BCM- 600 Electrocardiograph ( Ver 1.0 ) Operation Manual Head Office/Factory 687-5, Sangoan-ri, Hongcheon-eup, Hongcheon-gun, Gangwon-do, Korea (zip.250-804) Tel : +82-33-434-9041 Fax : +82-33-434-9043 Seoul Office/R&D 1406, Masters Tower, 553, Dohwa-dong, Mapo-gu, Seoul, Korea (zip.121-748) Tel : +82-2-714-29602 Fax : +82-2-714-2963 Customer Service Dept. Tel : +82-2-714-2962 Fax : +82-2-714-2963 Address of Region Representative ET MEDICAL DEVICES SPA VIA DE ZINIS 6, 38011 CAVARENO (TN) ITALY Tel : +39 0463 85 01 25 Fax : +39 0463 85 00 88

Upload: others

Post on 14-Aug-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

BCM-600Electrocardiograph (Ver 1.0)

Operation Manual

Head Office/Factory 687-5, Sangoan-ri, Hongcheon-eup, Hongcheon-gun,Gangwon-do, Korea (zip.250-804)Tel : +82-33-434-9041 Fax : +82-33-434-9043

Seoul Office/R&D1406, Masters Tower, 553, Dohwa-dong, Mapo-gu,Seoul, Korea (zip.121-748)Tel : +82-2-714-2960∼2 Fax : +82-2-714-2963

Customer Service Dept.Tel : +82-2-714-2962 Fax : +82-2-714-2963

Address of Region Representative ET MEDICAL DEVICES SPAVIA DE ZINIS 6, 38011 CAVARENO (TN) ITALY Tel : +39 0463 85 01 25 Fax : +39 0463 85 00 88

Page 2: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

OPERATION MANUAL

BCM-600Electrocardiograph (Ver 1.0)

(OCT. 26. 2010)

Page 3: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Table of Contents

1. Before using the product 3

2. Contacts 3

3. Product Guarantee 4

4. Symbols in the Manual 5

5. Precautions for use 6

6. Precations for Electrical Safety 9

7. Characteristics and Principles of BCM-600 10

8. Composition of BCM-600 11

9. How to set BCM-600 12

Precations for Settings 12

Checkpoints Before Use 12

Maintenance and Cleaning After Use 13

10. Components of BCM-600 14

Front & Operation Panel 14

Side & connection 15

11. Description of Product Symbols 16

Screen configuration 17

12. How to set BCM-600 19

How to use 20

13. MENU SETTING 20

EKG SETTING 20

Other Menu Setting 22

Default Set 25

15. PRINT 26

Page 4: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Before Using the Product 3

BCM-600

Before Using the Product

Contacts

Thank you for purchasing BCM-600

For safe and efficient operation, please read the manual to learn about the functions of the

product before use.

Please keep this manual nearby to assist you with operating the functions of the product.

The product should be used under the supervision of a medical- related certified professional.

The product is used for examining patient’s abnormal condition. For the safety of the patient, only

the components or the accessories recommended by our company should be used.

If you connect any accessories to the product that is not specified in the manual, please be sure to

notify our company or an agency authorized to distribute our products.

For various services and product information, please feel free to contact our salesperson at the

following number.

For any questions about use or setup of the product, please contact the following number.

Technical Support BIONIS Co.,Ltd. Tel: 82-2-714-2962

※ For reporting defects or abnormalities, please note the model name, product serial number,

purchasing date, and abnormal points before contacting us.

16. How to use RECORDS 28

17. Power & Battery Specifications 30

AC Power 30

Battery Specifications 30

18. EKG Measurement 32

How to connect ECG cable 32

Electrode Attachement 33

Electro Cardiography 33

19. Before Requesting After Sale Service 34

20. Product Specifications 36

21. Product Warranty 38

Page 5: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

4 Product Guarantee Symbols in the Manual 5

BCM-600

Product Guarantee Symbols in the Manual

The product you have purchased comes with a 1-year warranty from the purchasing date

covering inferior materials and the manufacturing process. For the disposable products, a

warranty is given for 6 months from the purchasing date or 6 months from its first use covering

inferior materials and the manufacturing process. This warranty cannot be applied to

consumption supplies such as the battery, external cables, cradle and disposable accessories.

