microwave photonics || rf and microwave photonics in biomedical applications
TRANSCRIPT
9
RF and Microwave Photonicsin Biomedical Applications
Afshin S. Daryoush
9.1 Introduction
In the last 20 years the field of microwave photonics has evolved due to unique features of
analogue fibre-optic systems and its applications in radio over fibre for telecommunications [1]
and optically controlled phased array antennas [2] for military applications, as has been
discussed in earlier chapters of this book. Recently, microwave photonics techniques have also
been extended to biomedical systems and this chapter presents two distinctive biomedical
imaging applications that employ these techniques. (Optics already lends its application to
laser Doppler anemometry, optical biopsy and optical molecular imaging, and phase micros-
copy.) The first application to be discussed is the design and implementation of optical
hydrophone for calibration of ultrasound transducers for frequencies up to 100MHz,which has
found applications in sub-millimeter wave imaging and therapeutic applications. The second is
the use of broadbandmodulated near infrared (NIR) light waves for quantifying blood flow and
cellular functionality using spectroscopy, which is to be applied to coagulation monitoring and
functional imaging with sub-centimetre spatial resolution using photon density waves. Both
techniques are first discussed in terms of the fundamental physical interaction of lightwaves
with biological tissues and the technical advantages that RF and microwave photonics could
bring to conventional imaging modalities.
9.1.1 Introduction to Optical Hydrophone
Only a decade ago, the highest ultrasound imaging frequencywas of the order of 7MHz. Today,
modern diagnosticmachines, particularly those designed for applications such as dermatology,
ophthalmology and microsurgery, operate at centre frequencies close to 15 or 20MHz [3].
Microwave Photonics: Devices and Applications Edited by Stavros Iezekiel
© 2009 John Wiley & Sons, Ltd. ISBN: 978-0-470-84854-8
Also, a majority of the currently available machines has integrated harmonic imaging
capability that can provide diagnostic images at twice the fundamental frequency and hence
increase the resolution twofold. However, ultrasound probes need to be calibrated in the
measurement bandwidth extending to at least eight times the centre frequency of the imaging
transducer [4, 5]. This frequency limit has been introduced to account for nonlinear propaga-
tion phenomena, which lead to distortion of the pressure–time waveform launched into the
examined tissue. Assuming the centre frequency of the imaging array to be about 12MHz, this
would require the sensitivity of the hydrophone probe to be determined in the vicinity of
100MHz [6] (Figure 9.1).
In order to acquire and reproduce faithfully the pressure–time waveforms, and also to
determine accurately the key acoustic parameters of the characterized acoustic field, the active
element of the probe has to be considerably smaller in comparisonwith the cross sections of the
acoustic beam profiles measured; alternatively, tedious spatial averaging correction models
have to be used [4, 7, 8]. In order to ensure half-wavelength sampling that is needed to eliminate
spatial averaging effects, the pressure-sensitive portion of the probe needs to be about 8 mm to
faithfully reproduce the field at 100MHz (assuming plane waves). Unfortunately, the physical
dimensions of available ultrasound probes typically have diameters of 500 mm [4–6, 9]. The
smallest available probe has a diameter of approximately 80 mm and this is still an order of
magnitude too large for 100MHz measurements.
The development of fibre-optic (FO) probes that are specifically designed to increase
sensitivity to ultrasound waves with a diameter of about 8 microns to minimize the spatial
averaging error up to 100MHz is discussed. In Section 9.2, various principles of sensing
acoustic waves using interaction of acoustic energy with intensity, phase and frequency of
lightwaves in the FOprobes are briefly outlined, and emphasis is given to the intensity detection
technique. Theoretical calculations of the laser source output power and acoustic pressure
amplitude needed to ensure a minimum acceptable signal-to-noise ratio are presented, and an
experimental evaluation of a custom-designed down-tapered gold-coated FO probe, which
meets bandwidths of up to 100MHz, is also explored.
9.1.2 Introduction to Optical Spectroscopy using NIR
Near infrared (NIR) spectroscopy is a new, noninvasive technique to analyse living tissue in
terms of absorption and reduced scattering coefficients, which could provide information about
disease-related functional and structural changes [10–12]. Three main categories of NIR
spectroscopy are time domain, frequency domain and continuous wave (CW) measurement
techniques [13]. The frequency-domain method has attracted interest from the biomedical
Figure 9.1 High-resolution imaging of biological tissues using short ultrasound pulses with a
bandwidth requirement of up to 100MHz
240 Microwave Photonics: Devices and Applications
research field for decades due to its low component cost, ease in separating absorption and
scattering parameters, and potential for real-time imaging.
In frequency-domain photon migration (FDPM) methods, diffuse photon density waves
(PDW)aregeneratedwhen light ismodulatedbyradio frequencysignals,whichpropagateswith
a wavelength of several millimetres to centimetres depending on the modulation frequency.
Amplitudeandphase informationof thediffusephotondensitywavesareused tomap theoptical
absorption and scattering properties of the medium. These optical properties in turn are used to
obtain haemaglobin concentration, blood volume and ‘absolute’ oxygen saturation [14, 15].
There are several advantages to multi-frequency instruments compared with single-fre-
quency instruments. Since most tissues have a layered structure and because photon penetra-
tion depth is less at a higher frequency due to a higher loss (cf. Figure 9.2), by sweeping the
modulation frequency one can have information for all layers in a single measurement. This
approach is very important in clinical measurements, where it is preferable to make a single
measurement and obtain as much information as possible, due to the calibration challenges.
The development of specialized optical systems with modulation capability of up to 3GHz
has also been demonstrated and spectroscopic information is conducted for both solid and
liquid phantoms. Moreover, the accuracy of the broadband extraction process is compared to
the single-frequency extraction for phantom resembling breast tissue, where the results of this
extraction are extended to future clinical imaging.
9.2 Approaches to Fibre-optic Based Acoustic Pressure Sensors
Fibre optics has been used in underwater acoustic sensing for a long time and is now being
extended to biomedical applications. Optical fibres have the following advantages over
Figure 9.2 Concept of diffused photon near-infrared spectroscopy in biological tissue depicting:
(a) optical absorption coefficient of oxygenated and de-oxygenated haemoglobin (Oxy-Hb andDeoxy-Hb)
and water in near-infrared region; (b) impact of modulation frequency on tissue penetration depth and
structure of banana-shaped photon scattering in tissue
RF and Microwave Photonics in Biomedical Applications 241
conventional acoustic sensing/imaging techniques: (i) immunity to electromagnetic interfer-
ence; (ii) small sensing area; (iii) small physical dimensions and low weight; (iv) large
bandwidth,and (v) high resistance to high temperature, corrosion by chemicals and adverse
climatic conditions. This section provides an overview of various fibre-optic sensing schemes
which have been used in acoustic sensing applications.
An electric wave travelling in a medium in the positive z-direction can be expressed by the
basic equation given below:
Eðz; tÞ ¼ E0cosðvt� kzþ u0Þ ð9:1Þwhere E0 is the electric field amplitude in V/m,v is the angular frequency in rad/s, and u0 is theinitial phase in radians.
In fibre-optical sensing, the physical phenomenon being sensed interacts with the fibre and
changes one or more of the above parameters associated with the electromagnetic field in or
around the fibre. Accordingly we can classify acoustic fibre-optic sensors as intensity
modulated sensors, frequency/wavelength modulated sensors and phase modulated sensors.
After a review of the best reported results, the latest ‘hero’ performance results of intensity
detection techniques are discussed by the author in Section 9.3.
9.2.1 Intensity Modulated Sensors
Acoustic pressure induces change in the intensity of light passing through an optical fibre.
These sensors can be reflection type, transmission type or total internal reflection based.
9.2.1.1 Reflection Type
In reflection type sensors, the incident pressure induces a change in the refractive index of the
sensing medium surrounding the fibre (Figure 9.3). This leads to a change in reflectance at the
fibre–water interface. The intensity of reflected light is thus modulated by incident pressure.
Figure 9.3 Reflection type fibre-optic acoustic sensor with intensity detection [16]. Reprinted from J. E.
Parsons, C. A. Cain, J. B. Fowlkes, ‘‘Cost-effective assembly of a basic fiber-optic hydrophone for
measurement of high-amplitude therapeutic ultrasound fields,’’ J. Acoust. Soc. Am., vol. 119, pp.
1432–1440, 2006. (� Copyright 2006, American Institute of Physics)
242 Microwave Photonics: Devices and Applications
This change in intensity is detected using a photodetector [16] where typically a responsivity of
�302 re 1VmPa and sensitivity of 0.9MPa has been reported.
9.2.1.2 Transmission Type
Another technique is the transmission type approach [17], where two single mode fibres are
placed tip-to-tip at a very small separation. An acoustic wave incident on the tip causes motion
of the fibres and hence changes the optical coupling efficiency (Figure 9.4). The transmitted
light intensity thus varies in accordance with the pressure. A modification of the above
approach is the schlieren technique [18], in which the region in between the two fibres is
occupied by a grating structure to enhance system sensitivity by monitoring changes in
coupling efficiency induced by diaphragm movement. A suitable sensitivity of these hydro-
phones is demonstrated for deep-sea applications over a frequency range of 100Hz to 1 kHz.
9.2.1.3 Frustrated Total Internal Reflection
Another modification to the above transmission sensor type includes the use of frustrated total
internal reflection at the edge of two polished fibres [19]. The fibres are angle polished and
placed at an angle such that all the modes undergo total internal reflection at the fibre–air–fibre
interface. Any vertical displacement of one of the fibres due to the applied acoustic wave
violates the total internal reflection condition and hence changes the amount of light coupled to
the other fibre. This sensor type has a bandwidth of 500Hz.
Intensity modulation schemes are simple and are not as affected by phase noise as
interferometric schemes. However, they are subject to high losses of optical energy since
most is lost in reflection or coupling loss. Thus the sensitivity of these systems is lower. External
amplification has to be provided in order to boost sensitivity.
9.2.2 Phase Modulated Sensors
The phase of incident light varies with acoustic pressure and can be detected using interfero-
metric techniques. When two coherent light waves which are shifted apart in phase are
Figure 9.4 Schematic of transmission type intensity modulated fibre-optic acoustic sensor [17]
RF and Microwave Photonics in Biomedical Applications 243
superimposed they undergo constructive and destructive interference to form an interference
pattern. In the interferometric sensing method, two arms of fibre are used and compared. One
arm acts as the reference arm while the other is the sensing arm. Any change in the relative
phase or optical path length of the light from the two arms leads to displacement of the
interference fringes. Michelson�s interferometer, Fabry–Perot interferometers and Mach–-
Zehnder interferometers (Figure 9.5) have been employed in the past to detect acoustic signals.
These phase modulated sensors have been extensively studied and can be classified as either
external or internal interferometric phase sensors.
9.2.2.1 External Interferometric Phase Sensors
Michelson and Mach–Zehnder interferometers have been extensively used as phase sensors.
Application of pressure to the sensing arm leads to changes in the refractive index of the sensing
fibrematerial aswell as changes in the physical dimensions of the fibre. This leads to phase shift
and subsequently fringe displacement. Sensitivity of 92 kPa/fringe displacement has been
reported [20]. This technique is, however, subject to fringe displacement related to tempera-
ture-inducedfibre indexvariation,which is of the order of 10 p.p.m./C in silica basedfibres [21].
This temperature-induced phase change leads to uncertainty in the measured pressure value
and is only useful at frequenciesmuch higher than the rate of thermal fluctuations in the sensing
arm. In such sensors, the fibre in the two arms of the interferometer can be free or wrapped
around a mandrel (cylinder) of suitable material. Previous studies indicate that higher
sensitivity can be obtained by using mandrel fibres [20, 22]. In this case, when a mandrel
of low Young�s modulus is subject to pressure, its length changes thus causing stretching or
Figure 9.5 Acoustic pressure sensor using Mach–Zehnder interferometric detection. Reproduced
from [20]
244 Microwave Photonics: Devices and Applications
compression of the fibre wound around it. This leads to changes in refractive index as well as
changes in the optical length of the fibre as mentioned above. The sensitivity of such structures
depends onmandrel geometry and the properties of themandrel material. Due to the low photo-
elastic coefficient of silica, a long length of fibre is needed for improved sensitivity. This,
however, leads to poor frequency response. As a solution, embedded single-mode fibre acoustic
sensors have been suggested [23], where the fibre is moulded on to the mandrel by using
material of appropriate Young�s modulus and coating thickness, as depicted in Figure 9.6. In
this case, sensitivity in the range of �328 to �338 dB re 1V/mPa has been reported over a
frequency range of 0.75–10 kHz.
