ners/bioe 481 lecture 10 x-ray imaging detectors · lecture 10 x-ray imaging detectors michael...

90
Health System RADIOLOGY RESEARCH Henry Ford NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science [email protected] [email protected]

Upload: vanminh

Post on 25-Jun-2018

219 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

Health System

RADIOLOGY RESEARCH

Henry Ford

NERS/BIOE 481

Lecture 10X-ray Imaging Detectors

Michael Flynn, Adjunct Prof

Nuclear Engr & Rad. Science

[email protected]

[email protected]

Page 2: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

2NERS/BIOE481 - 2017

General Models

Radiographic Imaging: Subject contrast (A) recorded by thedetector (B) is transformed (C) to display values presented (D)for the human visual system (E) and interpretation.

A

B

Radioisotope Imaging: The detector records the radioactivitydistribution by using a multi-hole collimator.

Page 3: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

3NERS/BIOE481 - 2017

Intro Acquisition methods

• Early Digital Radiography methods.

1. Photostimulable phosphor (PSP) imaging(often called Computed Radiography or CR)

Agfa Carestream

Fuji Konica

2. CCD camera with Gd2O2S / CsI

SwissRay Oldelft

Imaging Dynamics DIRA

• Current Digital Radiography Technologies.

3. Amorphous selenium with TFT arrays

Hologic Shimadzu

Siemens Fuji

4. Photodiode arrays coupled with CsI

GE Canon Carestream

Trixell (Siemens, Philips, Thompson)

Page 4: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

4NERS/BIOE481 - 2017

VI.C.1 – Computed Radiography, CR (15 charts)

1) Computed Radiography, CR

a) Photostimulable (storage) Phosphors

b) CR Reader

c) Advanced designs

d) Performance – MTF/DQE

e) artifacts

Page 5: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

5NERS/BIOE481 - 2017

CR history

Storage Phosphor Radiography

(computed radiography)

• The most common digital radiographic modality

• Invented by G.W. Luckey at Eastman Kodak Co. in 1975

• First commercial unit introduced in 1983 (Fuji FCR 101)

• Exposed cassettes areprocessed in a remote readerusing the same operationalparadigm as for processed film.

• Common in medicine & industry.

• Easily replaces film-screencassettes and film processors.

Page 6: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

6NERS/BIOE481 - 2017

• Formation of the latent image• Excitation of e- to the conduction band upon x-ray exposure

• Deposition of the e-’s in the energy traps within theconduction band created by the impurities in the phosphorstructure

• Reading the latent image• stimulation (detrapping) of the trapped e-s with a laser

beam and collecting the PSL signal

a - CR Latent Image

x-rays PSL signal

LaserConduction band

Valence band

Page 7: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

7NERS/BIOE481 - 2017

a - CR Photostimulable luminescence

Photostimulable luminescence (PSL)

• Emission of light (PSL signal) from amaterial (phosphor screen) in an excitedstate (exposed) upon stimulation by visiblelight (scanning laser beam)

• Phosphor BaFX:Eu (X = Br or I)

Page 8: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

8NERS/BIOE481 - 2017

b - CR scanned laser readout

Laser scanned plate readout

• A rotating polygon mirror scans a laser spot across a phosphor plate.

• The stimulated release of light is recorded and the position is registered.

PMT

Beam deflector

LaserSource

Light channeling guide

Plate translation: Sub-scan direction

Laser beam:Scan direction

Output Signal

Reference detector

Beam splitter

Stationary mirror

f-thetalens

Page 9: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

9NERS/BIOE481 - 2017

b - CR specifications.

Size / Type(inches)

Digital MatrixPixelSize(mm)

NyquistFrequency

(cylcles/mm)

File Size(Mbytes)

14 x 17 / ST 1760 x 2140 0.2 2.5 7.533

10 x 12 / ST 1760 x 2010 0.15 3.33 7.075

8 x 10 / ST 2000 x 2510 0.1 5 10.04

8 x 10 / HR 2000 x 2510 0.1 5 10.04

• Typical Signal:

• ~.01 mR minimum to ~10 mR maximum (104 range)

• 10 bit (1024) or 12 bit(4096) log signal recorded.

• Typical Sizes:

Page 10: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

10NERS/BIOE481 - 2017

• Phosphor plate in a standard cassette are exposed usingconventional ‘Buckey’ devices.

• Energy deposited in the plate forms a latent image.

• After a digital image is read, the plate is erased bydepleting all stored energy.

B - CR reader

X-raySystem

1Image

Reader

exposed

erased

2Image

Processing

Phosphorplate

Page 11: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

11NERS/BIOE481 - 2017

B - CR Remote QC stations

Remote QC Stations located nearexam rooms improves CR workflow.