We are not responsible for any physical, financial and mental loss or damage and direct or

indirect expenses caused by, or resultant from, accidental or special circumstances during the

use of the product. The warranty is limited to the repair or exchange of the product by the

authority of the company or the agency, for problems in materials or the manufacturing process

which occurred during use under normal circumstances.

The company has no legal responsibility for the confirmation of services, declarations or whole

warranty beyond the warranty specified above, which were supplied or will be supplied by the

agency or the staff of the company at their own discretion. The user cannot request the company

of services, declarations or whole warranty which were supplied or will be supplied by the

agency or the staff of the company at their own discretion based on the statements above.

The contents of this warranty represent all other warranties indicated or implied, and also

represent all kinds of service- related responsibilities of the product sellers for the product they

sell.

The product was manufactured under strict quality controls and the inspection process of the

company. The indemnification standard for product repair and exchange is according to the

Economic Planning Board notification “Regulation for the indemnification of consumer

damage”

Some symbols are used in this manual to help users or patients use this product safely and

appropriately and to prevent any risk to patients or damage to materials. Please read and

understand all warnings and precautions.

The “Warning” symbol indicates that a dangerous situation may occurresulting in fatal injury or death of the patient, or may incur material andfinancial damage when the warning is ignored.

The “Caution”symbol indicates that a situation may occur resulting ininjury or damage to the patient, though not life-threatening, when thecaution is ignored.

The “Note”symbol is used to inform you of important, but not dangerous,information regarding installation, use, and maintenance of the product.

◈ Use of this product is prohibited during MRI photographing. A fire may breakout from the

induced current, and the accuracy of the product and the MRI may be influenced by cross

interference.

◈ All cables should be set properly. Cables and hoses etc. should be kept away from the

patient, so that there is no possibility of the patient’s neck being entangled in them, and the

cables and lines should be well arranged for the safety of the staff in the hospital.

◈ The computerized ECG analysis should be judged by qualified medical professionals, and

it should not be the sole basis in determining necessity of treatment for the patient.

Each symbols mean :

WARNING

CAUTION

NOTE

WARNING

Page 6: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

6 Operational Do’s and Don’ts Operational Do’s and Don’ts 7

BCM-600

Operational Do’’s and Don’’ts

Do not operate or store the equipment under the following environments.

Locations exposed to chemicalor explosive gases.

Disassembling of equipmentshould be done only by authorizedpersonnel, otherwise we will not berequired to provide, any services.

When pulling out the power cord,be sure to hold the plug and pull,not the cord.

Do not plug in the power unit,until installation is completed,otherwise, damage may be causedto the equipment.

Make sure to prevent dust, especially metal debris, from penetrating in.

Avoid damp locations, and do not operate the equipment withwet hands.

Locations where the temperaturefluctuation is rather large.(Operational temperature range:10~45℃(C, Moisture level : 30~85%)

Locations where moisture levelincrease considerably or whereair is not ventilated properly.

Locations where sudden impactor vibration could be received.

Locations close to electricalheating apparatuses.

Locations where the machine isexposed to direct sunlight.

Standard operational conditions are as follows.Temperature : 10℃ ~ 45℃ (50 ~ 113℉)

Altitude : 70 ~ 106kpa

Humidity : 30 ~ 85%

Standard storage and transportation conditions are as follows.Temperature : -10℃ ~ 50℃ (14 ~ 122℉)

Altitude : 70 ~ 106kpa

Himidity : 20 ~ 95%

Page 7: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

8 Operational Do’s and Don’ts Precautions for Electric Safety 9

BCM-600

◈ The product may not work well when it is kept or operated beyond the temperature and

humidity specified.

◈ If you want to turn off the BCM-600 while you’re operating, please use the power switch

where is the front of the product.

◈ When you keep or operate the product in large temperature difference, product

malfunctioning can be happened because of humidity of battery pack.

Please operate product after keep it room temperature for three hours.

◈ Leakage, heating, ignition or rupture of the battery may cause fire or injury.

