non-contact thermometers emissivity -...
TRANSCRIPT
Non-contact thermometers – emissivity
Introduction
When working with non-contact thermometers the key term is emissivity ε. It
is defined as “the relative power of a surface to emit heat by radiation; the ratio of the
radiant energy emitted by a surface to that emitted by a blackbody at the same temperature”
[1]. The user needs to be aware that a non-contact thermometer is not measuring
temperature but radiated energy. In order to show temperature, the emissivity of the
measured object has to be known and correctly set on the thermometer. The purpose of this
experiment is to estimate emissivity of different surfaces and to estimate the error caused by
the wrong setting of emissivity (or fixed – some IR thermometers don’t allow to change
emissivity).
Tasks:
1. With contact thermocouple and
IR thermometer with fixed
emissivity 0.95 determine
emissivity of different surfaces.
2. Calculate absolute and relative
error of temperature reading
caused by the different
emissivity of the object.
3. Make a thermo graphic image
of the Al heat sink with the
thermocamera.
Used instruments:
Fixtures with resistors on a heatsink, Peltier element
IR thermometer with fixed emissivity 0,95
Multimeter Axiomet AX-18B with surface thermocouple probe
Thermocamera FLIR i50
Power supply 18210
Procedure description – gauge factor
The radiation of a black body is described by the Planck’s law as shown in Fig. 1 [3].
It relates the spectral radiance of a black body (the quantity of radiation) with its temperature
and wavelength of the radiation.
A blackbody emits total radiant power WB into a surrounding hemisphere given by [4]
4
BW T (1)
Where σ is Stefan-Boltzman constant and T is temperature in Kelvin.
Any other body can be characterized by a dimensionless parameter - emissivity
/ BW W (2)
Fig. 1 – Black body radiation [3]
It is the fraction of black body power emitted in the surrounding hemisphere.
Emissivity depends on the surface of the body and on its temperature. By definition it is 1 for
a black body.
The black body is an idealized concept. Real objects do not absorb all incident energy,
some part is reflected. They behave like gray bodies. Theirs emissivity is ε<1.
In order to correctly measure temperature
with an IR thermometer, the emissivity of the
measured object has to be known.
The method used here consist in measuring
the real object temperature with a contact
thermometer (thermocouple in this case).
The used IR thermometer has a fixed setting
of emissivity 0.95. Therefore its reading is correct only for this emissivity. The IR
thermometer measures the radiated energy 4
0,95 0,95IRW T (3)
Where T0,95 is the temperature shown on the IR thermometer with fixed emissivity
0,95.
The object radiated energy is a function of its temperature Tobj. 4
. . .obj obj objW T (4)
In order to determine object emissivity εobj. the object temperature Tobj. is measured with a
contact thermometer.
Then the object emissivity εobj. can be calculated as [5] 4
0,954 4
. . 0,95 0,95 . 0,95 4
.
obj obj obj
obj
TT T
T (5)
The temperatures have to be substituted in Kelvin!
Conclusions
In conclusions state the precision of thermocouple measurement from multi-meter
manual, discus the effect of accuracy on emissivity calculation. Also find the accuracy of
reading of the IR thermometer.
References
[1] Emissivity, online on <http://www.merriam-webster.com/dictionary/emissivity>,
accessed on 11.3.2013
[2] Emissivity Coefficients of some common Materials, online on <
http://www.engineeringtoolbox.com/emissivity-coefficients-d_447.html>, accessed on
11.3.2013
[3] Planck's law, online on < https://en.wikipedia.org/wiki/Planck's_law>, accessed on
11.3.2013
[4] Chandos, R. J., Chandos R.E. : Radiometric Properties of Isothermal Diffuse Wall
Cavity Sources, online on < http://www.electro-
optical.com/pdf/chandosemissivitypaper.pdf>, accessed on 11.3.2013
[5] A Review of the Physics for Emissivity Correction of Infrared Temperature
Measurements, online on < http://www.apogeeinstruments.co.uk/content/SI-
emissivitycorrection.pdf>, accessed on 11.3.2013
material Emissivity
Al – not oxidized 0,12 – 0,18
Fe - shiny 0,32 – 0,42
Cu 0,1 – 0,35
Graphite, coal 0,65 – 0,97
Ideal black body 1
Non-contact thermometers – IR
thermometer and camera
Introduction
Non-contact thermometers can be used to measure very high temperatures
impossible to be measured in a different way. However the user needs to be aware that a
non-contact thermometer is not measuring temperature but radiated energy. It is not a
simple point and measure device. Precautions need to be taken with IR thermometers. The
purpose of this task is to show their basic properties that need to be considered in order to get
a correct reading. The same rules apply for a thermo camera.
Tasks:
1. With the contact probe of IR thermometer
determine emissivity of the black, white
and silver surfaces on the test plate.
2. Measure the temperature distribution on
the plate for 2 given configurations of
heat sources (sources 1,2 or 3). Plot in a
surface plot. On each colored surfaces, set
the correct emissivity.
3. Take an image of the test plate with the
thermocamera
Used instruments:
Heated plate with Peltier elements
IR thermometer Fluke 576 + PC with software IRGraph
Thermocamera FLIR i50
Power supply Diametral P230R51D
Procedure description
The IR thermometer measures a sum of energies. The energy as seen by the IR
thermometer is composed of emitted energy (function of object temperature), reflected
energy, transmitted energy (energy passing through the object) and absorbed energy (not
shown in Fig. 2 – the energy absorbed between the object and IR thermometer – for small
distances, this can be neglected).
The emitted energy is a function of temperature as given by Planck’s law. In order to
measure the radiation of a body correctly emissivity has to be known. One possibility how
to measure emissivity was shown in the previous experiment.
When the IR thermometer has a possibility to adjust the emissivity, like the one used
here also the following procedure can be used.
1] Measure the object temperature with a contact probe of the IR thermometer.
This usually is a thermocouple, type K in our case.
2] Measure the temperature with the IR thermometer in the same spot, adjust the
emissivity on the IR thermometer until both temperature readings are the same.
Fig. 2 – IR thermometer measures a sum of
energies [1]
The infrared thermometer is not measuring in a single point. Its active area is a cone
with specified D: S ratio like the one shown in Fig. 3.
The further the object is apart, the bigger is the cone. The cone has to be filled
completely with the measured object in order to obtain a correct reading – Fig. 4. When the
object is not filling the whole cone, the IR thermometer measures partially the radiation of the
background and therefore the reading will be false.
For the used IR thermometer Fluke 576, the optical chart is shown in Fig. 7
Due to the used optics, the smallest spot
diameter is 19 mm for a distance of 1150 mm.
For this experiment, set the distance
1150 mm between the IR thermometer and
heated plate.
With the IR thermometer measure the
temperature in the center of each square on the
heated plate. Plot the result in a surface plot and
make a make an image of the plate with a
thermocamera.
During all experiment, consider that you are measuring a sum of radiation. The effects
of different emissivity can be significant. In Fig. 6 there is an object (a garden door) with
same temperature on its surface. There are metallic and wooded parts. As they have different
emissivity they seem to have different temperature. They also reflect differently the ambient
radiation. An example of how the reflected energy affects the result is shown in Fig. 7. This
picture was obtained on a mirror surface. The reflection is visible and masks completely the
real temperature of the object.
Fig. 3 – IR thermometer measures in a cone [2] Fig. 4 – Correct use of a IR thermometer [1]
Fig. 5 – IR thermometer measures a sum of
energies [1]
Conclusions
Assess the influence of different emissivity of the surfaces on the reading, when the
emissivity would not be set to a correct value. From the IR thermometer manual read the
accuracy of reading for both the IR and thermocouple. From response time in the manual
calculate time constant of the IR thermometer.