It can also be noted that the above phase modulation techniques are subject to random phase
fluctuation due to temperature drifts or environmental conditions and hence are limited in
performance by phase noise. These temperature drifts can be reduced by using a push–pull
configuration and tuneable resonant cavity techniques [24], where a feedback control circuit is
employed tomaintain the stability of the operating point on the interference pattern.However, a
large sensing area of the mandrel fibre sensors leads to poor spatial resolution making them
unsuitable for imaging applications.
9.2.2.2 Internal Interferometric Phase Sensors
In this case the interferometer is embedded within the fibre itself. Two distinct approaches are
discussed next.
Fabry–Perot (FP) InterferometersThe FP interferometer is one type of structure used for internal interferemetric phase sensors.
Previous work includes formation of an FP cavity by using 25 mm thick Parylene film deposited
on the edge of a straight-cleaved fibre from which the final 1 mm of jacket has been
removed [25]. Partially reflectingmirrors of the resonant structure consist of reflective coatings
of aluminum, where experimentally system sensitivity in terms of noise equivalent power
(NEP) of 10 kPa over a 25MHz bandwidth is observed. Fabry–Perot cavity resonators
consisting of 10 mm Parylene film surrounded by gold mirrors have been reported [27]. These
structures are fabricated at the tip of straight-cleaved and tapered fibres. The reflectivity can be
Figure 9.6 Schematic of an optical sensor structure wound on the mandrel hydrophone as a pressure
sensor arm of an interferometer. Reproduced from [22]
RF and Microwave Photonics in Biomedical Applications 245
controlled by the gold-coating thickness. Incident acoustic pressure modulates the optical
length of the Parylene film as well as the reflectivity of the gold creating an interference pattern
(Figure 9.7). The system has a nonflat frequency response of up to 50MHz and extrapolated
sensitivity of 2 kPa.
Multi-layer structuresThese are constructed based on the principle of periodic micro-interferometers as shown in
Figure 9.8 [27]. Two high-reflection subsystems, both consisting of several l/4 layers, are
connected by a central l/2 spacer layer. The sensor is made of 19 dielectric layers with
alternating high–low refractive indices of n¼ 2.3 (Nb2O5) and n¼ 1.48 (SiO2). The sensor
element has an overall thickness of d¼ 1.9 mm. The principle of ultrasound measurement is
based on the elastic deformation of the multi-layer system by an incident acoustic pressure and
the detection of the induced change in optical reflectance, DR. The system is operated at a
maximum sensitivity point (the point where the slope of change of reflectance with pressure is
N+1
(N+1)/2
L L L
0 2 (N+1)/2
Fibre
L L L
nidi 2nidi
I in
Iout
Acoustic pressure
Figure 9.8 Fibre-optic multilayer hydrophone, i¼ 1. . ..N, L¼ low index layer, H¼ high index layer,
p¼ pressure pulse [27]
Figure 9.7 Plane-cleaved fibre-optic sensor based on the principle of the FP interferometric technique.
Reproduced from [26]
246 Microwave Photonics: Devices and Applications
highest). The transfer function of this sensor shows a resonant peak at a frequency of 24MHz,
which is attributed to possible diffraction effects.
In intrinsic FP cavity structures, the operating point has to be stabilized at the maximum
slope point on the interference transfer function. Any drift in this operating point due to
temperature or other factors can lead to increased noise or a decrease in sensitivity. Due to
periodicity, the accurate measurement of differential phase change in phase modulated sensors
is difficult and hence such systems are not suited to most applications.
9.2.3 Frequency/Wavelength Modulated Sensors
Frequency or wavelength sensing mechanisms are an extension of phase sensitive fibre-optic
sensors and a few examples are discussed next.
9.2.3.1 External Bragg Cell Sensor
Ameasuring technique based on acoustically induced frequency modulation of light has been
reported in fibre sensors over a frequency range of 100–1200Hz [28]. This arrangement makes
use of an external Bragg cell which shifts the frequency (wavelength) of the laser source by
11MHz. The unshifted frequency component is incident on the sensing fibre which reacts to
acoustically induced strain and change in refractive index. This induces a change in the phase of
the optical signal and results in the frequency modulation of the signal. This frequency
modulated signal from the fibre sensor and frequency shifted component from the Bragg cell
are combined at the frequency discriminator, the output of which is proportional to the incident
acoustic pressure.
9.2.3.2 Fibre Bragg Grating Sensor
The most common type of wavelength modulated sensor makes use of the fibre Bragg grating
(FBG). Detection is based on an acoustically induced change in Bragg wavelength or intensity
modulation of light from the grating.When an acoustic wave is incident on an optical fibrewith
an FBG, both the refractive index and the FBG grating period undergo changes due to elasto-
optic effects (Figure 9.9) [29]. This change in refractive index in turn causes a shift in the FBG
wavelength resulting in wavelength modulation. The bandwidth of these sensors ranges from
0.1–5MHz.
Figure 9.9 Schematic of a fibre Bragg grating sensor with a required interaction length of at least
600 mm for sufficient sensitivity. Reproduced from [29]
RF and Microwave Photonics in Biomedical Applications 247
9.2.3.3 Distributed Grating Reflector
Another type of sensing technique based on an internal distributed Bragg reflector (DBR) has
been reported recently [31]. Detection of this fibre sensor is based on the modulation principle
of birefringence induced by the incident acoustic pressure in the fibre laser. Two 1550 nm
Er–Yb doped grating structures each of length 10mm and 3mm respectively are written inside
a single mode fibre (Figure 9.10). The separation distance between the two gratings is about
10mm and results in a dual polarized signal. Awavelength divisionmultiplexer, alongwith the
photodetector, monitors beat frequency between both the polarization modes. In the presence
of an acoustic signal, the change in birefringence modulates the beat frequency and additional
sidebands are obtained at the output. The beat frequency serves as an indication of the incident
acoustic pressure, while its amplitude indicates the excitation voltage. The minimum detect-
able pressure level was calculated to be 164 dB re mPa and 158 dB re mPa at 10 and 20MHz,
respectively. Such a sensor can detect ultrasound up to 40MHz though its sensitivity is
relatively poor and its signal-to-noise ratio versus frequency is nonuniform.
These types of sensors are localized and have smaller sensing areas as compared to phase
modulated schemes. They are also insensitive to random amplitude fluctuations. However,
frequency or phase drift due to temperature can affect the performance of such systems. Also,
the sensing aperture dimension is of the order of a fewmillimeters. Hence, they cannot be used
for high spatial resolution measurements and also have limited bandwidth.
9.3 Intensity Sensing Principle, Design and Realization
Intensity-detection based optical hydrophone uses measurement of the Fresnel reflectance
caused by the change in refractive index between the fibre tip and the surrounding fibre
medium. The index of refraction of water depends on the acoustic pressure as reported earlier
and a resultant reflectance versus acoustic pressure is plotted in Figure 9.11 [26]. Sensitivity of
the fibre-optic ultrasound hydrophone probe, S, is calculated from the expression S¼V0/p,
where, V0 is the output voltage of the ultrasound hydrophone probe for a given acoustic
pressure p. With a light source power of 50mW, the theoretical sensitivity of the fibre-optic
probe is calculated to be 4.3mV/MPa for uncoated fibres. Figure 9.12 shows a typical APD
Figure 9.10 Schematic of an optical hydrophone employing a distributed Bragg reflector fibre sensor.
Reproduced from [30] (� 2005 IEEE)
248 Microwave Photonics: Devices and Applications
output power at different acoustic pressure amplitudes and laser source power levels. The
�60 dBm plane corresponds to the calculated system noise floor, where it can be seen that at
100 kPa, to achieve an output signal-to-noise ratio of unity, a light source power level of 65mW
is needed. Also, with the level of 200mW, the minimum detectable acoustic pressure is about
34 kPa.
Figure 9.12 The photodetected output power from an APD as a function of different acoustic pressure
amplitudes and light source power levels as reported [26]. The �60 dBm plateau corresponds to the
calculated system RIN-dominated noise floor
-75
-70
-65
-60
-55
-50
-45
-40
1010.10.01
Acoustic pressure (MPa)
∆R
(d
B)
Figure 9.11 Reflectance change versus acoustic pressure, where a linear relationship is depicted
RF and Microwave Photonics in Biomedical Applications 249
As indicated in this reported analysis the achieved response of this fibre hydrophone sensor is
not competitive with commercial hydrophones, such as the needle hydrophone with a
sensitivity of 40mV/MPa (i.e. �262 dB re V/mPa). System sensitivity improvements are
achieved by coating the fibre tip with gold. Design of optimum gold-coating thickness at the
fibre tip is identified throughmaximization of reflectance sensitivity to acoustic pressure and is
described next.
9.3.1 Transmission Line Modelling of Coated Fibres
A simplified transmission line model was used to improve the understanding of the impact of
fibre coating on the overall sensitivity of the optical hydrophone. Based on this model, the
complex reflection coefficient (r) of coated fibre can be expressed as
r ¼ ndðnc � nwÞþ ðnwnc � n2dÞtanh gdndðnc þ nwÞþ ðnwnc þ n2dÞtanh gd
ð9:2Þ
where d is the coating thickness, nd is the complex refractive index of the coating, nw is the
refractive index of water, and nc is the refractive index of the fibre core. The complex
propagation constant in a lossy gold layer, g, is described in terms of the complex index of
refraction of gold, nd, and wavelength, l, as:
g ¼ jð2pðn� jkÞl
Þ: ð9:3Þ
Note that n is the real part of the complex refractive index andk is the extinction coefficient. The
effective Fresnel reflectance (R) is given by R¼ r.r�, and the dependence of reflectance on
pressure can be found by differentiating the reflectance expression with respect to pressure:
dR
dP¼ r:
dr*
dpþ r*
dr
dpð9:4Þ
where
dr
dp¼ 2ndðdnw=dpÞ½nwtanhgd þ ndðtanhgdÞ2 � ncnd � ncnwtanhgd�
½ndðnc þ nwÞþ ðncnw þ nd2Þtanhgd �2: ð9:5Þ
The analysis presented here can be extended to uncoated fibre by considering the coating
thickness to be zero in Equation (9.2). Reflectance (R) and variation of the reflection coefficient
with pressure, are now given by
R ¼ nc � nwð Þ2nc þ nwð Þ2 ð9:6Þ
and
dr
dp¼ � 2ncðdnw=dpÞ
ðnc þ nwÞ2: ð9:7Þ
250 Microwave Photonics: Devices and Applications
A comparison of the sensitivity of uncoated and coated fibre is presented in Figure 9.13 as a
function of coating thickness. In this calculation the complex index of refraction of gold has
been assumed to be 0.18� j2.21. Note that the calculation based on this simplified model
demonstrates an improvement of as much as 15 dB for a thickness of about 50 nm. Naturally a
more accurate model is required for down-tapered coated and uncoated fibres.
9.3.2 Finite Element Model (FEM) for Various Fibre Sensor Designs
In order to predict the performance of various fibre tip geometries, FEM simulations have been
performed using COMSOL. The physical dimensions of 0.1mm length and core diameter of
10 mm are considered for straight-cleaved optical fibre. Core and cladding refractive indices of
1.4456 and 1.4378 respectively are used for all fibres. A small tapering angle of 6� is consideredfor the etched fibre sensor to properly represent anisotropic etching. The FEM modelling of
gold-coated fibre is based on a gold-coating thickness of 100 nm with complex index of
refraction of 0.18� j0.31. The hybrid mode of HE11 is considered as the dominant mode and
the simulated power density profile of the field along the fibre length is depicted in Figure 9.14.
All the sensor tips have been immersed in water. The region outside the cladding is considered
as an absorbing boundary in order to reduce computation complexity. Simulation results for
straight cleaved, down-tapered uncoated and coated fibres have indicated power densities of
�46.7,�51.5 and�36.3 dBm/mm2 respectively, with amarked improvement in reflectance for
the coated fibre case.
9.3.3 Fabrication of Fibre Sensors
The probe sensitivity can be enhanced by increasing the reflected signal power. The fibre tip is
etched to the size of 7–10microns before it is coated with a thin layer of metallic material, such
10-1 100 101 102 1030
2
4
6
8
10
12
14
16
18
20
X: 52Y: 16.15
Impr
ovem
ent i
n se
nsiti
vity
, (dB
)
Coating thickness, (nm)
Figure 9.13 Improvement in sensitivity of a straight cleaved fibre with a thin gold coating. (The
classical appproach formodelling that is presented here is not totally accurate for thin films, but it is a good
approximation for a general understanding of this problem)
RF and Microwave Photonics in Biomedical Applications 251
Figure 9.14 FEM simulation of power density distribution along the direction of propagation of HE11
for various sensor designs using COMSOL; (a) straight cleaved fibre with mm cross-sectional diameter,
(b) tapered fibre with 7mm cross-sectional diameter, (c) tapered fibre with gold coating
as gold. The fibre tip is etched to a smaller diameter by wet chemical etching of the fibre, using
HF (50% by volume) solution. In this process the fibre being etched is connected to a light
source and detector, and the fibre tip is dipped in the HF solution for tip etching, while the
back reflected signal is continuously monitored. The etched fibre is then coated with a
252 Microwave Photonics: Devices and Applications
Figure 9.15 Sensing fibre tip images with 10�magnification: (a) down-tapered gold coated fibre and
(b) down-tapered uncoated fibre tip cross-sectional diameter, 7 mm
semitransparent film of gold. The thin gold layer was sputtered on the fibre tip with a
Cressington 108 sputtering machine, and the approximate thickness of the gold layer is
50–150 nm for sputtering times of 5–20 s. An optical image of the fabricated coated and
uncoated fibres is shown in Figure 9.15.
