molecular genetic pathology || molecular testing: regulatory issues
TRANSCRIPT
Molecular Testing: Regulatory Issues 44Frank S. Ong, Wayne W. Grody, and Kingshuk Das
Contents
44.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1092
44.2 Federal Regulation: ClinicalLaboratory Improvement Act . . . . . . . . 1092
44.2.1 Chronology of Clinical
Laboratory Improvement Act
Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1092
44.2.2 Laboratories Regulated by Clinical
Laboratory Improvement Act . . . . . . . . . . . 1092
44.2.3 Categories of Testing . . . . . . . . . . . . . . . . . . . . 1093
44.2.4 Clinical Laboratory Improvement Act
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1094
44.2.5 Proficiency Testing . . . . . . . . . . . . . . . . . . . . . . 1096
44.2.6 Facilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1097
44.2.7 Quality Control and Assurance . . . . . . . . . 1098
44.2.8 Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1102
44.2.9 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1103
44.2.10 Enforcement of Regulations . . . . . . . . . . . . 1104
44.3 Other Governmental Regulations andRegulatory Agencies . . . . . . . . . . . . . . . . . . . 1104
44.3.1 State Departments of Health . . . . . . . . . . . . 1105
44.3.2 Office of the Inspector General . . . . . . . . . 1105
44.3.3 Centers for Medicare and Medicaid
Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1105
44.3.4 United States Food and Drug
Administration, Office of In Vitro
Diagnostic Device Evaluation and
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1105
44.3.5 Equal Employment Opportunity
Commission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1106
44.3.6 Occupational Safety and Health
Administration . . . . . . . . . . . . . . . . . . . . . . . . . . . 1106
44.3.7 Environmental Protection Agency . . . . . . 1106
44.3.8 Nuclear Regulatory Commission . . . . . . . 1106
44.3.9 Department of Transportation . . . . . . . . . . . 1106
44.3.10 Health Insurance Portability and
Accountability Act . . . . . . . . . . . . . . . . . . . . . . 1107
44.3.11 Ban on Physician Self-referral
(Stark Law) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1107
44.3.12 Three-Day Rule . . . . . . . . . . . . . . . . . . . . . . . . . . 1107
44.4 Nongovernmental Agencies . . . . . . . . . . . 1107
44.4.1 College of American Pathologists . . . . . . 1107
44.4.2 Commission on Office Laboratory
Accreditation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1108
44.4.3 Joint Commission on Accreditation of
Healthcare Organizations . . . . . . . . . . . . . . . 1108
Further Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1108
F.S. Ong, MD
Department of Pathology and Laboratory Medicine,
Cedars-Sinai Medical Center, Los Angeles, CA, USA
W.W. Grody, MD, PhD
Divisions of Medical Genetics and Molecular Pathology,
Departments of Pathology and Laboratory Medicine,
Pediatrics, and Human Genetics, UCLA School of
Medicine, Los Angeles, CA, USA
K. Das, MD (*)
Department of Pathology and Laboratory Medicine,
UCLA David Geffen School of Medicine, Los Angeles,
CA, USA
L. Cheng, D.Y. Zhang, J.N. Eble (eds.), Molecular Genetic Pathology,DOI 10.1007/978-1-4614-4800-6_44, # Springer Science+Business Media New York 2013
1091
44.1 Introduction
• In response to reports of erroneous Pap smear
testing, subsequent public outcry, and skepti-
cism of quality standards in clinical laborato-
ries as a whole, in 1967, the Clinical
Laboratory Improvement Act (CLIA) was
passed, and the underpinnings of the first
consistent set of federal laboratory regulations
were laid, with the following consequences
– CLIA determined standards for all clinical
laboratories
– Legislation provided necessary oversight
of clinical laboratories (molecular diagnos-
tic laboratories comprising a minor numer-
ical subset) to ensure patient safety through
quality laboratory operations
– Clinical laboratories have become among
the most intensely regulated parts of the
entire system of healthcare in the United
States
– It is paramount to understand not only these
federal regulations but also state and local
– Regulations in order to establish and oper-
ate a modern, compliant molecular diag-
nostic laboratory
• The following chapter will provide an over-
view of the various regulatory issues facing
such laboratories. It is important to note that
the regulatory issues presented here are com-
mon to most clinical laboratories; however,
those issues most specific and pertinent to
molecular testing will be highlighted
44.2 Federal Regulation: ClinicalLaboratory Improvement Act
44.2.1 Chronology of ClinicalLaboratory Improvement ActRegulations
• CLIA 1967: CLIA was passed in 1967 in
response to questionable clinical laboratory
practices involving Pap smear testing.
However, several issues were identified with
this Act
– It was most applicable to only large
independent laboratories
– Oversight and enforcement of standards
was inconsistent
– Proficiency testing was found to be fairly
ineffective
– Many smaller laboratories, including
physician-operated laboratories, were
increasing in popularity, though largely
immune to such regulation
• CLIA ’88: In order to address these issues, in
1988, Congress enacted Public Law 100–578,
a revision of Section 353 of the Public Health
Service Act (42 U.S.C. 263a) that amended the
Clinical Laboratory Improvement Act of 1967
(also known as Clinical Laboratory Improve-
ment Amendments of 1988, or CLIA ’88)
requiring the Department of Health and
Human Services (HHS) to establish regula-
tions to ensure quality and reliable clinical
laboratory testing
– In 1990, rules were proposed
– In 1992, HHS published the final rules
(with comment), establishing CLIA ’88
regulations that describe requirements for
all laboratories performing clinical testing
– Since the 1992 final rules, there have been
numerous notices and regulations
published, revising CLIA ’88
44.2.2 LaboratoriesRegulated by ClinicalLaboratory Improvement Act
• There are two types of laboratories regulated
by CLIA, those meeting either definition listed
under 42 Code of Federal Regulations (CFR),
Section 493.2– . . .a facility for the biological, microbiological,
serological, chemical, immunohematological,
hematological, biophysical, cytological, patho-
logical, or other examination of materials
derived from the human body for the purpose
of providing information for the diagnosis, pre-
vention, or treatment of any disease or impair-
ment of, or the assessment of the health of,
human beings. These examinations also include
procedures to determine, measure, or otherwise
describe the presence or absence of various
substances or organisms in the body.
