name of department : causuality upgradation...

50
1 NAME OF DEPARTMENT : CAUSUALITY UPGRADATION COMPLIANCE STATMENT SHOULD ACCOMPANY EACH INSTRUMENT / EQUIPMENT SPECIFICATIONS S.No. Item Code No. Name of Equipment Specification 1. CU. 001 Shadowless Lights i) Ceiling track model Size : 45 cms ii) Double dome shadow less ceiling OT : Shadowless operating lamp LED double dome Ceilling LED operating light double lamphead should have more than 110 LED in major lamp head and more than 70 in satellite lamphead should offer excellent luminous efficiency light source and incredibly long life LED. LED operating light should have overlapping fields of light and multiple patterns. Minimize shadow and redirect the light to the operating area. Lamphead should be of modular design with five and three segment for optimized laminar flow compatability. Lamphead should be plastic moulded and should be slick for easy manuoverbility. Feather touch control should be on the lamphed to offer different level of lightinensity. Feather touch control should be on the lamphed to offer different level of temperature. Sterilize handle should be provided in the centre

Upload: others

Post on 20-Oct-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

  • 1

    NAME OF DEPARTMENT : CAUSUALITY UPGRADATION

    COMPLIANCE STATMENT SHOULD ACCOMPANY EACH INSTRUMENT / EQUIPMENT SPECIFICATIONS

    S.No. Item Code No.

    Name of Equipment Specification

    1. CU. 001 Shadowless Lights

    i) Ceiling track model Size : 45 cms

    ii) Double dome shadow less ceiling OT : • Shadowless operating lamp LED double dome • Ceilling LED operating light double lamphead

    should have more than 110 LED in major lamp head and more than 70 in satellite lamphead should offer excellent luminous efficiency light source and incredibly long life LED.

    • LED operating light should have overlapping fields of light and multiple patterns.

    • Minimize shadow and redirect the light to the operating area.

    • Lamphead should be of modular design with five and three segment for optimized laminar flow compatability.

    • Lamphead should be plastic moulded and should be slick for easy manuoverbility.

    • Feather touch control should be on the lamphed to offer different level of lightinensity.

    • Feather touch control should be on the lamphed to offer different level of temperature.

    • Sterilize handle should be provided in the centre

  • 2

    of lamphead for positioning and focusing. • Major lamphed intensity 150000 lux and minor

    90000 lux. • Life of LED more than 35000 hrs.

    Features : • Aerodynamic lamp head hammered out by CAD. • Advanced filter technology • Special diffuser. • Single halogen bulb 24V 150W with one back up

    bulb • Feather touch digital control for various level of

    light intensity. • Unique sterilizable handle for focusing • Low voltage supply unit.

    Technical specification :

    Lamphead 700 mm dia 1 bulb

    Light intensity 130000 Lux

    Power requirement 150 W

    Color temperature 4250 K

    Average life for halogen bulb 1000 working hrs

    Satellite Light head :

    Light intensity 500 mm dia 1 bulb

    Power requirement 80000 Lux

    Color temperature 150 W

    Average life for halogen bulb 2000 working hrs

  • 3

    2. CU. 002 Computerized Digital Radiography Unit

    Latest CR system compatible with standard X-ray machine consist of digital patient identification system preview and processing server workstation. Multi loading CR system should have following features.

    1. Image recording system (Cassettes & imaging plates) The following sizes of imaging plates cassettes should be supp along with unit. a) 14" x 17" - 8 nos b) 10" x 12" - 4 nos c) 8" x 10" - 4 nos

    2. Image reading (CR Reader) /Digital

    i) Should be able to process not less than 85 imaging (14"x17") hour.

    ii) Should have resolution of 6 piexels / mm for standard reading.

    iii) It should have an input and output cassette buffer / stacker multiplier cassettes.

    iv) Preset anatomical processing should be available.

    v) Acquisition and image transfer must be done at a resolution 12 bits / pixel.

    vi) It should have status display preferably on a LCD display indicating error status etc.

    3. Processing Server and workstation with 17" LCD monitor centralized patient study management it should have following features.

    i) Work station must be able to receive the CR images for digitzer.

  • 4

    ii) It should be DICOM ready form sent, receive and print facility protocol.

    iii) It should be possible to multi formatting the images on film printing.

    iv) 17" LCD monitor for patient study management.

    v) It should possible to write images on a CD.

    vi) Should be ready to accept images from work list and patient data and images another RIS / HIS (DICOM PROTOCOL)

    vii) Should have black border facility.

    viii) Should have software for printing and to create various layout and multiple formats on a single film for optimum utilization of film and presentability. The capability to customize user defined no. of formats and layout on the single film. There should be capability of printing the zoomed image along with overview of the main image on the same film. There should be capability of printing multiple patient images on one film and multiple image of same patient on one film.

    ix) Other features should include image gray scale reversal imaging flipping and rotating, image zooming, edge enhancement, latitude reduction, image noise reduction, gray scale saturation feedback.

    x) There should be provision for adding markers.

    4. Patient identification / preview

    i) Separate or inbuilt patient identification

  • 5

    system.

    ii) Cassettee identification and demographics should be standard.

    iii) It should be possible to identify the multiple x-ray unit separately.

    5. Advanced QC viewer with 19" LCD monitor: For reviewing of the images and should be a separate module for being able to be kept away from the Digitizer / CR reader in the room of choice it should have following features :

    i) Window levelling, annotation, printing, rotating, flipping panning, zooming all image post processing.

    ii) Should have hospital information system (HIS connectivity)

    iii) There should be provision for adding markers on the film.

    iv) It should be possible to write images on a CD.

    6. Dry imager (For film printing) The system should be supplied with dry imager (dry chemistry with a spatial resolution of 500 PPI/DPI or more. It should have contrast resolution of 12 bits / pixel or more. It should have online film sizes out of 8"x10:, 10"x12". The image should preferably come with standard filom sorter and output for sorting the films based on modality connected. It should have a normal through put of 100 films per hour for the largest size. Access time for first film should be 90 seconds or less. The imager should be DICOM compatible for receive, sent and print facility. The system should at least three sizes from the five sizes

  • 6

    to be loaded at any time. Printed status should be display for any error status etc.

    7. Interconnectivity : Interconnectivity between various CR modules should be ethernet based.

    8. Scalability : The CR system should have scope of adding advanced quality software for image processor, workstations connectivity to an DICOM archieve or image management system (PACS).

    3. CU.003 Boyle Apparatus I) Circle Absorber ii) Cylinders iii) Breathing systems iv) Bains circuit v) Vaporizer vi) B.P. apparatus

    - Stainless steel frame : - 2 Flow meters each with extend range. i) For oxygen - 0.1 to 2 and 2.5 to 15 liters / mm ii) For nitrous oxide - 0.05 to 1 and 1.25 to 10 liters / mm. - Oxygen and nitrous oxide. - Magills - Jackson Rees - Fluotex (Halothane) - EMO ether vaporizer 1. Anesthesia machine should a gas systems. Air / oxygen nitrous oxide with dual cascade flow meter for delivering low flow. 2. Anesthesia machine should have mechanical hypoxie guard. 3. Basal flow of 50 ml should be there. 4. System must be incorporated with each of 02 and nitrous oxide pin index and pipeline input connections for O2 N2O air. 5. Anaesthesia machine must be having integrated micro process controlled anaesthesia ventilator, circle system and passive scavenging system.