9.3.4 Experimental Set-up and Results
The fibre-optic hydrophone system is constructed with commercially available components
using singlemode FC/APC connectors. The system block diagram is shown in Figure 9.16, and
the system is composed of optical source, optical sensor, acoustic source and optical receiver
assemblies, as reported in [31]. The optical source is the 1550 nm distributed feedback (DFB)
laser (NEC NX8563LB) with an output power of �2 dBm for Ib¼ 30mA. The source is
coupled to a 10 dB optical coupler and the output from the 10% coupled arm is sent to the
erbium-doped fibre amplifier (NuPhotonics NP2000CORSV303500FCA1) which has an
optical gain of 40 dB and output power of up to 30 dBm. The output from the EDFA is
divided equally using a 2� 2 coupler with a 3 dB coupling factor. One of the optical outputs is
immersed in water as reference, while the other output is connected to the optical sensor. The
optical sensor is immersed in a water tank and placed at the focal point of a focused acoustic
transducer.
The acoustic transducer is a one-element transducer (Sonic Concepts H110AS/N 01) with
dual band operation at frequencies of 1.6MHz and 5.0MHz. It requires a radio frequency (RF)
impedance matching network and 50W over dual bands of 1.41–1.98MHz and 5.0–5.7MHz.
The transducer has an active diameter of 20 mm and it has a focal length of 34.52mm. An RF
power amplifier provides a maximum pulsed power level of 100Wwith 25% duty cycles. The
position of the acoustic source and optical hydrophone are controlled by a precision scanning
system from Onda Corporation. The system provides precise six-axis positioning and data
acquisition from any sensor in thewater tank for accurate measurements of acoustic fields. The
precision of each axis is repeatable within 12 mm and absolute accuracy of 25 mm over 30 cm.
Experiments have been performed using the various optical sensors discussed in Section 9.3.3.
The reflected optical energy is collected in a wide band amplified InGaAs detector (Thor Labs
PDA 10CF) with a responsivity of 0.95A/Wat 1500 nm and signal bandwidth of 150MHz. It
has a transimpedance gain of 5 kW and noise equivalent power of 12 pW/(Hz)1/2. Comparison
RF and Microwave Photonics in Biomedical Applications 253
Chi
p-ho
pe3
dB c
oupl
er
SMSC
A22
3RP
5005
FA
Wat
er ta
nk
OU
T
Ref
eren
ce
fibre
Sens
or fi
bre
Pow
eram
plifi
er a
nd
mat
chin
g ci
rcui
t
Ultr
asou
nd
trans
duce
r
Thor
labs
PIN
ph
oto-
dete
ctor
PD
A10
CF
Agi
lent
spec
trum
an
alys
er
E840
8A
50%
OU
T
50%
OU
T
ININ
NEC
DFB
La
ser 1
550n
m
NX
856
3LB
10%
OU
T
90%
OU
T
IN
10dB
cou
pler
SMSC
A22
3RP1
005F
A
Nu
phot
onic
s ED
FA
NP2
000C
OR
SB30
3500
FCA
1
Opt
ical
isol
ator
ISIL
PD55
SS9
0.76
m
0.38m
0.46 m
Not
con
nect
ed a
t pr
esen
t
Sign
al
gene
rato
rag
ilent
3325
A
Figure
9.16
System
block
diagram
[31]
254 Microwave Photonics: Devices and Applications
of fibre performance using three designed sensors depicted in Figure 9.14 has typically
indicated an 11 dB and 13 dB improvement for gold 5 second (about 50 nm thick) coated fibre
over straight cleaved and uncoated down-tapered etched fibre sensors, which corroborates the
simulated predictions from FEMmodelling. Themeasured reflected optical signals at different
EDFA power levels indicate a 2 dB variation for every dB variation of optical power. The noise
floor level also increases at the same rate, which indicates that the receiver noise is dominated
by amplified relative intensity noise (RIN) of laser source. Using common mode rejection of a
balun as a power combiner, the amplified RIN noise is cancelled by 14 dB leading to a shot-
noise dominated noise floor of the detection system [32]. The measured optimum gain
responsivity for 25 dBmoptical power ismeasured as�245 re 1V/mPawith average sensitivityas low as 150 Pa.Comparison of time domain pressure response of this optical hydrophonewith
commercially available needle hydrophones is depicted in Figure 9.17, where a modulating
tone burst at frequency of 1.5MHz is employed. The small active sensing area of <10 mm of
these sensors avoids spatial averaging leading to high spatial resolution required for diagnostic
applications. As depicted in Figure 9.18, the sensitivity of this optical hydrophone is flat at least
up to 60MHz, which is significantly better than the results achieved for the commercially
available membrane (polyvinylidenfluorid (PVDF)) hydrophone, even after spatial averaging
corrections [7, 8, 33, 34]. Comparison of the best reported data in each category of fibre-optic
hydrophone (i.e. intensity, phase, wavelength and frequency methods) is summarized in
Table 9.1. The reported experimental results (circa 2008) show that the designed gold coated
fiber provides the best performance in terms sensitivity, bandwidth, and responsivity.
8 10 12 14 16 18 20-3
-2
-1
0
1
2
3
4
Time, (µs)
Vol
tage
, (V
)
needle hydrophone
gold coated FOPH
Figure 9.17 Comparison of the measured pressure versus time response of the 5 seconds gold-coated
fibre-optic hydrophone (in solid line) against a commercially available needle hydrophone (in dash line)
to a 1.5MHz tone burst at acoustic pressure of 1 MPa
RF and Microwave Photonics in Biomedical Applications 255
100 MHz hydrophone calibration
-310
-300
-290
-280
-270
-260
-250
-240
-230
1009080 70 605040302010 0 Frequency (MHz)
PVDF hydrophone with TDS method
PVDF hydrophone with TGFA method
PVDF hydrophone with nonlinear model method
Current gold coated fibre optic probe hydrophone
Hyd
roph
one
sens
itiv
ity
(dB
re
1V/µ
Pa)
Figure 9.18 Comparison in sensitivity response versus frequency for the fibre-optic hydrophone and a
membrane hydrophone. Upper trace data for the 5 second (50 nm) gold-coated fibre-optic hydrophone up
to 60MHz in solid line and estimated response up to 100MHz in dashed line. Lower traces and data points
are for a commercially available 0.5mm diameter, 25mm thick, coplanar PVDF membrane hydrophone.
Two measurement methods of TDS and TGFA are employed to correct for spatial averaging errors of a
500mm wide hydrophone
Table 9.1 Summary of performance comparison of various acoustic pressure sensors.
Sensing technique Detection technique
Gain
responsivity B.W.
Minimum detect-
able pressure
Acousto-optic phase
change in mandrel
fibre [22]
External interfero-
metric phase
detection
Not reported 0.75–10 kHz 92� 27 kPa
Inrtinsic Fabry–Perot
resonant structure [27]
Intensity detection Not reported 20MHz 5 kPa
Multilayer resonant
structure [28]
Intensity detection �264 dB re
1V/mPa10 MHz Not reported
External Bragg cell
wavelength
modulation [24]
FM detector
frequency
detection
Not reported 100Hz
�1.2 kHz
1 kPa
Wavelength modula-
tion of distributed
Bragg reflector
fibre [30]
Intensity detection
of beat
frequencies
Not reported 20MHz 6.4 kPa
Intensity modulation of
reflected light
Intensity detection
of reflected light
�245 dB re
1V/mPaOver
60MHz
150 Pa in shot-noise
dominated
256 Microwave Photonics: Devices and Applications
9.4 Approaches to NIR Imaging
Shadow images created as light passes through the body were first proposed by Cutler in 1929
for medical imaging [35]; however, he found the low resolution of the images limited its
clinical application due to high scattering and absorption. In the past 20 years, significant
advancements in laser and detector technologies in the near-infrared (NIR) electromagnetic
spectrum have been driven by the long haul telecommunication industry; combined with a
better understanding of light propagation in tissue this has now led to renewed interest in optical
imaging of the human body as well as acquiring information about tissue optical and dynamic
properties noninvasively.
In NIR spectroscopy the main aim is to extract the optical properties (absorption and
scattering) of the living tissue. The absorption, ma, and reduced scattering, m0s, parameters of
tissue can provide information on a variety of physiological processes. Absorption information
is used to characterize the concentration of biological chromophores, such as haemoglobin,
which in turn indicates the physiological changes in blood [36]. Scattering information
quantifies the composition, density and organization of tissue structures, such as cells and
subcellular organelles [37, 38]. Therefore NIR techniques could ultimately provide informa-
tion about disease-related functional and structural changes in tissue.
Currently, three main categories of diffuse optical measurements have been developed: (i)
continuous wave (CW), (ii) time domain and (iii) frequency domain measurements. In
continuous wave (CW) systems, light sources emit light continuously at constant amplitude
(or are modulated at frequencies not higher than a few tens of kHz to reject ambient light using
synchronous detection schemes). CW systems measure only the amplitude decay of the
incident light. Time-domain, or time-resolved, systems introduce extremely short (picosecond)
incident pulses of light into tissue, which are broadened and attenuated by the various tissue
layers (e.g. shin, skull, cerebrospinal fluid and brain). A time-domain system detects the
temporal distribution of photons as they leave the tissue, and the shape of this distribution
provides information about tissue absorption and scattering. In frequency-domain systems, the
light source shines continuously but is amplitude-modulated at frequencies at least on the order
of tens of MHz. Information about the absorption and scattering properties of tissue are
obtained by recording the amplitude decay and phase shift (delay) of the detected signal with
respect to the incident signal. These techniques are each discussed next.
9.4.1 Continuous-wave (CW) Method
The absorption spectrumof tissue iswavelength dependent, which ismainly the contribution of
Hb (haemoglobyn), HbO and water, as depicted in Figure 9.2, where the absorption coefficient
is provided for blood concentration of 5% in whole tissue and 100% water. Measuring the
concentration of an absorbing species in a sample is accomplished by applying the Beer–
Lambert law [39], where the absorption of a sample at a given wavelength is directly
proportional to the concentration of the absorbing material, its extinction coefficient and the
path length of light through it. The Beer–Lambert law assumes that the medium is homoge-
neous, the incident light is collimated and reflection and scattering do not contribute to the loss
of the transmitted light. The Beer–Lambert law analytically expresses optical density (i.e.
absorbance) as
lnðI0=IÞ ¼ srd ¼ mad ð9:8aÞ
RF and Microwave Photonics in Biomedical Applications 257
or
log10ðI0=IÞ ¼ eCd ¼ mad
2:3¼ Optical Density ðODÞ ¼ absorbance ð9:8bÞ
where I0 is the incident intensity, I is the transmitted light intensity, s is the absorption cross
section, r is the number density of the absorbing molecules, C is the concentration of the
absorbing molecules (in mM), d is the path length (in cm), e is the extinction coefficient for asolutionofmolarconcentration(inmolar�1 cm�1),andma is theabsorptioncoefficient (incm
�1).
