optimisation and validation of pcr method for hla gene
TRANSCRIPT
Linköping University | Department of Physics, Chemistry and Biology
Bachelor’s Thesis, 16 hp | Educational Program: Chemical Biology
Spring term 2020 | LITH-IFM-G-EX—20/3892--SE
Optimisation and Validation of
PCR Method for HLA Gene
Expression to Enable PCR System
Transfer and Master Mix Change
Paulina Odlander
Examiner: Lars-Göran Mårtensson
Supervisor: Josephine Overmyr, Dynamic Code AB
i
Abstract
98% - 99,85% of patients diagnosed with celiac disease, gluten intolerance, carry either human
leukocyte antigen (HLA) genes DQ2 or DQ8, which is why detection of these serotypes is the
most relevant method in determining celiac disease risk or diagnosis. If suffering from celiac
disease, getting diagnosed and being put on a gluten restrictive or gluten-free diet is a health
profit. Not doing so can lead to deteriorating absorption of nutrients in the intestine leading to
malnutrition, and also long-term complications such as cancer, infertility, autoimmunity and
weakening of the bone tissue.
Health Tech company Dynamic Code AB provides a PCR test for determination of HLA DQ-
genes connected to development of celiac disease. The PCR method is probe based and in real
time and is at this time carried through on the, somewhat outdated, PCR instrument from
Thermo Fisher/Applied Biosystems called 7300 Real-Time PCR System. The run time for this
analysis on the instrument is 1 hour and 50 minutes. The Master Mix in use is TaqMan™ Gene
Expression Master Mix, from the same manufacturer.
Moving on to a more modern PCR instrument is a natural step for the company and is favourable
in several regards, one of them being the run time that will be cut by 50 minutes, allowing for
more samples to be analysed in the same amount of time. The objective is to move the HLA
analysis to Thermo Fisher’s QuantStudio™ 6 and 7 Flex Systems and at the same time change
the Master Mix to SolisFast® Probe qPCR Mix (Purple) from Solis BioDyne, in order to
achieve better accuracy as this Master Mix is more compatible with the latter instrument, along
with reducing reagent cost as it is less expensive.
In order to find the optimal primer and probe concentration for each target included in the HLA
analysis, their concentrations were varied and tested with the new Master Mix on the new
instrument.
PCR instrument transfer and Master Mix change was successful and validation experiments
showed a 98,9% accuracy for the new method compared to the original method.
ii
iii
Abbreviations
ABI 7300 or 7300 – Applied Biosystems 7300 Real-Time PCR System
HLA – Human leukocyte antibody
IPC – Internal positive control
PCR – Polymerase chain reaction
QS – QuantStudio™ 6 or 7 Flex System
Solis - SolisFAST® Probe qPCR Mix (Purple)
TaqMan - TaqMan™ Gene Expression Master Mix
iv
Table of Contents 1. Introduction ..................................................................................................................................... 1
1.1. Goal ......................................................................................................................................... 2
1.2. Purpose .................................................................................................................................... 2
2. Theoretical Background .................................................................................................................. 3
2.1. Celiac Disease ......................................................................................................................... 3
2.1.1. Human Leukocyte Antibody ............................................................................................... 3
2.2. Polymerase Chain Reaction ..................................................................................................... 5
2.2.1. The Components .................................................................................................................. 6
2.3. Internal Positive Control ......................................................................................................... 8
2.4. The Reaction............................................................................................................................ 8
2.4.1. Initial Denaturation .............................................................................................................. 9
2.4.2. Replication step ................................................................................................................... 9
2.4.3. Final extension step ........................................................................................................... 10
2.5. Amplification Plot ................................................................................................................. 10
2.6. Ct-value .................................................................................................................................. 11
2.7. Passive Reference Dye, Reporter Dye and Master Mix: Changes on Ct-values ................... 11
3. Methodology ................................................................................................................................. 13
3.1. DNA Samples ........................................................................................................................ 13
3.2. Negative and Positive Controls ............................................................................................. 13
3.2.1. Negative Controls .............................................................................................................. 13
3.2.2. Positive Controls ............................................................................................................... 13
3.3. Reaction Components ............................................................................................................ 14
3.3.2. PCR-mix: TaqMan™ Gene Expression Master Mix ........................................................ 15
3.3.3. PCR-mix: SolisFast® Probe qPCR Mix (Purple) .............................................................. 16
3.4. Oligonucleotide Mixes .......................................................................................................... 17
3.5. PCR Optimisation and Validation ......................................................................................... 18
4. Method .......................................................................................................................................... 20
4.1. Optimisation of Primers and Probes ...................................................................................... 20
4.2. Validation of Modified PCR Method .................................................................................... 20
5. Results ........................................................................................................................................... 21
5.1. Mix 1: DQA1_05................................................................................................................... 21
5.1.1. Mix 1: IPC ......................................................................................................................... 22
5.1.2. Mix 1: Probe 1 ................................................................................................................... 23
5.1.3. Mix 1: Probe 2 ................................................................................................................... 23
v
5.2. Mix 2: DQA1_02................................................................................................................... 24
5.2.1. Mix 2: DQB1_02 ............................................................................................................... 24
5.2.2. Mix 2: Probe 1 ................................................................................................................... 25
5.2.3. Mix 2: Probe 2 ................................................................................................................... 25
5.3. Mix 3: DQA1_03................................................................................................................... 26
5.3.2. Mix 3: Probe 1 ................................................................................................................... 27
5.3.3. Mix 3: Probe 2 ................................................................................................................... 28
5.4. Mix 4: DQB1*0302 B ........................................................................................................... 28
5.4.1. Mix 4: IPC ......................................................................................................................... 29
5.4.2. Mix 4: Probe 1 ................................................................................................................... 29
5.4.3. Mix 4: Probe 2 ................................................................................................................... 29
5.5. Validation .............................................................................................................................. 30
5.5.1. Mix 1 ................................................................................................................................. 30
5.5.2. Mix 2 ................................................................................................................................. 30
5.5.3. Mix 3 ................................................................................................................................. 30
5.5.4. Mix 4 ................................................................................................................................. 30
5.6. Final Primer and Probe Concentrations ................................................................................. 31
5.7. Final Affirmation of Genotypes ............................................................................................ 32
6. Discussion ..................................................................................................................................... 36
7. Conclusion ..................................................................................................................................... 38
8. Acknowledgments ......................................................................................................................... 39
9. References ..................................................................................................................................... 40
Appendix A ........................................................................................................................................... 42
1. PCR-plate Setups ....................................................................................................................... 42
Appendix B ........................................................................................................................................... 44
1. PCR Amplification Plots – Optimisation experiments .............................................................. 44
2. PCR Amplification Plots – Validation experiments .................................................................. 50
1
1. Introduction
The polymerase chain reaction (PCR) technique is still the most fundamental strategy being
used for gene amplification by scientists all over the world. The technique enables detection of
foreign pathogens e.g. virus and bacteria in biological material, such as human cells. PCR can
also be used to perform genotyping, which refers to the process of determining differences in
genes. Both gene amplification and genotyping are used for medical purposes in detection of
infectious diseases or when determining risk factors for development of genetically caused
disorders.
One genetically caused disorder in humans are in focus in this project - the risk for
genetically disposed coeliac disease. Through genotyping of the specific genes playing the big
role in development of the disease, it is possible to determine the potential risk for a specific
individual to develop it.
Besides the PCR instrument itself, one of the central reagents used in a PCR is the so-called
Master Mix, in which samples are placed, which contains some of the crucial components
needed to perform the reaction.
The health-tech company Dynamic Code AB performs PCR tests for determining risk for
coeliac disease on the 7300 Real-Time PCR System from Applied Biosystems (ABI 7300) which
is a part of ThermoFisher Scientific, a PCR instrument that has been on the market for many
years. Although the instrument still is reliable, switching to the more modern instruments
QuantStudio™ 6 and 7 Flex System (QS) from Applied Biosystems, run time for each analysis
can be cut almost in half, which means almost double the number of samples can analysed in
the same amount of time. By changing the instrument to the QS, it also becomes possible to
scale up the production thanks to the interchangeable well formats, from 96-well to 384-well,
which is not possible for the ABI 7300, an upscale which is relevant for Dynamic Code AB.
Besides this, the QS, compared to the ABI 7300, has a greater sensitivity in detecting small
changes in reactions and overall well-to-well and instrument-to-instrument data is improved.
Changing the PCR instrument will greatly improve the operation at Dynamic Code.
The two methods will be referred to as the ‘original method’, the one currently in use by the
company, carried through on the ABI 7300, and the ‘new method’ or ‘optimised method’, which
is going to be the optimised method, carried through on the QS.
Data obtained from a PCR can be greatly influenced by the Master Mix used in terms of
sensitivity, specificity, accuracy and dynamic range, and specific Master Mixes are optimised
for usage with specific instruments to achieve optimal accuracy in test results. The TaqMan™
Gene Expression Master Mix from Thermo Fisher (TaqMan) is optimised for and
recommended to use with ABI 7300 and SolisFAST® Probe qPCR Mix (Solis) from Solis
Biodyne has compatibility with the QS system. Thus, changing the Master Mix along with the
instrument transfer is essential to attain accurate and precise results. The new Master Mix is, in
addition to this, more affordable which together with the compatibility aspect makes the enzyme
change an important matter for the company.
2
1.1. Goal
The goal of the project is to successfully transfer analysis for detecting human leukocyte
antibody (HLA) genes, from an old PCR instrument to a latter one, changing the Master Mix
and optimising the method through adjustments of parameters, in specific, primer and probe
concentrations. The method will be validated, and results should be equal or at least comparable
to the original method.
The objective is to not have the Ct-values generated from the analysis with the new method
differ too far off from the corresponding Ct-values generated from the original method, so that
patient diagnosis based on Ct-values does not have to be re-evaluated and changed.
1.2. Purpose
The purpose of the project is to streamline the testing procedure by cutting down the time for
which a PCR run requires as well as cutting down on reagent cost. Moving on from the outdated
PCR machine to the more relevant version is also eagerly awaited and a natural transition to
make.
Table 1: Planned changes.