  • 7

    6. Anaesthesia ventillator should have EL display for displaying of inspired oxygen percentage tidal volume minute volume. Airway pressure and graphical display 7. Below must be ascending type 8. Electronic beep should be standard. 9. Both volume controlled mode with tidal volume compensation and pressure controlled mode should be available in the system. 10. Compression losses within below and circle absorber should be compresated automatically in the system. 11. System must have facility to context bain circuit and magill circuit bain should have facility to mechanical ventilator. 12. Machine parameter setting for the patients like tidal volume. Rate 1:E ratio airway pressure limit should be accessive directly through soft keys. 13. Circle absorber should have mechanicals. Airway pressure limiting valve and bags to vent switch. 14. Battery backup of 30 minutes. 15. Alarm for aponoe and patient circuit low airway pressure. 16. Machine to be fixed with temperature pressure and flow compensated halothane and isoflurane having key filling arrangements.

    4. CU.004 Defibrillator monitor Amotating Recorder : Documents time, date and energy selection. Heart rate : Monitoring continuous display. Three leads select : Paddle monitoring for rapid patient ECG assessment. Individual Batteries for greater reliability. Synchronizer : Reverts to a synchronizer mode. Mine energy setting : Five to 360 joules.

  • 8

    1. Capable of delivering 1 to 360 Joules of charge. Energy from 1b to 10 joules should have increment of 1 joule each. 2. Charging trime less than 10 seconds. 3. Synchronous and asynchronous modes for shock delivery. 4. Shock advisory adaptor for automatically shockable rhythm. 5. To store events of the defibrillator and painting of pre and post shock events. 6. Operable on mains and rechargeable battery with at least 90 minutes working back up for 75 shocks of 360 joules. 7. Standard accessories. 8. In addition to the standard accessories two extra sets of patient cable for each machine. 9. Trolley of standard company and make for each machine. 10. Other terms and conditions. Guarantee for two years At least six inspection and service visits during the guarantee period in addition to the service on calls subsequent to the breakdown. Should quote for comprehensive annual maintenance contract for five years after the expiry of the guarantee period. Quarterly service visit should be made during teh entire contact period a part from service on call during the breakdown. Breakdown to repair interval should not exceed five working days, failing which a fine of Rs. 500 per day shall be imposed. Supplied should furnish an undertaking to this effect. Optional : Upgradable to NIBP, SPO2 Respiratory rate, temperature and external pacemaker. SPECIFICATION FOR DEFIBRILLATOR : 1. Defib design should be with biphasic technology.

  • 9

    2. Energy selection should be adjustable between 2-200 joules. 3. Charge and discharge button should be available both on monitor and paddles. 4. Charging time for maximum energy level 200 J should be less than 5 sec. 5. Charge indicator : audio and visual. 6. Internal discharge when unit is turned off or automatically after time limit. 7. Built in test charging facility shall be available at any energy level 2-200 J against 50 ohms impedance. 8. Built in automatic external defibrilator with voice prompt. 9. Manual and AED operation. 10. Display of selected energy. 11. Built in thermal recorder. 12. It should offer synchronized cardio version. 13. Summary storage facility (INTERNAL MEMORY) 14. Sync and async modes shall be indicated on monitor and recorder. 15. Deibrillator should have option for pacing. Monitor :

    • Sweep speed : 25 mm/sec • Heart rate indicator 15-300 bpm • Alarm setting : upper limit 100-250 lower limit 30-100

    bpm • Hi and Low adjustable alarm setting • Adjustable ECG size 5 levels. • Monitor should have bright LCD display. • Display resolution (320*240 pixels) • Audio visual indication should be available for R wave

  • 10

    detection. External Pacing :

    • Pacing rate 30-180 pulse / min +/- 1.5% • Output current 10-200 mA (+/-5mA) monophasic • Pacing mode : Demand or fixed rate • Status indicator : ECG lead fault pace lead fault.

    Defibrillator stock should be delivered thru paddles as well as thru multifunction disposable electrode pads. General : AC input - 100 to 230 V ACL: +/- 15%, 50/60 Hz Power consumption - 100 VA Battery type - Rechargeable SMF lead acid Battery capacity - 12 V, 4.5 AH 7 hrs only monitoring 100 discharges of 360 joules. Specifications for multi parameter monitor Monitor should have high resolution active matrix SVGA TFT display. Display should be integrated and not less than 8.4" in size. Display resolution should be 800 x 600. Monitor should have adult pediatric and neonatal modes. Monitor should have options for integrated dual channel thermal recorder. 1) ECG : - 3 leads facility - HR ranges 15-300 bpm - HR accuracy +/- 1 bpm - Pacemaker detection and lead off failure alarms. - ECG should meet AAMIU standards and should comply with IEC. 2) Non invasive blood pressure :

  • 11

    - Auto manual and stat mode of operation - Measurement ranges 30-255 mm of Hg. - Accuracy +/- 5 mm of Hg. 3) Pulse Oximetry : - Measurement range 0-100% - Pulse range 30 - 300 bpm - Accuracy +/- 3% 4) General - Monitor should have more than 90 hr trends for all parameters. - Trends should be available in graphical and tabulr format. - Monitor should have 2 level of alarm monitoring. - Display should have intensity control for adjusting brightness. - Monitor should have in built facility for battery back up - Battery type : Li-lion and battery charge time : 4 Hrs to charge - Battery run time : More than 3.5 Hrs with 15 min NBP operation - Monitor should be truly portable weighing less than 32 kg - Monitor should be portable with carrying handle. Pulse Oximeter :

    • Should operate in AC mains and battery. • Should have Plethysmograph waveform. • Should have bright and luminescent back lit, LCD

    display for spo 2 and pulse rate. • Facility to vary display and contrast. • Should have alarm limits for pulse rate spo 2 alarm

    silence facility should have digital auto gain for low flow

  • 12

    signal pick up. • Should have 12 hours trend for pulse rate and spo2.

    Should have speaker volumeter facility. • Should have battery level indicator. • Should have a local technical support engineers for 24

    hours back up. • A comprehensive warranty for 3 years. Original supplier

    should quote AMC after 3 years. •

    5. CU.005 E.C.G. Machine LEAD SELECTION Leads : 12-lead selection Selector Positions : 1, II, III aVR, aVL, aVL, 1mV FREQUENCY RESPONSE : Frequency response : 0.05Hz to 100 Hz (-3dB) Time : 3.2 seconds (internally) adjustable to 1.6 seconds. Filter : Muscle filter at 36 Hz (-3dB) PATIENT SAFETY : Isolation : Floating input isolation Patient leakage current : Less than 10 UA (220 V/50Hz) SENSITIVITY : Sensitivity 5mm / mV, 10mm/mV, 20 mm/mV push button selectable Auxilary input : 10mm deflection for 30mV input. Scope output : Output voltage 500 mV at 1 mV input voltage gain 1. Calibration : 1mV push button, effective at input amplifier.

  • 13

    CMRR better than 100 dB. SPECIFICATION FOR MULTICHANNEL ECG MACHINE :

    Ø Brand new and or reputed make. Ø Opera table on mains and rechargeable battery capable

    of four hours working backup. Ø Twelve lead simultaneous acquistion. Ø Three channel rhythm strips. Ø ECG with full interpretation and reasoning. Ø In built laser printing. Ø Lead fail lead reversal alarm / message. Ø Memory to store and reprint 200 or more ECGs. Ø Myogram and line frequency filter. Ø Sensitivity 5/10/15/20 with automatic and manual

    selection capability. Ø Automatic baseline adjustment Ø Speed 25mm/sec. and 50 mm/sec. Ø Standard safety measures incorporated.

