MERCEM300ZMERCURY GAS ANALYZERADVANCED MEASUREMENT OF MERCURY IN FLUE GASES

Extractive gas analyzers

Page 2: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K 8013839/2014-08Subject to change without notice

22

Challenges of measuring mercuryContinuously measuring mercury in flue gases presents a number of particular challenges for measuring systems. Firstly, the Hg concentrations to be detected are smaller than other flue gas components by a factor of 1,000. Limit values in the lower μg range must be monitored with high accuracy. Further-more, any chemically bonded mercury must be converted into elemental mercury. Mercury analysis has already been carried out successfully for many years in waste incineration plants. In cement plants and power plants, the situation is different.

In plants where secondary fuels are used for power generation, Hg emissions also have to be monitored accurately as well.

Here too, this is done using continuous analyzer systems. The prevailing system conditions in cement and power plants, namely high levels of dust or high concentrations of interfer-ence components such as SO2, CO, and CO2 mean that the conversion and measuring processes are not always reliable.

The SICK solution SICK offers the best solution for creating measurement certainty and for avoiding long downtime periods, high mainte-nance requirements, and therefore high operating costs for the operator: MERCEM300Z for high-precision mercury measuring.

PROVEN MEASUREMENT TECHNOLOGY, IDEAL FOR REFERENCE PURPOSES

Reliably measuring mercury in flue gases is difficult and places a great deal of demand on measuring technology. The innovative MERCEM300Z mercury gas analyzer from SICK is able to take reliable measurements in situations where other measuring systems have had to give up.

MERCEM300Z measures using hot extraction: The flue gas is extracted using a sampling probe and is transported to the analyzer via a sample gas line. All components in contact with the sample gas, such as the probe, sample gas line, and converter in the analyzer are heated above the dew point.

Sample gas line

Sampling probe

Page 3: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K8013839/2014-08Subject to change without notice

33

Reliable measuring technology thanks to a combination of conversion and photometers With the MERCEM300Z gas analyzer, SICK has set a new stan-dard when it comes to analyzing mercury. No mercury leaks, no chemicals used – instead, we apply hot measuring technology, which has been proven over many years.

Measuring without moving parts – minimal maintenanceThe production of the measuring gas is based on the low-main-tenance ejector principle. What's more, taking the measure-ment itself does not require any moving parts. MERCEM300Z: The mercury gas analyzer characterized by its very low mainte-nance requirements and long maintenance intervals.

Drift check without expensive test gasAt the push of a button, the MERCEM300Z carries out a quick and effective drift check with an internal adjustment cell. Using test gas is not necessary.

Large variety of certified measuring rangesThe certified measuring ranges of the MERCEM300Z start from an extremely low 0 ...10 µg/m3 reach 0 ... 45/100 µg/m3, which is the current requirement, and go right up to the unusu-ally high 0 ... 1,000 µg/m3. This range covers the requirements for both emissions and process measuring.

Internal test gas generatorTo avoid the need for expensive test gases completely, an optional integrated test gas generator is available. With up to 3 predefined concentrations, this enables an automatic reference point check and adjustment (QAL3). This provides measurement certainty and means that money does not need to be spent on specialist technicians and costly, external testing equipment.

Page 4: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K 8013839/2014-08Subject to change without notice

44

MERCEM300Z MODELS

MERCEM300Z Indoor

For measurements in air-conditioned rooms at temperatures from +5 to +35 °C

MERCEM300Z

For measurements outdoors at temperatures from –20 to +50 °C

Optimized industrial design for measurements in harsh ambient conditions near to where the sample is taken – with integrated air conditioning.• Ambient temperature: –20 °C ... +50 °C.

Compact standard design for the installation in air-condi-tioned rooms. Possible to arrange additional cabinets on the left-hand side.• Ambient temperature: +5 °C ... +35 °C.

Page 5: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K8013839/2014-08Subject to change without notice

55

MERCEM300Z – proven industrial design

Hg conversion in high-temperature converterOwing to the principles involved, mercury can only be measured in its elemental state. Therefore, it is essential to convert all mercury compounds to their elemental states before actually determining the Hg concentration. And this is where the key advantages of the MERCEM300Z really lie: The conversion is carried out using a purely thermal method, simply by using high temperatures of about 1,000 °C – no chemicals or non-dura-ble catalyst materials are required.