Do not use batteries beyond those specified in this manual.

Do not short-circuit the battery and do not heat, disassemble, or dispose of it in a fire.

Do not mix using the used battery with new batteries , different kind of batteries, other

company batteries

Do not insert batteries with (+) and (-) ends reversed.

◈ When the product is not used for a long time, separate the battery from the product.

◈ To ensure the lifecycle and safety of battery, please discharge the battery and then charge

perfectly again every six months.

Also time-worn battery exceed 24 months should be replaced.

◈ When the product has been moved, make sure to turn it off and arrange the accessories in

order. Injured cords or cables may cause fire or electric shock.

Precautions for Electric Safety

To prevent electric noise during use, the product should be installed apart from dynamo, X-

ray equipment, broadcast equipment or portable cables. An inaccurate result may occur when

these equipment are placed near the product. BCM-600 requires an independent power circuit

and a stable earth connection. Sharing power with other electric equipment may result in

inexact results.

BCM-600 is classified as follows :

◈ Type 1 for electric shock, CF grade

◈ Please do not use this product near inflammable anesthetics and diluents.

◈ Calssification is Class IIb

◈ The noise level is “B” Class according to the IEC/EN 60601-1 (Safety of Electric Medical

Product), and the noise redemption is “B” Level according to the IEC/EN 60601-1-2

(Electromagnetic Compatibility Requirements)

◈ Protection grade against water is “IPX 0”.

Please check the following points before use.

Is the product supplied with the appropriate power? (100~240VAC)

Are all connecting parts (power cables or the product) properly connected?

Is the product completely grounded? (If not, noise may occur.)

Before turning the power on, is the accessory for measuring the target parameter well connected

to the product?

Page 8: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

10 Characteristics and Principles of BCM-600 Composition of BCM-600 (Accessories) 11

BCM-600

Characteristics and Principles of BCM-600 Composition of BCM-600 (Accessories)

BCM-600 is a 12-channel ECG recorder, measuring the patient’s electrocardiogram. BCM-600

provides Auto-Diagnosis and the required parameters for the patient’s ECG records & Diagnosis.

In addition, it can be used effectively in ECG chart management with the printed paper including

the patient’s information and user’s information

Considering the convenience of users, it is available to measure ECG, to save data, to apply the

filter, to extract the parameters and to get Auto-Diagnosis by pressing the button only one time,

and then, it prints in 4inch sized paper

Also, it can be easily used for an emergency or transport situation by built-in-battery and portable

weight

Features

12 channel ECG waves are configurated with 3channels and 6channels. They can be printed in

4inch sized or 2 inch sized paper

BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis

calculated automatically in the report form.

BCM-600 provides the various diagnosis by Auto-Diagnosis

BCM-600 can be used in transport by the built-in-rechargeable battery.

BCM-600 can be used effectively in ECG chart management with the printed paper including

the patient’s information and user’s information

BCM-600 can manage about 30pcs of data and print again after amendment.

Standard Accessories

Main 1 setEKG Cable

10 lead type (1EA)

Limp

(4EA)

Chest electrode

(6EA/1set)

ECG Gel (1EA) Power Cord

Adapter(1EA)

15V 3Ah

Operation Manual

(1EA)

Roll Paper

(1EA)

Ground Cable

(1EA)

Page 9: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

12 How to Set BCM-600 How to Set BCM-600 13

BCM-600

How to Set BCM-600

(3) Maintenance and Cleaning After UseA. Keep the product cleans by cleaning it using a soft cloth at least once a month. Do not use

thinner, ethylene, oxidant or lacquer since these may damage the product.

B. Keep the accessories clean of dust or foreign materials. For cleaning, use a soft cloth and

clinical alcohol.

C. Do not soak the accessories in liquid or detergents. Keep the product and accessories out of

any liquid.

D. For the protection of the environment, please contact the agency you purchased the product

from when you wish to discard the BCM-600

E. Our product can be used semipermanently. Please contact our Customer Service dept. or any

distributor and then replace the parts , when wasting parts such as battery, LCD or LED and

so on have a problem .