    Accessories : Ø Standard accessories Ø In addition two extra sets of ECG cables with chest

    electrodes and rubber strap limb electrodes instead of

    clamp electrodes.

    Ø Trolley of standard company and make for the machine.

    Other terms and conditions :

    Ø Three years guarantee. At least six periodic

    maintenance visits during the guarantee period. All calls

  • 14

    to be attended in case of breakdown during the

    guarantee period.

    Ø Should quote for comprehensive annual maintenance

    contract for five years after the expiry of the guarantee

    period.

    Ø Quarterly maintenance visits during the CAMC period.

    All calls to be attended in case of breakdown.

    Ø The break down call to repair interval should not

    exceed five working days, falling which a fine of Rs. 500

    per day shall be payable by the agency.

    Optional :

    Ø Internal modem to transmit and receive ECG from

    similar machine through telephone line.

    Ø Upgradable to vector cardiography / stress test system.

    6. CU.0006 Cardiac monitors for emergency wards

    - it should have

    - 3 leads facility

    - HR ranges 15-300 bpm

    - HR accuracy +/- 1 bpm

    - Pacemaker detection and lead off failure alarms.

    - ECG should meet AAMIU standards and should comply with

  • 15

    IEC.

    7. CU.0007 Surgical OT Instrument SPECIFICATION FOR GENERAL INSTRUMENT SET

    1. The instruments quoted should be of high quality and standard

    2. The Instruments should be imported and of CE or FDA

    certification.

    3. Copy of the CE Certificate or FDA certificate must be enclosed

    4. The instruments must be ISO certified and copy to be enclosed

    5. Sterilization Container and Instruments should be of the same

    parent company.

    6. The Sterilization containers should meet international standards

    and approved for steam sterilization procedures to EN 285: 2008

    and validated acc tp ISO 17665 Part1:2006

  • 16

    7. The Sterilization containers should offer superior filtration

    efficiency of 99.99997%.

    8. It should have an indicator wherein colour coated green means

    the container is sterile and the container is opened, the indicator

    should automatically change to red colour indicating "unsterile"

    9. It should have reusable microbial barrier instead of disposable

    filters. The microbial barrier should be easy to remove and clean.

    10. It should have lateral flow ducts at the top for flow of air.

    11. The instrument should remain sterile in the container and the

    container should be capable of being brought into the Operation

    Room without any essential packaging

    12. It should also consist of tray and silicone matt (for micro

    instruments).

    13. The container should be of the parent company.

  • 17

    Description GENERAL INSTRUMENT SET

    GENERAL SURG./MAJOR BASIC SET FOERSTER SPONGE FORCEPS GROSS-MAIER DRESS FCPS W RAT ALLIS FORCEPS 4/5 TEETH, 6" BACKHAUS TOWEL FORCEPS SCALPEL HANDLE NO 3 SCALPEL HANDLE NO 4 TC METZENBAUM SCISSORS CVD TC-DISSECTING SCISSORS, CVD., 17.5 CM TC LEXER-FINO SCISSORS CVD STANDARD SCISSORS BL-SH CVD DRESSING FORCEPS DRESSING FORCEPS TISSUE FORCEPS 1X2 T TISSUE FORCEPS 1X2 T DE BAKEY ATRAUMA FCPS 1,5 MM POTTS-SMITH TISS FCPS 1X2 T HALSTED-MOSQUITO FCPS STR HALSTED-MOSQUITO FCPS CVD OCHSNER HAEM FCPS 1X2 T STR OVERHOLT-GEISSENDOERFER FCPS OVERHOLT-GEISSENDOERFER FCPS MIKULICZ PERIT CLAMP FORCEPS VOLKMANN RETRACTOR SHARP VOLKMANN RETRACTOR SHARP KOERTE RETRACT SHARP 23,5 CM LANGENBECK RETRACTOR LANGENBECK RETRACTOR 50X11MM MIKULICZ ABDOM RETR 120X50MM SIMON RETRACTOR 115X22MM ROUX RETRACTOR FIG. 1 ABDOMINAL SPATULA 30 MM GROOVED DIRECTOR 14,5 CM PROBE WITH EYE, 5 1/2", 14,5 CM

  • 18

    MARTIN HERNIA DIRECTOR KOENIG LIGAT.CONDUCT. 19,5CM DESCHAMPS LIGAT NEEDLE BLUNT DESCHAMPS LIGAT NEEDLE BLUNT POOLE SUCTION TUBE 10 MM TC CRILE-WOOD NEEDLEHOLDER TC MAYO-HEGAR NEEDLEHOLDER TC MAYO-HEGAR NEEDLEHOLDER MIXING BOWL SATIN 80X40 MM MIXING BOWL SATIN 0,375 LTR NEEDLE CASE PERFORATED Sterilisation Container 600X300X160 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./MINOR BASIC SET GROSS-MAIER DRESS FCPS W RAT GROSS-MAIER DRESS FCPS W RAT ALLIS FORCEPS 5/6 TEETH, 6" BACKHAUS TOWEL FORCEPS TOWEL FORCEPS W CLIP SCALPEL HANDLE NO 3 SCALPEL HANDLE NO 4 KLING-EX BLADE REMOVER DRESSING FORCEPS DRESSING FORCEPS TISSUE FORCEPS 1X2 T TISSUE FORCEPS 1X2 T TAYLOR DRESS FCPS W SCRAPER METZENB-NELSON SCS BL-BL CVD METZENB-NELSON SCS BL-BL CVD STANDARD SCISSORS BL-SH STR TC LIGATURE SCISSORS CURVED 17 CM SERRAT IRIS SCISSORS CURVED STRABISMUS SCISSORS CVD

  • 19

    HALSTED-MOSQUITO FCPS 1X2 T HALSTED-MOSQUITO FCPS 1X2 T HALSTED-MOSQUITO FCPS CVD ROCHESTER-PEAN HAEM FCPS STR OVERHOLT-GEISSENDOERFER FCPS MIKULICZ PERITONEAL CLAMP KOCHER RETRACTOR BLUNT KOCHER RETRACTOR BLUNT CUSHING RETRACTOR 12MM 20 CM ROUX RETRACTOR FIG. 1 LANGENBECK RETRACTOR TC MAYO-HEGAR NEEDLEHOLDER TC DE BAKEY NEEDLEHOLDER DESCHAMPS LIGAT NEEDLE BLUNT KOENIG LIGAT.CONDUCT. 19,5CM DOYEN GROOVED DIRECT.14 CM KRONECKER LIGAT NEEDLE PROBE 16 CM/6-1/4" GUIDE NEEDLE LANC.TP CHAR 8 GUIDE NEEDLE CVD CHARR 12 GROSS-MAIER DRESS FCPS W RAT BOWL SATIN 0,3 LTR NEEDLE CASE PERFORATED KIDNEY DISH, 170X100X35 MM CONTAINER MICROSTOP 600X300X140 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./CHOLECYSTECTOMY INSTR. TC METZENBAUM-THOREK SCISSORS TC METZENBAUM-FINO SCISSORS BL-BL CVD POTTS-SMITH SCISSORS 40 DEG MIXTER HAEMOST FORCEPS FIG 2

