od hawkeyedp probes along with a 45˚ hawkeye ... - testex …
TRANSCRIPT
Date: November 3, 2010
Graphitization Tube Study
Author: Shawn Gowatski
Mirant provided 6 boiler tubes to TesTex that were suspected of containing graphitization. Four
of the tubes were removed from Bowline Station and two were removed from the Pittsburg Station.
The Balanced Field Electromagnetic Technique (BFET) was used to examine the tube samples. Curved
OD HawkeyeDP probes along with a 45˚ Hawkeye DP probe were used with the ProdigyII Electronics
(serial #09-3118-04)
Balanced Field Electromagnetic Technique Fundamentals:
The Balanced Field Electromagnetic Technique was developed to enhance the signal responses
produced from smaller defects such as cracks. The technology was originally used to detect cracking
from the outside surface to inspect welds in DA Tanks, pressure vessels, tube stubs, and any other areas
suspected to have cracking. In this technique, electromagnetic coils are wound and arranged in a
balanced state. This balanced state is achieved by placing coils in both the “x” and “z” geometries at
zero potential to each other. With the excitation coil in the “x” geometry and the sensor coil in the “z”
geometry, a differential signal is produced over defected areas. In detail, the alternating current
produced by the excitation coil is uniform and undisturbed when no defects are present. Conversely, the
current is interrupted when a defect is present and forced to travel around it in a distorted fashion. It is
this state of distortion that causes the coils to become unbalanced and thus producing an indication for
the user that signals a defect. This distortion signal response can be measured and a crack depth can be
calculated by comparison to a calibration standard. The sizing of graphitization will be extremely
difficult but some type of ranking is possible through the scanning some collected tube samples with
graphitization. At minimum, the defects found can be ranked or classified according to the signal
strength of the defect. Another key to the balanced field electromagnetic technique is the ability to
eliminate liftoff (and/or probe wobble) and noise from the signal. This is accomplished through a
special algorithm in which these unwanted elements are rotated away from the main signal.
With the technology based on electromagnetics, a polished surface is not required. Quality
readings can be acquired through coatings such as paint, epoxy, and rubber. The actual probe does not
need to be in contact with the test piece. Sometimes the probe is placed in a small cart with the probe
base set slightly above the test piece. The probe can be pulled quickly across the test specimen at a
speed up to 1 foot per second. The BFET method can test different types of metal by adjusting the test
frequencies which range from 100HZ to 30,000HZ.
During an inspection, the data is viewed in real-time on a computer screen. This allows defects
to be found and located during the data collection process which allows a plant to quickly prove-up any
suspect defects and to schedule repairs without having to wait for the completion of the job.
The picture above shows various OD HawkeyeDP probes used to test the samples. The handle on the probe can be modified
to fit the probe into tight areas.
The picture above shows the 45˚ HawkeyeDP.
Findings:
All six tubes were scanned initially with the 45˚ HawkeyeDP probe. This probe design allows
circumferential and axial cracks to be detected. The circumferential HawkeyeDP probes were used to
prove-up any defects found and also to test around the welds on the two Pittsburg samples. The tests
were performed at 100Hz. A total of 8 discontinuities were detected. Each area containing a
discontinuity was marked with a square and labeled by a number. Some of the discontinuities gave a
stronger response than others. TesTex did not have any calibration standards to size the responses from
these discontinuities. The discontinuities are graded in a relative ranking from the strongest to weakest
signal and are listed below. Indications 6, 7, and 8 are on the two Pittsburg tubes and the remaining
indications are on the Bowline samples.
Indications 3, 6, and 8 showed the strongest signal response.
Indications 1 and 5 gave a moderate signal response.
Indications 2, 4, and 7 gave a small but distinguishable signal response.
A Metallurgical evaluation was performed on all eight indications and graphitization was found in each
area. The table below lists the results of the metallurgical evaluation.
Area Number Station Damage Present
1 Bowline Chained graphitization through section
2 Bowline Chained graphitization through section
3 Bowline Chained graphitization through section
4 Bowline Chained graphitization through section
5 Bowline Chained graphitization through section
6 Bowline Chained graphitization through section
7 Pittsburg Chained graphitization at OD with scattered graphite nodules
throughout
8 Pittsburg Band of graphite midwall with scattered graphite nodules throughout
The table above shows the metallurgical evaluation results.
Waveforms and pictures for each indication follow.
The waveform above is from indication 1 that was collected The waveform above is from indication 1 that was collected
using a 45˚ HawkeyeDP. using a circumferential HawkeyeDP.
The Balanced Field Electromagnetic Technique (BFET) waveform is displayed in (5)
different windows. The bottom right window is the raw data. The bottom left window shows the data
processed. The middle left window is a simulated C-scan. The top left window is a zoomed in view of
the data and the top right window shows a color-coded 3D view of the data. Any sharp crack-like
defects will cause the waveform to rise up/down sharply.
The picture above is a traverse section from Indication 1 with chained graphitization at the 11 and 1 o’clock positions.
The picture above is from Indication 1 showing chained graphitization at higher magnification.
The picture above is from Indication 1 showing chained graphitization at higher magnification.
The waveform above is from indication 2 that was collected The waveform above is from indication 2 that was collected
using a 45˚ HawkeyeDP. using a circumferential HawkeyeDP.
The first picture above is from Indication 2 showing chained graphitization through wall and the second picture is from
Indication 3 showing chained graphitization through wall.
The waveform above is from indication 3 that was collected The waveform above is from indication 3 that was collected
using a 45˚ HawkeyeDP. using a circumferential HawkeyeDP.
The waveform above is from indication 4 that was collected The waveform above is from indication 4 that was collected
using a 45˚ HawkeyeDP. using a circumferential HawkeyeDP.
The first picture above is from Indication 4 showing chained graphitization through wall and the second picture is from
indication 5 showing chained graphitization through wall.
The waveform above is from indication 5 that was collected The waveform above is from a visible crack that was collected
using a 45˚ HawkeyeDP. using a circumferential HawkeyeDP.
The waveform above is from indication 6 that was collected The waveform above is from indication 6 that was collected
using a 45˚ HawkeyeDP. using a circumferential HawkeyeDP.
The first picture above is from Indication 6 showing chained graphitization through wall and the second picture is from
indication 5 showing chained graphitization near the OD.
The waveform above is from indication 7 that was collected The waveform above is from indication 8 that was collected
using a circumferential HawkeyeDP using a circumferential HawkeyeDP.
The pictures above show the graphite nodules as polished from Indication 8.
The picture above shows indications 1 and 2.
The picture above shows indications 3 and 4.
The picture above shows indication 5. There was some uneven scale on the surface, so the location of the signal was
sandblasted.