(1) Precautions for SettingsPay attention when setting BCM-600 for the following points

- Use BCM-600 under conditions of 10℃∼45℃ of circumstantial temperature and 30%∼85%

humidity.

- Check the connection status of the power cord.

- Do not connect multiple cords to the power supply.

- Place the main body on a flat area.

- If noise occurs, ground the equipment.

- Do not use any electric cords generating connecting noise.

- Take care since the product can be broken by mechanical shock.

- Get rid of dust or inflammable material near the product when setting it up.

(2) Checkpoints before UseCheck the following points before measuring the patient’s condition.

- Make sure there is no mechanical risk.

- Check the lead, power plug, and accessories connected on the outside.

Precautions Before Use① Do not use product near dynamo, X-ray equipment, broadcast equipment.

② Do not twist the cable.

③ Attach each Lead properly.

④ Keep the patient motionless with comfortable position while measuring

Page 10: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

14 Components of BCM-600 Components of BCM-600 15

BCM-600

Components of BCM-600

(1) Front & Operation Panel (2) Side & Connection

(1) AC Power Button

(2) Screen Shift Button

(3) FREEZE Button

(4) GAIN Button

(5) PRINT START Button

(6) SPEED Button

(7) EXIT Button

(8) MENU Button

(9) LCD SCREEN

(10) PAPER LOCK

(11) PAPER Cover

(12) AC Power LED

(13) DC Power LED

(1) Connecting part with Patient cable

(2) USB Port

(3) RS-232(For Upgrade)

(4) LAN Port (For Upgrade)

(5) Power Adapter

(6) Potential equalization stud

Page 11: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

16 Description of Product Symbols Description of Product Symbols 17

BCM-600

3. Description of Product Symbols(1) Screen Configuration

① WAVE Area : Display the graph of electrocardiogram

② Wave Property Area : Display the Speed and Size of electrocardiogram

③ Each measuring Mode icon : Display the status of measurement, printing and so on.(Refer the

following page (Table 1) Explanation of Screen Icons)

④ Time : Display the present time

⑤ HR : Display the Heart Rate of a Patient

(Fig1. Main Screen)

Symbol Comments Symbol Description

Type CF Product,Defibrillation protected

Power stand-by switch

Wave speed change

Attention, consult accompanying documents

Equipotentiality

Direct Current

Signal On/Off RS-232C Port

LAN Port

Screen Shift

Wave Freeze

Start Printing (Printer)

Wave size change

Initial Screen (Exit)

Signal Output

Use for Program Upgrade

Use for Central Monitoring

Page 12: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

How to Use BCM-600 19

BCM-600Electrocardiograph (Ver 1.0)

18 Description of Product Symbols

(Table 1). Explanation of screen icons

Icon Meaning How to display

Limb Electrode (R) lead check

Limb Electrode (L) lead check

Limb Electrode (F) lead check

Chest Electrode (C1) lead check

Chest Electrode (C2) lead check

Chest Electrode (C3) lead check

Chest Electrode (C4) lead check

Chest Electrode (C5) lead check

Chest Electrode (C6) lead check

Auto diagnosis

Printing

Waveform stop

AC in use

Battery in use

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Normal (Green) Abnormal(Red)

Flickers at 1-second interval.

Flickers at 1-second interval.

Flickers at 1-second interval.

Fixed.

Fixed.

How to Use BCM-600

(1) How to Use

Step 1) Turn on the main power switch at the back of the product. Check the light-on status of the

AC LED in the front, and push the power switch 1~2 seconds to turn it on.(If any

problems occur, refer to the “How to solve the problems” section in the manual)

Step 2) If you want to change the setting value etc. during use, push the Menu button to change

the setting values per each measurement parameter

Step 3) When the measurement sensor is connected to the patient, the information of the patient

electrocardiogram is displayed as values and waveforms on the LCD screen

Push the Power button 1~2 seconds when turn it off and then “Power off” massage will be

displayed.

