BCM-600Electrocardiograph (Ver 1.0)
Operation Manual
Head Office/Factory 687-5, Sangoan-ri, Hongcheon-eup, Hongcheon-gun,Gangwon-do, Korea (zip.250-804)Tel : +82-33-434-9041 Fax : +82-33-434-9043
Seoul Office/R&D1406, Masters Tower, 553, Dohwa-dong, Mapo-gu,Seoul, Korea (zip.121-748)Tel : +82-2-714-2960∼2 Fax : +82-2-714-2963
Customer Service Dept.Tel : +82-2-714-2962 Fax : +82-2-714-2963
Address of Region Representative ET MEDICAL DEVICES SPAVIA DE ZINIS 6, 38011 CAVARENO (TN) ITALY Tel : +39 0463 85 01 25 Fax : +39 0463 85 00 88
OPERATION MANUAL
BCM-600Electrocardiograph (Ver 1.0)
(OCT. 26. 2010)
Table of Contents
1. Before using the product 3
2. Contacts 3
3. Product Guarantee 4
4. Symbols in the Manual 5
5. Precautions for use 6
6. Precations for Electrical Safety 9
7. Characteristics and Principles of BCM-600 10
8. Composition of BCM-600 11
9. How to set BCM-600 12
Precations for Settings 12
Checkpoints Before Use 12
Maintenance and Cleaning After Use 13
10. Components of BCM-600 14
Front & Operation Panel 14
Side & connection 15
11. Description of Product Symbols 16
Screen configuration 17
12. How to set BCM-600 19
How to use 20
13. MENU SETTING 20
EKG SETTING 20
Other Menu Setting 22
Default Set 25
15. PRINT 26
Before Using the Product 3
BCM-600
Before Using the Product
Contacts
Thank you for purchasing BCM-600
For safe and efficient operation, please read the manual to learn about the functions of the
product before use.
Please keep this manual nearby to assist you with operating the functions of the product.
The product should be used under the supervision of a medical- related certified professional.
The product is used for examining patient’s abnormal condition. For the safety of the patient, only
the components or the accessories recommended by our company should be used.
If you connect any accessories to the product that is not specified in the manual, please be sure to
notify our company or an agency authorized to distribute our products.
For various services and product information, please feel free to contact our salesperson at the
following number.
For any questions about use or setup of the product, please contact the following number.
Technical Support BIONIS Co.,Ltd. Tel: 82-2-714-2962
※ For reporting defects or abnormalities, please note the model name, product serial number,
purchasing date, and abnormal points before contacting us.
16. How to use RECORDS 28
17. Power & Battery Specifications 30
AC Power 30
Battery Specifications 30
18. EKG Measurement 32
How to connect ECG cable 32
Electrode Attachement 33
Electro Cardiography 33
19. Before Requesting After Sale Service 34
20. Product Specifications 36
21. Product Warranty 38
Electrocardiograph (Ver 1.0)
4 Product Guarantee Symbols in the Manual 5
BCM-600
Product Guarantee Symbols in the Manual
The product you have purchased comes with a 1-year warranty from the purchasing date
covering inferior materials and the manufacturing process. For the disposable products, a
warranty is given for 6 months from the purchasing date or 6 months from its first use covering
inferior materials and the manufacturing process. This warranty cannot be applied to
consumption supplies such as the battery, external cables, cradle and disposable accessories.
We are not responsible for any physical, financial and mental loss or damage and direct or
indirect expenses caused by, or resultant from, accidental or special circumstances during the
use of the product. The warranty is limited to the repair or exchange of the product by the
authority of the company or the agency, for problems in materials or the manufacturing process
which occurred during use under normal circumstances.
The company has no legal responsibility for the confirmation of services, declarations or whole
warranty beyond the warranty specified above, which were supplied or will be supplied by the
agency or the staff of the company at their own discretion. The user cannot request the company
of services, declarations or whole warranty which were supplied or will be supplied by the
agency or the staff of the company at their own discretion based on the statements above.