References
[1] Fluke 576 Precision Infrared Thermometer – Users Manual, online on <
http://www.myflukestore.com/crm_uploads/fe_576_users_manual.pdf>, accessed on
11.3.2013
[2] $15 Infrared Thermometer, online on <
http://forums.anandtech.com/showthread.php?t=2049940>, accessed on 11.3.2013
Fig. 6 – Effects of different emissivity Fig. 7 – Mirror image on reflective surface
Emissitivy Submitted by:
Task: Date: 16.1.2014
It is important to wait for steady state. Check with the IR thermometer that the temperature is constant at 1 point.
black painted Al heatsink - current 3A black painted Al heatsink - current 5A Peltier element - current 0,5A
thermocouple (°C) 39,6 thermocouple (°C) 55,1 thermocouple (°C) 60,6 20,8
IR thermometer (°C)39,2
IR thermometer (°C)56,8
IR thermometer (°C)60,9 16,7
emissivity (-) 0,95 emissivity (-) 0,97 emissivity (-) 0,95 0,8981
absolute error (°C) -0,4 absolute error (°C) 1,7 absolute error (°C) 0,3
relative error (%) -1 relative error (%) 3 relative error (%) 0
original shiny Al heatsink - current 3A original shiny Al heatsink - current 5A Heatsink thermal image:thermocouple (°C) 39,8 thermocouple (°C) 53,2
IR thermometer (°C) 24,3 IR thermometer (°C) 26,8
emissivity (-) 0,78 emissivity (-) 0,68
absolute error (°C) -15,5 absolute error (°C) -26,4
relative error (%) -64 relative error (%) -99
Resistor with heatsink - current 2A Human skin
thermocouple (°C) 94,0 thermocouple (°C) 38,4
IR thermometer (°C) 105,0 IR thermometer (°C) 35,7
emissivity (-) 1,07 emissivity (-) 0,92
absolute error (°C) 11,0 absolute error (°C) -2,7
relative error (%) 10 relative error (%) -8
Used instruments:
Conclusions:
1. With contact thermocouple and IR thermometer with fixed emissivity 0,95 determine emissivity of
different surfaces
2. Calculate absolute and relative error of temperature reading caused by the different emissivity of the
object
In conclusions state the precision of thermocouple measurement from multimeter manual, discus the effect of accuracy on emissivity calculation. Also
find the accuracy of reading of the IR thermometer.
3. Make a thermographic image of the Al heatsink with the thermocamera
4
0,95
. 0,95 4
.
obj
obj
T
Tε ε= ⋅
page 1/1
Czech Technical University in PragueFaculty of Mechanical Engineering
Department of Instrumentation and Control Engineering
Technická 4
166 07, Prague 6
Czech republic
Certificate Of CalibrationCertificate No: 1
Model name: IR-101
Manufacturer: Europe supplies Ltd.
Description: Portable IR thermometer
Temperature range: -20°C to 300°C
Serial No: Z2-02001372/000
Calibration date: 17.10.2013
Calibration due: 17.10.2014
Calibration interval: 1 year
Test report
Set value Reading Absolute error Relative error Specification
°C °C °C %
40 44 4 9,09 ±2°C nebo ±2 % (platí větší hodnota)
80 84 4 4,76 ±2°C nebo ±2 % (platí větší hodnota)
Reference standards used
Serial No Manufacturer Model Name Description Calibrated
130209108 Thermoworks IR-500
PORTABLE IR
CALIBRATOR
(BLACKBODY TARGET) 11.7.2013
Laboratory Enviroment:
Temperature: 22 °C
Humidity: 62%
Pressure: 745 torr
Distance of IR
thermometer to
calibrator 30 mm
Calibrated by: Martin Novak Date: 16.1.2014
Non-contact thermometers - IR thermometer and camera Submitted by:
Task: Date:
2. Measure the temperature distribution on the plate for 2 given configurations of heat sources (sources 1,2 or 3). Plot in a surface plot. On each colored surfaces, set the correct emissivity.
3. Take an image of the test plate with the thermocamera
It is important to wait for steady state. Check with the IR thermometer that the temperature is constant at 1 point.
Temperature distribution on the test plate - configuration 1 Emmissitity:Y/X 1 2 3 4 5 6 7 8 black 0,96
1 24,2 24,4 25,1 25,9 26,2 25,7 25,3 25,0 white 0,96
2 24,3 24,5 25,6 27,2 27,4 26,5 25,6 25,0 silver 0,96
3 24,3 24,6 26,2 29,9 33,4 28,0 26,0 25,1
4 24,2 24,7 26,1 29,3 33,9 28,8 26,2 25,2
5 24,1 24,5 25,3 26,6 27,7 26,6 25,6 24,9
Temperature distribution on the test plate - configuration 2
Y/X 1 2 3 4 5 6 7 8 Used instruments:1 24,4 24,4 24,6 24,7 25,1 25,6 26,2 26,8
2 24,8 24,8 25,9 25,0 25,2 26,2 27,5 28,3
3 25,6 26,3 26,6 25,9 25,3 26,5 30,3 32,0
4 26,0 27,7 30,5 26,5 25,9 27,3 30,4 33,5
5 25,7 26,6 27,2 26,2 25,8 26,6 28,1 29,0
Charts:place charts on separate sheet
Conclusions:
1. With the contact probe of IR thermometer determine emissivity of the black, white and silver surfaces on
the test plate.
Assess the influence of different emissivity of the surfaces on the reading, when the emissivity would not be set
to a correct value. From the IR thermometer manual read the accuracy of reading for both the IR and
thermocouple. From response time in the manual calculate time constant of the IR thermometer.
Non-contact thermometers - IR thermometer and camera charts:
Thermocamera image: Thermocamera image:
1
2
3
4
51 2 3 4 5 6 7 8
y (-)
temperature (°C)
x (-)
Temperature distribution - configuration 1
20,0-25,0 25,0-30,0 30,0-35,0
1
2
3
4
51 2 3 4 5 6 7 8
y (-)
temperature (°C)
x (-)
Temperature distribution - configuration 2
25,0-30,0 30,0-35,0
Force – gauge factor
Introduction
One of the basic tasks in mechanical engineering is to measure force or torque.
Strain gauges are used almost solely for this purpose. There are two kinds of strain gauges.
Wire (or foil) strain gauges and semiconductor strain gauges. Wire strain gauges have lower
sensitivity but have better linearity and are not so strongly dependent on temperature like
semiconductor strain gauges. Only foil strain gauges will be used in this task. As the
resistance change with the applied force is very small, a bridge has to be used.
Tasks:
1. Determine Gauge factor K of a
foil strain gauge. Repeat the
experiment 3 times, calculate
average Gauge factor.
2. Measure and plot dependence of
bar bend on applied force y = f (G)
and bridge output voltage on
applied force V =g (G).
3. Test temperature influence – as the
last DEMO.
Used instruments:
Fixture with bar + strain gauges – ¼ bridge configuration
Geometric Quantities Measurement System – INTRONIX – NX 3030 + position
sensor (LVDT-probe ± 0,5 mm – accuracy 1 %),
Digital Precision Measuring Amplifier – SCOUT 55 + strain gauge quarter bridge
with the compensating gauge strain RK = RM = 120 Ω
The loading system with 6 pcs metal discs (0,63 kg ± 1% each)
Resistive decade Rc = 567 k Ω + R, where R … 0-100 kΩ, (1%).
Procedure description – gauge factor
Resistive strain gauges change electrical resistance R with applied strain. Strain is the
relative elongation ε = Δl/l. The basic resistance R for a strain gauge without load is typically
120 Ω.
R lK
R l
(1) Where K is so called “gauge factor” (“the deformation
sensitivity” – the main gauge strain parameter).
The experiment is shown in Fig. 2. The fixture is arranged so as to apply a constant
bending torque M0 between supports. The relative elongation of bar surface filaments can be
calculated as
2
0
0 4
r
yh
EW
M
l
l
(2) Where y is the bar deflection, E is Young's modulus and h,
b, r are bar dimensions.
Fig. 1 - Strain gauge bridge
Equation (2) is valid for a rectangular
bar as
,6
2
0
bhW (r >>y) (3)
8
12
0rM
EJy (4)
3
12
h bJ (5)
A shunt resistor Rc is connected in
parallel to the strain gauge – Fig. 1,
the change of resistance ΔR is
(parallel combination of resistance)
0
2
0
0
00
RR
R
RR
RRRR
cc
c
(6)
As the value of Rc is known, by substituting into (1) the gauge factor K is 2
0
0 4c
R rK
R R yh
where R0 = 120 Ω (7)
The procedure is the following (repeat this experiment 3 times):
1) Apply a known load - all 6 disks = 6x 0,63 kg, measure the bar deflection on
the – INTRONIX – NX 3030 system and the bridge output voltage from the SCOUT 55
system. Don't forget to note the initial bar deflection when the beam is unloaded. Use
this as a reference (ZERO) value.
2) Remove the disks and achieve the same bridge output voltage by connecting a
parallel resistor (decade) to the stain gauge - shunt resistor Rc.
