multi-slice ct or multidetector ct (mdct) 1991 · slice thickness: single detector array scanners...

69
62 Multi-slice CT or MultiDetector CT (MDCT) 1991 Multi-slice CT or MultiDetector CT (MDCT) 1991 Multiple rows of fan beam detectors Wider fan beam in axial direction Table moves much faster Substantially greater throughput

Upload: others

Post on 16-Jul-2020

9 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

62

Multi-slice CT or MultiDetector CT (MDCT) 1991

Multi-slice CT or MultiDetector CT (MDCT) 1991

Multiple rows of fan beam detectors Wider fan beam in axial direction Table moves much faster Substantially greater throughput

Page 2: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

63

Multi-slice CTMulti-slice CT

Page 3: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Multi-slice CT

64

Page 4: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Multiple detector arrays

Set of several linear detector arrays, tightly abutted

Use solid-state detector arrays Slice width is determined by the detectors, not

by the collimator (although collimator does limit the beam to the total slice thickness)

65

Page 5: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

66

Page 6: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Multiple detector arrays (cont.)

3rd generation multiple detector array with 16 detectors in the slice thickness dimension and 750 detectors along each array uses 12,000 individual detector elements

4th generation scanner would require roughly 6 times as many detector elements; consequently currently planned systems use 3rd generation geometry

67

Page 7: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Slice thickness:single detector array scanners

Determined by the physical collimation of the incident x-ray beam with two lead jaws

Width of the detectors places an upper limit on slice thickness

For scans performed at the same kV and mAs, the number of detected x-ray photons increases linearly with slice thickness

Larger slice thicknesses yield better contrast resolution (higher SNR), but the spatial resolution in the slice thickness dimension is reduced

68

Page 8: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Slice thickness:multiple detector array scanners

In axial scanning (i.e., with no table movement) where, for example, four detector arrays are used, the width of the two center detector arrays almost completely dictates the thickness of the slices

For the two slices at the edges of the scan, the inner side of the slice is determined by the edge of the detector, but the outer edge is determined either by the outer edge of the detector or by the collimator penumbra, depending on collimator adjustment

69

Page 9: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

70

Page 10: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Slice thickness: MDA (cont.)

In helical mode, each detector array contributes to every reconstructed image Slice sensitivity profile for each detector array needs to be

similar to reduce artifacts Typical to adjust the collimation so that the focal spot –

collimator blade penumbra falls outside the edge detectors Causes radiation dose to be a bit higher (especially for small slice

widths) Reduces artifacts by equalizing the slice sensitivity profiles

between the detector arrays

71

Page 11: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Detector pitch/collimator pitch

Pitch is a parameter that comes into play when helical scan protocols are used

In a helical scanner with one detector array, the pitch is determined by the collimator

Collimator pitch = table movement (mm) per 360-degree rotation of gantry / collimator width (mm) at isocenter

Pitch may range from 0.75 (overscanning) to 1.5 (faster scan time, possibly smaller volume of contrast agent)

72

Page 12: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Pitch (cont.)

For scanners with multiple detector arrays, collimator pitch is still valid

Detector pitch = table movement (mm) per 360-degree rotation of gantry / detector width (mm)

For a multiple detector array scanner with N detector arrays, collimator pitch = detector pitch / N

For scanners with four detector arrays, detector pitches running from 3 to 6 are used

73

Page 13: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Multi-detector planes

74

Page 14: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Multi-detector planesNew Technology

Page 15: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

GE QXi (multi-detector CT) acquires four interweaving helices simultaneously.e.g., 4 x 5 mm slice = 20 mm total scan width

4-slice in one rotation

76

Page 16: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Definitions of Pitch

Old definition: Table travel per rotation

P = slice thickness

New definition:

Table travel per rotationP’=

Total nominal scan width

77

Page 17: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

GE QXi High Quality (HQ) vs High Speed (HS)

Pitch = 15mm/20 mm =0.75

Pitch = 30mm/20 mm =1.5

20 mm

15 mm table travel

30 mm table travel

20 mm

78

Page 18: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

79

Typical characteristics of CT

1972 1980 1990 2000

Minimum scan time 300 s 5-10 s 1-2 s 0.3-1s

Data acquired per 360° 57.6 kB 1 MB 2MB 42 MB

Data per spiral sequence - - 24-48 MB 200-500 MB

Image matrix 802 2562 5122 5122

Power (generator) 2 kW 10 kW 40 kW 60 kW

Slice thickness 13 mm 2-10 mm 1-10 mm 0.5-5 mm

Page 19: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Toshiba Aquilion ONE CT

