Optimised extraction of heterocyclic aromatic amines from blood using hollow fibre membrane liquid-phase microextraction and triple quadrupole mass spectrometry Cooper, K., Jankhaikhot, N., & Cuskelly, G. (2014). Optimised extraction of heterocyclic aromatic amines from blood using hollow fibre membrane liquid-phase microextraction and triple quadrupole mass spectrometry. Journal of Chromatography A, 1358, 20-28. DOI: 10.1016/j.chroma.2014.06.061 Published in: Journal of Chromatography A Document Version: Peer reviewed version Queen's University Belfast - Research Portal: Link to publication record in Queen's University Belfast Research Portal Publisher rights © 2014 Elsevier. This manuscript version is made available under a Creative Commons Attribution-NonCommercial-NoDerivs License (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits distribution and reproduction for non-commercial purposes, provided the author and source are cited. General rights Copyright for the publications made accessible via the Queen's University Belfast Research Portal is retained by the author(s) and / or other copyright owners and it is a condition of accessing these publications that users recognise and abide by the legal requirements associated with these rights. Take down policy The Research Portal is Queen's institutional repository that provides access to Queen's research output. Every effort has been made to ensure that content in the Research Portal does not infringe any person's rights, or applicable UK laws. If you discover content in the Research Portal that you believe breaches copyright or violates any law, please contact [email protected]. Download date:15. Feb. 2017

Upload: others

Post on 08-Aug-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Optimised extraction of heterocyclic aromatic amines from bloodusing hollow fibre membrane liquid-phase microextraction andtriple quadrupole mass spectrometryCooper, K., Jankhaikhot, N., & Cuskelly, G. (2014). Optimised extraction of heterocyclic aromatic amines fromblood using hollow fibre membrane liquid-phase microextraction and triple quadrupole mass spectrometry.Journal of Chromatography A, 1358, 20-28. DOI: 10.1016/j.chroma.2014.06.061

Published in:Journal of Chromatography A

Document Version:Peer reviewed version

Queen's University Belfast - Research Portal:Link to publication record in Queen's University Belfast Research Portal

Publisher rights© 2014 Elsevier. This manuscript version is made available under a Creative Commons Attribution-NonCommercial-NoDerivs License(https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits distribution and reproduction for non-commercial purposes, provided theauthor and source are cited.

General rightsCopyright for the publications made accessible via the Queen's University Belfast Research Portal is retained by the author(s) and / or othercopyright owners and it is a condition of accessing these publications that users recognise and abide by the legal requirements associatedwith these rights.

Take down policyThe Research Portal is Queen's institutional repository that provides access to Queen's research output. Every effort has been made toensure that content in the Research Portal does not infringe any person's rights, or applicable UK laws. If you discover content in theResearch Portal that you believe breaches copyright or violates any law, please contact [email protected].

Download date:15. Feb. 2017

Page 2: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Accepted Manuscript

Title: Optimised extraction of heterocyclic aromatic aminesfrom blood using hollow fibre membrane liquid-phasemicroextraction and triple quadrupole mass spectrometry

Author: Kevin M. Cooper Natcha Jankhaikhot GeraldineCuskelly

PII: S0021-9673(14)00987-XDOI: http://dx.doi.org/doi:10.1016/j.chroma.2014.06.061Reference: CHROMA 355540

To appear in: Journal of Chromatography A

Received date: 16-4-2014Revised date: 3-6-2014Accepted date: 19-6-2014

Please cite this article as: K.M. Cooper, N. Jankhaikhot, Optimised extraction ofheterocyclic aromatic amines from blood using hollow fibre membrane liquid-phasemicroextraction and triple quadrupole mass spectrometry, Journal of ChromatographyA (2014), http://dx.doi.org/10.1016/j.chroma.2014.06.061

This is a PDF file of an unedited manuscript that has been accepted for publication.As a service to our customers we are providing this early version of the manuscript.The manuscript will undergo copyediting, typesetting, and review of the resulting proofbefore it is published in its final form. Please note that during the production processerrors may be discovered which could affect the content, and all legal disclaimers thatapply to the journal pertain.

Page 3: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 1 of 28

Accep

ted

Man

uscr

ipt

1

Optimised extraction of heterocyclic aromatic amines from blood 1

using hollow fibre membrane liquid-phase microextraction and triple2

quadrupole mass spectrometry3

4

Kevin M. Cooper*, Natcha Jankhaikhot, Geraldine Cuskelly5

6

Institute for Global Food Security, School of Biological Sciences, Queen’s University 7

Belfast, David Keir Building, 18-30 Malone Road, Belfast, BT9 5BN, United Kingdom8

9

Email addresses: [email protected]; [email protected]; [email protected]

11

*Corresponding author at: Institute for Global Food Security, School of Biological 12

Sciences, Queen’s University Belfast, David Keir Building, 18-30 Malone Road, Belfast, 13

BT9 5BN, United Kingdom. Tel.: +44 (0)28 90976562; Fax: +44 (0)28 90976513; Email: 14

[email protected]

16

17

Abstract18

Heterocyclic aromatic amines (HCA) are carcinogenic mutagens formed during cooking 19

of proteinaceous foods, particularly meat. To assist in the ongoing search for 20

biomarkers of HCA exposure in blood, a method is described for the extraction from 21

human plasma of the most abundant HCAs: 2-Amino-1-methyl-6-phenylimidazo(4,5-22

b)pyridine (PhIP), 2-amino-3,8-dimethylimidazo[4,5-f]quinoxaline (MeIQx) and 2-amino-23

3,4,8-trimethylimidazo[4,5-f]quinoxaline (4,8-DiMeIQx) (and its isomer 7,8-DiMeIQx),24

using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 25

2.5 cm lengths of porous polypropylene fibres impregnated with organic solvent to 26

facilitate simultaneous extraction from an alkaline aqueous sample into a low volume 27

acidic acceptor phase. This low cost protocol is extensively optimised for fibre length, 28

extraction time, sample pH and volume. Detection is by UPLC-MS/MS using positive 29

mode electrospray ionisation with a 3.4 min runtime, with optimum peak shape, 30

sensitivity and baseline separation being achieved at pH 9.5. To our knowledge this is 31

the first description of HCA chromatography under alkaline conditions. Application of 32

fixed ion ratio tolerances for confirmation of analyte identity is discussed. Assay 33

Page 4: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 2 of 28

Accep

ted

Man

uscr

ipt

2

precision is between 4.5 and 8.8% while lower limits of detection between 2 and 5 34

pg/mL are below the concentrations postulated for acid-labile HCA-protein adducts in 35

blood.36

37

38

Keywords:39

Heterocyclic aromatic amine, PhIP, MeIQx, Hollow fibre membrane liquid-phase 40

microextraction, Human plasma, UPLC-MS/MS.41

42

43

Page 5: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 3 of 28

Accep

ted

Man

uscr

ipt

3

1. Introduction43

44

Heterocyclic aromatic amines (HCAs) are formed during the cooking of proteinaceous 45

foods, particularly meat and fish, which provide creatin(in)e and other precursors such 46

as amino acids, sugars or other aldehydes [1]. Their formation in the parts per billion 47

concentration range is highly dependent upon the type of food and degree of cooking;48

therefore making estimation of dietary exposure to HCAs difficult [2]. The past 30 years 49

have seen extensive investigation into HCAs, in terms of their production, metabolism50