Photometric measurement using the Zeeman effectThe MERCEM300Z measures the Hg concentration in flue gas photometri-cally by means of Zeeman atomic ab-sorption spectroscopy (Zeeman AAS): An Hg discharge lamp emits light in the element-specific wavelength. A strong magnetic field generated around the lamp splits the spectral line into several components which form the measuring and reference wavelengths simultaneously (Zeeman effect). The advantage here is that there are no moving parts and that cross sensitivi-ties, lamp deterioration, and contami-nation are compensated for directly.

Patented direct measurement: Photometric measurement direct in the hot converterThe highlight of the MERCEM300Z: The photometric determination of the mer-cury content occurs directly in the hot converter without the need to change the sample further. This makes the process fast and continuous, and ensures that the conversion rate is steady. This means that the overall Hg concentration in the flue gas can be determined reliably. Chemicals and catalyst materials are not necessary. This direct measurement, which is patented by SICK, makes the MERCEM300Z a truly unique mercury gas analyzer. What's more, the resulting benefits make it exceptionally valuable for the operator.

High-temperature converter

Ejector

Optical unit

Electronics

Automatic test gas generator (option)

Page 6: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K 8013839/2014-08Subject to change without notice

66

MERCEM300Z – PREDESTINED FOR FLUE GAS MEASURING

Mercury is attracting more and more attention globally when it comes to monitoring emissions. Plant operators are faced with increasingly large challenges thanks to new annual limits in Europe and new legislation on measuring mercury in cement plants and coal-fired power plants in the USA.

In many cases, the permitted limits for flue gases are already below 10 µg/Nm3. Power plants, incineration plants, or cement kilns, which use fossil fuels or household goods as a fuel must all take responsibility.

The fuel used, e.g., lignite or bituminous coal in power plants or waste materials in waste incineration plants, is almost solely responsible for mercury entering the process. While in power plants, the level of mercury emissions is almost constant, these emissions fluctuate greatly at waste incineration plants due to the non-uniform distribution of mercury in the fuel.

However, the same requirement applies to all types of plants: In order to comply with the new limits, it is vital to invest in an opti-mized gas cleaning system and in even more accurate emission measuring technology. The flue gas cleaning units in the major-ity of existing plants do not have a cleaning stage designed spe-cifically for mercury. As a result, an increasing number of other options are used to discharge mercury from the process, e.g., adding doped activated carbon upstream of the electrostatic precipitator or adding precipitating agents in the washer.Measuring the mercury in the raw gas, upstream of the washer or the electrostatic precipitator, offers significant benefits.

Page 7: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K8013839/2014-08Subject to change without notice

77

Plant safetyIf high concentrations of mercury are present in the process over an extended period of time, this can cause the flue gas washer to become overloaded with mercury and can ultimately lead to the entire plant becoming contaminated. In the worst case scenario, this could bring the whole plant to a standstill. This can be avoided if high concentrations of mercury are detected in the raw gas using an accurate measurement at an early stage and the appropriate countermeasures are imple-mented.

Cost optimizationUsing activated carbon and precipitating agents to remove mercury from the process is expensive. Excessive amounts of both of these are often added to the process for safety reasons. By measuring the amount of mercury in the raw gas, it is pos-sible to accurately determine the required quantity of activated carbon and precipitating agents and therefore save money.

Complying with the Hg limits in clean gasTo be compliant with the new, lower emissions limits, it is vital to respond at process level. Measuring mercury in the raw gas and implementing the appropriate countermeasures in the process can ensure that the emissions limits are not exceeded.

Higher requirements in raw gasThe requirements for continuous mercury measuring systems in raw gas are significantly higher in comparison to measuring emissions:

• Higher dust loads up in flue gas• Higher concentrations of interference components (such as

SO2 or HCl)• Accurate detection of mercury concentration peaks (Hg

peaks) up to the mg range• Fast response times to be able to react to Hg peaks.• No carry-over effects

In the gas conditioning process, the mercury must be converted into metallic mercury safely and reliably as, if this conversion is not done correctly, this leads to results which are too low and therefore a less accurate measurement of the total mercury. At the same time, suitable processes must be implemented in order to compensate for the increased gas concentration of the cross sensitivity components. This is where an almost continu-ous measurement with an additional amalgamation process step is also not enough as it does not respond to possible Hg peaks quickly enough. The MERCEM300Z mercury gas analyzer from SICK offers considerable benefits thanks to its use of Zeeman absorption spectroscopy (Zeeman AAS). It is even pos-sible to compensate for high concentrations of contaminants continuously and immediately by using this technology. The pat-ented direct measurement enables immediate detection of Hg raw gas concentrations in the thermal converter, meaning that you can respond to fluctuating concentrations at an early stage.