The contents of this warranty represent all other warranties indicated or implied, and also
represent all kinds of service- related responsibilities of the product sellers for the product they
sell.
The product was manufactured under strict quality controls and the inspection process of the
company. The indemnification standard for product repair and exchange is according to the
Economic Planning Board notification “Regulation for the indemnification of consumer
damage”
Some symbols are used in this manual to help users or patients use this product safely and
appropriately and to prevent any risk to patients or damage to materials. Please read and
understand all warnings and precautions.
The “Warning” symbol indicates that a dangerous situation may occurresulting in fatal injury or death of the patient, or may incur material andfinancial damage when the warning is ignored.
The “Caution”symbol indicates that a situation may occur resulting ininjury or damage to the patient, though not life-threatening, when thecaution is ignored.
The “Note”symbol is used to inform you of important, but not dangerous,information regarding installation, use, and maintenance of the product.
◈ Use of this product is prohibited during MRI photographing. A fire may breakout from the
induced current, and the accuracy of the product and the MRI may be influenced by cross
interference.
◈ All cables should be set properly. Cables and hoses etc. should be kept away from the
patient, so that there is no possibility of the patient’s neck being entangled in them, and the
cables and lines should be well arranged for the safety of the staff in the hospital.
◈ The computerized ECG analysis should be judged by qualified medical professionals, and
it should not be the sole basis in determining necessity of treatment for the patient.
Each symbols mean :
WARNING
CAUTION
NOTE
WARNING
Electrocardiograph (Ver 1.0)
6 Operational Do’s and Don’ts Operational Do’s and Don’ts 7
BCM-600
Operational Do’’s and Don’’ts
Do not operate or store the equipment under the following environments.
Locations exposed to chemicalor explosive gases.
Disassembling of equipmentshould be done only by authorizedpersonnel, otherwise we will not berequired to provide, any services.
When pulling out the power cord,be sure to hold the plug and pull,not the cord.
Do not plug in the power unit,until installation is completed,otherwise, damage may be causedto the equipment.
Make sure to prevent dust, especially metal debris, from penetrating in.
Avoid damp locations, and do not operate the equipment withwet hands.
Locations where the temperaturefluctuation is rather large.(Operational temperature range:10~45℃(C, Moisture level : 30~85%)
Locations where moisture levelincrease considerably or whereair is not ventilated properly.
Locations where sudden impactor vibration could be received.
Locations close to electricalheating apparatuses.
Locations where the machine isexposed to direct sunlight.
Standard operational conditions are as follows.Temperature : 10℃ ~ 45℃ (50 ~ 113℉)
Altitude : 70 ~ 106kpa
Humidity : 30 ~ 85%
Standard storage and transportation conditions are as follows.Temperature : -10℃ ~ 50℃ (14 ~ 122℉)
Altitude : 70 ~ 106kpa
Himidity : 20 ~ 95%
Electrocardiograph (Ver 1.0)
8 Operational Do’s and Don’ts Precautions for Electric Safety 9
BCM-600
◈ The product may not work well when it is kept or operated beyond the temperature and
humidity specified.
◈ If you want to turn off the BCM-600 while you’re operating, please use the power switch
where is the front of the product.
◈ When you keep or operate the product in large temperature difference, product
malfunctioning can be happened because of humidity of battery pack.
Please operate product after keep it room temperature for three hours.
◈ Leakage, heating, ignition or rupture of the battery may cause fire or injury.
Do not use batteries beyond those specified in this manual.
Do not short-circuit the battery and do not heat, disassemble, or dispose of it in a fire.
Do not mix using the used battery with new batteries , different kind of batteries, other
company batteries
Do not insert batteries with (+) and (-) ends reversed.
◈ When the product is not used for a long time, separate the battery from the product.
◈ To ensure the lifecycle and safety of battery, please discharge the battery and then charge
perfectly again every six months.
Also time-worn battery exceed 24 months should be replaced.
◈ When the product has been moved, make sure to turn it off and arrange the accessories in
order. Injured cords or cables may cause fire or electric shock.