[3], formation of adducts with DNA [4] and protein [5], their quantification [6] and 51

implications for human health.52

53

2-Amino-1-methyl-6-phenylimidazo(4,5-b)pyridine (PhIP), 2-amino-3,8-54

dimethylimidazo[4,5-f]quinoxaline (MeIQx) and 2-amino-3,4,8-trimethylimidazo[4,5-55

f]quinoxaline (4,8-DiMeIQx) (Fig. 1) are three of the most abundant HCAs present in 56

cooked meat and fish [7,8] although many others have been identified [9]. There is 57

strong evidence from animal and in vitro studies of the carcinogenic and mutagenic58

properties of the HCAs although in vivo evidence correlating HCA dietary intake and 59

incidence of cancers can be equivocal [10,11]. To overcome the limitations of 60

estimating HCA intake by food frequency questionnaires, direct measurement of the 61

HCAs, their metabolites or their DNA or protein adducts in vivo is necessary. The 62

methodology of analysis of PhIP and its metabolites has been reviewed by Teunissen 63

and colleagues [12] who concluded that LC-MS/MS was clearly the detection method of 64

choice for sensitive qualitative and quantitative analyses of this most abundant of HCAs65

in biological matrices. Sample pretreatment for HCA analysis usually involves protein 66

precipitation, liquid-liquid extraction (LLE) or solid phase extraction (SPE).67

68

Hollow Fibre Membrane Liquid-Phase Microextraction (HF-LPME) techniques, which 69

employ porous membrane fibres to support an organic solvent during extraction of an 70

aqueous sample, were first introduced by Pedersen-Bjergaard and Rasmussen in 199971

[13] and have received considerable attention for analyses of environmental 72

contaminants and pharmaceuticals and related substances in body fluids, as reviewed 73

by Lee and colleagues [14]. More widespread adoption of HF-LPME is possible, 74

particularly in combination with LC-MS/MS detection techniques [14]. The application of 75

HF-LPME techniques to extraction of HCAs has been limited, yet the traditional 76

Page 6: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 4 of 28

Accep

ted

Man

uscr

ipt

4

extraction methods for HCAs (usually LLE and/or SPE) are prime candidates for 77

transfer to HF-LPME techniques. HF-LPME was first applied to extraction of PhIP from 78

urine and plasma by a group in Lund University, Sweden [15] who then expanded this 79

to eleven HCAs [16] and metabolites of PhIP in urine [17], proposing urinary PhIP as a 80

possible biomarker of exposure to dietary PhIP [18]. This extraction technique has also 81

been applied to the detection of HCAs in barbecued meats [19].82

83

Despite considerable research into the fate of HCAs in vivo, with much now focussing 84

on adducts to DNA, a reliable marker for HCA exposure (either circulating free or 85

adducted to blood proteins such as serum albumin or haemoglobin) remains elusive.86

There is a need for further validated HCA extraction techniques from blood products to 87

assist in this search.88

89

The three-phase HF-LPME system [14] described in the current study uses a porous 90

polypropylene hollow fibre impregnated with a small volume of organic solvent (the91

supported liquid membrane phase). An acidic aqueous acceptor phase fills the lumen of 92

the hollow fibre. The third phase is the alkaline aqueous sample (donor phase) 93

containing the weakly basic HCA analytes into which the fibre is immersed. Extraction is 94

by diffusion based on pH differences and is effectively a simultaneous double liquid-95

liquid extraction from alkaline sample to organic phase to acidic acceptor phase.96

97

Hollow fibres require preparation by the operator prior to use. We have noted that the 98

literature employing hollow fibres for extractions from low volume biological samples 99

does not always provide clear descriptions of the procedures involved. Therefore, in the 100

current study we have focussed on some of the detailed practicalities of preparing and 101

handling hollow fibres in addition to the optimisation and validation of the extraction 102

protocol and LC-MS/MS detection of HCAs in human plasma.103

104

2. Materials and methods105

106

2.1. Reagents and samples107

Reference standards 2-amino-1-methyl-6-phenylimidazo[4,5-b]pyridine (PhIP), 2-108

amino-3,8-dimethyl-imidazo [4,5-f]quinoxaline (MeIQx), 2-amino-3,4,8-trimethylimidazo109

[4,5-f]quinoxaline (4,8-DiMeIQx), 2-amino-3,7,8-trimethylimidazo [4,5-f]quinoxaline (7,8-110

Page 7: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 5 of 28

Accep

ted

Man

uscr

ipt

5

DiMeIQx) and deuterium labelled internal standards (I.S.) D3-PhIP, D3-MeIQx and D3-111

4,8-DiMeIQx were obtained from Toronto Research Chemicals (North York, ON, 112

Canada). Mixed standard solutions prepared in LCMS grade methanol were stored at 113

4°C. Unless stated, all other chemicals were obtained from Sigma-Aldrich (Dorset, UK).114

115

Blood samples, obtained from the research project FoodCAP funded by the World 116

Cancer Research Fund (Grant ID 2010/255), were from healthy volunteers who had 117

fasted for 10-12 h to ensure removal of free circulating HCAs from serum. This study 118

was conducted according to the guidelines laid down in the declaration of Helsinki and 119

all procedures involving human subjects were approved by the School of Medicine,120

Dentistry and Biomedical Sciences Research Ethics Committee of Queen’s University 121

Belfast. Each participant provided written, informed consent prior to blood donation.122

123

2.2. Apparatus124

Accurel® PP 300/1200 polypropylene hollow fibre membranes (300 μm wall thickness, 125

1200 μm inner diameter and 0.2 μm pore size, F-No-5129, manufactured by Membrana 126

GmbH, Wuppertal, Germany) were kindly donated by the Danish Meat Research 127

Institute (Roskilde, Denmark). Sample extractions were carried out in LC-GC certified 128

clear glass 2 ml HPLC microvials (12 x 32 mm) with screw neck caps containing 129

bonded pre-slit PTFE/silicone septa obtained from Waters Corporation (Manchester, 130

UK), and also used 18 gauge by 2.5 cm (0.8 mm needle outer diameter) syringe 131

needles (Sigma Aldrich) and 8 x 3 mm PTFE stirrer magnets (Scientific Laboratory 132

Supplies, Nottingham, UK).133

134

An Acquity I-class UPLC® binary pump and sample management system (Waters 135

Corporation, Manchester, UK) coupled to a Xevo TQS tandem mass spectrometer 136

(Waters Corporation), both controlled by MassLynx™ software, were used for sample 137

extract analysis. The mass spectrometer operated under positive electrospray 138

ionisation mode (ESI). Data acquisition was in Multiple Reaction Monitoring mode 139

(MRM) with a total run time of 3.4 min. Data analysis was performed using Waters 140