The Beer–Lambert relation holds true when specular reflection or scattering does not
contribute to the loss of transmitted light. This is clearly not the case in tissue. When the
scattering length is shorter than, or comparable to, the absorption length, the optical properties
cannot be accurately determined using the Beer–Lambert law. The first attempts at diagnostic
imaging using optical radiation revealed that multiple scattering occurs when visible to
near-infrared light propagates through tissue and blurs features below the surface. As a
consequence, any measurement of the transmitted intensity through more than a few milli-
meters of tissue is dominated by scattered light. The scattering characteristic of tissues is
commonly expressed in terms of the transport (or reduced) scattering coefficient (correspond-
ing to isotropic scattering),
m0s ¼ msð1� gÞ ð9:9Þ
where ms is the scattering coefficient and g is the anisotropy factor of scattering equal to the
average cosine of the single-scattering phase function [40]. In order to correct for the multiple-
scattering effect in the tissue, a modified Beer–Lambert law is introduced,
OD ¼ � log10I
I0¼
Xi
eiCiLBþG ð9:10Þ
where L is the path length (in cm), B is a path-length factor, which accounts for increases in the
photon path length caused by tissue scattering, G is the measurement geometry factor, and
index ‘‘i ’’ represents the ith chromophore. Parameters e and L remain constant, andB andG are
assumed to be constant. Therefore the change in optical density is given by
DOD ¼ � log10Ifinal
Iinitial¼
Xi
eiDCiLB: ð9:11Þ
By considering the contribution of only two chromophores, Hb andHbO, the above equation
becomes:
DODl ¼ ðelHbOD½HbO� þ elHbD½Hb�ÞLBl ð9:12Þwhere [HbO] and [Hb] are themolar concentrations of oxy- and deoxy-haemoglobin for a l thatindicates a particular optical wavelength. The changes in oxy- and deoxy-Hb concentrations
(and therefore the change in total Hb concentration) are assumed to be wavelength indepen-
dent. This assumption could be invalid, if different wavelengths sample different volumes of
tissue with different haemoglobin concentrations.
By measuring DOD at two wavelengths (l1 and l2) and using the known extinction
coefficients of oxy-haemoglobin (eHbO) and deoxy-hemoglobin (eHb) at those wavelengths,
258 Microwave Photonics: Devices and Applications
we can then determine the concentration changes of oxyhaemoglobin and deoxyhaemoglo-
bin [41],
D Hb½ � ¼ el2Hbo DODl1
Bl1 � el1Hbo DODl2
Bl2
ðel1Hbel2HbO � el2Hbel1HbOÞLð9:13aÞ
D HbO½ � ¼ el1Hb DODl2
Bl2 � el2Hb DODl1
Bl1
ðel1Hbel2HbO � el2Hbel1HbOÞL: ð9:13bÞ
In addition to the Beer–Lambert law and the model presented here, a more rigorous theory
for the migration of photons through tissue has been developed based on the radiative transport
equation [42]. This approach recognizes that near-infrared photons in tissue essentially
undergo a random walk because the scattering probability is much greater than the absorption
probability, and therefore their propagation through tissue can be described by a diffusion
equation. The photon diffusion equation is [13, 42, 43]:
1
v
@Fðr; tÞ@t
�Dr2Fðr; tÞþmaFðr; tÞ ¼ Sðr; tÞ; ð9:14Þ
whereF(r, t) is the photon fluence at position r and time t and the photon fluence is proportional
to the optical intensity. S(r, t) is the source distribution of photons. D¼ 1/[3(ma þ m0s)] is the
photon diffusion coefficient, m0s is the reduced scattering coefficient, ma is the absorption
coefficient and v is the speedof light in themedium.Note that the absorption coefficient is related
to the extinction coefficient and the concentration as ma¼ eC. For a combination of the
haemoglobin chromophores,
ma ¼ eHbO½HbO� þ eHb½Hb�: ð9:15ÞEquation (9.14) accurately models the migration of light through highly scattering media
provided that the probability of scattering is much greater than the absorption probability. Note
that all factors in (9.14) are wavelength-dependent. Solutions of the photon diffusion equation
can be used to predict the photon fluence (or intensity) detected for typical diffuse measure-
ments. Assuming that concentration changes are both global and small, the solution of the
photon diffusion equation for a semi-infinite medium is
DOD¼�logFFinall
FInitial
¼1
2
3m0s
mInitiala
� �1=2
1� 1
ð1þLð3m0Initials mInitial
a Þ1=2" #
ðeHbOD HbO½ �þeHbD Hb½ �ÞL
ð9:16ÞThe solution of the photon diffusion equation for representative tissue geometry (Equa-
tion (9.16)) tells us that the modified Beer–Lambert law is reasonable for tissues with spatially
uniform optical properties when the chromophore concentration does not change significantly
(i.e. D[X]/[X]� 1). The path length factor B in Equation (9.12) is given by
B ¼ 1
2
3m0s
mInitiala
� �1=2
1� 1
ð1þ Lð3m0Initials mInitial
a Þ1=2" #
ð9:17Þ
RF and Microwave Photonics in Biomedical Applications 259
for a semi-infinite medium. This shows that B depends on tissue scattering, the initial
chromophore concentration, the extinction coefficient (and thus B is wavelength dependent)
and the optode separation. In practice, the validity of the assumption thatB is independent ofma
and L has often been ignored sinceB is in general empirically determined and the changes inma
are typically small.
The quantity of oxygen in blood is often expressed as the haemoglobin oxygen saturation (S),
which is defined as
S ¼ ½HbO�½HbO� þ ½Hb� � 100% ¼ ½HbO�
½HbT� � 100%: ð9:18Þ
This expresses the percentage of the total oxygenated haemoglobin.
A typical CW imager is shown in Figure 9.19, where light sources are driven by the drive
circuitry and emit near-infrared light into tissue. The diffused and attenuated light is collected
and converted to an analogue electrical signal by the photodetector. Finally, the amplified
signal goes through theA/D converter so that a computer can be used to process and display the
data. As shown in Figure 9.19, the light source is at the second stage of the open loop, so that the
quality of the light source is vital to the whole system. Generally, there are three choices for
light sources inCWimagers:white light (such as tungsten light bulbs), lasers and light-emitting
diodes (LED). Light spectrum purity and light intensity output are two important parameters.
White light has been extensively used with interference filters at 760 nm and 850 nm to detect
blood volume and deoxygenation changes [39]. Lasers are ideal light sources for many
applications due to their excellent spectral purity and collimation. The linewidth of the isolated
wavelength is less than 1 nm.A laser beam focuses all the light energy into a very small area and
over a very small wavelength bandwidth, with potential for tissue damage even though its
power is much less than that of a white light source. This is why its power is limited to less than
0.1mWby the Food andDrugAdministration�s (FDA�s) law (type I) when laser light is applied
to humans. Thus, it will be difficult to satisfy the light intensity requirement of a CW imager.
LED spectral purity is about 30 nm and it is good for a CW imager. More light intensity can be
utilized since LEDs illuminate the tissue more diffusely than a laser but more like a point light
source than white light, and with less heat. Stability of light intensity is another important
Figure 9.19 Typical block diagram of a single channel CW imager [42].
260 Microwave Photonics: Devices and Applications
requirement of a CW imager. Both white light and LEDs have a drift of light power and thus
require 2–3 minutes for warming up, but the laser diode operates more stably.
A typical experimental result by an LED imager [42] is shown in Figure 9.20. The probewas
placed on the lateral side of the lower right-hand side of the leg. Seated baseline measurements
were made for 1 minute. Seated exercise began by the subject doing 60 repetitions of toe
extension (pointing the toe as far as possible). This seated exercise recruited the extensor
muscles to a greater extent than the flexors which are more activated during plantar flexion or
walking.
Figure 9.20 Blood volume and deoxygenation changes during cycling exercise are presented during
physical exercise and rest periods. A grey scale coding (black for increase, light grey color is for moderate
decrease, and dark grey for significant decrease) is employed to provide approximate changes in blood
volume and deoxyngenation levels (in unit aM) compared to the initial condition represented in grey
colour [42] (Reprintedwith permission fromY. Lin, et al., ‘‘Noninvasive, low-noise, fast imaging of blood
volume and deoxygenation changes inmuscles using light-emitting diode continuous-wave imager,’’ Rev.
Sci. Instrum. 73(8), 3065–3074 (2002). Copyright 2002, American Institute of Physics)
RF and Microwave Photonics in Biomedical Applications 261
9.4.2 Pulsed-time or Time-resolved Method
Time-resolved reflectance spectroscopy (TRS) is a novel nondestructive method for the
complete optical characterization of highly diffusive media, that is for the evaluation of the
absorption coefficient ma and the reduced scattering coefficient m0s. TRS is gaining acceptance
in biomedicine for the noninvasive investigation of biological tissues [44–46] since a short light
pulse injected into a turbid medium experiences absorption and scattering during photon
propagation. Moreover, the diffusely reflected pulse is attenuated, broadened and delayed.
Consequently the best fit of its time distribution with a theoretical model of light propagation
allows the simultaneous evaluation of both optical coefficientsma andm0s by probing bulk rather
than superficial properties. Furthermore, useful information on internal quality of tissue can be
gathered.
When a narrow collimated pulsed light beam is normally incident on the surface of a semi-
infinite or finite homogeneous tissue slab, the diffuse photon fluence rate F(r, t) satisfies thediffusion equation (Equation (9.9)). The fluence rate can be accurately calculated using
Equation (9.14) if ma � m0s and if the point of interest is far from sources or boundaries. For
a short pulse from an isotropic point source that is represented by a delta function of
s(r,t)¼ d(0,0), it may be shown that in an infinite medium the solution of Equation (9.14) is
Fðr:tÞ ¼ vð4pDvtÞ� 3=2exp � r2
4Dvt�mavt
� �: ð9:19Þ
One can use this Green�s function to solve the semi-infinite problem by making two further
assumptions. First, assume that all the incident photons are initially scattered at a depth
z0 ¼ ðm0sÞ� 1
so that the actual source term becomes the simple delta function described above.
The second assumption is that F(r,t)¼ 0 on the physical boundary z¼ 0. As discussed by
Eason et al. [47], this boundary condition can be met by adding a negative or image source of
photons to the infinite medium. The fluence rate per incident photon can then be written in
cylindrical coordinates as the sum of contributions from the two sources:
Fðr; z; tÞ ¼ cð4pDctÞ� 3=2expð�mactÞ exp � ðz� z0Þ2 þ r2
4Dct
" #� exp � ðzþ z0Þ2 þ r2
4Dct
" #( ):
ð9:20ÞThe number of photons reaching the surface per unit area per unit time, jJ(,0,t)j, can be
calculated from Fick�s law as:
jJðr; 0; tÞj ¼ �DrFðr; z; tÞjz¼0 ð9:21Þwhich leads to the final expression for the reflectance R(r, t):
Rðr; tÞ ¼ jJðr; 0; tÞj ¼ ð4pDcÞ� 3=2z0t
� 5=2expð�mactÞexp � r2 þ z204Dct
� �: ð9:22Þ
For the case where r2 � z20, also note that
d
dtlnRðr; tÞ ¼ � 5
2t�macþ
r2
4Dct2: ð9:23Þ
262 Microwave Photonics: Devices and Applications
The observation that
limt!¥
d
dtlnRðr; tÞ ¼ �mac ð9:24Þ
leads to the suggestion that the absorption coefficient of the tissue can be determined from the
asymptotic slope of the ln R(, t) versus t curve. The transport scattering coefficient m0s can also
be determined from the lnR(, t) versus t curve by noting that the slope is zero at tmax, the time of
maximum detected signal. Solving Equation (9.23) yields the expression:
m0s ¼ 1
3r24mac
2t2max þ 10ctmax
� ��ma ð9:25Þ
Therefore, both optical properties of a semi-infinite slab of tissue could, in principle, be
obtained from Equations (9.24) and (9.25) by measuring the diffusely reflected light some
distance from the source as a function of time. A superior signal-to-noisewould be obtained by
integrating the reflected light over some larger area.
The TRS instrumentation consists of both hardware and signal processing software. The
laser pulse scanning mammography, as shown in Figure 9.21 and developed by PTB [48],
measures time-resolved transmittance through the female breast, which is gently com-
pressed between two parallel glass plates. When the source fibre and the detector fibre
bundle are scanned in tandem across the breast, optical properties can be extracted from the
measured photon density versus time. The proposed mammograph is equipped with two
excitation channels at 670 nm and 785 nm and one detection channel. The output pulse
trains of two picosecond laser diodes are multiplexed in time, and optical mammograms are
simultaneously recorded by detecting the transmitted photons by a fast photomultiplier.
Distributions in the times of flight are recorded for 100ms at each scan position by high-
throughput time-correlated single photon counting electronics at count rates of up to 1MHz.
The optical mammograms are recorded along cranio-caudal and medio-lateral projections
within 3 to 5 minutes each as 1000–2000 scan positions are typically sampled at a step size
of 2.5mm.
Mammograms were generated from a variety of parameters derived from recorded
distributions of times of flight, such as photon counts in selected time windows. By
analysing photon counts in a late time window, changes in absorption can be imaged
qualitatively, whereas photon counts in an early time window are most sensitive to changes
in scattering. In Figure 9.22 the tumour shows up as reduced transmittance in the late time
window, whereas the cyst is clearly seen in the image representing photon counts in the first
time window.
