Page 1: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Orthopedic Radiology

Dr. Terence A. Perrault D.C., D.A.C.B.R.

Professor of Clinical Sciences

University of Bridgeport

College of Chiropractic

Page 2: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic Imaging Modalities

Dr. Terence A. Perrault D.C., D.A.C.B.R.

Director of Radiology

University of Bridgeport College of Chiropractic

Page 3: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiographyUtilizes ionizing radiation to penetrate organic matter

Xrays are simply EM radiation of higher intensity/energy than visible light (Shorter wavelength, higher frequency)

Casts shadow of dense structures on a film

Page 4: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Page 5: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Recognizes only 5 densities:

Air (Gas)

Fat

Water

Bone (Calcium)

Metallic

Page 6: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Need mixture of all 4 physiologic densities to get diagnostic film.

Metallic density degrades image (in most cases)

Need minimum of 2 views at 90 degrees to each other for localization of structures.

Page 7: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Patient 1

Page 8: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Patient 1

Page 9: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Patient 2

Page 10: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Patient 3

Page 11: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Patient 4

Page 12: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Plain film radiography

Advantages:

Availability

Quick imaging

Relatively inexpensive

Good screening tool

Disadvantages:

Ionizing radiation

Insensitivity

Shows only structure

Poor tissue differentiation

Page 13: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Contrast radiography

Metallic density degrades images In most cases

Contrast material (barium or iodine based) often used to opacify tubular structures for visualization

Page 14: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Contrast radiography

Page 15: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Patient #3

Patient 3

Page 16: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Contrast radiography

Patient 3

Page 17: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Utilizes xray beam (ionizing radiation)

Encircles patient with xray beam and radiation detectors measure attenuation of beam.

Gives cross sectional images of the region of interest

Page 18: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 19: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 20: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography All images are digital

and stored in computer

This digital information can be manipulated later to enhance certain tissues

Creates bone and soft tissue “windows”

Page 21: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography Image

reconstruction

Digital grids can be reconstructed in many planes

Page 22: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography Imaging

reconstruction via computer reformatting can produce images in multiple planes

Page 23: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 24: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 25: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 26: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 27: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Page 28: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Bone window Gas window

Page 29: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Patient #1

Patient 1

Page 30: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Patient 1

Page 31: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

“Star” defect resulting from metallic artifact

Page 32: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Patient #3

Patient 3

Page 33: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography

Patient 3

Page 34: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Computed Tomography Advantages:

Highly sensitive

Quick acquisition times

Lung / G.I. Imaging

Available

Disadvantages:

Ionizing radiation

Soft tissue differentiation not as good as MRI

Relative cost

Structure not function

Page 35: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imaging(MRI)

Fat is very high in free hydrogen and gives off a very high signal

Muscle gives off varying amounts of signal based upon various physiologic properties

Bone, tendons, and ligaments have hydrogen bound in crystalline-like lattice and unable to be manipulated by RF,.. No signal

Page 36: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imaging

Alterations of free hydrogen (water) content in tissues affect their ability to “relax” from

RF pulse and re-align into the main magnetic field

These relaxations are designated T1 and T2

Each tissue has different values for T1 & T2 relaxation

Page 37: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imaging Bone, tendons and ligaments

Have hydrogen bound into crystalline-like lattice

Unable to manipulate with magnetic fields or RF pulses

Therefore no signal given off

Appears black on images

Flowing blood also does not give an image “Flow void” more about this later…,

Page 38: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

MR Imaging Parameters

Type TR (msec)

TE (msec)

CSF FAT

T1 weighted

Short (400-800)

Short (20-25)

Dark Bright

T2 weighted

Long (1500-2000)

Long (>60)

Bright Less Bright

PD Intermed

Long(1500-3000)

Medium (30-50)

Gray Gray

Page 39: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Manipulation

T1-weighted Lumbar spine T2-weighted Lumbar spine

Page 40: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imaging

2 Major types of MR scanners

Air core (closed)

Solid Core (open)

Page 41: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic resonance imagers

Air core (closed MR unit)

Page 42: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic resonance imagers

Solid Core (Open MR unit)

Page 43: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Patient #1

Patient 1

Page 44: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imaging

Artifact resultingFrom metallic Fragment distortingMagnetic field

Patient 1

Page 45: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imaging Artifact

Page 46: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Patient 2 This is the plain film

xray from earlier in the lecture

Did you see any injuries

Patient 2

Page 47: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic resonance imaging

Patient 2

Page 48: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic resonance imaging

Patient 3

Page 49: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic resonance imaging