TargetLynx™ software. MS source settings were as follows: capillary voltage 0.5 kV,141

source temperature 150°C, desolvation temperature 650°C, cone nitrogen gas flow 150 142

L/h, desolvation nitrogen gas flow 1000 L/h. Separation of HCAs was carried out on an 143

Acquity BEH C18 1.7 m UPLC analytical column (50 x 2.1 mm) equipped with an in-144

Page 8: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 6 of 28

Accep

ted

Man

uscr

ipt

6

line filter unit (0.2 µm, 2.1 mm) (Waters Corporation), maintained at 40°C. A binary 145

gradient mobile phase was applied at a flow rate of 0.8 mL/min, phase A being 5 mM 146

ammonium formate pH 9.5 (aq) and phase B being acetonitrile. The rapid gradient 147

profile was: (1) 0-0.2 min, held at 93% A, (2) 0.2-2 min, falling linearly to 75% A, (3) 148

2.01-2.2 min, held at 70% A, (4) 2.25-2.75 min, held at 50% A, (5) 2.8-3.4 min, held at 149

93% A. The UPLC purge wash was acetonitrile:water (10:90, v/v) and the wash solution150

was acetonitrile:water (50:50, v/v). Injection volume was 5 µl. Table 1 provides details of 151

HCA fragmentation transitions and other MS/MS conditions.152

153

2.3. Preparation of hollow fibres154

The optimised conditions for extraction of HCAs from plasma using HF-LPME are 155

described below. Description of the method optimisation study follows in section 2.6. 156

Porous Accurel® PP 300/1200 polypropylene hollow fibre membranes were cut into 2.5157

cm lengths and one end was heat-sealed using a hot soldering iron – the fibre tip was158

touched lightly onto the hot iron surface for 1-2 sec and then immediately squeezed159

repeatedly using fine tipped metal tweezers to form a seal of approximately 3-4 mm160

length. The fibre was cleaned by soaking in acetone for 10 min before drying at 37°C161

for approximately 15 min (in a glass or paper container, not plastic, to avoid damage by 162

residual acetone). Sealed fibres were stored in a capped glass tube prior to use. If 163

condensation of acetone is evident during storage, fibres should be re-dried at 37°C. 164

Immediately prior to use, fibres were prepared for extraction by filling the lumen of the 165

hollow fibre with acidic acceptor phase and filling the pores of the fibre membrane with 166

organic solvent as follows. An 18 gauge hypodermic needle fitted to a 1 mL disposable 167

plastic syringe containing acidic acceptor solution (0.1 M sulphuric acid) was inserted 168

carefully into the unsealed end of the fibre ensuring the membrane was not punctured 169

and a strong seal was achieved (use of a fibre with different internal diameter would 170

require a different gauge needle). Acceptor solution was injected firmly into the lumen of 171

the fibre until droplets were clearly visible on the outer surface of the porous fibre and 172

no leakage was evident from the sealed end. The fibre was removed from the needle173

and a clean hypodermic needle protruding through the pre-slit septum of a microvial 174

screw cap was carefully inserted. Holding the needle Luer-Lok connector, the fibre was175

then dipped into 1-octanol for 30 sec, allowing the organic solvent to fill the membrane176

pores. Excess 1-octanol on the fibre surface was removed by manually shaking the 177

fibre in deionised 18Ω water for 30 sec. The prepared fibre could then be conveniently 178

Page 9: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 7 of 28

Accep

ted

Man

uscr

ipt

7

immersed in a sample vial, allowing the screw cap to be sealed and the height of the 179

fibre adjusted through the pre-slit septum without removing or touching the fibre. Each 2180

mL sample vial contained an 8 x 3 mm stirrer magnet and the needle height was 181

adjusted to avoid the stirrer damaging the sealed end of the fibre during extraction (Fig. 182

2).183

184

2.4. Sample extraction185

Plasma samples (0.2 mL) were placed in 2 mL glass HPLC microvials and fortified with 186

50 pg/mL internal standard (I.S.; 10 µl of 1 ng/mL mixed deuterium labelled HCAs D3-187

PhIP, D3-MeIQx and D3-4,8-DiMeIQx) for validation and routine analyses. For 188

optimisation of the method the I.S. were added to samples after extraction by addition 189

of 10 µl to the HPLC microvial insert containing the recovered acidic acceptor phase to 190

enable calculation of HCA extraction efficiency. For optimisation and validation of the 191

method, mixed standard HCAs were added to samples prior to extraction at 30 pg/mL192

(60 µl of 100 pg/mL mixed PhIP, MeIQx, 7,8-DiMeIQx and 4,8-DiMeIQx). Samples were 193

made alkaline by addition of 1.3 mL of 0.5 M NaOH and a 8 x 3 mm stirrer magnet was194

placed in each microvial. A prepared hollow fibre was immersed in each sample as 195

described above (Fig. 2) and vials were placed in a Perspex microvial rack on a single 196

position magnetic stirrer at room temperature for 5 h, stirring at 550 rpm. Use of a 197

microvial rack avoided the need for a multiple-position magnetic stirrer. After extraction 198

the fibre was removed from the sample with the needle still attached. The fibre sealed 199

end was cut off with sharp scissors and an air filled 1 mL disposable syringe used to 200

expel the acceptor phase (typically 20 µl from a 2.5 cm fibre) into a pre-weighed 200 µl 201

glass insert inside a 2 mL HPLC microvial. The recovered liquid was weighed and an 202

equal volume of 0.1 M NaOH was added to neutralise the acidic accepter phase. The 203

neutralised extract was then made up to a final volume of 100 µl by addition of 30 mM 204

ammonium formate (aq.) pH 9.5 to ensure compatibility with the LC-MS/MS mobile 205

phase.206

207

2.5. Calibration208

LC-MS/MS solvent calibration standards were prepared by addition of I.S. and 209

increasing volumes of 100 pg/mL mixed standard HCAs in microvials. Solvent was 210

evaporated to dryness under nitrogen and standards were reconstituted in 20 µl 211

Page 10: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 8 of 28

Accep

ted

Man

uscr

ipt

8

acetonitrile and 80 µl of 30 mM ammonium formate pH 9.5 before transferring to 200 µl 212

glass microvial inserts.213

214

2.6. Method optimisation215

The optimum conditions for extraction using HF-LPME are highly dependent upon the 216

target compound and the fibre being used. Each new extraction method requires217

specific optimisation to take account of the structure of the analytes and the gauge of 218

fibre employed. The critical variables assessed in this optimisation study were the 219

length of hollow fibre, the molarity of NaOH used to dilute the plasma sample, the final 220

volume of diluted sample and the extraction time. The four HCA analytes being 221

extracted from plasma included three compounds based on an imidazo-quinoxaline 222

skeleton (MeIQx, 4,7-DiMeIQx, 7,8-DiMeIQx) and one based on a phenyl-imidazo-223

pyridine skeleton (PhIP). Plasma from a single volunteer was used during method 224

optimisation. Plasma was fortified with HCAs prior to extraction under various 225

conditions (duplicate samples for each condition) and I.S. added after extraction.226