Procedure description – dependence on bend
Apply gradually 0 to 6 disks to the beam and measure the bar deflection and bridge
output voltage. Then gradually remove the disks, measure again in order to determine if there
is any hysteresis.
Conclusions
Calculate sensitivity for dependence of bar bend on applied force y = f (G) and bridge
output voltage on applied force V =g (G). Assess the influence of temperature and the
importance of temperature compensation.
Fig. 2 – Gauge factor experiment
Force – ¼ and ½ bridge
Introduction
¼ bridge uses one sensor. In order to increase sensitivity 2 sensors can be used
in a ½ bridge configuration. As it was seen from the previous experiment, there is a high
dependence on temperature for a strain gauge. It is primarily working as a thermometer and
only when we correctly compensate for temperature changes, it is measuring strain.
Therefore, a temperature compensation strain gauge has to be used at all times. It is the
same strain gauge type, on a same temperature, but not loaded with the measured strain. For a
½ bridge 2 temperature compensating strain gauges have to be used.
Tasks:
1. Determine sensitivity of ¼ and
½ strain gauge bridge
configuration.
2. Compare sensitivities for
compressive and tensile stress
– compare sensitivities in the
I. and III. quadrant. Interpolate
with a straight line with the
linear regression method
Used instruments:
2x fixture with strain gauge half bridge, R0 = 120Ω ± 0,35 ‰, K – factor 2,08 ± 1%,
type 10/120 LY 11 for the steel
Precision Measuring Amplifier – SPIDER (acc. class 0,1)
PC with software for SPIDER amplifier
Procedure description
The condition of bridge balance is
3241 RRRR (8)
When the bridge is balanced, the output voltage Uv = 0
¼ bridge has 1 measuring strain gauge + 1 compensating in the neighboring branch
½ bridge has 2 measuring strain gauges + 2 compensating strain gauges
Theoretically it has a double sensitivity compared to ¼ bridge, the compensating
gauges have to be placed:
- In opposite branch – if the measuring gauges measure the same stress (tensile –
tensile, or compressive – compressive)
- In neighboring branches – if the measuring gauges measure opposite stress
(compressive – tensile)
Full bridge has 4 measuring strain gauges; two have to measure tension, two
compressions. It is self temperature compensating and has theoretically 4 times
sensitivity compared to ¼ bridge.
On the SPIDER unit, the strain gauge bridges are connected to channel 4 and channel 5.
Fig. 3 – ¼ and ½ bridge configurations with temperature
compensation
4th-channel = ½ bridge (in reality it is a full bridge, only 2 strain gauges are active
in this task), configuration name is TEN55 in the software
5th-channel = ¼ bridge (in reality it is a full bridge, only 1 strain gauge is active in
this task), configuration name is TEN55 in the software
The procedure is the following:
Change deflection with step 1 mm from – 5 mm to + 5 mm and measure bridge
output voltage Uv for both ¼ and ½ bridge.
If non-zero value appears for deflection 0 try the RESET button in the software.
DON´T SET the deflection BY MEANS of the MICROMETRICAL SCREW
ALONE!! It is not a motion screw. Instead, bend lightly the beam with a thermally non-
conductive material (pen, pencil, …) , move the micrometrical screw to the desired position
and then release the beam.
Conclusions
Compare sensitivity of ¼ and ½ bridge configurations, compare sensitivities in the I.
and III. quadrant.
Torque
Introduction
Strain gauges can also be used to measure torque. They have to be placed in a
special configuration on the object – placed under a 45° angle. Typically only full bridges are
used in order to obtain the highest possible sensitivity and to assure temperature
compensation.
Tasks:
1. Calibrate the torque sensor -
measure and plot
dependence of bridge output
voltage on applied torque
V = f(T). Use load force in
range m = (0 – 1000) g. 2. Brake the motor and with
the calibration curve
measure the torque.
Used instruments:
Fixture for torsion load with DC motor, strain gauges (full bridge) and brake
Power supply Diametral P230R51D
Digital voltmeter Agilent 34461A
Procedure description
Calibration
1) Make sure, the DC motor is unplugged from the 12 V power supply
2) Screw on the lever-arm
3) Turn on the power supply and voltmeter
4) Hook different weights on the lever arm, from weight and lever-arm length
calculate torque. The calibration curve is the dependence of bridge output voltage
on applied torque
5) UNSCREW THE LEVER-AMR
Torque
1) MAKE SURE THE LEVER-AMR IS UNSCREWED
2) Connect the DC motor to the 12 V power supply, the motor will start to turn
3) By tightening the nut on the brake, brake the motor to 3 different values of torque.
With the previously obtained calibration curve determine the torque.
Conclusions
Discuss the advantages/disadvantages of a torque sensor with rings and brushes. Find,
describe and reference one industrial sensor for measuring torque on a static (non-rotating)
shaft and one on a rotating shaft. Make a table with sensor parameters. Compare prices.
Fig. 4 – Arrangement of the torque experiment
Force - gauge factor Submitted by:
Task: Date:
2. Measure and plot dependence of bar bend on applied force y = f (G) and bridge output voltage on applied force V =g (G).
3. Test temperature influence – as the last DEMO
Gauge factor
No. FU load [
N]
y deflection
[mm]
RC = R +
567 kW
K [ - ] Kawg [ - ]
1 37,1 0,21 657 2,0596
2 37,1 0,21 652 2,0754
3 37,1 0,21 657 2,0596
dependence of bar bend on applied force y = f (G) and bridge output voltage on applied force V =g (G).
Number of used 0,63
kg discs
Applied
force G (N) ↑ ↓ ↑ ↓ Charts:loading unloading loading unloadin
gplace charts on separate sheet
0
0,000
0 0 0 0,007 0,007 0,007
1 6,180 0,008 0,008 0,008 0,041 0,041 0,041 Sensitivities:2 12,361 0,0015 0,015 0,00825 0,076 0,076 0,076
3 18,541 0,023 0,023 0,023 0,112 0,112 0,112 0,001227 (V/N)
4 24,721 0,03 0,031 0,0305 0,148 0,148 0,148
5 30,902 0,038 0,038 0,038 0,182 0,18 0,181
6 37,082 0,046 0,045 0,0455 0,217 0,217 0,217
0,0056632 (mm/N)
Conclusions: Assess the influence of temperature and the importance of temperature compensation.
average average
1.Measure and plot dependence of strain = f(T) and bridge output voltage V0 = g(T) as a
function of torsion load T on a steel tube.
2,06
Bridge output V [V] y deflection [mm]
2
0
04
c
R rK
R R yh=
+
=∆
∆=
G
Vc
F
=∆
∆=
G
ycy
Force - gauge factor - charts:
0
0,005
0,01
0,015
0,02
0,025
0,03
0,035
0,04
0,045
0,05
0,000 5,000 10,000 15,000 20,000 25,000 30,000 35,000 40,000
V (
V)
G (N)
Bridge output voltage V as function of applied force G
0
0,05
0,1
0,15
0,2
0,25
0,000 5,000 10,000 15,000 20,000 25,000 30,000 35,000 40,000
y (
mm
)
G (N)
Bar bend y as function of applied force G
Force - 1/4 and 1/2 bridge Submitted by:
Task: Date: ########
-5 -4 -3 -2 -1 0 1 2 3 4 5
¼ - bridge -0,48 -0,39 -0,3 -0,21 -0,12 0 0,08 0,16 0,23 0,31 0,39
½- bridge -0,98 -0,78 -0,63 -0,43 -0,24 0 0,16 0,32 0,48 0,64 0,79
Sensitivities: I. Quadrant III. Quadrant
0,078 (V/mm) 0,096 (V/mm)
0,158 (V/mm) 0,196 (V/mm)
Charts: place charts on separate sheet
Conclusions:Compare sensitivity of ¼ and ½ bridge configurations, compare sensitivities in the I. and III. quadrant.
1.Measure and plot dependence of strain = f(T) and bridge output voltage
V0 = g(T) as a function of torsion load T on a steel tube.
