Introduction to Nuclear Magnetic Resonance
• Topics– Nuclear spin and magnetism– Resonance behavior and the Larmor Frequency
• Larmor frequency• flip angle
– Energy Absorption and Emission• NMR spectroscopy • Energy absorption in tissue (safety issues)• Relaxometry
– T1,T2,T2* relaxation
![Page 2: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/2.jpg)
Nuclear Magnetism• Nucleons (protons, neutrons) have a quantum property
known as spin.• Nucleons have been shown to obey Fermi statistics, and thus
have a maximum spin magnitude of 1/2 Bohr magneton. (spin=1/2)
• In the absence of a magnetic field, nuclear spin is not an observable
• In the presence of a homogeneous magnetic field, the energy of the nucleus depends on the relative orientation of the magnetic field and the nuclear spin vector
![Page 3: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/3.jpg)
• M: net magnetization from collection of nuclei• At thermal equilibrium, a sample of N protons in a static
field B0 will have magnetisation
• at room temperature is very small.
Electron paramagnetism dominates nuclear paramagnetismkT3h22
kTIIhN
kTomBh
kTomBhmhNM
3)1(22
)/exp(
)/exp(
Bo
![Page 4: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/4.jpg)
single voxel
net magnetization
Individual nuclear spins
![Page 5: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/5.jpg)
Nuclear Magnetic Resonance
• It is very difficult to observe static nuclear magnetism at room temperatures
• Resonance techniques can dramatically amplify effects
• Note: uncertainty principle: N~1 in radiofrequency regime of NMR, N is very large, s.t. a classical description is valid
![Page 6: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/6.jpg)
• At thermal equilibrium, M is aligned with static field• application of a perturbative field nuclei
experience a torque
• if applied field is rotating about Bo at angular
frequency , (recall that torque is the angular analogue of force: F=ma= ; = )
eBM
dtdP
dtdL
e1 BM
dtdM
dtdL
![Page 7: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/7.jpg)
Bo
B1
Be
Bo = Original static field
BoB1 = Applied perturbing field
Be = Bo + B1 = resulting “effective” field
![Page 8: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/8.jpg)
• Rotating Bapp B1= B1(t)
Bloch Equation
• switching to rotating frame of M ( )
)(e)()( tBtMdt
tdM
dtd
dtd
))(e()()( tBtM
dttdM
![Page 9: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/9.jpg)
• When , M is at rest in rotating frame
• there are two conditions (i.e. solutions) M parallel to Be (only when Be = B0)
=-Be=o Larmor frequency
o is the frequency at which M rotates about Be (~ B0)
0dt
dM
![Page 10: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/10.jpg)
Resonance
• Application of perturbative field at t=0 causes precession of M about Be (net field)
• Resonance occurs when 1=o, since 1 will appear to be stationary in the frame of M
![Page 11: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/11.jpg)
M
B1
M
B1
Net force
![Page 12: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/12.jpg)
• After a time t, the angle of M with respect to B0 is:
t=B1t flip angle
M
B1
Net force M
B1
![Page 13: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/13.jpg)
Nuclear Magnetic Resonance:Properties in Matter
• Energy Absorption– In matter, resonance frequency depends on
magnetic field at the nucleus• in complex molecules, electron moments will alter the
field seen by the nucleus (chemical shift)Absorption spectrum is a reflection of the chemical
composition
![Page 14: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/14.jpg)
Nuclear Magnetic Resonance:Properties in Matter
• Relaxation– After we have delivered energy to the nuclei
in our sample at the Larmor frequency, there are two possible ways for the sample to lose this energy (back to lowest energy state):
• spontaneous emission • induced emission
![Page 15: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/15.jpg)
• Spontaneous emission:– negligible effect at RF frequencies (dominant at visible
frequencies)
• Induced emission– Energy emission requires interaction of the nucleus with
its external environment
The nature of energy emission depends strongly on the environment of the excited nucleus (Relaxation)
3
![Page 16: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/16.jpg)
• NMR Spectroscopy is the study of the chemistry of matter using the NMR absorption spectrum
• Relaxometry is the study of the chemistry of matter using the NMR relaxation properties.
MRI generates tissue contrast based (mostly) on NMR relaxation differences.
![Page 17: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/17.jpg)
NMR in tissue• Protons in water molecules are the dominant nuclear species in
the human body• At 1.5T, 10-6 more protons are aligned with the static field than
anti-aligned at room temperature very small magnetic moment.
• Proton Resonance frequencies: =4257 Hz/gauss0.5T 21.28 MHz1.0T 42.57 MHz1.5T 63.86 MHz (Channel 3!)