For some operators, it is currently sufficient to measure the trends for Hg concentrations in the raw gas. A key factor here is the simultaneous monitoring of several plants. SICK also provides the ideal solution for this: With the optional measuring point switchover, up to two measuring points can be operated with one gas analyzer. It is possible to switch between the measuring points manually or in parametrized cycles if this is requested by the customer.

Page 8: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K 8013839/2014-08Subject to change without notice

88

MERCEM300Z ExTRACTIVE GAS ANALYZERS

Product descriptionThe MERCEM300Z mercury measuring system monitors Hg emissions in flue gases with high reliability within the smallest measuring ranges. Due to its rugged housing, MERCEM300Z is suit-

able for use in harsh industrial environ-ments. The complete extractive system is designed to meet national and interna-tional regulations and directives.

At a glance• Accurate measurement of “total

mercury” directly in a thermal converter (patented)

• Measuring operation without using consumables

• Very low maintenance gas sampling using an ejector pump – no moving parts

• Integrated adjustment cell for automatic drift correction

• Automatic adjustment of the entire measuring system with a built-in test gas generator (optional)

• Modular design with the entire system

Your benefits• Reliable results of the actual measur-

ing values of elemental Hg and Hg compounds in gases

• Very low operating expenses• Minimum maintenance expenditure• Long-term stability minimizes techni-

cian time requirements due to self-adjusting measuring system

• Measuring certainty using the fully automated adjustment with test gas

• Convenient and fast access for easy service and user-friendly remote diagnosis

MERCM300ZExTRACTIVE GAS ANALYZERSMERCEM300ZSubject to change without notice

.

INNOVATIVE MEASUREMENT OF MERCURY IN FLUE GASES

Additional information

Fields of application . . . . . . . . . . . . . . .9

Detailed technical data . . . . . . . . . . . .9

Ordering information . . . . . . . . . . . . .10

Dimensional drawings . . . . . . . . . . . .10

15267certified

14181certified

- www.mysick.com/en/MERCEM300ZFor more information, just enter the link and get direct access to technical data, CAD design models, operating instructions, software, application examples and much more.

Page 9: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K8013839/2014-08Subject to change without notice

99

ExTRACTIVE GAS ANALYZERS MERCEM300Z

Fields of application• Emission monitoring in incineration of domestic and

hazardous waste• Emission monitoring in combustion of sewage sludge or

hospital waste

• Emission monitoring in cement plants• Emission monitoring in power stations

Detailed technical dataThe exact device specifications and performance data of the product may deviate from the information provided here, and depend on the applica-tion in which the product is being used and the relevant customer specifications.

System

Measured values Hg

Performance tested measurands Hg

Measurement principles Zeeman atomic absorption spectroscopy

Gas flow rate 150 l/h ... 400 l/h

Measuring ranges

Hg 0 ... 1 ppb / 0 ... 100 ppb

Certified measuring ranges

Hg 0 ... 10 µg/m³ / 0 ... 45 µg/m³ / 0 ... 100 µg/m³ / 0 ... 1,000 µg/m³

Sensitivity drift < 3 % within the maintenance interval, relative to measuring range full scale

Zero point drift < 3 % within the maintenance interval, relative to measuring range full scale

Detection limit < 2 % relative to measuring range end value

Process temperature ≤ +1,300 °C

Process pressure 850 hPa ... 1,100 hPa

Ambient temperature MERCEM300Z: –20 °C ... +50 °CMERCEM300Z Indoor: +5 °C ... +35 °C

Storage temperature –20 °C ... +40 °C

Ambient pressure 850 hPa ... 1,100 hPa

Ambient humidity ≤ 80 % Relative humidity; non-condensing

Conformities EN 14181EN 152672001/80/EC2000/76/EC

Electrical safety CE

Enclosure rating MERCEM300Z: IP 55MERCEM300Z Indoor: IP 43

Analog outputs 2 outputs: 0/4 ... 20 mA, 500 Ω Electrically isolated; additional outputs if using I/O modules (option)

Digital outputs 4 relay contacts: 50 V, 4 A Electrically isolated; additional outputs if using I/O modules (option)

Digital inputs 4 inputs: 24 V, 0.3 A Electrically isolated; additional outputs if using I/O modules (option)

Interfaces CAN (internal system bus)Ethernet

Bus protocol Modbus TCP

Indication LC display

Input Functional keys

Operation Via LC-display or software SOPAS ET

Model Aluminium cabinet

Page 10: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K 8013839/2014-08Subject to change without notice