• Worklist access to patient order

• Identification of view

• Image approval (processing adj.)

• Exam completion

CR

QC – Rm 1 QC – Rm 2QC – Rm 2

Page 12: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

12NERS/BIOE481 - 2017

c - CR new methods: scanhead readout (1D)

Improved readout speedand a more compactdesign has been recentlyachieved using scannedlinear array stimulationand readout.

From Seibert, 2007

Page 13: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

13NERS/BIOE481 - 2017

c - CR new methods: needle phosphor material

Introduced recently by Agfa, CsBr:Eu2+ needleshaped storage phosphor crystals permit thickerphosphor screens for same light blur in comparisonwith traditional granular phosphor screens.

Illustration from:

• US Patent 6967339

• 22-NOV-2005

• Agfa-Gevaert,Mortsel (BE)

Page 14: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

14NERS/BIOE481 - 2017

c - CR: needle vs granular phosphor

AGFA ADC (granular) AGFA DX (needle)

Needle screen is thicker with betterx-ray absorption and lower noise.

Page 15: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

15NERS/BIOE481 - 2017

d. Image Blur from Light Diffusion

1. X-rays are absorbed in storagephosphors (light blue).

2. A focused laser beam (red)scatters within the screenbefore stimulating light release.

3. Released light (blue) furtherscatters before leaving thesurface of the screen.

4. Emitted light is detected by alarge area photo-detectors (PD)

PD PD

X-RAYS

• Blur in computed radiography results from laser light lateraldiffusion prior to stimulating light emission.

• For the same screen thickness, needle phosphor screens haveless blur than granular phosphors. However, needle phosphorsscreens are often thicker for improved absorption.

• No further blur results from diffusion of the emitted light.

Page 16: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

16NERS/BIOE481 - 2017

d - CR Physical Performance KCARE Report #06004

Agfa Healthcare DX-S

March 2006

Modulation Transfer Function (MTF) and DetectiveQuantum Efficiency (DQE) for a scanhead/needlephosphor system in relation to conventional CR.

Page 17: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

17NERS/BIOE481 - 2017

e - CR artifacts

CR specific artifacts

• Incorrect image processing

• Segmentation error (mispositioning)

• White specks (unclean screens)

• White traces (cracked screens)

• Vertical bands (dirty light guide)

• Multiple images (non-erased screens)

• Moiré pattern (due to the grid)

• Aliasing

• Print artifacts

Page 18: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

18NERS/BIOE481 - 2017

e - CR Dust artifact Dust on a CR screen blocks laser readout.Periodic screen cleaning is required.

Page 19: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

19NERS/BIOE481 - 2017

e - CR Grid artifactA moire artifact resulting from thelaser scan direction being nearly parallelto a grid with too large a spacing.

Page 20: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

20NERS/BIOE481 - 2017

VI.C.2 – CCD systems, (8 charts)

2) Phosphor - CCD systems

a) Lens coupled design

b) Light collection efficiency.

c) Methods for improved light collection

d) Charge Coupled Devices, CCD

e) Folded optics

f) Gantry mount

g) Focus problems

h) Small FOV designs (fiber coupled)

Page 21: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

21NERS/BIOE481 - 2017

a - Lens coupled CCD camera Light from a scintillation phosphorscreen is recorded using a lens and afocal plane CCD sensor

Page 22: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

22NERS/BIOE481 - 2017

b - Lens coupled CCD camera Lens couples system suffer frompoor light collection efficiency

Liu, Karellas .., Med Phys, 21(7), 1994

Collection efficiency, Lambertian emitters.

114222

Mf

Tg

n

fn = 2 lens aperture number

M = 10 minification ratio

T = 0.9 transmission

g2 = 0.0005

Page 23: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

23NERS/BIOE481 - 2017

b - Excess noise from poor light collection

Early designs showedexcessive noise dueto poor lightcollection efficiencythat resulted in asmall number ofelectrons beingproduced for eachabsorbed x-ray.