In the clinical trials, mammograms are recorded with the transmitting fibre and receiving
fibre bundle facing each other (on-axis geometry). Whereas lesions can be localized laterally
with sufficient precision in optical mammograms recorded in this way (localization of lesions
in two dimensions), the position of lesions along the compression direction cannot be inferred
from measurements taken using the on-axis arrangement. A promising approach to gain
information about the location of the lesion under investigation along the compression
direction (three-dimensional localization) is to record optical mammograms at several lateral
offsets between the transmitting fibre and detecting fibre bundle (off-axis geometry) and to
analyse the shifts of features in mammograms.
RF and Microwave Photonics in Biomedical Applications 263
9.4.3 Intensity-modulated or Frequency-domain Method
Light propagation in scattering media can be described by the Boltzmann transport equation.
Under specific approximations and limitations, the Boltzmann transport equation is simplified
to the time-dependent standard diffusion equation (SDE) as shown in Equation (9.14).
Analytical solutions for the SDE have been described for a variety of boundary and initial
conditions. In the case of a sinusoidal point source modulated at a frequency of f¼v/2p, theinfinite medium solution is given by [13, 48]:
fðr; tÞ ¼ Adc
4pD
expð� r=dÞr
þ Aac
4pD
expð� krealrÞr
� exp � iðkimagr�vtÞ� � ð9:26Þ
where Adc and Aac are the DC and RF components of the source respectively, d is the DC
penetration depth, and kreal and kimag are the real and imaginary components of the photon
density wave (PDW) complex wave number. By convention, kreal governs PDW amplitude
Mode-locked laser 785nm
Mode-locked laser 670nm
Photodetector (photomultiplier)
Time-correlated single-photon counting
Signal processor (computer)
~ 400ps
~ 100ps
Optical fibre
X-Y Scanner
x
z
y
Fibre bundle
~ 2.4ns
Dispersed short optical pulses after
flight through breast
Figure 9.21 Schematic diagram of the initial PTB time-domain scanning optical mammography, where
two trains of optical short pulses are combined before injection into the breast tissue with a black circle
representing the tumour/cyst. Reproduced from [48]
264 Microwave Photonics: Devices and Applications
attenuation and kimag describes PDW phase propagation. The complex wave vector, I¼ krealkimag, is dependent on the absorption and reduced scattering parameters, as well as the source
modulation frequency and the velocity of light in the medium:
kreal ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffi3
2mam
0s
r1þ v
cma
� �2" #1=2
þ 1
8<:
9=;
1=2
ð9:27Þ
kimag ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffi3
2mam
0s
r1þ v
cma
� �2" #1=2
� 1
8<:
9=;
1=2
: ð9:28Þ
For reflectance or transmittance measurements performed in infinite media, PDW phase lag
and amplitude attenuation relative to the source are
Qlagðr;vÞ ¼ kimagr ð9:29Þ
Aattðr;vÞ ¼ exp½ � krealðvÞr�4pDr
: ð9:30Þ
For semi-infinite geometries, in reflectance mode and accounting only for the fluence term,
PDW phase lag and amplitude attenuation are given as follows:
Figure 9.22 TRS generated mammograms of a breast (cranio-caudal projection, l¼ 785 nm) contain-
ing a tumour (invasive ductal carcinoma) and a cyst as a function of scan distance of x in cm: (a) reciprocal
photon count in N8, the eighth of 10 consecutive time windows, (b) photon counts in N1 the first time
window. Reproduced from [48]
RF and Microwave Photonics in Biomedical Applications 265
Qlagðr;vÞ ¼ kimagðvÞr0 � arctanIMAG
REAL
� �; ð9:31Þ
Aattðr;vÞ ¼ Air
4pDðREAL2 þ IMAG2Þ1=2; ð9:32Þ
where real and imaginary parts of the received signal are analytically related to operating
frequency and physical dimensions by:
REAL ¼ exp½ � krealðvÞr0�r0
� cos kimagðvÞðr0b � r0Þ� � exp½ � krealðvÞr0b�
r0bð9:33Þ
IMAG ¼ sin kimagðvÞðr0b � r0Þ� � exp½ � krealðvÞr0b�
r0b: ð9:34Þ
Note that distances r0 and r0b are expressed by:
r0 ¼ ½ðm0sÞ� 2 þ r2�1=2; ð9:35Þ
r0b ¼ 4
3m0s
1þReff
1�Reff
þ 1
m0s
� �2
þ r2
" #1=2
: ð9:36Þ
Air is the net amplitude response of the instrument (due to source power, detector gain, etc.) and
Reff is the effective reflection coefficient.Reff represents the fraction of light that has reached the
surface and is reflected back into the medium.
The frequency-domainmeasurement technique provides both optical tissue parameters such
as TRS and is therefore more advantageous than the CW imaging system. From the three
techniques discussed so far, the frequency domain is most compatible with advances in radio
technologies and hence ismore cost effective than TRS. There are several approaches to realize
frequency-domain detection, such as the I/Q system, the phased array system and the
broadband frequency-domain system.
9.4.3.1 I/Q System
Figure 9.23 depicts the basic form of an I/Q system [49, 50], where two in-phase and quadrature
phase signals are employed for the detection of optical parameters. The phase difference and
amplitude attenuation between the reference oscillator and the signal pathway is detected. The
phase-shifted path involves a laser diode, an optical detector, an appropriate amplifier and a
narrowband filter. Thus the outputs of the I/Q detector are the sine and cosine components of
phase and amplitude and thus require trigonometric computation by nonlinear analogue
circuitry or by conversion to the digital domain and the use of a look-up table or other similar
method. Quadrature imbalance in the I/Q detector is typically 0.3� in phase and 0.5 dB in
magnitude, but anyunwanted signalwill be detected and cause variableDCoffsets in the output
that need to be distinguished from the DC sine/cosine output.
266 Microwave Photonics: Devices and Applications
9.4.3.2 Phased-array System
To achieve high sensitivity in the detection of small objects embedded in a scattering medium,
dual interfering sources in a phased-array configuration have been explored experimentally and
theoretically [51, 52]. The analytical solution of the dual-source system can be derived by
applying the superposition principle of a linear system by summing over the solutions to the
single source system.Complex terms ofA1exp(�jvt) andA2exp(�jvt þ Df) are employed to
represent the pair of sources. The total field is equal to the superposition of the two independent
solutions of those terms as represented by:
Ftotalðr; tÞ ¼ F1ðrs1 ; tÞþF2ðrs2 ; tÞ: ð9:37ÞPhoton-density waves generated from a pair ofmodulated optical sources that are excited by
180� out-of-phase RF signals, have been shown byKnuttel et al. [53] to interfere destructively.A null exists along the line of symmetry in the phased-array geometry, where a detector
placement along the null-line detects a very small perturbation to the symmetric environment.
The goal of this initial application was to desensitize the detector at the excitation surface (in a
reflection mode measurement) to effects near the surface. It was shown that the phase
measurement provided greater sensitivity for absorbers at larger depths than for the single-
source case (cf. Figure 9.24). In this way, the approach showed promise for localization by
providing improved depth information with a reflection measurement on the planar source
boundary.
A block diagram of a frequency-domain phased-array system is shown in Figure 9.25. A 50
MHz oscillator is modulated in the single sideband (SSB) mode by a 2 kHz sine wave and the
upper sideband (USB) from the transmitter is selected and split into 0� and 180� by an RF
splitter. The RF signals then modulate the two laser diodes respectively with an optical
modulation near 90%. The optical signal after passing through the diffused medium and
experiencing scattering due to the presence of an object is detected by a photomultiplier tube
(PMT). The output of the PMT goes through two separate channels for detection of amplitude
and phase. The automatic gain control (AGC) voltage of the receiver is used to indicate the
amplitude of the RF signal and the phase information is obtained from the phase meter.
A dual-source scanner that collects optical parameters of cylindrical absorbers using a
phased array system is depicted in Figure 9.26(a). Two sizes of objects of 1 cm and 0.1 cm
Turbid medium
under test
Laser diode
Opticalsplitter
Single RF tone
Photo- detector
IQ demodulator
Signal processor
Figure 9.23 Block diagram of I/Q system based on the detection of optical parameters (reduced
scattering and absorption) of a medium under test. Reproduced from [50]
RF and Microwave Photonics in Biomedical Applications 267
Phase perturbation
0.01
0.1
1
10
100
0.10.010.001Contrast - delta µa (cm-1)
Ph
ase
shif
t (º
)
Simulation-SS
Simulation-DS
Detection noise
Experiment - SS
Experiment - DS
Figure 9.24 Phase perturbation effect in a phased-array system due to a contrast agent used in an intra-
lipid tank, where experimental and simulation results of a signal source (SS) and dual sources (DS) are
compared as the absorption parameter ma is changed. Reproduced from [51]
50MHz oscillator
SSB modulator
SP
LIT
2kHz phase detector
2kHz oscillator
2kHz
filter
ADC
50MHz SSB receiver
ADC
50MHz SSB receiver
x
z
y
1Hz
0°
180°
Detector PMT
S1
S2
Object
1Hz
USB
Phase detection
AGC control
Amplitude detection
Figure 9.25 Block diagram of a phase-sensitive phased-array detection system for the optical
parameters of an object in an intra-lipid tank. Reproduced from [51]
268 Microwave Photonics: Devices and Applications
diameters are used in this experiment. The two-dimensional scanning of the homogeneous
intra-lipid loaded by the cylindrical object is collected using a measurement system similar to
Figure 9.25 in terms of amplitude and phase. However, modern automatic network analysers
could be employed for accurate amplitude and phase measurements.
The experimental results are depicted in Figure 9.26(b) and (c). Consider the homogeneous
case first. When the dual source is positioned toward either side of the domain, the boundary
selectively absorbs photons from the nearest source, disrupting the interference line. As the
dual source approaches the centre of the input scan line, the mutual absorbing effects of the
boundaries begin to balance, allowing equivalent contributions from both sources that are out
of phase to reach the output plane, manifested as a sharp amplitude null and a 180� shift inphase. When a strong absorber is introduced at the centre of the region (1 cm in diameter), an
additional balancing of absorbed light occurs between each boundary and the absorber.
Therefore, at a point between the boundary and the absorber, the loss to the boundary and
the heterogeneity are somewhat balanced and a smaller null line is detected. Likewise, when
the sources are centred about the heterogeneity, a large-amplitude null and phase step are
detected.
As explained in [52], the positions and sizes of the nulls in the magnitude are related to the
nature of the absorber and the proximity of the boundaries. The larger absorber has the
narrowest central null and satellite nulls that are closer to the outer boundary; all nulls are
Figure 9.26 Experimental results of a dual source phased array: (a) typical phased-array geometry,
(b) amplitude and (c) phase response, when two-dimensional scanning is performed over a cylindrical
absorber. Reproduced from [52]
RF and Microwave Photonics in Biomedical Applications 269
sharper owing to the larger effective absorption by the heterogeneity. The phase responses also
reflect the size of the absorber. All three curves have a 180� phase step in the centre. Over theremainder of the domain, the phase is smooth. The other abrupt phase changes are artificial
phase wraps (as phase plots go through phases larger than 180� or smaller than –180� in
automatic network analysers) and could be corrected with modulo of 2p. Notice that the
magnitudes and phases for the cases with the absorbers asymptotically approach the homoge-
neous case. If the domain were infinite in this scan arrangement, one would see a magnitude
minimum when the detection plane was an appreciable distance on either side of the absorber
and alsowhen it was centred at the absorber, as well as a reduction in null depth just off centre.
Finally, note that the magnitude approaches a maximum as the minimum and maximum scan
positions are approached. Data points do not go all the way to the boundary, where light
intensity at the detector would decrease as a result of losses through the boundary. The
experimental results show the power of differential detection, where small perturbations can be
clearly identified as the line of symmetry is destroyed.
9.4.3.3 Broadband Frequency-domain System
The development of phase-sensitive phased arrays can benefit from the accuracy of modern
vector network analysers, which bring in the capability of accurate broadband measurement
using calibration procedures. There are several advantages of broadband characterization of
tissue. For example, most tissues normally have a layered structure and since photon
penetration depth depends on modulation frequency, as frequency-domain photon migration
(FDPM) is evaluated over a broad bandwidth, intensity-modulated light can be used to quantify
multi-layer tissue absorption (ma) and reduced scattering ðm0sÞ parameters at discrete wave-
lengths, as depicted in Figure 9.2 for low (MHz) and high (GHz)RF frequencies. This feature is
extremely useful in clinical experiments where, due to calibration difficulties, it is preferable to
make a single measurement and obtain as much information as possible [10, 14, 54].