2. Compare sensitivities for compressive and tensile stress – compare sensitivities in
the I. and III. quadrant. Interpolate with a straight line with the linear regression method
y deflection [mm]
V [V]
=∆
∆=
y
Vc
2/1
=∆
∆=
y
Vc
4/1 =∆
∆=
y
Vc
4/1
=∆
∆=
y
Vc
2/1
Force - 1/4 and 1/2 bridge
-1,2-1
-0,8-0,6-0,4-0,2
00,20,40,60,81
-5 -4 -3 -2 -1 0 1 2 3 4 5
V (
V)
y deflection [mm]
Output voltage of a 1/4 and 1/2 bridge
¼ - bridge ½- bridge
Torque Submitted by:
Tasks: Date:
Charts:
Calibration
lever-
arm
(mm): 150
load Force Torque Voltage
m (g) (N) T (Nm) V (mV)
0 0 0,0 0,636 0,00
200 1,962 0,3 0,903 0,27
400 3,924 0,6 1,165 0,53
600 5,886 0,9 1,417 0,78
800 7,848 1,2 1,675 1,04
1000 9,81 1,5 1,930 1,29
Braking
# Voltage Torque
(-) V (mV) T (Nm)
1
2
3
Závěr:
1.Calibrate the torque sensor - measure and plot dependence of bridge output voltage on
applied torque V = f(T). Use load force in range m = (0 – 1000) g.
2. Brake the motor and with the calibration curve measure the torque.
Discuss the advantages/disadvantages of a torque sensor with rings and brushes. Find,
describe and reference one industrial sensor for measuring torque on a static (non-rotating)
shaft and one on a rotating shaft. Make a table with sensor parameters. Compare prices.
0,0
0,2
0,4
0,6
0,8
1,0
1,2
1,4
1,6
0,0 0,5 1,0 1,5 2,0 2,5
V0
(m
V)
T (Nm)
Bridge output voltage V0 as a function of torsional load force F
Position – LVDT
Introduction
The LVDT (Linear Variable Differential Transformer) is a linear displacement position
sensor. It transfers the movement of the transformer core into the output signal. The output signal is a
voltage signal; in your case the used sensor has build in electronics to transfer the voltage to current
output. Sensors with current output have a limit of maximal resistance that can be connected to
sensors output. The ideal load for a current output sensor is resistance zero; the maximal load is
limited by the available voltage for output. When the output loop resistance is increasing, the output
compensates by increasing output voltage so as to keep constant current. When the output is already on
the maximal voltage, the output current starts to drop. The maximal resistance connected to the current
output limits the wiring length between the sensor and gauge.
Tasks:
1. Measure and plot the static
characteristic I = f(x) for load
resistor RZ = 0.
2. For constant position (x = 100
mm) determine the maximum
value of Rz, until which the
output current remains constant
and the sensitivity of the
transducer.
Used instruments:
Fixture with LVDT
DC ampere meter
Decade resistor Rz
Procedure description
The industry standard for current output is 0 – 20 mA or 4 – 20 mA. It is defined by ANSI/ISA–
50.00.01–1975 (R2002) standard and has wide industrial usage. The advantage is that the reading is not
dependent on the wire length between the sensor and gauge. The limitation of wire length is imposed by
the connecting wire resistance and by the maximal load resistance of the transmitter.
Fig. 1 – Inside an LVDT
A typical connection of
such sensor in a controller system
is shown in Fig. 2. In order to
push a constant current through
the loop (dependent only on the
measured variable), the
transmitter is adjusting its output
voltage.
When the output loop
resistance is increasing (e.g. by
changes of temperature, oxidation
etc.), the output compensates by
increasing output voltage so as to
keep constant current. When the
output is already on the maximal
voltage, the output current starts
to drop.
Conclusions
Calculate the maximal allowed length of AWG 36 [1] copper wire between sensor and 250 Ω
load (gauge) when the ANSI/ISA–50.00.01–1975 (R2002) standard [2] defines maximal load resistance
600Ω.
Discuss advantages and disadvantages of sensors with current and voltage output.
References
[1] American wire gauge, online on < https://en.wikipedia.org/wiki/American_wire_gauge>,
accessed on 31.3.2013
[2] ANSI/ISA–50.00.01–1975 (R2002), Compatibility of Analog Signals for Electronic Industrial
Process Instruments, online on <http://www.isa.org>, accessed on 31.3.2013
[3] Understanding 4-20 mA Current Loops, Application note, BAPI, rev. 10/05/06, online on <
http://www.bapihvac.com/CatalogPDFs/I_App_Notes/Understanding_Current_Loops.pdf>,
accessed on 31.3.2013
Fig. 2 – Typical connection of current loop [3]
Position – Resistive sensor
Introduction
Resistive position sensors are simple, robust and reliable. They have a voltage output;
the ideal load resistance is infinity. Practically the load resistance should be as high as possible, so a
digital voltmeter should be used. Otherwise, the linearity of the sensors’ steady state characteristic is
compromised.
Tasks:
1. Measure and plot the static
characteristic V2=f(x) with a
digital and an analog voltmeter.
2. Compute the maximal error in
% for K=1.2 and compare with
measured value of the error
from graph.
3. Compute the value of K so that
maximal error is less than 1 %.
Used instruments:
Fixture with 5k resistive position sensor Vishay SFERNICE 115L 14E 502 W06017, 5 kΩ, 330
mm
Analog DC voltmeter, range 6 V, 1kΩ/V
Digital multimeter, Ri = 10 MΩ
5 V power supply
Procedure description
The output voltage from a resistive sensor can be calculated from the voltage divider
2
2 2
2
2 1 2 1 21
2
L
L L
L L L
L
R R
R R R RV V V
R R R R R R R RR
R R
(1)
2 0R xR 1 0 2 0 0 01R R R R xR x R 0
LRK
R
(2)
After substitution and simplification, output voltage is
2
1
K xV V
x x K
(3)
Fig. 3 – Task schematic diagram
Relative error is
2
21
% 100 1001
x xV x V
V x x K
(4)
The relative error is zero for K -> . The relative error in this case is always negative (we measure
always lower voltage than would correspond to linear dependence.
Maximal relative error is for
0 2 / 3xx
(5)
Conclusions
Find and reference at least two
industrial versions of resistive position
sensors. Make a table with range,
resistance, accuracy and price.
Fig. 4 – Influence of K on output voltage
Position – Repeatability and accuracy
Introduction
Repeatability is the closeness of the agreement between the results of successive
measurements of the same measurand carried out under the same conditions of measurement [4]. These
conditions are called "repeatability conditions". Repeatability conditions include the same measurement
procedure, the same observer, the same measuring instrument used under the same conditions, the same
location, and repetition over a short period of time. Repeatability may be expressed quantitatively in
terms of the dispersion characteristics of the results.
Tasks:
1. Determine repeatability of the x-axis
of a 3D printer. Check repeatability
in two different points; make 10
measurements in every point.
2. Calculate sample mean and standard
deviation of position in both checked
points.
Used instruments:
3D printer + PC
Variable resistance position sensor
VISHAY SFERNICE - 115L 14E 502
W06017 - TRANSDUCER, LINEAR,
350MM RANGE, 5KOHM
Digital Multimeter Agilent 34461A,
6½ Digit
Procedure description
The x-axis of the 3D printer is
equipped with a variable resistance position
sensor. The resistance (position) is evaluated with the Agilent 34461A digital multimeter.
1) Launch the 3D printer controller software 3D COM Terminal with an icon on the desktop
2) Select port COM4 (“Komunikační port”) and open it (“Otevřít”)
3) Align the printer axes with the button “Zarovnat osy”
4) Wait until alignment is finished
5) In the “G-kód” window write command “G3 X1000” and send with button “Odeslat”. This
will move x-axis to position 1000. Wait until the movement is finished; record the resistance
from the multimeter.
6) In the “G-kód” window write command “G3 X2000” and send with button “Odeslat”. This
will move x-axis to position 2000. Wait until the movement is finished; record the resistance
from the multimeter.
7) Repeat steps 5) – 6) nine more times
Fig. 5 – Repeatability and accuracy
Sample mean: Standard deviation
1
1 n
i
i
x xn
2
1
1
1
n
i
i
s x xn
(6)
Conclusions
In conclusions discus whether the 3D printer axis is accurate and repeatable. Discus the obtained
accuracy and repeatability compared to the desired accuracy 0.5 mm.
