NREL is a national laboratory of the U.S. Department of Energy, Office of Energy Efficiency and Renewable Energy, operated by the Alliance for Sustainable Energy, LLC. Molecular Beam Mass Spectrometry for Analysis of Condensable Gas Components Tar Analysis Workshop 19 th EU Biomass Conference Berlin Daniel Carpenter June 8 th 2011

Upload: others

Post on 03-Sep-2019

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NREL is a national laboratory of the U.S. Department of Energy, Office of Energy Efficiency and Renewable Energy, operated by the Alliance for Sustainable Energy, LLC.

Molecular Beam Mass Spectrometry for

Analysis of Condensable Gas Components

Tar Analysis Workshop

19th EU Biomass Conference

Berlin

Daniel Carpenter

June 8th 2011

Page 2: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Background

MBMS analysis technique developed and

used by NREL for >30 years (originally to

study prompt thermochemical phenomena):

1. Line-of-sight extraction of vapor-phase sample from

high T, ambient P environment into mass

spectrometer (Psource/PStage1 ≥ 104)

2

0.30 mmorifice

1.0 mmskimmer

Stage 1 Stage 2 Stage 3

10-5 Torr 10-7 Torr10-3 Torr

440 L/s 520 L/s 260 L/s

Formation of molecular beam

2. Supersonic expansion,

rapid cooling/rarefaction

preserves sample

without condensation or

reaction

3. Mass analysis provides

instantaneous chemical

fingerprint of sample

4. Useful for tar and alkali

metal vapor analysis

Page 3: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Transportable instrument for process monitoring

•Constructed field-deployable MBMS (~1995)

•Mass spectrometer cart: ~1 m2, 300 kg

•On-board, PC-based sample handling and calibration control (temperature, pressure, flow control, gas & liquid standards)

3

1

dP

P

vent

mass spectrometer

filters liquid standard

dilution gas, calibration gas, scrubbed gas

2

3

4

5 6

7

8

9

= 450°C

1. Sample manifold

2.Flow control valve

3.Orifice plate flow meter

4.Sampling orifice

5.Condenser

6.Coalescing filter

7.Pressure control valve

8.Sample pump

9.Dry test meter

Page 4: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Results and experience

4

m/z

0 50 100 150 200 250 300 350 400

Inte

nsity

(arb

.)

5.0e+7

1.0e+8

1.5e+8

100 150 200 250 300

1.0e+7

2.0e+7

3.0e+7

16 (methane)

40 (argon)

108

110

122

132

144 158

160

7894

108

650 C

m/z

0 50 100 150 200 250 300 350 400In

ten

sity

(a

rb.)

5.0e+7

1.0e+8

1.5e+8

2.0e+8

2.5e+8

100 150 200 250 300

0.0

2.0e+7

4.0e+7

6.0e+7

8.0e+7

16 (methane)

40 (argon)

116

128

152

16678

128

178

178

202

216

228

252

276

290

104142

875 C

Mass spectra observed with MBMS during wood gasification at 650° C and 875°C

•Provides real-time, continuous, and robust process monitoring of hot, untreated process gas

•Near-universal detection (typical sensitivity ~1 ppmv)

•Reproducible and stable with routine maintenance

•Learning curve is steep, complex system

•Quantitation is somewhat cumbersome (requires careful injection of liquid standard for eachspecies of interest and good measurement of wet volumetric flow)

•Compares well with EU Tar Protocol, but estimations of “gravimetric” tar difficult with limited standard set (injection of heavy tar standards into hot oven problematic due to solvent vaporization and capillary plugging)

•Expensive (approx. $300K U.S.)

•Most valuable during plant startup and initial product gas characterization, less valuable for routine analysis

Page 5: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Current status and future work

5

•System works, but can be made smaller, lighter, more energy efficient

with recent equipment advancements (vacuum pumps, electronics)

•Phase-sensitive detection (molecular beam chopper) could be added to

increase sensitivity by >100x

•Commercial MBMS systems are now available (Extrel CMS, Hiden

Analytical) that could easily be outfitted for the field

References:•Evans, R.J., Milne, T.A., “Molecular Characterization of the Pyrolysis of Biomass. 1. Fundamentals” Energy & Fuels 1987, 1 (2), 123-137.

•Carpenter, D.L., Deutch, S.P., French, R.J. “Quantitative Measurement of Biomass Gasifier Tars Using a Molecular-Beam Mass

Spectrometer: Comparison with Traditional Impinger Sampling” Energy & Fuels 2007, 21, 3036-3043.

•Dayton, D.C., French, R.J., Milne, T.A., “Direct Observation of Alkali Vapor Release during Biomass Combustion and Gasification. 1.

Application of Molecular Beam/Mass Spectrometry to Switchgrass Combustion” Energy &Fuels 1995, 9 (5), 855-865.

