osu rhr apparatusair supply assembly: 127-220v, 3 phase, 50hz, 30 amps /120-208v, 3 phase, 60hz, 30...

2
[email protected] +44 (0) 1903 784 336 OSU RHR APPARATUS The instrument measures the heat release properties of aircraft interior materials. The Apparatus package includes the following: All stainless steel construction 2 Specimen Holders Extensive control panel Calibration rig with centre and corner transducers Mass flow controller for calibration Data acquisition system Computer with installed software to automatically run calibration & calculate test results Water Cooled Air Supply System consisting of: • Motor/Blower to deliver required air pressure • Variable Frequency Drive With FRI Filter • Water cooled heat exchanger • Water separator and automatic drain • Air pressure gauge and safety relief valve • Venturi pipe for pressure measurement • Flux transducer calibration rig cooling, including flow meter, regulator, solenoid valve cut off • Plus all necessary wall mounting and pipework, including fittings for the above PLEASE NOTE: The U.S.A. Federal Aviation Administration requires all test apparatus used to perform this Appendix F Part IV test must be approved by the FAA to determine physical conformance. All OSU instruments are fully inspected by the FAA at the manufacturing site and are delivered complete with a certificate of conformity issued by the FAA. FAR Part 25 Appendix F Part IV Boeing BSS 7322 Airbus AITM 2.0006 FEATURES SPECIFICATIONS

Upload: others

Post on 28-Jul-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: OSU RHR APPARATUSAir Supply Assembly: 127-220V, 3 phase, 50Hz, 30 amps /120-208V, 3 phase, 60Hz, 30 amps Ambient Temperature: Operating 10 C to 35 C Apparatus Dimensions: 1550mm(w)

[email protected] +44 (0) 1903 784 336

OSU RHR APPARATUS

• Theinstrumentmeasurestheheatreleaseproperties

ofaircraftinteriormaterials.TheApparatuspackage

includesthefollowing:

• Allstainlesssteelconstruction

• 2SpecimenHolders

• Extensivecontrolpanel

• Calibrationrigwithcentreandcornertransducers

• Massflowcontrollerforcalibration

• Dataacquisitionsystem

• Computerwithinstalledsoftwaretoautomaticallyrun

calibration&calculatetestresults

• WaterCooledAirSupplySystemconsistingof:

•Motor/Blowertodeliverrequiredairpressure

•VariableFrequencyDriveWithFRIFilter

•Watercooledheatexchanger

•Waterseparatorandautomaticdrain

•Airpressuregaugeandsafetyreliefvalve

•Venturipipeforpressuremeasurement

•Fluxtransducercalibrationrigcooling,includingflow

meter,regulator,solenoidvalvecutoff

•Plusallnecessarywallmountingandpipework,

includingfittingsfortheabove

PLEASE NOTE:

TheU.S.A.FederalAviationAdministrationrequiresalltestapparatususedtoperformthisAppendixFPartIVtestmustbeapprovedbytheFAAtodeterminephysicalconformance.AllOSUinstrumentsarefullyinspectedbytheFAAatthemanufacturingsiteandaredeliveredcompletewithacertificateofconformityissuedbytheFAA.

FAR Part 25 Appendix F Part IVBoeing BSS 7322Airbus AITM 2.0006

FEATURES

SPECIFICATIONS

Page 2: OSU RHR APPARATUSAir Supply Assembly: 127-220V, 3 phase, 50Hz, 30 amps /120-208V, 3 phase, 60Hz, 30 amps Ambient Temperature: Operating 10 C to 35 C Apparatus Dimensions: 1550mm(w)

[email protected] +44 (0) 1903 784 336

TECHNICAL DATA

Gas Supply:Commercialmethanewithapurityof99%

Air Supply (if customer supply): Clean,dryandfilteredairataflowrateof0.05m3/sec.Airtemperaturetobe21-24°C

SERVICES REQUIRED

Electrical: MainFrame(includingheatingelements):208-220V,singlephase,

50/60Hz,50amps

AirSupplyAssembly:127-220V,3phase,50Hz,30amps/120-208V,

3phase,60Hz,30amps

Ambient Temperature: Operating10°Cto35°C

Apparatus Dimensions:1550mm(w)x700mm(d)x2110mm(h)

OSU RHR APPARATUS