References
[4] Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results,
Appendix D, online on < http://physics.nist.gov/Pubs/guidelines/appd.1.html>, accessed on
9.11.2013
Position – LVDT Submitted by:
Task: Date:
x [mm] 0 15 30 45 60 75 90 105 120 135 150
RZ=0 Ω I [mA] 19,41 19,24 17,73 16,34 14,85 13,29 11,82 10,34 8,85 7,35 5,71
Sensitivity: -0,1 mA/mm Used instruments:
Charts:
Conclusions:
1. Measure and plot the static characteristic I = f(x) for load resistor RZ = 0.
2. For constant position (x = 100 mm) determine the maximum value of Rz, until
which the output current remains constant and the sensitivity of the transducer.
Calculate the maximal allowed length of AWG 36 [1] copper wire between sensor and 250 Ω load (gauge) when
the ANSI/ISA–50.00.01–1975 (R2002) standard [2] defines maximal load resistance 600Ω. Discuss advantages
and disadvantages of sensors with current and voltage output.
0
5
10
15
20
25
0 15 30 45 60 75 90 105 120 135 150
Ou
tpu
t cu
rren
t (m
A)
angular position α (°)
RZ=0 W
RZ=0 W
Position – Resistive sensor Submitted by:
Task: Date:
Used instruments:
x [°] 0 30 60 90 120 150 180 210 240 270 300 330
V2[V] (digital) 0,001 0,367 0,784 1,199 1,624 2,047 2,473 2,889 3,313 3,739 4,164,6
V2[V] (analog) 0 0,3 0,51 0,69 0,86 1,725 2,05 2,4 2,8 3,25 3,754,4
Charts:
Maximal error: K for error less than 1 %:
Conclusions:
2. Compute the maximal error in % for K=1.2 and compare with
measured value of the error from graph.
Find and reference at least two industrial versions of resistive position sensors. Make a table with range, resistance,
accuracy and price.
3. Compute the value of K so that maximal error is less than 1 %.
1.Measure and plot the static characteristic V2=f(x) with a digital and an analog
voltmeter.
0
0,5
1
1,5
2
2,5
3
3,5
4
4,5
0 50 100 150 200 250 300 350
Ou
tpu
t vo
lta
ge
V2
(V
)
α (°)
V2[V] (digital)
V2[V] (analog)
Position – Repeatability and accuracy Submitted by:
Task: Date:
experiment 1 2 3 4 5 6
pos. X100 1,137 1,139 1,14 1,139 1,139 1,14
pos. X500 1,975 1,975 1,973 1,974 1,975 1,975
Sample mean: Used instruments:X100 1,1390
X500 1,9745
Standard deviation:
X100 1,0954E-03
X500 8,3666E-04
Conclusions:
1. Determine repeatability of the x-axis of a 3D printer. Check repeatability in two
different points; make 10 measurements in every point.
In conclusions discus whether the 3D printer axis is accurate and repeatable. Discus the obtained accuracy and
repeatability compared to the desired accuracy 0.5 mm.
2. Calculate sample mean and standard deviation of position in both checked points.
Position – resolver
Introduction
Resolver is an angular position sensor. It is an absolute sensor within one
revolution. The construction is similar to an electrical motor. It has typically one rotor
winding and two stator windings. The rotor winding is powered with AC voltage with
frequency 2 to 2 kHz. The two stator windings are perpendicular to each other; AC voltage is
induced through inductive coupling from the rotor. The dependence of stator voltages on
position is sinusoidal. Both stator windings’ signals are required to use the resolver as
absolute sensor from 0 to 360°.
Task:
1. Measure the RMS voltage
of both stator windings in
dependence on angle
V1=f(α) , V2=g(α).
Measure the angle α by
step of 15° for one whole
revolution.
Used instruments:
Fixture with resolver
Oscilloscope with signal
generator
Procedure description
Resolver is electrical machine. It has one rotor winding two stator windings. The stator
windings are perpendicular to each other. The rotor winding is supplied from sinusoidal
power supply by frequency approximately 10 kHz. This is called reference voltage.
_ max sin( )gen genV V t (1)
The fixture is supplied from the oscilloscope generator in this task. In an industrial
system, the signal has to be provided.
The change of mutual inductance between rotor and stator is sinusoidal with
dependence on angle α. Induced voltages from rotor to both stator windings follow the same
dependence. Because the stator windings are mutually shifted, the induced voltages are
shifted too. As the shift is 90°, when one stator voltage is maximal (the mutual inductance
between this stator winding and rotor winding is maximal), the induced voltage in the second
stator winding is minimal (the mutual inductance between this winding and rotor winding is
minimal).
1 1_ max sin( ) sin( )V V t (2)
2 2_ max sin( ) cos( )V V t (3)
Fig. 1 – Task schematic diagram
To evaluate the actual angle, one has to measure both stator voltages. Moreover the
dependence is sinusoidal. As we desire from many sensors a linear steady state characteristic,
the signal has to be processed.
The electronics is called “Resolver to digital” converter [1]. It samples both stator
voltages with Analog/digital converters and uses a look-up table do convert to position. The
output is then a digital signal.
The recommended procedure for this task is to read both stator voltages from the
oscilloscope as peak-peak voltages and convert them to RMS voltage with the equation
(valid only for sinusoidal voltage)
/ 2 2RMS p pV V (4)
Conclusions
In conclusions compare resolver and selsyn(synchro). Find and reference at least one
“Resolver-to-digital” converter. Find out the frequency of the reference signal and resolution
(positions per revolution).
References
[1] AD2S1220, Variable Resolution, 10-Bit to 16-Bit R/D Converter with Reference
Oscillator, online on < http://www.analog.com/static/imported-
files/data_sheets/AD2S1210.pdf>, accessed on 7.4.2013
Dimensions – Ultrasonic thickness sensor
Introduction
Ultrasonic thickness sensors are basically composed from a transmitter and a
receiver. The transmitter produces a pulse; this is reflected at the object boundary and
detected by the receiver. The time delay between transmission and reception is proportional to
object thickness and speed of sound in the material. Ultrasonic thickness sensors are used in
places where traditional thickness meters, like a caliper cannot be used e.g. wall thickness of a
closed box. The same principle is used also to detect cracks in materials.
Tasks:
1. Measure the thickness of
enclosed steel plates in five
points. Verify thickness by
micrometer in the same
points.
2. Measure the wall thickness
profile of two steel tubes in
step of 10 mm, measured
on tube perimeter. Draw
thickness profile in polar
coordinates.
Used instruments:
Ultrasonic thickness meter DIO 570
Procedure description
It is necessary to lubricate by grease (e.g. INDULONA) the measured elements to
improve contact with the ultrasonic probe.
Measure the plate thickness in 5
indicated points. Verify at the same points
with a micrometer.
For steel tube wall thickness
measurement, place the probe
perpendicularly to the surface.
Wipe clean the ultrasonic probe
carefully after measurement is finished.
Demo: Measure the dimensions of the laboratory with an ultrasonic distance meter and
laser distance meter.
Conclusions
Discuss errors in the measurement of wall tube thickness that impact accuracy.
Propose a solution. Compare the accuracy of the ultrasonic and laser distance meter. Find and
reference one ultrasonic and one laser distance meter, compare theirs accuracies, ranges and
usage.
Fig. 2 – Principle of ultrasonic thickness sensor
Fig. 3 – Points to be measured on steel plates
3D scanner
Introduction
A 3D scanner is used to create 3D models from real world objects. It can easily
be built with a laser beam projector, camera and rotational table. The object to be scanned is
placed on the rotational table. The laser beam projector projects a line on the object. The
camera records the line shape as the object is rotated. From the known angle between laser
and camera, the 3D model of the object can be reconstructed.
Task:
1. Scan a polystyrene head
Used instruments:
Rotational table with
polystyrene head
Web camera + PC
Laser water level
Description of principle
The camera sees the image in
its image plane. The angle Θ between
the laser line projector axis and camera
axis is known. The distance from the
axis of rotation r can be calculated as follows
/ sin( )r x (5)
Where x is the position of the laser point as
seen by the camera.
This procedure is repeated for all vertical
points of the laser line.
Procedure description
1) Launch the program “3DScanner Gui”. Be
patient, the start takes a while. 2) With the button “Configure webcam”
configure the webcam. Select device
winvideo, device ID 1, format
“YUV2_640x480”
3) Launch preview with button “Start preview”
4) Turn on the laser water level
5) Dependent on the current light conditions,
adjust the line detection threshold to get a
clear laser line. The processed image with the
laser line is obtained with the button
“Capture”
6) Turn on the power supply, adjust voltage to set
speed 1 revolution/minute (voltage around
Fig. 4 – Principle of 3D scanner [2]
Fig. 5 – Principle of 3D scanner, image of
head from [5]
4V)
7) Start the 3D image capture with the button “Timer ON”. The 3D image is made from
120 images, captured with period 0.5s. When satisfied with the image, stop the capture
with the “Timer OFF” button and save the image to a file.