Patient 4

Page 50: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic resonance imaging

Advantages:

No ionizing radiationExcellent soft tissue demonstrationVery sensitize to tissue changesShows structure AND function

Disadvantages:

Slow acquisition timesExpensiveAvailabilityUncomfortableClaustrophobicMagnetic shielding problemsHeavy machinery

Page 51: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imagingwith Fat Saturation

Page 52: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Imagingwith Flexion/Extension

Page 53: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

MRI pulse sequences

Collection of specific imaging parameters selected for a scan of a patient

Typical musculoskeletal exam contains 3-6 sequences in various planes

Page 54: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Pulse Sequence Strengths & WeaknessesSequence Strength Weakness

Spin Echo T1Anatomic detail, Fat, Subacute Hemorrhage, Marrow, Menicus,

Contrast

ST edema, other fluid

Fast Spin Echo T2

Marrow path when fat-sat. used, Good for pts with metal hardware,

Fluid

Poor marrow w/o fat-sat

Gradient Echo T2*Fibrocartilage, Loose bodies & Hemorrhage(susceptibility effects)

Poor marrow, metallic hardware

Short Tau Inversion recovery

(STIR)

Marrow & ST Pathology d/t fat suppression

Not to be used with contrast

Fluid Attenuation Inversion Recovery

(FLAIR)

Similar to STIR mostly for Brain and Neuro tissues

Page 55: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Arteriography (MRA)

Utilizing the “flow void” of blood to produce an image.

It is possible via the software to digitally eliminate the signal of most/all tissues with exception of flowing blood

Page 56: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Arteriography (MRA)

This will produce the appearance of only the vascular structures and any abnormalities

The images appear as if contrast material was injected without the invasiveness of arteriography

Page 57: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Flow Void

Antenna

Page 58: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Flow Void

Antenna

Page 59: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Flow Voids

Page 60: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Arteriography (MRA)

Utilizing the “flow void” to produce an image

Page 61: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Magnetic Resonance Arteriography (MRA)

Middle aged male patient with chronic, progressive neck pain and no response to treatment

Special thanks to Dr. Terry Yochum

Page 62: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Bone scan (scintigraphy)

Utilizes radioactive pharmaceutical injected into blood stream

Agent accumulates in regions of increased blood flow and increased bone metabolism

Patient is scanned, and “hot spots”demonstrate the areas of accumulation

Page 63: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Bone scan (scintigraphy)

Page 64: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Bone scan (scintigraphy)

Page 65: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Bone scan (scintigraphy)

Page 66: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Bone scan (scintigraphy)

Lung perfusion scans

Page 67: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Scintigraphy (Bone scan)

Dx: Osteoid Osteoma

Page 68: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Bone scan (scintigraphy)

Advantages:

Highly sensitive to bone changes

Shows function changes early

Relatively cheap procedure

Disadvantages:

Ionizing radiation

Poor specificity for lesions

Invasive procedure Infection/allergies

Normal “hot spots” Open epiphyses

Page 69: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

SPECT scan

Single photon emission computed tomography

Combination of bone scanning with ability of CT to give tomographic “slices”through body

Page 70: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

SPECT scanner

Page 71: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

SPECT scan

Pars interarticulares stress reaction (fracture?) in young athlete

Page 72: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Osteoid Osteoma

Page 73: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

SPECT-CT Scan

Page 74: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound Soundwaves penetrate surface of

patient

Waves are reflected back to surface as soundwaves pass through different densities of tissues

Reflected waves are recorded and an image constructed

Page 75: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound Soundwaves reflected

back based upon changes in densities (interface).

Allows evaluation of fluid accumulation as well as fibrotic changes in tissue

Page 76: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound Realtime images

Allows multiplanar imaging

Areas in motion

Page 77: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound

Page 78: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound

Page 79: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound

Page 80: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound

Page 81: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound

Page 82: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound

Tenosynovitis of the tibialis posterior tendon

Page 83: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic Ultrasound

OUCH!!

Page 84: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic Ultrasound

Page 85: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

Diagnostic ultrasound Advantages:

Real time imaging

Good soft tissue resolution

No ionizing radiation

Relatively inexpensive

Shows early edema, and fibrotic changes

Disadvantages:

Very difficult to interpret

Not all areas thoroughly researched (spine)

Shadowing effect

Page 86: Orthopedic Radiology - files.bridgeport.edu · Magnetic Resonance Imaging (MRI) Fat is very high in free hydrogen and gives off a very high signal Muscle gives off varying amounts

This ends our modalities review, thanks your time