227

2.7. Method validation228

Validation of the optimised extraction method was carried out using plasma from a 229

single volunteer, seven aliquots being fortified with 30 pg/mL HCA mixed standards 230

and 50 pg/mL I.S. prior to extraction on each of 3 days. Extracted HCAs were quantified 231

against solvent calibration curves. Plasma fortified with I.S. only was included as a 232

negative control. Samples fortified with progressively lower HCA concentrations were 233

extracted in order to estimate limits of detection and quantification.234

235

236

3. Results and discussion237

238

Development of this HF-LPME technique demonstrated the convenience of leaving the 239

hollow fibre attached to a hypodermic needle during extraction, allowing easy handling 240

and recovery of acceptor phase when extracting a limited sample volume. The fibre 241

need not be touched after immersion in the sample as the needle Luer-Lok acts as a 242

convenient handle for the remainder of the procedure. Other authors have immobilised 243

the fibre on a wire during extraction [17], however this requires the fibre to be 244

transferred after extraction from the wire to a new needle for recovery of the acceptor 245

Page 11: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 9 of 28

Accep

ted

Man

uscr

ipt

9

phase, increasing the number of handling steps and the risk of cross-contamination. For 246

larger liquid sample volumes, a longer fibre may be used and both ends sealed before 247

complete immersion of the fibre in the sample without any support. This approach has 248

been used at the University of Seville for extraction of pharmaceuticals from urine and 249

environmental water samples [20,21]. The use of HPLC microvials with pre-slit septa 250

screw caps as sample containers proved to be a simple way to secure the fibres and 251

needles during extraction and a convenient way to adjust fibre depth in the sample. 252

Acceptor phase could even be recovered from the fibre without removing the needle 253

from the pre-slit septum. A little practice in preparation of the fibres, particularly the 254

sealing of the ends with a soldering iron, yielded consistent results, with less than 5% of 255

fibres leaking from the sealed end when filling with acceptor phase (these fibres were 256

discarded).257

258

3.1. Method optimisation259

260

3.1.1. LC-MS/MS optimisation261

Several sub-2 µm, octadecylsilyl (C18-based) UPLC columns manufactured by Waters 262

were assessed for the separation of the four HCAs (HSS T3, HSS, BEH C18, CSH C18263

and AccQTag Ultra) in addition to a Phenomenex Kinetex pentafluorophenyl (PFP, 2.6264

µm) column. All were assessed under acidic mobile phase A conditions ranging from 265

pH 3.5 to 6.4, while BEH C18 and Kinetex PFP were also assessed at alkaline pH 8.0 266

to 9.5. Notably, the BEH column yielded the best peak shape, sensitivity (peak 267

intensity) and baseline separation of the DiMeIQx isomers at pH 9.5 - the natural pH of 268

5 mM ammonium formate, avoiding the need for pH adjustment of mobile phase A (Fig. 269

3). Separation of HCAs has traditionally been achieved under acidic LC conditions 270

[1,12], on the principle that mobile phase pH should be lower than the analyte pKa (<pH 271

5 for the HCAs) in order to fully protonate the HCA amine groups prior to positive mode 272

electrospray ionisation. However, baseline chromatographic separation of the DiMeIQx 273

isomers is sensitive to pH and is incomplete under the commonly used pH 4.7 or lower 274

[22,16]. Holland and colleagues [23] unusually employed a mobile phase ranging from 275

pH 6.8 to 7.85 to separate HCAs, including 4,8-DiMeIQx, in hydrolysed urine but the 276

degree of chromatographic separation from 7,8-DiMeIQx was not described. In the 277

current study, use of pH 9.5 and the UPLC gradient described above (5mM aqueous 278

ammonium formate and acetonitrile) facilitated baseline separation of MeIQx (tR 1.07279

Page 12: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 10 of 28

Accep

ted

Man

uscr

ipt

10

min), 7,8-DiMeIQx (1.31 min), 4,8-DiMeIQx (1.38 min) and PhIP (2.21 min) with a total 280

gradient runtime of 3.4 min and typical peak widths of 3.4-3.8 sec (Fig. 4). Ammonium 281

formate was employed as an ion pairing agent, in keeping with Bianchi and colleagues282

[24] who demonstrated better HCA peak shapes with formate than with acetate. It may 283

be that investigators, on observing improved HCA peak shape as mobile phase pH was 284

lowered below pH 4.7, have not previously studied the benefits of LC conditions closer 285

to or higher than neutral. Nevertheless, as stated by Bianchi and colleagues [24], 286

“depending on the specific purpose and design of the experiment, fine adjustments for 287

pH and mobile phase concentration are always recommended to achieve optimal 288

separation of HCAs”. To our knowledge this is the first report of chromatographic 289

separation of HCAs under alkaline LC-MS/MS conditions. The benefits of alkaline 290

mobile phase conditions with positive mode ionisation are clearly compound dependent, 291

as shown by Gerssen and colleagues [25] who observed improved recovery from 292

shellfish of the marine toxin azaspiracid-1 using pH 11 LC conditions compared with pH293

2.6 under positive ESI, attributing this to altered matrix suppression effects, whilst other 294

toxins performed better under acidic conditions. Kipper and colleagues [26] also found 295

that optimum signal intensities and peak separation of several antibiotics under positive 296

ESI were achieved at pH 9.297

298

The choice of fragmentation transitions for identification of the isomers of DiMeIQx is 299

important. Both 4,8-DiMeIQx and 7,8-DiMeIQx share m/z 228>213 as their most intense 300

transition. In the absence of demonstrable chromatographic separation, studies which 301

use this peak for quantification of DiMeIQx (for example [23,27]) risk misidentification of 302

the isomers. It is advisable to use the m/z 213 fragment as the qualitative (confirmatory)303

ion and to use the less intense, but essentially unique, fragments m/z 212 and m/z 131 304

for reliable quantification of 4,8-DiMeIQx and 7,8-DiMeIQx respectively (Table 1 and 305

Fig. 4).306

307

3.1.2. Fibre length optimisation308

The bar charts in Fig. 5 illustrate the proportion of HCAs extracted under various 309

conditions from plasma fortified with 30 pg/ml mixed HCAs. Data are normalised to 310

percentages of the maximum HCA concentration extracted under the assessed 311

conditions. HF-LPME extraction conditions were as described above in sections 2.3 and 312

2.4, with each of the following four variables being independently optimised.313

Page 13: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 11 of 28

Accep

ted

Man

uscr

ipt

11

314

Fig. 5a demonstrates the influence of the length of hollow fibre used in a fixed sample 315

donor volume of 1.5 ml, with 0.5 M NaOH diluent and 5 h extraction time. A 1 cm length 316

of fibre clearly provided insufficient acidic acceptor phase for successful extraction (8-317

14 µl recovered) due to loss of lumen volume following sealing of the fibre and insertion 318

of the supporting needle. A 2.5 cm length was convenient for the 2 ml HPLC vials. 319

Using a pair of 2.5 cm fibres in a single sample vial yielded twice the volume of 320

acceptor phase following extraction (45 µl) but only a marginal increase in PhIP and 321

