medical oxygen sensors general specifi cations · 2 3 medical oxygen sensors technical specifi...

2
EnviteC-Wismar GmbH a Honeywell Company Internet: E-mail: Alter Holzhafen 18, 23966 Wismar, Germany www.envitec.com [email protected] Phone: +49 (0) 3841 / 360-1 Phone: +49 (0) 3841 / 360-200 Sales Fax: +49 (0) 3841 / 360-222 Medical Oxygen Sensors general specifications Widest range of compatible sensors for leading manufactures in the world. Measurement range 0-100 % oxygen Accuracy and repeatability < 1 % vol. O 2 when calibrated at 100 % O 2 Zero offset < 0.5 % vol. O 2 in 100% N 2 , applied 5 minutes Linearity error < 3 % relative Cross interference Meets EN ISO 21647 requirements Influence of humidity - 0.03 % rel. per % RH at 25°C Influence of pressure proportional to change in oxygen partial pressure Influence of mechanical shock < 1% relative after a fall from 1m Operating temperature 0° C to 50° C Temperature compensation built-in NTC compensation (depends on type) Effect of temperature between +25 °C and +40 °C: 3 % relative error compensation (steady state) between 0 °C and +50 °C: 8 % relative error Operating humidity 0-99 % RH non condensing Long term output drift < 1 % vol oxygen per month typically < - 15 % relative over lifetime Storage temperature -20 to +50 °C Recommended storage +5 to +15 °C Recommended load 10 kOhms Warm-up time < 30 minutes, after replacement of sensor Weight approximately 28 grams approximately 43 grams OOM107 series Meet EN ISO 21647, Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003 All specifications are applicable at standard conditions: 1013 hPa, 25 °C dry ambient air Medical Oxygen Sensors for life supporting systems Part-No. 001-33-00000011-b • 10/2008 • Technical information is subject to change without notice! © 2008 Honeywell International Inc. oxygen_sensors_10_2008_final.ind1-2 1-2 oxygen_sensors_10_2008_final.ind1-2 1-2 05.11.2008 12:16:08 Uhr 05.11.2008 12:16:08 Uhr

Upload: phunghuong

Post on 15-Jul-2019

222 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Medical Oxygen Sensors general specifi cations · 2 3 Medical Oxygen Sensors technical specifi cations Oxygen sensor part number Output signal in air Response time T 90% Nominal sensor

EnviteC-Wismar GmbH

a Honeywell Company Internet: E-mail:

Alter Holzhafen 18, 23966 Wismar, Germany www.envitec.com [email protected]: +49 (0) 3841 / 360-1 Phone: +49 (0) 3841 / 360-200 Sales Fax: +49 (0) 3841 / 360-222

Medical Oxygen Sensors general specifi cations

Widest range of compatible sensors for

leading manufactures in the world.

Measurement range 0-100 % oxygenAccuracy and repeatability < 1 % vol. O2 when calibrated at 100 % O2Zero offset < 0.5 % vol. O2 in 100% N2, applied 5 minutesLinearity error < 3 % relativeCross interference Meets EN ISO 21647 requirements Influence of humidity - 0.03 % rel. per % RH at 25°C Influence of pressure proportional to change in oxygen partial pressureInfluence of mechanical shock < 1% relative after a fall from 1mOperating temperature 0° C to 50° CTemperature compensation built-in NTC compensation (depends on type)Effect of temperature between +25 °C and +40 °C: 3 % relative errorcompensation (steady state) between 0 °C and +50 °C: 8 % relative errorOperating humidity 0-99 % RH non condensingLong term output drift < 1 % vol oxygen per month typically < - 15 % relative over lifetime

Storage temperature -20 to +50 °C

Recommended storage +5 to +15 °C

Recommended load ≥ 10 kOhms

Warm-up time < 30 minutes, after replacement of sensor

Weight approximately 28 grams approximately 43 grams OOM107 seriesMeet EN ISO 21647, Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003All specifications are applicable at standard conditions: 1013 hPa, 25 °C dry ambient air

Medical Oxygen Sensors for life supporting systems

Par

t-N

o. 0

01-3

3-00

0000

11-b

• 1

0/20

08 •

Tec

hnic

al in

form

atio

n is

sub

ject

to c

hang

e w

ithou

t not

ice!