Precautions for Electric Safety
To prevent electric noise during use, the product should be installed apart from dynamo, X-
ray equipment, broadcast equipment or portable cables. An inaccurate result may occur when
these equipment are placed near the product. BCM-600 requires an independent power circuit
and a stable earth connection. Sharing power with other electric equipment may result in
inexact results.
BCM-600 is classified as follows :
◈ Type 1 for electric shock, CF grade
◈ Please do not use this product near inflammable anesthetics and diluents.
◈ Calssification is Class IIb
◈ The noise level is “B” Class according to the IEC/EN 60601-1 (Safety of Electric Medical
Product), and the noise redemption is “B” Level according to the IEC/EN 60601-1-2
(Electromagnetic Compatibility Requirements)
◈ Protection grade against water is “IPX 0”.
Please check the following points before use.
Is the product supplied with the appropriate power? (100~240VAC)
Are all connecting parts (power cables or the product) properly connected?
Is the product completely grounded? (If not, noise may occur.)
Before turning the power on, is the accessory for measuring the target parameter well connected
to the product?
Electrocardiograph (Ver 1.0)
10 Characteristics and Principles of BCM-600 Composition of BCM-600 (Accessories) 11
BCM-600
Characteristics and Principles of BCM-600 Composition of BCM-600 (Accessories)
BCM-600 is a 12-channel ECG recorder, measuring the patient’s electrocardiogram. BCM-600
provides Auto-Diagnosis and the required parameters for the patient’s ECG records & Diagnosis.
In addition, it can be used effectively in ECG chart management with the printed paper including
the patient’s information and user’s information
Considering the convenience of users, it is available to measure ECG, to save data, to apply the
filter, to extract the parameters and to get Auto-Diagnosis by pressing the button only one time,
and then, it prints in 4inch sized paper
Also, it can be easily used for an emergency or transport situation by built-in-battery and portable
weight
Features
12 channel ECG waves are configurated with 3channels and 6channels. They can be printed in
4inch sized or 2 inch sized paper
BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis
calculated automatically in the report form.
BCM-600 provides the various diagnosis by Auto-Diagnosis
BCM-600 can be used in transport by the built-in-rechargeable battery.
BCM-600 can be used effectively in ECG chart management with the printed paper including
the patient’s information and user’s information
BCM-600 can manage about 30pcs of data and print again after amendment.
Standard Accessories
Main 1 setEKG Cable
10 lead type (1EA)
Limp
(4EA)
Chest electrode
(6EA/1set)
ECG Gel (1EA) Power Cord
Adapter(1EA)
15V 3Ah
Operation Manual
(1EA)
Roll Paper
(1EA)
Ground Cable
(1EA)
Electrocardiograph (Ver 1.0)
12 How to Set BCM-600 How to Set BCM-600 13
BCM-600
How to Set BCM-600
(3) Maintenance and Cleaning After UseA. Keep the product cleans by cleaning it using a soft cloth at least once a month. Do not use
thinner, ethylene, oxidant or lacquer since these may damage the product.
B. Keep the accessories clean of dust or foreign materials. For cleaning, use a soft cloth and
clinical alcohol.
C. Do not soak the accessories in liquid or detergents. Keep the product and accessories out of
any liquid.
D. For the protection of the environment, please contact the agency you purchased the product
from when you wish to discard the BCM-600
E. Our product can be used semipermanently. Please contact our Customer Service dept. or any
distributor and then replace the parts , when wasting parts such as battery, LCD or LED and
so on have a problem .
(1) Precautions for SettingsPay attention when setting BCM-600 for the following points
- Use BCM-600 under conditions of 10℃∼45℃ of circumstantial temperature and 30%∼85%
humidity.
- Check the connection status of the power cord.
- Do not connect multiple cords to the power supply.
- Place the main body on a flat area.
- If noise occurs, ground the equipment.
- Do not use any electric cords generating connecting noise.
- Take care since the product can be broken by mechanical shock.
- Get rid of dust or inflammable material near the product when setting it up.