Change from Change to
PCR instrument Applied Biosystems 7300
Real-Time PCR System
(Thermo Fisher)
QuantStudio™ 6 and 7 Flex
Systems (Thermo Fisher)
Run time 1 hour 50 minutes 1 hour
Master Mix TaqMan™ Gene Expression
Master Mix from Thermo
Fisher
SolisFast® Probe qPCR Mix
(Purple) from Solis BioDyne
Passive Reference Dye ROX Mustang purple
DNA polymerase AmpliTaq Gold DNA
Polymerase
SolisFAST® DNA Polymerase
Probe used TaqMan® TaqMan®
Instrument compatibility 7300 System QuantStudio™
3
2. Theoretical Background
2.1. Celiac Disease
Celiac disease is a chronic autoimmune disease that causes ingestion of gluten protein to start
an inflammation that damages the intestinal tract of the small intestine, which can lead to
deteriorating absorption of nutrients in the intestine leading to malnutrition, and also long term
complications such as cancer, infertility, autoimmunity and weakening of the bone tissue
(Megiorni et al. 2009). Since celiac disease is a genetic condition, the spread of individuals
carrying the genetic variation that causes celiac disease is uneven distributed throughout the
world, although it exists on all continents (Lindfors et al. 2019) with an average occurrence of
approximately 1%-2% in Europe (Catassi et al. 2007). Although the disease is common, a vast
majority of genetically predisposed individuals do remain undiagnosed, ostensibly due to a
vague symptom picture or a complete lack of symptoms (Megiorni et al. 2008). There is a health
profit in getting diagnosed and being put on a gluten restrictive or gluten-free diet, which to this
date is the only effective treatment (Tack et al. 2010), which is why Dynamic Code is providing
a test for detection of celiac disease.
The test detects the presence of DQ2.5-, DQ2.2- and DQ8-alleles (genetic predisposition),
by gene amplification through PCR. Specific antibodies for tTG-IgA and total-IgA in the
sample (to rule out igA-sufficency) is also detected through enzyme-linked immunosorbent
assay (ELISA)-based analysis. Anti-tissue transglutaminase, i.e. Anti-tTG, and anti-endomysial
antibodies, i.e. EmA, is detected through serologic screening, e.g. ELISA, and the presence of
these indicates on an active phase of celiac disease in the patient (Megiornia et al. 2008).
2.1.1. Human Leukocyte Antibody
The majority of the genetic risk factors for developing celiac disease lie within the over 200
genes located in the region for the major histocompatibility complex (MHC), in humans titled
human leukocyte antibody (HLA), on chromosome 6.
The HLA complex, which is the most gene-dense region in the entire genome, codes for
around ten thousand different cell surface peptides, i.e. antigens, especially located on dendritic
cells, B lymphocytes and macrophages, that can be recognised by T cells that produce immune
responses to fight foreign pathogens, thus playing a crucial role in the immune system (Shen
et. al, 2010; Williams 2001). If immune cells recognise self-antigens as foreign antigens,
autoimmune diseases, such as celiac disease, can occur.
The MHC is divided into three class regions, I, II and III. Within class II lies loci for HLA DQ-
genes, which will be in focus in this report. The HLA-DQ gene complex furthermore consists
of two genes, HLA-DQA1 and HLA-DQB1. Further, the gene complex and the genes combine
in different serotypes and haplotypes, some of them listed in Table 2.
4
Table 2: Genetic variants in the HLA gene complex that can cause celiac disease.
Name Type Serotypes Haplotypes
HLA-DQ Gene complex 2, 8 -
HLA-DQA1 Gene - 03, 05
HLA-DQB1 Gene - 02, 0302
Table 3: Combinations of allelic variants of the DQA1 and DQB1 alleles and the risk the combinations compose in
developing celiac disease (Megiorni et. al 2007).
HLA-DQ heterodimer Risk for developing
celiac disease
DQ2.5 and DQ8 Very high
DQ2.5 (with a double dose of DQB1*02) Very high
DQ8 High
DQ2.5 (single dose of DQB1*02) High
DQ2.x (double dose of DQB1*02) High
DQ2.x (single dose of DQB1*02) Low
DQX.5 Extremely low
DQX.x Extremely low
In almost all cases, genetically predisposed individual’s express HLA-DQ2 or HLA-DQ8
variants (Table 3), which makes detection of these serotypes the most relevant method in
determining celiac disease risk or diagnosis (Megorini et. al 2008; Brown, Guandalini, Semrad
& Kupfer 2019; Cecilio & Bonatto 2015). Both genes code for α/β heterodimers involved in
mediating the activation of CD4+, gluten-specific T-cells, with gut-homing potential possibly
leading to lymphocytic elimination of epithelial cells (Jabri & Sollid 2017).
The two HLA genes, DQ2 and DQ8, account for approximately 98% - 99,84% of patient
diagnosis, based on a large European collaborative study, one American study and one Italian
study (Karell et al. 2003; Megiorni et al. 2009; Pietzak et. al 2009), and approximately 95% of
patients carry allelic variants DQA1*05 and DQB1*02 in DQ2 gene, or DQA1*03 and
DQB1*0302 in DQ8 gene (Megiorni et. al 2007), although heterozygosity for either of the DQ2
or DQ8 alleles also induce a risk. Detection of the sensible DQ genes does not predict
development of celiac disease but their absence means a negative prediction of the disease to
close to 100% (Wolters & Wijmenga 2008).
The HLA complex is the most convoluted and polymorphic genetic system in the entire human
genome to ever be discovered. One single locus can be possessed by (for HLA-A, HLA-B,
HLA-C and DRB1) hundreds of different allelic variants (Williams 2001), each variant
determining which antigens will be presented on the cell surface. Because of the redundancy in
allelic variants, the chance that an individual will possess the same allele on both chromosomes
is small, resulting in a wide variation of genotypes within the human population. Furthermore,
the two alleles operate in co-dominance, making the variance even more complex. Some of
these allelic variants (presented in Table 2 and Table 3) are known to cause celiac disease.
HLA class I and II genes show significant sequence homology due to their mutual descend
from the same common ancestor as a result of former gene duplication events. The genes have
closely related pseudogenes (defect, non-functional genes) which errors prevent successful
transcription and translation. These facts make primer designing for PCR diagnosis a critically
important matter to accomplish locus specific amplification in molecular diagnosis. HLA allele
identification is to be considered the most complex problem in molecular diagnostics, with
more than 1300 known HLA alleles (Williams 2001).
5
2.2. Polymerase Chain Reaction
The polymerase chain reaction (PCR) allows for exponential amplification of specific DNA
fragments through repeating of cycles under specific conditions (Harris 1998). The reaction is
continued until desired amount of product is acquired, often around 40 cycles. After 40 cycles,
one single DNA molecule could theoretically have generated over a thousand billion copies (240
= 1,099,511,627,776). Although, in reality, amplification efficiency is not always 100% but
rather 90-105% (Johnson et. al 2013).
PCR is an artificial mimic of when a cell, in vivo, replicates its DNA. Although a cell uses
the aid of helicase to separate the double strand and RNA primase to add primers to the strand,
the procedure is not far off from what occurs during a PCR.
PCR has several fields of applications within molecular biology, such as gene expression,
genotyping (detecting), mutagenesis, sequencing, cloning, etc. In this study, the DNA yield is
used to detect variants of the HLA genes. In fact, the study of allelic sequence variation within
the HLA locus via PCR was in 1986, only the year after the PCR method first was invented,
the inception to forensic analysis based on PCR (Saiki et. al 1986).
The technique advanced into what is called quantitative PCR (qPCR) (also known as real time
PCR but not to be confused with reverse transcriptase PCR as both tend to be abbreviated RT-
PCR). qPCR allows for real time analysis of quantity of PCR product by measuring relative
fluorescence emitted from unspecific fluorescent dye or fluorescent-labelled target-specific
probes, which will be directly proportional to the amount PCR product being produced (more
on probe-based qPCR under ‘Probe’) (Maddocks & Jenkins 2016).
From here on, mentions of “PCR” in this report will be referring to quantitative PCR.
Figure 1: Real-time quantitative PCR (qPCR) and TaqMan® probe chemistry mechanism.
6
2.2.1. The Components
Needed to carry out a quantitative polymerase chain reaction are, besides the PCR instrument
itself and the PCR plate (often 96-well) or PCR tube, following reagents: template DNA,
primers, dNTPs, Mg2+, reaction buffer, polymerase and nuclease free water, and some sort of
reporter molecule. These components together are what is commonly called a Master Mix.
First of all, a template of what is actually being replicated and amplified is necessary. This is
often referred to as the ‘sample’. Template DNA can be derived from a eukaryotic cell, in that
case often in the shape of linear DNA originated in the nucleus, or from a prokaryotic cell in
the shape of a circular bacterial plasmid, or even a virus (the template can in this case even be
derived from RNA and must be converted into complementary DNA through reverse
transcription)(Pfaffl 2010; Maddocks & Jenkins 2016; National Laboratory of Enteric
Pathogens et. al 1991).
Either form, the genomic material of interest must be extracted from whatever cell or capsid
it resides in and purified, before it can be used as a template in PCR analysis.
In a cell, when single stranded DNA is exposed and ready to be transcribed, a primer needs to
be attached to the strand to act as an indicator on where in the genome the transcription is to be
initiated and simply as a primer for nucleotides to further be added to during elongation of the
strand. The primer, a short nucleic acid strand, is added by the enzyme RNA primase in the cell.
The same is true for DNA amplification through PCR: it is only possible under the aid of
primers, a forward and a reversed. This is where synthetic primers, also called oligonucleotides,
come in use.
These are synthetic DNA segments, around 15-30 nucleotides in length, designed to
complementary attach to those particular parts in the template DNA where the start of
transcription is desired (Harris 1998). The primers will run in the opposite direction to the
template DNA, and in turn also run in opposite directions to each other. This means one primer
will run in the forward direction, from the 3’ end, and one primer also 3’ to 5’ but in the reverse
direction to the other primer.
Amplification of two or more target sequences simultaneously is also possible through
multiplex PCR by adding multiple sets of primers to the Master Mix along with different probes
to detect each sequence through its unique fluorescence signal (Basu 2015).
Attaining PCR product is rarely the problem, it is attaining specific PCR product that is the
tricky part. Primer design is a crucial part in PCR (especially in multiplex PCR where several
targets are being amplified in a single reaction) as primers need to be specific enough to not
attach to unwanted parts in the genome (or even to another primer in the mixture, called a primer
dimer) so that random regions in the DNA get amplified. Each primer pair (forward primer and
reverse primer) should enable effective amplification of strictly one amplicon. Very short
primers are often not specific as they risk to complement not only the target sequence and thus
misprime, while long primers mean hybridisation rate will be reduced and by extension, yield
will be small (Basu 2015).
When primers have attached to the exposed single stranded DNA, it is time for the
hybridisation, that is, the annealing of complementary nucleic acids to the template DNA to
extend the forming DNA strand. The hybridisation is catalysed, both in the cell and in a
polymerase chain reaction, by an enzyme called a polymerase (Ishino & Ishino 2014).
The polymerase commonly used in PCR, Taq polymerase, is a thermostable DNA
polymerase isolated from the extreme thermophile bacteria Thermus aquaticus which, thanks
7
to its resistance to high temperatures (while DNA denatures during denaturation steps in PCR,
Taq polymerase stays intact and active), small size, and absence of nuclease activity, makes it
applicable for the field of application (Chien et. al 1976).