MeIQx recovery and required twice the time and consumable materials to prepare. 322

Longer fibres had a detrimental effect on MeIQx recovery and had to be folded once or 323

twice to be fully submerged in the sample, risking damage to the integrity of the 324

supported organic layer during extraction. Therefore, a single 2.5 cm fibre was used for 325

all further extractions.326

327

3.1.3. Donor NaOH molarity optimisation328

Fig. 5b demonstrates that HCA extraction efficiency by HF-LPME is influenced by the 329

molarity of NaOH used to create the alkaline sample conditions for extraction (0.2 ml 330

plasma sample was diluted with 1.3 ml NaOH). At least 0.5 M NaOH was required to 331

achieve maximum analyte recovery (1 M in the case of MeIQx) with extraction efficiency 332

dropping when more concentrated NaOH was used. This is in contrast to Busquets and 333

colleagues [17,18] who suggested that PhIP or its metabolites could be extracted from 334

urine using HF-LPME with greater signal to noise ratio when mildly acidic conditions 335

were employed (pH 5.5) compared with a donor pH greater than 10. However, this 336

observation was not consistent across Busquets’ studies. It is also in contradiction of 337

the conventional use of alkaline sample conditions for the extraction of a weakly basic 338

compound such as PhIP (pKa=5.6) into an organic solvent followed by back-extraction 339

into an acidic acceptor phase [15]. Furthermore, the current study shows that simply 340

raising sample pH above 10 may not be optimal for HCA extraction. Observed pH in 341

duplicate diluted plasma samples prior to extraction were as follows: 0.01 M NaOH (pH 342

11.0), 0.05 M (pH 12.5), 0.2 M (pH 13.2), 0.5 M NaOH (pH 13.5), 1 M (pH 13.6) and 2343

M (pH 13.7). It can be seen from Fig. 5b that even at sample pH 12.5 (0.05 M NaOH), 344

extraction of HCAs may still be less than 50% of maximum. A NaOH molarity of 0.5 M 345

(pH 13.5) was employed as a compromise for the optimal extraction of the quinoxaline 346

and pyridine HCAs. The use of 0.5 M NaOH is in agreement with the studies of HCA 347

Page 14: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 12 of 28

Accep

ted

Man

uscr

ipt

12

extraction from urine by HF-LPME emanating from Lund University, Sweden 348

[15,16,17,18]. However, low parts per billion concentrations of PhIP and 4,8-DiMeIQx 349

were also successfully extracted from cooked meats by HF-LPME using 0.05 M NaOH 350

[19], although extraction efficiency was not described.351

352

3.1.4. Donor volume optimisation353

Fig. 5c illustrates the influence of the final volume of a 0.2 ml plasma sample diluted 354

with 0.5 M NaOH prior to extraction using a 2.5 cm hollow fibre. Donor volume did not 355

influence the volume of acidic acceptor phase recovered (typically 22 µl from a 2.5 cm 356

fibre) but the dilution effect on the sample resulted in poorer extraction into the fixed 357

acceptor phase volume. A final donor volume of 1.5 ml was chosen as optimum for the 358

current system.359

360

3.1.5. Extraction time optimisation361

Fig. 5d illustrates how optimum extraction time using HF-LPME is dependent upon the 362

chemical characteristics of the analytes. PhIP (a phenyl-imidazo-pyridine) reached 363

maximum recovery after 5 h, while the three imidazo-quinoxaline compounds extracted 364

more slowly, with recovery still rising after 7 h. A compromise protocol of 5 h was 365

adopted to cover all analytes and to facilitate completion of a sample batch extraction 366

within a single working day.367

368

This 5 h extraction time is in keeping with Ramos Payan and colleagues [20] who 369

extracted fluoroquinolone antibiotics from bovine urine and environmental water 370

samples in a 5.5 h timescale using HF-LPME. The same research group similarly 371

extracted sulphonamide antibiotics from human urine in 6 h, but extraction of non-372

steroidal anti-inflammatory drugs was achieved within 20 min [21], illustrating again the 373

need to optimise HF-LPME techniques for each analyte of interest. A review of 374

environmental and bioanalytical applications of HF-LPME [14] also highlighted the 375

variable extraction times and recoveries achievable by this technique, depending upon 376

analyte chemistry and sample type. Extraction times were often 1 h or less, but 377

recoveries were consequently incomplete.378

379

380

381

Page 15: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 13 of 28

Accep

ted

Man

uscr

ipt

13

3.2. Method validation382

Solvent calibration curves were linear in the range equivalent to 5-80 pg/mL plasma383

(10-160 pg/mL reconstitution solvent, or 50-800 fg on-column), coefficients of 384

determination R2 typically being greater than 0.998 for all HCAs. The mean recovery 385

and precision of the HF-LPME extraction method are shown in Table 2. Data are based 386

on 21 replicates of human plasma fortified with 30 pg/mL HCAs prior to extraction (7 387

replicates extracted on each of 3 days). Mean observed HCA concentrations (recovery) 388

were greater than 92% of fortified concentrations for all four HCAs. Precision (RSD) 389

was below 9% in all cases, both within and between runs, demonstrating satisfactory 390

performance at physiologically relevant HCA-adduct concentrations in human blood [28,391

29]. Estimated lower limits of detection and quantification based on signal-to-noise 392

ratios greater than 3 and 10 respectively (using Peak-to-Peak calculations on 393

unsmoothed raw data) in plasma fortified with HCAs prior to extraction are shown in 394

Table 2. Limits are based on the lowest intensity peak for each analyte. Limits for PhIP 395

are higher than for the quinoxaline HCAs due primarily to the low intensity of the 396

secondary, confirmatory transition peak for PhIP (m/z 225.2>183.2). HF-LPME assay 397

sensitivity and performance are similar to those reported by Lezamiz and colleagues 398

[15] who studied only PhIP in plasma. 399

400

Confidence in the identification of analytes is important when measuring low 401

concentrations of compounds in complex biological matrices, particularly for analyte 402

groups like the HCAs which share common structures and fragmentation patterns. The 403

presence of a second transition peak to confirm analyte identity should be a 404

prerequisite, although this is not always the case in the published literature of HCAs. 405

Furthermore, in the present study the ratio of confirmatory to primary quantitation peaks406

(ion ratio) was monitored in every sample to ensure agreement with the same ratio in 407

calibration standards. Guidance on the tolerances to apply to compliance of ion ratios 408

with their standards was taken from the document laying out the required analytical 409

performance of methods in the veterinary pharmaceuticals field: Commission Decision410

2002/657 [30]. While this European Commission document sets out to ensure animal-411

derived food products are free of harmful residues and is not directly applicable to the 412

study of natural carcinogens such as the HCAs, it is the opinion of the authors that 413

similar performance criteria should be applied to analysis of suspected carcinogens 414

whenever possible. Consequently, we applied the tolerances as defined in Decision 415