(2) Checkpoints before UseCheck the following points before measuring the patient’s condition.
- Make sure there is no mechanical risk.
- Check the lead, power plug, and accessories connected on the outside.
Precautions Before Use① Do not use product near dynamo, X-ray equipment, broadcast equipment.
② Do not twist the cable.
③ Attach each Lead properly.
④ Keep the patient motionless with comfortable position while measuring
Electrocardiograph (Ver 1.0)
14 Components of BCM-600 Components of BCM-600 15
BCM-600
Components of BCM-600
(1) Front & Operation Panel (2) Side & Connection
(1) AC Power Button
(2) Screen Shift Button
(3) FREEZE Button
(4) GAIN Button
(5) PRINT START Button
(6) SPEED Button
(7) EXIT Button
(8) MENU Button
(9) LCD SCREEN
(10) PAPER LOCK
(11) PAPER Cover
(12) AC Power LED
(13) DC Power LED
(1) Connecting part with Patient cable
(2) USB Port
(3) RS-232(For Upgrade)
(4) LAN Port (For Upgrade)
(5) Power Adapter
(6) Potential equalization stud
Electrocardiograph (Ver 1.0)
16 Description of Product Symbols Description of Product Symbols 17
BCM-600
3. Description of Product Symbols(1) Screen Configuration
① WAVE Area : Display the graph of electrocardiogram
② Wave Property Area : Display the Speed and Size of electrocardiogram
③ Each measuring Mode icon : Display the status of measurement, printing and so on.(Refer the
following page (Table 1) Explanation of Screen Icons)
④ Time : Display the present time
⑤ HR : Display the Heart Rate of a Patient
(Fig1. Main Screen)
Symbol Comments Symbol Description
Type CF Product,Defibrillation protected
Power stand-by switch
Wave speed change
Attention, consult accompanying documents
Equipotentiality
Direct Current
Signal On/Off RS-232C Port
LAN Port
Screen Shift
Wave Freeze
Start Printing (Printer)
Wave size change
Initial Screen (Exit)
Signal Output
Use for Program Upgrade
Use for Central Monitoring
How to Use BCM-600 19
BCM-600Electrocardiograph (Ver 1.0)
18 Description of Product Symbols
(Table 1). Explanation of screen icons
Icon Meaning How to display
Limb Electrode (R) lead check
Limb Electrode (L) lead check
Limb Electrode (F) lead check
Chest Electrode (C1) lead check
Chest Electrode (C2) lead check
Chest Electrode (C3) lead check
Chest Electrode (C4) lead check
Chest Electrode (C5) lead check
Chest Electrode (C6) lead check
Auto diagnosis
Printing
Waveform stop
AC in use
Battery in use
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Normal (Green) Abnormal(Red)
Flickers at 1-second interval.
Flickers at 1-second interval.
Flickers at 1-second interval.
Fixed.
Fixed.
How to Use BCM-600
(1) How to Use
Step 1) Turn on the main power switch at the back of the product. Check the light-on status of the
AC LED in the front, and push the power switch 1~2 seconds to turn it on.(If any
problems occur, refer to the “How to solve the problems” section in the manual)
Step 2) If you want to change the setting value etc. during use, push the Menu button to change
the setting values per each measurement parameter
Step 3) When the measurement sensor is connected to the patient, the information of the patient
electrocardiogram is displayed as values and waveforms on the LCD screen
Push the Power button 1~2 seconds when turn it off and then “Power off” massage will be
displayed.
