mutareal® factor ii
TRANSCRIPT
MutaREAL Factor II
Version 2.0 / November 2018
KF291632
KF291696
32
96
Only for in vitro diagnostics
Real-Time PCR Kit
Real-Time PCR kit for the analysis of mutation G20210Ain the FII gene based on the FRET technology
on LightCycler 1.5, 2.0 and 480.
®
Immundiagnostik AG, Stubenwald-Allee 8a, 64625 Bensheim, Germany www.immundiagnostik.com Tel.: +49 (0)6251/ [email protected] Fax: +49 (0)6251/ 849430
2
© 2018 Immundiagnostik AG / Version 2.0
MutaREAL® Factor II
Table of Contents
1 Intended Use 3
2 Introduction 3
3 Concept of the Assay 3
4 Kit Components 4
5 Required Materials 4
6 Storage and Handling 4
7 Considerations and Precautions 4
8 Test Procedure 5
8.1 PCR Preparation 5
8.2 PCR Protocol 6
9 Evaluation 6
10 Troubleshooting 7
11 Test Limitations 7
3
© 2018 Immundiagnostik AG / Version 2.0
MutaREAL® Factor II
1 Intended Use
The Factor II Real-Time PCR Kit is a FRET-based test for the analysis of the mutationG20210A in the FII gene.
2 Introduction
Prothrombin (Factor II) is the precursor of the active coagulation enzyme Thrombin(Factor IIa), which has a key position in the haemostasis and thrombosis. It convertsfibrinogen to fibrin for the blood coagulation, stimulates the aggregation of blood plateletsand activates the coagulation factors V, VII and XIII. The base substitution G20210A isassociated with an increased level of prothrombin and thus with an increasedsusceptibility to venous thrombosis.
References:De Stefano et al., The New England Journal of Medicine, 1999, 341:801-6Gerhardt et al., The New England Journal of Medicine, 2000, 342:374-80Franco et al., British Journal of Haematology, 1999, 104:50-4
3 Concept of the Assay
This sequence specific detection assay is based on fluorescence resonance energytransfer (FRET). The assay contains two specific primers flanking the target region aswell as two hybridization probes. One of the hybridization probes binds to a specific targetregion known to carry the gene mutation of interest. The second hybridization probe bindsto a sequence in close proximity to the first probe, not covering the mutation. In order to achieve FRET, one of the hybridization probes is labeled with a donor-fluorophor, the other one is labeled with an acceptor-fluorophor. If both hybridizationprobes are in immediate proximity, the donor-fluorophor transfers energy to the acceptorfluorophor following excitation. Due to this energy transfer, the acceptor-fluorophor emitslight of a longer wavelength. Since energy transfer can only occur if both hybridizationprobes are bound to the target sequence, the amount of hybridized probe pairs, andthereby the fluorescence signal proportionally increases to the amount of amplified PCRproduct. Following amplification, genotyping is then performed via melting curve analysis.For this purpose, the temperature is slowly increased after a short denaturation step andthe dissociation behavior of the probe is monitored by constantly measuring the emittedfluorescence signal. If the targeted mutation is not present, the fluorescence signaldecreases only at high temperatures since the perfectly matched hybridization probedissociates late due to its high binding affinity to the target region. However, in thepresence of the mutation, the fluorescence signal decreases earlier with increasingtemperatures since the mismatched and therefore less stable probe dissociates earlierfrom the target region.
4
© 2018 Immundiagnostik AG / Version 2.0
MutaREAL® Factor II
4 Kit Components
Factor II Real-Time PCR-Kit Volume
Reagents 32-rxn 96-rxn
Enzyme Mix (blue) 438 µL 3 x 438 µL
Detection Mix (yellow) 368 µL 3 x 368 µL
Positive Control (red) 15 µL 45 µL
Negative Control (green) 50 µL 150 µL
5 Required Materials
Required Materials - not provided:Roche LightCycler® 1.5, 2.0 or 480 Real-Time PCR-Systemo The CE conformity is only given when one of the above mentioned
components is used.LightCycler® capillaries, Roche or 96-well plate/stripes (white)
LightCycler® cooling block, Roche or cryo container for PCR reaction tubes
Pipettes (0.5 – 200 µL)o 0.5 - 10 µLo 10 - 200 µL
1.5 mL reaction tubes
6 Storage and Handling
All reagents should be stored at -20 °C.
Avoid several freeze / thaw cycles of the reagents (if necessary prepare aliquots).
The detection mixes have to be protected from exposure to light.