Page 16: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 14 of 28

Accep

ted

Man

uscr

ipt

14

2002/657 to the ion ratios in all samples: if ion ratios did not agree with the same ratios 416

in calibration standards to within 20 to 30% (dependent on the magnitude of the ratio), 417

the identity of the analyte peak could not be confirmed. Ion ratios greater than 0.5 are 418

permitted a tolerance of ±20% relative to calibration standards, 25% tolerance is 419

applied for ratios between 0.2 and 0.5, 30% for ratios of 0.1-0.2, and 50% for ratios 420

below 0.1. For example, the ratio of the peak areas of MeIQx daughter ions 131.1/199.1 421

(Table 1) was typically around 0.4, permitting this ratio in a sample to be within 25% of 422

the mean ratio of the same peaks in calibration standards.423

424

3.3. Application to blood samples425

Preliminary studies demonstrated that the HF-LPME plasma extraction method may 426

also be applied to serum and whole blood, although whole blood matrix effects caused 427

an additional 30-40% signal suppression and quantitative accuracy was adversely 428

affected, with I.S.-corrected recoveries being around 85% for the quinoxaline HCAs and 429

around 72% for PhIP (data not shown), compared with 93-99% and 92% respectively in 430

plasma (Table 2).431

432

Human plasma samples were obtained from the World Cancer Research Fund project 433

FoodCAP. Samples were from volunteers whose dietary HCA intake was estimated434

using WISP nutritional analysis software on the basis of 7 day food diaries and the US 435

National Cancer Institute’s CHARRED database436

(http://dceg.cancer.gov/tools/design/charred). Volunteers fasted for 10-12 h before 437

providing blood samples to ensure removal of free circulating HCAs from their plasma. 438

Unsurprisingly, no HCAs were detected in these samples using the optimised HF-LPME439

extraction method, even from volunteers with a nominally high HCA intake of greater 440

than 1 µg/day.441

442

Adducts of HCAs bound covalently to blood proteins such as haemoglobin and serum 443

albumin have, since the 1990s, been proposed as potential biomarkers of exposure to 444

HCAs (as reviewed by [2,31]). However, the suitability of protein adducts as reliable 445

indicators of potential carcinogenic damage by HCAs is still unproven [31].446

Furthermore, Magagnotti and colleagues [29] detected PhIP released by mild acid 447

hydrolysis from purified haemoglobin and serum albumin, demonstrating differences 448

between meat consumers and vegetarians in the range 7-67 pg/mL for PhIP-albumin 449

Page 17: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 15 of 28

Accep

ted

Man

uscr

ipt

15

adducts. However, this work has not been replicated since, despite ongoing studies in 450

the field.451

452

The current HF-LPME method was applied to plasma samples from high dietary HCA 453

intake volunteers in an attempt to measure protein-bound HCAs following their release 454

by the acid hydrolysis protocol of Magagnotti and colleagues [29]. Plasma was diluted 455

to a final concentration of 0.1 M hydrochloric acid and incubated at 80°C for 1 h before 456

adjusting the pH with NaOH and extracting by HF-LPME as described above. No acid-457

labile HCAs were detected under these conditions, demonstrating the need for further 458

work on the release of labile HCA adducts from blood proteins.459

460

461

4. Conclusions462

463

The described HF-LPME method presents a convenient and low cost technique for 464

extraction of carcinogenic heterocyclic aromatic amines from plasma. LC-MS/MS assay 465

sensitivity is in the range postulated for acid-labile HCA-protein adducts in blood [29], 466

with satisfactory assay precision and recovery. The hollow fibre membrane liquid-phase467

microextraction technique benefits from very low organic solvent usage compared with 468

conventional liquid-liquid extraction (LLE) and low cost compared with solid phase 469

extraction methods (SPE). To our knowledge this is the first description of a validated 470

HF-LME protocol for extraction of a group of HCAs from plasma and also of their 471

chromatographic separation under alkaline conditions.472

473

Extensive optimisation of the HF-LPME protocol demonstrates the need for such 474

microextraction techniques to be carefully optimised for each analyte of interest. A final 475

compromise protocol is described, balancing the variable recoveries between different 476

analytes and enabling a batch of approximately 20 samples to be completed in a 477

normal working day. It should be noted that samples can be left unattended during the 478

lengthy HF-LPME extraction step, freeing the operator for other duties, unlike manual 479

LLE and SPE techniques. This technique represents an additional, low cost extraction 480

tool in the ongoing search for biomarkers of exposure to carcinogenic heterocyclic481

aromatic amines in plasma.482

483

Page 18: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 16 of 28

Accep

ted

Man

uscr

ipt

16

484

Acknowledgments485

486

This research was part-funded by the World Cancer Research Fund UK, Grant ID 487

2010/255. Grateful thanks are expressed to Dr Kirsten Jensen of the Danish Meat 488

Research Institute, Roskilde for the kind gift of the hollow fibre membranes, and to Dr 489

Sarah Brennan and Prof Jayne Woodside of Queen’s University Belfast and all 490

participants in the FoodCAP project for blood donations.491

492

493

References494

[1] K. Cheng, F. Chen, M. Wang, Heterocyclic amines: Chemistry and health, Mol. Nutr. Food 495Res. 50 (2006) 1150-1170. 496

[2] J. Alexander, R. Reistad, S. Hegstad, H. Frandsen, K. Ingebrigtsen, J.E. Paulsen, G. 497Becher, Biomarkers of exposure to heterocyclic amines: approaches to improve the exposure 498assessment. Food and Chemical Toxicology, Food Chem. Toxicol. 40 (2002) 1131-1137. 499

[3] R. Turesky, F. Guengerich, A. Guillouzo, S. Langouet, Metabolism of heterocyclic aromatic 500amines by human hepatocytes and cytochrome P4501A2, Mutat. Res. Fundam. Mol. Mech. 501Mutag. 506 (2002) 187-195. 502

[4] H. Schut, E. Snyderwine, DNA adducts of heterocyclic amine food mutagens: implications 503for mutagenesis and carcinogenesis, Carcinogenesis 20 (1999) 353-368. 504

[5] L. Peng, S. Dasari, D.L. Tabb, R.J. Turesky, Mapping serum albumin adducts of the food-505borne carcinogen 2-amino-1-methyl-6-phenylimidazo[4,5- b ]pyridine by data-dependent 506tandem mass spectrometry, Chem. Res. Toxicol. 25 (2012) 2179-2193. 507

[6] M. Murkovic, Analysis of heterocyclic aromatic amines, Anal. Bioanal. Chem. 389 (2007) 508139-146. 509

[7] A. Lynch, S. Murray, N. Gooderham, A. Boobis, Exposure to and activation of dietary 510heterocyclic amines in humans, Crit. Rev. Oncol. 21 (1995) 19-31. 511

[8] Y. Zhang, C. Yu, J. Mei, S. Wang, Formation and mitigation of heterocyclic aromatic amines 512in fried pork, Food Addit. Contam. Part A Chem. Anal. Control Expo. Risk Assess. 30 (2013) 5131501-1507. 514