R

L

F

C1

C2

C3

C4

C5

C6

Page 13: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

All of the Menu can be controlled by Menu button.The way to control is as below

Step 1) To change the default setting, press the Menu button once, then you can see the below

display

After moving to Menu you want to change, by rolling to left or right the Menu button, you can

select the Menu by press (the triangle on the right side of Menu means that there are more

menu)

Step 2) Each Menu can be changed by the above mentioned way

(by rolling to right or left the menu button and press)

(1) EKG setting

MENUMENU ▶

PRINT ▶

PATIENT ▶

SETUP ▶

EXIT

MENUMENU ▶

PRINT ▶

PATIENT ▶

SETUP ▶

EXIT

EKGLEAD ▶

SPEED ▶

GAIN ▶

FILTER ▶

MUSCLEBASELINENOTCHReturn

Electrocardiograph (Ver 1.0)

20 MENU Setting MENU Setting 21

BCM-600

MENU SettingEKG Lead change

▶ To change the EKG Lead Wave setting, Roll the menu buttons to right or left and

select by press it

Lead of BCM-600 provides the 4 kinds of screen

EKG channel conversion can be done by press the Lead key

- 4 different configurations are as below;

MENUMENU ▶

PRINT ▶

PATIENT ▶

SETUP ▶

EXIT

EKGLEAD ▶

SPEED ▶

GAIN ▶

FILTER ▶

MUSCLEBASELINENOTCHReturn

LEAD- 3X1- 3X2- 6X1- 6X2Return

3 X 1 3 X 2

6 X 1 6 X 2

Page 14: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

22 MENU Setting MENU Setting 23

BCM-600

(2) Other Menu setting

▶ Please refer the below table for the other menu

◈ The choice of the filter may distort the original signal Therefore, you’d better use

appropriately only in case the recorded signal is bad so that the distortion of ECG signal can

be reduced

And we recommend you to apply the Baseline filter and Notch Filter all the time and to

apply appropriately Muscle filter

150Hz of filter is recommend in case of printing Diagnosis

3 x 13 x 26 x 16 x 2

Select the Configuration type ofEKG wave to be displayed

(same way in printing)

Changeable Range DescriptionEKG Menu

LEAD

12.5 mm/sec25.0 mm/sec50.0 mm/sec

Select EKG Wave processingspeed

SPEED

5 mm/mV10 mm/mV15 mm/mV

Select EKG Wave SizeGAIN

40 Hz100 Hz150 Hz

OFF

Select EKG FilterFILTER

ONOFF

Muscle Artifact filter settingMUSCLE

ONOFF

Baseline filter settingBASELINE

50 Hz60 HzOFF

If there is input, the eliminationfrequency is set.

NOTCH

MONITORDIAGNOSIS

FREEZE

MONITOR : Patient’sinformation, WAVE display

DIAGNOSIS : Patient’sinformation, WAVE, Diagnosis

informationFREEZE : Patient’s information,

WAVE output

Changeable Range DescriptionPRINT Menu

MODE

2 ~ 9 sec. Time control for the WAVE printWAVE SIZE

1 ~ 3 stageControl the Output Density(1-Dull,2-Normal,3-Dark)

BRIGHTNESS

30 pcsRefer the How to use Record on

page 27RECORDS

◈ The print processing speed, size, and channel settings comply with the setting value on

the LCD screen.

To print what are on the screen, as they are, it makes users to get more accurate

judgments

In order to change the print processing speed, size and cannel setting, Press the Menu

button to EKG menu, where you can change the setting value

◈ on Freeze in Print Mode, you can print as they are ,after checking the waves on the screen

◈ In case of no paper, "PAPER OUT" appears and you can not print

Open the printer paper cover 4 inches of the paper to be printed facing up, please insert.

Perforations, past the end of the paper should be positioned.

◈ Do not print greater than consecutive 3 times

◈ There is possibility of fire,

Do not use printing for 5min. after consecutive 3 times of printing .

Page 15: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

24 MENU Setting MENU Setting 25

BCM-600

Changeable Range DescriptionPRINT Menu

NAME

ID

AGE

WEIGHT

HEIGHT

Within 16 characters

Within 16 characters

OFF, 0 ~ 110 age

OFF, 0~300Kg

OFF, 0~250cm

Patient’s name setting

Patient’s Serial no. setting

Patient’s age setting

Patient’s weight setting

Patient’s Height setting

SEXFEMALE

MALEOFF

Patient’s gender setting

Changeable Range DescriptionSETUT Menu

HOSPITAL

DATE&TIME

Initialize

LANGUAGE

Within 16 characters

Year,Month,Day,Hour,Min.