7 Considerations and Precautions
The regulations and principles for working in a biomolecular laboratory have to be strictlyfollowed.
All steps have to be performed in an uninterrupted manner
All PCR reagents have to be cooled while in use
The DNA purity (A260/A280 ratio) should be between 1.8 and 2.0
5
© 2018 Immundiagnostik AG / Version 2.0
MutaREAL® Factor II
8 Test Procedure
8.1 PCR Preparation
Gently thaw all components on ice. Mix them gently (do not vortex) and spin them downbefore use. Keep in mind to protect the detection mixes from exposure to light. During thePCR setup all reagents have to be kept cool.
For the amplification one PCR reaction tube (type dependent on the instrument used) isneeded per sample plus two additional tubes for the negative and positive controls. Thefollowing table shows the volume of each reagent per sample. For the analysis amastermix should be prepared for the number of samples (incl. negative and positivecontrol) (N) plus 10 % to compensate inaccuracies. The mastermix should be prepared inthe same order as indicated in the table.
Reagent Volume per 25 µLReaction
Master Mix Volume
Detection Mix (yellow) 10.5 µL 10.5 µL * (N + (N * 0.1))
Enzyme Mix (blue) 12.5 µL 12.5 µL * (N + (N * 0.1))
Thoroughly mix the mastermix by pipetting up and down or inverting (do not vortex)and shortly spin down. Aliquot 23 µL into each PCR tube. For the negative control add 2 µL of the provided negative control (green).
For the positive control add 2 µL of the provided positive control DNA (red).
Add 2 µL of each sample DNA to the corresponding PCR tube.
LightCycler® 1.5 and 2.0: Close the capillaries with the lid, transfer them into theLightCycler® carousel and centrifuge in the LightCycler® centrifuge (if a table centrifuge isused, centrifuge the capillaries in the adapters of the cooling block for 15 s at 3000 rpm).Subsequently transfer the carousel into the LightCycler® and use the PCR protocoldescribed in 8.2.
LightCycler® 480: Close the wells with the sealing-foil and centrifuge the plate for 15 s at2000 rpm. Subsequently transfer the plate into the LightCycler® 480 and use the protocoldescribed in 8.2.
6
© 2018 Immundiagnostik AG / Version 2.0
MutaREAL® Factor II
8.2 PCR Protocol
Step Temperature[°C]
Time [s] Ramp Rate[°C/s]
Cycles Acquisition
Initial Denaturation 94 120 max. 1 x None
Denaturation 94 10 max.
45 x
None
Primer Annealing 55 25 max. Single
Elongation 72 25 max. None
Melting Curve
94 30 max. 1 x None
40 30 max. 1 x None
80 0 0,1 - 0,2* 1 x Constant
Cooling 40 30 max. 1 x ---
*depending on the number of selected filters for the LightCycler® 480, it can be necessary to adjustthe acquisitions per °C.
9 Evaluation
For the evaluation of the melting curves add an analysis of the type "genotyping". Therebythe derivation of the fluorescence curve is generated. The detection wavelength is640 nm.
Temperature mutated allele: 53,0 °C (+/-2 °C) Temperature wildtype allele: 60,0 °C (+/-2 °C)
The following graphic shows the typical results for the possible genotypes: pink curve -negative control, violet and blue curve - homozygous wildtype, green Kurve -heterozygous mutation, black curve - homozygous mutation.
The provided positive control contains a template heterozygous for the point mutationG20210A.
7
© 2018 Immundiagnostik AG / Version 2.0
MutaREAL® Factor II
10 Troubleshooting
Problem Solution
No or low fluorescence peakwith positive control orsamples
Check PCR-program of the real time PCR instrument inuse and repeat with corrected protocol.
The Detection Mix was thawn / frozen more than twice orstored longer than four days at 2-8 °C. Repeat analysiswith a fresh aliquot or new Detection Mix.
Quality of DNA template is not sufficient. Use freshlyextracted DNA and measure the concentration/puritybefore use.
The Detection Mixes were not protected from light. Repeatanalysis with a fresh aliquot or new PCR reagents.
11 Test Limitations
The accuracy of genetic testing is never 100%. However, an accuracy of more than 98%was determined based on the validation data. The attending physician is free to use thetest results as a guidance in the decision making process in terms of diagnosis andtherapy. However, these recommendations are based on genetic test outcomes and needto be interpreted in the context of medical history and known familiar risks of eachindividual patient. The attending physician is fully responsible for the final diagnosis andtreatment.