Taq polymerase binds to the 3’ end of the primer and synthesises approximately 1 kb each
minute (2 kb the first, initial minute) at an optimal elongation temperature of 70-80 °C (Lorenz
2012), through extension of both primers in 3’ to 5’ direction to the template.
Figure 2: Primers (blue and orange) annealed to template DNA (bold), and Taq polymerase elongating the chain in 3' to
5' direction to the template DNA (arrows).
Deoxynucleotide triphosphates (dNTPs) come in the form of dATP, dTTP, dGTP and dCTP,
each representing one of the four nitrogenous bases adenine, thymine, guanine and cytosine,
and each binding through hydrogen bonding to its corresponding base in the template DNA.
These are the building blocks required to build the new DNA string and need to be in equal
concentration in the Master Mix (Harris 1998).
For proper function, the polymerase needs magnesium cations, which act as a cofactor to the
enzyme as the ions, through interactions with the active site in DNA polymerase and the dNTPs,
catalyses the formation of phosphodiester bonds between the dNTP and the primer on the 3’
end. In addition, the magnesium ions stabilise the structure through bindings and decreases the
negative charge.
As for the other components in the Master Mix, the concentration of magnesium ions can
greatly impact the outcome of the PCR product. An excessive amount of magnesium may lead
to non-specific amplicons due to increase in polymerisation activity, but decrease in overall
reliability and consistency, while an insufficient amount of magnesium will do the opposite and
reduce polymerisation but increase infallibility.
Contribution of magnesium ions to the mixture often come in the form of MgCl2 or MgSO4
and are usually included in the reaction buffer (Slack et. al 2011).
KCl and Tris are often main reagents in the reaction buffer, which key tasks are to contain the
pH balance and maintain ion levels. pH value for the buffer at room temperature is 8.2 to 9 but
when temperature increases, pH for Tris (and thus the buffer) decreases. When the reaction
temperature rises to 72 °C, which is optimal extension temperature for Taq polymerase, pH
decreases to 7.0-7.5, which also is the optimal pH for Taq (Harris 1998). This way the buffer
maintains optimal reaction conditions throughout the PCR.
8
As stated by the name, quantitative PCR gives the opportunity for quantitative analysis of DNA,
i.e. it is possible to determine how much amplicon has been created in real time throughout the
reaction, in comparison to regular PCR, where yield has to be analysed post reaction through,
for example, gel electrophoresis.
Measurement of quantity of the amplicon in real time is made possible through fluorescent
labelling of the PCR product, either with fluorescent dye or fluorescent probes. Fluorescence
emission is measured after each cycle and the signal is proportional to the amount of PCR
product generated.
As opposed to dye-based qPCR, where fluorescent dye binds non-specifically to DNA and
therefore gives non-specific signals, probe-based qPCR is designed to complementary bind to
desirable sequences in the DNA. Just like primer designing is a crucial part of a successful PCR,
so is probe designing. Properly designed probes enable accurate and dependable detection,
which is the basis for qPCR (Basu 2015).
The probe is tagged with a fluorescent reporter. Its emission gives rise to fluorescence
resonance energy transfer (FRET), which is the basis of real-time detection. On the probe, close
to the reporter, is a quencher molecule. When the quencher is present, fluorescens emission
from the reporter is quenched. Then Taq polymerase extends the DNA chain, it hydrolyses the
probe with its exonuclease activity. When the reporter dissociates from the primer and distance
itself from the quencher, fluorescence is emitted and recorded by the detector. Every tagged
molecule sends out its own signal which means signal generated is in direct proportion to the
number of molecules being synthesised – the more molecules, the more intense signal (Wong
et. al 2015).
2.3. Internal Positive Control
An internal positive control (IPC) is sometimes included in the PCR mix so as to detect any
PCR inhibiting substances in the DNA samples (naturally present or added as a result of
contamination in the lab during sample processing and nucleic acid purification), a faulty
sampling or otherwise degradation of the DNA, minimising the risk for false-negative results
or decreased sensitivity (Schrader et. al 2012).
The control sequence is amplified along with the other target sequences and detected with
a different probe to rule out failed amplification in cases where the target sequence could not
be detected.
2.4. The Reaction
The reaction consists of three stages, as displayed in Figure 3: an initial denaturation,
performed just at the beginning of the reaction, followed by a cycling of three temperatures
during the replication step, and ended with a final extension.
9
Figure 3: A typical PCR program with its cycles and temperatures.
2.4.1. Initial Denaturation
During the initial denaturation, double-stranded DNA denatures under high temperature to form
two single strands of DNA. The temperature is set to around 94 °C to 98 °C. For G-C rich
template DNA, the temperature is set closer to 98 °C due to a phenomenon called base-stacking,
which bases guanine and cytosine contributes to. Base-stacking energy contributes to DNA
stabilisation which is why G-C rich DNA will require a higher temperature to denature
(Yakovchuk et. al 2006). Initial denaturation temperature is also dependent on the optimal
temperature for the DNA polymerase activity.
Formation of secondary structures within the template DNA due to unbroken hydrogen
bonds may prevent proper primer annealing and disruption of the polymerase. However, too
high melting temperature may also cause the polymerase to denature, resulting in lower product
yield.
The initial denaturation, performed once before the cycling begins, is of importance because
it ensures that all double stranded DNA is separated into single strands so that primers can
anneal to the target sequence. The high temperature also helps with inactivating possible
nucleases in the sample, that otherwise risk to degrade DNA and primers (Lorenz 2012).
2.4.2. Replication step
DNA amplification takes place during the cycling process in the replication step. Denaturation,
primer annealing and extension/elongation is alternated in thermal cycles, around 40 of them.
After denaturation for approximately a minute, temperature is lowered to around 50-65 °C
(5 °C below the melting temperature for the primers is optimal), thus entering the annealing
step. Now, primers have the time to anneal to their target sequences on the DNA template and
extension of the new DNA strand can begin. The temperature is once again increased, this time
to around 72 °C, to meet the optimal annealing temperature for the polymerase. During the
elongation, the polymerase extends the primer in 5’ to 3’ direction, creating a complementary
DNA strand to the template.
After each cycle, the amount of product created is doubled (at least theoretically). Cycling
is ended when desired amount of product is reached (Lorenz 2012).
10
2.4.3. Final extension step
When desired amount of product is reached, the final extension takes place, during which
uncompleted amplicon is given extended time to finish. During this time, if Taq polymerase is
used, the addition of 3’ adenine residue overhang is added to the amplicons – a useful
modification for some subsequent steps when gene cloning (Lorenz 2012).
2.5. Amplification Plot
Figure 4: A typical amplification plot showing eight amplification curves generated from five samples. Two target
sequences were amplified, “purple” and “orange”. All samples showed substantial amplification for target “purple” and
all of them did at some point cross the threshold value, 0.05. Only in three samples was target “orange” amplified, though
not efficient enough for any of the samples to cross the threshold value, 0.2.
From the amplification plot generated by the QuantStudio™ Real-Time PCR Software, each
sample will appear as a sperate curve (as long as they give rise to any fluorescence emission)
and plotted with the number of cycles in the PCR on the x-axis against the normalised
fluorescence intensity on the y-axis. Rn (normalised reporter) is the fluorescence of the reporter
dye divided by the fluorescence of a passive reference and ΔRn is given when subtracting the
baseline (the initial cycles where little change in fluorescence signal is reported) from Rn.
In Figure 4, two thresholds have been set, one to 0.2 and one to 0.05 (orange and purple).
The thresholds are specific to the targets analysed. In this particular case, five samples were
analysed for two target sequences. All the five samples showed successful amplification of one
of the target sequences but only three of those samples also showed amplification of the other
target sequence. Although, none of these three samples (orange) had a high enough amplicon
concentration to cross the 0.2 threshold.
The higher concentration of the copied target sequence in the beginning of the reaction, the
sooner the fluorescence signal will be detected. It will require less cycles of amplification to
detect a signal from a sample that contain high concentrations of the target sequence.
11
2.6. Ct-value
Ct-values are obtained from a table in the PCR software. The Ct-value for a sample equals the
x-value (cycle number) at which the curve crosses the threshold and is a relative measure of the
targets’ concentration. If a sample generates a low Ct-value, thus showing substantial
amplification in an early cycle, it means the concertation of the target sequence initially was
high, compared to a sample with a higher Ct-value. A sample with a high Ct-value needed more
cycles to amplify the target sequence, hence the target sequence initially existed in a lower
concentration.
2.7. Passive Reference Dye, Reporter Dye and Master Mix: Changes on Ct-values
A passive reference dye’s function is to provide an internal fluorescence to act as a reference to
the reporter dyes to be normalised against. Normalisation is needed to account for fluctuations
in fluorescence due to changes in volume or concentration.
Depending on the Master Mix composition, there is a great chance fluorescence signals will
vary, even though there is no change in concentration of probe, reference dye or template DNA.
This is because factors such as pH and salt concentrations within the master mixes can vary and
thereby affect the fluorescence signals generated. This is displayed in Figure 5, accessed from
Applied Biosystems application note “Real-time PCR: Understanding Ct”, which can be
downloaded through their website.
Although, in this project, changing the Master Mix also meant changing the passive
reference dye, from ROX to MP, and since the Rn value is defined as the ratio of the reporter
dye (FAM/NED/VIC) divided by the reference dye (ROX/MP), this change in reference dye
most definitely will have an impact on the Rn values. Having the Rn values changed (and
thereby also the Ct-values) does not mean sensitivity of the reaction was compromised.
Figure 5: Difference in fluorescence (VIC and ROX) due to different Master Mixes. The y-axis shows the emission
intensity and the x-axis shows the emission wavelength of the fluorophore. Picture from ABI’s “Real-time PCR:
Understanding Ct”.
12
As fluorescence intensity may vary with different Master Mixes, so may the Ct-values, even
though the sensitivity of the Master Mixes is the same (Figure 6). It is important to notice that
variations in Ct-values do not reflect the performance of the reaction, however, when
implementing a different method of analysis, it is beneficial to not have the Ct-values change
since patient diagnosis is based on Ct-values and that would mean that algorithms and manual
evaluations will have to be changed, or else diagnosis will be incorrect.
Figure 6: Ct-values (x-intercept) differ between the two Master Mixes even though the target concentration was identical.
Threshold are in both reactions set the same. Ct-value for Master Mix A is 14 and Ct-values for Master Mix B is around
12,5. Picture from ABI’s “Real-time PCR: Understanding Ct”.