1092 F.S. Ong et al.
– Laboratories seeking payment under the
Medicare and Medicaid Programs
• Virtually all clinical laboratories, including
molecular diagnostic labs, will fit into at least
one, if not both categories, and therefore fall
under CLIA regulation
• Includes any laboratory falling under federal
jurisdiction (although regulations may be
modified for these laboratories)
• Notable exceptions to CLIA regulation are
– Laboratories performing only forensic
testing, including molecular testing labora-
tories performing only forensic DNA
analysis
– Laboratories testing human specimens that
do not meet the aforementioned criteria in
42 CFR, Section 493.2; typically academic
or private sector research laboratories,
many of which employ molecular genetic
methodologies
– Laboratories certified by Substance Abuse
andMental Health Services Administration
(SAMHSA), typically providing drug test-
ing under SAMHSA regulations
• However, all other testing performed by
such laboratories is subject to CLIA reg-
ulation, as necessary
• With regard to molecular testing, any
pharmacogen(etic/omic) testing, if used
for diagnostic purposes and/or seeking
Medicare/Medicaid reimbursement for
such testing, may be subject to CLIA
regulation
44.2.3 Categories of Testing
• CLIA regulations recognize three categories
of laboratory testing (waived, moderate
complexity, and high complexity) and
enforce regulations specific to these
categories
44.2.3.1 Waived Testing• This category consists of tests simple enough
that they require minimal regulatory oversight
by CLIA. According to 42 CFR Sec. 493.15,
these tests are either
– “Cleared by FDA for home use” or
– “Employ methodologies that are so simple
and accurate as to render the likelihood of
erroneous results negligible” or
– “Pose no reasonable risk of harm to the
patient if the test is performed incorrectly”
• Common examples of such testing include
many urine pregnancy tests (visual interpreta-
tion) and blood glucose monitoring devices
cleared by the FDA for home use
– The list of waived testing is determined and
maintained by HHS
– At the time of this writing, there are no
waived molecular genetic tests available,
although methodologies exist that may
make such testing possible
44.2.3.2 Moderate and High ComplexityTesting
• All other testing is divided into tests of moderate
and high complexity, according to grading
based on seven criteria (see 42CFRSec. 493.17)
– Knowledge required to perform test
– Training and experience required to per-
form test
– Reagents and material preparation
– Characteristics of operational steps
– Calibration, quality control, and profi-
ciency testing materials
– Test system troubleshooting and equip-
ment maintenance
– Interpretation and judgment
• A subcategory of moderate complexity testing
is known as provider-performed microscopy
procedures (PPM). To be considered PPM
procedure, test must meet several criteria
(42 CFR Sec. 493.19)
– Performed by either a physician or other
professional healthcare provider, on own
patient or patient of group practice or on
specimen obtained during visit
– Procedure must be moderate complexity
– Test is performed with microscope (bright
field or phase contrast)
– Specimen is labile, or delay in performing
test could compromise accuracy
– Control materials not available to monitor
entire testing process
44 Molecular Testing: Regulatory Issues 1093
– Limited specimen handling/processing
required
– Testing limited to list determined by
HHS (examples include direct wet
mount microbiology procedures, urine
sediment examination, qualitative semen
analysis)
• At the time of this writing, only 12 tests listed
as PPM procedures, with no molecular diag-
nostic assay listed; however, above criteria do
not preclude PPM approval for future molec-
ular diagnostic testing
44.2.3.3 Determination of TestingComplexity
• Testing manufacturers submit supporting data
to the US Food and Drug Administration
(FDA) for determination based upon above
criteria
• For noncommercial systems, laboratory or
group involved submits such application for
determination by Centers for Disease Control
and Prevention (CDC)
• If a test is not listed in the Federal Register
notices, it is automatically categorized highcomplexity, until applicable determination
process completed
– All molecular diagnostic testing at the time
of this writing categorized as moderate or
high complexity
– Nature of molecular diagnostic testing, as
an evolution of basic/translational research,
generally obviates new testing be catego-
rized high complexity
• A clinical laboratory may perform any of three
types of testing, or any combination thereof, if
CLIA certificate permits (see below)
44.2.4 Clinical LaboratoryImprovement Act Certification
• Any laboratory regulated by CLIA must have
current (unrevoked or unsuspended) CLIA
certificate
• There are five types of CLIA certificates
– Certificate of waiver
– Certificate for PPM procedures
– Certificate of registration
– Certificate of compliance
– Certificate of accreditation
• Applications for CLIA certificate submitted
on behalf of every laboratory location, with
following exceptions
– If laboratory is not at fixed location (such as
a mobile lab or temporary facility), single
application may be filed on behalf of pri-
mary site
– Nonprofit, federal, state, or local govern-
ment laboratories offering limited menu of
moderate complexity and/or waived tests
(�15 total tests), for public health pur-
poses, can file one application
– Multiple laboratories within hospital sys-
tem, located in contiguous buildings on
same campus or same physical location/
street address, under common direction
can file single application
44.2.4.1 Certificate of Waiver• Laboratories performing only waived testing
must carry such certificate (as noted earlier, no
molecular diagnostic tests currently catego-
rized as waived)