  • 20

    BENGOLEA HEAM FCPS 1X2 T CVD FOERSTER SPONGE FORCEPS FOERSTER SPONGE FORCEPS BLAKE GALL STONE FORCEPS CVD DESJARDINS GALL ST FORC FIG1 DOYEN ATRAU INTEST FORC 23CM DE BAKEY ATRAUMA FCPS 2.0 MM DE BAKEY ATR.FORC.1,5MM 24CM BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV BAKES GALL DUCT DILATOR OLIV GALLSTONE SCOOP, NO. 5, 15 MM, 32 CM GALLSTONE SCOOP, NO. 000, 2.7 MM, 32 CM GALLSTONE SCOOP, NO. 00, 3.4 MM, 32 CM MIXING BOWL SATIN 80X40 MM MIXING BOWL SATIN 0,375 LTR CONTAINER MICROSTOP 600X300X140 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT SUPPLEMENT SET FOR GASTRO-INTESTINAL SCALPEL HANDLE, NO. 4L, LONG, 20.5 CM TC-DISSECTING SCISSORS, CVD., 26 CM TC-SCISSORS, MAYO, STR., 17 CM TISSUE FORCEPS, 1X2 T., 25 CM ATR. FORCEPS, DE BAKEY, 2 MM, 25 CM ATR. FORCEPS, DE BAKEY, 3.5 MM, 25 CM

  • 21

    TC-NEEDLEHOLDER, DE BAKEY, 26.5 CM RETRACTOR, FRITSCH, 45X75 MM, 25.5 CM RETRACTOR, MIKULICZ, 125X50 MM, 25 CM RETRACTOR, MIKULICZ, 90X35 MM, 25 CM ABDOM. SPATULA, HABERER, 37/45 MM, 30 CM FORCEPS, KOCHER, 1X2 T., STR., 26 CM FORCEPS, BENGOLEA, CVD., 26 CM ATR. ORGAN SEIZING FORCEPS, 25 CM ATR. TISSUE FORCEPS, ALLIS, 20 CM ATR. TISSUE FORCEPS, ALLIS, 24.5 CM TISSUE FORCEPS, ALLIS, 5X6 T., 20 CM ATR. KIDNEY PEDI. FORCEPS, GUYON, 23 CM ATR. RESECTION FORCEPS, FEHLAND, 24 CM ATR. INTEST. FCPS., KOCHER, STR., 25 CM ATR. INTEST. FCPS., KOCHER, CVD., 25 CM ATR. INTEST. FCPS., DOYEN, CVD., 18 CM CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./RECTAL SURGERY INSTR. SCALPEL HANDLE NO 3L TC DE BAKEY NEEDLEHOLDER TC METZENBAUM SCISSORS CVD ZENKER DISSECTING FCPS ZENKER DISSECTING FCPS OVERHOLT-GEISSENDOERFER FCPS KELLY ABDOM RETRACT 190X38MM KELLY ABDOM RETRACT 190X57MM DE BAKEY ATR.FORC.2,7MM 30CM ALLIS ATRAUMA FORCEPS 30 CM RESANO RECTUM CLAMP DESCHAMPS LIGAT NEEDLE BLUNT

  • 22

    DESCHAMPS LIG NDL F. R-HAND BRUNNER LIGATURE NEEDLE FIG2 BRUNNER LIGATURE NEEDLE FIG2 MARTIN HERNIA DIRECTOR KOENIG LIGAT.CONDUCT. 19,5CM BRUNNER LIGAT CONDUCTOR FERZ FERROZELL LIG CONDUCTOR CVD CONTAINER MICROSTOP 600X300X140 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GEN.SURG./INCISION+DRAINAGE INSTRUMENTS SCALPEL HANDLE NO 3 DRESSING FORCEPS TISSUE FORCEPS 1X2 T GROSS-MAIER DRESS FCPS W RAT SCHEDE BONE CURETTE FIG 000 SCHEDE BONE CURETTE FIG 00 SCHEDE BONE CURETTE FIG 0 SCHEDE BONE CURETTE FIG 1 SCHEDE BONE CURETTE FIG 2 SCHEDE BONE CURETTE FIG 3 TC MAYO-HEGAR NEEDLEHOLDER STANDARD SCISSORS BL-SH CVD HALSTED-MOSQUITO FCPS CVD CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./SUTURE SET SMALL SCALPEL HANDLE NO 3 TISSUE FORCEPS 1X2 T

  • 23

    ADSON DRESSING FCPS SERRATED MICRO-MOSQUITO HAEM FCPS CVD GROSS-MAIER DRESS FCPS W RAT TC CRILE-WOOD NEEDLEHOLDER MIXING BOWL SATIN 40X19 MM STANDARD SCISSORS BL-SH CVD MICROSTOP MINISET CONTAINER 310X189X81MM TRAY, PERFORATED, 260X130X50 MM LOGISTIC FRAME, RED, F. CONTAINER IDENTIFICATION LABLE, W. TEXT, W/O HOLE SILICON MATT

    GEN.SURG./VARICOSE VEIN STRIPPING SET SCALPEL HANDLE NO 3 DISP SCALPEL FIG18 GROSS-MAIER DRESS FCPS W RAT NABATOFF VAR VEIN PROBE SET HALSTED-MOSQUITO FCPS CVD STRABISMUS SCISSORS CVD IRIS SCISSORS CURVED DOYEN GROOVED DIRECT.14 CM KOENIG LIGAT.CONDUCT. 19,5CM DESCHAMPS LIGAT NEEDLE SHARP LIGATURE NEEDLE FLEXIBLE ROCHESTER-PEAN HAEM FCPS STR CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GEN.SURG./MAMMA AMPUTATION INSTR. BRAUN VULS FCP 15.5 CM 4X4 T CZERNY VUL FCP 20.5 CM 4X4 T TC METZENBAUM-FINO SCISSORS BL-BL CVD OCHSNER HAEM FCPS 1X2 T STR

  • 24

    ROCHESTER-PEAN HAEM FCPS CVD BABY-ADSON HAEM FORCEPS KOCHER RETRACTOR SHARP ADSON TISSUE FORCEPS 1X2 T CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./TRACHEOTOMY INSTRUMENT GROSS-MAIER DRESS FCPS W RAT GROSS-MAIER DRESS FCPS W RAT ALLIS FORCEPS 4/5 TEETH, 6" SCALPEL HANDLE NO 3 SCALPEL HANDLE NO 7K SEMKEN FORCEPS 1X2 T DE BAKEY ATRAUMA FCPS 1,5 MM STANDARD SCISSORS BL-SH CVD STRABISMUS SCISSORS CVD LEXER-FINO SCISSORS CVD POTTS-SMITH SCISSORS 40 DEG MICRO-MOSQ HAEM FCPS 1X2T S MICRO-MOSQ HAEM FCPS 1X2T C MINI GEMINI FORCEPS 18 CM ITERSON TRACHEA RETRACTOR BL MINI-LANGENB.RETRACTOR 20X6 DESMARRES SADDLE RETR.16MM WULLSTEIN RETRACT SHARP 3/3T LABORDE TRACHEAL DILATOR TC CRILE-WOOD NEEDLEHOLDER TC DE BAKEY NEEDLEHOLDER MIXING BOWL SATIN 80X40 MM NEEDLE CASE ROUND 65X15 MM CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED

  • 25

    CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./AMPUTATION INSTR. VOLKMANN RETRACTOR SHARP LEXER CHISEL 15X230MM MALLET WITH PLASTIC HANDLE BONE FILE LAMBOTTE RASPATOIRE 15X210MM SCHEDE BONE CURETTE FIG 00 SCHEDE BONE CURETTE FIG 0 CHARRIERE AMPUT SAW WITH 2BL PERCY AMTUTATING RETRACTOR RUSKIN RONGEUR FORCEPS CONTAINER MICROSTOP 600X300X140 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./TYROID INSTR.SET DE BAKEY ATRAUMA FCPS 1,5 MM DE BAKEY ATRAUMA FCPS 1,5 MM ZENKER RETRACTOR 35X 8MM ZENKER RETRACTOR 60X12MM ZENKER RETRACTOR 75X12MM HEISS HAEMOST FORCEPS FIG 2 HEISS HAEMOST FORCEPS FIG 3 BABY-ADSON HAEM FORCEPS BABY-MIXTER HAEM FORCEPS OCHSNER HAEM FCPS 1X2 T STR MICRO-MOSQUITO HAEM FCPS CVD TC METZENBAUM-FINO SCISSORS BL-BL CVD TC METZENBAUM-FINO SCISSORS BL-BL CVD TC METZENBAUM-FINO SCISSORS BL-BL CVD KOENIG LIGAT.CONDUCT. 19,5CM KOENIG LIGAT.CONDUCT. 19,5CM DESCHAMPS LIGAT.NEEDLE BLUNT

  • 26

    DESCHAMPS LIGATURE NEEDLE BL LIGATURE NEEDLE FLEXIBLE YASARGIL SUCTION TUBE 2 MM YASARGIL SUCTION TUBE 2.5 MM GUIDE NEEDLE LANC.TP CHAR 8 CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT LONG INSTR.(SPLEEN/LIVER RUPTURE) TC METZENBAUM SCISSORS CVD TC METZENBAUM-FINO SCISSORS BL-BL CVD DRESSING FORCEPS TISSUE FORCEPS 1X2 T DE BAKEY ATR.FORC.1,5MM 24CM ROCHESTER-PEAN HAEM FCPS CVD OVERHOLT-GEISSENDOERFER FCPS OVERHOLT-GEISSENDOERFER FCPS ZENKER DISSECTING FCPS SATINSKY ATRAUMA FORCEPS SATINSKY ATRAUMA FORCEPS HARKEN ATRAUMA FORCEPS BRUNNER RETRACTOR 180X30 MM DEAVER RETRACTOR 38MMX12" DEAVER RETRACTOR 50MM X 12" KELLY ABDOM RETRACT 190X38MM ABDOMINAL SPATULA 50 MM TC MAYO-HEGAR NEEDLEHOLDER TC DE BAKEY NEEDLEHOLDER BRUNNER LIGATURE NEEDLE FIG1 BRUNNER LIGATURE NEEDLE FIG2 BRUNNER LIGAT CONDUCTOR FIG1 BRUNNER LIGATURE NEEDLE FIG2 DESCHAMPS LIGAT.NEEDLE SHARP DESCHAMPS LIGAT.NEEDLE BLUNT

  • 27

    KOENIG LIGAT.CONDUCT. 19,5CM KOENIG LIGAT.CONDUCT. 19,5CM CONTAINER MICROSTOP 600X300X140 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./SKIN GRAFT SET SCHINK DERMATOME VOLKMANN BONE CURETT FIG000 VOLKMANN BONE CURETT FIG 0 VOLKMANN BONE CURETT FIG 1 VOLKMANN BONE CURETT FIG 2 VOLKMANN BONE CURETT FIG 3 VOLKMANN BONE CURETT FIG 4 VOLKMANN BONE CURETT FIG 5 SCALPEL HANDLE NO 7K MICRO-MOSQ HAEM FCPS 1X2T S ADSON TISSUE FORCEPS 1X2 T ADSON DRESSING FCPS SERRATED TC HEGAR-BAUMGARTNER NDLH MIXING BOWL SATIN 80X40 MM MIXING BOWL SATIN 0,375 LTR STRABISMUS SCISSORS CVD STANDARD SCISSORS BL-SH CVD METZENB-FINO SCS BL-BL CVD IRIS SCISSORS CURVED CONTAINER MICROSTOP 600X300X140 MM TRAY, PERFORATED, 480X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GENERAL SURG./VENAE SECTIO INSTR. SCALPEL HANDLE NO 3 TISSUE FORCEPS 1X2 T MICRO FORCEPS 1X2 T 0,6 MM

  • 28

    MICRO FORCEPS 0,6 MM MICRO-MOSQUITO HAEM FCPS CVD MICRO-MOSQ HAEM FCPS 1X2T S BABY-ADSON HAEM FORCEPS BABY-ADSON HAEM FORCEPS TOHOKU TOWEL FORCEPS GROSS-MAIER DRESS FCPS W RAT KOCHER RETRACTOR SHARP LANGENBECK RETRACTOR ROUX RETRACTOR FIG 2 WEITLAN LOKT RETR BLUNT 3/4T BROM VEIN HOOK LARGE DOYEN GROOVED DIRECT.14 CM KOENIG LIGAT.CONDUCT. 19,5CM DESCHAMPS LIGAT.NEEDLE BLUNT COOPER LIGATURE NEEDLE FIG 2 TC MAYO-HEGAR NEEDLEHOLDER ALLIS FORCEPS 4/5 TEETH, 6" CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITHOUT HOLE SILICON MATT GEN.SURG./WOUND DRESSING SET LARGE STANDARD SCISSORS BL-SH CVD DRESSING FORCEPS TISSUE FORCEPS 1X2 T GROSS-MAIER DRESS FCPS W RAT ROCHESTER-PEAN HAEM FCPS STR VOLKMANN BONE CURETT FIG0000 VOLKMANN BONE CURETT FIG 00 VOLKMANN BONE CURETT FIG 0 KIDNEY BOWL 275X150X45 MM SILICON MATT GEN.SURG./WOUND DRESSING SET SMALL STANDARD SCISSORS BL-SH CVD

  • 29

    TISSUE FORCEPS 1X2 T KIDNEY BOWL 250X140X40 MM SILICON MATT GEN.SURG./INSTR.FOR HAND SURG.(EXTRAS) LEAD HAND, CHILD SIZE LEAD HAND, ADULT SIZE RADOLF NAIL FORCEPS FINGER RING SAW FORCEPS IDEAL FINGER NAIL DRILL CPL SILICON MATT GENERAL SURG./SUTURE SET LARGE SCALPEL HANDLE NO 3 DRESSING FORCEPS TISSUE FORCEPS 1X2 T MICRO-MOSQUITO HAEM FCPS CVD GROSS-MAIER DRESS FCPS W RAT RETRACTOR FINE BLUNT 16 CM WEITLAN LOKT RETR BLUNT 2/3T TC MAYO-HEGAR NEEDLEHOLDER TC LEXER-FINO SCISSORS CVD STRABISMUS SCISSORS CVD STANDARD SCISSORS BL-SH CVD MIXING BOWL SATIN 80X40 MM ADSON TISSUE FORCEPS 1X2 T MICROSTOP MINISET CONTAINER 310X189X81MM TRAY, PERFORATED, 260X130X50 MM LOGISTIC FRAME, RED, F. CONTAINER IDENTIFICATION LABLE, W. TEXT, W/O HOLE SILICON MATT GENERAL SURG./HEMORRHOIDAL SET BACKHAUS TOWEL FORCEPS GROSS-MAIER DRESS FCPS W RAT SCALPEL HANDLE NO 3 METZENB-FINO SCS BL-BL CVD STANDARD SCISSORS BL-SH CVD