In addition, a change in Master Mix and/or passive reference dye might also lead to an increase
in standard deviation of the Ct-values of the samples in the PCR assay, which is not favourable
as that will result in lower confidence when distinguishing between target concentrations with
small differences.
In the original method with ABI 7300 and TaqMan, the passive internal reference dye was ROX
which is supposed to give superb precision on Applied Biosystems qPCR instruments (such as
the 7300), according to the manufacturer ThermoFisher. In the new method with QS and Solis,
Mustang Purple (MP) was used as passive reference. In each duplex reaction, regardless
method, either FAM, VIC or NED dyes were used to detect the two target sequences in each
assay.
Table 4: Passive reference dye and reporter dye in use for each method and PCR instrument. Emission maxima for each
dye is specified.
Instrument Passive reference
dye
Reporter dye type 1 Reporter dye type 2 Reporter dye type 3
Applied Biosystems
7300
ROX (~610 nm) FAM (~520 nm) NED (~575 nm) VIC (~554 nm)
QuantStudio 6 or 7 MP (~654 nm) FAM NED VIC
13
3. Methodology
3.1. DNA Samples
The samples used in the experiments was already extracted and purified DNA obtained from
Dynamic Code’s costumers and samples were picked randomly for the experiments where they
all throughout remained anonymous (i.e. could never be linked to a person).
Purification of DNA was done with Quick-DNA™ 96 Kit from the brand Zymo Research
which is a technique based on a series of lysis buffer wash steps on a silica-based plate and
where the finished product finally is eluted through the silica into an elution plate. After eluted,
the product is ready to be directly placed into a PCR-mix in a PCR-plate.
3.2. Negative and Positive Controls
Controls are used to supervise the method of analysis to make sure it works properly and to
warn against possible errors as well as for follow up of the method over a longer period of time.
To better understand where in the process contamination took place, or where possible
inhibitors were introduced in the sample, negative and positive controls can be specified into
extraction negative and positive and amplification negative and positive. This makes it possible
to more or less pinpoint where in the process contamination was introduced and what steps
needs to be redone. For example, if the extraction negative control was contaminated, the entire
extraction process and the following amplification process has to be redone, but if only the
amplification negative control failed, solely the amplification has to be redone.
For this project, only amplification negatives and positives were of relevance since no
extraction was done.
With all analyses involved in this project, a negative control along with a positive control
unique for each mix (four of them), were amplified together with the rest of the samples.
Amplification negative and positive were placed directly to the PCR-plate along with the other
samples right before amplification.
3.2.1. Negative Controls
Negative controls verify sample-to-sample contamination, PCR contamination and reagent
contamination. The negative control should only contain those reagents necessary for the
polymerase chain reaction, but no template. It should be ‘blank’ and should not be showing any
kind of amplification in the amplification plot. If there is considerable noise from the negative
control sample, the control was at some point in the process contaminated and chances are all
or some of the other samples on that plate also were, meaning none of the results can be trusted.
In the negative control wells, 3 µl ddH2O was placed instead of the 3 µl template DNA.
Accepted Ct-values for the negative controls are everything >35 or “undetermined” (not enough
amplification to cross the given threshold).
3.2.2. Positive Controls
Positive controls verify that the DNA yield is good enough to draw any conclusions from. If
the positive control, which should be of known concentration, is ‘as positive’ as expected, it
14
affirms that DNA was not degraded by inhibitors thereby risking false-negative or faulty results,
and that in all the reaction was carried through completely and correctly.
In the positive control wells, 3 µl purified DNA, extracted with Quick-DNA™ 96 Kit, from
specific individuals of Dynamic Code’s staff positive for the given genotypes being analysed,
was placed instead of the 3 µl template DNA. At some point in the project, amplification
positive related to two of the mixes ran out and a new positive control was created and used.
The new positive control was created through DNA pooling of equal amounts of samples
previously analysed that showed strong amplification (corresponding to the Ct-values in table
of the target genes).
Accepted Ct-values for the positive controls are the same as what is considered a positive
result for costumer samples (Table 5).
Table 5: Accepted Ct-values for the positive controls.
Gene Ct-value
DQA1*02 <29
DQA1*05 <27
DQB1*02 <28
DQA1*03 <30
DQB1*0302 <27
DQB1*0302B# <27
IPC <32
3.3. Reaction Components
The total reaction volume was always 15 µl per well. 12 µl of these were the PCR-mix (Master
Mix) and 3 µl were the DNA sample (or 3 µl ddH2O if negative control/3 µl positive control
DNA template if positive control).
PCR-mixes consists of double-distilled water (from here on labelled only as ‘water’), HLA-
mix (1-4 depending on reaction) and Master Mix, either TaqMan™ Gene Expression Master
Mix (original method) or SolisFast® Probe qPCR Mix (Purple) (new method). Volumes vary
depending on the Master Mixes stock concentration where one of them (specified further down)
is more concentrated and needs to be diluted with more water.
The double distilled water, ddH2O, is deionized water with a resistance better or equal to
18MΩcm and it is obtained from a dispenser called ELGA PureLab Flex and aliquoted into
sterile Eppendorf tubes before use.
3.3.1. HLA-mix
What is labelled as “HLA-mix” is the oligonucleotide-mix containing the primers and probes.
The oligo-mix was for this optimisation project the most central part of the method optimisation
where volumes of the different components (Table 9) were varied and tested.
Each one of the four oligo-mixes contains primers and probes for two target sequences (so
called duplex). A total of six unique targets are amplified in pairs in this analysis:
15
Table 6: Target genes amplified in each of the four HLA-mixes.
HLA-mix 1 DQA1_05 IPC
HLA-mix 2 DQA1_02 DQB1_02
HLA-mix 3 DQA1_03 DQB1_0302
HLA-mix 4 DQB1*0302 B IPC
Mix 4 is a confirmative analysis performed on any samples positive for only HLA-DQ8
(DQA1_03 & DQB1_0302) that have been analysed with mixes 1-3 prior.
3.3.2. PCR-mix: TaqMan™ Gene Expression Master Mix
In the original method used for HLA gene amplification via qPCR within Dynamic Code,
TaqMan Master Mix is used and samples are placed in the ABI 7300 instrument.
Standard protocol for the TaqMan PCR mixes is as followed (12 µl per sample, 15 µl in
total when template is added):
Table 7: Protocols for PCR-mixes based on the TaqMan Master Mix, original method.
Mix 1 (DQA1_05 & IPC)
Stock concentration End concentration Volume
TaqMan™ Gene
Expression Master Mix
2X 1X 7,5 µl
HLA-mix 1 25X 1X 0,6 µl
ddH2O 3,9 µl
Mix 2 (DQA1_02 &DQB1_02)
Stock concentration End concentration Volume
TaqMan™ Gene
Expression Master Mix
2X 1X 7,5 µl
HLA-mix 2 25X 1X 0,6 µl
ddH2O 3,9 µl
Mix 3 (DQA1_03 & DQB1_0302)
Stock concentration End concentration Volume
TaqMan™ Gene
Expression Master Mix
2X 1X 7,5 µl
HLA-mix 3 25X 1X 0,6 µl
ddH2O 3,9 µl
Mix 4 (DQB1*0302 B & IPC)
Stock concentration End concentration Volume
TaqMan™ Gene
Expression Master Mix
2X 1X 7,5 µl
HLA-mix 4 25X 1X 0,6 µl
ddH2O 3,9 µl
The polymerase included in TaqMan is called AmpliTaq Gold DNA Polymerase, UP (Ultra
Pure). Passive internal reference is ROX dye.
16
3.3.3. PCR-mix: SolisFast® Probe qPCR Mix (Purple)
When running the reaction with Solis Master Mix, composition of PCR-mix components is
slightly changed because of the 5X-concentration (compared to TaqMan’s 2X-concentration),
therefore more water is added.
Standard protocol for the Solis PCR mixes is as followed (12 µl per sample, 15 µl in total
when template is added):
Table 8: Protocols for PCR-mixes based on the Solis Master Mix, new method.
Mix 1 (DQA1_05 & IPC)
Stock concentration End concentration Volume
SolisFast® Probe qPCR
Mix (Purple)
5X 1X 3,0 µl
HLA-mix 1 25X 1X 0,6 µl
ddH2O 8,4 µl
Mix 2 (DQA1_02 &DQB1_02)
Stock concentration End concentration Volume
SolisFast® Probe qPCR
Mix (Purple)
5X 1X 3,0 µl
HLA-mix 2 25X 1X 0,6 µl
ddH2O 8,4 µl
Mix 3 (DQA1_03 & DQB1_0302)
Stock concentration End concentration Volume
SolisFast® Probe qPCR
Mix (Purple)
5X 1X 3,0 µl
HLA-mix 3 25X 1X 0,6 µl
ddH2O 8,4 µl
Mix 4 (DQB1*0302 B & IPC)
Stock concentration End concentration Volume
SolisFast® Probe qPCR
Mix (Purple)
5X 1X 3,0 µl
HLA-mix 4 25X 1X 0,6 µl
ddH2O 8,4 µl
The polymerase included in Solis is called SolisFAST® DNA Polymerase. Passive internal
reference is MP (Mustang Purple).
17
3.4. Oligonucleotide Mixes
Six different sequences were, together with an internal positive control targeting the human
RNA polymerase II gene in reaction 1 and 4, amplified to express and detect HLA-DQ2 and
HLA-DQ8 in samples analysed for celiac disease.
Each mix is duplexed and each target has one forward primer (FP1 and FP2), one reversed
primer (RP1 and RP2) and one probe (P1 and P2), as listed below. In tables below are
concentrations and volumes optimised for the original method, with TaqMan Master Mix run
on ABI 7300.
Table 9: Protocols for oligonucleotide mixes (HLA-mixes), original method.