[9] W. Ni, L. McNaughton, D.M. LeMaster, R. Sinha, R.J. Turesky, Quantitation of 13 515heterocyclic aromatic amines in cooked beef, pork, and chicken by liquid chromatography-516electrospray ionization/tandem mass spectrometry, J. Agric. Food Chem. 56 (2008) 68-78. 517

[10] T. Sugimura, K. Wakabayashi, H. Nakagama, M. Nagao, Heterocyclic amines: 518Mutagens/carcinogens produced during cooking of meat and fish, Cancer Sci. 95 (2004) 290-519299. 520

Page 19: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 17 of 28

Accep

ted

Man

uscr

ipt

17

[11] D. Tang, O.N. Kryvenko, Y. Wang, S. Trudeau, A. Rundle, S. Takahashi, T. Shirai, B.A. 521Rybicki, 2-Amino-1-methyl-6-phenylimidazo[4,5-b]pyridine (PhIP)-DNA adducts in benign 522prostate and subsequent risk for prostate cancer, Int. J. Cancer 133 (2013) 961-971. 523

[12] S.F. Teunissen, H. Rosing, A.H. Schinkel, J.H.M. Schellens, J.H. Beijnen, Review on the 524analysis of 2-amino-1-methyl-6-phenylimidazo[4,5-b]pyridine and its phase I and phase II 525metabolites in biological matrices, foodstuff and beverages, J. Chromatogr. B Analyt. Technol. 526Biomed. Life Sci. 878 (2010) 3199-3216. 527

[13] S. Pedersen-Bjergaard, K.E. Rasmussen, Liquid-phase microextraction with porous hollow 528fibers, a miniaturized and highly flexible format for liquid-liquid extraction, J. Chromatogr. A 5291184 (2008) 132-142. 530

[14] J. Lee, H.K. Lee, K.E. Rasmussen, S. Pedersen-Bjergaard, Environmental and 531bioanalytical applications of hollow fiber membrane liquid-phase microextraction: A review,532Anal. Chim. Acta 624 (2008) 253-268. 533

[15] J. Lezamiz, T. Barri, J.Å. Jönsson, K. Skog, A simplified hollow-fibre supported liquid 534membrane extraction method for quantification of 2-amino-1-methyl-6-phenylimidazo[4,5-535b]pyridine (PhIP) in urine and plasma samples, Anal. Bioanal. Chem. 390 (2008) 689-696. 536

[16] F.U. Shah, T. Barri, J.A. Jönsson, K. Skog, Determination of heterocyclic aromatic amines 537in human urine by using hollow-fibre supported liquid membrane extraction and liquid 538chromatography-ultraviolet detection system, J. Chromatogr. B Analyt. Technol. Biomed. Life 539Sci. 870 (2008) 203-208. 540

[17] R. Busquets, J.Å. Jönsson, H. Frandsen, L. Puignou, M.T. Galceran, K. Skog, Hollow fibre-541supported liquid membrane extraction and LC-MS/MS detection for the analysis of heterocyclic 542amines in urine samples, Mol. Nutr. Food Res. 53 (2009) 1496-1504. 543

[18] R. Busquets, H. Frandsen, J.A. Jönsson, L. Puignou, M.T. Galceran, K. Skog, 544Biomonitoring of dietary heterocyclic amines and metabolites in urine by liquid phase 545microextraction: 2-amino-1-methyl-6-phenylimidazo[4,5-b]pyridine (PhIP), a possible biomarker 546of exposure to dietary PhIP, Chem. Res. Toxicol. 26 (2013) 233-240. 547

[19] M.D. Aaslyng, L. Duedahl-Olesen, K. Jensen, L. Meinert, Content of heterocyclic amines 548and polycyclic aromatic hydrocarbons in pork, beef and chicken barbecued at home by Danish 549consumers, Meat Sci. 93 (2013) 85-91. 550

[20] M. Ramos Payán, M.Á. Bello López, R. Fernández-Torres, J.A.O. González, M. Callejón 551Mochón, Hollow fiber-based liquid phase microextraction (HF-LPME) as a new approach for the 552HPLC determination of fluoroquinolones in biological and environmental matrices, J. Pharm. 553Biomed. Anal. 55 (2011) 332-341. 554

[21] M. Villar Navarro, M. Ramos Payán, R. Fernández-Torres, M.A. Bello López, Hollow fiber 555liquid-phase microextraction and determination of nonsteroidal anti-inflammatories by capillary 556electrophoresis and sulfonamides by HPLC in human urine, Biomed. Chromatogr. 27 (2013) 557246-253. 558

[22] E. Barcelo-Barrachina, E. Moyano, M.T. Galceran, J.L. Lliberia, B. Bago, M.A. Cortes, 559Ultra-performance liquid chromatography-tandem mass spectrometry for the analysis of 560heterocyclic amines in food, J. Chromatogr. A 1125 (2006) 195-203. 561

Page 20: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 18 of 28

Accep

ted

Man

uscr

ipt

18

[23] R. Holland, J. Taylor, L. Schoenbachler, R. Jones, J. Freeman, T. Miller, B. Lake, N. 562Gooderham, R. Turesky, Rapid biomonitoring of heterocyclic aromatic amines in human urine 563by tandem solvent solid phase extraction liquid chromatography electrospray ionization mass 564spectrometry, Chem. Res. Toxicol. 17 (2004) 1121-1136. 565

[24] F. Bianchi, A. Careri, C. Corradini, L. Elviri, A. Mangia, I. Zagnoni, Investigation of the 566separation of heterocyclic aromatic amines by reversed phase ion-pair liquid chromatography 567coupled with tandem mass spectrometry: The role of ion pair reagents on LC-MS/MS sensitivity,568J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 825 (2005) 193-200. 569

[25] A. Gerssen, M.A. McElhinney, P.P.J. Mulder, R. Bire, P. Hess, J. de Boer, Solid phase 570extraction for removal of matrix effects in lipophilic marine toxin analysis by liquid 571chromatography-tandem mass spectrometry, Anal. Bioanal. Chem. 394 (2009) 1213-1226. 572

[26] K. Kipper, K. Herodes, I. Leito, L. Nei, Two fluoroalcohols as components of basic buffers 573for liquid chromatography electrospray ionization mass spectrometric determination of antibiotic 574residues, Analyst 136 (2011) 4587-4594. 575

[27] Y. Zhang, C. Lin, G. Fang, J. Mei, X. Wang, S. Wang, Tandem solid phase extraction 576coupled to LC-ESI-MS/MS for the accurate simultaneous determination of five heterocyclic 577aromatic amines in processed meat products, Eur. Food Res. Technol. 234 (2012) 197-205. 578

[28] K.H. Dingley, K.D. Curtis, S. Nowell, J.S. Felton, N.P. Lang, K.W. Turtletaub, Cancer 579Epidemiol. Biomark. Prev. 8 (1999) 507-512.580

[29] C. Magagnotti, F. Orsi, R. Bagnati, N. Celli, D. Rotilio, R. Fanelli, L. Airoldi, Int. J. Cancer 58188 (2000) 1-6.582