R
L
F
C1
C2
C3
C4
C5
C6
All of the Menu can be controlled by Menu button.The way to control is as below
Step 1) To change the default setting, press the Menu button once, then you can see the below
display
After moving to Menu you want to change, by rolling to left or right the Menu button, you can
select the Menu by press (the triangle on the right side of Menu means that there are more
menu)
Step 2) Each Menu can be changed by the above mentioned way
(by rolling to right or left the menu button and press)
(1) EKG setting
MENUMENU ▶
PRINT ▶
PATIENT ▶
SETUP ▶
EXIT
MENUMENU ▶
PRINT ▶
PATIENT ▶
SETUP ▶
EXIT
EKGLEAD ▶
SPEED ▶
GAIN ▶
FILTER ▶
▶
▶
▶
MUSCLEBASELINENOTCHReturn
Electrocardiograph (Ver 1.0)
20 MENU Setting MENU Setting 21
BCM-600
MENU SettingEKG Lead change
▶ To change the EKG Lead Wave setting, Roll the menu buttons to right or left and
select by press it
Lead of BCM-600 provides the 4 kinds of screen
EKG channel conversion can be done by press the Lead key
- 4 different configurations are as below;
MENUMENU ▶
PRINT ▶
PATIENT ▶
SETUP ▶
EXIT
EKGLEAD ▶
SPEED ▶
GAIN ▶
FILTER ▶
▶
▶
▶
MUSCLEBASELINENOTCHReturn
LEAD- 3X1- 3X2- 6X1- 6X2Return
3 X 1 3 X 2
6 X 1 6 X 2
Electrocardiograph (Ver 1.0)
22 MENU Setting MENU Setting 23
BCM-600
(2) Other Menu setting
▶ Please refer the below table for the other menu
◈ The choice of the filter may distort the original signal Therefore, you’d better use
appropriately only in case the recorded signal is bad so that the distortion of ECG signal can
be reduced
And we recommend you to apply the Baseline filter and Notch Filter all the time and to
apply appropriately Muscle filter
150Hz of filter is recommend in case of printing Diagnosis
3 x 13 x 26 x 16 x 2
Select the Configuration type ofEKG wave to be displayed
(same way in printing)
Changeable Range DescriptionEKG Menu
LEAD
12.5 mm/sec25.0 mm/sec50.0 mm/sec
Select EKG Wave processingspeed
SPEED
5 mm/mV10 mm/mV15 mm/mV
Select EKG Wave SizeGAIN
40 Hz100 Hz150 Hz
OFF
Select EKG FilterFILTER
ONOFF
Muscle Artifact filter settingMUSCLE
ONOFF
Baseline filter settingBASELINE
50 Hz60 HzOFF
If there is input, the eliminationfrequency is set.
NOTCH
MONITORDIAGNOSIS
FREEZE
MONITOR : Patient’sinformation, WAVE display
DIAGNOSIS : Patient’sinformation, WAVE, Diagnosis
informationFREEZE : Patient’s information,
WAVE output
Changeable Range DescriptionPRINT Menu
MODE
2 ~ 9 sec. Time control for the WAVE printWAVE SIZE
1 ~ 3 stageControl the Output Density(1-Dull,2-Normal,3-Dark)
BRIGHTNESS
30 pcsRefer the How to use Record on
page 27RECORDS
◈ The print processing speed, size, and channel settings comply with the setting value on
the LCD screen.
To print what are on the screen, as they are, it makes users to get more accurate
judgments
In order to change the print processing speed, size and cannel setting, Press the Menu
button to EKG menu, where you can change the setting value
◈ on Freeze in Print Mode, you can print as they are ,after checking the waves on the screen
◈ In case of no paper, "PAPER OUT" appears and you can not print
Open the printer paper cover 4 inches of the paper to be printed facing up, please insert.
Perforations, past the end of the paper should be positioned.
◈ Do not print greater than consecutive 3 times
◈ There is possibility of fire,
Do not use printing for 5min. after consecutive 3 times of printing .
Electrocardiograph (Ver 1.0)
24 MENU Setting MENU Setting 25
BCM-600
Changeable Range DescriptionPRINT Menu
NAME
ID
AGE
WEIGHT
HEIGHT
Within 16 characters
Within 16 characters
OFF, 0 ~ 110 age
OFF, 0~300Kg
OFF, 0~250cm
Patient’s name setting
Patient’s Serial no. setting
Patient’s age setting
Patient’s weight setting
Patient’s Height setting
SEXFEMALE
MALEOFF
Patient’s gender setting
Changeable Range DescriptionSETUT Menu
HOSPITAL
DATE&TIME
Initialize
LANGUAGE
Within 16 characters
Year,Month,Day,Hour,Min.
