Clinical Chemistry SolutionBA-800M
BA-800MClinical Chemistry Solution
Technical Speci�cations
System Function:Throughput: 800 photometric tests per hour
1200 T/H with ISE
Sample Handling: Sample capacity:
Sample volume:
Sample probe:
Total sample capacity is 440 (140 positions on the sample tray and 300 positions on the Sample Delivery Module)25 cooled positions for calibrators and controls1.5 ~35 uL, step at 0.1 uL22 uL for ISELiquid level detection, clot detection and collision protection
Reagent Handling:Reagent tray: 120 positionsReagent volume: 15~300 µL, step at 0.5 µLReagent probe: Liquid level detection, bubble detection and
collision protection
Barcode and Data Transmission:Codabar, ITF (Interleaved Two ofFive), Code128, Code39, UPC/EAN, Code93; Bi-directional LIS interface transmission
Reaction System:Reaction volume: 100~360 µLTemperature: 37˚C with 0.1˚C �uctuation
Optical System:Light source: Photometer: Wavelength:
Absorbance range:
Halogen tungsten lampReversed optics, grating photometry340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm, 800nm 0~3.4Abs(10mm conversion)
ISE Module (Optional):Principle: Indirect K+, Na+ Cl-
Control and Calibration: Calibration mode: Linear (one-point, two-point and multi-point),
Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
Control rules: Westgard multi-rule, Twin plot
Operation Unit:Operation system: Windows® XP Professional/Home SP2 or above
Windows® 7/ Window® 8
1430 lbWeight:BA-800M:
Dimension:BA-800M: 40 in x 47 in x 91 in (D x H x L)
P/N: ENG-BA800M-FDA-210285x10-20160621
Sample tray:
Innovative
Coaxial reagent diskUnique coaxial reagent disks design making
reagent replacement time saving and convenient.
The semi-opened cover ensures operator safety
during operation.
Reagent bubble detectionSystem provides su�cient reagent aspiration
through liquid level and reagent bubble detection
technology.
Dot light sourceThe dual focus of forward lighting is possible via
the dual-diaphragm and dual-lens technology
which creates a high intensity focused light.
Thus, lower the minimum reaction volume and
enhances measuring accuracy.
Water quality monitorThe system employs resistance principle. It
provides premium DI Water quality for assay
analysis and reduces potential contamination.
Cost-e�cient
Large sample capacityTotal sample capacity is 440 (140 positions on the
sample tray and 300 positions on the Sample
Delivery Module). The Sample Delivery Module
automatically delivers samples to the analytical unit.
The large sample capacity coupled with 165
washable permanent cuvettes allows the operator
hours of walk-away time.
Low reagent consumption
Uniquely designed reagent bottles maximize
reagent usage and reduce residual volume.
One key STATDedicated One-Key STAT test button, STAT sample
position, and STAT sample rack provide prompt
response to the highest priority samples.
Continuous reagent loading and unloadingTwo separate buttons independently control each
reagent carousel; it ensures safe and continuous
reagent replacement during testing.
Indirect ISELow sample volume, high ISE analysis throughput, and
cost-elective electrodes. Each electrode supports up to
30,000 total test volume or three months.