• However, when waived molecular diagnostic
testing does become available, a laboratory
operating under this certificate must notify
HHS (or designee) before
– Performing and reporting any nonwaived
testing
– Within 30 days of any change in owner,
name, location, or director
• Certificate of waiver valid for maximum of
2 years
44.2.4.2 Certificate for Provider-Performed MicroscopyProcedures
• Certificate for PPM procedures required for
– Laboratories performing PPM procedures
– Certificate of waiver laboratories intending
to perform PPM procedures
• As already noted, currently no molecular diag-
nostic testing falls under category of PPM
procedures
1094 F.S. Ong et al.
• Laboratories must notify HHS (or designee)
before
– Performing and reporting results for
moderate and/or high complexity testing
– Within 30 days of any change in owner,
name, location, or director
• Certificate for PPM procedures valid for max-
imum of 2 years
44.2.4.3 Certificate of Registration• Required of
– Laboratories performing moderate and/or
high complexity testing
– Laboratories issued certificate of waiver or
certificate for PPM procedures intending to
perform moderate and/or high complexity
testing
• Pending inspection by HHS to demonstrate
compliance with regulations governing mod-
erate and/or high complexity testing
• Certificate of registration valid for maximum
of 2 years (or until inspection conducted,
whichever is shorter)
• Laboratory must notify HHS (or designee)
within 30 days of any change in owner,
name, location, director, or technical supervi-
sor (high complexity testing only)
44.2.4.4 Certificate of Compliance• Certificate of compliance may be issued after
successful HHS inspection of laboratory,
authorizing waived, PPM procedures, moder-
ate complexity, high complexity, or any com-
bination of testing thereof
• Laboratories operating under this certificate
must notify HHS (or designee) within
30 days of any changes in owner, name, loca-
tion, director, or technical supervisor (high
complexity testing only) and
– Within 6 months after performing testing
not included on certificate
– Within 6 months after any deletions or
change in test methodology for any test on
certificate
• Certificate of compliance valid for maximum
of 2 years
44.2.4.5 Certificate of Accreditation• To perform waived, PPM procedures, moder-
ate complexity, high complexity, or any com-
bination of testing thereof, a laboratory can
pursue a certificate of accreditation (in lieu of
or in addition to certificate of compliance)
• Lab must secure a certificate of registration
(valid for a period of no more than 2 years),
carrying valid certificate of compliance,
and then
– Become accredited by private nonprofit
entity approved by HHS for this purpose
– Provide HHS with proof of accreditation
(within 11 months of receiving certificate
of registration or, if certificate of compli-
ance is held, prior to expiration of that
certificate)
• Lab must notify HHS and HHS-approved
accreditation program within 30 days of
change in owner, name, location, director, or
supervisor (high complexity testing only) and
– Notify the HHS-approved accreditation
program no later than 6 months after dele-
tion or change of testing methodologies
– Notify HHS-approved accreditation pro-
gram no later than 6 months after
performing testing for which laboratory is
not accredited (for reevaluation of
compliance)
– Maintain HHS-approved accreditation,
including allowing accreditation program
to release relevant data to HHS for scrutiny,
and allow HHS validations and inspections
as well
• Certificate of accreditation is valid for nomore
than 2 years
44.2.4.6 Department of Health andHuman Services-ApprovedLaboratory AccreditationPrograms and ClinicalLaboratory Improvement ActExemption
• The Centers for Medicare and Medicaid Ser-
vices (CMS) oversee the CLIA program on
behalf of HHS. The aforementioned certificate
of accreditation process may be thought of as
a CMS-approved laboratory accreditation
44 Molecular Testing: Regulatory Issues 1095
program “deeming” a laboratory compliant
with relevant CLIA regulations
– Such CMS-approved nonprofit laboratory
accreditation organizations are described
as having CLIA “deeming authority”
• Laboratory can also be exempted from CLIA
requirements by CMS, if licensed or approved
in certain states. These states are described as
having been granted an “exemption” that
allows licensed/approved laboratories in
those states exemption from CLIA require-
ments, as long as certain criteria are met
• For private nonprofit laboratory accreditation
organizations to be granted “deeming author-
ity” or state licensure/approval programs to be
granted “exemptions,” CMS must approve
their respective regulations to be at least as
stringent as CLIA regulations. Such authority
given by CMS cannot exceed 6 years. A list of
such organizations made publically available
by CMS. At the time of this writing, the fol-
lowing states have exempt status
– New York
– State of Washington
• The following private nonprofit accreditation
organizations have “deeming authority”
– AABB (formerly the American Associa-
tion of Blood Banks)
– American Osteopathic Association
– American Society for Histocompatibility
and Immunogenetics
– College of American Pathologists (CAP)
– Joint Commission on Accreditation of
Healthcare Organizations (JCAHO)
– COLA (Commission on Office Laboratory
Accreditation)
44.2.5 Proficiency Testing
• Every laboratory conducting nonwaived test-
ing must engage in proficiency testing (PT)
program approved by CMS for each CLIA
regulated specialty/subspecialty and each
CLIA regulated analyte/test for which it is
certified or seeks certification
• The following are specialties/subspecialties
regulated by CLIA, which encompass
analytes/tests regulated by CLIA and thus
have specific CMS-prescribed PT
requirements
– Microbiology (specialty): bacteriology,
mycobacteriology, mycology, parasitol-
ogy, and virology
– Diagnostic immunology: syphilis serology
and general immunology
– Chemistry: routine chemistry, endocrinol-
ogy, and toxicology
– Hematology: no subspecialties
– Pathology: cytology (limited to gyneco-
logic examinations)
– Immunohematology: ABO group and
D (Rho) typing, unexpected antibody
detection, compatibility testing, and anti-
body identification
• Molecular testing is not recognized as spe-
cialty/subspecialty by CLIA; therefore, there
are no specific CLIA requirements for molec-
ular PT; however
– Laboratories performing nonwaived testing
not regulated by CLIAmust adhere to alter-
native performance assessment for such
tests/analytes, including molecular testing
– Minimum guidelines require alternative
assessment twice per year for each such
assay
44.2.5.1 Proficiency Testing Guidelines• Proficiency testing samples must be treated in
same manner as patient specimens
– Within regular clinical patient workload
– By routine testing personnel
– Using routine laboratory methods
– Without communicating with any other
laboratory regarding PT samples (until
after PT event due date)
– If laboratory has multiple sites performing
test, no communication is allowed regard-
ing PT results (until after PT event due date)
– Must not send samples or aliquots to any
other laboratory for analysis that laboratory
is certified to conduct (violation can result
in revocation of certificate)