Conclusions
In conclusions discus the influence of object color on the sensor reading, discuss the
difference between CCD and PSD technology. Made and estimate of precision of this simple
scanner.
References [2] 3D Scanning Basics, online on < http://www.etc.cmu.edu/projects/plastico-
fantastico/?p=295>, accessed on 7.7.2013
[3] Polhemus FastSCAN 3D Laser Scanner, online on <
http://www.youtube.com/watch?v=SyzgBycPxyw>, accessed on 7.4.2013
[4] Make a 3D Laser Scanner, online on < http://www.youtube.com/watch?v=SPywgDBjM1Y>,
accessed on 7.4.2013
[5] Human Head Modeling – 3DS Max, online on < http://www.tutorius.net/2010/03/human-
head-modeling-3ds-max/comment-page-1/>, accessed on 7.4.2013
Fig. 6 – 3D Scanner GUI
Position – resolver Submitted by:
Task: Date:
α [°] V1 p-p [V] V2 p-p [V] V1 RMS [V] V2 RMS [V]
0 0,620 0,500 0,219 0,177
15 0,480 0,640 0,170 0,226
30 0,320 0,740 0,113 0,262
45 0,100 0,800 0,035 0,283 Used instruments:60 0,124 0,800 0,044 0,283
75 0,320 0,740 0,113 0,262
90 0,400 0,540 0,141 0,191
105 0,580 0,460 0,205 0,163 Charts:
120 0,760 0,280 0,269 0,099
135 0,720 0,080 0,255 0,028
150 0,720 0,100 0,255 0,035
165 0,680 0,300 0,240 0,106
180 0,580 0,440 0,205 0,156
195 0,440 0,580 0,156 0,205
210 0,280 0,660 0,099 0,233
225 0,100 0,710 0,035 0,251
240 0,092 0,720 0,033 0,255
255 0,280 0,660 0,099 0,233
270 0,440 0,560 0,156 0,198
285 0,560 0,440 0,198 0,156
300 0,660 0,280 0,233 0,099
315 0,720 0,084 0,255 0,030
330 0,710 0,160 0,251 0,057 Conclusions:345 0,660 0,290 0,233 0,103
360 0,560 0,450 0,198 0,159
1. Measure the RMS voltage of both stator windings in dependence on angle V1=f(α) ,
V2=g(α). Measure the angle α by step of 15° for one whole revolution.
In conclusions compare resolver and selsyn(synchro). Find
and reference at least one “Resolver-to-digital” converter.
Find out the frequency of the reference signal and resolution
(positions per revolution).
( )/ 2 2RMS p pV V
−
= ⋅
0,000
0,050
0,100
0,150
0,200
0,250
0,300
0 100 200 300
Ou
tpu
t v
olt
ag
e (
V)
angular position α (°)
V1 RMS [V]
V2 RMS [V]
Dimensions – Ultrasonic thickness sensor Submitted by:
Task: Date:
Used instruments:
steel plate 1 Charts:
point 1 2 3 4 5
ultrasonic (mm) 5,4 5,3 5,3 5,3 5,3
micrometer (mm) 5,20 5,17 5,23 5,17 5,20
steel plate 2point 1 2 3 4 5
ultrasonic (mm) 4,8 4,8 4,8 4,9 4,9
micrometer (mm) 4,72 4,74 4,75 4,72 4,72
tube 1 - wall thickness
point 1 2 3 4 5 6 7 8 9 10 11 12 13 14
ultrasonic (mm) 3,1 2,9 3,1 3,1 3,6 3,6 3,6 3,6 3,4 3,4 3,3 3,4 3,4 3,3
tube 2 - wall thicknesspoint 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
ultrasonic (mm)
steel box: wall thickness :
Conclusions:
1. Measure the thickness of enclosed steel plates in five points. Verify thickness by
micrometer in the same points.
2. Measure the wall thickness profile of two steel tubes in step of 10 mm, measured
on tube perimeter. Draw thickness profile in polar coordinates.
Discuss errors in the measurement of wall tube thickness that impact accuracy. Propose a solution. Compare the accuracy of the
ultrasonic and laser distance meter. Find and reference one ultrasonic and one laser distance meter, compare theirs accuracies,
ranges and usage.
0
1
2
3
4
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Tube wall thickness profile in mm
3D scanner Submitted by:
Task: Date:
3D image:
Precision estimation: 5 mm
Conclusions:
1. Scan a polystyrene head
In conclusions discus the influence of object color on the sensor reading, discuss the difference between
CCD and PSD technology. Made and estimate of precision of this simple scanner.
Pressure - calibration
Introduction
When calibrating an instrument, readings from two gauges are compared. One
is the calibrated one; the other has to use a definition principle for the calibrated physical
property. In case of pressure the definition is force acting on a known area (used in e.g.
deadweight tester) or equivalent to this - hydrostatic pressure. The reading from the definition
instrument is then taken as “correct”. The calculated deviations are then used to add/subtract a
correction from the meter reading in order to obtain the correct pressure.
Tasks:
1. Draw task´s block diagram
2. Calibrate deformation manometer in 11
points of whole range including zero and
maximal pressure 100 kPa.
3. Plot graph of absolute errors.
Used instruments:
manometer to be calibrated = deformation
manometer with strain gauge transducer,
accuracy class = 0,5
liquid mercury manometer (well type)
control deformation manometer with Bourdon tube – including valve for pressure
setting, accuracy class = 0,6
Procedure description
Many industrial pressure gauges use the deformation principle. Pressure is transduced
into a deformation of a Bourdon element, spring bellows or a diaphragm. Although this
principle is robust and reliable, it is not a definition principle for pressure and all such
gauges need to be calibrated. An example of a diaphragm pressure gauge is shown in Fig. 1.
The position of the diaphragm is measured. Either mechanically like in Fig. 1, or by means of
e.g. strain gauge or capacitive sensors.
The calibration gauge used in this task is a
well type manometer shown in Fig. 2.
Liquid column height h is read from the
scale.
The pressure is then
p g h (1)
Where g is gravitational acceleration and ρ
is density of the used liquid.
Mercury (Hg) is used in our manometer.
Table 1 - Hg density in dependence on temperature
Fig. 1 – Diaphragm pressure gauge
Fig. 2 – Well type manometer
t(°C) 0 10 20 30
p (kg.m3) 13595.1 13570.4 13545.7 13521.2
Fig. 3 – Correct reading Fig. 4 – Avoid parallax error
Conclusions
In conclusion discuss what properties are required for a calibration instrument (from
accuracy and range point of view). Find, describe and reference at least two either industrial
or automotive applications of deformation pressure sensors.
Pressure - verification
Introduction
For instrument verification a more precise instrument is required. Its reading is
compared with the device under test. None of both uses the definition principle so it is
different from calibration. Both instruments need to have approximately the same range; the
more precise instrument needs a higher accuracy. The reading from the more precise
instrument is then taken as “correct”. Verification serves to check whether the meter can be
trusted, usually in multiple points of the scale.
Tasks:
1. Draw task´s block diagram
2. Verify accuracy class of deformation
manometer. Make verification by
comparing pressure with control
manometer in 5 points in the whole range
of pressure - including maximal pressure
(but not zero pressure).
3. Create verification protocol.
Used instruments:
tested deformation manometer (diaphragm,
Tp = 1,5)
Smart pressure difference transducer
Honeywell ST3000, type STD924,
accuracy 0.1%. Variable range, basic range
100" inch H2O, with handheld unit
Honeywell SFC
24 VDC power supply
current supply for 4 - 20 mA current loop
control deformation manometer with Bourdon tube – including valve for pressure
setting, accuracy class = 0,6
Procedure description
The goal is to compare pressure from both meters, the verified one with the more
precise one. The verification protocol will contain calculated values in Pa and absolute
and relative errors. Take the reading from the Smart sensor as the correct ones.
Conclusions
In conclusion state whether the tested deformation manometer has the declared
accuracy class Tp = 1.5 or not. Discuss the difference between calibration and verification.