OFF

ON

English

Hospital name setting

Date/time setting

Initialization

(Default Set referrence)

DEMO MODE setting

Language Select

DEMO

(3) Default setting

▶ The default settings of the values can be used, without the need of changing the setting values

of each parameter.

LEAD

SPEED

GAIN

FILTER

MUSCLE

BASELINE

NOTCH

MODE

WAVE SIZE

BRIGHTNESS

NAME

ID

AGE

WEIGHT

HEIGHT

SEX

HOSPITAL

DEMO

LANGUAGE

- 6 x 2

12.5 mm/sec

10 mm/mV

40 Hz

OFF

OFF

50 Hz

MONITOR

3 sec

2(Normal)

PATIENT

ID-001

OFF

OFF

OFF

OFF

BIONICS

OFF

ENGLISH

Item Name NeonateParameter

EKG

PRINT

SETUP

PATIENT

Page 16: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

26 PRINT PRINT 27

BCM-600

◈ In case of no paper, "PAPER OUT" appears and you cannot print

Open the printer paper cover 4 inches of the paper to be printed facing up, please insert.

Perforations, past the end of the paper should be positioned.

◈ Do not print greater than consecutive 3 times.

There is possibility of fire, do not use printing for 5min. after consecutive 3 times of

printing .

◈ When you save a printed data, don't remove an USB Flash Memory until the save progress

is complete.

◈ Please use an USB Flash Memory listed in 'BCM-600 Viewer Manual'. If not, it may

cause malfunction of the product.

PRINT

The print processing speed, size, and channel settings comply with the setting value on the LCD

screen. To print what are on the screen, as they are, it makes users to get more accurate judgments.

In order to change the print processing speed, size and cannel setting, press the Menu button to

EKG menu, where you can change the setting value.

If you use an USB Flash Memory, you can save a printed data to an USB Flash Memory. When

you save a printed data to an USB Flash Memory, print information will not be printed. (You can

use it in case of no paper)

For more information about USB SAVE, please refer to the PC Program manual provided by our

company.

Please refer the below table for the contents of other PRINT Menu.

PRINT Menu Changeable Range Description

MODEMONITOR

DIAGNOSISFREEZE

Print Menu SettingMONITOR : Patient's information,

WAVE displayDIAGNOSIS : Patient's information,

WAVE, Diagnosis informationFREEZE : Patient's information,

WAVE output

WAVE SIZE 2 ~ 9 sec. Time control for the WAVE print

BRIGHTNESS 1 ~ 3 stageControl the Output Density(1-Dull,2-Normal,3-Dark)

Refer the How to use Record on page 27RECORDS 30 pcs

Page 17: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

28 How to use RECORDS How to use RECORDS 29

BCM-600

How to use RECORDS

Select RECORDS on Print Menu by

press Menu button

1. Press Menu button after moving toMenu wanted to select by rolling theMenu button to right or left

2. In order to change the print properties,press the Menu button after moving tothat properties by rolling the Menubutton to right or left

3. To print, select Print Menu.And to move to no.1, select “CANCEL”

4. Press EXIT button to close Menu

MENUMENU ▶

PRINT ▶

PATIENT ▶

SETUP ▶

EXIT

PRINTMODE ▶

WAVE SIZE ▶

BRIGHTHESS ▶

RECORDS ▶

Return

(1) The data entry area already printed

(2) Print Key

(3) Cancel Key

(4) Print properties

◈ BCM-600 has stored the late 30pcs of printed data

Page 18: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

30 Power & Battery Specifications Power & Battery Specifications 31

BCM-600

◈ When the battery is removed, all data can be reset. Only in case the product is kept for long

time, Remove the battery from the product

◈ For the protection of the environment, please contact the agency you purchased theproduct from when you wish to discard the battery.

