Recent Applications
• New Column Extraction Applications at SRS– UTEVA method for Pu/U removal for Pu/U oxides (Impurity
assay)
– New UTEVA method for Pu and U-Isotope Dilution Mass spectrometry in mixed oxides (strip Pu separately using 3M HNO3-0.2MHF)
– Actinides in soil using Diphonix Resin-microwave digestion
– Sequential column extraction in Bioassay Lab for Pu, Np, Am, U plus enhanced Sr method using cartridge technology
– Pu, Am in fecal samples using Diphonix Resin-microwave digestion and TEVA+TRU Resin
![Page 3: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/3.jpg)
Actinides in Urine• Current Urine Methods
– Pu-acidified 500 mL samples/evaporated -anion resin
– U-acidified 50 mL samples/evaporated -anion resin
– Pu/Am/Sr- 500 mL calcium phosphate/ TRU + SR Resin
• Need more efficient, consistent, sequential methods with better Th-228 removal; larger aliquots for special urine samples
![Page 4: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/4.jpg)
Bioassay Urine Methods
• New Methods– Expand use of calcium phosphate precipitation (500 ml to 1000+
mL)
– Improve Th-228 removal
– Enhance vacuum column extraction with cartridge work
– Pu (Np when Pu-236 tracer used) on TEVA Resin
– Pu, Np, U, Am using TEVA/TRU column or cartridge technology
– Pu, Np, U, Am, Sr using TEVA/TRU +SR columns or cartridge technology
![Page 5: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/5.jpg)
Bioassay Application
• Advantages– Faster, more consistency, less acid waste (0.9 L
vs. 6 L per batch); reduced labor costs; sequential analysis; better alpha peak resolution
– Better detection limit for uranium-500 mL+ sample (vs. current 50 mL sample size)
– Cartridge technology more efficient; eliminates large sequential load solutions (evaporation steps)
![Page 6: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/6.jpg)
Bioassay Implementation
Status• Implementation over next 3 to 6 months
• Pu, Np-TEVA -implemented-10/99
• Pu, Np, Sr -TEVA /SR cartridge or column
• Pu, Np, Am -TEVA+TRU cartridge or column
• Pu, Np, Sr, U,Am -TEVA+TRU cart. + SR column (add 2nd small TEVA column to ensure complete Th-228 removal)
![Page 7: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/7.jpg)
Challenges
• Optimize nitrate levels for Pu, Np on TEVA and Sr on SR Resin work using cartridges
• Achieve adequate Th-228 removal for Pu/U work
– successful with second 1 ml TEVA column
• Interface with current electroplating practices/evaluate cerium fluoride ppt
• Improved Pu stripping from TEVA Resin (NH4I or Ti+3)
• Ensure adequate Pu, Np reduction w/o Fe+2 to enable TRU cartridge below TEVA for urine samples
![Page 8: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/8.jpg)
Sample Preparation
• Tracer addition (Pu-242 or Pu-236, U-232, Am-243)
• Routine calcium phosphate precipitation
– 3 mmol Ca (120 mg )+ 15 mmol (NH4)2HPO4
• Precipitate/centrifuge/redissolve/ash
• TEVA+TRU: Pu, Np, Am, U, Sr– Redissolve in 6 mL of 5M to 6M HNO3
– Add 6 mL of 2 to 2.5 M Al(NO3)3 -scrubbed using UTEVA – after valance adj., no additional acidity adjustment needed
• TEVA+SR: Pu, Np, Sr– Redissolve in 5 mL of 5M HNO3
– Add 5 mL of 2 to 2.5 M Al(NO3)3 -scrubbed using UTEVA
– after valance adj., add 3 mL 16M HNO3 (adjust acidity to 4.5M )
![Page 9: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/9.jpg)
Sample Preparation (contd.)
Valence options to adjust Pu and Np for TEVA:
A. 1) Add 1 mL 1.5M ferrous sulfate + 1 mL 1.5M ascorbic acid, wait 3 min.
2) Add 1 mL 3.5M to 4M sodium nitrite
For TRU cartridge below TEVA (no iron added) :