To appreciate fully the advantages of operation at various frequencies to extract optical
properties of a layered tissue, the amplitude and phase of received scattered signals are
simulated as a function of frequency employing Equations (9.31) and (9.32). Simulations are
provided for a homogenous phantom resembling breast tissue and calculations are conducted
for reduced scattering and absorption parameters. The simulation results are rendered in plots
shown in Figure 9.27, indicating a greater amplitude and phase change for the same optical
parameters at higher frequencies. Note that at the same condition, the higher-frequency signal
attenuates more compared to the low-frequency signal, which indicates that the penetration
depth of the diffuse photon density wave over tissue depends on the modulation frequency and
itwill bemore shallow at frequencies approaching 1GHz as opposed to being deeper into tissue
at frequencies approaching 100MHz. This result affirms that spectroscopic information could
be extracted about various layers of the multi-layer tissue structures. As shown in Figure 9.2, a
receivedmodulated signal at GHz range can yield the upper layer (fat layer) information, while
the photon density wave modulated in MHz range can penetrate both the fat layer and muscle
layer so that it will bring the information combined with the fat layer and muscle layer. Hence,
by taking a multi-frequency measurement, one can identify this multi-layer structure and
extract the optical parameters on both layers.
However, as attenuation increases at the higher frequency, a reduction in signal-to-noise
ratio is experienced, which leads to a greater potential for phase measurement error. This issue
270 Microwave Photonics: Devices and Applications
is highlighted in Figure 9.28(a), where measurements of phase have a larger spread as the
amplitude decreases. Moreover, for typical clinical concentrations of absorbing dyes (such as
nigrosin), a higher attenuation is also measured. Therefore to achieve broadband detection, a
flat response over a wide frequency range of laser sources and detectors is crucial. This flat
frequency response allows one to speed up the extraction process without going through the
two-tier calibration processes of the network analyser and multi-wavelength optical link
system. Moreover, accurate measurements of amplitude and phase over various broadband
systems require careful design of high optical sources (i.e. flat laser driver, comparable laser
diode responsivities and reduced laser diode RIN levels) and the optical receiver (i.e. good
responsivity over NIR wavelengths of interest, flat trans-impedance gain and broadband RIN
cancellation techniques). Some of these aspects are well established in the microwave
-6
-5
-4
-3
-2
-1
0
120010008006004002000
Frequency (MHz)
Amplitude attenuation
(dB)
µ ′s = 8/cm µ ′s = 9/cm µ ′s = 10/cm
µ ′s = 8/cm µ ′s = 9/cm µ ′s = 10/cm
(a)
0
10
20
30
40
50
120010008006004002000
Frequency (MHz)
Phase shift (º)
(b)
Figure 9.27 Frequency dependence for different phantom parameters: (a) amplitude change and
(b) phase shift when the reduced scattering coefficient ðm0sÞ changes from 7 cm�1 to 8, 9 and 10 cm�1
respectively. The simulation assumes that the absorption coefficient is constant at 0.04 cm�1. The infinite
boundary condition and source-detector separation of 3 cm are applied
RF and Microwave Photonics in Biomedical Applications 271
photonics community and relevant examples are to be discussed next as part of a broadband
system realization.
9.5 Design and Realization Challenges of Broadband NIR OpticalSpectroscopy Systems
As was discussed earlier, microwave photonics can also be applied to optical spectroscopy,
where diffused photon waves will suffer different levels of scattering and absorption as a
function of various optical wavelengths for tissue depending on the level of oxygenated and
Figure 9.28 Impact of operation at higher frequency on the phase measurement accuracy. (a)
Comparison of measured (dot) and theoretical (line) phase-delay expectation of a broadband frequen-
cy-domain instrument. (b) Comparison of the measured (dot) versus expected (line) attenuation due to an
absorbing dye (nigrosin) at a modulation frequency of 500MHz and a wavelength of 782 nm [13].
Reprinted with permission from T. H. Pham, O. Coquoz, J. B. Fishkin, E. Anderson, and B.Tromberg,
‘‘Broad Bandwidth Frequency Domain Instrument for Quantitative Tissue Optical Spectroscopy,’’ Rev.
Sci. Inst. 71, 2500–2513 (2000). � 2000, American Institute of Physics
272 Microwave Photonics: Devices and Applications
de-oxygenated haemoglobin and tissue structure. Moreover, since scattering increases at a
highermodulation frequency, then the diffused photonwave penetration in the tissue is less and
multi-level data could be collected from a microwave modulation frequency of 1GHz
compared to 100MHz.Microwave photonics techniques are explored here for the development
of custom-designed optical transmitters to be employed for broadband spectroscopic sys-
tems [55, 56].
9.5.1 Broadband Multi-frequency Instrument
The overall system block diagram of the proposed broadband NIR spectroscopy system is
shown in Figure 9.29. Different laser diodes (677 nm, 785 nm, 830 nm and 977 nm) are
sequentially modulated using RF power from a network analyser using an SP4T electrical
switch. A custom-designed laser driver modulates high-power laser diodes with a flat
frequency response up to 1GHz, while an optical switch is used for different source positions
(N positions). Detected light from different detection positions (M positions) on the turbid
medium is collected by a fibre bundle and received by an optical receiver (e.g. APD C5658,
Hamamatsu, Inc.). The network analyser receives the RF signal and forward or backward
scattering parameters are extracted using the measured amplitude and phase at each wave-
length. The first design is based on high-power laser diodes packaged in a transistor outline
Figure 9.29 Simplified block diagram of a broadband multi-wavelength NIR spectroscopy system
using a commercially available (HP 8753ES) vector network analyser. The four-wavelength system is
based on high-power fibre-coupled laser diodes that are driven by a broadband high-current laser driver,
electromechanical optical switches forM source andN detector positions, and a balanced optical receiver.
The source and detector are separated for distances of up to 3 cm in a turbid medium
RF and Microwave Photonics in Biomedical Applications 273
(TO) can (cf. Figure 9.30). DC biasing of the laser diodes for low- and high-power is achieved
by a power transistor (Darlington, TIP125), where a DC biasing current of up to 1.2 A can be
achieved. The active laser driver is employed to provide an RF driver design requirement of
45mA in the laser junction. This drive RF current to the laser diode is achieved using an HJ-
FET from NEC (NE6510179A). Details of this design are discussed elsewhere [54] and a
photograph of the four wavelength optical sources and broadband high-current laser driver
boards placed in front of the APD optical receiver is shown in Figure 9.30. A pre-emphasis
circuit based onRC speed-up is used to overcome the bandwidth limitation of the transistor and
laser diode at higher frequencies and extend the system operation up to 1GHz. Insertion loss
and phase of the system (S21) in free space is depicted in Figure 9.31. Note a flatness of�1.5 dB
and phase linearity of �5� over a decade bandwidth is observed.
An optical link is established using the 785 nm optical source, free-space optical attenuator
(OD 1), and the optical receiver (APD, C5658, Hamamatsu, Inc.) to evaluate the performance
of the active driver for a TO-can package. The overall optical link performance is shown in
Figure 9.32, where a linear phase response is observed; however, the amplitude is attenuated at
higher frequencies at a rate of 25 dB/decade. This performance indicates that there is a
bandwidth limitation due to the laser diode operating at 785 nm (typically 40 dB/decade
after relaxation oscillation frequency) in addition to package parasitics associated with the
TO can (typically 20 dB/decade after the corner frequency). A similar behaviour is required
for the other three wavelengths to avoid calibration corrections using two-tier de-embedding
of the measured data to extract the optical parameters of the turbid medium under test
properly.
Figure 9.30 Overall view of the four-wavelength laser diode transmitter assembly and optical receiver.
An APD is shown on the left and four laser diodes are mounted on the driver on the right. Custom-made
four-laser diode drivers are in a single compact form
274 Microwave Photonics: Devices and Applications
Figure 9.32 Transmission response of the custom-designed optical link using a laser at 785 nm,where a
frequency roll-off is observed due to the bandwidth limitation of a TO-can and high-power laser
Figure 9.31 Transmission response of a custom-designed high-current laser driver. Note the flatness of
the frequency response, especially at high frequencies. This response is obtained by an RC speed-up
circuit, which pumpsmore current at higher frequencies to compensate for the bandwidth limitation of the
high-current transistor and semiconductor laser
RF and Microwave Photonics in Biomedical Applications 275
The TO-can package parasitics are eliminated when the approach described above is
extended to a fully integrated optical transmitter with a C-submount package. For operation
at higher frequencies of up to 3GHz, the TO can will contribute to significant roll-off, which
cannot be corrected by using RC speed-up circuits. Therefore, a C-submount laser diode has
been chosen because of its good output power capability while satisfying the high-speed
performance. A laser diode at 820 nm available on a C-submount is employed in the active
driver module, where a monitoring photodiode and TE cooler have been assembled to monitor
the output optical power and keep the temperature of the laser diode constant as depicted in
Figure 9.33(a). As seen in this picture of the optical transmitter, each component is properly
Figure 9.33 Depiction of a packaged broadband optical source: (a) photograph of the optical source
module at 820 nm with an active driver circuit; (b) measured frequency response of the laser driver over
the frequency range from 100MHz to 3GHz
276 Microwave Photonics: Devices and Applications
labelled and a 50/125 mmmultimode fibre is employed for light coupling. A custom-designed
laser diode driver can provide up to 1ADCbias current for a laser diode and has a flat frequency
response up to 3GHz (cf. Figure 9.33(b)). An optical link is established using the active optical
transmitter, an OD optical attenuator and an optical receiver (APD, C5658, Hamamatsu, Inc.).
The overall optical link performance is shown in Figure 9.34, where a flat frequency response is
observed except around 580MHz. This discrepancy at 580MHz is due to a resonance in the
active driver portion of the optical transmitter. This link performance is used to calibrate the
link performance through biological tissue and the extraction of scattering and absorption
parameters of phantom experiments.
-20
-10
0
10
20
30
40
50
1000900800700600500400300200100
Frequency (MHz)
Am
plitu
de (
dB)
-200
-150
-100
-50
0
50
100
150
200
Pha
se (
º)
Amplitude
Phase
Figure 9.34 The overall optical link performance with flat gain and linear phase at all frequencies,
except for a self-resonance frequency of about 580MHz
Optical transmitter
Optical receiver (APD)
ANA
Intralipid solution
Figure 9.35 Experimental set-up for intra-lipid solution measurements
RF and Microwave Photonics in Biomedical Applications 277
9.5.2 Intralipid Experiments using a Broadband System
To demonstrate the advantages of a broadband system in terms of a greater spatial resolution
and sensitivity at higher frequencies, results of a number of experiments are discussed here.
These experiments are conducted using a liquid phantom (i.e. intralipid), which is a container
primarily filled withmilk as shown in Figure 9.35. The system amplitude and phase response to
optical parameter changes of the liquid phantom are measured. The experimental set-up is
shown in Figure 9.35. The optical transmitter fibre and the receiver fibre are submerged in the
intralipid solution. The distance between the transmitter fibre tip and the receiver fibre tip is
3 cm. Before adding more intralipid, the amplitude and phase response were measured as a
baseline. The results of three separate experiments are reviewed next.
-20
-18
-16
-14
-12
-10
-8
-6
-4
-2
0
2
1400120010008006004002000
Frequency (MHz)
Amplitude difference (dB)
Amp1
Amp2
Amp3
Amp4
Amp5
Amp6
Amp7
Amp8
Amp9
(a)
0
10
20
30
40
50
60
70
80
90
1400120010008006004002000
Frequency (MHz)
Phase shift (º)
Phase1
Phase2
Phase3
Phase4
Phase5
Phase6
Phase7
Phase8
Phase9
(b)
Figure 9.36 System response to the scattering coefficient changes of a liquid phantom: (a) normalized
amplitude attenuation; (b) normalized phase shift
278 Microwave Photonics: Devices and Applications
9.5.2.1 Experiment 1
A small portion of intralipid was added to the solution to change the scattering parameter
and the new amplitude and phase were recorded. Equal amounts of intralipid were added
eight times and the amplitude and phase changes were recorded each time. The amplitude
and phase responses were collected over the frequency range from 100MHz to 1.2 GHz and
the relative performance with respect to a baseline are shown in Figure 9.36, where the rapid
decrease in the amplitude after 1GHz is due to the frequency response of the APD. Also, a
response notch between 558MHz and 761MHz is observed, which is caused by the
frequency response of the custom-designed laser driver. The phase response is hard to read
since the phase difference between each curve is very small compared to the phase shift for
different frequencies.
The normalized data plotted in Figure 9.36 clearly shows that with larger intralipid
concentrations – and hence larger scattering coefficients – there is an increase in amplitude
attenuation and phase shift. From the figure one can also tell that at each intralipid addition,
higher-frequency data always show a larger signal change. The benefit of higher-frequency
measurement is shown more clearly in Experiments 2 and 3.
9.5.2.2 Experiment 2
The system amplitude and phase response to absorbers with different absorption coefficients
was measured. The set-up is shown in Figure 9.37, which is similar to the set-up of Experiment
1 except that cylindrical absorbers with different absorption coefficients were placed in the
middle of the distance between the transmitter fibre and the receiver fibre. Three absorbers with
the same diameter of 4.6mm and different absorption parameters (i.e. absorber1
absorber2 < absorber3) were used. The absorber used was India ink, a well-known reference
material.