◈ Contact the agency you purchased battery in case battery should be changed.◈ To ensure the lifecycle and safety of battery, please discharge the battery and then charge

perfectly again every six months. Also time-worn battery exceed 24 months should be replaced.

C. Natural lifetime for spontaneous discharge of battery

The table below shows the lifetime for spontaneous discharge of the battery when the

product is unused.

D. Type of Battery

The same sort of battery should be used for replacement.

- Model : 3S2P(LS18650)

- Type : LI-Ion Battery

E. Time for the Battery exchange

The lifetime of the battery gets shorter if it is not 100% recharged for use

9 months

6 months

3 months

0℃ ~ 20℃

21℃ ~ 30℃

31℃ ~ 45℃

Charge / Discharge ration Number for use

1200 times

500 times

200 times

100 % Charge / discharge

50 % Charge / discharge

30 % Charge / discharge

Power & Battery Specifications

◈ During operation using AC power, noise may occur in the waveform. If this is the case,ground the product soundly.

◈ An AC power connection in a wet or damp area may cause electric shock or fatal damageto the product.

◈ When you connect the adapter will charge the battery automatically

2. Battery SpecificationsWhen the AC power is disconnected and the switch is turned on, the power is supplied using the

battery, signaled by the orange light on the battery LED.

Both AC and DC power is used for BCM-600. Basically, AC power is used and a rechargeable

battery is available for portable use.

1. AC Power (110~240VAC)If the AC power is connected to the main body of the product, a light-green color lights up on the

AC LED on the front. If the product is not being used, the power changes automatically to the

recharging mode.

A. Signs for Battery

A. Type of Adaptor

- Model : MW160 (KA1503F02)

- Input : 100~240V(AC), 50~60Hz- Output : 15V(DC), 3Ah

- Manufacturer: BridgePower Corp

Battery status Explanation

B. Time for 1 ea of battery

Recharging time : 9 H(OP Mode), 6H (Power Off Mode)

Use time :240 minutes

Battery full

After use for 50 ~ 60 minutes

Requiring recharging

Low battery requiring recharging

Storage temperature Life for spontaneous discharge

Page 19: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

32 EKG Measurement EKG Measurement 33

BCM-600

EKG Measurement

A. How to Connect the ECG Cable

◈ With the use of an electric pacemaker, do not touch the patient, table, or other equipment.

◈ Check if the ECG cable is damaged before taking the ECG measurement. (A damaged

ECG cable cannot protect the patient.)

◈ Do not use disposable electrodes. (They may cause noise.)

When the patient is on stable condition, lay him down on the bed.

Then spread EKG gel all over the parts where the electrodes are to be attached.

Please confirm the EKG signal on LCD screen after attaching the patient electrodes to the patient

Please rub the skin with the gel, the alcohol or the water before measuring the patient

Attach the patient’s electrode to the sites shown below.

B. Electrode AttachmentCheck the status of the electrode attachment

When you measure a patient's condition

- Make sure if there are mechanical hazards or not

- Check the cable connected to External and accessories.

- Check all of the measuring instruments to measure the patient

C. Electro CardiographyAfter attaching electrodes to the patient, Turn the power switch

Set the filter, Wave size, printing speed, channel composition, printing time, diagnosis.

Input the patient information

Check the state of ECG signal and the state of Lead through the LCD screen.

Check the electrode connection if the signal is in bad condition

BCM-600 records the patient’s ECG by press printer button

BCM-600 analyzes and prints the cumulated ECG data for 10 seconds.

“Please wait” will be appeared in case the ECG data are cumulated within 10seconds.

Press printer button in order to stop working for ECG analysis or to stop printing

V1 = C1

V2 = C2

V3 = C3

V4 = C4

V5 = C5

V6 = C6

RA = R

LA = L

R.L = N

L.L = F

Page 20: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Before Requesting After-Sales Service 35

BCM-600

Before Requesting After-Sales Service

This section shows how to treat simple problems encountered during use.

Please check the following points before requesting after-sales service

▣ Problems in Display

Sign How to treat

▣ Problems in Measurement ECG

Sign How to treat

The ECG waveform is notdisplayed.