![Page 18: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/18.jpg)
NMR Absorption in Tissue
• RF energy at the Larmor frequency will be absorbed by water protons in tissue
• MRI scanner: 16 Kilowatt RF transmitter• Dosage: “Specific Absorption Rate (SAR)”
– mass normalized rate of RF energy coupling to biologic tissue (watts/kg)
![Page 19: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/19.jpg)
Specific Absorption Rate
• Depends on:– frequency– pulse sequence (shape of RF pulse,repetition
time, pulse width)– RF coil– Volume of tissue in coil (i.e. exposed)– resistivity of tissue– geometry (spherical vs. cylindrical volume…)
![Page 20: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/20.jpg)
Specific Absorption Rate
• Regulated by the FDA– 0.4 W/kg averaged over the whole body, or 8.0
W/kg peak SAR in any 1g of tissue, and 3.2 W/kg averaged over the head
– RF energy insufficient to produce a 1o C rise in core temp. and localized heating less than 38o C in the head, 39o C in the trunk, and 40o C in the extremities (except pts. with impaired circulation)
![Page 21: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/21.jpg)
Specific Absorption Rate
221
2 RHSAR tissue density
weightTR
2angle flip2
oH
SAR
Practically:
![Page 22: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/22.jpg)
2.5 W/kg
1 W/kg
Specific Absorption Rate
• RF Heating occurs mostly at the surface
![Page 23: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/23.jpg)
NMR Relaxation• Energy emission occurs through interaction
with environment– time evolution:
– Free Induction Decay solution:
)()()( toBtMdt
tdM
)1
exp(1oMzMTt
))cos(2
exp(oMxyM
toTt
![Page 24: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/24.jpg)
• Longitudinal relaxation
• Transverse relaxation
MzMo 1 exp tT1
1
T2* 1
T2 1
T 2
MxyMo
2
exp t
T*
)cos( to
![Page 25: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/25.jpg)
NMR Relaxation• T1 relaxation
– time constant of recovery of longitudinal component of magnetization
– physics• reflection of spin thermal interactions with the environment
(i.e. the lattice)• induced emission: molecules moving near the Larmor
frequency will induce relaxation– pure water: molecular motion too fast long T1
– solids: molecular motion too slow long T1
– tissue: molecular motion near Larmor freq short T1
• Field strength: fraction of protons moving near Larmor frequency decreases with Ho T1 increases with Ho
![Page 26: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/26.jpg)
NMR Relaxation• T2 Relaxation
– Time constant of disappearance of transverse magnetization
– Geometry dictates that T1 is a part of T2 (as longitudinal component grows, transverse component decays)
T2 is always greater or equal to T1
tenhancemen1
1
2
1TT
![Page 27: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/27.jpg)
NMR Relaxation• T2 Relaxation (cont’d)
– physics:• Induced emission from interactions with immediate surroundings (spin-
spin interactions)• Each nucleus experiences slight, temporary changes in local field due to
slow interactions with other nuclei. This causes temporary changes in Larmor frequency leading to permanent phase dispersion
• Field strength: change in Larmor frequency doesn’t affect much
• T1 versus T2 in tissue– T1 and T2 roughly correlate (e.g. low T1 implies low T2)– T1 = ~5 T2
![Page 28: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/28.jpg)
x
y
x
y
x
y
x
y
Spin Dephasing after excitation
Bo
![Page 29: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/29.jpg)
NMR Relaxation
• Relaxation times (msec)
0.5T 1.0T 1.5T T1 T2 T1 T2 T1 T2
Gray Matter 650 100 800 100 900 100
Muscle 550 50 700 50 880 50
Fat 200 80 250 80 270 80
![Page 30: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/30.jpg)
NMR Relaxation
• T2 versus T2*
– True T2: decay of transverse magnetization due to “natural” processes at the molecular level
– T2*: the observed or effective decay of transverse
magnetization due to magnetic field inhomogeneity and susceptibility effects
2
1
2
1*2
1TTT
![Page 31: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/31.jpg)
• Longitudinal relaxation
• Transverse relaxation
MzMo 1 exp tT1
1
T2* 1
T2 1
T 2
MxyMo
2
exp t
T*
)cos( to
![Page 32: Nmr intro1](https://reader038.vdocuments.site/reader038/viewer/2022110218/58ecc6021a28ab162a8b45a7/html5/thumbnails/32.jpg)
T1 contrast:Inversion-recovery
T2 contrast:Spin Echo
T2* contrast:
Gradient echo
T1
FID
T2
T2*
T2*