Page 1: Mounting Media by Dr Iram

Mounting Media

Dr Iram Iqbal

Page 2: Mounting Media by Dr Iram

Aim

To define and discuss different type of mounting medias ,their technique for mounting sections ,and the qualities of the ideal mountant.

Page 3: Mounting Media by Dr Iram

sequence

Defination Qualities of an ideal mountant Classification of mounting media Cover glasses Properties of the good mounted

sections Technique for mounting sections Labelling & and storage of slides

Page 4: Mounting Media by Dr Iram

MOUNTING OF SECTIONS

After staining and before examination, sections should be mounted with a suitable medium under cover slips .

To facilitate ease of handling and storage and to prevent damage to the section.

The liquid used for mounting slides is known as “Mountant”.

Page 5: Mounting Media by Dr Iram

Qualities of an ideal mountant

An ideal mounting medium must fulfill following criteria; It should fulfill the basic

requirement of protecting from dust and damage.

It should preserve the cell characteristics and quality of the stain.

It should be colourless and transparent .

Page 6: Mounting Media by Dr Iram

AN IDEAL MOUNTANT The medium should have

a refractive index as close as possible to that of the glassi,e,; 1.5 .

It should not cause stains to “bleed”or fade

It should not crack or appear granular on setting.

It should be non sticky and harden quickly.

Page 7: Mounting Media by Dr Iram

Cont…

It should not shrink back from the edges of the cover glass.

It must be free flowing. It must be free from bubbles. It should be resistant to

contamination Once set, the mountant should

remain stable.

Page 8: Mounting Media by Dr Iram

Types of mounting Media

Aqueous based mounting

media

Resinous mounting media .

Page 9: Mounting Media by Dr Iram

AQUEOUS MOUNTANT

Apathy,s Medium (RI 1.52)

Farrant,s Medium (RI 1.43)

Fructose (Levulose) Syrup (RI 1.47)

Glycerin Jelly (RI i.47)

Highman,s Medium (RI 1.52)

Page 10: Mounting Media by Dr Iram

RESINOUS MOUNTANTS Canada balsam (RI 1.52) Cover bond (RI I.53) DePeX (RI 1.52) Gurr,s Neutral Mounting Medium(RI 1.51

) Histoclad( RI 1.54) Permount (RI 1.526) Pro-Texx (RI 1.495) Technicon Resin (RI 1.62) UVinert (RI 1.517) XAM (RI 1.52)

Page 11: Mounting Media by Dr Iram

It is important to note also that the clearing agent used must be compatible with the mounting medium, or the sections must be thoroughly dried prior to mounting.

Page 12: Mounting Media by Dr Iram

Aqueous Mounting Media

Aqueous mounting media are used for mounting sections from distilled water.

Page 13: Mounting Media by Dr Iram

AQUEOUS MOUNTING MEDIA

It is used in cases where the stain would be decolourized or removed by alcohol and xylene

Aqueous mountants require the addition of a bacteriostatic agent such as phenol to prevent the growth of molds

Once the medium has hardened, a permanent seal can be obtained by ringing the coverglass with clear nail polish

Page 14: Mounting Media by Dr Iram

Apathy’s Medium (RI 1.52)It is a very useful medium for mounting

sections forfluorescent microscopy.

Gum Arabic 50g Sucrose 50g Distilled water 50ml Thymol 1 crystal

This medium remains fluid in storage andhardens by evaporation .Should be stored in well stoppered bottle.

Page 15: Mounting Media by Dr Iram

Farrant`s Medium (RI1.43)

Gum Arabic 50gGlycerin 50mlDistilled water 50mlSodium merthiolate 0.025g

This medium remains fluid in storage there for more convenient than glycerin jelly.

Its disadvantage lies in the fact that It takes longer to set. Store in a well stoppered bottle.

Page 16: Mounting Media by Dr Iram

Fructose (Levulose) Syrup (RI 1.47)Recommended as a temporary mountant

and forspecial techniques.

Fructose 75g Distilled water 25ml

Glycerin Jelly (RI 1.47)An excellent routine mountant for fat stains.

Gellitin 10g Distilled water 60ml Glycerin 70ml Phenol 0.25g

The gelatin causes this medium to solidify and it

must be melted before use.

Page 17: Mounting Media by Dr Iram

This is easily done by placing it in the oven or a water bath at 60 C Sometime prior to use.

Do not shake or stir the medium during melting as this will produce bubbles that are difficult to remove.

Page 18: Mounting Media by Dr Iram

Highman`s Medium (RI 1.52)This medium is recommended for use with the metachromatic dyes especially methyl violet.

Gum Arabic 20g Sucrose 20g Potassium acetate 20g Sodium merthiolate 10ml Distilled water 40ml

This medium remains fluid in storage and hardens by evaporation. Store in a well stoppered bottle.

Page 19: Mounting Media by Dr Iram

RESINOUS MOUNTING MEDIA

Resinous mountants are either natural or synthetic resins dissolved in solvents such as benzene, toluene or xylene.

Page 20: Mounting Media by Dr Iram

Canada balsam It is very expensive

natural resin.this is prepared by collecting the resin exuded by Abies Balsamica (the Balsam fir) and dilute it in the solvent eg, xylene.

Page 21: Mounting Media by Dr Iram

Canada balsam As canada balsam does not mix

with water mounting in it implies the use of sequence of dehydration,starting with Low grade alcohols followed by High grade alcohol Absolute alcohol Mixed cleaning agent plus alcohol Cleaning agent Cleaning agent mixed with xylene Pure xylene and Balsam dissolved in xylene.