  • 30

    OCHSNER HAEM FCPS 1X2 T STR HALSTED-MOSQUITO FCPS CVD MAYO-HEGAR NEEDLE HOLDER RUDD HAEMORRHOIDAL-LIGATOR 27,5 CM RUDD LOADING CONE ONLY RUBBER RINGS FOR HAEMORRHOID.-LIGATOR MCGIVNEY HAEMORRHOIDAL FORCEPS 19 CM CONTAINER MICROSTOP 300X300X140 MM TRAY, PERFORATED, 260X260X50 MM COLOR-TAG, RED CODING LABEL, WITH TEXT, WITOUT HOLE SILICON MATT

    8. CU. 0008 Orthopedic O.T. Instrument To be purchased as a set

    1. Leur Bone Rongeur Single Action Nibbler - 150 mm 2.Stille Ruskin bone rongeur double action Nibbler - a) Straight - 230 mm 3. North field bone rongeur double actionh nibbler. - a) Curved - 230 mm 4.Tudor Edward Bone cutter double action. - 260 mm 5.Liston double action bone cutter a) Straight - 265 mm b) Curved - 265 mm 6. Humbys skin grafting knives fiber handle - 300 mm 7.Silvers skin grafting knives fiber handle. - 190 mm 8.Andresons tendon tunneling forceps a) Straight - 200 mm b) Curved - 275 mm 9. Amputations saw - 24 cm Blade 10.Finger saw - 10 cm Blade 11.Ombreanee mallet - 550 gm

  • 31

    12.Mallet with rubber head - 13.Hajek mallet fiber handle - 140 gm 14. Bone File - 220 mm One side file cut Other side Rasp cut 15.Buck bone lever - 160 mm 16.Wagner bone lever - 10 x 220 mm 17.Verbragge muller bone lever - 44 x 240 mm 18.angenbeck bone lever - 230 mm 19.Lane bone holding forceps a) With Rachet - 330 mm b) Without Rachet - 330 mm 20.Fergusson bone holding forceps - 210 mm 21.A.O bone holding forceps a) Small b) Medium c) Large 22.A.O. Plate holding forceps a) Small b) Medium c) Large 23.J J Many Bone cum plate holder a) Small b) Medium c) Large 24Volkmans double ended curette - 230 mm 25. Stille Osteotomes : a) Straight - 6 x 180 mm - 12 x 200 mm - 18 x 200 mm

  • 32

    - 24 x 200 mm - 19 x 200 mm b) Curved - 6 x 180 mm - 12 x 200 mm - 18 x 200 mm - 24 x 200 mm

    26.

    Stile chisels a) Straight b) Curved

    6 x 200 mm 12 x 200 mm 18 x 200 mm 24 x 200 mm 6 x 200 mm 12 x 200 mm 18 x 200 mm 24 x 200 mm

    27.

    Stille Bone Gouges 12 x 200 mm 15 x 200 mm

    28 Bristow Periosteal elevators 220 mm

    29.

    Farabuef Raspatories Periosteal Elevators a) Straight b) Curved

    155 mm 155 mm

    30 Mayo Heagars Needle Holder a) Straight b) Curved

    150 mm 200 mm

    31 Metzenbauim scissors a) Straight b) Curved

    180 mm 180 mm

    32 Mayos Scissors a) Straight

    160 mm

  • 33

    b) Curved 160 mm

    33 Dressing Scissors Blunt x Blunt a) Straight

    150 mm

    34 Littauer ligature scissors 135 mm

    35 Adson Dissecting Forceps a) Plain b) Tooth

    125 mm 125 mm

    36 Standard Dissecting forceps a) Plain b) Tooth

    150 mm 150 mm

    37 Gillies dissecting forceps a) Plain b) Tooth

    180 mm 180 mm

    38 Mayo safety pin instrument holder 140 mm

    39 Backhaus towel clips 110 mm

    40 Allis tissue forceps 200 mm

    41 Sponge holding forceps 200 mm

    42 Mixtures ligature forceps 200 mm

    43 Kochers forceps a) Straight b) Curved

    200 mm 200 mm

    44 Halsted mosquito forceps a) Straight b) Curved

    125 mm 125 mm

  • 34

    45 Morris retractor 30 x 40 mm

    220 mm

    46 Langenbeck retractor 90 x 20 mm

    215 mm

    47 Single skin hook retractor a) Sharp b) blunt

    160 mm 160 mm

    48 Seen miler skin retractor 165 mm

    49 B.P. Handle a) No. 3 to fit Blade 10-15 b) No. 7 to fit Blade 10-15 c) No. 31 to fit Blade 10-15 d) No. 4 to fit Blade 18-25 e) No. 4L to fit Blade 18-25

    50 Adson baby Retractor, Blunt 4x 3 Promps

    145 mm

    51 Mollison retractor sharp 4 x 4 prongs 150 mm

    52 Power Drilling System with Seprate hand piece for drilling saw and reaming, along with flexible reamer set. Cutting heads from 8 mm 15 14 with. 5 mm increments. Extra flexible cord

    53 American pattern suction tip with stillet Size 5

    54 Screw Driver Hexagonal a) Fiber handle small 3.5 mm b) Fiber handle large 4.5 mm c) Self retaining 3.5 mm d) Self retaining 4.5 mm

  • 35

    55 English Plier cum cutter

    56 K wire bender with cutter a) Plier cum bender cum cutter 7" b) Plier cum bender cum cutter 9"

    57 Pointed reduction clamp a) Small b) Medium c) Large

    150 mm 210 mm 240 mm

    58 Giggly wire a) Hooks b) Imported Giggly wire

    59 Depth gauge with mm marking a) 3.5 mm b) 4.5 mm

    60 Stainless steel scale a) 6 Inches b) 12 Inches

    61 Stainless steel bowl a) Large b) Medium c) Small

    62 Kidney Tray

    63 Pneumatic Tourinquet with pump with pressure gauge with 3 size cuff.

    64 Sterilization box large 24 inches

    65 Mayo Hegars Needle Holder with Tungsten carbide tip with gold plated

    150 mm 180 mm

  • 36

    rings

    66 Metzenbaum scissors with tungsten carbide tip with gold plated rings a) Straight b) Curved

    145 mm 145 mm

    67 Mayos scissors with Tungsten carbide tip with gold plated rings a) Straight b) Curved

    145 mm 145 mm

    68 Adson Dissecting forceps with tungsten carbide tip with gold plated end a) Plain b) Tooth

    120 mm 120 mm

    69 Fine dissecting forceps with tungsten carbide tip with gold plated end a) Plain b) Tooth

    145 mm 145 mm\

    09. CU. 0009 Neuro surgical O.T. Instrument

    • Hudson brace non glare 248 mm made of non rusted stainless steel screw tightening.

