Director of CTAssociate Professor of Radiology
MGHemail: [email protected]
Director of CTAssociate Professor of Radiology
MGHemail: [email protected]
New Concept in CT Image New Concept in CT Image Reconstruction (Reconstruction (ASiRASiR) in The ) in The
AbdomenAbdomen
HARVARDMEDICAL SCHOOL
![Page 2: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/2.jpg)
Disclosures
Research Grant Support from GE Healthcare
![Page 3: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/3.jpg)
11%
26%
14% 49%
Percentage of RadiationExposure
ContributionContribution of CT to Medical Radiation
![Page 4: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/4.jpg)
Estimated Effective Doses From Typical Abdominal Radiology Exams
Estimated Effective Doses From Typical Abdominal Radiology Exams
![Page 5: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/5.jpg)
Appropriate UtilizationAppropriate Utilization
Two Pillars of Radiation Dose Reduction
Two Pillars of Radiation Dose Two Pillars of Radiation Dose ReductionReduction
Optimization of CT
Protocols
Optimization of CT
Protocols
+
![Page 6: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/6.jpg)
General StrategiesGeneral Strategies
Faster gantry rotation reducesradiation dose. However, noiseis increased if the other factorsare constant
Decreasing kVp reduces dose and increases tissue contrast; increasing mAs could reduce noise
Off-centering increases dose
Splitting scan volumesincreases dose
“Extra” images = Extra dose
Collimation, beam pitch, table speed
Automatic exposure control
![Page 7: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/7.jpg)
Factors Affecting Radiation DoseFactors Affecting Radiation Dose
DECREASE IN
• Tube current • Tube potential • Gantry rotation time *• Scan length• Overlap scanning• Number of scan phases
DECREASE IN
• Tube current • Tube potential • Gantry rotation time *• Scan length• Overlap scanning• Number of scan phases
INCREASE IN
• Pitch (Table speed) *•Beam collimation
INCREASE IN
• Pitch (Table speed) *•Beam collimation
*Provided other scanning factors are kept constant
*Provided other scanning factors are kept constant
![Page 8: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/8.jpg)
NI-15
NI-35
NI-15
NI-35
Low Dose CT Challenges: Image Quality
Low Dose CT Challenges: Low Dose CT Challenges: Image QualityImage Quality
![Page 9: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/9.jpg)
Low Dose ImagingLow Dose Imaging• Standard Filtered back projection (FBP) treats projection data as if coming form an idealized system
• ASIR ( Adaptive statistical iterative recon.) • FBP with added modeled based noise suppression (ASIR Factor)
![Page 10: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/10.jpg)
x-ray source model
image voxel model
detector model
statistics model
object model
synthesized projection
measured projection
system optics
current estimation of
the object
object update candidate
system statistics
• Iterative reconstruction is a time consuming process.• System optics modeling (orange box) is computationally expensive• In ASIR, we model only the system statistics (green box) and achieve
the dose reduction benefit
Technology Overview
Courtesy: Bob Becket, GE Healthcare
![Page 11: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/11.jpg)
Technology Overview
Data Statistics
Nonlinear Smoothing
System ModelRaw Data
MBIR Image
CostFunction
Optimization(Recon)
• Optimal use of spatially-varying statistical model for stable, low noise, and high resolution IQ
Courtesy: Bob Becket, GE Healthcare
![Page 12: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/12.jpg)
Adaptive Statistical Iterative Reconstruction (ASIR)
• Noise reduction reconstruction method to improve the signal‐to‐noise
