IMPACT ASSESSMENT OF STABILIZATION DEVICES ON CLABSI Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors and Masters in Nursing Science. His initial 8 years in the art of nursing was in the specialty of burns/trauma but for the past 26 years has focused on Vascular Access as a clinical/education specialty. Mark, as Senior Vascular Access Specialist, along with a great team; assist with the education of vascular access by hosting Vascular Access Residency program. Mark was honored to serve as the 2014-2015 Board President of the Association for Vascular Access (AVA) after serving as Director at Large for 4yrs and past Scientific Meeting Chairperson. Mark has presented local, national, and international presentations related to all aspects of vascular access. Mark is married and they are the proud adopted parents of the best dogs in the world! (seen is upcoming slide!)

Upload: others

Post on 02-Mar-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

IMPACT ASSESSMENT OF STABILIZATION DEVICES ON CLABSI Mark Rowe MNSc, RNP, VA-BC®

Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors and Masters in Nursing Science. His initial 8 years in the art of nursing was in the specialty of burns/trauma but for the past 26 years has focused on Vascular Access as a clinical/education specialty. Mark, as Senior Vascular Access Specialist, along with a great team; assist with the education of vascular access by hosting Vascular Access Residency program.

Mark was honored to serve as the 2014-2015 Board President of the Association for Vascular Access (AVA) after serving as Director at Large for 4yrs and past Scientific Meeting Chairperson. Mark has presented local, national, and international presentations related to all aspects of vascular access.

Mark is married and they are the proud adopted parents of the best dogs in the world! (seen is upcoming slide!)

Page 2: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Financial Disclosure • Mark Rowe

• Past-President, Association for Vascular Access (AVA)

• Employer: University of Arkansas for Medical Sciences (UAMS)

• Independent Consult/Speaker:

– Interrad Medical, Inc.

– Ethicon, Inc.

– Becton Dickinson and Company, Inc.

– 3M, Inc.

– I will not discuss off label use and/or investigational use in my presentation

Page 3: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Impact Assessment of Securement Devices on CLABSI

Mark Rowe MNSc, RNP, VA-BC®

Sr. Vascular Access Specialist

University of Arkansas for Medical Sciences

CE Code (Attendee Use Only): _______________________________

Page 4: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Objectives • Learner will:

– Understand the difference between subcutaneous engineered securement device (SESD) and

adhesive engineered securement device (AESD)

– Understand retrospective methodology used to accumulate data

– Understand the risk reduction comparison, as related to securement device selection, as it related to CLABSI

Page 5: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Our Story at University Arkansas Medical Sciences

• 500+ all Private beds

• Only level 1 Trauma center in state

• Only high risk birth center in state

• 7 Institutes on campus

Page 6: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Do We Look Alike? Patient Visit 2017

ED Visits 60,861

Surgery Cases 19,262

Outpatient Visits 485,121

Infusion Visits 44,655 (122.3/day)

Vascular Access 2017

Vascular Access Procedures 2,603

PICCs 1,748

Ultrasound PIVs 668

Chest Procedures 187

Page 7: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Anyone Remember This?

Page 8: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

What is your experience with Suture?

Page 9: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Suture • Multiple punctures to tissue creating infection risk

• Variation in technique as broad as inserter base

• Does not prevent movement long term

• Associated with safety issues, skin tears

• Hinders care and maintenance

• May require replacement, and additional punctures

Page 10: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

What is your experience with Adhesive Engineered Securement Devices?

Page 11: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Adhesive Engineered Securement Devices

• Evidence suggests frequent migration & dislodgement up to 20%

• May damage skin, Medical Adhesive Related Skin Injury

• Care and maintenance

– Vulnerable to movement & catheter loss

– Must replace with each dressing change

– Inconsistency of care with patient transitions • Are they being replaced?

• Are they available in community care?

– Material costs over time are burdensome • Are the costs covered for the patient after insertion?

Page 12: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

What is your experience with Subcutaneous Engineered Securement Devices?

Page 13: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

• Learning curve

• Change is not easy (Never Is!)

• Perceived claims vs substantiated truths

– Pain

– Bleeding

– Looks barbaric

– Causes INFECTIONS!?! (BUT DOES IT?)

What is your experience with Subcutaneous Engineered Securement Devices?

Page 14: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Our Team & Impact Assessment

Page 15: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Methods • Retrospective data analysis of 3 years of hospital PICC data

• Analysis of CLABSI was segmented by: – Patient demographics: age, gender, diagnosis

– Placement Arm

– Device Type

– Dwell Time

– Inserter Type

– CLABSI Organisms

– Securement Type

Page 16: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Sources of CLABSI Contaminated

Infusate Fluid or Medication

Extrinsic Sources

Skin Organisms Endogenous Flora Extrinsic Sources –

Pistoning/Migration Invading wound

Contamination of Device Prior to

Insertion Rare post- bundle

Contamination of Catheter Hub

Extrinsic Sources Endogenous Flora Hematogenous Seeding

From distant infection

Page 17: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Departmental PICC Data Bedside Vascular Access

Using SESD

Interventional Group

Using AESD

2015 1827

2016 1795

2017 1688

TOTAL 5310

2015 272

2016 215

2017 203

TOTAL 690

Page 18: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

0

2

4

6

8

10

12

14

16

18

2015 2016 2017

Total CLABSI by Gender

Male Female

Mean Age 51 y

Mean Age 51 y

Mean Age 51 y

Page 19: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Oncology 44%

Medicine 24%

Gastronterology 16%

Surgery 11%

Orthopedic 5%

PICC PATIENT DIAGNOSIS DISTRIBUTION

Page 20: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

0

2

4

6

8

10

12

14

16

18

2015 2016 2017

CLABSI by Insertions Side Arm

Right Left

Page 21: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

0

2

4

6

8

10

12

14

16

18

2015 2016 2017

CLABSI Occurance by Lumens

Single Lumen Double Lumen Triple Lumen (OSH)

Page 22: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

0

10

20

30

40

50

60

70

80

2015 2016 2017

Mean Dwell Days for PICC CLABSI by Securement Type

SESD AESD

Page 23: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Candida 59%

Kleb 9%

Entracoc 8%

Ecoli 8%

VRE 8%

Staph Epi 8%

2015 CLABSI ORGANISMS

Page 24: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Candida 22%

Enterab 11%

Entracoc 11%

Ecoli 22%

VRE 34%

2016 CLABSI ORGANISMS

Page 25: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

Candida 6%

Kleb 18%

Entracoc 23%

Ecoli 23%

MRSA 6%

Pseudo 6%

Strep 6%

VRE 6%

Acinetobacter 6%

2017 CLABSI ORGANISMS

Page 26: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

2015

SESD Infections with SESD AESD Infections with AESD

1.47%

0.44%

Page 27: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

2016

SESD Infections with SESD AESD Infections with AESD

1.40%

0.33%

Page 28: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

2017

SESD Infections with SESD AESD Infections with AESD

1.97%

0.77%

Page 29: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

3 Year Total

SESD Infections with SESD AESD Infections with AESD

1.59%

0.51%

Page 30: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

CLABSI per 100 Securement Devices

-0.44

-0.50

-0.49 -0.48

Relative Risk Reduction

Page 31: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors

In Conclusion • SESD does not increase chance of CLABSI…all indications

(Relative Risk) is that it decreases!

• SESD has become international recognized as a patient centered securement

• Securement device will continue to evolve… but so must we!

• The risk of having a CLABSI with an SESD is consistently about half as much for a Subcutaneous device as it is with an Adhesive securement device.