    • Single precision out wire pack of 25 pcs • Costing perforator original model 12mm, 15 mm. • Cushing spherical burrs 13mm, 18 mm. • Demartel gigli saw guide. • Dandy scap hemostatic forceps curved on side 1x2 • Adsons micro forceps 1 x 2 teeth 150 mm. • Intervertebral disc biop[sy forcepos love gruewald 3mm

    x 10 mm.Straight • Intervertebral disc biopsy forceps love gruewald 3 mm x

  • 37

    10 mm • straight b) up c) down • Adson self retaining retrator blunt 330mm • Scoville reptured disc curette straight 254 mm long. • Codman leksell laminectomy ronger double action

    regular curve 245 mm, 8 mm wide jaw and 5 mm jaw. • Long tooth forceps fine 200 mm • Delicate lahey scissors, curved 146 mm. • Malls bipolar forceps, Bayonet length 178 mm. • Weltaner self retaining retraction sharp 3 x 4 teeth 250

    mm 10" (Imported). • Mayo hegar needle holder 152 mm tungsten carbide tip

    periosteal penny backer elevator. • Saw handle set • Cushing Bayonet tissue forceps, dissecting end 197

    mm • Skull Trephine with depth guard retractable centering

    drill 51 x 95 mm tungsten steel blade. • Skull Trephine with depth guard retractable centering

    drill 51 x 95 mm tungsten steel blade. • Subcutaneous tunneler for VP shunt adult and

    paediatric brain cannula. • Yasargil spring book for galea fixation with buldog

    clamp 410 mm • Adson haemostatic forceps curved. • Zaufal, Jansen Rongers : Curved double action 170 m. • Bachaus towel clips 138 mm • Hudson cerebellar, extension 102 mm. • Ray ring curette horizontal 45 degree 180 mm • Kerrison pituitary punch 1300 up cutting, 250 mm, 2

  • 38

    mm. • Perisoteal penny backer elevator.

    10. CU. 0010 Orthopediac splints /traction To be purchased as a set

    • Each set consisting of following in mentioned quantity. • All the below mentioned items will form one set and any

    thing short will disqualify compliance certificate is must. • Thomas splint, iron made, polished, ring cushioned,

    (Small, medium, large, extra large). • Bohler Braun Splint, Iron made, Polished with three

    pulleys in assembled condition. • Crammer wire splints (Small, medium, large, extra

    large). • Steinman pin (4 mm, 4.5 mm) made up of stainless

    steel with trocar tip. • Denhams pin (4.5 mm) • K wires (2mm, 2.5 mm, 3mm) • Bohler stir up • Skeletal traction kit (Consisting of protector cap for pins,

    cords for tying, spreader for cords, cord to hang wt.) • Wooden block 4'x4'x4' • Traction pulley stand made up of iron with variable

    angle for multiple pulleys to go on foot end bar of the bed.

    • Plastic pulleys with hooks. • ASHE brace. • Cervical collar Philadephe type (small, medium, large) • Clavicle brace (small, medium, large) • Adhesive skin traction kit with spreader and cord for

    traction (below knee, kit for kids)

  • 39

    • Non adhesive traction kit sponge, padded with spreader and cord for traction and clastic bandage for applying the traction (Below knee, above knee)

    • Iron weight for traction and stand for hanging weight total of 10 pounds.

    • Head halter traction kit. • Crutch field tong with guarded drill bit • Over head beam with all pulley attachment.

    11. CU. 0011 Plaster table • Table made of iron 6" length x 3" width • Rounded pipes • SS top 16 G • Powder coated • Breakable foot end • With all attachments for spica table including perineal

    post & foot traction posts. • With two step ladder • Two table mattresses with water proof covering

    12. CU. 0012 Haomo dialysis Machine HAEMODIALYSIS M/C WITH RO PLANT SPECIFICATION OF HAEMODIALYSIS PLANT A. General The haemodialysis unit should have microprocess control and capable of providing the following features : 1.1 Acetate and bicarbonate dialysis 1.2 Volumetric ultrafiltration 1.3 Sodium and UF profilings

  • 40

    1.4 Central deliver system : HD concentrate 1.5 Machine should have an option of upgradable to link up with central monitoring and data collection system. B. Graphical user interface 2.1 Should have a keyboard function keys and an enlarged and high resolution color screen for dialysis data display. 1. Blood circuit 1.1. Vascular access should have double needle with single needle as option 1.2 Blood pump flow rate range : 15-600 ml / min for haemodialysis in 5ml / min increment Accuracy : +10% 1.3 Heparin pump Infusion rate : 0.1-10 ml / hr in 1 ml /hr increment 1.4 Pressure monitoring and alarms Venous pressure monitoring with alarm Arterial pressure monitoring with alarm 1.5 Air detection should be able to detect air bubble and microbubble. 2. Dialysis circuit 2.1 Dialysate flow rate : 0,300,500,800 ml / min (user selectable) Accuracy +10% 2.2 Temperature control and alarms 2.3 Blood leak alarm should be able to detect blood leakage. Alarm should activate for blood loss rate not greater than 0.5 ml / min in to dialysate at maximum flow of haemotrocit about 20-25%. 2.4 It should be capable of an line preparation of bicarbonate dialysate fluid.

  • 41

    C. Dialysis parameter display The equipment should digitally display the following parameters : 3.1 Arterial pressure 3.2 Venous pressure 3.3 Blood flow rate 3.4 Dialysate conductivity 3.5 TMP 3.6 UF Volume 3.7 UF rate 3.8 Remaining treatment time 3.9 Heparin infusion rate 3.10 Alarm operation in text format D. Online Clearance Monitoring : Should have built in device for measurement and monitor effect of effective urea clearance (K, dialysis dose (Kt/V) and plasma sodium (Na) automatically during the treatment. E. Battery Backup : The equipment should be able to operate and monitor the extra corpeal circuit without interruption for at least 15 min. in case of AC power failure by back up battery.

    • F.R.O Plant • RO plant should have capacity of 500 litres per hour. • Water coming from RO plant should meet AAMI

  • 42

    standard water for haemodialysis application. • It should have the facility to infect both chemical and by

    least. • disinfection and cleaning process should be fully

    automatic without extra. • end user handling during the whole disinfection

    process. • Warranty period of 1 year should be provided. • Warranty period of 1 year should be provided. • AMC rates for 1 year after expiring of warranty should

    be given. • Operation manual to be provided. • Equipment to be installed in hospital. • List of 10-15 days unit / hospital consumer to whom

    equipment has been supplied during the last 5 years in M.P.

    • Pipeline from raw water tank to RO feed pump then RO to treat water.

    • tank, then tank to each dialysis machine including tank. • Two SS transfer pump one working and one stand by.

    13. CU. 0013 Endoscopy with different probes

    1-A : Should have following features and specifications :

    Field of View : 140 degree or more

    Direction of View : 0 degree forwarding viewing

    Depth end outer diameter : 9.8 mm or less

    Insertion tube outer diameter : 9.8 mm or less

    Tip bending range : Up 2100, Down 900, left & right 1000

    Working length : 1030 mm or more

  • 43

    Channel inner diameter : 2.8 mm or more

    Minimum visible distance of Instrument used thru channel

    : 3 mm or closer from distal end.

    Slimmer / light weight and fully immersible in disinfectant solution and compatible with leak testing device with auto regulated air flow and pressure two light guide, 4 or more no of remote control switches on control body. Silicon free Air / water and suction valve buttons with separate inlet ports for better air and water transmission. Instrument should be computable with narrow band imaging / FICE / I-Scan and other high advance high definition technology. 1-B : Light Source & Video Processor : Should have following features. XENON light source 300 W or more with spare lamp facility (halogen light), capable of generating special band of light (narrow band imaging / FICE / I-Scan) helpful for minute mucosal diagnosis. Automatic IRIS control, picture displays with image size adjustment (3-4 different size of image display), electronic zoom upto 1.5 x or more. Automatic brightness control, force air cooling, white balancing. On screen color control, compact and light weight, medical (high definition) grade 14" monitor (RGB), Video output : RGB output for compatible with RGB monitor. 1-C Computer : Window 7 with multimedia DVD writer and Drive, Scroll Mouse, 500 GB HHD or more, special software for recording endoscopic procedures, 15" colour monitor, Photo quality inkjet printer, UPS with 1 hour back up. 1-D ; Trolly for all above instrument

    14. CU. 0014 Rotary microtome 1. Rotary Microtome (Spencer type) 1. Manual Rotary microtome with knife holder and specimen holder and all accessories. 2. With imported heavy duty knives 200 & 240 mm

  • 44

    3. Range of thickness 0.5 micron to 60 micron increment 1 micron to 10 micron 4. Horizontal advance of specimen 30 mm and vertical stroke 60 mm. 5. Universal knife holder base for different knives along with disposable knife holder with safety finger protection guard. 6. Maximum specimen size 50x60x40 mm. 2. Fully motorized programmable Rotary Microtome : 1. Fully automatic microtome for variable specimen retraction and sectioning. 2. Two separate programmes for trimming and sectioning. 3. Speed control through cutting window. 4. Section thickness setting from 1 micron to 99 micron in 1 micron increments. 5. Section thickness from 0.5 micron to 99 micron. 6. Three sectioning modes one manual and two motorized (continuous and separate).