Mix 1 (DQA1_05 & IPC)
Oligo Tag Stock
concentration
End
concentration
Volume
HLAmix1_FP1 100 µM 300 nM 45 µl
HLAmix1_RP1 100 µM 300 nM 45 µl
HLAmix1_P1 FAM 100 µM 200 nM 30 µl
HLAmix1_mix4_FP2 100 µM 50 nM 7,5 µl
HLAmix1_mix4_RP2 100 µM 50 nM 7,5 µl
HLAmix1_mix4_P2 NED 100 µM 200 nM 30 µl
ddH2O 435 µl
Mix 2 (DQA1_02 & DQB1_02)
Oligo Tag Stock
concentration
End
concentration
Volume
HLAmix2_FP1 100 µM 300 nM 45 µl
HLAmix2_RP1 100 µM 300 nM 45 µl
HLAmix2_P1 VIC 100 µM 200 nM 30 µl
HLAmix2_FP2 100 µM 300 nM 45 µl
HLAmix2_mix3_RP2 100 µM 300 nM 45 µl
HLAmix2_P2 FAM 100 µM 200 nM 30 µl
ddH2O 360 µl
Mix 3 (DQA1_03 & DQB1_0302)
Oligo Tag Stock
concentration
End
concentration
Volume
HLAmix3_FP1 100 µM 300 nM 45 µl
HLAmix3_RP1 100 µM 300 nM 45 µl
HLAmix3_P1 VIC 100 µM 200 nM 30 µl
HLAmix3_FP2 100 µM 300 nM 45 µl
HLAmix2_mix3_RP2 100 µM 300 nM 45 µl
HLAmix3_P2 FAM 100 µM 200 nM 30 µl
ddH2O 360 µl
18
Mix 4 (DQB1*0302 B & IPC)
Oligo Tag Stock
concentration
End
concentration
Volume
HLAmix1_FP1 100 µM 300 nM 45 µl
HLAmix1_RP1 100 µM 300 nM 45 µl
HLAmix1_P1 FAM 100 µM 200 nM 30 µl
HLAmix1_mix4_FP2 100 µM 50 nM 7,5 µl
HLAmix1_mix4_RP2 100 µM 50 nM 7,5 µl
HLAmix1_mix4_P2 NED 100 µM 200 nM 30 µl
ddH2O 435 µl
The concentrations above are what have been optimised in order to make the HLA analysis run
properly with the Solis Master Mix and QS instruments.
3.5. PCR Optimisation and Validation
Although most manufacturers provide solid and robust assays designed to meet all necessary
criteria, ThermoFisher included, assays need to be adjusted depending on the selection of
instrument, reagents, thermal PCR conditions and primer and probe concentrations.
Although optimisation can be done through adjustment of many different components, it is
favourable to maintain standard buffer conditions and instead solely focus on modifying primer
and/or probe concentrations, which will affect primer binding kinetics.
Primer design optimisation is sometimes a valid way of approach, for example if primer
dimers are a problem, however, in this project, focus was solely on optimisation through
modifications in primer/probe concentrations. Primers were already design and working
properly.
The approach was a one-factor-at-a-time approach where only one of the components
(forward and reverse primer for target one, forward and reverse primer for target two, probe for
target one or probe for target two) concentration was varied while remaining components had
a fixed concentration. For DQ-targets, the original primer concentration, 300 nM, was tested
together with 100 nM and 900 nM. For IPC, 100 nM and 300 nM were tested for primers, and
for each probe, 200 nM, 100 nM and 400 nM were tested.
Figure 7: Overview of experiments on primer (blue squares) and probe (grey squares) optimisation during the project. The
so called one-factor-at-a-time approach. Probe 1 is connected to target 1 and probe 2 is connected to target 2.
19
When all mixes were completely optimised, each mix was placed on a full PCR-plate and
compared to the original method by placing those same samples on yet another PCR-plate
which was analysed according to the original, already optimised method. If results added up
with the already validated method, the optimisation was successful.
Figure 8: Overview of validation experiments of the four optimised mixes.
20
4. Method
4.1. Optimisation of Primers and Probes
An oligonucleotide mix (oligo-mix) was prepared with the concentrations being evaluated in
the particular experiment. When doing this, the original protocol for preparing oligo-mixes
(Table 9) was used as a guideline and volumes of the components were adjusted to create the
desired concentrations.
The PCR-mixes were prepared, one for each concentration being examined in the particular
experiment. For experiments on ABI 7300 according to the original method, the protocols in
Table 7 were followed, and for experiments on QS 6 or 7, the protocols in Table 8 were
followed. The previously prepared oligo-mix was combined with the Master Mix and water, 12
µl for each sample in total.
12 µl of the PCR-mix was pipetted into each well of a PCR-plate. 3 µl template DNA was then
pipetted in each well, adding up to a total of 15 µl reaction solution per well. One amplification
positive control and one amplification negative control was added for each different
concentration. See appendix A for PCR-plate setup.
Initially, one full PCR-plate containing all four HLA-mixes (24 wells of each mix) was pipetted
and analysed according to the original method, and one with the new Solis Master Mix on the
QS, i.e. two identical plates but with different Master Mixes and placed in different instruments.
The samples and their positions on the plates matched up between the two plates, so as to make
a direct comparison of the results possible. This experiment was conducted to get an overview
of the differences right away from just running the experience non modified, with a different
Master Mix and on a different instrument.
The experiments following this initial experiment was conducted based on the one-factor-at-
a-time approach described in chapter 3.6.
4.2. Validation of Modified PCR Method
After optimisation of each primer and each probe in all the mixes was completed, validation
experiments were performed, one for each of the four HLA-mixes, where the final
concentration for each of the components (primers and probes) was tested on a full PCR-plate.
The same samples were placed on two PCR-plates, one according to the original method and
one according to the optimised method. This was done for each of the four mixes.
The data obtained from the validation experiments were compared, and the Ct-values obtained
from each method were inserted into an algorithm that determines final HLA diagnosis. The
algorithm compared the values from each method, and determined final diagnosis (positive or
negative for the DQ-genes, or not proven) so as to see if final results were the same regarding
method. If so, the goal of the project was reached – final results were the same disregarding of
method.
21
5. Results
From the initial experiment, where the original method was compared to the new method, but
with no modifications to any of the primer or probe concentrations, the differences in Ct-values
between the methods were calculated (Solis QS – TaqMan 7300). The values varied from a
median of -1,8 (for DQB1_0203 in mix 3) up to 6,6 (for DQB1_02 in mix 2).
Table 10: Mean and median of the difference between Ct-values from each sample on QS versus 7300 before optimisation.
Mix 1 Mix 2 Mix 3 Mix 4
DQA1_05 IPC DQA1_02 DQB1_02 DQB1_0203 DQA_03 DQB1*0203 B IPC
Mean 4,207 2,866 -0,239 6,603 -1,812 2,563 3,913 3,286
Median 4,119 3,018 0,794 6,563 -1,897 2,631 5,744 3,635
Also found from this experiment was the fact that 12 of the 88 costumer samples on the plate
that was valued as “undetermined” (was not assigned a Ct-value due to too little amplification)
on ABI 7300, was in fact detected on QS, although with high Ct-values. An indication on a
greater sensitivity on the QS compared to ABI 7300.
5.1. Mix 1: DQA1_05
Figure 9: Amplification plot for mix 1 DQA_05 300 nM (red), 100 nM (pink) and 900 nM (purple), and IPC 50 nM
(purple). Probe 1 200 nM and probe 2 200 nM.
In general, the shape of the curves is similar with some shifting along the x and y axis between
the different concentrations. The 900 nM concentration of DQA1_05 did result in the highest
amplitude amongst the three different concentrations, especially higher than 100 nM, but also
slightly higher than 300 nM. Although, when studying the different concentrations individually,
it seems like DQA1_05 samples with 900 nM concentration did somewhat suppress the IPC
signals (purple) the most. Regarding Ct-values, samples appear in pretty much the same range
when it comes to concentrations 300 nM and 900 nM. Ct-values for the 100 nM samples was
much higher than both higher concentrations, indicating on a less effective amplification.
22
The IPC amplification was, in general, far too low when using the original concentration from
Table 9 of 50 nM. IPC should be in the Ct-range 23-32 while Ct-values from this experiment
were 33 and above. Not only was the amplification delayed, also the efficiency was poor. No
clear exponential phase or even linear phase was achieved. IPC primer concentration had to be
increased drastically in order to get more sensible amplification.
For DQA1_05 in Mix 1, 900 nM concentration was at this point considered to be the most
optimal and following optimisation experiments on Mix 1 was based on this concentration for
DQA1_05.
5.1.1. Mix 1: IPC
Figure 10: Amplification plot for mix 1 IPC 100 nM (green) and 300 nM (blue), and DQA1_05 900 nM. Probe 1 200 nM
and probe 2 200 nM.
When IPC primer concentration was increased to 100 nM (green) and 300 nM (blue), signals
were stronger and appearance of the curves made more sense. There was a significant difference
between the stronger and the weaker concentration, both in amplitude and in amplification
efficacy, where samples analysed with the higher concentration were detected with higher
signals as well as after fewer cycles.
However, the goal is not to obtain the most efficient amplification of the IPC, since the purpose
of the internal control is to indicate whether any inhibitors might be present in the sample. If
the IPC concentration is made too high, results of a quantitative amplification might be faulty
as the signals could be exaggerated due to the excessive concentration.
Although, at this point, 300 nM IPC was considered giving the best results but would later be
reconsidered due to the fact just stated.
23
5.1.2. Mix 1: Probe 1
Figure 11: Amplification plot for mix 1 probe 1 in 100 nM (yellow), 200 nM (red) and 400 nM (grey), DQA1_05 900 nM,
IPC 300 nM, signals are hidden. Probe 2 200 nM.
For probe 1, connected to DQA1_05 in mix 1, signals were stronger the higher the concentration
of the probe. Also, Ct-values decreased when concentration increased, i.e. samples were
amplified earlier in the PCR when probe concentration was higher.
400 nM gave objectively speaking slightly better results, but 200 nM was not far off. Therefore,
due economic reasons, 200 nM was chosen.
5.1.3. Mix 1: Probe 2
Figure 12: Amplification plot for mix 1 probe 2 in 100 nM (orange), 200 nM (lilac) and 400 nM (blue), DQA1_05 900
nM, IPC 300 nM, probe 1 200 nM.
In this case, regarding probe 2 connected to IPC in mix 1, the higher the concentration the better
the result was not the case.
The lowest probe concentration, 100 nM, clearly gave the worst results – amplification
started many cycles into the reaction, fluorescence signals were weak and there was a large
scattering between samples regarding Ct-values.
24
Figure 13: Amplification plot for mix 1 probe 2 in 100 nM (orange), 200 nM (lilac) and 400 nM (blue), DQA1_05 900 n,
IPC 300 nM, probe 1 200 nM. Only samples 1-8 are previewed for concentrations 200 nM and 400 nM.
If comparing a smaller selection of samples of 200 nM and 400 nM (Figure 13), it becomes
clearer as to which one of these concentrations was more successful.
200 nM was considered the optimal concentration for probe 2.
5.2. Mix 2: DQA1_02
Comparing the three variations of concentrations of the forward and reversed primers for
DQA1_02 target in mix 2 (see appendix B for amplification plots), 100 nM had by far the worst
amplification. While six samples from the 300 nM concentration and five samples from the 900
nM concentration was clearly positive, none of the samples from the 100 nM was successfully
amplified; they appear very late into the reaction and amplification curves do not adapt desirable
shape.
When comparing 300 nM to 900 nM, there is a better distinction between the two clusters of
curves in the lower concentration, 300 nM. When comparing samples individually, one can see
that DQB1_02 signals have been supressed by raising the concentration on DQA1_02 primers.