• Any laboratory receiving PT samples
from another laboratory must report to
CMS
1096 F.S. Ong et al.
– All records must be kept for at least 2 years
from PT event
• If laboratory has certificate suspended or
limited due to PT performance failure on
at least one specialty/subspecialty and
analyte/test
– Must perform satisfactorily on two consec-
utive PT events before certificate reinstate-
ment and Medicare/Medicaid approval
addressed
– Suspension/limitation of certificate and/or
Medicare/Medicaid penalties are at least
6 months in duration
44.2.5.2 Alternative PerformanceAssessment
• If specialty/subspecialty or analyte/test is not
listed within CLIA guidelines, then alternative
performance assessment must be performed in
lieu of CMS-approved PT programs. This is
the case for all molecular testing at the time of
this writing
– Recommended to participate in other PT
programs, whenever available
– Programs available for limited number of
molecular genetic analytes (CAP PT
programs)
• Molecular PT samples often do not chal-
lenge entire testing process (i.e., sample
may be purified DNA; therefore, it
does not interrogate nucleic acid
extraction)
• Rare genetic disorders present further
challenge: adequate samples may be dif-
ficult to procure; few laboratories will
perform such testing
• Recommended strategies for alternative per-
formance assessment in molecular testing
– Because external quality assessments avoid
bias of internally based assessments,
recommend
• Available PT programs
• Interlaboratory exchange
• Acquire externally derived PT materials
– However, if such options are unavailable
• Repeat testing of blinded (internal)
samples
• Blinded testing of known external
samples
• Sample exchange with research facility
or foreign laboratory
• Split samples for method comparison
with different instrument or method
• Interlaboratory data comparison
44.2.6 Facilities
• Laboratories performing nonwaived testing
must meet certain criteria for facility quality
– Adequate space, ventilation, and utilities
– Procedures and controls to minimize
contamination of specimens, equipment,
and reagents
• For molecular testing
– This includes unidirectional workflow
for nucleic acid amplification
procedures (not contained in closed
systems)
– Separate areas for specimen prepara-
tion, reagent preparation, amplifica-
tion, and detection
– Appropriate equipment, reagents, and
supplies
– Compliance with applicable federal, state,
local regulations
– Adequate and accessible safety procedures
to protect from physical, chemical,
biochemical, electrical hazards, and
biohazards
– Records, slides, blocks, and tissues stored
under proper conditions for preservation
– Test requisitions and authorizations
retained for at least 2 years
– Test procedures retained for at least 2 years
after procedure discontinued
– Retain quality control and patient result
records for at least 2 years
– Retain analytical system performance
records for lifetime of use (no less than
2 years total)
– Retain proficiency testing records for at
least 2 years
– Retain quality assurance records for at least
2 years
44 Molecular Testing: Regulatory Issues 1097
– Retain (or be able to retrieve) test reports
for at least 2 years after reporting (pathol-
ogy reports for at least 10 years)
– If slides/blocks/tissues are retained by the
molecular diagnostic laboratory, cytology
slides must be retained for at least 5 years
from examination, histopathology slides
for at least 10 years from examination,
blocks for at least 2 years from examina-
tion, and tissue until a diagnosis is made
• Note that several professional practice
guidelines recommend, although not
required by CLIA, that molecular
testing specimens should be retained as
long as they are stable, and at least until
next PT event, to allow retesting of
specimens in the event assay error is
identified
– If laboratory closes operation, all retained
samples and records must be maintained
for the required time frames listed
44.2.7 Quality Control and Assurance
• Laboratories conducting nonwaived testing
must develop and document quality control
and assurance procedures that monitor the
preanalytic, analytic, and postanalytic phases
of testing, in addition to general laboratory
practices, and provide for continuous quality
improvement
44.2.7.1 General Laboratory Quality• The laboratory must
– Ensure confidentiality of patient
information
– Ensure specimen identification and
integrity
– Document problems and complaints and
necessary investigations
– Identify communication issues between
laboratory and end users
– Develop policies on personnel competency
– Identify tests/analytes not evaluated by a
CMS-approved PT program (at least twice
per year) or for which CMS-approved PT
program is unavailable
– Document and review any problems, cor-
rective actions, effectiveness of corrective
actions, and any necessary revisions to
policies/procedures
• All of the above guidelines are useful for any
clinical laboratory, including those
conducting molecular testing
44.2.7.2 Preanalytic Phase• All laboratories performing nonwaived testing
must
– Receive written/electronic request for
patient testing (from individual authorized
to request such testing)
• Laboratory may accept verbal test
requests if written/electronic authoriza-
tion received within 30 days (or docu-
mentation of efforts to obtain)
• Test requisition must list
– Name and other suitable identifiers of
requestor, including (as necessary)
contact person for reporting alert
value(s)
– Patient name or unique patient
identifier
– Sex and age or date of birth of patient
– Test(s) to be performed
– Source, as necessary
– Date (and time, if necessary) of
specimen collection
– Additional necessary information for
interpretation
– Note that patient chart or medical
record may suffice as test requisition,
provided it is available to laboratory at
time of test, and CMS (upon request)
– Information transcribed accurately
– Establish specimen submission, handling,
and referral policies
– Establish and maintain preanalytic quality
control and assurance procedures
• For molecular testing, the majority of labora-
tory errors occur in preanalytic phase
– Major culprit thought to be inappropriate
test ordering, likely resulting from inade-
quate information of those ordering testing
(including clerks, nursing staff, and
clinicians)
1098 F.S. Ong et al.
• Users may not understand limitations/
specifications of test, risking
misinterpretation
• Users may incorrectly order test, due to
inappropriate indication or method
– Therefore, although not required by CLIA,
it is common professional practice that lab-
oratory convey relevant test information
clearly and accurately to users, including
• Intended use of test
• Indications for testing
• Test method
• Analytical and clinical specifications
• Limitations of test
– Also, although not required by CLIA, it is
common professional practice for labora-
tory to request information relevant to test
analysis and interpretation, particularly for
molecular genetic testing, for example
• Gender
• Ethnicity
• Selected clinical/laboratory information
• Pedigree or family history
• Informed consent and/or pretest genetic
counseling (recommended prior to
testing for many heritable conditions,
required by law in many states)
44.2.7.3 Analytic Phase• All laboratories conducting nonwaived testing
must
– Maintain a written procedure manual for
each assay performed
– Select appropriate test systems and estab-
lish and maintain criteria for proper opera-
tion of the test systems
– Establish criteria and maintain proper stor-
age of reagents and specimens
• Error rates in analytic phase of testing for
molecular assays reported to range between
0.06% and 0.12% (comparable to nongenetic
testing)
– Errors reportedly encompass all phases of
analytic process, including specimen
handling
– For instance, mutations or polymorphisms
leading to false-positive or false-negative
results have been reported for some
common molecular genetic tests (such
as cystic fibrosis mutations and HFE-associated hereditary hemochromatosis
testing)
– Such interferences are difficult to predict or
evaluate during method evaluation, as they
are specific by definition from specimen to
specimen, and exert effects on assay spe-
cific to each variant
Analytical Test Validation or Verification
• All laboratories must establish or verify per-
formance specifications for test systems
implemented after April 24, 2003
– To verify and document performance spec-
ifications of unmodified, FDA-cleared/
approved test systems, a laboratory must
demonstrate comparable performance
specifications to those obtained by the man-
ufacturer for the following parameters
• Accuracy
• Precision
• Reportable range
• Reference intervals – verify that manu-
facturer’s stated intervals appropriate
for laboratory-specific patient
population
– To establish and document performance
specifications (for a modified FDA-
cleared/approved test; or test system not
subject to FDA clearance or approval,
such as a laboratory-developed test, or test
system without provided specifications),
the laboratory must determine the follow-
ing parameters, as applicable
• Accuracy
– Note that for a molecular test evalu-
ating multiple targets, all variants
should be included
– For rare variants alternative samples
such as cell lines or synthetic nucleic
acid constructs may be necessary
• Precision
• Analytical sensitivity
• Analytical specificity (including
interfering substances)