    15. CU. 0015 Tissue processor

    AUTOMATIC TISSUE PROCESSOR 1. Fully automatic tissue processor with capacity of 200 cassettes per run. 2. LCD display with 4 lines, Digital programming up to 9 programmes. All controls should be operated by feather touch keys. 3. Should have 12 heated stations with (3 wax baths) with cover. 4. 2 litre glass jars with fume hood with continuous agitation. 5. Programmable 12 stage timing sequence for each stage for duration 1 min to 9 hrs, in steps of one minute with delay time upto 99 hrs. 6. Automatic wax bath (PID temperature controller and PT

  • 45

    100 sensors) with facility of vacuum operation and protection hood.

    16. CU. 0016 Microscope Binocular

    SPECIFICATIONS : BINOCULAR LED MICROSCOPE 1. Optical system : Universal infinity corrected optical system, anti fungus. 2. Nosepiece : Revolving nosepiece 3. Coarse / fine focusing knob 4. Stage : Mechanical fixed stage (120 x 132 mm) with specimen holder. 5. Eyepiece tube: Binocular tube (30 degree inclination and 48-75 mm IPD) 6. Condenser : Abde condenser with objective guide marking position. Factory fitted. Numerical aperture : 1.25. 7. Eyepiece lens : 10x (anti fungus) factory fitted 8. Objectives : a. Plan Achromat 4x b. Plan Achromat 10x c. Plan Achromat 40x d. Plan Achromat 100x 9. Illumination :LED illumination source voltage : 100-240 volts AC Imported Binocular Research Microscope : 1. Optical system : Universal infinity corrected optical system, anti fungus. 2. Nosepiece : Quadruple revolving nosepiece. 3. Co-axial coarse / fine focusing knob. 4. Stage : Mechanical fixed stage (120 x 132mm) with specimen holder. 5. Eyepiece tube : Binocular tube (30 degree inclination and

  • 46

    48-75 mm IPD) Rotable 360 6. Condenser : Abde condenser with objective guide marking position. Factory fitted. Numerical aperture 1.25 7. Eyepiece lens : 10x (anti fungus) factory fitted 8. Objectives : a. Plan Achromat 4x b. Plan Achromat 10x c. Plan Achromat 40x d. Plan Achromat 100x 9. Illumination :Halogen : 12-20/W illumination source. 10. Voltage : 100-240 volts AC Research Microscope with Epi-fluorescne attachment with microscope digital camera:

    • Color corrected infinite optical system Anti fungus. • Microscope stand with coaxial focusing control knobs,

    upper stage drive stop. • Siedntopf type binocular trinoccular eyepiece tube 30 o

    rotatable 360. • Wide field eye piece 10x/18 mm with rubber shield

    (pair) • Quintuple revolving nose piece with click stop. • Objectives :

    o CCIS plan achromatic PL 4/NAO 13 o CCIS plan achromatic phase Ph 10/Na0.25 o CCIS plan achromatic Ph 40/NO.65 spring

    loaded. o CCIS plan achromatic Ph 100/NA0.1.25 spring

    loaded. • Phase contrast 5 position turret condenser for phase

  • 47

    10x40x100xBF & DF. • Centering telescope • Sub stage illuminator 6v/30w helogen lamp power supply

    100-240 CE.

    • Epi-flourescent attachment with filter cassette. • Lamp house HBO 100 W collector lens. • Mercury lamp socket for HBO 100 W, Starter unit HBO 100

    W. • Lamp centering tool. • mercury lamp HG 100 W • High resolution digital camera with 1/2 CMOP chip. • +USB 2.0 PC connection. • Real time live image resolution 5.0 mega pixel 2580x1944

    pixel. • 16 mm lens macro viewing tube, calbration slide. • Microscope eyepieces adaptors (28mm, 30mm, 34mm,

    35mm) • Image soft ware an instant image capturing real time full

    screen image programmed interval capturing video capturing. All measurements in micro, inch, mm, length, angle etc.

    • Facility for on line file sending. • Image assembly at high magnification. • Image capturing at multi focal depths.

    Branded Pc computer : Pentium 1 core chipsets with 4 GB rams 300 GB HDD, DVD writer with graphic card and high resolution 19" TFT Monitor with key board and mouse.

    17. CU. 0017 Digital Binocular Microscope with CCTV

    1 CCIS EFN Plan Achromat objectives2 2 Ergonomic Design on Continous use 3 Built in High Resolution 3.0 MP Live imaging chip 4 Enhanced optical and digital performance 5 Swing out Condenser

  • 48

    (A ) Microscope- Digital compound CCIS Binocular Optics- WF 10X/30mm Eyepiece on seidentopf Head (IPD 55-75mm) Quintuple Revolving Reversed Mounted Nosepiece for Five Objective EFN Plan Achromate 4X, 10X, 40X. 100 XS-Oil Rock and pinion focusable centerable 0.9/ 1.25 NA Abbe condensers Magnification- 40X, 100X, 400X, 1000X Total Optical I IIumination- 6 V/30W Quartz Halogen with intensity control Focusing Coaxial Coarse & fine with Shuttle Control Power Universal Voltage And built in Transformer CAMERA Imaging Chip ½ Built in 5.0 MP CMOS Live Resolution Max 2048x1536 Data Link USB2.0 Note: This microscope should be upgradeable for :-

    LED Fluorescence, Polarization, Dark Field, Phase Contrast, Teaching Head (For 2 People)

    (B) Specification for Microscope Digital Camera Model 2000 with Imaging Software Consisting of: C Mont Adaptor, High resolution Digital Camera with 1/3” CMOS Chip, 2 MP, USB 2.0 PC connection, Real time Live Image resolution, 1600x1200 pixel, 16 mm lens, Macro viewing tube, Calibration Slide, Microscope eyepiece adapters (28mm, 30mm, 34mm, 35mm). Software for both A & B would be as given below Software Included-(Function) Instant image Capturing. Real time full screen image, Programmed interval Captures, Video Capture by Time Settings. Easy Measurement calibration, Measurement in microns , inches, millimeters, Length Measurements, Ellipse, Rectangle, irregular Shape Measurements , Perimeter, Radius, Circumference Measurements, Angle Measurements, Magnifier (zoom) function, Distant image Sharing, On line files

  • 49

    sending/receiving, Sound recording, image amalgamation, Automatic image amalgamation. Image Adjustment Effects. Automatic cell Counting, Date Export, Report generating and Print Out. Interactive File format, NP3 compatible, Calibration Slide. Note: Microscope, camera and software must be Manufactured by same Company.

    DEAN GANDHI MEDICAL COLLEGE

    BHOPAL

  • 50