Page 24: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

24NERS/BIOE481 - 2017

c - Lens coupled CCD cameraFour methods have been used to improvethe light collection efficiency

1. Oriented screens withforward light emission.

2. Large aperturecondensor lens.

fn < 1.0

3. Large areasensor forreduced M.

Yu & Boone, Med.Phys. 24(4), 1997

g2 for forward emission

4. Improved sensorquantumefficiency.

Page 25: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

25NERS/BIOE481 - 2017

d - Large area, high QE CCD sensors

Large area, full frame transfersensors are made by EastmanKodak Co. with a thin filmsurface that promotes shortwavelength absorption

39 mm x 50 mm Kodak 22 MP sensor

From

Hammamatsu

Page 26: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

26NERS/BIOE481 - 2017

e - Lens coupled CCD camera

More compact designs are achievedby using a mirror (folded optics)

Front surface mirror

Page 27: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

27NERS/BIOE481 - 2017

g - Gantry Mounted CCD detector

Imaging Dynamic Company Ltd (IDC)

Calgary, AB, Canada

www.imagingdynamics.com

Page 28: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

28NERS/BIOE481 - 2017

g -Blur artifacts

While current designs do notsuffer from added noise, thesystems are susceptible to blurartifacts due to the lenscoupling being out of focus.

Blurred

Page 29: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

29NERS/BIOE481 - 2017

VI.C.3 – Flat panel Digital Radiography, DR (20 charts)

3) Digital Radiography, DR

a) DR detector panels

b) DR systems integration

c) Signal preprocessing

d) Direct vs Indirect

e) Artifacts

Page 30: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

30NERS/BIOE481 - 2017

a. Detectors: DR

Flat panel digitalradiographydetectorsintegrate theabsorption ofradiation and theelectronic readoutin a single panel

Electronic circuits made ofamorphous silicon form thinfilm transisters (AM-TFT)that read charge created byx-rays. The AM-TFTtechnology is similar to thatused in common LCD displays

Human hair for size reference

Amorphous Silicon Flat Panel Detectors

Page 31: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

31NERS/BIOE481 - 2017

Indirect: Each absorbed x-ray produces a burst of light in aphosphor screen. These light photons are absorbed bythe photodetector and converted to electrical charge.

Direct: Each absorbed x-ray produces charge directly in aconductive material. This burst of charge is rapidlycollected and stored on a capacitor.

a. Detectors: Direct & Indirect

GLASS SUBSTRATE

PHOTOCONDUCTORELECTRODE

CAPACITOR TFT SWITCH

- -- - -- -

+ ++ + ++ +

PHOTODETECTOR TFT SWITCH GLASS SUBSTRATE

SCINTILLATOR

Page 32: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

32NERS/BIOE481 - 2017

• The charge collected ateach pixel in a row ismeasured and convertedto a digital number.

• All rows are sequentiallyaddressed until theentire image is read anddigitized.

a. Detectors: TFT panel readout

RO

WSE

LEC

TIO

NC

ON

TR

OL

PREAMP & ADC CHANNELS

IMAGE ACQUISITION PROCESSOR

Page 33: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

33NERS/BIOE481 - 2017

a. Detectors: DR pixels

• The row addresscontrol circuits areused to connect allcapacitors in a rowto the readout lines.

• While the row isactive, preamplifiersconvert the chargeon each pixel to avoltage forconversion using ananalogue to digitalconvertor (ADC)

Page 34: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

34NERS/BIOE481 - 2017

a. Detectors: Pixel details

At each pixel, amorphous silicon thin film circuitsform a capacitor to store the signal charge and aswitching transistor to activate readout

Page 35: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

35NERS/BIOE481 - 2017

b. DR Integrated xray generators

Integration of the x-ray generator control with theimage acquisition allows the views and technicalparameters to be programmed for all exam orders

PACS WorklistOrder Code

TechniquekVpmAS (timed)grid

ViewsAPLatTownes

Page 36: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

36NERS/BIOE481 - 2017

c - DR ‘For Processing’ Data

RAW data from the detector is pre-processedto produce an image suitable for processing.

* DICOM UID 1.2.840.10008.5.1.4.1.1.1.1.1

BADPIXELS

RAW

DARK

DR FORPROCESSING

LOG

GAIN

* DICOM SOP ClassFor ProcessingDigital X-rayImage Storage

LINEAR

Page 37: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

37NERS/BIOE481 - 2017

c – Bad pixels

• Pixels with high or low values or with excessive noise

• Values corrected by interpolation from neighbors

• There are presently no requirements to report badpixel statistics as a part of DR system purchase.

450 x 200 region

Page 38: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

38NERS/BIOE481 - 2017

c – Dark image

• The signal recorded whenno x-rays are incident onthe detector is referred toas the ‘dark image’ or‘offset image’.

• Most detectors produce asignal that linearly increasefrom the offset value ofeach pixel as x-ray incidentexposure is increased.

• Dark image values aresusceptible to drift andoften have high thermaldependence.