– Note that in addition to common
exogenous or endogenous
44 Molecular Testing: Regulatory Issues 1099
interfering substances, other
alleles, homologous sequences,
and contaminants may need to be
assessed on a case-by-case basis
for positive/negative interference
• Reportable range
• Reference intervals; for molecular
genetic testing, this may consti-
tute the reference sequence
• Any other performance character-
istic required for test performance
(such as clinical sensitivity/
specificity)
• Test system maintenance must be performed
and documented at least according to manu-
facturer’s recommendations
– If the test system is laboratory-developed
or such protocols are not available, the lab-
oratory must establish such maintenance
protocols to ensure reliable test results
Calibration Procedures
• Calibration and calibration verification are
necessary to ensure and document accuracy
of testing and should interrogate the entire
reportable range of the test system. The labo-
ratory must
– Follow the manufacturer’s recommenda-
tions using provided materials (if possible)
or
– Use laboratory-established acceptance
criteria and frequency (based on method
verification or establishment) using trace-
able (to reference method or material) cal-
ibration materials
• Include at least a minimal, a midpoint,
and maximum value near upper limit of
the reportable range
– Must occur at least every 6 months and
whenever any of the following occur
• Complete change of reagents
• Major preventive maintenance or repair
of test system
• Control materials detect potential fail-
ure mode, and other measures have not
identified a source
• More frequent calibration verification
required
Quality Control Procedures
• Each test system must utilize control proce-
dures that monitor entire analytical process, to
detect real-time failure modes, as well as
developing trends that may signal future fail-
ure. These control procedures should be devel-
oped in accordance with each test’s
verification or establishment of performance
characteristics. Quality control (QC) should
also determine long-term accuracy and
precision
• QCs should be performed
– In same manner as patient specimens
– According to specifications verified or
established in laboratory
– At least once each day, patient testing is
performed
• For quantitative testing, two control
materials of different concentrations
• For qualitative testing, a negative and
positive control
– For testing involving extractions, at least
one control should interrogate the extrac-
tion procedure
– After complete change of reagents
– After major preventive maintenance or
repair
– With different lot of calibrator material, if
calibrators used as control material
– All control procedures should be
documented
• Specific QC issues for molecular testing
– For molecular testing involving
amplification
• At least two controls are required
• If reaction inhibition is of concern,
include control to sensitive to inhibition
(spike-in, or endogenous)
– Although not required by CLIA, it is com-
mon professional practice
• To minimize or detect contamination by
nucleic acid target or amplicon
– At least one no template control
(NTC) sample should be tested,
starting with amplification step
– Preferable to test NTC through entire
analytical process (i.e., starting with
extraction step)
1100 F.S. Ong et al.
• If multiple genotypes to be assayed,
rotate positive controls over time, to
avoid excessive number of controls
per run
• For rare variants with no natural
controls available, use artificial or cell
line controls
• For sequencing assays, bidirectional
sequencing can be confirmatory
• If using pre-extracted DNA control
(thus not controlling for extraction),
can assay surrogate for DNA extraction
in each sample such as exogenous spike-
in control or internal control gene(s)
• If laboratory develops nucleic acid con-
trols, must avoid cross-contamination of
patient specimens and reagents
Comparison of Test Results
• Between laboratory methods: laboratories
performing same test using different methods
or same method at multiple sites must perform
comparison testing twice per year to establish
and document relationship
• Between results: laboratories must identify
potential testing inconsistencies, based on
criteria such as patient age, sex, diagnosis,
and/or other test parameters or previous results
44.2.7.4 Postanalytic Phase• All laboratories that perform nonwaived test-
ing must monitor their postanalytic systems,
including
– Test reporting
• Accuracy and timeliness
• Reports must indicate patient identifica-
tion, name and address of laboratory
location, date reported, date performed,
specimen source, and result and allow
access to reference intervals
• Reports can only be released to autho-
rized persons, person responsible for
using test result (if applicable), and lab-
oratory that requested test
• Theremust be amechanism for reporting
alert values to responsible entities
• Reports must be maintained and readily
accessible
• The laboratorymust have a procedure formon-
itoring, assessing, and correcting postanalytic
systems and documenting these activities
Postanalytic Issues Specific to Molecular
Testing
• Most frequently encountered error is misinter-
pretation of test results by healthcare pro-
viders, often due to misunderstanding of
testing limitations
– As mentioned earlier, this can be remedied
in preanalytical arena
• Clinical validity and utility difficult to assess
because of rapid evolution of testing
– Validity challenging with regard to
genotype–phenotype correlations because
novel/emerging targets are less character-
ized, in addition to the presence of variants
of uncertain significance
– Clinical utility difficult to determine as
clinical impact of emerging diagnostics
often requires more time and sample
numbers, as opposed to analytical valida-
tion of assays
• Because of the implications of test results
from molecular testing, laboratories should
consider including additional information not
required by CLIA, in clear, concise, and infor-
mative fashion, including
– Test indication
– Test method
– Test specifications and limitations
– Current standard nomenclature, as well as
commonly used terms (to avoid confusion
regarding results)
– Detailed interpretation
– Genetic counseling recommendation, when
applicable
– Implications of test results for family
members
– Updated or amended reports as new infor-
mation becomes available
• Previously reported deleterious muta-
tions may be reclassified as benign
variants/polymorphisms, or vice versa
• Laboratories may consider placing
a system to amend reports and contact
patients and providers, in the event of
44 Molecular Testing: Regulatory Issues 1101
such changes, and to maintain
a database of variants to identify those
patients more readily
– Retain reports for at least 25 years, as
opposed to minimal CLIA guidelines
• Molecular genetic diagnoses are essen-
tially a lifetime diagnosis that may need
to be referenced and also may affect
future relatives of the proband; also,
new knowledge may surface that will
alter original diagnosis
Release of Molecular Testing Reports
• As required by CLIA, laboratories may
only release test results to authorized persons,
the person responsible for using the test
results (if applicable), and the laboratory
that initially requested the test (42 CFR
493.1291)
– Note that molecular genetic test results
may be necessary for the healthcare of fam-
ily members of patients
• Healthcare provider of family member
would need to receive authorization
from patient to request patient test
information
• Laboratory should request and confirm
patient’s authorization before releasing
results
44.2.8 Personnel
• Laboratories performing nonwaived testing
are regulated with respect to all personnel
involved in testing
44.2.8.1 Provider-PerformedMicroscopy Procedures
• The laboratory must have
– Qualified director (42 CFR Sec. 493.1357)
• Qualified to manage and direct the per-
sonnel and perform PPM procedures
• Carrying current license as laboratory
director issued by state, if required
• Either a physician, midlevel practitioner
(authorized by the state to practice inde-
pendently), or a dentist
• Who directs no more than five
laboratories
– Qualified testing personnel (42 CFR
Sec. 493.1363)
• Who possess a current license issued by
the state, if required
• Either a physician, midlevel practitioner
(authorized by the state to practice inde-
pendently), or a dentist
44.2.8.2 Moderate Complexity Testing• The laboratory must have
– A qualified director (42 CFR
Sec. 493.1405)
• Responsible for overall operation and
administration of laboratory, including
employment of competent personnel
• Accessible to laboratory for onsite, tele-
phone, or electronic consultation
• Who may direct no more than five
laboratories
– A qualified technical consultant (42 CFR
Sec. 493.1411)
• Responsible for technical and scientific
oversight of laboratory
• Available to provide onsite, telephone,
or electronic consultation
– A qualified clinical consultant (42 CFR
Sec. 493.1417)
• Who provides consultation regarding
appropriateness of testing and interpre-
tation of results
• Available to provide clinical consulta-
tion to clients
• Qualified testing personnel (42 CFR
Sec. 493.1423)
• Responsible for specimen processing,
test performance, and reporting results
44.2.8.3 High Complexity Testing• Laboratories performing high complexity test-
ing must have
– A qualified director (42 CFR
Sec. 493.1443)