The amplitude and phasewere again normalizedwith respect to the baseline data and relative
results are plotted in Figure 9.38. The result shows that the one-wavelength NIR spectroscopy
system can easily identify these three absorbers and themulti-frequencymeasurement gives us
Figure 9.37 Experimental set-up for measuring optical parameters of a turbid medium dominated by
different absorption parameters
RF and Microwave Photonics in Biomedical Applications 279
a clear image that the absorber with higher absorption coefficient causes more amplitude
attenuation and phase shift.
The data in Figure 9.38 also show the benefit of employing higher-frequencymeasurements.
To make this argument more obvious, the amplitude and phase response at 144MHz
(a frequency which has quite often been reported as part of single frequency I/Q receiver)
and 1GHz (a high frequency that is being advocated by the author) are compared in Figure 9.39.
At 144MHz, the amplitude attenuation caused by the three absorbers is less than 0.1 dB and the
phase shift is less than 2.2�, whereas at 1GHz the amplitude attenuation is around 1.5 dB and
phase shifts around 9�, which proves that higher-frequencymeasurement has higher sensitivity
to detect absorption change in a turbid medium.
-1.8
-1.6
-1.4
-1.2
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
1400120010008006004002000
Frequency (MHz)
Amplitude attenuation
(dB)
Ph
ase
shif
t (º
)
Absorber1
Absorber2
Absorber3
(a)
(b)
0
2
4
6
8
10
12
1400120010008006004002000
Frequency (MHz)
Absorber1
Absorber2
Absorber3
Figure 9.38 System response to different absorbers: (a) normalized amplitude attenuation; (b)
normalized phase shift
280 Microwave Photonics: Devices and Applications
9.5.2.3 Experiment 3
The system amplitude and phase response to absorbers (e.g. India ink of a set concentration)
with different sizes of container were measured. The set-up is identical to the set-up shown in
Figure 9.37. This time, two cylindrical absorbers with the same absorption coefficient but of
different diameters (straws of 4.6mm and 5.5mm) were placed in the middle of the distance
between the transmitter fibre and the receiver fibre. The distance between these two fibres was
still 3 cm. The normalized amplitude and phase responses of these two absorbers are shown in
Figure 9.40. As expected, the 5.5mm absorber causes more amplitude attenuation and phase
shift than the 4.6mm absorber. Again, the higher-frequency measurement is more sensitive to
-1.6
-1.4
-1.2
-1
-0.8
-0.6
-0.4
-0.2
0
321
Absorber #
Amplitude attenuation
(dB)
144MHz
1GHz
(a)
0
1
2
3
4
5
6
7
8
9
10
321Absorber #
Phase shift (º)
144MHz
1GHz
(b)
Figure 9.39 Sensitivity of the measured signal to the absorption parameter of three different absorbers
at 144MHz and 1GHz: (a) amplitude change and (b) phase shift
RF and Microwave Photonics in Biomedical Applications 281
the size difference, which indicates that the higher-frequency measurement has better spatial
resolution than the low-frequencymeasurement. The ability to distinguish the 4.6mmdiameter
from the 5.5mm diameter is only observable at frequencies approaching 1000MHz in
amplitude.
9.5.3 Extraction of Phantom Optical Parameters using a Broadband System
A phantom made from a polymer whose optical parameters resemble biological tissue (i.e.
breast) is used in these experiments. The set-up is shown in Figure 9.41. The solid phantom has
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
1400120010008006004002000
Frequency (MHz)
Amplitude attenuation
(dB)
Size1: 4.6mm
Size2: 5.5mm
(a)
0
1
2
3
4
5
6
7
1400120010008006004002000
Frequency (MHz)
Phase shift
(º)
Size1: 4.6mm
Size2: 5.5mm
(b)
Figure 9.40 System response to absorbers with different sizes: (a) normalized amplitude attenuation
and (b) normalized phase shift
282 Microwave Photonics: Devices and Applications
homogeneous optical parameters, that is an absorption coefficient and a scattering coefficient.
The transmitter fibre and the receiver fibre are placed on the same surface of the solid phantom.
The amplitude and phase responses at two distances (d1¼ 1 cm, and d2¼ 1.5 cm) were
measured.
The amplitude attenuation and phase difference of the two distances have been calculated as
follows
Amplitude attenuation ðdBÞ ¼ Amplituded1ðdBÞ�Amplituded2ðdBÞPhase difference ðdegreeÞ ¼ Phased1ðdegreeÞ� Phased2ðdegreeÞ
and the results are shown in Figure 9.42. To extract the optical parameters, a Matlab program
was developed based on the relationship between the amplitude and the phase of the received
signal and optical parameters of the phantom as shown in Equations (9.31) and (9.32). The
measured data are employed in the program to extract the absorption and reduced scattering
coefficients. Measurements of relative amplitude and phase changes for two distances are
shown in Figure 9.42.
This measurement result could be used for optical parameter extraction. In a single-
frequency instrument (I/Q system), only the amplitude and phase signal related to this specific
frequency can be acquired, and hence one can extract absorption and reduced scattering
coefficients based on only one frequency. We call this kind of extraction single-frequency
extraction. However, with a frequency sweptmode using amulti-frequency instrument, such as
a network analyser, the amplitude and phase signals of multi-frequencies can be measured at
the same time. The measured results can be fitted to a theoretical model and the absorption and
scattering coefficients are extracted; we call this process broadband extraction. The advantage
of this process is that one can not only obtain information from various layers, but the
performance anomaly due to the in-frequency resonance can be averaged out or even removed.
To demonstrate the strength of this broadband frequency domain system over the CW
measurement system, an experiment is conducted using a known phantom resembling
breast tissue. In this experiment, 101 frequency points have been chosen between 100MHz
and 1000MHz and both broadband extraction and single-frequency extraction have been
performed. The optical parameters are extracted by relating the measured amplitude and
Optical transmitter
Optical receiver (APD)
ANA
Solid phantom
Fibres
Figure 9.41 Experimental set-up for solid phantom measurements
RF and Microwave Photonics in Biomedical Applications 283
phase of the scattered light from the phantom using Equations (9.31) and (9.32). The
single-frequency extraction of absorption and scattering parameters as a function of
frequency is compared and depicted in Figure 9.43 against the manufacturer�s data. Note
that the extracted single-frequency results show an average error of 35.2% for ma and
23.6% for m0s.
This extraction error is typical of the inaccuracy encountered in I/Q demodulators, when
optical loss is significant. The benefits of the broadband measurement system become more
evident when extraction is correlated over a broader bandwidth. For broadband extraction,
0
2
4
6
8
10
12
14
16
18
20
120010008006004002000
Frequency (MHz)
Amplitude attenuation
(dB)
(a)
0
5
10
15
20
25
30
35
40
45
50
120010008006004002000
Frequency (MHz)
Phase difference
(º)
(b)
Figure 9.42 Measurements of relative amplitude and phase change as a function of frequency for a solid
phantom at two distances: (a) amplitude change and (b) phase shift
284 Microwave Photonics: Devices and Applications
different frequency ranges have been selected to extract the optical coefficients, e.g. 300 MHz
range (100 400MHz, 400 700MHz, 700 1000MHz), 450MHz range (100 550MHz,
550 1000MHz) and 900MHz (100MHz 1GHz) range. All the extracted values are
summarized in Table 9.2. As shown in this table, for a larger frequency range, the extracted
values tend to be closer to the manufacturer�s values. Moreover, the accuracy of broadband
extraction is much better than single-frequency extraction.
0
0.04
0.08
0.12
0.16
0.2
0.24
0.28
900700500300100
Manufactured value
Extracted valueµa
(cm-1)
µs′(cm-1)
(a)
(b)
Frequency (MHz)
0
3
6
9
12
15
18
900700500300100
Frequency (MHz)
Manufactured value
Extracted value
Figure 9.43 Extracted ma (a) and m0s (b) by using single-frequency extraction, which results in an
average error of 35.2% for ma and 23.6% for m0s
RF and Microwave Photonics in Biomedical Applications 285
9.6 Conclusions
Microwave photonics techniques have become established for telecommunication and radar
systems in the last 20 years; however, the benefits of RF andmicrowave photonic technology in
biomedical applications have not been fully explored. This chapter has presented two separate
applications of microwave photonics to medical imaging. In particular, it is stipulated that a
higher-spatial resolution and broadband ultrasound imaging could be developed using an
optical hydrophone, where development of a fibre-optic hydrophone probe was described.
Although the probe was designed to operate at frequencies up to 100MHz it has not yet been
tested at 100MHz frequency. Once fully developed, the FO probe will constitute an effective
measurement tool allowing the need for spatial averaging corrections to be eliminated.A power
budget calculation of the fibre sensor indicated that a relatively high-power (1000mW) laser
source is essential to achieve a sufficiently high signal-to-noise ratio. Experiments validated the
analytical results and broadband-measured sensitivity of about 500mV/MPa was achieved in
the frequency range up to 60MHz without any spatial averaging correction required. This
performance at the moment supersedes the commercially available needle hydrophones.
Microwave photonics techniques are also applied to the eventual development of a
broadband NIR spectroscopy system to achieve real-time imaging of biological tissues with
a millimeter scale spatial resolution. In particular, an active driver system, which was
developed for a TO-can based laser system operating at wavelengths of 677 nm, 785 nm,
830 nm and 977 nm, was extended to C-submount based optical transmitters. In particular, an
active laser driver using a C-submount laser diode at 850 nm was developed using an active
laser driver for a high-power laser diode. The laser driver is realized using hybrid techniques.
The optical system was designed for a flat frequency response up to 3GHz, where spatial
resolution of a few millimeters is predicted. Both broadband extraction and single-frequency
extraction have been performed to extract optical parameters ma and ms for a phantom that
resembles breast tissue as a turbid media. Results show that the accuracy of broadband
extraction is much better than single-frequency extraction.
Table 9.2 Broadband extraction and percentage error results.
Extracted
values (cm�1)
Manufacturer�svalues (cm�1)
Error
(cm�1)
Error
(%)
Frequency range from 100MHz 400MHz ma 0.0488 0.045 0.0038 8.44
m0s 11.0769 10 1.0769 10.77
Frequency range from 400MHz 700MHz ma 0.0487 0.045 0.0037 8.22
m0s 9.0779 10 0.9221 9.22
Frequency range from 700MHz 1GHz ma 0.0436 0.045 0.0014 3.11
m0s 9.4321 10 0.5679 5.68
Frequency range from 100MHz 500MHz ma 0.0499 0.045 0.0049 10.89
m0s 10.2672 10 0.2672 2.67
Frequency range from 500MHz 1GHz ma 0.0479 0.045 0.0024 5.33
m0s 9.4655 10 0.5345 5.35
Frequency range from 100MHz 1GHz ma 0.0488 0.045 0.0038 8.44
m0s 9.6711 10 0.329 3.29
286 Microwave Photonics: Devices and Applications
Acknowledgements
Material covered in this chapter is due to the contribution of a number of colleagues, past and
present graduate students. The intellectual influence of, and fruitful discussions with, my
colleagues Prof. Emeritus Britton Chance, Prof. Peter Lewin and Prof. Kambiz Pourrezaei
from the School of Biomedical Engineering, Drexel University, is greatly recognized. Finally,
without the tireless work ofmy students, the required understanding and scientific research that
has now led to the writing of this chapter would not have been feasible. I would like to
acknowledge the contributions of Dr Sumet Umchid, Chenpeng Mu, Rupa Gopinath, Karthik
Srinivasan andDo-YoonKim to this chapter. Finally, support of theNational Institute of Health
(NIH) is greatly appreciated.
References
[1] Papers in Front End Opto-Electronics for Future Radio Communications, Workshop WM1, 2004 IEEE Radio
and Wireless Conference, September 2004.
[2] Papers in Beamforming Techniques for Active Phased Array Antennas Based Communication Satellites, in the
IEEE 1999 International Microwave Symposium, Anaheim, CA, USA, June 1999.
[3] M. Berson, J. M. Gr�egoire, F. Gens, J. Rateau, F. Jamet, L. Vaillant, F. Tranquart and L. Pourcelot, ‘‘High
frequency (20 MHz) ultrasonic devices: advantages and applications’’, European Journal of Ultrasound,
vol. 10(1), pp. 53–63, 1999.
[4] AIUM, ‘‘Acoustic output measurement standard for diagnostic ultrasound equipment’’, Laurel, MD, USA, 1998.
[5] FDA, Revised FDA 510(k) ‘‘Information for manufactures seeking marketing clearance of diagnostic ultrasound
systems and transducers’’, September 1997.
