operations & maintenance manual - sutron corporation · pad at the site with three anchor bolts set...

13
Siphon Tipping Bucket 5600-0625 & 5600-0725 Operations & Maintenance Manual Part No. 8800-1206 Revision – C August 16, 2016

Upload: others

Post on 30-Jan-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

  • Siphon Tipping Bucket 5600-0625 & 5600-0725

    Operations & Maintenance

    Manual

    Part No. 8800-1206 Revision – C

    August 16, 2016

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 2

    TABLE OF CONTENTS

    1. Introduction ........................................................................................................................................ 3

    1.1 Overview ............................................................................................................................ 3

    2. Specifications ...................................................................................................................................... 3

    3. Installation .......................................................................................................................................... 4

    3.1 Setup & Wiring ................................................................................................................... 4

    3.2 Mounting Hole Pattern....................................................................................................... 5

    4. Theory of Operation ........................................................................................................................... 6

    5. Calibration .......................................................................................................................................... 7

    5.1 Laboratory Calibration........................................................................................................ 7

    5.2 Field Calibration ................................................................................................................. 8

    5.3 Maintenance ...................................................................................................................... 8

    5.4 Calibration Drip Setup ........................................................................................................ 9

    6. Ordering Guide ................................................................................................................................. 10

    7. Training & Warranty ......................................................................................................................... 11

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 3

    1. Introduction

    1.1 Overview

    This Tipping Bucket Rain Gage is used to measure rainfall volume and/or rate. Rain entering through a funnel assembly with an 8.08 inch (205mm) orifice passes through a debris filtering screen and is funneled into a siphon that regulates flow and into one side of the tipping bucket assembly inside the gage. The bucket tips when a given amount of water, determined by gage calibration, has been collected. The model 5600-0625-2 tips for each 0.2mm collected; the Model 5600-0625-5 tips for each 0.5mm; the Model 5600-0625-6 tips for each 0.01 inch (0.254 mm). As the bucket tips it causes a magnet to pass by a ruggedized reed switch, momentarily closing the switch. The tipping of the bucket brings a second bucket into position under the funnel, ready for filling. After the rain water is measured, it is directed into drain tubes that allow it to exit out holes in the base of the gage; the exit holes are 3/8 NPT to allow for tubing or screen fittings.

    Materials used in construction of the instrument are corrosion resistant. All external openings of the gage are protected from insects and foreign material by mesh screen.

    2. Specifications

    Type Tipping Bucket

    Sensitivity/Resolution 5600-0625-2 5600-0625-5 5600-0625-6

    1 tip per 0.2 mm 1 tip per 0.5 mm 1 tip per 0.254 mm (0.01 inch)

    Orifice size 205 mm (8.08 inch) dia.

    Accuracy ±2% from 25-500 mm/hr (1-19.7 in/hr)

    Bearings Sapphire jeweled bearings with hard stainless steel shaft

    Insect protection Mesh screens

    Capacity 700 mm/hr (27.5 inch/Hr) with ±5%

    Output 0.1 second switch closure

    Switch type Single reed switch

    Size Diameter Height

    228 mm (9 inch) dia. 482 mm (19 inch)

    Weight 3kg (6.6 lbs)

    Temperature range, operating 0° to 60°C (32°F to 140°F)

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 4

    3. Installation

    3.1 Setup & Wiring

    1. This instrument is thoroughly tested and fully calibrated at the factory and is ready for installation. If return to the factory is necessary please contact the Sutron Customer Service department at (703) 406-2800 for a Return Material Authorization Number.

    2. The gage must be mounted level and adequately protected from high winds. Errors of greater than 5% can occur due to rainfall blowing over the lip of the orifice.

    3. Install the rain gage at a distance away from surrounding obstructions two to four times the obstacle's height. Such obstacles may block the normal rainfall path into the gage or rainfall may blow off the obstructions and into the gage.

    4. The gage should be securely mounted by using the mounting holes provided by the mounting feet. A recommended method of mounting the gage is to secure the gage pour a small (12" square) concrete pad at the site with three anchor bolts set for the hole pattern of the mounting feet.

    5. Loosen the 3 screws attaching the outer housing of the rain gage to the tipping bucket mechanism. Rotate the top half clockwise and lift up to remove outer housing of rain gage.

    6. Using the bulls-eye level on the tipping bucket mechanism, level and securely mount the gage.

    7. Fish a two conductor cable through the PG gland in the base and connect the two conductor cables between the Switch thumb screw and the screw on the housing to the right of the switch. Do this by unscrewing both screws from housing to create a gap about the thickness of the wire, stripping 1/2" of insulation off of both wires, wrap the stripped part around each screw clockwise between the gap, tighten down both screws to clamp the wire in place. For SDI-12 connections use a 3 conductor cable in the PG gland to SDI-12 interface and 2 short pieces of wires connected to switch and TB in and GND.