Electrocardiograph (Ver 1.0)

34 Before Requesting After-Sales Service

▣ Problems in Printer (OPTION)

Sign How to treat

1. Press the power button during 1~2 seconds 2. Check if “Power off” message is shown in display.

1. Check if there is any equipment with a strong magnetismaround the product.

2. Check if the cable connection is stable.3. Check if the product is grounded.

The power of the product isnot turned off.

The waveform is not normal.

The power of the productturns off suddenly during use.

The power of the product isnot turned on.

1. Check if the AC power is supplied..2. Press and hold the power button for 2 seconds

1. Check if the power is off due to battery release.2. Check if AC power is supplied to the product.

1. Check if the electrode attached to the patient is stable.

1. Check if the cable connection is stable.

2. Check if the ECG cable is damaged.

Severe noise on thewaveform.

No printing

1. Check if the printer icon blinks on the screen.

2. Check if the paper is set normally.

3. Check if the printer door is closed.

Noise during printing

1. Check if the roller is contaminated with foreign materials

or if the gear is out of position.

2. Check if the problem is due to a paper jam (change the

paper)

Page 21: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

36 Product Specifications Product Specifications 37

BCM-600

Product Specifications

1) Features

- LCD Graphic Display 800 X 480 dots, 3x1/3x2/6x1/6x2

- Channel with base line movement possibility

- Standard 6 channels

- Real time heart rate display

- Real time display and print out

- Various patient information

- (Patient Name, Sex, Age, Weight, Height)

- High Resolution Thermal Printer

- Defibrillator protection

- Rechargeable Battery Operation

2) Specifications

- Diagnostic Electrocardiography

Input circuit : Floating input

Leads : Standard 12 leads

Input impedance : 20Mohm min

Input circuit current : 0.05 A max

Calibration voltage : 1mV with ± 2%

CMR : 100dB or more

Polarization voltage : ± 350mV min

Time constancy : 3.2 seconds (0~+20%)

Sensitivity : 5, 10, 15 mm/mV ± 5%

Frequency response : 0.05 ~ 100Hz (13dB)

Filters : AC50/60 Hz, -20dB or lower, Muscle tremor 35 to 45 Hz, -3db(-6db/oct)

A/D conversion : 12bits

Sampling rate : 2msec

3) Physical Specifications

- Power requirements : AC 100 ~ 240V, 50/60Hz

- Power consumption : 50VA(MAX)

- Battery: LI-Ion , 11.1V , 4400mAh

- Continuous operating time with battery : Max 4Hours

- Dimension : 327(W) X 206(L) X 87(H) mm (W x D x H)

- Weight : 1.75Kg(Incl. battery)

4) Interface

- RS-232C port

- LAN Port (Program upgrade)

- USB Port (Data storage)

5) Operating Environment

- Temperature : (Operating : 10℃ ~ 45℃ / Storage : -10℃ ~ 50℃)

- Humidity : (Operating : 30 ~ 85% / Storage : 20 ~ 95%)

- Atmospheric Pressure : (Operating : 70 to 106 kPa / Storage : 70 to 106 kPa)

6) Product specification listed on the manual are subject to change according to the our

design change procedure without notice.

Page 22: Operation Manual - BIONICSebionics.co.kr/bbs/download2.asp?filename=BCM-600... · I BCM-600 prints the ECG, HR, PR interval, QRS interval, QT interval, QTc, P-R-T-axis calculated

Electrocardiograph (Ver 1.0)

Product Warranty

Product Name

Model Code

Manufacture Approval No.

Manufacture Approval Date

Manufacture Serial No.

Warranty Period

Purchase Date

Name ofHospital

Address

Name

Tel

Thank you for using BCM-600

This product is a medical device.

Only the products that have Passed our extensive and thorough quality tests are

offered to our customers

Our policies on repairing, exchange and refund complies with “Regulation on

Consumer Protection & Compensation Standards” issued by Economic

Planning Board of Korea

Name of Distributor

Name of Manufacture

CustomerInformation

Electrocardiograph

BCM-600

No. 1936

One year after the purchase date

Date Month Year

Tel) Fax)

BIONICS Co., Ltd.