[6] E. G. Radulescu, P. A. Lewin, J. Wojcik and A. Nowicki, ‘‘Calibration of ultrasonic hydrophone probes up to
100MHz using time gating frequency analysis and finite amplitude waves’’, Ultrasonics, vol. 41, pp. 247–254,
2003.
[7] R. A. Smith, ‘‘Are hydrophones of diameter 0.5 mm small enough to characterize diagnostic ultrasound
equipment?’’ Phys. Med. Biol., vol. 34(11), pp. 1593–1607, 1989.
[8] G. R. Harris, ‘‘Medical ultrasound exposure measurements: update on devices, methods, and problems’’,
Ultrasonics Symposium, pp. 1341–1352 (1999).
[9] J. Staudenraus and W. Eisenmenger, ‘‘Fibre-optic probe hydrophone for ultrasonic and shock-wave measure-
ments in water2, Ultrasonics, vol. 31(4), pp. 267–273, 1993.
[10] E.M. Sevick, B. Chance, J. Leigh and S. Nioka, ‘‘Quantitation of Time-resolved and Frequency-resolved Optical
Spectra for the Determination of Tissue Oxygenation’’, Anal. Biochem., vol. 195(2), pp. 330–351, 1991.
[11] B. Beauvoit, T. Kitai and B. Chance, ‘‘Contribution of the Mitochondrial Compartment to the Optical Properties
of the Rat Liver: A Theoretical and Practical Approach’’, Biophys. J., vol. 67(6), pp. 2501–2510, 1994.
[12] S. Thornsen andD. Tatman, ‘‘Physiological and Pathological Factors ofHumanBreastDisease that Can Influence
Optical Diagnosis2, Ann. N. Y, Acad. Sci., vol. 838, pp. 171–193, 1998.
[13] T. H. Pham, O. Coquoz, J. B. Fishkin, E. Anderson and B. Tromberg, ‘‘Broad Bandwidth Frequency Domain
Instrument for Quantitative Tissue Optical Spectroscopy’’, Rev. Sci. Inst., vol. 71, pp. 2500–2513, 2000.
[14] B. J. Tromberg, B, O. Coquoz, J. B. Fishkin, T. Pham, E. R. Anderson, J. Buffer et al., ‘‘Non-invasive
Measurements of Breast Tissue Optical Properties Using Frequency-domain Photon Migration’’, Philos. Trans.
R. Soc. Lond. B. Biological Sciences, vol. 352(1354), pp. 661–668, 1997.
[15] A. Duncan, T. L. Whitlock, M. Cope and D. T. Delpy, ‘‘A Multi-wavelength, Wideband, Intensity Modulated
Optical Spectrometer for Near Infrared Spectroscopy and Imaging’’, SPIE, vol. 1888, pp. 248–257, 1993.
[16] J. E. Parsons, C. A. Cain and J. B. Fowlkes, ‘‘Cost-effective assembly of a basic fiber-optic hydrophone for
measurementofhigh-amplitudetherapeuticultrasoundfields’’,J.Acoust.Soc.Am.,vol.119, pp.1432–1440,2006.
[17] W. B. Spillman, Jr. and R. L. Gravel, ‘‘Moving fiber-optic hydrophone’’,Optics Letters, vol. 5, No. 1 pp. 30–31,
January 1980.
[18] W. B. Spillman, Jr., ‘‘Multimode fiber-optic hydrophone based on a schlieren technique’’,Appl. Opt., vol. 20, pp.
465–470, 1981.
RF and Microwave Photonics in Biomedical Applications 287
[19] W. B. Spillman, Jr. and D. H. McMahon, ‘‘Frustrated-total-internal-reflection multimode fiber-optic hydro-
phone’’, Appl. Opt., vol. 19, pp. 113–117, 1980.
[20] G. B. Hocker, ‘‘Fiber-optic sensing of pressure and temperature’’,Appl. Opt., vol. 18, No. 9, pp. 1445–1448, 1978.
[21] A.S. Daryoush, H.W. Li, M. Kaba, G. Bouwmans, D. Decoster, J. Chazelas and F. Deborgies, ‘‘Passively
Temperature Stable Opto-electronic Oscillators Employing Photonic Crystal Fibers?’’, Journal of the European
Microwave Association, vol. 3, Issue 3, pp. 201–209, September 2007.
[22] A. M. Yurek, ‘‘Status of Fiber Optic Acoustic Sensing’’,Optical Fiber Sensors Conference, vol. 8, pp. 338–341,
1992.
[23] N. Lagakos and J.A. Bucaro, ‘‘Linearly Configured Embedded Fiber-optic Acoustic Sensor’’, Journal of
Lightwave Technology, vol. 11, pp. 639–642, April 1993.
[24] G.E.McDearmon, ‘‘TheoreticalAnalysis of a Push-Pull FiberOpticHydrophone’’, J. LightwaveTech., vol. LT-5,
pp. 647–652, 1987.
[25] P. C. Beard,A.M.Hurrell, andT.N.Mills, ‘‘Characterization of a polymer film optical fiber hydrophone for use in
the range 1 to 20 MHz: A comparison with PVDF needle and membrane hydrophones’’, IEEE Transactions on
Ultrasonics, Ferroelectrics, and Frequency Control, vol. 47, pp. 256–264, 2000.
[26] P. Morris, A. Hurrell and P. Beard, ‘‘Development of 50 MHz Fabry-Perot type fiber optic hydrophone for the
characterization of medical ultrasound fields’’, Proceedings of the Institute of Acoutics, vol. 18, pp. 717–725,
2006.
[27] V. Wilkens, C. Koch, and W. Molkenstruck, ‘‘Frequency response of a fiber-optic dielectric multilayer
hydrophone’’, Ultrasonics Symposium, IEEE, vol. 2, pp. 1113–1116, 2000.
[28] J. A. Bucaro and T. R. Hickman, ‘‘Measurement of sensitivity of optical fibers for acoustic detection’’, Applied
Optics, vol. 18, No. 6, pp. 938–940, 1979.
[29] P. Fomitchov and S. Krishnaswamy, ‘‘Response of a Fiber Bragg-Grating Ultrasound Sensor’’, Optical
Engineering, vol. 42, No 4, pp. 956–963, 2003.
[30] Bai-Ou Guan, Hwa-Yaw Tam, Sien-Ting Lau and Helen L. W. Chan, ‘‘Ultrasonic hydrophone Based on
Distributed Bragg Reflector Fiber Laser’’, IEEE Photonics Technology Letters, vol. 17, no. 1, pp. 169–171,
January 2005.
[31] R. Gpoinath, P. Arora, G. Gandhi, L. Bansal, A.S. Daryoush, P.A. Lewin and M. El-Sherif, ‘‘Broadband Fiber
Optic Hydrophone Sensors for Ultrasound Applications’’, MWP 2008 Proc., International Topical Meeting on
Microwave Photonics, Gold Coast, Australia, 2008.
[32] K. Srinivasan,‘‘Noise Cancelled Optical Receivers in Fiber Optic Hydrophone up to 100MHz?’’, M.S. Thesis,
Drexel University, USA, 2007.
[33] Sumet Umchid,‘‘Development of Calibration Techniques for Ultrasonic Hydrophone Probes in the Frequency
Range from 1 to 100 MHz’’, Ph.D. Dissertation, Drexel University, Philadelphia, PA, USA, 2007.
[34] P.A. Lewin, C.Mu, S.Umchid,A.Daryoush andM.El-Sherif, ‘‘Acousto-optic, PointReceiverHydrophoneProbe
for Operation up to 100 MHz’’, Ultrasonics, vol. 43, Issue 10, pp. 815–821, December 2005.
[35] M. Cutler, ‘‘Transillumination of the breast’’, Surg. Gynecol. Obstet., vol. 48, pp. 721–727, 1929.
[36] E. M. Sevick, B. Chance, J. Leigh and S. Nioka, ‘‘Quantitation of time-resolved and frequency-resolved optical
spectra for the determination of tissue oxygenation’’, Anal. Biochem., vol. 195(2), pp. 330–351, 1991.
[37] B. Beauvoit, T. Kitaiand B. Chance, ‘‘Contribution of the mitochondrial compartment to the optical properties of
the rat liver: a theoretical and practical approach’’, Biophys. J., vol. 67(6), pp. 2501–2510, 1994.
[38] S. Thornsen and D. Tatman, ‘‘Physiological and pathological factors of human breast disease that can influence
optical diagnosis’’, Ann. N. Y, Acad. Sci,. vol. 838, pp. 171–193, 1998.
[39] A. Zourabian, A. Siegel et al., ‘‘Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse
oximetry’’, J. Biomed. Opt., vol. 5(4), pp. 391–405, 2000.
[40] S. R. Arridge, ’’Optical tomography in medical imaging’’, Inverse Probl., vol. 15, pp. 41–93, 1999.
[41] D. A. Boas et al., ‘‘The accuracy of near imfrared spectroscopy and imaging during focal changes in cerebral
hemodynamics’’, NeuroImages, vol. 13, pp. 76–90, 2001.
[42] Y. Lin et al., ‘‘Noninvasive, low-noise, fast imaging of blood volume and deoxygenation changes inmuscles using
light-emitting diode continuous-wave imager’’, Rev. Sci. Instrum., vol. 73(8), pp. 3065–3074, 2002.
[43] Y. Yang, H. Liu, X. Li and B. Chance, ‘‘Low-cost frequency-domain photon migration instrument for tissue
spectroscopy, oximetry, and imaging’’, Opt. Eng., vol. 36(5), pp. 1562–1569, 1997.
[44] M. S. Patterson et al., ‘‘Time resolved reflectance and transmittance for the non-invasive measurement of tissue
optical properties’’, Applied Optics, vol. 28, pp. 2331–2336, 1989.
288 Microwave Photonics: Devices and Applications
[45] S. L. Jacques, ‘‘Time-resolved reflectance spectroscopy in turbid tissues’’, IEEETrans. Biomed.Eng., vol. 36, pp.
1155–1161, 1989.
[46] B. C. Wilson and S. L. Jacques, ‘‘Optical reflectance and transmittance of tissues: principles and applications’’,
IEEE J. Quantum Electron., vol. 26, pp. 2186–2199, 1990.
[47] G. Eason et al., ‘‘The theory of the backscattering of light by blood’’. J. Phys., vol. 11, pp. 1463–1479, 1978.
[48] D. Grosenick, H. Wabnitz, H. Rinneberg, K. T. Moesta and P. Schlag, ‘‘Development of a time-domain optical
mammograph and first in vivo applications’’, Appl. Opt., vol. 38, pp. 2927–2943, 1999.
[49] B. Chance et al., ‘‘Phase measurement of light absorption and scatter in human tissue’’, Rev. Sci. Instrum.,
vol. 69(10), pp. 3457–3481, 1998.
[50] T. Tu et al., ‘‘Analysis on performance and optimization of frequency-domain near-infrared instruments’’,
J. Biom. Opt., vol. 7(4), pp. 643–649, 2002.
[51] Y. Chen, C. P. Mu, X. Intes and B. Chance, ’’Signal-to-noise analysis for detection sensitivity of small absorbing
heterogeneity in turbidmediawith single-source anddual-interfering-source’’,Opt. Express, vol. 9, pp. 212–224,
2001.
[52] M. Erickson et al., ‘‘Comparison of sensitivity for single-source and dual-interfering-source configurations in
optical diffusion imaging2, J. opt. Soc. Am. A, vol. 14(11), pp. 3083–3092, 1997.
[53] A. Knuttel, J. M. Schmitt and J. R. Knutson, ’’Improvement of spatial resolution in reflectance near-infrared
imaging by laser-beam interference’’, in ‘Time-Resolved Laser Spectroscopy in Biochemistry III’, J. R. Lakowicz
(Ed.), Proc. SPIE vol. 1640, pp. 405–416, 1992.
[54] A. Duncan, T. L. Whitlock, M. Cope and D. T. Delpy, ‘‘A multi-wavelength, wideband, intensity modulated
optical spectrometer for near infrared spectroscopy and imaging’’, SPIE, vol. 1888, pp. 248–257, 1993.
[55] C. Mu, D-Y Kim, U. Sunar, K. Pourrezaei and A. Daryoush, ‘‘Multi-wavelength NIR system for spectroscopy of
biomedical tissues’’, MWP 2003 Proc., International Topical Meeting on Microwave Photonics, Budapest,
Hungary, pp. 275–278, 2003.
[56] DoyoonKim,‘‘Design of laser diode driver for diffused photon near infrared imaging applications’’,M.S. Thesis,
Drexel University, Philadelphia, PA, USA, 2002.
RF and Microwave Photonics in Biomedical Applications 289