1 01 0

MERCEM300Z ExTRACTIVE GAS ANALYZERS

Dimensions (W x H x D) 1,038 mm x 1,744 mm x 744 mm (MERCE300Z with cooling device)806 mm x 2,156.5 mm x 605 mm (MERCEM300Z Indoor)

Weight ≤ ± 250 kg

Power supply

Power consumption Without integrated test gas generator ≤ 2,100 WWith integrated test gas generator ≤ 3,100 W

Auxiliaries

Instrument air: ≤ 2,500 l/h 6 ... 7 bar; particle size max. 1 µm; oil content max. 0.1 mg/m³; pressure dew point max. –30 °C

Test gas: ≤ 500 l/h Max. 0.5 bar; accuracy ±2%; water content 5 ... 30 vol.-%

Sample connections Sample gas inlet: Swagelok 6 mm

Auxiliary connections Test gas: Swagelok 6 mmInstrument air: Swagelok 10 mmExhaust gas outlet: Swagelok 10 mm

Corrective functions Internal adjustment cuvetteIntegrated test gas generator (option)Drift correction (QAL3) and check of the complete system

Integrated components Integrated instrument air conditioning (option)Integrated test gas generator (option)

Ordering informationOur regional sales organization will help you to select the optimum device configuration.

Dimensional drawings (Dimensions in mm (inch))

MERCEM300Z system

744 (29.29)

196(7.72)

351 (13.82)

1,74

4 (6

8.66

)

1,43

8 (5

6.61

)

1,00

8 (3

9.69

)

1,55

1 (6

1.06

)

551

(21.

69)

100

(3.9

4)

1,038 (40.87)

Page 11: MERCEM300Z MERCURY GAS ANALYZER - SICK

M E R C M 3 0 0 Z | S I C K8013839/2014-08Subject to change without notice

1 11 1

ExTRACTIVE GAS ANALYZERS MERCEM300Z

Product

Applications

Literature

Service

Accessories

Software

Search online quickly and safely – with the SICK “Finders”

Efficiency – with the e-commerce tools from SICK

Product Finder: We can help you to quickly target the product that best matches your application.

Applications Finder: Select the application description on the basis of the challenge posed, industrial sector, or product group.

Literature Finder: Go directly to the operating instructions, technical information, and other literature on all aspects of products from SICK.

These and other “Finders” at - www.mysick.com

Find out prices and availability: Determine the price and possible delivery date of your desired product simply and quickly at any time.

Request or view a quote: You can have a quote generated online here. Every quote is confirmed to you via e-mail.

Order online: You can go through the ordering process in just a few steps.

WWW.MYSICK.COM – SEARCH ONLINE AND ORDER

MERCEM300Z Indoor system

2,15

6.5

(84.

90)

1,832 (72.13)

1,427 (56.18)

994 (39.13)944 (37.17)900 (35.43)

683 (26.89)643 (25.31)603 (23.74)563 (22.17)523 (20.59)483 (19.02)443 (17.44)

100 (3.94)

0

605 (23.82) 806 (31.73)

173(6.81)

303(11.93)

72(2.83)

Page 12: MERCEM300Z MERCURY GAS ANALYZER - SICK

SICK AG | Waldkirch | Germany | www.sick.com

SICK AT A GLANCESICK is a leading manufacturer of intelligent sensors and sensor solutions for industrial applications. With more than 6,500 employees and over 50 subsidiaries and equity investments as well as numerous repre-sentative offices worldwide, we are always close to our customers. A unique range of products and services creates the perfect basis for controlling processes securely and efficiently, protecting individuals from ac-cidents and preventing damage to the environment.

We have extensive experience in various industries and understand their processes and requirements. With intelligent sensors, we can deliver exactly what our customers need. In application centers in Europe, Asia and North America, system solutions are tested and optimized in accordance with customer specifications. All this makes us a reliable supplier and development partner.

Comprehensive services round out our offering: SICK LifeTime Services provide support throughout the machine life cycle and ensure safety and productivity.

For us, that is “Sensor Intelligence.”

Worldwide presence:Australia, Austria, Belgium/Luxembourg, Brazil, Czech Republic, Canada, China, Denmark, Finland, France, Germany, Great Britain, Hungary, India, Israel, Italy, Japan, Mexico, Netherlands, Norway, Poland, Romania, Russia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, USA

Detailed addresses and additional representatives - www.sick.com

8013

839/

2014

-08

∙ 3PM

S/IT

L (2

015-

02) ∙

Pre

USm

od e

n43