multigas analysis - spacelabs healthcare
TRANSCRIPT
MULTIGAS ANALYSIS
Advanced Technology for
Speed, Accuracy and Reliability
• PHASEINgasanalysisbenchwarmsupandbegins providingfullaccuracyofgasconcentrationdataand agentidentificationinlessthan20seconds.Automatic zerocalibrationatstartup,andthenonceevery8hours ofcontinuoususe,eliminatesmultiplezeroingmessages andthedropoffofcriticaldata.Alight-emittinggasinlet providesastatusindicatortotheuser,detectingthe presenceofaNomolinesamplinglineandproviding color-codedinformationaboutbenchstatusand thealarmsystem.
• Servomexparamagneticoxygensensorprovidesinspired oxygen(FiO2),expiredoxygen(FeO2)andinstantaneousO2 values.Anon-depletingsensor,itdoesnotrequirereplacement.
• Nomoline“no-moisture”gassamplinglineprovideswater andfiltrationmanagementwithouttheneedforwatertraps. Achoiceofsingleandmultipleuseversionsavailable.
Speed and Performance in a Small Package
SpacelabsHealthcare’sMultigasModuleincorporatesadvancedtechnology– includingthePHASEINgasanalysisbench,theNomoline™(no-moisture)gas samplinglineandtheServomexO2sensor–forincreasedperformance andreliabilityinmonitoringCO2,N2O,O2andanestheticagents.
92518MultigasModule
ThisMultigasModuleiscapableofidentifyinganddisplayingagentvaluesinlessthan 20secondsafterstartup,andissuitable foruseonallpatientpopulations–fromnewbornthroughadult–duetoitslowsampleflowrate(50ml/min). Acompactsolutionfor yourcompletegasanalysisneeds,itcanbeintegrateddirectlyintoSpacelabspatientmonitoringsystems.
www.spacelabshealthcare.com
Spacelabs Healthcare Company Headquarters5150 220th Ave. SE, Issaquah, WA 98029PO Box 7018, Issaquah, WA 98027-7018
Telephone: +1 425-657-7200In North America: +1 800-522-7025Fax: +1 425-657-7212
© 2011 Spacelabs HealthcareSpecifications subject to change without notice. Ultraview SL and Xprezzon are trademarks of Spacelabs Healthcare. Other brands and product names are trademarks of their respective owners.
Product may not be available in all locations. Check with your local Spacelabs representative.
M1122-de11
For all Patient Populations – Neonate through Adult
TheMultigasModule’shighlyefficientPHASEINgasanalysis benchusesasampleflowrateofjust50ml/min–allowing side-streamgassamplingonallpatientpopulations.
Compact & Compatible
ForintegrationwithSpacelabsXprezzon™andUltraviewSL® patientmonitoringsystems.
Performance Specifications
WarmUp <20secondsforconcentrationreporting,automaticagentidentification,andfullaccuracyspecification
SampleLineFlow50ml/min±10ml/min
Occlusion Automaticallydetectsandattemptstoclear samplelineocclussions
SuspendSampling InSuspendMode,sensorscontinuetooperatebutpumpstopsandwaveformsandnumericzonesarecleared,allowingsensorstoremainwarmedup
Comprehensive Gas Analysis
• RELIABLECO2&multigasmonitoring• FASTO2technology• AUTOagentidentification -Isoflurane -Desflurane -Sevoflurane -Enflurane -Halothane
LightEmittingGasInlet(LEGI)StatusIndicatorNomolineGasSamplingLine
SystemOK
AnestheticAgentPresent
AlarmWarning