near real time safety air solutions products€¦ · 40/60 carbograph ‐1 is a product of camsco...
TRANSCRIPT
Products
Near Real Time Safety Air and Process Monitoring Solutions
• Optics Kit• Selective Sampler™• Stream Selector with Continuous Vacuum Option
• Heavy Vapor Sample Lines• Solutions for difficult analytical requirements
Optics Kit Overview
Near Real Time Safety Air and Process Monitoring Solutions
• New coupling between the detector block and MINICAMS® PMT
• A series of focusing lenses mounted inside a specially machined PMT coupling
• A concave mirror mounted on the back of the detector block
Optics Kit Overview
Near Real Time Safety Air and Process Monitoring Solutions
Optics Kit Operation
Near Real Time Safety Air and Process Monitoring Solutions
• To the analyst, operations of the modified MINICAMS® are completely transparent
• Determine proper PMT voltage set point for analyte concentration, typically lower
• The increased sensitivity will allow for decreased cycle times
Standard MINICAMS®
Near Real Time Safety Air and Process Monitoring Solutions
CAMSCO 60/80 Haysep‐D – 30pg on column, PMT @ 900
Last Year before Improvements
Near Real Time Safety Air and Process Monitoring Solutions
CAMSCO 60/80 Haysep‐D – 30pg on column, PMT @ 900
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit
CAMSCO 60/80 Haysep‐D – 30pg on column, PMT @ 900
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit – The Sorbent Matters
Empty PCT – Only N2 flowing through column. Column heater on at 40 seconds and off at 160 seconds. PMT @ 900
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit – The Sorbent Matters
CAMSCO 60/80 Haysep‐D – After 30 minutes of conditioning, Charcoal filter installed on inlet. PMT @ 900
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit – The Sorbent Matters
CAMSCO 40/60 Carbograph‐1 – After 30 minutes of conditioning, Charcoal filter installed on inlet. PMT @ 900
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit – The Sorbent Matters
CAMSCO 40/60 Carbograph‐1 – 30pg on column, PMT @ 900
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit – The Sorbent Matters
CAMSCO 40/60 Carbograph‐1 – This is approx. 1pg on column. Tube cleanliness and trapping efficiency make this possible. PMT@1000
Near Real Time Safety Air and Process Monitoring Solutions
Current Optics Kit – The Sorbent Matters
CAMSCO 40/60 Carbograph‐1 – This is approx. 500 femtograms on column. PMT@1000
Optics Kit Performance
Near Real Time Safety Air and Process Monitoring Solutions
Coupling Signal Noise S/NImprovement
Factor
Standard (no lenses) 113 4 28 0
Before Improvements 888 4 222 7.92
Current Design 3246 4 811.5 28.98
NRT Methodologies now offers the Optics Kit for bench top GC’s.
Optics Kit Safety Performance
Near Real Time Safety Air and Process Monitoring Solutions
Airborne Exposure Limits in mg/m3
AGENT STEL WPL GPLGA 0.0001 0.00003 0.000001GB 0.0001 0.00003 0.000001GD 0.00005 0.00003 0.000001GF 0.00005 0.00003 0.000001VX 0.00001 0.000001 0.0000006
Optics Kit Safety Performance
Near Real Time Safety Air and Process Monitoring Solutions
• This improved sensitivity will allow for much faster cycle times• VX STEL in ≤ 5 minutes at lower PMT voltage• G’s WPL in ≤ 5 minutes at lower PMT voltage• VX WPL in ≤ 8 minutes• G’s GPL in ≤ 8 minutes• VX GPL in ≤ 10 minutes
Optics Kit Testing Performance
Near Real Time Safety Air and Process Monitoring Solutions
• Improved sensitivity and data points for the following:• Vapor generation• Permeation• Collective Protection• Smear and clear off gassing
Optics Kit Testing Performance
Near Real Time Safety Air and Process Monitoring Solutions
• Faster cycle times for extremely low concentrations provides the following: • A more precise time as to when breakthrough
occurred• More data for off gassing tests• More data to determine trends or fixture
problems
Optics Kit 140 Second Cycle Time
Near Real Time Safety Air and Process Monitoring Solutions
• NRT Methodologies developed a 140 second method for a customer able to quantitate TEP as low as 0.00005 mg/m3 from as far away as 100’• Innovative sample lines were developed by
Quetron Systems and NRT • The customer has made this method the new
standard by which data is collected for field testing
140 Second CycleApprox. 25pg Recovered Through 100’ Line During a Field Trial
Near Real Time Safety Air and Process Monitoring Solutions
Optics Kit Testing Performance
Near Real Time Safety Air and Process Monitoring Solutions
• The amount of times a bubbler or sorbent tube can be extracted is far exceeded by the amount of quantifiable analyses a MINICAMS® can produce in the same sample time
The Selective Sampler™
Near Real Time Safety Air and Process Monitoring Solutions
Selective Sampler™ Overview
Near Real Time Safety Air and Process Monitoring Solutions
• It is an external accessory that increases analytical selectivity by adding a differing stationary phase column before the MINICAMS® column
• No modifications need to be made to the MINICAMS® for operation
• Does not affect cycle time• Can be used for GC and non‐GC based analytical systems
Selective Sampler™ Overview
Near Real Time Safety Air and Process Monitoring Solutions
• The instrument is fitted with any 0.32 mm capillary column at the time of manufacture.
