ndt in calibration system

22
Di sampaikan pada Seminar GMF, 24 November 2009 Calibration in NDT Equipment

Upload: riyan-ngudiarjo

Post on 15-Sep-2015

25 views

Category:

Documents


1 download

DESCRIPTION

NDT In Calibration System

TRANSCRIPT

  • Calibration in NDT EquipmentDi sampaikan pada Seminar GMF, 24 November 2009

  • Six Most Common NDT Methods Visual Eddy Current Ultrasonic Magnetic Dye Penetrant Radiography

  • Visual InspectionMost basic and common inspection method.

    Tools include fiberscopes, borescopes, needs calibration on the tips for measurement accuracy

  • Visual CalibrationVerification Block 4 Measurement Targets on each Block:Forward StereoSide StereoForward ShadowSide Shadow

    Each target allows a distance measurement of 0.1 Inch or 1 MM. Traceable to NIST Standards

  • Eddy Current Testingmagnetic field Crack detection Coating thickness measurements Conductivity measurements

  • EC Calibration

  • Ultrasonic Inspection (Pulse-Echo) High frequency sound waves are introduced into a material and they are reflected back from surfaces or flaws.

    Reflected sound energy is displayed versus time, and inspector can visualize a cross section of the specimen showing the depth of features that reflect sound and need calibration for horizontal linearity.

  • Ultrasonic Inspection with Normal Probe

  • Ultrasonic Inspection with Angle Probe

  • UT Calibration Display Quality of MeasurementHorizontal LinearityScreen Height LinearityAmplitude Control Linearity

    Traceable to ASTM E-317 and ASME V Standards

  • Magnetic Particle Inspection

  • MT Calibration Performance Quality of MeasurementLifting 10 40 LbsAmmeter readingTime exposure

    Traceable to ASME V Standards

  • Dye Penetrant InspectionStep 1. Pre-CleaningEnsure surface is very Clean normally with the use of a solvent

  • Dye Penetrant InspectionStep 2. Apply penetrantAfter the application of the penetrant the penetrant is normally left on the components surface for approximately 15 minutes (dwell time). The penetrant enters any defects that may be present by capillary action

  • Dye Penetrant InspectionStep 3. Clean off penetrantAfter sufficient penetration time (dwell time) has be given the penetrant is removed, care must be taken not to wash any penetrant out off any defects present

  • Dye Penetrant InspectionStep 3. Apply developerAfter the penetrant has be cleaned sufficiently a thin even layer of developer is applied. The developer acts as a contrast against the penetrant and allows for reverse capillary action to take place

  • Dye Penetrant Inspection

  • PT CalibrationSupport Performance Quality of CalibrationBlack LightUV meterDwell timer

    Traceable to ASME V art. 7 Standards

  • Radiography InspectionX - RaysElectrically generatedGamma RaysGenerated by the decay of unstable atoms

  • Radiography Inspection

  • RT CalibrationSupport Performance Quality of CalibrationFilm DensitometerLux meterSurvey meter

    Traceable to ASTM E-1079 Standards (densitometer)