Display Window = 0-20

Digital Fluoroscopy dark image

Page 39: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

39NERS/BIOE481 - 2017

c - Offset/Gain correction

• Dark Image - ID

Obtained by averaging many images obtainedwith no xray input to the detector.

• Gain Image - IG

Obtained by averaging many images obtainedwith a uniform x-ray fluence.

• Uniformity correction is performed subtracting thedark offset and adjusting for gain differences.

ICOR = (IRAW – ID) {k/ (IG – ID)}• Log transformation using a Log look-up table allows

this to be performed with a subtraction.

IFP = log (IRAW – ID) - log(IG – ID) - K

Page 40: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

40NERS/BIOE481 - 2017

c – log image values

m(s)

The linear signal is approximatelyproportional to the incident x-rayintensity;

I(x,y) a Io exp{- P(x,y)}

P(x,y) = m(s)ds

The log signal is proportional tothe line integral.

Ln(I(x,y)) a -P(x,y) +Ln(Io)

Small perturbations cause thesame image value change whetherin high or low transmission regions

I1FP a P1(x,y) + DP

I2FP a P2(x,y) + DP

s

s

P1

P2

Page 41: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

41NERS/BIOE481 - 2017

c - DR ‘For Presentation’ Data

Image processing is applied to improve thevisual presentation of low contrast details.

* DICOM UID 1.2.840.10008.5.1.4.1.1.1.1.1

DR FOR PROCESSING

* DICOM SOP ClassFor Processing

DX Image Storage

DR FOR PRESENTATION

† DICOM UID 1.2.840.10008.5.1.4.1.1.1.1

† DICOM SOP ClassFor Presentation

DX Image Storage

http://dicom.nema.org/

Note: L12 will describe image processing

Page 42: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

42NERS/BIOE481 - 2017

d. Direct vs Indirect DR

thin film transistor

Oriented

CsI

reflective layer

indirect DR

dielectriclayer

electrode

a-selenium

thin film transistor

+-

direct DR

• Indirect DR: lateral light transport in scintillator screens causes blur. Forthe same screen thickness, this is less for oriented fiber screens comparedto granular screens. However, oriented fiber screen are typically thicker toachieve improved absorption with similar blur

• Direct DR: Experiments with x-ray micro beams from a synchrotron haveestablished that there is negligible blue in a direct DR detector. Charge istransported with minimal lateral movement.

Page 43: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

43NERS/BIOE481 - 2017

d. DR MTF KCARE report number 05078, October 2005.

Digital Detectors for General Radiography

DIRECT

INDIRECT

Page 44: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

44NERS/BIOE481 - 2017

d. DR DQE KCARE report number 05078, October 2005.

Digital Detectors for General Radiography

DIRECT

Shimadzu

New 1 mm Se Detector

Page 45: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

45NERS/BIOE481 - 2017

d. DR MTF Hunter, Med. Phys., 2012

Measured and calculated K-fluorescence effects on theMTF of an amorphous-selenium based CCD x-ray detector

Radiation transportreduces the MTF

High resolution Se DRsystems (25 micron pixels)are limited by radiationtransport effects

Scott

,S

PIE

MI,2

014

Am

orp

hous

sele

niu

md

irect

dete

cti

on

CM

OS

dig

ital

x-r

ay

imager

wit

h2

5m

icro

npix

elpit

ch

92 mm Se, 40 kVp W

Se K-edge = 12.66 keV

Page 46: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

DR vs CR

DR systems offer several advantages relative to CR:

• Improved image quality (i.e MTF, NPS, DQE)

• Integrated system provides better productivity.

This has led to a decrease in the utilization of CR.

46NERS/BIOE481 - 2017

NOTE: TheUS Medicaresystem willbegin reducingpayments forexamsperformed onCR systemsbeginning in2018.

Page 47: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

47NERS/BIOE481 - 2017

e. Detectors: DR artifacts

DR specific artifacts

• Pixel defects:• Circuit defects can make a single pixel or an entire

column/row unresponsive to x-rays.

• These defects are generally corrected byinterpolation.

• Excessive correction can mask small image details.

• Pixel gain variation:

• For the same x-ray exposure, each pixel mayproduce a slightly different digital signal:

• Area variations

• Amplifier gain variations

• A uniformity correction is done to prevent fixedpattern noise in the image.

• Improper correction adds noise in addition to thenormally present quantum mottle.

Page 48: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

48NERS/BIOE481 - 2017

c – New Bad pixels

• New pixel defects candevelop in DR panelsthat are in service.

• Frequent gaincalibration can helpdetect newly developedproblems.

• The defects shown tothe right were reportedby the radiologistinterpreting the study.