• Responsible for overall operation of lab-
oratory, including employment of com-
petent personnel
1102 F.S. Ong et al.
• If qualified, may perform duties of the
technical supervisor, clinical consultant,
general supervisor, and/or testing per-
sonnel or delegate these responsibilities
• Must be accessible to laboratory to pro-
vide onsite, telephone, or electronic
consultation
• May direct no more than five
laboratories
– A qualified technical supervisor (42 CFR
Sec. 493.1449)
• Responsible for the technical and scien-
tific oversight of the laboratory
• Accessible to the laboratory to provide
onsite, telephone, or electronic
consultation
– A qualified clinical consultant (42 CFR
Sec. 493.1455)
• Provides consultation regarding appro-
priateness of testing ordered and
interpretation of results
• Available to provide consultation to
clients
– A qualified general supervisor (42 CFR
Sec. 493.1461)
• Responsible for day-to-day supervision
or oversight of the laboratory operation
and personnel
• Accessible to testing personnel at all
times; testing is performed to provide
onsite, telephone, or electronic
consultation
– Qualified testing personnel (42 CFR
Sec. 493.1489)
• Responsible for specimen processing,
test performance, and results reporting
44.2.9 Inspection
• Basic inspections: all laboratories issued
a CLIA certificate and CLIA-exempt labora-
tories must permit CMS or a CMS agent to
conduct validation and/or compliance inspec-
tions, which may include
– Testing of samples or performing
procedures
– Interviews of personnel
– Observation of testing
– Examination of all records
44.2.9.1 Laboratories Issueda Certificate of Waiver orProvider-PerformedMicroscopy
• Not subject to biennial inspections
• However, CMS or a CMS agent may conduct
an inspection at any time to
– Determine if laboratory constitutes “immi-
nent and serious risk” to public health
– Evaluate complaints
– Determine whether laboratory performing
tests outside scope of certificate
– Collect information regarding appropriate-
ness of waived or PPM procedures
44.2.9.2 Laboratories Issueda Certificate of Compliance
• Laboratory issued certificate of registration
must permit initial inspection to assess labo-
ratory compliance with applicable CLIA
regulations
• CMS or a CMS agent may conduct subsequent
inspections on biennial basis (or with other
frequency, based on compliance history)
44.2.9.3 Inspection of ClinicalLaboratory Improvement Act-Exempt Laboratories or ThoseRequesting or Issueda Certificate of Accreditation
• CMS or a CMS agent may conduct validation
inspection of any accredited or CLIA-exempt
laboratory at any time
– CMS or a CMS agent may conduct com-
plaint inspection of CLIA-exempt labora-
tory (or one requesting or issued certificate
of accreditation) at any time upon receiving
complaint
– If validation or complaint inspection results
in noncompliance
• Laboratory-issued certificate of accred-
itation is subject to full review
• CLIA-exempt laboratory is subject to
appropriate enforcement actions under
state licensure program
44 Molecular Testing: Regulatory Issues 1103
44.2.10 Enforcement of Regulations
• The Clinical Laboratory Improvement Act of
1967, amended by CLIA ’88, set forth require-
ments for all laboratories performing diagnos-
tic testing on human specimens and
– Required federal certification of such
laboratories
– Granted the Secretary of HHS broad
authority for enforcement, allowing
• Intermediate sanctions
• Suspension, limitation, or revocation of
certification
• Civil suit (for laboratory activity posing
“significant hazard” to public health)
• Imprisonment or fine (for conviction of
intentional violation of CLIA
regulations)
44.2.10.1 Basis of Sanctions• Violations of CLIA regulations are detected by
– Deficiencies found by CMS (or its agents)
– Proficiency testing results
• One or more of the following sanctions
imposed upon finding “condition level” viola-
tions (42 CFR Sec. 493.1806 for additional
information), based on one or more following
factors
– Deficiencies pose “immediate jeopardy”
– Nature, incidence, severity, and duration of
deficiencies
– Repeated offenses
– Accuracy, extent, and availability of labo-
ratory records, with regard to deficiency
– Relationship of multiple deficiencies
– Overall compliance history
– Intended outcome of sanctions
– Whether laboratory has taken opportunity
to correct deficiencies
– Recommendation by state agency as to
appropriate sanction(s)
44.2.10.2 Types of Sanctions• Principal sanctions
– Three principal CLIA sanctions
• Suspension
• Limitation
• Revocation of CLIA certificate
• Alternative sanctions
– A directed plan of correction
– State onsite monitoring
– Civil money penalty
– Civil suit
– Criminal sanctions: individual convicted of
intentionally violating CLIA regulation
may be imprisoned or fined
• Additional sanctions
– For laboratories approved for Medicare
reimbursement
• Principal sanction: cancellation of labo-
ratory approval for Medicare
reimbursement
• Alternative sanctions
– Suspension of payment for subset of
tests performed on or after sanction-
ing date
– Suspension of payment for all tests
performed on or after sanctioning
date
• Medicare consequences
– Suspension or revocation of any CLIA cer-
tificate leads to CMS cancellation of labo-
ratory approval for Medicare
reimbursement
– Limitation of any CLIA certificate leads to
CMS limitation of Medicare approval to
specialties/subspecialties still authorized
by certificate
• Medicaid consequences
– Reimbursement under state plans limited if
CLIA certificate and/or state licensure/
approval is limited
44.3 Other GovernmentalRegulations and RegulatoryAgencies
• In addition to the Department of Health and
Human Services (HHS) and its CMS branch
for overseeing CLIA regulations, other
branches of federal and state governments
have regulatory impact on laboratories – and
HHS and its branches (including CMS) have
other regulatory oversight of clinical laborato-
ries, in addition to CLIA regulations.