©

200

8 H

oney

wel

l Int

erna

tiona

l Inc

.

oxygen_sensors_10_2008_final.ind1-2 1-2oxygen_sensors_10_2008_final.ind1-2 1-2 05.11.2008 12:16:08 Uhr05.11.2008 12:16:08 Uhr

Page 2: Medical Oxygen Sensors general specifi cations · 2 3 Medical Oxygen Sensors technical specifi cations Oxygen sensor part number Output signal in air Response time T 90% Nominal sensor

2 www.envitec.com 3

Medical Oxygen Sensors technical specifi cations

Oxygen sensorpart number Output signal in air Response time

T 90% Nominal sensor life-

time Electrical interface

OOM101 46μA – 63μAno temperature compensation < 12 seconds ≥ 500 000 % volume

oxygen hours gold plated slip rings

OOM102 9mV – 14mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours3 pin Molex® con-

nector

OOM102-1 9mV – 14mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours3.5mm mono Jack

OOM103 9mV – 13mVtemperature compensated < 5 seconds ≥ 500 000 % volume

oxygen hours3 pin Molex® con-

nector

OOM103-1 9mV – 13mVtemperature compensated < 5 seconds ≥ 500 000 % volume

oxygen hours3.5mm mono Jack

OOM103-1M 9mV – 13mVtemperature compensated < 5 seconds ≥ 500 000 % volume

oxygen hoursSwitchcraft® mini

power Jack

OOM104 24μA – 32μAno temperature compensation < 12 seconds ≥ 750 000 % volume

oxygen hours gold plated slip rings

OOM105 Teledyne® TED range < 5 seconds ≥ 500 000 % volume oxygen hours

Molex® plug 4P4C

OOM106 9mV – 13mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours3 pin Molex® con-

nector

OOM107 170μA – 230μAno temperature compensation < 12 seconds ≥ 250 000 % volume

oxygen hours gold plated slip rings

OOM107-2 170μA – 230μAno temperature compensation < 12 seconds ≥ 250 000 % volume

oxygen hoursFlying leads with pin-connectors

Oxygen sensorpart number Output signal in air Response time

T 90% Nominal sensor life-

time Electrical interface

OOM109 9mV to 13mVtemperature compensation < 300 msec. ≥ 200 000 % volume

oxygen hours 3pin molex®

OOM109-EM* 130mV to 190mV*temperature compensation < 150 msec.* ≥ 200 000 % volume

oxygen hoursAMP 100411-0

connector*

OOM109-EMLF* 130mV to 190mV*temperature compensation < 150 msec.* ≥ 200 000 % volume

oxygen hoursAMP 100411-0

connector*

OOM109-LF 9mV to 13mVtemperature compensation < 300 msec. ≥ 200 000 % volume

oxygen hours 3pin molex®

OOM110 10mV – 12mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours modular Jack 6P4C

OOM111 11mV – 13mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours 3mm stereo Jack

OOM112 30mV – 40mVtemperature compensated < 12 seconds ≥ 500 000 % volume

oxygen hours gold plated slip rings

OOM201 24μA – 32μA (Dual Cathode)no temperature compensation < 12 seconds ≥ 500 000 % volume

oxygen hours gold plated slip rings

OOM202 13mV – 16mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours 3 pin molex®

OOM202-1 13mV – 16mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hours 3.5mm mono Jack

OOM202-2 9mV – 13mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hoursfl ying leads with

3pin female molex® connector

OOM202-2S 9mV – 11.5mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume

oxygen hoursAMP MATE-N-LOK /

2 circuit

OOM204 9mV – 13.5mV (dual cathode)temperature compensated < 12 seconds ≥ 500 000 % volume

oxygen hours 3 pin molex®

Use the advantages:• accurate and reliable response

• resistant to N2O

• excellent signal stability

• high product quality

• short lead times

• technical support

* in connection with E.R.O.S. – Module 01

oxygen_sensors_10_2008_final.indAbs1:2-Abs1:3 Abs1:2-Abs1:3oxygen_sensors_10_2008_final.indAbs1:2-Abs1:3 Abs1:2-Abs1:3 05.11.2008 12:16:11 Uhr05.11.2008 12:16:11 Uhr