Page 22: Mounting Media by Dr Iram

toluene or benzene could be used instead of xylene ,but these solvent are equally toxic and dangerous.

Development of some synthetic is substitutes for balsam doesn't solve the problem. they are equally expensive, that need the same steps and use the same toxic solvents.

There are less toxic and less dangerous substances but they are expensive.

Page 23: Mounting Media by Dr Iram

Canada balsam has

an RI of1.52 very

near of the glass

slides and cover

slips

The mounting

media must always

have an RI higher

than the mounted

sample.

Page 24: Mounting Media by Dr Iram

Canada balsam (RI 1.52)This may be purchased as a readymade solution of 60% resin by weight in xylene or in the dry form.

If it is obtained in the dry form, it is made up as a 55 – 70% solution in xylene.

A few chips of marble in the bottom of the bottle will help reduce acidity.

Coverbond (RI 1.53)This is Harleco Synthetic resin (H.S.R.) dissolved as a 60% solution by weight in toluene.it may be purchased s dry resin and may be dissolved in toluene or xylene.

Page 25: Mounting Media by Dr Iram

DePeX (RI 1.52)

Gurr`s DePeX is a polysterene resin dissolved as a 20% solution in xylene. Because this mountant tends to shrink on hardening, it should be added to excess. Overflow is easily removed with a scalpel after drying is complete.

Gurr`s Neutral Mounting Medium (RI 1.51)This is a mixture of coumarone and other resins as 76% solution in cineol. This mountant is a rather viscous solution.

Histoclad (RI 1.54)Clay Adam’s Histoclad is a 60% solution of a synthetic resin in toluene.

Permount (RI 1.528)Fisher’s Permount is a 60% solution of naphthalene polymer in toluene.

Page 26: Mounting Media by Dr Iram

Pro-Texx (RI 1.495)Pro-Texx from Lerner Laboratories is a mounting medium of neutral pH and with an antioxidant additive to preserve stain quality. It is soluble in both toluene and xylene.

Technicon Resin (RI 1.62)This mountant from the Technicon Corporation is a 60% solution of couramone-indene polymer in benzene and xylene in equal parts.

It has a tendency to form bubbles on hardening due to high volatility of benzene.

UVinert (RI 1.517)A patented mounting medium that is non fluorescent from Gurr.

Page 27: Mounting Media by Dr Iram

Properties Of Good Mounted Sections

Present a neat appearance. There should be no excess

mountant on or around the coverglass.

There should be no bubbles under the cover glass.

Page 28: Mounting Media by Dr Iram

Usual causes of messy slides are

Too much mountant Too thick a mountant that has not

been spread to the edges of the cover glasse

If the mountant is too thick it can be diluted with either toluene or xylene.

Page 29: Mounting Media by Dr Iram

Coverglasses

Glass coverslips are used to protect specimens from physical damage and to keep them from drying out when exposed to air after staining.

Page 30: Mounting Media by Dr Iram

COVERGLASSES Coverglasses are

either square or rectangular

Glass coverslips are used to protect specimens from physical damage

They come in different sizes and widths

Page 31: Mounting Media by Dr Iram

Cont…

They are either

22mm or 24mm

wide and come

in four different

lengths such as,

22mm, 30mm,

40mm or 50mm

Page 32: Mounting Media by Dr Iram

Coverglasses Coverglasses are manufactured to a

specified thickness to complement the optical specifications of microscope objectives lenses(the recommended thickness is indicated on the barrel of the lens and is normally 0.17 mm).

Various thicknesses are available (with some variation between manufacturers) and are designated

Page 33: Mounting Media by Dr Iram

Cont…

Various thicknesses are available No. 1 (0.13 - 0.17 mm) No. 1½ (0.16 - 0.19 mm) No. 2 (0.19 - 0.25 mm)

Page 34: Mounting Media by Dr Iram

TECHNIQUE FOR MOUNTING SECTIONS

Select the appropriately sized coverslip and place on a white paper sheet.

Place a drop of mountant in the centre of the coverslip

Invert the slide (section face down) over the coverslip and with one end resting on the paper sheet; gradually lower the other end until the mountant touches the section.

Mountant will spread quickly over the section, between slide and coverslip.

Page 35: Mounting Media by Dr Iram

Cont… The slide, with coverslip attached, is

then turned upright. Any trapped air is gently squeezed out

whilst aligning the coverslip. The mountant is allowed to set. The time required will depend upon

the particular agent used. If the result is inadequate, slides are

returned to the solvent (appropriate to the mountant) to have the coverslip removed and the process repeated

Page 36: Mounting Media by Dr Iram

Other techniques of mounting

Place the slide on a level surface, and apply a drop of mountant using the dispenser rod

Hold the cover slip at a 45° angle to the surface of the slide, and allow the bottom edge to touch the drop of mountant. When the drop has spread along the edge of the slip, let go of the slip and allow the Histomount to spread slowly

Page 37: Mounting Media by Dr Iram

Cont…. The stained section

on the slide is mounted under a thin glass coverslip. The technician takes care to keep any bubbles from being trapped in the resinous mounting medium.

Page 38: Mounting Media by Dr Iram

Cont…

Excess mounting medium may be removed while wet with a tissue, or with a razor blade when dry.

Histomount will dry sufficiently to be read in 30 minutes.

Page 39: Mounting Media by Dr Iram

LABELLING & STORAGE OF SLIDES

After staining and mounting apply a paper label with:

Name of tissue date laboratory

number staining method