In some samples, the DQB1_02 is barely detected at all.
Considering all this, 300 nM was chosen as the most optimal primer concentration for
DQA1_02.
5.2.1. Mix 2: DQB1_02
When comparing the three concentrations for the second target in mix 2, DQB1_02, what is
most obvious is the spread amongst the signals in the lowest concentration, 100 nM. The non-
concurrent signals along with poor fluorescence signal rules out this concentration to be the
best.
25
Figure 14: Amplification plot for mix 2 DQB1_02 in 300 nM (green) and DQB1_02 900 nM (blue), probe 1 and probe 2
both 200 nM.
When looking at 300 nM and 900 nM, displayed together in the figure above, there are no
apparent differences. Therefore, due to economic reasons, 300 nM was considered the optimal
primer concentration of DQB1_02.
5.2.2. Mix 2: Probe 1
The most apparent difference between the different concentrations on probe 1 connected to
DQA1_02 is the spread in Ct-values amongst the samples. Raising the concentration up to 400
nM seems to have resulted in non-current signals amongst the samples. In addition, the non-
specific amplification visible in the far right of the plot (see appendix B) was a bigger issuer
with this concentration where many of those non-specific curves crossed the threshold quite
early in the reaction.
For 100 nM, the non-specific amplicons did barely cross the threshold, and if they did, it was
at a high Ct-value. Results were not much better for 300 nM; therefore 100 nM was chosen for
probe 1.
5.2.3. Mix 2: Probe 2
100 nM concentration was not successful for probe 2. Amplification was just not efficient
enough.
400 nM clearly resulted in stronger signals of DQB1_02, but at the same time DQA1_02 signals
were supressed. Over all, curves in the plot showing probe 2 concentration 200 nM showed
great results and thus, this concentration was chosen as the optimal.
26
5.3. Mix 3: DQA1_03
Figure 15: Amplification plot for mix 3 DQA1_03 in 100 nM (red), 300 nM (yellow) and 900 nM (blue), and DQB1_0302
300 nM (grey), probe 1 and probe 2 both 200 nM.
Moving on to mix 3, even displaying all three concentrations in the same plot, one can clearly
notice the differences between them.
100 nM primer concentration resulted in very poor amplification of DQA1_03 where all
samples positive for DQA1_03 was detected late and with weak fluorescence.
300 nM and 900 nM curves adapt the same appearance and amplification seems to be efficient.
However, there is a significant difference in at what point samples start amplifying as well as
when they intercept with the threshold. Considering these aspects, 900 nM amplification did
perform better and decision was at this point made to move forward with this concentration.
5.3.1. Mix 3: DQB1_0302
Figure 16: Amplification plot for mix 3 DQB1_0302 in 100 nM (red), 300 nM (green) and 900 nM (blue), and DQA1_03
900 nM (yellow), probe 1 and probe 2 both 200 nM.
27
Generally, between all three concentrations, amplification of non-specific sequences was a
bigger problem than with previously targets, especially when using 300 nM and 900 nM primer
concentration.
Regarding the signals that are obviously DQB1_0302, the target sequence, there is a correlation
in primer concentration, amplitude and cycle number. The higher the concentration, the higher
the amplitude and the lower the cycle number.
However, there was no greater difference between 300 nM and 900 nM, as both primer
concentration performed well and due to economic reasons, 300 nM was therefore considered
the optimal concentration for DQB1_0302.
5.3.2. Mix 3: Probe 1
Figure 17: Amplification plot for mix 3 probe 1 in 100 nM (green), 200 nM (yellow) and 400 nM (lilac), DQA1_03 900
nM and DQB1_0302 300 nM, probe 2 200 nM.
The lowest concentration of probe 1 connected to DQA1_03, 100 nM, in mix 3 clearly gave the
weakest fluorescence signals and also the highest cycle numbers. In addition, when looking at
the targets individually, 100 nM made non-specific amplification signals more wide spread than
the higher concentrations. Non-specific amplification was reduced with the higher probe
concentrations.
28
Figure 18: Amplification plot for mix 3 probe 1 in 200 nM (yellow) and 400 nM (lilac), DQA1_03 900 nM and
DQB1_0302 300 nM, probe 2 200 nM.
DQA1_03 signals from the two higher concentrations were similar, in fact pretty much
overlapping each other, see Figure 18. Due to economic reasons, the lower concentration, 200
nM was chosen.
5.3.3. Mix 3: Probe 2
Results from amplification with all three probe concentrations are similar and all yield strong
signals implying an efficient amplification. 100 nM was considered the optimal concentration
for probe 2 in mix 3.
5.4. Mix 4: DQB1*0302 B
IPC signals from all three mixes were similar and was therefore not considered when evaluation
the results from the primer concentrations of DQB1*0302 B in mix 4.
As in most other previous cases, the lowest primer concentration did perform unsatisfactory in
regards to efficiency.
29
Figure 19: Amplification plot for mix 4 DQB1*0302 B in 300 nM (lilac) and 900 nM (blue), IPC has been hidden in the
plot, probe 1 and probe 2 both 200 nM.
300 nM and 900 nM are displayed together in the amplification plot above to show the
accordance with each other. With both concentrations, there is a clear distinction between what
are non-specific amplification signals and what are DQB1*0302 B signals. The target signals
are clustered nicely together and amplification efficiency was good. Even though the sample
size is small, at least within these samples 300 nM signals appear at a lower Ct and with even
more narrow spread.
300 nM was chosen as the most optimal concentration for primers connected to this target.
5.4.1. Mix 4: IPC
IPC in mix 4 was initially only tested in 300 nM. This concentration was considered too high
for the IPC as curves appeared with Ct-values as low as the Ct-values, and explained in chapter
5.1.1., the goal is not to have IPC signals as high as possible, at as low Ct-values as possible.
In a later experiment, 100 nM was confirmed as the optimal IPC primer concentration, but in
the following experiments regarding mix 4, 300 nM IPC is what is being used.
5.4.2. Mix 4: Probe 1
Generally, in this assay signals were strong and amplification efficient. The lowest
concentration clearly gave the weakest signals and highest Ct-values. Results obtained from
200 nM and 400 nM were similar in regards to signal strength and especially in regards to Ct-
values where all samples of these concentrations fall within the same range.
Due to economic reasons 200 nM was chosen to move forward with.
5.4.3. Mix 4: Probe 2
For the probe connected to the IPC in mix 4, 400 nM showed the worst performance out of the
three concentrations. IPC signals were discordant and a real plateau phase was not reached
30
before the 40 cycles in the reaction, if compared to both 100 nM and 200 nM, where samples
gather and plateau around Ct 36.
Considering the two lower concentrations, results are more satisfying in 200 nM samples
where amplitude is higher and spacing along the threshold is slightly smaller. Results are not
bad with 100 nM probe either, but since 200 nM was chosen for this probe in mix 1, choosing
the same concentration in this mix will ease laboratory work and reduce complications when
preparing the mixes.
5.5. Validation
5.5.1. Mix 1
When comparing the two targets within mix 1 run with the new optimised method on QS with
the old method on 7300, Ct-values from each method align well.
Ct-values of DQA1_05 range from 21,3 to 26,7 on 7300, and between 22,9 and 28,4 on QS.
For IPC the interval is 23,2 to 33,4 on 7300, and 24,8 to 35,6 on QS.
5.5.2. Mix 2
Samples positive for DQA1_02 range from 22,2 to 27,7 on 7300 and from 22,9 to 27,4 on QS.
For DQB1_02 the interval is 21,4 to 26,8 on 7300 and 22,6 to 27,7 on QS. Hence, there is a
great accordance between the Ct-values obtained from the different methods, for both targets
included in mix 2.
5.5.3. Mix 3
Besides the deviating sample at around Ct 36 in the DQA1_03 plots, samples range from 22,8
to 34,7 on 7300 and from 23,5 to 28,6 on QS. If the deviating sample at 30,0 on 7300 is
disregarded, the sample range is 22,8 to 28,5. This particular sample is “undetermined” on QS.
For DQB1_0302, Ct-values for positive samples fall within the interval 21,0 to 25,3 on 7300
and 21,2 to 26,5 on QS.
5.5.4. Mix 4
Samples positive for DQB1*0302 B in mix 4 fall within the Ct range 19,3 to 25,5 on 7300 and
21,0 to 26,4 on QS. For the IPC, values are 25,2 to 36,2 on 7300 and 22,3 to 32,7 on QS.
After optimisation of the primer and probe concentrations along with lowering of some of the
thresholds, the differences between Ct-values from each sample obtained from both methods
was calculated (Solis QS – TaqMan 7300). The medians of the differences are presented in
Table 11, together with corresponding medians calculated from data before optimisation.
31
Table 11: Medians of the difference between Ct-values from each sample on QS versus 7300 after optimisation and
adjustments to the thresholds compared to before optimisation and with old thresholds.
Mix 1 Mix 2 Mix 3 Mix 4
DQ
A1_05 IPC
DQ
A1_02
DQ
B1_02
DQ
B1_0203
DQ
A1_03
DQB1
*0203 B IPC
Median
before: 4,119 3,018 0,794 6,563 -1,897 2,631 5,744 3,635
Median
after:
0,504
0,404
0,368
-0,453
0,374
0,794
0,566
-3,678
Table 12: Mean values of medians from Table 11, before and after optimisation for all targets and for only DQ-targets.
Before optimisation After optimisation
Mean difference DQ-targets 2,992 0,359
Mean difference all targets 3,076 -0,140
5.6. Final Primer and Probe Concentrations
Mix 1 (DQA1_05 & IPC)
Oligo End
concentration,
original method
End
concentration,
new method
HLAmix1_FP1 300 nM 300 nM*
HLAmix1_RP1 300 nM 300 nM*
HLAmix1_P1 200 nM 200 nM
HLAmix1_mix4_FP2 50 nM 100 nM*
HLAmix1_mix4_RP2 50 nM 100 nM*
HLAmix1_mix4_P2 200 nM 200 nM
Mix 2 (DQA1_02 & DQB1_02)
Oligo End
concentration
End
concentration,
new method
HLAmix2_FP1 300 nM 300 nM
HLAmix2_RP1 300 nM 300 nM
HLAmix2_P1 200 nM 100 nM
HLAmix2_FP2 300 nM 300 nM
HLAmix2_mix3_RP2 300 nM 300 nM
HLAmix2_P2 200 nM 200 nM
32
Mix 3 (DQA1_03 & DQB1_0302)
Oligo End
concentration
End
concentration,
new method
HLAmix3_FP1 300 nM 300 nM*
HLAmix3_RP1 300 nM 300 nM*
HLAmix3_P1 200 nM 200 nM
HLAmix3_FP2 300 nM 300 nM
HLAmix2_mix3_RP2 300 nM 300 nM
HLAmix3_P2 200 nM 100 nM
Mix 4 (DQB1*0302 B & IPC)
Oligo End
concentration
End
concentration,
new method
HLAmix1_FP1 300 nM 300 nM
HLAmix1_RP1 300 nM 300 nM
HLAmix1_P1 200 nM 200 nM
HLAmix1_mix4_FP2 50 nM 100 nM*
HLAmix1_mix4_RP2 50 nM 100 nM
HLAmix1_mix4_P2 200 nM 200 nM
Concentrations marked with an asterisk (*) were changed following further evaluation of the
results, after the experiments were already carried through. For example, primer concentration
for DQA1_05 in mix 1, named HLAmix1_FP1 and HLAmix1_RP1 in the table above, was first
set to 900 nM and further optimisation of the mix was based on this concentration. It was later
on changed to 300 nM. However, all validation experiments, chapter 5.5., were conducted with
the re-evaluated concentrations as shown in the tables above.