OFF
ON
English
Hospital name setting
Date/time setting
Initialization
(Default Set referrence)
DEMO MODE setting
Language Select
DEMO
(3) Default setting
▶ The default settings of the values can be used, without the need of changing the setting values
of each parameter.
LEAD
SPEED
GAIN
FILTER
MUSCLE
BASELINE
NOTCH
MODE
WAVE SIZE
BRIGHTNESS
NAME
ID
AGE
WEIGHT
HEIGHT
SEX
HOSPITAL
DEMO
LANGUAGE
- 6 x 2
12.5 mm/sec
10 mm/mV
40 Hz
OFF
OFF
50 Hz
MONITOR
3 sec
2(Normal)
PATIENT
ID-001
OFF
OFF
OFF
OFF
BIONICS
OFF
ENGLISH
Item Name NeonateParameter
EKG
SETUP
PATIENT
Electrocardiograph (Ver 1.0)
26 PRINT PRINT 27
BCM-600
◈ In case of no paper, "PAPER OUT" appears and you cannot print
Open the printer paper cover 4 inches of the paper to be printed facing up, please insert.
Perforations, past the end of the paper should be positioned.
◈ Do not print greater than consecutive 3 times.
There is possibility of fire, do not use printing for 5min. after consecutive 3 times of
printing .
◈ When you save a printed data, don't remove an USB Flash Memory until the save progress
is complete.
◈ Please use an USB Flash Memory listed in 'BCM-600 Viewer Manual'. If not, it may
cause malfunction of the product.
The print processing speed, size, and channel settings comply with the setting value on the LCD
screen. To print what are on the screen, as they are, it makes users to get more accurate judgments.
In order to change the print processing speed, size and cannel setting, press the Menu button to
EKG menu, where you can change the setting value.
If you use an USB Flash Memory, you can save a printed data to an USB Flash Memory. When
you save a printed data to an USB Flash Memory, print information will not be printed. (You can
use it in case of no paper)
For more information about USB SAVE, please refer to the PC Program manual provided by our
company.
Please refer the below table for the contents of other PRINT Menu.
PRINT Menu Changeable Range Description
MODEMONITOR
DIAGNOSISFREEZE
Print Menu SettingMONITOR : Patient's information,
WAVE displayDIAGNOSIS : Patient's information,
WAVE, Diagnosis informationFREEZE : Patient's information,
WAVE output
WAVE SIZE 2 ~ 9 sec. Time control for the WAVE print
BRIGHTNESS 1 ~ 3 stageControl the Output Density(1-Dull,2-Normal,3-Dark)
Refer the How to use Record on page 27RECORDS 30 pcs
Electrocardiograph (Ver 1.0)
28 How to use RECORDS How to use RECORDS 29
BCM-600
How to use RECORDS
Select RECORDS on Print Menu by
press Menu button
1. Press Menu button after moving toMenu wanted to select by rolling theMenu button to right or left
2. In order to change the print properties,press the Menu button after moving tothat properties by rolling the Menubutton to right or left
3. To print, select Print Menu.And to move to no.1, select “CANCEL”
4. Press EXIT button to close Menu
MENUMENU ▶
PRINT ▶
PATIENT ▶
SETUP ▶
EXIT
PRINTMODE ▶
WAVE SIZE ▶
BRIGHTHESS ▶
RECORDS ▶
Return
(1) The data entry area already printed
(2) Print Key
(3) Cancel Key
(4) Print properties
◈ BCM-600 has stored the late 30pcs of printed data
Electrocardiograph (Ver 1.0)
30 Power & Battery Specifications Power & Battery Specifications 31
BCM-600
◈ When the battery is removed, all data can be reset. Only in case the product is kept for long
time, Remove the battery from the product
◈ For the protection of the environment, please contact the agency you purchased theproduct from when you wish to discard the battery.