1104 F.S. Ong et al.
The following are selected federal and state
regulatory agencies that have significant impact
on the modern molecular diagnostic laboratory
44.3.1 State Departments of Health
• As described in Section III under CLIA certi-
fication, various state Departments of Health
exert a spectrum of regulatory influence over
clinical laboratories
• New York state and state of Washington have
been granted “exemption” by CMS
– Clinical laboratories licensed in the state of
New York or Washington is eligible for
CLIA-exempt status
44.3.2 Office of the Inspector General
• The Office of the Inspector General (OIG) is
a division of the Department of Health and
Human Services
• Established in 1976, involved in auditing and
identifying inefficiency, fraud, and abuse of
HHS programs, including Centers for Medi-
care and Medicaid Services
• Impact on clinical laboratories is to identify
fraud and abuse of CMS with respect to clin-
ical laboratory testing, particularly issues of
reimbursement and medical necessity of
testing
– Molecular diagnostics is a rapidly evolving
field, so medical necessity of this type of
testing is often not well disseminated, and
therefore reimbursements are at risk for
denial without a certain amount of
vigilance
44.3.3 Centers for Medicare andMedicaid Services
• A division of HHS, together the CMS are the
largest healthcare entity in the United States
• The Medicare and Medicaid programs were
enacted in 1965, as Title XVIII and Title
XIX of the Social Security Act, targeting
healthcare to individuals aged 65 or older,
impoverished youths lacking parenting
(including their caretakers), the blind, and
those with disabilities
• Currently Medicare primarily covers the
elderly; however, it also provides coverage
for some persons with disability or end-stage
renal disease
• Medicaid primarily covers low-income indi-
viduals through cooperation with state
programs
• The impact of CMS through CLIA has already
been presented; however, CMS also sets reim-
bursement rates for laboratory testing that not
only impacts its own healthcare system but
also is referenced by many other third party
payers
– Because of the reagent and labor costs of
most molecular diagnostic assays, as well
as difficulties with reimbursement of such
leading edge technology and testing, CMS
reimbursement rates are vital to the opera-
tion of the modern molecular diagnostic
laboratory
44.3.4 United States Food and DrugAdministration, Office of InVitro Diagnostic DeviceEvaluation and Safety
• The Food and Drug Administration (FDA) is
a division of HHS, and its Office of In Vitro
Diagnostic Device Evaluation and Safety
(OIVD) was founded in 2002 to consolidate
regulation of in vitro diagnostic devices
(IVDs) and test kits
• The OIVD regulates both in-home and labo-
ratory IVDs
• As noted earlier in Section III, the OIVD also
regulates commercial in vitro diagnostic test-
ing complexity (waived vs. moderate/high
complexity) for CMS
• Of the three divisions (Chemistry and Toxi-
cology Devices, Immunology and Hematol-
ogy Devices, and Microbiology Devices), it
is the Division of Immunology and Hematol-
ogy Devices (DIHD) that regulates the
44 Molecular Testing: Regulatory Issues 1105
majority of IVDs and test kits that would be
encountered in a molecular diagnostic
laboratory
44.3.5 Equal EmploymentOpportunity Commission
• Created by Title VII of the Civil Rights Act of
1964 (amended by Congress in the Equal
Employment Opportunity Act of 1972), the
Equal Employment Opportunity Commission
(EEOC) enforces federal laws prohibiting dis-
crimination against applicants or employees
on the basis of race, religion, sex, age, disabil-
ity, or genetic information
• Most employers with 15+ employees, labor
unions, and employment agencies are covered
by EEOC
• Hiring of personnel for molecular diagnostics
laboratories is subject to the applicable EEOC
regulations
44.3.6 Occupational Safety and HealthAdministration
• A division of the United States Department of
Labor, the Occupational Safety and Health
Administration (OSHA) was created by Con-
gress with the Occupational Safety and Health
Act of 1970
• This administration promotes and enforces
healthy and safe workplace standards to pro-
tect employees
• As a place of employment, molecular diagnos-
tic laboratories must adhere to OSHA stan-
dards to protect their employees, which are
the single greatest asset to any laboratory
44.3.7 Environmental ProtectionAgency
• The Environmental Protection Agency (EPA)
was established in 1970 because of concerns
over pollution and the physical environment
• In order to protect the environment, research,
monitoring, standards, and enforcement were
enacted
• The molecular diagnostic laboratory must dis-
pose of environmentally hazardous laboratory
materials in accordance with EPA standards
– Commonly occurring such materials
include ethidium bromide, formalin, and
xylene
44.3.8 Nuclear RegulatoryCommission
• US Nuclear Regulatory Commission (NRC)
was established in 1974 to create and enforce
federal guidelines ensuring safe use of radio-
active materials (for nonmilitary purposes
only) and protect the public and environment
• NRC provides oversight through licensing,
inspections, and enforcement of regulations
of nuclear products such as nuclear power
plants, nuclear medicines, and other radioac-
tive materials (such as research reagents)
• Although most molecular diagnostic laborato-
ries have replaced radioactive reagents such as
Southern blot probes and radiolabeled nucle-
otides, with fluorescent or chemiluminescent
analogs, if a radioactive material is employed
in the laboratory, its use is governed by NRC
guidelines
44.3.9 Department of Transportation
• The Department of Transportation (DOT) was
established in 1966 to provide efficient, eco-
nomical, and environmentally conscious
national transportation system
• Primary agency in federal government regard-
ing safety, adequacy, and efficiency of trans-
portation system
• DOT is responsible for regulating transport of
biohazard materials (including human
specimens)