80

320-slice (320 x 0.5 mm), 16cm gantry rotation, Year product introduced: 2007, 7.5 MHU, more than a $1 million

Page 20: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Toshiba Aquilion ONE Vision Edition

81

640-slice. 0.275 sec rotation, 16cm gantry rotation, Year product introduced: 2012

Page 21: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Micro CT A miniaturized design The X-rayed measuring field, usually as small as 2cm3

for material testing and analysis, medical applications are on their way to taking center stage (analysis of trabecular structures in bones)

82

Page 22: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Dual Energy CT Single Source or Dual Source

83

Page 23: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Dual Energy CT

84

Page 24: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Dual Source CT

85

Page 25: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Dual Source CT

86

Page 26: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

Dual Source CTDetector 2 x Stellar detectorNumber of slices 2 x 128Rotation time 0.28 s‐1

Temporal resolution 75 ms-1, heart-rate independentGenerator power 200 kW (2 x 100 kW)kV steps 70, 80, 100, 120, 140 kVIsotropic resolution 0.33 mmCross-plane resolution 0.30 mmMax. scan speed 458 mm/s1 with Flash SpiralTable load up to 307 kgGantry opening 78 cm

87

Page 27: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

SPECT-CT

88

Page 28: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

SPECT-CT

89

Page 29: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

PET-CT

90

Page 30: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

PET-CT

4D PET-CT Image

91

Page 31: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

PE

T-C

T

92

Page 32: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

93

Measure Intensity of a Pencil Beam

X-Ray Source

Radiation Detector

Page 33: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

94

Principle of X-Ray CT

In one plane, obtain set of line integrals for multiple view angles

Reconstruct cross-sectional views

Detector

Linear scan

Angular scan

Object

Page 34: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

95

Pixels & Voxels

Page 35: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

96

Digital ImageDigital Image

2-dimensional array of individual image points calculated

each point called a pixel picture element

each pixel has a value value represents x-ray

transmission (attenuation)

Page 36: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

97

Pixels & Voxels

Pixel is 2D component of an image

Voxel is 3D cube of anatomyVolume Element

CT reconstruction calculates attenuation coefficients of Voxels

CT displays CT numbers of Pixels as gray shades

Page 37: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

98

Pixel & Voxel Size

Voxel depth same as slice thickness

Pixel dimension field of view / matrix size

FOV = 30 cm256 pixels 30 cmPixel size = ------------

256 pixels

Pixel size = 0.117 cm = 1.17 mm

Page 38: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

99

Attenuation Equation forMono-energetic Photon Beams

I = Ioe-x

I = Exiting beam intensityIo = Incident beam intensitye = constant (2.718…) = linear attenuation coefficient

•property of•absorber material•beam energy

x = absorber thickness

MaterialIo

I

x

For photons which are neither absorbed nor scattered

Page 39: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

100

Example Beam Attenuation

Using equation to calculate beam intensity for various absorber thicknesses ( = .223)

1cm100 80

I = Ioe-x

100*e-(0.223)(1) = 80-20%

Page 40: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

101

Example Beam Attenuation

Using equation to calculate beam intensity for various absorber thicknesses ( = .223)

1cm 1cm100 80 64

I = Ioe-x

100*e-(0.223)(2) = 64

-20% -20%

Page 41: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

102

Example Beam Attenuation

Using equation to calculate beam intensity for various absorber thicknesses ( = .223)

1cm 1cm 1cm100 80 64 51

I = Ioe-x

100*e-(0.223)(3) = 51

-20% -20% -20%

Page 42: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

103

Example Beam Attenuation

Using equation to calculate beam intensity for various absorber thicknesses ( = .223)

1cm 1cm 1cm 1cm100 80 64 51 41

I = Ioe-x

100*e-(0.223)(4) = 41

-20% -20% -20% -20%

Page 43: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

104

More Realistic CT Example Beam Attenuation for non-uniform Material 4 different materials 4 different attenuation coefficients