5.7. Final Affirmation of Genotypes
After evaluation of the results obtained from the validation experiments, decision was made to
lower some of the thresholds to better match the Ct-values obtained with the new method with
those obtained with the old one. Thresholds were adjusted according to Table 13.
Table 13: Thresholds according to the original method and the adjusted thresholds after optimisation.
Target Old threshold New threshold
DQA1_02 0,1 0,1
DQA1_03 0,1 0,1
DQA1_05 0,2 0,1
DQB1_02 0,2 0,1
DQB1_0302 0,2 0,1
IPC 0,05 0,02
After thresholds were adjusted, each Ct-value, from all mixes on both instruments, was inserted
into an Excel sheet programmed to calculate the final genotyping result (positive, negative or
not proven) for each DQ-target, also taking IPC value into account. The algorithm presents the
conclusive HLA-diagnosis (positive/negative for HLA-DQ2 and HLA-DQ8).
The final genotype results for HLA-DQ2 and HLA-DQ8 is presented in the list below.
33
QS QS 7300 7300
Sample
Name
HLA-
DQ2
HLA-
DQ8
HLA-
DQ2
HLA-DQ8
1 + - + -
2 - - - -
3 - - - -
4 - - - -
5 - - - -
6 - - - -
7 + - + -
8 - - - -
9 - - - -
10 - - - -
11 - - - -
12 - - - -
13 - - - -
14 - - - -
15 - - - -
16 - - - -
17 - - - -
18 - - - -
19 - - - -
20 - - - -
21 - - - -
22 - - - -
23 + - + -
24 + - + -
25 + - + -
26 - - - -
27 - - - -
28 - - - -
29 - - - -
30 - - - -
31 - - - -
32 + - + -
33 - - - -
34 - - - -
35 - - - -
36 - - - -
37 - - - -
38 - - - -
39 - - - -
40 - - - -
41 + - + -
42 - - - -
43 - - - -
44 - - - -
34
45 + - + -
46 - - - -
47 - - - -
48 - - - -
49 - - - -
50 + - + -
51 - - - -
52 + - + -
53 + - + -
54 - - - -
55 + - + -
56 + - + -
57 - - - -
58 - - - -
59 - - - -
60 + - + -
61 Invalid
result
Invalid
result
- -
62 - - - -
63 - - - -
64 - - - -
65 - - - -
66 + - + -
67 - - - -
68 + - + -
69 - - - -
70 + - + -
71 - - - -
72 - - - -
73 - - - -
74 - - - -
75 + - + -
76 - - - -
77 + - + -
78 - - - -
79 - - - -
80 - - - -
81 + - + -
82 + - + -
83 - - - -
84 - - - -
85 - - - -
86 - - - -
87 - - - -
88 + - + -
89 - - - -
90 - - - -
35
91 + - + -
92 - - - -
93 + - + -
94 - - - -
All but one of the 94 samples were genotyped the same regardless method. The deviant sample,
sample 61, was classified as invalid on the QS but was classified as negative on the 7300 for
both targets, DQ2 and DQ8.
The reason for the invalid result lies in the evaluation method in use by Dynamic Code.
Sample 61 obtained a Ct-value of 33,715 for the IPC on QS, and 31,868 on 7300. The sample
was otherwise negative for all targets. A negative result for all targets in combination with an
IPC <32 makes the result invalid and a new sample has to be taken and analysed.
36
6. Discussion
Out of 16 primer and probe concentrations up for optimisation in this project, four were
ultimately changed: the IPC probe in mix 1 and mix 4, and two probe concentrations, connected
to mix 2 and mix 3, whereby one of them was cut in half and one was doubled. Over all,
changing the Master Mix together with the PCR instrument was possible without any major
complications, through primer and probe concentration optimisation and through adjustments
of the PCR thresholds. If thresholds were not adjusted, the existing evaluation protocol for
diagnosis of HLA would not be appliable and would have to be remodelled.
Generally, when calculating the median of the difference in Ct-values between QS and 7300,
values obtained from QS were higher for each target but the IPC in mix 4, yielding a positive
median. The mean of the medians was for all targets combined -0,140, meaning QS Ct-values
were higher. Lowering the threshold will automatically make curves intercept with it earlier,
resulting in lower Ct-values, thus solving the problem with the overall higher values of the QS.
When studying Table 11 and Table 12, containing the median differences between Ct-values,
QS – 7300, for each mix before and after optimisation, the differences are generally lower after
optimisation. The Ct-difference between the methods started off being 3,076 and were after
optimisation -0,140, meaning the project was successful in changing PCR instrument and
Master Mix without having the Ct-values change too much so as to still be able to use the
already elaborated patient diagnosis evaluation system.
The one sample that achieved different final HLA diagnosis, Sample 61, was classified as
invalid on the QS but got a negative test result on the 7300. The reason for it being invalidated
when analysed on the QS is because it was negative for all HLA-targets in combination with
having a Ct-value higher than 32 for the IPC (33,72). This combination will, according to the
evaluation premises, result in an invalid test result after which a new test has to be sampled and
analysed. This is because a test with a too low IPC signal is not consider reliable as it might be
an indication of PCR inhibitors or a faulty sampling and lack of DNA in the sample. Giving
false negative results is a risk when analysing such tests.
Sample 61 received a Ct-value of 31,87 on the 7300, which is right under the limit for what
is accepted for the IPC. Hence, the possibility for the sample to be invalidated on the 7300 was
not far off.
Although this sample generated different end results on the two instruments, in this
particular case, that would not result in a faulty diagnosis. This patient is clearly negative for
all DQ-targets, regarding method, but if analysed on the QS, the patient would have to redo the
test for confirmation reasons. Hence, no critical consequences would follow this error.
One limitation of the study is the fact that only one validation experiment per mix was
conducted. Even though sample quantity was quite large, 94 samples, the experiment would
have to be executed again to account for possible measurement aberrations due to stochastic
variations and changes in conditions. For example, reaction conditions might vary between runs
in where heating of the block holding the samples inside the instrument not always carries out
identically. Besides temperature variations between runs, variations in detection is possible,
depending on the strength of the excitation lamp in the instrument.
In addition, there are practical aspects in the laboratory work pre analysation to take into
consideration, where volume variations when pipetting might be the biggest source of error.
37
When handling such small volumes as 3 µl, a minor divergence in the volume might have a
massive effect on outcome.
Reagent quality is another factor not to be neglected that will compromise the results.
Inequality in reagent freshness depending on how the reagent was stored and for long it has
been stored will influence the outcome.
Repeating the experiments but with different conditions and circumstances is necessary in
the future as that will give a better overall view of how the method actually performs.
Conducting only one experiment will only tell you how the method performed in those specific
conditions under those exact circumstances.
38
7. Conclusion
Optimisation of the new method to enable PCR instrument transfer and Mater Mix change
through primer and probe concentration manipulation was considered successful. The goal of
not having existing patient diagnosis validation algorithms changed after the method change
was also reached.
Based on the validation experiments, HLA diagnosis results were the same regardless of method
in 98,9% of the cases. Only one out of 94 samples had conflicting results. It did not obtain
different diagnosis, as it was negative in both methods, but was invalidated in the new method
because of a too high IPC value.
Compared to the old method, the new method allows for 83% more samples to be analysed in
the same amount of time thanks to the instrument transfer. The change in Master Mix means a
decrease in reagent costs by 21% (comparing the current market price for a 5x1 ml batch
SolisFast® Probe qPCR Mix and TaqMan™ Gene Expression Master Mix, according to each
manufacturers’ websites at current date).
39
8. Acknowledgments
First, I want to thank Dynamic Code for giving me this opportunity, and especially every single
one of my co-workers in the laboratory, who made my time spent there memorable and who
were always open for discussion or support with whatever I encountered during this project.
I want to thank Josephine Overmyr, my supervisor, for taking me on during this project.
Thanks to Malin Farnebäck, who were involved in designing the project and who helped me to
come on stream with it from the beginning. Malin also provided much valuable input and
guidance throughout the project.
A special thanks to Maria Lundström, who taught me everything I needed to know regarding
both practical work in the laboratory, as well as theoretically, and who were always there for
me to answer any of my questions or concerns.
Thank you, Robin Östman, my opponent, for noteworthy comments and questions regarding
my work.
Thank you, Lars-Göran Mårtensson, my examiner, for valuable input and feedback on my
report.
Finally, thank you mother, for your never-ending support.
40
9. References
1. Basu, C. (2015). PCR Primer Design.
Humana Press. Methods in Molecular
Biology.
2. Brown, N. K., Guandalini, S., Semrad, C.,
& Kupfer, S. S. (2019). A Clinician’s Guide
to Celiac Disease HLA Genetics. American
Journal of Gastroenterology: October 2019 -
Volume 114 - Issue 10 - p 1587-1592.
3. Catassi, C., Kryszak, D., Louis- Jacques,
O., Duerksen, D. R., Hill, I., Crowe, S. E.,
Brown, A. R., Procaccini, N. J., Wonderly, B.
A., Hartley, P., Moreci, J., Bennett, N.,
Horvath, K., Burk, M., & Fasano, A. (2007).
Detection of Celiac Disease in Primary Care:
A Multicenter Case-Finding Study in North
America. American Journal of
Gastroenterology, Blackwell Publishing.
4. Cecilio, L. A., & Bonatto, M. W. (2015).
The Prevelance of HLA DQ2 and DQ8 in
Patients With Celiac Disease, in Family and
in General Population. ABCD, arq. bras. cir.
dig. vol.28 no.3
5. Chien, A., Edgar, D. B., & Trela, J. M.
(1976). Deoxyribonucleic acid polymerase
from the extreme thermophile Thermus
aquaticus. Journal of bacteriology, 127(3),
1550–1557.