Indirect DR

Page 49: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

49NERS/BIOE481 - 2017

c – Gain image

• The linear gain may slightly differ from pixel to pixel.

• These variations produce fixed pattern noise.

Uniform radiation exposure

Page 50: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

50NERS/BIOE481 - 2017

VI.C.4 – Radiographic Detective Efficienty - Theory (10 charts)

4) Large area detective effiency, DQE(0)

a. energy deposition distribution

b. signal

c. noise

d. DQE(0)

e. Signal difference and relative contrast

Page 51: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

51NERS/BIOE481 - 2017

a - P(e,E)dE for a Selenium Photoconductor on glass

p(e,E)de

e , E

Page 52: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

52NERS/BIOE481 - 2017

a - P(e,E)dE for a Selenium Photoconductor on glass

p(e,E)de

e , E

TFT substrate glass

Page 53: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

53NERS/BIOE481 - 2017

IV.C.3 – Ideal image detector – energy integrating type

• An ideal energy integrating detector will record a signal equal tothe total energy of all photons incident on the detector surface.

• The detected signal for an ideal energy integrating detector, Se,can be written as:

Where

• t is the exposure time, sec

• f is the photon fluence rate, photons/mm2/sec,

• Ad is the effective area of a detector element.

max

0

)(

E

Ede dEEtAS

The majority of actual radiographic detectors areenergy integrating; however, they are not ‘ideal’.

From Lecture 05

Page 54: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

• As a first step in deriving an expression for an actualdetector, we write first the expression for monoenergeticx-rays.

• This can be used to define a signal detection efficiency.

54NERS/BIOE481 - 2017

H

iiE eS

1

E

deEeepS EE

0

,

EH

SESE

EdeEeepE

SE

S

E

EE

0

),()(

b - Signal - actual detector, energy integrating - MONO-ENERGETIC.

Note3: In a Monte Carlo analysis, the signal isdetermined by summing the energydeposited for H incident x-rays. Theefficiency is trivially determined from this.

EdE tA

Note2:

For simplicity,we now write:

Note1: SE and FE

are used to denotemonoenergetic at E

Page 55: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

55NERS/BIOE481 - 2017

c - Noise - actual detector, energy integrating - MONO-ENERGETIC.

• The noise can be similarly written as a second moment integral.

• And the noise transfer efficiency similarly defined.

Note that this efficiency is the variance transfer.

E

EdeEepeE

0

2 ,2

22

2

2

)(2

0

),( EdeEepeE

E

E

EE

Note: In a Monte Carlo analysis, thenoise transfer and relatedefficiency is determined byaccumulating the square of theenergy deposited by each x-ray.

E E

i

H

e2 2

1

EE

E H

22

2

Page 56: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

56NERS/BIOE481 - 2017

• For multiple channels which linearly accumulate signal, theenergy absorption and noise power efficiencies are additive.

• For detectors with linear signals, the signal and noise aredetermined by separate integrals incorporating the efficiency.

dEEEEkV

s

dEEEESkV

S

2

0

2 2

0

d - Signal and noise for a SPECTRUM of Xray Energies.

EEEE

EEEE

n

nssss

222221

21

Note1: F(E) is nowused to denote theenergy spectrum.

Page 57: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

57NERS/BIOE481 - 2017

Signal and Noisefor a

monoenergetic xray beam

S E

E E

E s E

S E

E

E

2 2 2

NEQ (i.e. SNR2) and DQEfor a

monoenergetic x-ray beam

SNEQ

DQE

E

S

SE

ES

E

E

E

E

E

2

2

2

2

2

2

d - SNR and DQE - MONO-ENERGETIC

DQEE was first reported by Swank in term of the moments, Mn,of the energy deposition distribution, I = M1

2/M2M0

Swank RK, Absorption and Noise in X-ray Phosphors, J. Appl. Phys., 1973

Page 58: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

58NERS/BIOE481 - 2017

S S

kV

E E E dE

ss

kVE E E dE

0

2 2

0

2

ideal

s

kV

kV

ideal

Q

SDQE

dEEE

dEEE

Q

22

0

2

2

0

)0(

d - SNR and DQE for a SPECTRUM of Xray Energies.

Signal and Noisefor a

a spectrum of xray energies

NEQ (i.e. SNR2) and DQEfor a

a spectrum of xray energies

with S and sS determinedfrom the integrals on the left.

Separate integral for signal and noise arerequired for a spectrum of energies.

This was not made clear in Swank 73.

