morgan lens chart rev - quadmed
TRANSCRIPT
Step 1: INSERTION: Instill topical ocularanesthetic, if available.
*Allows Lens to “float” over cornea and sclera.**Recommendation based on pH: Tears approximately 7.1, Normal Saline 4.5 to 7.0,
Lactated Ringers 6.0 to 7.5. ***Use only when indicated.
Morgan Lens Instructional ChartInstructions for using the Morgan Lens for continuous medication or lavage to the cornea and conjunctiva.
Quality CertifiedISO 9001 & 13485
Solution Mode with Morgan Lens Rate Frequency
Step 2: Attach Morgan Lens Delivery Set, I.V.,set-up or syringe using solution and rate ofchoice; START FLOW.*
Step 5: REMOVAL: CONTINUE FLOW, havepatient look up, retract lower lid—hold position.
Step 3: Have patient look down, insert MorganLens under upper lid. Have patient look up,retract lower lid, drop lens in place.
Step 6: Slide MorganLens out; TERMINATE FLOW.Step 4: Release the lower lid over MorganLensand adjust flow. Tape tubing to patient’s foreheadto prevent accidental lens removal. Absorb outflowwith the Medi-Duct. DO NOT RUN DRY.
Ocular injury due to acid burns orsolvents, gasoline, detergents, etc.
Alkali burns
Non-embedded foreign bodies
Foreign body sensation with novisible foreign body
Routine pre-operative
Eyelid surgery
Severe infection
Call toll free to place an order and inquire about our training materials.This easy process is demonstrated in our free In-Service Video or
PowerPoint Presentation CD. Just give us a call at 1-800-423-8659. We’ll be happy to send you one.
P.O. Box 8719, Missoula, MT 59807 USATo order: 800-423-8659Tel: 406-728-2522Fax: 406-728-9332www.morganlens.comemail: [email protected]
Lactated Ringer’s** I.V. Solution
Lactated Ringer’s** I.V. Solution
Lactated Ringer’s** I.V. Solution
Lactated Ringer’s** I.V. Solution
Lactated Ringer’s** I.V. Solution with suitableantibiotic and steriod***
20 cc sterile solution
10 cc of preferred ocular antiseptic
Morgan Lens Delivery Setor I.V. set-up
Morgan Lens Delivery Setor I.V. set-up
Morgan Lens Delivery Setor I.V. set-up
20 cc syringe
10 cc syringe
Morgan Lens Delivery Setor I.V. set-up
Morgan Lens Delivery Setor I.V. set-up
500 ml rapid/free flow. Reassessand continue at slower rate.
2000 ml rapid/free flow. Reassess.Continue at 50 ml/hour or 15 drops/minute.
500 ml rapid/free flow. Reassess andcontinue at slower rate.
Slowly without force.
Slowly without force.
4 drops/minute.
50 ml/hour or 15 drops/minute.
Once. Repeat as necessary.
Once. Repeat as necessary.
Once. Repeat once if necessary.
Once.
During entire procedure.
Continuous for 70 hours, then10-hour intervals untilmarked improvement.
Continuous until pH ofcul-de-sac is returned to neutrality.
Morgan Lens Uses