    Figure 1: Tipping Bucket Output

    Figure 2: SDI-12 Output

    8. Pull extra cable back out the PG gland and tighten the gland to secure the cable in place.

    9. Place the cover back on making sure to align with the screws in the base. Make a slight count-clockwise rotation once the cover is on the base and secure it in place by tightening all three screws in the base.

    10. Be sure to remove any packing material and the bucket retainers from the instrument.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 5

    3.2 Mounting Hole Pattern

    Use three 5/16-18 bolts to secure tipping bucket to installation platform.

    Drain holes are 3/8” NPT tapped for optional outlet filters or drainage tubing.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 6

    4. Theory of Operation

    The 205 mm (8.08 inch) diameter orifice was chosen from a trade-off between the sampling area and the

    size of the tipping bucket mechanism. The large funnel at the orifice helps prevent evaporation inside the

    instrument.

    Precipitation enters the funnel inside the instrument and remains there until there is sufficient accumulation

    to trigger the siphon. This is about equal to 4mL worth of precipitation. Once the siphon is triggered, it

    releases all the water in the funnel at a constant rate and is directed to one of two tipping buckets. When

    one bucket fills, its weight tips the next bucket into position, and empties the first. At the same time, a .1

    second momentary contact closure occurs. This closure initiates an event accumulating device. The water

    drains through the bottom of the instrument.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 7

    5. Calibration

    5.1 Laboratory Calibration

    Model Number Calibration Grams per Tip

    5600-0625-2 & 5600-0725-2 0.2 mm of rain 6.62 grams

    5600-0625-5 & 5600-0725-5 0.5 mm of rain 16.54 grams

    5600-0625-6 & 5600-0725-6 0.254 mm (0.01”) of rain 8.40 grams

    1. To calibrate the tipping bucket rain gage, remove the outer cover by first loosening the three screws at the base. BE SURE THE GAGE IS LEVEL, using the built-in bulls-eye level as an indicator.

    2. To make the first rough adjustment, draw water into a B-D 20cc syringe and weigh out to grams indicated in the table above for the required model number on an ACCULAB (or equal) Model 121 precision weighing gage. Be sure to zero out the weight of the empty syringe before weighing the water. With the bucket adjustment screws located on the bottom of tipping bucket, adjusted all the way counter-clockwise, transfer the water from the syringe into one side of the bucket. Rotate the opposite side adjusting screw clockwise until the bucket tips. Repeat these steps for the second bucket.

    3. To make the final calibration adjustments, reassembly the cover to the tipping bucket and devise a

    source of water (Calibration Drip Setup, page 8) with a uniform flow rate of approximately 50mm/hr (2"/hr). Direct this flow onto the wall of the small funnel. After the gage has tipped several times and the buckets are wet, collect the water in two containers, one under each drain tube. (Styrofoam cups work well) Carefully insert the collection containers between tips and collect exactly 5 tips of each bucket. Remove the containers. Measure the amount of water in each container by weight using the ACCULAB weighing gage described above. (Be sure to zero out the weight of the empty collection containers) The total weight of both containers should be from 82.32 grams to 85.68 grams, or + 2% of 84 grams if calibrating for a -6 unit.

    4. Adjustments may be made to the stops located under the buckets. Raising the stop of the bucket will

    reduce the catch for that bucket and consequently lowering the stop will reduce the catch for that bucket and consequently lowering the stop will increase the catch. NOTE: The stop for a bucket is located under the opposite bucket. Make only small corrections at a time (1/4 turn per test). Repeat the calibration procedure until the gage is within specifications.

    5. The above procedure is for a rate of 50mm/hr (2"/hr). The gage may be calibrated at other rates of

    interest.

    6. After calibration, note date and place of calibration and reinstall the outer cover and screws. Be sure the cover is correctly lined up.

    7. If the gage has been removed from a field location for calibration be sure the gage is installed and

    leveled properly.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 8

    5.2 Field Calibration

    To check for calibration in the field, use a Nova Lynx, or equal, Rain Gauge Field calibrator Model 260-2595. An alternate is Allweatherinc, Model 60103.