References
[1] Pressure transmitters lower plant lifecycle costs, online on
<http://www.myflukestore.com/crm_uploads/fe_576_users_manual.pdf >, accessed on
31.3.2013
Fig. 5 – Smart pressure sensor [1]
Humidity
Introduction
Humidity is closely related to pressure and temperature. Absolute humidity is
the mass of water vapor - mh20 - present in the air water vapor mixture [2]. Relative humidity
is the ratio of the partial pressure of water vapor in an air-water mixture to the saturated vapor
pressure of water at a prescribed temperature [3]. Relative humidity is usually expressed in
per cent and abbreviated by φ or RH.
Tasks:
1. Measure transient response of
electronic hygrometers, determine time
constant.
2. Measure the air humidity in the lab
with a psychrometer
Used instruments:
Climatic chamber
Kir Sensirion Sensirion EKH-4 with sensors SHT21 (0 – 100 % RH ; ±2.0 %) ,
SHT71 (0 – 100 % RH ; ±3.0 %) , SHT21 (0 – 100 % RH ; ±1.8 %)
PC + software EHK viewer
aspiration psychrometer with mercury thermometers, aspiration psychrometer with
thermistor temperature transducers
Procedure description – electronic hygrometers
One of the principles used in electronic hygrometers is based on changes of capacity
or resistance of a dielectric material. The dielectric is Al2O3. The dielectric is placed between
one porous and one non-porous electros. The humidity from the air can enter into the
dielectric material through the porous electrode and will cause changes of capacity/resistance.
This change is evaluated. In our case the build-in electronic communicates through I2C bus.
Steps – electronic hygrometers
1) If not done so by the previous group – pull out the sensor from the climatic chamber.
2) Turn on the power supply for the EHK-4 it, on the PC start EHK4 viewer.
3) On the power supply for the ultrasonic humidifier set voltage 20V, connect the power
supply to the humidifier, and turn on the fan (voltage 9,2V).
4) Wait 3 minutes to reach steady state. In the meantime, use a psychrometer to
measure air humidity in the lab.
5) In menu File -> Log to file select the file where the recorded data will be stored.
On the PC start the data recording with the START button.
6) Insert the sensors into the climatic chamber, measure transient response, remove the
sensors from the climatic chamber and measure the transient response as well. The
recording is stopped with the STOP button.
Conclusions – electronic hygrometers
Explain why the time constant is different for adsorption and desorption.
Fig. 6 – Principle of Al2O3 hygrometer
Procedure description – psychrometer
The principle of psychrometer is
shown in Fig. 7. The instrument is composed
from 2 thermometers. One is called dry
(measures the temperature of the air where
humidity is to be measured), the other is called
wet (measures temperature of air with relative
humidity 100 %). From temperature
difference Δt and dry thermometer reading
tdry, relative humidity RH can be found in a
psychrometric chart or table.
Conclusions – psychrometer
Discuss differences in RH reading from
different psychrometers.
References
[2] Absolute Humidity of Air, online on < http://www.engineeringtoolbox.com/absolute-
humidity-air-d_681.html>, accessed on 31.3.2013
[3] Relative humidity, online on < https://en.wikipedia.org/wiki/Relative_humidity>, accessed on
31.3.2013
Fig. 7 – Principle of psychrometer
Pressure - calibration Submitted by:
Task: Date:
2. Calibrate deformation manometer in 11 points of whole range including zero and maximal pressure 100 kPa.
3. Plot graph of absolute errors.
Block diagram:deformation
manometer
(kPa)
well type
manometer
(mm Hg)
well type
manometer
(kPa)
absolute
error
(kPa)
4,8 37,0 4,9 -0,1
10 77,0 10,2 -0,2
20 151,0 20,1 -0,1
30 226,0 30,0 0,0
40 303,0 40,3 -0,3 Charts:50 379 50,4 -0,4
60 452 60,1 -0,1
70 529 70,3 -0,3
80 604,0 80,3 -0,3
90 680,0 90,4 -0,4
100 755,0 100,3 -0,3
lab. air temperature t(°C) 22,0
lab. barometric pressure(kPa) 96292
Used instruments:
Conclusions:
1. Draw task´s block diagram
In conclusion discuss what properties are required for a calibration instrument (from accuracy and range point
of view). Find, describe and reference at least two either industrial or automotive applications of deformation
pressure sensors.
Laboratory conditions:
-0,4
-0,4
-0,3
-0,3
-0,2
-0,2
-0,1
-0,1
0,0
0 20 40 60 80 100 120
absolute error (kPa)
absolute error (kPa)
Pressure - verification Submitted by:
Task: Date:
3. Create verification protocol.
Verificaton protocol: Block diagram:deformation
manometer
(mm H2O)
deformation
manometer
(kPa)
"Smart"
pressure
reference
(kPa)
absolute
error
(kPa)
relative
error (%)
500 4,9 6,1 -1,2 -24,2
1000 9,8 12,5 -2,7 -27,5
1500 14,7 19,2 -4,5 -30,3
2000 19,6 27,0 -7,4 -37,8
2500 24,5 34,2 -9,6 -39,3 Charts:
lab. air temperature t(°C) 22,0
lab. barometric pressure(kPa) 96292,0
Used instruments:
Conclusions:
1. Draw task´s block diagram
In conclusion state whether the tested deformation manometer has the declared accuracy class Tp = 1.5 or not.
Discuss the difference between calibration and verification.
2. Verify accuracy class of deformation manometer. Make verification by comparing pressure with
control manometer in 5 points in the whole range of pressure - including maximal pressure (but not
zero pressure).
Laboratory conditions:
-45,0
-40,0
-35,0
-30,0
-25,0
-20,0
-15,0
-10,0
-5,0
0,0
0,0 5,0 10,0 15,0 20,0 25,0 30,0
relative error (%)
relative error (%)
Humidity Submitted by:
Tasks: Date:
Measured transient responses: Used instruments:
SHT25 SHT71 SHT75
lab. air temperature t(°C)
lab. barometric pressure(kPa) lab air humidity - measured with psychrometers
Conclusions
Explain why the time constant is different for adsorption and desorption. Discuss differences in RH reading from different psychrometers.
1. Measure transient response of electronic hygrometers, determine time constant.
2. Measure the air humidity in the lab with a psychrometer
time constant - adsorption
Laboratory conditions:time constant - desorption
30
40
50
60
70
80
90
100
15:00:43 15:01:26 15:02:10 15:02:53 15:03:36 15:04:19 15:05:02
RH
(%
)
time
SHT25
SHT71
SHT75
DC tachogenerator
Introduction
DC generator is in many ways similar to a DC motor/generator. Its main parts
are a permanent magnet, coils, commutator and brushes. As the rotor rotates, the coils are
provided with variable magnetic flux and voltage is induced. The induced voltage is AC. The
commutator acts as a mechanical rectifier. The limited number of commutator lamellas limits
the achievable accuracy. The output voltage is a function of rotor speed. As it depends also on
tachogenerator load, a defined load resistor has to be connected to the output.
Task:
1. Measure steady state
characteristic of a DC
tachogenerator with an
unloaded and loaded
output with RL = 25
kΩ. Calculate
sensitivities for both
cases.
2. Hand draw schematic
diagram of the task
Used instruments:
Fixture with DC motor, DC tachogenerator 80V/1000 RPM, optical speed sensor
Power supply Manson ED-613
Voltmeter UNI-T (range 1000 V dc)
Counter TR-525B/D009
Procedure description
Control the DC motor speed by changing the power supply voltage. Read the correct
speed (frequency) from the counter. Repeat experiment twice – first without the load resistor
24 kΩ, then with the load resistor. Calculate sensitivities.
Conclusions
Discuss what happens with the output voltage when the DC tachogenerator is loaded.
Discuss what limits the accuracy of a DC tachogenerator. Find, describe and reference at least
one industrial DC tachogenerator.
Fig. 1 – Schematic diagram of the laboratory set-up
AC tachogenerator
Introduction
The AC tachogenerator is an electrical machine. Its output voltage and
frequency is a function of speed. Both of them can be used to measure speed.
Tasks:
1. Measure amplitude and
output voltage as a
function of speed for the
AC tachogenerator.