6. Harris, E. (1998). A Low-Cost Approach to
PCR. Appropriate Transfer of Biomolecular
Techniques. Oxford University Press.
7. Ishino, S. & Ishino, Y. (2014). DNA
polymerases as useful reagents for
biotechnology - the history of developmental
research in the field. Frontiers in
microbiology, 5, 465.
8. Jabri, B. & Sollid, L. M. (2017). T cells in
celiac disease. Journal of immunology
(Baltimore, Md. : 1950) vol. 198,8 (2017):
3005-3014.
9. Johnson, G., Nolan, T., & Bustin, S. A.
(2013). Real-time quantitative PCR,
pathogen detection and MIQE. Methods Mol.
Biol. 943, 1–16.
10. Karell, K., Louka, A. S., Moodie, S. J.,
Ascher, H., Clot, F., Greco, L., Ciclitira, P.J.,
Sollid, L. M., & Partanen, J. (2003). HLA
types in celiac disease patients not carrying
the DQA1*05-DQB1*02 (DQ2) heterodimer:
results from the European Genetics Cluster
on Celiac Disease. Hum Immunol. 2003
Apr;64(4):469-77
11. Lindfors, K., Ciacci, C., Kurppa, K.,
Lundin, K. E. A., Makharia, G. K., Mearin,
M. L., Murray, J. A., Verdu, E. F., &
Kaukinen, K. (2019). Coeliac disease. Nature
Reviews Disease Primers 5:3.
12. Lorenz T. C. (2012). Polymerase chain
reaction: basic protocol plus troubleshooting
and optimization strategies. Journal of
visualized experiments: JoVE, (63), e3998.
13. Maddocks, S. & Jenkins, R. (2016).
Understanding PCR: A Practical Bench-Top
Guide. Academic Press.
14. Megiorni, F., Mora, B., Bonamico, M.,
Nenna, R., Di Pierro, M., Catassi, C., Drago,
S., & Mazzilli, M. C. (2007). A rapid and
sensitive method to detect specific human
lymphocyte antigen (HLA) class II alleles
associated with celiac disease. Clinical
Chemistry and Laboratory Medicine (CCLM)
| Volume 46: Issue 2.
15. Megiorni, F., Mora, B., Bonamicob, M.,
Barbatob, M., Nennab, R., Maiellab, G.,
Lullia, P., & Mazzilli, M. C. (2009). HLA-DQ
and risk gradient for celiac disease. Elsevier
41
Inc. on behalf of American Society for
Histocompatibility and Immunogenetics.
16. National Laboratory of Enteric Pathogens,
Bureau of Microbiology, Laboratory Centre
for Disease Control (1991). The polymerase
chain reaction: An overview and development
of diagnostic PCR protocols at the LCDC.
The Canadian journal of infectious diseases,
2(2), 89–91.
17. Pfaffl, M. W. (2010). The ongoing
evolution of qPCR. Methods, Volume 50,
Issue 4, April 2010, Pages 215-216.
18. Pietzak, M. M., Schofield, T. C.,
McGinniss, M. J., & Nakamura, R. M. (2009).
Stratifying Risk for Celiac Disease in a Large
At-Risk United States Population by Using
HLA Alleles. Clinical Gastroenterology &
Hepatology, September 2009, Volume 7 (9),
p 966–971.
19. Saiki, R., Bugawan, T., Horn, G. et al.
(1986). Analysis of enzymatically amplified β-
globin and HLA-DQα DNA with allele-
specific oligonucleotide probes. Nature 324,
163–166.
20. Schrader, C., Schielke, A., Ellerbroek, L.,
& Johne, R. (2012). PCR inhibitors –
occurrence, properties and removal. Journal
of Applied Microbiology, 113: 1014-1026.
21. Shen, C-m., Zhu, B-f., Deng, Y-j., Ye, S-
h., Yan, J-w., Yang, G., Wang, H-d., Qin, H-
x., Huang, Q-z., & Zhang, J-j (2010). Allele
Polymorphism and Haplotype Diversity of
HLA-A, -B and -DRB1 Loci in Sequence-
Based Typing for Chinese Uyghur Ethnic
Group. PLoS ONE 5(11): e13458.
22. Slack, G. S., Low, J., Kaminski, G., &
Yirrell, D. (2011). Optimising PCR reactions;
remember magnesium? Journal of Clinical
Virology, Volume 52, Issue 1, September
2011, Pages 63-64.
23. Tack, G. J., Verbeek, W. H. M., Schreurs,
M. W. J., & Mulder, C. J. J. (2010). The
spectrum of celiac disease: epidemiology,
clinical aspects and treatment. Nature
Reviews Gastroenterology & Hepatology
volume 7, pages 204–21.
24. Williams, T, M., (2001) Human
Leukocyte Antigen Gene Polymorphism and
the Histocompatibility Laboratory. Journal of
Molecular Diagnostics, Vol. 3, No. 3, August
2001.
25. Wolters, V. & Wijmenga, C. (2008)
Genetic background of celiac disease and its
clinicalimplications. The American Journal
of Gastroenterology 2008;103:190–5.
26. Wong, W., Farr, R., Joglekar, M.,
Januszewski, A., & Hardikar, A. (2015).
Probe-based Real-time PCR Approaches for
Quantitative Measurement of microRNAs.
Journal of visualized experiments : (98),
52586.
27. Yakovchuk, P., Protozanova, E., & Frank-
Kamenetskii, M. D. (2006). Base-stacking
and base-pairing contributions into thermal
stability of the DNA double helix. Nucleic
acids research, 34(2), 564–574.
42
Appendix A
1. PCR-plate Setups
Figure 20: PCR-plate setup for experiments when varying concentrations of one component. Concentration 1 in yellow,
concentration 2 in teal and concentration 3 in orange. Template DNA is identical between the different concentrations.
Figure 21: PCR-plate setup for the initial experiment when all four mixes were analysed. Mix 1 in pink, mix 2 in blue,
mix 3 in purple and mix 4 in green. The same samples and setup were analysed according to the original method as well
as with the new Master Mix on the new instrument.
43
Figure 22: PCR-plate setup for the validation experiments. One full plate of each mix was pipetted, above shows mix 1,
and analysed according to the original method as well as the new, optimised method.
44
Appendix B
1. PCR Amplification Plots – Optimisation experiments
Figure 23: Amplification plot for mix 2 DQA1_02 in 100 nM (red) and DQB1_02 300 nM (green), probe 1 and probe 2
both 200 nM.
Figure 24: Amplification plot for mix 2 DQA1_02 in 300 nM (blue) and DQB1_02 300 nM (green), probe 1 and probe 2
both 200 nM.
45
Figure 25: Amplification plot for mix 2 DQA1_02 in 900 nM (light green) and DQB1_02 300 nM (dark green), probe 1
and probe 2 both 200 nM.
Figure 26: Amplification plot for mix 2 DQB1_02 in 100 nM (red) and DQA1_02 300 nM (orange), probe 1 and probe 2
both 200 nM.
Figure 27: Amplification plot for mix 2 DQB1_02 in 300 nM (green) and DQA1_02 300 nM (orange), probe 1 and probe
2 both 200 nM.
46
Figure 28: Amplification plot for mix 2 DQB1_02 in 900 nM (blue) and DQA1_02 300 nM (orange), probe 1 and probe 2
both 200 nM.
Figure 29: Amplification plot for mix 2 probe 1 100 nM (red), DQA1_02 300 nM, DB1_02 300 nM (green), probe 2 200
nM.
Figure 30: Amplification plot for mix 2 probe 1 200 nM (lilac), DQA1_02 300 nM, DB1_02 300 nM (green), probe 2 200
nM.
47
Figure 31: Amplification plot for mix 2 probe 1 400 nM (blue), DQA1_02 300 nM, DB1_02 300 nM (green), probe 2 200
nM.
Figure 32: Amplification plot for mix 2 probe 2 in 100 nM (brown), DQB1_02 300 nM and DQA1_02 300 nM (blue),
probe 1 100 nM.
Figure 33: Amplification plot for mix 2 probe 2 in 200 nM (green), DQB1_02 300 nM and DQA1_02 300 nM (blue),
probe 1 100 nM.
48
Figure 34: Amplification plot for mix 4 probe 1 in 100 nM (yellow), 200 nM (orange) and 400 nM (grey), DQB1_0302 in
300 nM, IPC in 300 nM (hidden in the plot), probe 2 200 nM.
Figure 35: Amplification plot for mix 4 DQB1*0302 B in 100 nM (orange), IPC has been hidden in the plot, probe 1 and
probe 2 both 200 nM.
Figure 36: Amplification plot for mix 2 probe 2 in 400 nM (teal) and DQA1_02 300 nM (blue), probe 1 100 nM.
49
Figure 37: Amplification plot for mix 3 probe 2 in 100 nM (blue), 200 nM (grey) and 400 nM (orange), DQB1_0302 300
nM and DQA1_03 900 nM, probe 1 200 nM.
Figure 38: Amplification plot for mix 4 IPC 300 nM (purple), DQB1*0302 B 100 nM (orange), 300 nM (lilac) and 900
nM (blue), probe 1 and 2 200 nM.
Figure 39: Amplification plot for mix 4 probe 2 in 100 nM (yellow), IPC 300 nM and DQB1*0302 B 300 nM (orange),
probe 1 200 nM.
50
Figure 40: Amplification plot for mix 4 probe 2 in 200 nM (purple), IPC 300 nM and DQB1*0302 B 300 nM (orange),
probe 1 200 nM.
Figure 41: Amplification plot for mix 4 probe 2 in 400 nM (grey), IPC 300 nM and DQB1*0302 B 300 nM (orange),
probe 1 200 nM.
2. PCR Amplification Plots – Validation experiments
Figure 42: Amplification plot for DQA1_05 on ABI 7300.
51
Figure 43: Amplification plot for DQA1_05 on QS.
Figure 44: Amplification plot for IPC on ABI 7300.
Figure 45: Amplification plot for IPC on QS.
52
Figure 46: Amplification plot for DQA1_02 on 7300.
Figure 47: Amplification plot for DQA1_02 on QS.
Figure 48: Amplification plot for DQB1_02 on 7300.
53
Figure 49: Amplification plot for DQB1_02 on QS.
Figure 50: Amplification plot for DQA1_03 on 7300.
Figure 51: Amplification plot for DQA1_03 on QS.
54
Figure 52: Amplification plot for DQB1_0302 on 7300.
Figure 53: Amplification plot for DQB1_0302 on QS.
Figure 54: Amplification plot for DQB1*0302 B on 7300.
55
Figure 55: Amplification plot for DQB1*0302 B on QS.
Figure 56: Amplification plot for IPC on 7300.
Figure 57: Amplification plot for IPC on QS.