•Carpenter, D.L., Bain, R.L., Davis, R.E., Dutta, A., Feik, C.J., Gaston, K.R., Jablonski, W., Phillips, S.D., Nimlos, M.R., “Pilot-Scale

Gasification of Corn Stover, Switchgrass, Wheat Straw, and Wood: 1. Parametric Study and Comparison with Literature” Ind. Eng. Chem.

Res. 2010, 49 (4), 1859-1871.

NREL/National Bioenergy Center: http://www.nrel.gov/biomass

DOE Biomass Program: http://www.eere.energy.gov/biomass

Page 6: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

•High-resolution mass spectrometry

•Diode laser spectroscopy

•Laser ablation, REMPI/TOF

•Advanced gas chromatography

Additional syngas analysis methods

under development at NREL

6

Supplemental Slides

Page 7: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

High-resolution mass spectrometry

7

Mass spectrum of scrubbed, wood-derived syngas observed with high-resolution mass spectrometer

m/z

0 20 40 60 80 100

Re

lati

ve

In

ten

sit

y (

%)

0

20

40

60

80

100

•Resolution of 5000 (m/ m) enables “accurate mass” measurements (elemental compositions)

•Mass range: 1-3000 amu

•Sensitivity: 30 pg/ L (ppt)

•Capable of CI and direct insertion probe operation

Limitations:

•Capillary inlet may limit throughput of high-mass compounds

•Magnet stability vs. ambient temperature

•Ion source robustness?

44-CO2

m/z

15.8 16.0 16.2 16.4 16.6 16.8 17.0

m/z

27.96 27.98 28.00 28.02 28.04

15.9877-O+

16.0239-CH4+

16.9957-OH+

16.9957-NH3+ 27.9912-CO+ 28.0024-N2

+

28.0276-C2H4+

JMS-GCmate II (JEOL Ltd., Japan) high-resolution magnetic sector mass spectrometer installed in TCPDU

•Motivation: analysis of very low levels of DOE-targeted syngas impurities, e.g. NH3, HCl, H2S

•Modified inlet system for continuous monitoring of syngas (still working on heated capillary inlet for tars)

•Can resolve at m/z… 16: NH2+/CH4

+/O+

17: OH+/NH3+/13CH4

+

27: 13C2H2+/HCN+/C2H3

+

28: CO+/N2+/C2H4

+

Page 8: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Diode laser spectroscopy

8

•Based on near-IR absorption spectroscopy

• Detection limit: 0.1 ppmv

•Response time ~ 2 sec

•Expandable to multispecies detection at roughly $5000/component

Limitations:

•Susceptible to absorption interferences, especially water vapor

•High resolution spectroscopy unknown for most species

Multipass pattern observed on 2” mirror using 660 nm alignment laser. Effective pathlength~ 49 meters.

DFB-DL Herriott cell

Ramp

generator

oscilloscope

BD

DFB-DL Herriott cell

Ramp

generator

oscilloscope

BD

Photograph showing Herriott cell (top) and schematic of diode laser spectrometer. BD – balanced detector, DFB-DL – distributed feedback diode laser Contact: David Robichaud ([email protected])

Page 9: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Laser ablation, REMPI/TOF

9

355 nm Laser Ablation Beam

He Carrier Gas Under Free Jet Expansion

Linear translator Biomass

vapor

Biomass sample (wood section/leaf/disk)

REMPI: λ = 278 nm (4.46 eV)

• Laboratory unit for fundamental studies

• Highly selective towards lignin pyrolysis products (see below)

•Screening of catalysts for biomass pyrolysis

•Study effects of mineral salts on pyrolysis of lignin

•Classification of lignin form different biomass feedstocks

Sample inlet

Contact: Calvin Mukarakate ([email protected])

Rev. Sci. Instrum. 82, 033104 (2011)

Page 10: Molecular Beam Mass Spectrometry for Analysis of ... · •Phase-sensitive detection (molecular beam chopper) could be added to increase sensitivity by >100x •Commercial MBMS systems

NATIONAL RENEWABLE ENERGY LABORATORY

Advanced gas chromatography

10

Wasson-ECE Instrumentation (Ft. Collins, CO) customized GC system for analysis of hot syngas up to 325°C.

•Customized Agilent 6890N & 7890A gas chromatographs

•Accepts heated sample to 325°C

•Dual FID detectors (hydrocarbons through C20)

•Dual TCD detectors (permanent gases, tracer gases)

•Nitrogen chemiluminescence detector – NCD (9 N compounds, including NH3, HCN, pyrrole, pyridine at 0.05 ppmv)

•Sulfur chemiluminescence detector –SCD (16 S compounds, including H2S, mercaptans, thiophene at 0.1 ppmv)

Limitations:

•Some compounds not detectable at 325°C

•Analysis time – fast analysis: 8 min., detailed analysis: 40 min.

auxiliary isothermal ovens

Isothermal injector ovens

sample oven

heater/valve control

NCD & SCD