W ater Fill Level Acu-sInQ Enzymatic Dosing System For use in the SPD and GI Prolystica 2X Concentrate Enzymatic Presoak and Cleaner STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 USA 440-354-2600 800-548-4873 www.steris.com Operate/Procedures Fill the sink to the desired level based on STERIS line level sticker. Select the appropriate fill capacity on the Acu-sInQ. (Half/Full) Press the Start/Scroll pushbutton. Acu-sInQ System will automatically dispense correct ratio of chemistry into the sink. The LCD screen will display the temperature of the solution (water + chemistry). The ideal temperature ranges for optimal performance are 90-120º F. Immerse and soak devices/scopes until all organic matter is dissolved and removed. A minimum of one minute is recommended. For instruments with excessive soil, extend soak time. Mechanically (manually) clean devices/scopes with appropriate brushes, sponges, etc. and irrigate any channels with solution according to device manufacturer instructions and recommended procedure by professional organization such as SGNA, AORN and IAHCSMM. Thoroughly rinse instruments and equipment, aspirating water through all channels to remove Prolystica 2X Concentrate Enzymatic Presoak and Cleaner. If the solution is out of range, the temperature will blink on the LCD screen alerting to refresh the solution to achieve the appropriate temperature. Discard enzymatic solution after each use or when visibly soiled. Document # M3402EN.2010-01, Rev. A Printed 04/2010, 1000 ©2010 STERIS Corporation. All Rights Reserved. Printed in the USA. For more information, please contact STERIS in the United States at 800-548-4873, in Canada at 800-661-3937 or visit our website at www.steris.com. Ensure the Prolystica 2X Concentrate Enzymatic Presoak and Cleaner is fully primed within the Acu-sInQ system and the chemistry container is not empty. Ensure the chemistry tubes are affixed to the back of the sink with a suction cup. Ensure the temperature probe wiring on the Acu-sInQ is fully positioned into the sink directly above the sink bottom. 1 B B A A C C Before you start! Materials 2 Disclaimer: This wall chart is for quick reference only and is not a substitute for the applicable operations manual and instructions for use. All personnel involved in the use of these products must carefully review and comply with the warnings, cautions and instructions set forth in the applicable operations manual and/or instructions for use.

Upload: others

Post on 26-Mar-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Operate/Procedures - Infrastructure BC · channels to remove Prolystica 2X Concentrate Enzymatic Presoak and Cleaner. • If the solution is out of range, the temperature will blink

Water Fill Level

Acu-sInQ Enzymatic Dosing SystemFor use in the SPD and GI

Prolystica 2X Concentrate

Enzymatic Presoak and Cleaner

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 • USA440-354-2600 • 800-548-4873www.steris.com

Operate/Procedures• Fill the sink to the desired level based on STERIS line level sticker.

• Select the appropriate fill capacity on the Acu-sInQ. (Half/Full)

• Press the Start/Scroll pushbutton.

• Acu-sInQ System will automatically dispense correct ratio of chemistry into the sink.

• The LCD screen will display the temperature of the solution (water + chemistry).

• The ideal temperature ranges for optimal performance are 90-120º F.

• Immerse and soak devices/scopes until all organic matter is dissolved and removed. A minimum of one minute is recommended.

• For instruments with excessive soil, extend soak time.

• Mechanically (manually) clean devices/scopes with appropriate brushes, sponges, etc. and irrigate any channels with solution according to device manufacturer instructions and recommended procedure by professional organization such as SGNA, AORN and IAHCSMM.

• Thoroughly rinse instruments and equipment, aspirating water through all channels to remove Prolystica 2X Concentrate Enzymatic Presoak and Cleaner.

• If the solution is out of range, the temperature will blink on the LCD screen alerting to refresh the solution to achieve the appropriate temperature.

• Discard enzymatic solution after each use or when visibly soiled.

Document # M3402EN.2010-01, Rev. A Printed 04/2010, 1000©2010 STERIS Corporation. All Rights Reserved. Printed in the USA.

For more information, please contact STERIS in the United States at 800-548-4873, in Canada at 800-661-3937 or visit our website at www.steris.com.

• Ensure the Prolystica 2X Concentrate Enzymatic Presoak and Cleaner is fully primed within the Acu-sInQ system and the chemistry container is not empty.

• Ensure the chemistry tubes are affixed to the back of the sink with a suction cup.

• Ensure the temperature probe wiring on the Acu-sInQ is fully positioned into the sink directly above the sink bottom.

1

B

B

A

A

C

C

Before you start! Materials

2

Disclaimer: This wall chart is for quick reference only and is not a substitute for the applicable operations manual and instructions for use. All personnel involved in the use of these products must carefully review and comply with the warnings, cautions and instructions set forth in the applicable operations manual and/or instructions for use.