#1 #2 #3 #4

1 2 4

Io I

x

I = Ioe-(+++)x

xk

k

eII

0

IIx

kk

0ln

IIdxx 0ln)(

Page 44: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

105

Page 45: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

106

Reconstruction:Solve for ’s

16 22 11 1017

22

12

10

15

13

11 12 13 14

21 22 23 24

31 32 33 34

41 42 43 44

Page 46: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

107

Real Problem Slightly More Complex

11 12 13 14

21 22 23 24

31 32 33 34

41 42 43 44

24 13 15 22 16

35

13

22

9

14512 values

512values

Page 47: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

108

Effect of Beam Energy on Attenuation

Low energy photons more easily absorbed High energy photons more penetrating All materials attenuate a larger fraction of low

than high energy photons

Material100 80

Higher-energymono-energeticbeam

30Material

Lower-energymono-energeticbeam

100

Page 48: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

109

Mono vs. Poly-energetic X-ray Beam Equations below assume Mono-energetic x-

ray beam

#1 #2 #3 #4

1 2 4

Io I

x

I = Ioe-(+++)xI = Ioe-x

Page 49: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

110

Mono-energetic X-ray Beams

Available from radionuclide sources Not used in CT because beam intensity much

lower than that of an x-ray tube

Page 50: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

111

X-ray Tube Beam High intensity Produces poly-energetic beam

#1 #2 #3 #4

1 2 4

Io I

x

I = Ioe-(+++)xMono-energetic beam equation!

Page 51: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

112

Beam Hardening Complication Attenuation coefficients n depend on beam energy!!! Beam energy incident on each block unknown Four ’s, each for a different & unknown energy

1 2 4

1cm 1cm 1cm 1cm

I = Ioe-(+++)x

Page 52: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

113

Beam Hardening Complication

Beam quality changes as it travels through absorber greater fraction of low-energy photons removed from

beam Average beam energy increases

1cm 1cm 1cm 1cm

Fewer PhotonsBut higher avg

kV than A

Fewer PhotonsBut higher avg

kV than B

A B

Fewer PhotonsBut higher avg

kV than C

C D

Fewer PhotonsBut higher avg

kV than D

E

Page 53: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

114

Reconstruction

Scanner measures “I” for thousands of pencil beam projections

Computer calculates tens of thousands of attenuation coefficients one for each pixel

Computer must correct for beam hardening effect of increase in average beam energy from one side of

projection to other

I = Ioe-(++++)x

Page 54: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

115

CT Number (The Hounsfield Unit)

Calculated from reconstructed pixel attenuation coefficient

t - W)HU= CT # = 1000 ------------

W

Where:t = linear attenuation coefficient for tissue in pixelW = linear attenuation coefficient for water

Caculate CT # for Water. Answer: 0Caculate CT # for Air. Answer: -1000

Page 55: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

116

CT Numbers for Special Stuff

Bone: +1000 Water: 0 Air: -1000

t - W)CT # = 1000 ------------

W

Page 56: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

117

The Hounsfield scale

Page 57: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

118

Digital Image MatrixDigital Image Matrix

125 25 311 111 182 222 176

199 192 85 69 133 149 112

77 103 118 139 154 125 120

145 301 256 223 287 256 225

178 322 325 299 353 333 300

Page 58: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

119

Numbers / Gray ShadesNumbers / Gray Shades

Each number of a digital image corresponds to a gray shade for one pixel

Page 59: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

120

Digital to Analog Conversion(D to A)

Computer reconstructs digital image set of numbers

Computer displays analog image

125 25 311 111 182 222 176

199 192 85 69 133 149 112

77 103 118 139 154 125 120

145 301 256 223 287 256 225

178 322 325 299 353 333 300

Page 60: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

121

Analog vs. Digital Images

Analog continuous gray

shade information Digital

Discrete gray shade information

Page 61: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

122

Digital Image FormationDigital Image Formation

Clinical ImageScreen Wire Mesh

Page 62: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

123

Digital Image Formation:Sampling

Digital Image Formation:Sampling

Place mesh over image

Assign each square (pixel) a value based on density

Pixel values form the digital image

120

-10

-650

Page 63: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

124

Digital Image Formation:Sampling

Digital Image Formation:Sampling

Each pixel assigned a value

Value averages entire pixelAny spatial variation

within a pixel is lostThe larger the pixel,

the more variation120

-10

-650

Page 64: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

125

Digital Image FormationDigital Image Formation The finer the mesh (sampling), the more accurate the

digital rendering

Page 65: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

126

What is this?What is this?

12 X 9 Matrix

Page 66: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

127

Same object, smaller squaresSame object, smaller squares

24 X 18 Matrix

Page 67: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

128

Same object, smaller squaresSame object, smaller squares

48 X 36 Matrix

Page 68: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

129

Same object, smaller squaresSame object, smaller squares

96 X 72 Matrix

Page 69: Multi-slice CT or MultiDetector CT (MDCT) 1991 · Slice thickness: single detector array scanners Determined by the physical collimation of the incident x- ray beam with two lead

130

Same object, smaller squaresSame object, smaller squares

192 X 144 Matrix