8kg(NiMHbattery)
Operating temperature 0ºCto+50ºC
Storage temperature -30ºCto+65ºC
Humidity 80%RHat40ºCoperation 95%RHat40ºCstorage
Altitude 1000m
Safety DesignedinaccordancewithIEC61010
EMC LightindustrialIEC61326-1ClassB,CISPR22,CISPR16-1andCISPR16-2
18
VCM100D/VCM80D Voltage check metersTechnical SpecificationTheVCM100DandVCM80DareOTSaccessoriesthatenablethevoltageoutputoftheoiltestsettobecheckedbymeasuringtheactualvoltageattheHVelectrodes.ThecalibrationmetersfitintheoiltestsetchamberinplaceofastandardvesselandaredesignedtoensurethattheloadontheOTStransformerissimilartotheloadduringanoildielectricstrengthtest.
Meter type Proprietary100kV/80kVMeggerOTSvoltmeterwith4seven-segment digitdigitalreadout
Accuracy at 23 °C 2.0%plus2digitsto50kVrms 2.5%plus2digits>50kVrmsto75kVrms 3.0%plus2digits>75kVrmsto100kVrms(VCM100Donlyabove80kV)
Temperature range Operational 0°Cto40°C
Storage -30°Cto+70°C
Humidity range Operational 50%RHat20°C
Storage 93%RHat40°C
Safety ThemeterisonlyapprovedforusewithMeggeroiltestsets
Dimensions (W x H x D) 250mmx150mmx115mm
Weight 1.8kg
Storage and transportationTheOTS80PB/60PBareprecisioninstrumentthatgeneratesHVto80kVandneedstobetransportedandstoredcarefully.AtransportationcaseisavailableasanoptionandrequiredbycustomerswantingtotransporttheOTS80PB/60PBbyroad,railorair.Asoft,shouldercarrycaseisprovidedasanoptionwiththeOTS80PB/60PB.Itopensuptofacilitaterunningatestwithoutremovaloftheinstrumentandprovidestwosidepouchesforstorage.
Theunitshouldbestoredinaroomorareawheretheenvironmentiswithinthetemperaturerangeof-30ºCto65ºCandmaximumhumidityof95%at45ºC.
IftheOTS80PB/60PBhasbatteriesfitteditshouldbefullychargedatleastonceeverythreemonthstomaintainfullbatterycapacity.
19
Only Megger supplied accessories can be used with the OTS range. Non approved equipment cannot be guaranteed to work with the OTS.
Factory fitted accessoriesn Internalprintern Battery–eitherLeadacid(OTS80PBonly),NiMHbatteryornobattery(linepoweronly)optionn Internal12Vchargerand12Vvehiclecabletochargebatteries(nottooperatetheinstrument)n Electrodesets–IEC,ASTMandfullelectrodesetcontainingoneofeachtypeofelectroden Vessellidmountedimpellerforusewith400mlvesseln Powerlead
Optional accessories & sparesn Transportationcasen VCM100D/VCM80Dvoltagecheckmetersn Vesselassembly400mln Vesselassembly150mln Vessellidmountedimpellerforusewith400mlvesseln ElectrodesetsIEC156n ElectrodesetASTMD877/1816n Fullelectrodeset–containsallelectrodesn Printerpaper
20
Repair and warranty
TheTestsetcircuitcontainsstaticsensitivedevices,andcaremustbetakeninhandlingtheprintedcircuitboard.Iftheprotectionofaproducthasbeenimpaireditshouldnotbeused,andbesentforrepairbysuitablytrainedandqualifiedpersonnel.Theprotectionislikelytobeimpairedif,forexample,theproductshowsvisibledamage,failstoperformtheintendedmeasurements,hasbeensubjectedtoprolongedstorageunderunfavourableconditions,orhasbeenexposedtoseveretransportstresses.
New products are guaranteed for 1 year from the date of purchase by the user.Note: Any unauthorized prior repair or adjustment will automatically invalidate the warranty.
InstrumentrepairandsparepartsForservicerequirementscontact:–
MeggerLimited or Megger ArchcliffeRoad ValleyForgeCorporateCenter Dover 2621VanBurenAvenue Kent,CT179EN Norristown,PA194 England U.S.A.
Tel:+44(0)1304502243 Tel:+1(610)676-8579 Fax:+44(0)1304207342 Fax:+1(610)676-8625
oranapprovedrepaircompany.
Approved repair companiesAnumberofindependentcompanieshavebeenapprovedforrepairworkonmostMeggerproducts,usinggenuineMeggerspareparts.ConsulttheAppointedDistributor/Agentregardingspareparts,repairfacilitiesandadviceonthebestcourseofactiontotake.
Returning a test set for repairIfreturningatestsettothemanufacturerforrepair,itshouldbesentfreightpre-paidtotheappropriateaddress.Acopyoftheinvoiceandofthepackingnoteshouldbesentsimultaneouslybyairmailtoexpediteclearancethroughcustoms.Arepairestimateshowingfreightreturnandotherchargeswillbesubmittedtothesender,ifrequired,beforeworkontheinstrumentcommences.
21
22
AUSTRALIAMegger Pty LimitedUnit 1, 11-21 Underwood RoadNSW 2140T +61 (0)2 9397 5900F +61 (0)2 9397 5911E [email protected]
dUbAIMeggerDubaiUnited Arab EmiratesT +971 4 456 9057E [email protected]
CANAdA Megger110 Milner Avenue Unit 1Scarborough Ontario M1S 3R2T +1 416 298 9688 (Canada only)T +1 416 298 6770F +1 416 298 0848E [email protected]
FRANCE Megger SARLZ.A. Du Buisson de la Couldre23 rue Eugène Henaff78190 TRAPPEST +33 (0)1 30.16.08.90F +33 (0)1 34.61.23.77E [email protected]
dEUTSChLANd Megger GmbHObere Zeil 261440 OberurselT 06171-92987-0F 06171-92987-19E [email protected]
INdIA Megger (India) Pvt Limited501 Crystal Paradise MallOff Veera Desai RoadAndheri(w)Mumbai - 400053T +91 22 26740468F +91 22 26740465
SoUTh AFRICAPO Box 22300Glen Ashley 4022DurbanSouth AfricaT +27 031 5646578F +27 031 5637990
SvERIgEMegger Sweden ABEldarvägen 4Box 2970SE-187 29 TÄBYT +46 8 510 195 00F +46 8 510 195 95E [email protected]
SChwEIzMegger AGOb. Haselweg 6305727 Oberkulm, AargauT +41 62 768 20 30F +41 62 768 20 33
ThAILANdMegger30/46 Moo 13Nongprue, BanglamungChonburi 20150T +66 860103395E [email protected]
UNITEd KINgdomMegger LimitedArchcliffe Road, DoverKent CT17 9ENT +44 (0)1 304 502101F +44 (0)1 304 207342E [email protected]
USA Megger4271 Bronze Way, Dallas,Texas 75237-1019T +1 800 723 2861 (USA ONLY)T +1 214 333 3201F +1 214 331 7399E [email protected]
USA MeggerValley Forge Corporate Centre2621 Van Buren AvenueNorristown, PA 19403 T +1 610 676 8500F +1 610-676-8610 (USA ONLY) E [email protected]
USA Megger4064 State Highway 6 SouthCollege Station, TX 77845T +1 979 690 7925F +1 979 690 0276E [email protected]
ThisinstrumentismanufacturedintheUnitedKingdom.Thecompanyreservestherighttochangethespecificationordesignwithoutpriornotice.
Meggerisaregisteredtrademark