2. Hand draw schematic
diagram of the task
Used instruments:
Power supply Diametral
P230R51D
Voltmeter UNI-T
Oscilloscope GW
INSTEK GOS-620FG
Counter Goldstar FC-2015
Tachometer DT-2236
Fixture with DC motor and AC tachogenerator
Procedure description
Control the DC motor speed by changing the power supply voltage (variable part). Do not
change the second power supply voltage. Read the correct speed from the optical tachometer.
Measure amplitude and output voltage as a function of speed for the AC tachogenerator from
0 to 5000 RPM.
Conclusions
Discuss linearity of voltage and frequency, calculate the number of pole-pairs of the
AC tachogenerator.
Fig. 2 – Arrangement of the experiment
IRC, Hall sensor, stroboscope
Introduction
The IRC (Incremental encoder) is a precise optical sensor. It has typically a
resolution of several thousand of pulses per one revolution. It is a relative sensor. The Hall
sensor uses the variations of a magnetic field in order to detect position (or speed) of a
ferromagnetic mark. The distance between the ferromagnetic wheel and the sensor is most
important. If not set correctly, the sensor will not output a correct signal.
Tasks:
1. For speed 2000 RPM
record oscilloscope
signals 1 and 2 from IRC.
Reverse direction of
rotation and record once
more. Based on those 2
signal propose an.
2. Record the Hall sensor
output signal frequency
for different distance of
the sensor from the
wheel. State for what
distances the sensor is measuring correctly.
Used instruments:
Power supply Manson EP-613
Voltmeter UNI-T
4 channel oscilloscope GW INSTEK GDS-2104
Multimeter Agilent 34401A (used as a frequency-meter)
Stroboscope TR5555
Fixture with DC motor, IRC, Hall sensor
caliper
Fig. 3 – Arrangement of the experiment
Procedure description
By adjusting the voltage on the power supply set motor speed to 2000 RPM. Use the on the
fixture available DC tachogenerator to set approximately the speed. DC tachogenerator
constant is 2V/1000 RPM.
Use the stroboscope to set the precise speed of 2000 RPM.
1. IRC On channel 1 and 2, visualize output signals 1 and 2 from the IRC for both directions of
rotation. Draw the signal for both cases and propose an algorithm able to distinguish left and
right rotation.
2. Hall sensor Synchronize the oscilloscope from channel 3 (with the TRIG button and the on-screen menu)
to visualize the Hall sensor signal. By adjusting the distance of the Hall sensor (pink sensor)
from the ferromagnetic wheel record the signal frequency with a counter and check with the
oscilloscope is the signal is OK (no pulses are missing).
!!!When adjusting the distance STOP THE MOTOR by
turning of the power supply. !!!
Conclusions
Find, describe and reference at least one industrial Hall sensor for speed. Describe its
application and compare available speed range with inductive speed sensor
Accelerometer
Introduction
A multi axis MEMS (Micro Electro Mechanical System) accelerometer is today a
common component of many devices. It is used to control the rotation of a mobile phone
screen or fires the airbag in a crash. Many accelerometers are based on capacitive sensing and
is the inertial force acting on a mass when acceleration is applied. The movement of the mass
is the measured.
Task:
1. Set 5 different inclination
angles of the
accelerometer platform.
Record acceleration in all
axes, calculate pitch and
yaw
Used instruments:
Development board XTRINSIC-
SENSORS-EVK on a inclined
platform (2 axes)
Arduino UNO + servos
PC + Hyperterminal software
Power supply 5 V
Procedure description:
1) Turn on the PC, with an icon on the
desktop run the software
“Hyperterminál”. In menu Soubor -
> Otevřít open profile „Akcelerometr“ (communication speed is 115200 bps). Open
the serial port with icon „Zavolat“.
2) Turn on the 5 V power supply. Push the reset button on XTRINSIC-SENSORS-EVK
board, a text will be shown in the Hyperterminál window.
3) In the Hyperterminal window enter command „S2“. Send with ENTER. The reading
will start in the window and show acceleration in all axes.
4) The second board– Arduino – controls the servos. With a screwdriver set the variable
resistors do different positions, this sets the position of servos (inclination of the
platform) in two axes. Set 5 different inclinations, record acceleration in all axes.
5) Deduce equations for pitch α and yaw β calculations – use figures 5 to 7
6) Calculate pitch α and yaw β of the platform for all 5 inclinations.
Fig. 4 – Principle of capacitive accelerometer [1]
Fug. 5 – Axes orientation
Conclusions
Discuss why differential capacitive sensing is used and in what application such a
system measuring two axial inclination could be used.
References
[1] Rob O'Reilly, Kieran Harney, Analog Devices Inc., Alex Khenkin Sensors, Sonic
Nirvana: MEMS Accelerometers as Acoustic Pickups in Musical Instruments, online
on < http://www.sensorsmag.com/sensors/acceleration-vibration/sonic-nirvana-mems-
accelerometers-acoustic-pickups-musical-i-5852>, accessed 29.11.2013
[2] Tom Lecklider, Measuring Motion, online on <
http://www.evaluationengineering.com/articles/200609/measuring-motion.php >,
accessed 29.11.2013
Fig. 6 – Platform not inclined Fig . 7 – Inclination in XZ axis - pitch
Fig . 8 – Inclination in YZ axis - yaw Fig . 9 – Accelerometer under a microscope [2]
DC tachogenerator Submitted by:
Task: Date:
Charts:speed (RPM) V (V) - unloaded V (V) - loaded
with 24k
0 0 0
500 45 42
1000 88 81
1500 133 122
2000 177 163
2500 222 203
3000 265 245
3500 308 280
4000 340 325
4500 400 364
5000 443 401
Used instruments: Schematic diagram:
Conclusions:
1. Measure steady state characteristic of a DC tachogenerator with an unloaded and loaded output with RL = 25
kΩ. Calculate sensitivity for both cases.
2. Hand draw schematic diagram of the task
Discuss what happens with the output voltage when the
DC tachogenerator is loaded. Discuss what limits the
accuracy of a DC tachogenerator. Find, describe and
reference at least one industrial DC tachogenerator.
0
100
200
300
400
500
0 1000 2000 3000 4000 5000
ou
tpu
t vo
lta
ge
(V
)
speed (RPM)
Steady state characteristics
V (V) - unloaded V (V) - loaded with 24k
AC tachogenerator Submitted by:
Task: Date:
Charts:speed (RPM) V (V) f (Hz)
0 0 0
1000 38 50
2000 77 105
3000 114 151
4000 153 200
5000 191 250
Used instruments:
Schematic diagram:
Conclusions:Discuss linearity of voltage and frequency, calculate the number of pole-pairs of the AC tachogenerator
1. Measure amplitude and output voltage as a function of speed for a AC tachogenerator
2. Draw schematical diagram of the task
0
50
100
150
200
250
300
0 1000 2000 3000 4000 5000vo
lta
ge
(V
), f
req
ue
ncy
(H
z)
speed (RPM)
Steady state characteristics
V (V) f (Hz)
IRC, Hall sensor, stroboscope Submitted by:
Task: Date:
IRC signal - 2000 RPM - left: IRC signal - 2000 RPM - right:
Hall sensor: Used instruments:distance sensor -
whell (mm)
frequency (Hz)working correctly (yes/no)
1,0 550,0 yes
1,4 550,0 yes
1,8 550,0 yes
2,0 550,0 yes
2,5 - no
3,0 - no
Conclusions:
1. For speed 2000 RPM record osciloscope signals 1 and 2 from IRC. Reverse direction of rotation and record
once more. Based on those 2 signal propose an algorithm to distingush left and right rotation
Find, describe and reference at least one industrial Hall sensor for speed. Describe its application and compare available speed range with
inductive speed sensor
2. Record the Hall sensor output signal frequency for different distance of the sensor from the wheel. State for what
distances the sensor is measuring correctly.
Accelerometer Submitted by:
Tasks: Date:
position
acceleration in X:
(mg)
acceleration in
Y: (mg)
acceleration in
Z: (mg)
pitch α (°) yaw β (°)
1 -101 -150 -980 -5,9 -8,7
2 -252 -150 -958 -14,7 -8,9
3 -300 -235 -900 -18,4 -14,6
4 -95 248 -939 -5,8 14,8
5 23 105 -973 1,4 6,2
Used instruments:
Conclusions:
1. Set 5 different inclination angles of the accelerometer platform. Record acceleration in all axes, calculate
pitch and yaw
Discuss why differential capacitive sensing is used and in what application such a system measuring two
axial inclination could be used.