B. 1) Add 0.5 mL 1.5M sulfamic acid , 2mL 1.5M ascorbic acid, wait 3 to 5 min.
2) 2 mL 4 M sodium nitrite, ensuring excess nitrite
![Page 10: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/10.jpg)
![Page 11: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/11.jpg)
Pu, Np on TEVA RESIN (URINE)
![Page 12: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/12.jpg)
TEVA Pu Tracer Recoveries
500 mL urine sample/ Pu-242 tracer= 1.25 dpm / One TEVA Column Fe+AA/+NO2
%Recovery (CeF3 microprecipitation) % Recovery (Electroplating* )
1) 110 1) 84.4
2) 93.3 2) 72.4
3) 92.6 3) 69.3
4) 95.2 4) 69.6
5) 101.5 5) 79.8
6) 99.3 6) 84.5
7) 97.7 7) 79.1
8) 115.4 8) 85.5
9) 107.9 9) 84.8
10) 106.8 10) 77.0
11) 101.6 11) 82.5
12) 102.6
Avg. = 102.0% Avg. = 79.0
*Add 4 mL 0.02M H2SO4 to enhance F removal during solution cleanup for plating
![Page 13: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/13.jpg)
TEVA- Np Spike Recoveries
500 mL urine sample/ Np-237 spike= 1.40 dpm
Single TEVA column Valence adj. : Fe+AA/+`NO2
% Np-237 Recovery (electroplating)
1) 101.6
2) 99.5
3) 71.3 4) 104.5
5) 85.8
6) 78.3
7) 85.9 8) 81.3
9) 77.8
10) 75.3 11) 91.9
12) 105.3
Avg. = 88.2%
![Page 14: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/14.jpg)
Pu, Np/Am, U, Sr on TEVA/TRU RESIN (URINE)
![Page 15: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/15.jpg)
Pu on TEVA RESIN (URINE or FECAL)(2nd Column to Remove all Th-228)
![Page 16: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/16.jpg)
TEVA (w/ 2nd Col.) -Pu Tracer Recoveries
500 mL urine sample/ Pu-242 tracer= 1.25 dpm
With 2nd 1 mL TEVA Column to ensure complete Th-228 removal when U-232 added
1+2: Fe+AA/+NO2
3-7: SA+AA/NO2
%Recovery (CeF3 microprecipitation) % Recovery (Electroplating* )
1) 93.2 1) 66.0
2) 90.0 2) 75.4
3) 84.1 3) 72.7
4) 90.0 4) 71.0
5) 86.0
Avg. = 89.3% 6) 80.9
7) 73.0
Avg. = 75.0%
![Page 17: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/17.jpg)
![Page 18: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/18.jpg)
![Page 19: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/19.jpg)
TRU Resin -Am Tracer Recoveries500 mL urine sample/ Am-243 tracer= 1.55 dpm / TRU cartridge after TEVA SA+AA/+NO2 / load TEVA and TRU at same time/remove TRU cart./elute Am
% Am-243 Recovery (Electroplating* )1) 93.2
2) 92.1
3) 107.4
4) 70.3
5) 102.4
6) 103.0
7) 100.2
8) 103.3
9) 102.6
10) 94.7
Avg. = 96.9%
![Page 20: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/20.jpg)
TRU Resin-U Tracer Recoveries
500 mL urine sample/ U-232 tracer= 0.554 dpm / TRU cartridge after TEVA SA+AA/+NO2 / load TEVA and TRU at same time/remove TRU cart./elute U
% U-232 Recovery (Electroplating* )
1) 97.9
2) 74.1
3) 85.6
4) 102.9
5) 90.6
6) 83.1
7) 57.7
8) 81.0
9) 80.4
10) 93.3
Avg. = 84.7%
![Page 21: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/21.jpg)
Pu/Sr On TEVA/SR RESIN (URINE)
![Page 22: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/22.jpg)
TEVA-Pu/Sr Cartridge Recoveries
• 500 mL urine sample Sr-90 spiked in samples: 206 dpm
Measured dpm % Sr-90 Spike Recovery
171 83.0
172 83.5
155 75.2
160 77.7
173 84.0
160 77.7
159 77.2
Avg 164.3 79.8
Sr-90 Spiked after precipitation:
Measured dpm % Sr-90 Spike Recovery
184 89.3%
188 91.3%
Avg 186 90.3 %
![Page 23: New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site](https://reader036.vdocuments.site/reader036/viewer/2022062517/56649f385503460f94c54ba8/html5/thumbnails/23.jpg)
Summary
• New implementation of rapid column methods in bioassay lab will result in significant time/cost savings
• Pu and Np together with Pu-236 tracer
• Better Th-228 removal using 2nd 1 mL TEVA col. when U-232 tracer added
• Pu, Np and Sr on TEVA+Sr cartridge
• Pu, Np and Am, U on single two stage TEVA-TRU column with Sr-90 on SR Resin after evaporation