◈ Contact the agency you purchased battery in case battery should be changed.◈ To ensure the lifecycle and safety of battery, please discharge the battery and then charge
perfectly again every six months. Also time-worn battery exceed 24 months should be replaced.
C. Natural lifetime for spontaneous discharge of battery
The table below shows the lifetime for spontaneous discharge of the battery when the
product is unused.
D. Type of Battery
The same sort of battery should be used for replacement.
- Model : 3S2P(LS18650)
- Type : LI-Ion Battery
E. Time for the Battery exchange
The lifetime of the battery gets shorter if it is not 100% recharged for use
9 months
6 months
3 months
0℃ ~ 20℃
21℃ ~ 30℃
31℃ ~ 45℃
Charge / Discharge ration Number for use
1200 times
500 times
200 times
100 % Charge / discharge
50 % Charge / discharge
30 % Charge / discharge
Power & Battery Specifications
◈ During operation using AC power, noise may occur in the waveform. If this is the case,ground the product soundly.
◈ An AC power connection in a wet or damp area may cause electric shock or fatal damageto the product.
◈ When you connect the adapter will charge the battery automatically
2. Battery SpecificationsWhen the AC power is disconnected and the switch is turned on, the power is supplied using the
battery, signaled by the orange light on the battery LED.
Both AC and DC power is used for BCM-600. Basically, AC power is used and a rechargeable
battery is available for portable use.
1. AC Power (110~240VAC)If the AC power is connected to the main body of the product, a light-green color lights up on the
AC LED on the front. If the product is not being used, the power changes automatically to the
recharging mode.
A. Signs for Battery
A. Type of Adaptor
- Model : MW160 (KA1503F02)
- Input : 100~240V(AC), 50~60Hz- Output : 15V(DC), 3Ah
- Manufacturer: BridgePower Corp
Battery status Explanation
B. Time for 1 ea of battery
Recharging time : 9 H(OP Mode), 6H (Power Off Mode)
Use time :240 minutes
Battery full
After use for 50 ~ 60 minutes
Requiring recharging
Low battery requiring recharging
Storage temperature Life for spontaneous discharge
Electrocardiograph (Ver 1.0)
32 EKG Measurement EKG Measurement 33
BCM-600
EKG Measurement
A. How to Connect the ECG Cable
◈ With the use of an electric pacemaker, do not touch the patient, table, or other equipment.
◈ Check if the ECG cable is damaged before taking the ECG measurement. (A damaged
ECG cable cannot protect the patient.)
◈ Do not use disposable electrodes. (They may cause noise.)
When the patient is on stable condition, lay him down on the bed.
Then spread EKG gel all over the parts where the electrodes are to be attached.
Please confirm the EKG signal on LCD screen after attaching the patient electrodes to the patient
Please rub the skin with the gel, the alcohol or the water before measuring the patient
Attach the patient’s electrode to the sites shown below.
B. Electrode AttachmentCheck the status of the electrode attachment
When you measure a patient's condition
- Make sure if there are mechanical hazards or not
- Check the cable connected to External and accessories.
- Check all of the measuring instruments to measure the patient
C. Electro CardiographyAfter attaching electrodes to the patient, Turn the power switch
Set the filter, Wave size, printing speed, channel composition, printing time, diagnosis.
Input the patient information
Check the state of ECG signal and the state of Lead through the LCD screen.
Check the electrode connection if the signal is in bad condition
BCM-600 records the patient’s ECG by press printer button
BCM-600 analyzes and prints the cumulated ECG data for 10 seconds.
“Please wait” will be appeared in case the ECG data are cumulated within 10seconds.
Press printer button in order to stop working for ECG analysis or to stop printing
V1 = C1
V2 = C2
V3 = C3
V4 = C4
V5 = C5
V6 = C6
RA = R
LA = L
R.L = N
L.L = F
Before Requesting After-Sales Service 35
BCM-600
Before Requesting After-Sales Service
This section shows how to treat simple problems encountered during use.
Please check the following points before requesting after-sales service
▣ Problems in Display
Sign How to treat
▣ Problems in Measurement ECG
Sign How to treat
The ECG waveform is notdisplayed.