Page 59: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

e – Signal difference and relative contrast

59NERS/BIOE481 - 2017

In lecture L05, we saw that a small object which perturbs the attenuationof the surrounding material results has a relative contrast given by thedifference between the attenuation coefficient of the small detail (i.e.the target material) and the attenuation coefficient of the surroundingmaterial (i.e. the background material).

When considering poly-energetic beams, the energy dependenttransmission and detector signal efficiency must be accounted for whendetermining Cr . In the derivation, there is a transmission term of theform exp(-µ(E)dt) that occurs within the integrations over pathlength andenergy that lead to the signal in the target region, St, and the signal in thebackground region, Sb. For a small target, the attenuation differencecomes from the difference in µ(E) over the region of the target, µt(E) -µb(E) . When [µt(E) - µb(E)]dt is small, these equation yield;

� � = � � − � � � �

� � =∫ � � � � � � � � � − � � � � � � �

∫ � � � � � � � �

From Lecture 05

Page 60: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

e – Signal difference and relative contrast

60NERS/BIOE481 - 2017

If we define the effective attenuation as a weighted average of m(E) ,

then Cr can be expressed as,

The effective attenuation coefficient, meff, is defined with respect to:

• An x-ray spectrum incident on a target, f(E).

• A detector with energy dependent absorption, hs(E).

� � = � �� � � − � �

� � � � �

� � � � =∫ � � � � � � � � � �

∫ � � � � � � � �

Note: Some textbooks may define an effective attenuation coefficientthat applies to other problems. For example, the transmittedexposure may be considered rather than a detector signal.

Page 61: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

61NERS/BIOE481 - 2017

VI.C.5 – Rapid Sequence Acquisitions & fluoroscopy (16 charts)

5) Rapid Sequence Acquisitions & fluoroscopy

a. Traditional fluoroscopy with Image Intensifiers.

b. Pulsed Digital fluoroscopy

c. Digital Angiography

Page 62: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

62NERS/BIOE481 - 2017

VI.C.5.a – Historical fluoroscopy

• As was noted in L01, real timeviewing of x-ray images(fluoroscopy) dates backalmost to the discovery of thex-ray.

• Excessive radiation dose wasencountered in Edison’s earlywork on fluorscopy.

• To reduce dose, fluoroscopywas done in dark rooms afterthe radiography had becomedark adapted (10-15 minutes).

• This continued until about1950 when image amplificationdevices became available.

1933 photograph takenat the Mayo Clinic

Page 63: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

63NERS/BIOE481 - 2017

VI.C.5.a – X-ray Image Amplification

• X-ray Image Intensifies use CsI scintillations to detect x-ray and aphotocathode to convert the light into electrons.

• Electrons are accelerated towards an output phosphor using high voltage,~30 kV. The electron path is controlled by focusing electrons such that theimage on the output phosphor is a mirror image of the input.

• Amplification is achieve by both minification and electron kinetic energy.

• The output phosphor image is then recorded with a TV camera or cine film.

www.alt-vision.com

Page 64: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

64NERS/BIOE481 - 2017

VI.C.5.a – X-ray Image Amplification

Front and back view of the glass vacuumtube of a Philips x-ray image intensifier.

Page 65: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

65NERS/BIOE481 - 2017

VI.C.5.a – GI Fluoscopy

Gastro-Intestinal (GI)fluorscopy exams havebeen done since the midfifties. The exam istraditionally done with atilt table. Barium loadedcontrast is administeredorally or rectally.

Fluoroscopy spot film from apatient with Crohn’s disease.

II

Page 66: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

66NERS/BIOE481 - 2017

VI.C.5.a – Rapid Sequence Angiography

Articulated C-armdevices provide flexiblepositioning forangiographic studies.

Pulsed image acquisitions at 2 to4 frames per second depict the

flow of iodinated contrastinjected in veins or arteries.

Page 67: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

67NERS/BIOE481 - 2017

VI.C.5.b – Pulsed Digital Fluoroscopy

Modern system now use solid state x-ray image detectors for pulsedfluoroscopic and rapid sequence acquisitions. The detector assembly ismore compact that the image intensifier systems previously used.

From Cowen 2008

Page 68: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

68NERS/BIOE481 - 2017

VI.C.5.b – Digital Fluoroscopy

CsI scintillators with TFTphotodiode arrays aresupplemented with flash lampassemblies to rapidly erasethe residual charge thatwould othewise create lag inthe image response.

From Cowen 2008

Digital PanelImage Intensifier

The photodiodearray geometryproduces imageswith no distortion.

The electronoptics of I.I.systems results indistortioninfluenced bymagnetic fields.