• The column can be changed anytime based on analytical requirements
• The instrument footprint is 7”H X 9”D X 11”W and weighs 10 lbs.
Selective Sampler™ Operation
Near Real Time Safety Air and Process Monitoring Solutions
• It uses the same nitrogen, vacuum, and air (optional) as the MINICAMS®
• It uses the same PCT as the host system• It uses a custom electrical cable to synchronize with the MINICAMS® purge and sample times
• It receives flow rate signal from the MFC connected to the MINICAMS®
• Selective Sampler flow rate determines the mass delivered to the MINICAMS® column
Selective Sampler™ Operation
Near Real Time Safety Air and Process Monitoring Solutions
• The instrument features 20 X 4 backlit LCD with an integrated 6‐button key pad
• Agent gate timing is adjusted here
Selective Sampler™ Operation
Near Real Time Safety Air and Process Monitoring Solutions
• Adjusting the agent gate• Allow the Selective Sampler to synchronize with the MINICAMS®
• When the MINICAMS® is sampling the Selective Sampler is purging and vice versa
• Using the user interface, adjust the agent gate so that only the analyte of interest elutes from the column
• This will eliminate or dramatically reduce the amount of interferences during analysis
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® VX/GB challenge –prior to calibration—resulted in concentration readings of 1.27 STEL for VX and 1.29 STEL for GB.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® sample of ALO boot in a 5‐gal container resulted in concentration readings of 62.5 STEL for VX a severe interference.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® sample of ALO boots in 5‐gal container through the Selective Sampler resulted in concentration readings of 0.11 STEL for VX.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® sample of ALO boots in 5‐gal container through Selective Sampler and a simultaneous VX/GB challenge into Selective Sampler.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® analysis of burning TEP, however the column is saturated with other compounds from combustion causing flame out.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® analysis of burning TEP, the Selective Sampler has only allowed TEP to be sampled by the MINICAMS.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® analysis of burning MeS, interferences delivered to the FID caused a false positive.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
MINICAMS® analysis of burning MeS, the Selective Sampler has made an FID very selective.
Selective Sampler™ Performance
Near Real Time Safety Air and Process Monitoring Solutions
• This improved selectivity greatly reduces or eliminates interferences for the following:• Safety air monitoring• Headspace monitoring• Remediation programs• Test programs that require interferences and
agent exposure to test articles at the same time
Stream Selector with Continuous Vacuum Option
Near Real Time Safety Air and Process Monitoring Solutions
Stream Selector with Continuous Vacuum Option
Near Real Time Safety Air and Process Monitoring Solutions
• Each 3‐way valve can have a steady vacuum• This ensures the sample lines do not stagnate or
oversaturate• A more representative sample is delivered to the
MINICAMS®• All permeation cups are under the same
conditions all the time
Stream Selector with Continuous Vacuum Option
Near Real Time Safety Air and Process Monitoring Solutions
• This is a much more reliable alternative• This design uses no rotating parts to select the
stream• Two units have over 10 years operation at the
Center for Domestic Preparedness at the Anniston Chemical Depot without repairs
• If repairs are required, they are simple
Quetron Systems Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
Quetron Systems Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
Quetron Systems Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
• Electrostatic dissipation of entire sample line body• Galvanically grounded gas interface ports• Hermetically sealed body and electrical components• Hermetically sealed gas interface ports (both ends)• Only sample line authorized by CMA for use in
munition storage igloos (explosive gas hazard)• No other sample line manufacturer utilizes distal end
pre‐heater
Quetron Systems Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
• The Quetron Heavy Vapor Sample Line is an integral part of the 140 second method• The distal end pre‐heater easily vaporizes more
persistent compounds like triethylphosphate (TEP) and GF
• The extreme inertness and kinetic reactivity of the transport tube provides excellent compound transfer
• Distal end challenges at 100’ passed after only 70 second sample cycle
140 Second CycleApprox. 25pg Recovered Through 100’ Line During a Field Trial
Near Real Time Safety Air and Process Monitoring Solutions
Quetron Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
• This line provides a more efficient derivitization for Lewisite at the distal end pre‐heater• No requirement for a reactor probe or reactor
tube• EDT is delivered to the pre‐heater were the
derivitization occurs• No carry‐over between distal end injections• Less maintenance
Quetron Systems Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
QC injection of 1.5ng Lewisite after 3 calibration injections through 100’ line.
Quetron Systems Heavy Vapor Sample Line
Near Real Time Safety Air and Process Monitoring Solutions
Clear cycle after four consecutive 1.5ng injections from the end of the line.
Near Real Time Safety Air and Process Monitoring Solutions
AcknowledgementsMINICAMS® is a registered trademark of CMS Field Products
The Selective Sampler™ is trademarked by Cahaba Micro, LLCStream Selector with Continuous Vacuum Option is a product of Cahaba Micro, LLC
Heavy Vapor Sample Line is a product of Quetron Systems
40/60 Carbograph ‐1 is a product of CAMSCO
NRT Methodologies, Inc. is the distributor for the Selective Sampler™, the Stream Selector with Continuous Vacuum Option, and the Heavy Vapor Sample Line.