• Relies on the accurate modeling of the distribution of noise in the data
Kole et al 2006 Phys Med Biol
Superior Image Quality with Less Noise and 10‐80% less dose
![Page 13: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/13.jpg)
ASIR: Noise Index
ASIR Selection Noise Index Adjustment
HIRES
30% X 1.2 X 1.3
40% X 1.3 X 1.4
50% X 1.4 (50% Dose) X 2
IRFBPASIR
IR40
FBP60
ASIR40%
![Page 14: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/14.jpg)
How to Apply in PracticeBody
Weight Routine NI Low Dose NI ASIR
<135 lbs 28 36 30%
136‐200 lbs 32 40 30%
>201 lbs 35 42 30%
![Page 15: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/15.jpg)
10 ASIR
100 ASIR
20 ASIR
30 ASIR 40 ASIR
60 ASIR 70 ASIR
NO ASIR
50 ASIR
![Page 16: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/16.jpg)
0 ASIR
20 ASIR
40 ASIR
50 ASIR
80 ASIR
30 ASIR
60 ASIR 70
ASIR
ASIR Selection and Image Quality
![Page 17: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/17.jpg)
ATTENUATION VALUES AND ASIR
Mean HU‐ No significant difference
Standard Deviation (SD) ‐ 10 ASIR increment ‐ 0.7HU reduction, (p<0.001)
ASIR Mean SD0 106.9 15.1
10 106.8 13.820 106.7 12.530 106.7 11.340 106.7 1050 106.6 8.860 106.6 7.570 106.5 6.480 106.5 5.390 106.5 4.4100 106.5 3.7
![Page 18: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/18.jpg)
ASIR ApplicationLow dose Low Contrast : Routine Abd‐Pel CT/Venous phase
Lower ASiR (20‐40%)
Low dose high contrast: CTA /Stone/CTC
High ASiR 50‐60% or greater
![Page 19: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/19.jpg)
Radiation Dose – 11.6mSv
Jan 10 2008
Radiation Dose – 5.1 mSv
55% DoseReduction
Dec 20 2008
23 yrs old with Crohn’s Disease. Baseline and FU CT exam
40% ASIR
Crohn’s Disease
![Page 20: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/20.jpg)
DLP 1382.07mSv
DLP 74411.16mSv 81%
July 2 2007 Feb 26 2009
263 lbs
60% ASIR
Kidney Stone
![Page 21: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/21.jpg)
FBP ASIR-60%
5.2 mSv 2.07 mSv
Liver Donor‐ CTC
![Page 22: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/22.jpg)
CTA
Low dose scansNI
NI ASIR
Non contrast 15 30 50
Arterial study 28 32 40
Delays 20 30 50
![Page 23: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/23.jpg)
Routine CT Protocol
% Dose Reduction
*Routine Abdomen
21‐58%
*Stone disease 23‐81%
*CT Enterography
6‐58%
MGH ExperienceMGH Experience‐‐ASIR BenefitsASIR BenefitsCTA Protocol % Dose
ReductionCT AAA (3 phase) 21‐58%
*CTA – Donors (3phase)
35‐54%
CT PE/DVT 7‐64%
*presented at RSNA 2009
![Page 24: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/24.jpg)
FBP ASIR
15 mSv 8.9 mSv
260lbs
Large Body Habitus
![Page 25: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/25.jpg)
Do’s and Dont’s of ASIR
Do’s
Apply ASIR technique after corresponding dose reduction
In multiphasic examinations, apply higher ASIR levels on noncontract studies
In patients undergoing multiple examinations, increase ASIR incrementally on follow up studies
Dont’s
Do not apply higher (> 50 %) ASIR technique on low contrast diagnostic exams
![Page 26: New Concept in CT Image Reconstruction (ASiR) in The ......Standard Deviation (SD) ‐10 ASIR increment ‐0.7HU reduction, (p](https://reader035.vdocuments.site/reader035/viewer/2022081602/614922cc9241b00fbd675d05/html5/thumbnails/26.jpg)
Summary
Low dose CT scans with ASIR technique allows substantial radiation dose reduction for abdominal CT examinations of varying indications.
ASIR enabled CT scans provide diagnostically acceptable image quality comparable to or better than FBP enabled CT scans.
Dual benefits of improved Image Quality and dose for patients with large body habitus
The radiation dose benefits are particularly beneficial in patients undergoing multiple CT scans or multi‐phase exams.