• All specimens received and referred by clini-
cal laboratories must adhere to DOT
guidelines
1106 F.S. Ong et al.
44.3.10 Health Insurance Portabilityand Accountability Act
• The Health Insurance Portability and
Accountability Act (HIPAA) of 1996 (Public
Law 104–191), an amendment to the Internal
Revenue Code of 1986, aimed to improve
portability and continuity of health insurance;
decrease inefficiency, fraud, and abuse in
health insurance and healthcare; promote
healthcare savings accounts; improve accessi-
bility to long-term care; and simplify admin-
istration of health insurance
• Onebenefitof thisActwasa reformtohealthcare
information management known as the HIPAA
Privacy Rule, safeguarding patient “protected
health information (PHI)” from unauthorized
written, oral, or electronic dissemination
• This Act is the basis for many current individ-
ual healthcare information standards and must
be followed by all clinical laboratories
44.3.11 Ban on Physician Self-referral(Stark Law)
• A provision in the Omnibus Budget Reconcil-
iation Act of 1989 (OBRA 1989), known as
“Stark I” (named for Congressman Pete Stark,
who sponsored bill), barred physicians from
self-referring Medicare patients for laboratory
services (effective 1992)
• There have been several additional provisions
(known as Stark II and III), now among other
changes governing self-referral for Medicare
and Medicaid patients
• Directors of molecular diagnostic laboratories
must understand these provisions, especially
those who are also clinicians (and thus may
refer testing to a laboratory they direct), in
order to avoid unintentional wrongdoing
44.3.12 Three-Day Rule
• Omnibus Budget Reconciliation Act of 1990,
OBRA 1990, Pub. L. 101–508; Preservation of
Access to Care for Medicare Beneficiaries and
Pension Relief Act of 2010, Pub. L. 111–192,
Section 102
• These laws require that diagnostic services
(including laboratory testing), or other ser-
vices related to an inpatient admission, pro-
vided by the admitting hospital (or an entity
owned/operated by the hospital) for admitted
patient during the 3 days preceding date of
admission is considered part of the inpatient
stay (for admission to a hospital paid under
inpatient prospective payment system,
“IPPS”). The payment window for admission
to a hospital not paid by IPPS (such as psychi-
atric, inpatient rehabilitation, long-term care,
children’s, and cancer hospitals) is 1 day pre-
ceding the date of admission
• These laws state that reimbursement for testing
performed in a molecular diagnostic laboratory
under any of the circumstances listed above
would be included on a DRG system
44.4 Nongovernmental Agencies
• In addition to governmental organizations and
regulations, there are many nongovernmental
agencies that impact molecular diagnostic lab-
oratories. An exhaustive list is beyond the
scope of this writing; however, three signifi-
cant organizations will be introduced
44.4.1 College of AmericanPathologists
• The CAP, a nonprofit organization, was formed
in 1947 to promote pathology education,
research, practice, and service for patients,
physicians, hospitals, and the general public
• As noted earlier, CAP is a CLIA-deemed
authority
• CAP accreditation recognized by JCAHO
• Offers the largest proficiency testing program
in the United States
• Offers a wide range of proficiency testing pro-
grams for molecular diagnostic laboratories,
including genetic and somatic mutation testing
• Offers an accreditation program tailored to
molecular diagnostic laboratories
44 Molecular Testing: Regulatory Issues 1107
44.4.2 Commission on OfficeLaboratory Accreditation
• COLA is a nonprofit organization originally
founded in 1988 to assist physician office lab-
oratories (POLs) comply with CLIA
• COLA laboratory accreditation program
granted CLIA deeming authority in 1993
• COLA accreditation recognized by JHACO
• Expanded program to include accreditation of
hospital and independent laboratories
• Accredits following types of laboratories:
physician office, community hospital, mobile
clinic, Veterans Administration, and Depart-
ment of Defense
• CMS-approved to accredit the following spe-
cialties: chemistry, hematology, microbiol-
ogy, immunology, and immunohematology/
transfusion services
44.4.3 Joint Commission onAccreditation of HealthcareOrganizations
• Founded in 1951, JCAHO is an independent
and nonprofit organization that evaluates and
accredits healthcare organizations
• To maintain accreditation, a healthcare orga-
nization undergoes onsite surveys at least
every 3 years, with laboratories surveyed at
least every 2 years
• In surveying a healthcare organization, JCAHO
typically surveys the laboratory as well
• Has accreditation programs for following
healthcare organizations: hospital, home
care, behavioral health, laboratory, ambula-
tory (including office-based surgery and pri-
mary care medical home), long-term care, and
critical access hospitals
• Has CLIA-deemed authority, and is CMS-
approved to accredit all CLIA specialties
• JCAHO accreditation also satisfies select state
licensure and insurer requirements for
laboratories
Further Reading
Centers for Medicare and Medicaid Services (http://www.
cms.gov/)
Chronology of CLIA regulations (http://wwwn.cdc.gov/
clia/chronol.aspx)
CLIA (http://wwwn.cdc.gov/clia/pdf/PHSA_353.pdf, http://
wwwn.cdc.gov/clia/regs/toc.aspx)
CMS-deemed authorities and Exempt States (http://www.
cms.gov/CLIA/13_Accreditation_Organizations_and
_Exempt_States.asp#TopOfPage)
COLA (http://www.cola.org/)
College of American Pathologists (CAP, http://www.cap.
org/apps/cap.portal)
Environmental Protection Agency (http://www.epa.gov/)
Good Laboratory Practices for molecular genetic testing
for heritable diseases and conditions (http://www.cdc.
gov/mmwr/pdf/rr/rr5806.pdf)
Health Insurance Portability and Accountability Act
(HIPAA, http://www.hhs.gov/ocr/privacy/hipaa/under-
standing/index.html)
Joint Commission on Accreditation of Healthcare Orga-
nizations (JCAHO, http://www.jointcommission.org/)
List of CLIA waived testing (https://www.cms.gov/CLIA/
downloads/waivetbl.pdf)
Listing of provider-performed microscopy procedures
(https://www.cms.gov/CLIA/downloads/ppmp.list.pdf)
New York State Department of Health (http://www.
health.ny.gov/)
Occupational Safety and Health Administration (http://
www.osha.gov/index.html)
Office of Inspector General, United States Department of
Health and Human Services (http://oig.hhs.gov/)
Searchable database of CLIA testing (http://www.accessdata.
fda.gov/scripts/cdrh/cfdocs/cfCLIA/search.cfm)
Stark Rule Phase III (http://www.gpo.gov/fdsys/pkg/FR-
2007-11-15/pdf/07-5655.pdf)
State of Washington Department of Health (http://www.
doh.wa.gov/)
Three Day Rule (https://www.cms.gov/AcuteInpa-
tientPPS/08a_Three_Day_Payment_Window.asp)
United States Department of Transportation (http://www.
dot.gov/)
United States Equal Employment Opportunity Commis-
sion (http://www.eeoc.gov/)
United States Food and Drug Administration (http://www.
fda.gov/)
United States Nuclear Regulatory Commission (http://
www.nrc.gov/)
1108 F.S. Ong et al.