[30] European Commission, Commission Decision 2002/657/EC of 12 August 2002 583implementing Council Directive 96/23/EC concerning the performance of analytical 584methods and the interpretation of results, Off. J. Eur. Comm. L221 (2002) 8-36.585

[31] R.J. Turesky, L. Le Marchand, Metabolism and biomarkers of heterocyclic aromatic amines 586in molecular epidemiology studies: Lessons learned from aromatic amines, Chem. Res. Toxicol. 58724 (2011) 1169-1214. 588

589

Page 21: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 19 of 28

Accep

ted

Man

uscr

ipt

19

589FIGURE CAPTIONS590

591

Fig. 1. Structures of heterocyclic aromatic amines.592

593

594Fig. 2. Format of hollow fibre membrane liquid-phase microextraction vials (2.5 cm 595

sealed fibre, 18 gauge needle, pre-slit septum cap and 1.5 mL diluted sample with 8 596

mm magnetic stirrer in 2ml microvial).597

598

599Fig. 3. ESI-UPLC-MS/MS MRM chromatograms of primary quantitation peaks of 4,8-600

DiMeIQx and 7,8-DiMeIQx (eluting first from a 5cm BEH C18 UPLC column) illustrating 601

the beneficial effects on peak shape, sensitivity and baseline separation of increasing 602

mobile phase pH. Analyte m/z transitions and peak heights are listed.603

604

605Fig. 4. ESI-UPLC-MS/MS MRM chromatograms of primary quantitation peaks of 606

heterocyclic aromatic amines and their deuterated internal standards extracted by HF-607

LPME from human plasma fortified at 30 pg/mL (I.S. at 50 pg/mL). Baseline separation 608

of DiMeIQx isomers is achieved at pH 9.5. Analyte names, m/z transitions and peak 609

heights are listed.610

611

612Fig. 5. Optimisation of the extraction of four heterocyclic aromatic amines from plasma613

by hollow fibre membrane liquid-phase microextraction. Data are means of duplicate 614

extractions of plasma fortified at 30 pg/mL. Four HF-LPME variables were optimised -615

A: length of hollow fibre, B: molarity of NaOH diluent, C: donor (final sample) volume,616

and D: extraction time.617

618

619

620

Page 22: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 20 of 28

Accep

ted

Man

uscr

ipt

Optimised extraction of heterocyclic aromatic amines from blood using hollow fibre

membrane liquid-phase microextraction and triple quadrupole mass spectrometry

Kevin M. Cooper*, Natcha Jankhaikhot, Geraldine Cuskelly

Highlights

A Hollow Fibre Liquid Microextraction technique for heterocyclic amines is

described.

HF-LPME extraction of carcinogenic HCAs from plasma is extensively

optimised.

Optimum UPLC-MS/MS chromatography and positive mode ES ionisation

achieved at pH 9.5.

*Highlights Cooper

Page 24: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 22 of 28

Accep

ted

Man

uscr

ipt

Figure 2 Cooper colour version

Page 25: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 23 of 28

Accep

ted

Man

uscr

ipt

Figure 2 Cooper b&w version

Page 26: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 24 of 28

Accep

ted

Man

uscr

ipt

Time1.90 1.95 2.00 2.05 2.10 2.15 2.20 2.25 2.30

%

0

100

1.90 1.95 2.00 2.05 2.10 2.15 2.20 2.25 2.30

%

0

100

Time1.95 2.00 2.05 2.10 2.15 2.20 2.25 2.30 2.35

%

0

100

1.95 2.00 2.05 2.10 2.15 2.20 2.25 2.30 2.35%

0

100

Time1.90 1.95 2.00 2.05 2.10 2.15 2.20 2.25 2.30

%

0

100

1.90 1.95 2.00 2.05 2.10 2.15 2.20 2.25 2.30

%

0

100

Time1.85 1.90 1.95 2.00 2.05 2.10 2.15 2.20 2.25

%

0

100

1.85 1.90 1.95 2.00 2.05 2.10 2.15 2.20 2.25

%

0

100

Time1.70 1.75 1.80 1.85 1.90 1.95 2.00 2.05 2.10

%

0

100

1.70 1.75 1.80 1.85 1.90 1.95 2.00 2.05 2.10

%

0

100

pH 8.0 pH 6.4 pH 9.5 pH 4.7 pH 5.7 3.6e5 2.0e5 7.1e4 4.0e4 2.9e4

7,8-DiMeIQx 228.2>131.1

4,8-DiMeIQx

228.2>212.0

3.9e5 1.9e5 4.6e4 8.8e4 3.1e4

Figure 3 Cooper

Page 27: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 25 of 28

Accep

ted

Man

uscr

ipt

Time1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

100

1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

100

1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

100

1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

100

1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

100

1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

100

1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

%

0

1001.38

1.38

1.31

2.21

2.21

1.06

1.07

D3-4,8-DiMeIQx 231.23 > 213.15 5.8e5

4,8-DiMeIQx 228.17 > 213.03 1.8e5

7,8-DiMeIQx 228.17 > 131.08 2.2e5

D3-PhIP 228.16 > 210.14 2.3e6

PhIP 225.16 > 210.12 7.0e5

D3-MeIQx 217.14 > 199.18 5.3e5

MeIQx 214.17 > 199.13 2.7e5

Figure 4 Cooper

Page 29: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 27 of 28

Accep

ted

Man

uscr

ipt

Table 1

Heterocyclic aromatic amine UPLC-MS/MS fragmentation conditions.

Analyte tR (min)

Primary

transition (m/z)

Confirmatory

transition (m/z)

Collision energy

(V)

PhIP 2.21 225.2>210.1 225.2>183.2 26 / 28

D3-PhIP 2.21 228.2>210.1 - 28

MeIQx 1.07 214.2>199.1 214.2>131.1 24 / 36

D3-MeIQx 1.07 217.1>199.2 - 26

7,8-DiMeIQx 1.31 228.2>131.1 228.2>213.1 36 / 24

4,8-DiMeIQx 1.38 228.2>212.0 228.2>213.1 35 / 26

D3-4,8-DiMeIQx 1.38 231.2>213.1 - 24

Table 1 Cooper

Page 30: Optimised extraction of heterocyclic aromatic amines from ...25 using Hollow Fibre Membrane Liquid-Phase Microextraction. This technique employs 26 2.5 cm lengths of porous polypropylene

Page 28 of 28

Accep

ted

Man

uscr

ipt

Table 2

Validation of HCA extraction by HF-LPME from plasma fortified with 30 pg/mL HCAs (n=21; 7 replicates on 3 days).

PhIP MeIQx 7,8-DiMeiQx 4,8-DiMeIQx

Mean concentration (pg/mL) 27.6 29.8 27.8 28.4

Mean recovery (%)a 92.0 99.4 92.5 94.6

Mean within day RSD (%) 5.4 4.5 6.4 7.1

Between day RSD (%) 7.5 4.6 7.7 8.8

Limit of detection (pg/mL) 5 3 2 2

Limit of quantification (pg/mL) 15 10 7 7

a I.S.-corrected recovery

Table 2 Cooper