Electrocardiograph (Ver 1.0)
34 Before Requesting After-Sales Service
▣ Problems in Printer (OPTION)
Sign How to treat
1. Press the power button during 1~2 seconds 2. Check if “Power off” message is shown in display.
1. Check if there is any equipment with a strong magnetismaround the product.
2. Check if the cable connection is stable.3. Check if the product is grounded.
The power of the product isnot turned off.
The waveform is not normal.
The power of the productturns off suddenly during use.
The power of the product isnot turned on.
1. Check if the AC power is supplied..2. Press and hold the power button for 2 seconds
1. Check if the power is off due to battery release.2. Check if AC power is supplied to the product.
1. Check if the electrode attached to the patient is stable.
1. Check if the cable connection is stable.
2. Check if the ECG cable is damaged.
Severe noise on thewaveform.
No printing
1. Check if the printer icon blinks on the screen.
2. Check if the paper is set normally.
3. Check if the printer door is closed.
Noise during printing
1. Check if the roller is contaminated with foreign materials
or if the gear is out of position.
2. Check if the problem is due to a paper jam (change the
paper)
Electrocardiograph (Ver 1.0)
36 Product Specifications Product Specifications 37
BCM-600
Product Specifications
1) Features
- LCD Graphic Display 800 X 480 dots, 3x1/3x2/6x1/6x2
- Channel with base line movement possibility
- Standard 6 channels
- Real time heart rate display
- Real time display and print out
- Various patient information
- (Patient Name, Sex, Age, Weight, Height)
- High Resolution Thermal Printer
- Defibrillator protection
- Rechargeable Battery Operation
2) Specifications
- Diagnostic Electrocardiography
Input circuit : Floating input
Leads : Standard 12 leads
Input impedance : 20Mohm min
Input circuit current : 0.05 A max
Calibration voltage : 1mV with ± 2%
CMR : 100dB or more
Polarization voltage : ± 350mV min
Time constancy : 3.2 seconds (0~+20%)
Sensitivity : 5, 10, 15 mm/mV ± 5%
Frequency response : 0.05 ~ 100Hz (13dB)
Filters : AC50/60 Hz, -20dB or lower, Muscle tremor 35 to 45 Hz, -3db(-6db/oct)
A/D conversion : 12bits
Sampling rate : 2msec
3) Physical Specifications
- Power requirements : AC 100 ~ 240V, 50/60Hz
- Power consumption : 50VA(MAX)
- Battery: LI-Ion , 11.1V , 4400mAh
- Continuous operating time with battery : Max 4Hours
- Dimension : 327(W) X 206(L) X 87(H) mm (W x D x H)
- Weight : 1.75Kg(Incl. battery)
4) Interface
- RS-232C port
- LAN Port (Program upgrade)
- USB Port (Data storage)
5) Operating Environment
- Temperature : (Operating : 10℃ ~ 45℃ / Storage : -10℃ ~ 50℃)
- Humidity : (Operating : 30 ~ 85% / Storage : 20 ~ 95%)
- Atmospheric Pressure : (Operating : 70 to 106 kPa / Storage : 70 to 106 kPa)
6) Product specification listed on the manual are subject to change according to the our
design change procedure without notice.
Electrocardiograph (Ver 1.0)
Product Warranty
Product Name
Model Code
Manufacture Approval No.
Manufacture Approval Date
Manufacture Serial No.
Warranty Period
Purchase Date
Name ofHospital
Address
Name
Tel
Thank you for using BCM-600
This product is a medical device.
Only the products that have Passed our extensive and thorough quality tests are
offered to our customers
Our policies on repairing, exchange and refund complies with “Regulation on
Consumer Protection & Compensation Standards” issued by Economic
Planning Board of Korea
Name of Distributor
Name of Manufacture
CustomerInformation
Electrocardiograph
BCM-600
No. 1936
One year after the purchase date
Date Month Year
Tel) Fax)
BIONICS Co., Ltd.