Page 69: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

69NERS/BIOE481 - 2017

VI.C.5.b – Pulsed Digital Fluoroscopy

Dynamic Solid StateDetectors, DSD, can havereduced DQE at lowexposures due tocapacitance noise in thephotodiode array readout.

Image Intensifier TV systems,IITV, can have reduced DQE athigh exposure due to instrumentnoise in the readout cameras..

From Cowen 2008

Page 70: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

70NERS/BIOE481 - 2017

VI.C.5.b – Pulsed Digital Fluoroscopy

Fluoroscopy with continuous radiation results in blurred images with blur.Fluoroscopy with short pulses of radiation reduces blur for the individual framesof the exam. Recorded cine images can be stopped with and show good detail.

Continuous Fluoro Pulsed Fluoro

Images From P. Rauch

Page 71: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

71NERS/BIOE481 - 2017

VI.C.5.b – Pulsed Digital Fluoroscopy

Continuous Fluoro Pulsed Fluoro

Images From P. Rauch

These cine clips illustrate the difference between continuous andpulsed fluoroscopy using a toy car traveling on a circular track.

Page 72: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

72NERS/BIOE481 - 2017

VI.C.5.b – Pulsed Digital Fluoroscopy

Images From P. Rauch

A radiograph of a clockmechanism details the gears

with low quantum noise.

Page 73: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

73NERS/BIOE481 - 2017

VI.C.5.b – Pulsed Digital Fluoroscopy

Images From P. Rauch

Individual fluoroscopic frames have higher noise due to the low exposure perframe. These examples have the same exposure/frame but different frame rates.

When shown as a movie clip, the higher frame rate has less noisebecause of temporal averaging by the human visual system.

Wilson DR,Perception of Temporally Filtered X-rayFluoroscopy Images, Med. Phys. v21, pp245-256, 1994

Page 74: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

74NERS/BIOE481 - 2017

VI.C.5.c –Digital Angiography

• Vascular disorders such as aneurisms and occlusions can bediagnosed using a rapid sequence of images obtained during theinjection of iodinated contrast material.

• The sequence at the left demonstrates the flow of the contrastmaterial through the right and left femoral arteries.

• On the right, the first frame has been subtracted from all frames

Page 75: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

75NERS/BIOE481 - 2017

VI.C.5.c –Digital Angiography

CerebralAngiogram

Arterial

Phase

Page 76: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

76NERS/BIOE481 - 2017

VI.C.5.c –Digital Angiography

CerebralAngiogram

Perfusion

Phase

Page 77: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

77NERS/BIOE481 - 2017

VI.C.5.c –Digital Angiography

CerebralAngiogram

Venous

Phase

Page 78: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

78NERS/BIOE481 - 2017

VI.C.6 – Photon Counting Detectors (6 charts)

6) Photon Counting Detectors

a. CdTe detectors

b. Si strip detector

c. Energy weighted signals

Page 79: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.a CdTe detectors

Photon-counting, direct conversion, x-ray sensor.

• CdTe X-ray sensor, 0.65 mm thick

• 30 mm x 25 mm per module (1 to 8 modules).

• 60 micron hexagonal pitch

• Energy thresholds for each pixel

• CMOS ASIC

79NERS/BIOE481 - 2017

http://www.pixirad.com/

The PiXirad technology wasdeveloped at INFN, Italy (i.e.the Italian National Institutefor Nuclear Physics).

Page 80: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.a CdTe detectors

• The CdTe sensor (Acrorad Co. Ltd, Japan)is a Schottky diode with electron collection.

• Very low leakage current @400-500V bias.

• Hexagonal array charge readous

80NERS/BIOE481 - 2017

The reported material properties areconsistent with those in Lecture 08.

Page 81: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.a CdTe detectors

• Charge sharing causes theMTF to be significantlyreduced from the aperturefunction MTF associated withideal charge collection.

81NERS/BIOE481 - 2017

Vedantham, Med. Phys., May 2016.

• Row and column pixels areinterpolated from the threehexagonal signals surround eachpixel in a square array.

• The interpolation has minimaleffect on the MTF.

Page 82: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.a CdTe detectors

• The energy resolutionmeasure from a cluster of1000 pixels demonstrated adependence on eneryconsistent with the statisticalcharge variation.

• The equivalent noise charge ofthe ASIC is 50 electrons.

82NERS/BIOE481 - 2017Vincenzi, J. Inst., April 2015.

Over half of the counts are from chargesharing (6-15 keV) which can producemultiple counts in different pixels from oneincident x-ray. This results from eitherradiation transport or charge diffusion.

Page 83: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.a CdTe detectors

Dual Energy photon-counting, direct conversion,x-ray sensor for digital x-ray imaging.