    Bring at least 3 liters of roughly room temp water to the gage site for this test. Below is a table for Nova Lynx calibrator with 946 ml of water. Note that exceeding the instrument’s specified accuracy range with a very high simulated rain rate will result in an under counting of the rain. A simulated rain rate exceeding the specified accuracy range defeats the benefit of the siphon resulting in water continuously running into the buckets. With water continuously running in, a little bit of water is lost during every tip transition. This error is more pronounced with smaller bucket sizes since there are more tips for a given amount of water. You can see the result of exceeding the specified accuracy range when testing at 26 in/hr in the table below.

    Orifice size Inches/hour mm/hr 5600-0625-2 5600-0625-5 5600-0625-6 Test Time

    1/32” 2 50.8 140-146 tips 56-58 tips 110-114 tips ~35 min

    1/16” 6 152.4 140-146 56-58 110-114 13 min

    3/32” 12 304.8 140-146 56-58 110-114 6 min

    1/8” 26* 660.4* 128-142 54-56 104-108 2.5 min

    *Rain rate exceeds the specified calibration range.

    1. For a calibration check at 2"/hr rate, fill the bottle with room temperature water until the meniscus level reaches the center of the reference line of 946 ml. Note that Nova Lynx calibrator embossed line may not be in the correct location.

    2. Screw the orifice assembly onto the bottle and slip the holder over with the feet pointing upward.

    3. While holding the calibrator over the gage, flip the calibrator assembly so that water begins flowing out into the rain gage funnel. Do not allow any water to drip outside the funnel.

    4. Set the calibrator in the funnel shown in the diagram. The feet of the holder should position themselves firmly in the funnel, supporting the bottle nearly vertical above the funnel opening.

    5. In about half an hour the water stream should cease. Tap or shake the bottle slightly to make sure all the water is out.

    6. Repeat the above test about three times, recording the number of tips after each test. (Note: The number of tips during each test can vary ± 1 count or so because of the residual water left in one bucket and/or siphon at the end of the preceding test.

    7. For 2"/hr rain rate the number of tips should be from 110 to 114 for -6 model. If the values fall below or above, remove the gage from the field and recalibrate using the Laboratory Calibration Check. Note that the siphon will prevent some water from going through the system so your tip count may reduce by 1 depending on whether you start with the unit completely dry or not.

    5.3 Maintenance

    Maintenance is limited to cleaning debris out of the input orifice and output drains.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 9

    5.4 Calibration Drip Setup

    Water Container

    Tipping Bucket Mechanism

    Flow Meter -- Gilmont

    Instruments or Equivalent

    Adjusts Flow Rate

    F i g . 2

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 10

    6. Ordering Guide

    Sutron Tipping Buckets & Accessories 5600-0625-2 .2mm Tipping bucket

    5600-0725-2 .2mm Tipping bucket with SDI-12 output

    5600-0625-5 .5mm Tipping bucket

    5600-0725-5 .5mm Tipping Bucket with SDI-12 output

    5600-0625-6 .01 inch (0.254 mm) Tipping Bucket

    5600-0725-6 .01 inch (0.254 mm) Tipping Bucket with SDI-12 output

    SDI-TB-1 SDI-12 Interface electronics, comes standard on -0725-X units. You can field upgrade an -0625-X model into a -0725-X.

    9500-0209 8 inch Stainless Steel Filter screen

    6211-1024-1 2” Pipe mounting kit

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 11

    7. Training & Warranty

    Dear Customer:

    Thank you for making the important decision to purchase Sutron equipment. All Sutron equipment is

    manufactured and tested to the highest quality standards as set by Sutron’s Quality Assurance Department. Our

    Customer Service Representatives have years of experience with equipment, systems and services. We have

    electronic technicians with field and applications experience, not limited to technical school training.

    Sutron Equipment Repairs

    Sutron maintains a Repair Department at the Virginia factory (22400 Davis Drive, Sterling, VA 20164).

    Average turn-around time for repairs ranges from ten to thirty days after Sutron receives equipment for

    repair.

    Prior to returning any equipment for repair, please call Customer Service at (703) 406-2800 for a Return

    Material Authorization (RMA) number. Then send the faulty equipment back to the Virginia factory**.

    Sutron Customer Service will repair Sutron-manufactured equipment sent to us for repair within 30 days of

    the time we receive the item or that repair is free of charge.*

    Please provide our Customer Service Representative with your email address when receiving the RMA

    number so that we can email you when your equipment is received at Sutron and again when it is shipped

    back to you.

    Expedited repairs can be completed within one (1) week for an additional expedite fee of $200.00.

    Repaired equipment is warranted for a period of 180 days after the repair shipment date.***

    We appreciate your cooperation in achieving a quick turn-around by promptly providing cost approval

    decisions, Purchase Order and credit card information once we have supplied you with a repair estimate.

