morgan lens chart rev - quadmed

1
Step 1: INSERTION: Instill topical ocular anesthetic, if available. *Allows Lens to “float” over cornea and sclera. **Recommendation based on pH: Tears approximately 7.1, Normal Saline 4.5 to 7.0, Lactated Ringers 6.0 to 7.5. ***Use only when indicated. Morgan Lens Instructional Chart Instructions for using the Morgan Lens for continuous medication or lavage to the cornea and conjunctiva. Quality Certified ISO 9001 & 13485 Solution Mode with Morgan Lens Rate Frequency Step 2: Attach Morgan Lens Delivery Set, I.V., set-up or syringe using solution and rate of choice; START FLOW.* Step 5: REMOVAL: CONTINUE FLOW, have patient look up, retract lower lid—hold position. Step 3: Have patient look down, insert Morgan Lens under upper lid. Have patient look up, retract lower lid, drop lens in place. Step 6: Slide Morgan Lens out; TERMINATE FLOW. Step 4: Release the lower lid over Morgan Lens and adjust flow. Tape tubing to patient’s forehead to prevent accidental lens removal. Absorb outflow with the Medi-Duct. DO NOT RUN DRY. Ocular injury due to acid burns or solvents, gasoline, detergents, etc. Alkali burns Non-embedded foreign bodies Foreign body sensation with no visible foreign body Routine pre-operative Eyelid surgery Severe infection Call toll free to place an order and inquire about our training materials. This easy process is demonstrated in our free In-Service Video or PowerPoint Presentation CD. Just give us a call at 1-800-423-8659. We’ll be happy to send you one. P.O. Box 8719, Missoula, MT 59807 USA To order: 800-423-8659 Tel: 406-728-2522 Fax: 406-728-9332 www.morganlens.com email: [email protected] Lactated Ringer’s** I.V. Solution Lactated Ringer’s** I.V. Solution Lactated Ringer’s** I.V. Solution Lactated Ringer’s** I.V. Solution Lactated Ringer’s** I.V. Solution with suitable antibiotic and steriod*** 20 cc sterile solution 10 cc of preferred ocular antiseptic Morgan Lens Delivery Set or I.V. set-up Morgan Lens Delivery Set or I.V. set-up Morgan Lens Delivery Set or I.V. set-up 20 cc syringe 10 cc syringe Morgan Lens Delivery Set or I.V. set-up Morgan Lens Delivery Set or I.V. set-up 500 ml rapid/free flow. Reassess and continue at slower rate. 2000 ml rapid/free flow. Reassess. Continue at 50 ml/hour or 15 drops/minute. 500 ml rapid/free flow. Reassess and continue at slower rate. Slowly without force. Slowly without force. 4 drops/minute. 50 ml/hour or 15 drops/minute. Once. Repeat as necessary. Once. Repeat as necessary. Once. Repeat once if necessary. Once. During entire procedure. Continuous for 70 hours, then 10-hour intervals until marked improvement. Continuous until pH of cul-de-sac is returned to neutrality. Morgan Lens Uses

Upload: others

Post on 03-Dec-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Step 1: INSERTION: Instill topical ocularanesthetic, if available.

*Allows Lens to “float” over cornea and sclera.**Recommendation based on pH: Tears approximately 7.1, Normal Saline 4.5 to 7.0,

Lactated Ringers 6.0 to 7.5. ***Use only when indicated.

Morgan Lens Instructional ChartInstructions for using the Morgan Lens for continuous medication or lavage to the cornea and conjunctiva.

Quality CertifiedISO 9001 & 13485

Solution Mode with Morgan Lens Rate Frequency

Step 2: Attach Morgan Lens Delivery Set, I.V.,set-up or syringe using solution and rate ofchoice; START FLOW.*

Step 5: REMOVAL: CONTINUE FLOW, havepatient look up, retract lower lid—hold position.

Step 3: Have patient look down, insert MorganLens under upper lid. Have patient look up,retract lower lid, drop lens in place.

Step 6: Slide MorganLens out; TERMINATE FLOW.Step 4: Release the lower lid over MorganLensand adjust flow. Tape tubing to patient’s foreheadto prevent accidental lens removal. Absorb outflowwith the Medi-Duct. DO NOT RUN DRY.

Ocular injury due to acid burns orsolvents, gasoline, detergents, etc.

Alkali burns

Non-embedded foreign bodies

Foreign body sensation with novisible foreign body

Routine pre-operative

Eyelid surgery

Severe infection

Call toll free to place an order and inquire about our training materials.This easy process is demonstrated in our free In-Service Video or

PowerPoint Presentation CD. Just give us a call at 1-800-423-8659. We’ll be happy to send you one.

P.O. Box 8719, Missoula, MT 59807 USATo order: 800-423-8659Tel: 406-728-2522Fax: 406-728-9332www.morganlens.comemail: [email protected]

Lactated Ringer’s** I.V. Solution

Lactated Ringer’s** I.V. Solution

Lactated Ringer’s** I.V. Solution

Lactated Ringer’s** I.V. Solution

Lactated Ringer’s** I.V. Solution with suitableantibiotic and steriod***

20 cc sterile solution

10 cc of preferred ocular antiseptic

Morgan Lens Delivery Setor I.V. set-up

Morgan Lens Delivery Setor I.V. set-up

Morgan Lens Delivery Setor I.V. set-up

20 cc syringe

10 cc syringe

Morgan Lens Delivery Setor I.V. set-up

Morgan Lens Delivery Setor I.V. set-up

500 ml rapid/free flow. Reassessand continue at slower rate.

2000 ml rapid/free flow. Reassess.Continue at 50 ml/hour or 15 drops/minute.

500 ml rapid/free flow. Reassess andcontinue at slower rate.

Slowly without force.

Slowly without force.

4 drops/minute.

50 ml/hour or 15 drops/minute.

Once. Repeat as necessary.

Once. Repeat as necessary.

Once. Repeat once if necessary.

Once.

During entire procedure.

Continuous for 70 hours, then10-hour intervals untilmarked improvement.

Continuous until pH ofcul-de-sac is returned to neutrality.

Morgan Lens Uses