• CdTe – CMOS detector (12 bits)

• 0.75 mm – 2.00 mm CdTe thickness

• 100 micron detector elements

• 1536 x 128 detector array

• Up to 1000 fps

83NERS/BIOE481 - 2017

Scintillator

http://www.xcounter.se/

Page 84: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.a CdTe detectors

The Xcounter CdTe detector processes thesignal from each detector element toimprove image quality. Anti Charge Sharing(ACS) compares the charge from each pixelto that of it’s neighbors to prevent multiplecounts due to fluorescent radiation emittedfrom Cd or Te atoms (see above).

An energy threshold is used toaccept only charge depositions above thedesignated threshold.

Both ACS and energy thresholdprocessing prevents correlated noise andproduces a flat noise power spectrum (NPS).

84NERS/BIOE481 - 2017

Ji X, SPIE Medical Imaging,vol. 10132, Mar. 2017.

Anti Charge Sharing (ACS)

Page 85: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.b Si strip detector

Philips MicroDose Mammography System

• Si Strip Detector

• 50 micron pixels

• Tungsten anode, 0.5mm Al filter

• Energy weighted signal

• High energy depositions

• Low energy depositions

85NERS/BIOE481 - 2017 Note: Philips purchased MicroDose from Sectra in 2011

Page 86: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.b Si strip detector

Aslund et. al. , Medical Physics, 2007

Physical characterization of a scanningphoton counting digital mammographysystem based on Si-strip detectors

86NERS/BIOE481 - 2017

The noise power spectrum (NPS) for the Si stripdetector is reported by normalization to theideal NPS (see Lecture 7).

N(w) = NPS(w) x Qi

Fig. 4. Average of 12 measures over an input exposure range from 21to 2006 microGray.

Page 87: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.b Si strip detector

87NERS/BIOE481 - 2017

Fredenberg et. al. , NIM A, 2010

Energy resolution of a photon-counting siliconstrip detector

Fig. 7. Energy response on monochromatic delta peaks.The plottedpeaks are for the experimental detector. Energy resolution (dE/E) isshown for the experimental detector, and for an improved detectorwith high AC efficiency and low threshold spread and electronic noise.

Page 88: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.b Si strip detector

The simulated CNR improvement isshown as a function of weight. The CNRimprovement and optimal weight werelarger for higher tube voltages. A weightof ~1.8 is near optimal for all cases.

50 micron calcification in breast tissue.18 keV threshold @ 26 kV20 keV threshold @ 32 kV22 keV threshold @38 kV

88NERS/BIOE481 - 2017

Fredenberg 2010, Fig 1

Berglund et. al. , SPIE MI, 2014

Energy weighting improves the imagequality of spectral mammograms:Implementation on a photon-countingmammography system.

_________________________________

“We have implemented and evaluatedso-called energy weighting on acommercially available spectral photon-counting mammography system. Apractical formula for calculating theoptimal weight from pixel values wasderived. Computer simulations andphantom measurements revealed thatthe contrast-tonoise ratio was improvedby 3%–5%, and automatic imageanalysis showed that the improvementwas detectable in a set of screeningmammograms.”

Page 89: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

A generalized expression for the detector signalis a weighted sum of the discrete fluence spectrum.

• wi = 0 is the case of a photon counting detector.

• wi = E is the case of an energy integrating detector

Using an approach similar to Tapiovaara1985, Cahn shows that a moreoptimal weight will optimize the contrast to noise, CNR, for a smalltarget (t) in a uniform background (b).

For high Z targets, the weights

will be proportional to E-3 dueto photoelectric absorbtion.

VI.C.6.c Energy Weighted Signals

Cahn et. al. , Medical Physics, 1999

Detective quantum efficiency dependence on x-ray energy weighting in mammography.

“..transmitted low energy photons carry more contrast information than transmitted photons ofhigher energy. Energy-integrating detectors will put a weight factor to each photon proportional toits energy and the weighting will thus be contrary to the information content.”

89NERS/BIOE481 - 2017

� = � � � ∅� �

� � �

� � ∝ 1 − �� � � �

� � � � �� � �

� � = � � − � � � �From Lecture 05:

� � ∝ � � �� − � � �

� � � , � � � � � � �

Page 90: NERS/BIOE 481 Lecture 10 X-ray Imaging Detectors · Lecture 10 X-ray Imaging Detectors Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science ... radiography detectors integrate

VI.C.6.c Energy Weighted Signals

Cahn et. al. , Medical Physics, 1999

Detective quantum efficiency dependence on x-ray energy weighting in mammography.

90NERS/BIOE481 - 2017