    Customer Phone Support: 8:00 am to 8:00 pm EST, Monday through Friday

    Customer Service Representatives routinely handle a wide variety of questions every day. When equipment

    questions arise, please feel free to contact me or one of Sutron’s Customer Service Representatives. We are

    available from 8:00 am to 8:00 pm EST Monday through Friday and are happy to take your calls. The main

    Customer Service number is (703) 406-2800, extension #5. After 5 pm EST Monday through Friday, please use

    extension #6.

    We typically answer most datalogger, transmitter, sensor and interface questions on the first call. If we cannot

    quickly answer a question on an interface, we will work with you until we find a solution.

    Sometimes a problem is application related. Although we pride ourselves on handling 95% of application related

    questions over the phone, we maintain constant contact with our Integrated Systems Division and Engineering

    Division for additional assistance.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 12

    On-Site Visits

    Of course not all problems can be fixed over the phone. Sometimes a customer needs an on-site technician to

    identify site related problems or troubleshoot a network. Sutron provides these services at a reasonable cost.

    Please call for details.

    Training

    Training is an important part of the Sutron Customer Service Success policy. With proper training, you will learn

    how to take advantage of all the benefits and tools that Sutron equipment provides. We are available for, and

    frequently provide, on-site introductory training at our Customers’ facilities. We also hold three (3) day training

    seminars approximately four times per year at our Sutron headquarters in Sterling (near Washington, DC).

    Contact me or your Sutron Regional Sales Manager for details.

    If you would like to learn more about Sutron products, please email [email protected]. Thank you again for

    your business,

    Paul Delisi, Customer Service Manager

    [email protected]

    * Note: The 30 day guarantee applies only to Sutron-manufactured products. The 30 day guarantee does not apply to

    the following: bulk shipments of 10 or more repair items, third party or buy-ship products, shipments awaiting

    payment authorizations, cases where delays are caused by international customs, shipping or other regulations.

    Additionally, the 30 day guarantee is exclusive of domestic and international transit times.*

    ** Transportation charges for equipment returns are the responsibility of the Customer.

    *** See Warranty .Any products repaired or replaced under Warranty will be warranted for the balance of the Warranty

    period or for a period of 180 days from the repair shipment date, whichever is greater.

  • Sutron Corporation Tipping Bucket Rev. C Operations & Maintenance Manual 8/17/2016 pg. 13

    COMMERCIAL WARRANTY

    SUTRON MANUFACTURED EQUIPMENT

    THE SUTRON CORPORATION WARRANTS that the equipment manufactured by its Manufacturing Division shall

    conform to applicable specifications and shall remain free from defects in workmanship and material for a

    period ending two years from the date of shipment from Sutron’s plant.

    Sutron’s obligation under this Warranty shall be limited to repair at the factory (22400 Davis Drive, Sterling, VA

    20164) or, at Sutron’s option, replacement of the defective product. In no event shall Sutron be responsible for

    incidental or consequential damages, whether or not foreseeable or whether or not Sutron has knowledge of

    the possibility of such damages. This Warranty shall not apply to products that have been damaged through

    negligence, accident, misuse or acts of nature such as floods, fires, earthquakes, lightning strikes, etc.

    Sutron’s liability, whether in contract or in tort, arising out of warranties or representations, instructions or

    defects from any cause, shall be limited exclusively to repair or replacement parts under the aforesaid

    conditions.

    Sutron requires the return of the defective electronic products or parts to the factory to establish claim under

    this Warranty. The customer shall pre-pay transportation charges to the factory. Sutron shall pay transportation

    charges for the return of the repaired equipment to the customer when the validity of the damage claim has

    been established. Otherwise, Sutron will pre-pay shipment and bill the amount to the customer. All shipments

    shall be accomplished by best-way surface freight.

    Sutron shall in no event assume any responsibility for repairs or alterations made other than by Sutron. Any

    products repaired or replaced under this Warranty will be warranted for the balance of the Warranty period or

    for a period of 180 days from the repair shipment date, whichever is greater. Products repaired at cost will be

    warranted for 90 days from the date of shipment.

    NON-SUTRON MANUFACTURED EQUIPMENT

    The above Warranty applies only to products manufactured by Sutron. Equipment provided, but not

    manufactured by Sutron, is warranted and will be repaired to the extent of and according to the current terms

    and conditions of the respective equipment manufacturers.

    EXTENDED WARRANTY AND ON-SITE MAINTENANCE

    Extended warranty and on-site maintenance contracts are available. Price quotations may be obtained from

    Sutron Customer Service Representatives.