medical oxygen sensors general specifi cations · 2 3 medical oxygen sensors technical specifi...
TRANSCRIPT
EnviteC-Wismar GmbH
a Honeywell Company Internet: E-mail:
Alter Holzhafen 18, 23966 Wismar, Germany www.envitec.com [email protected]: +49 (0) 3841 / 360-1 Phone: +49 (0) 3841 / 360-200 Sales Fax: +49 (0) 3841 / 360-222
Medical Oxygen Sensors general specifi cations
Widest range of compatible sensors for
leading manufactures in the world.
Measurement range 0-100 % oxygenAccuracy and repeatability < 1 % vol. O2 when calibrated at 100 % O2Zero offset < 0.5 % vol. O2 in 100% N2, applied 5 minutesLinearity error < 3 % relativeCross interference Meets EN ISO 21647 requirements Influence of humidity - 0.03 % rel. per % RH at 25°C Influence of pressure proportional to change in oxygen partial pressureInfluence of mechanical shock < 1% relative after a fall from 1mOperating temperature 0° C to 50° CTemperature compensation built-in NTC compensation (depends on type)Effect of temperature between +25 °C and +40 °C: 3 % relative errorcompensation (steady state) between 0 °C and +50 °C: 8 % relative errorOperating humidity 0-99 % RH non condensingLong term output drift < 1 % vol oxygen per month typically < - 15 % relative over lifetime
Storage temperature -20 to +50 °C
Recommended storage +5 to +15 °C
Recommended load ≥ 10 kOhms
Warm-up time < 30 minutes, after replacement of sensor
Weight approximately 28 grams approximately 43 grams OOM107 seriesMeet EN ISO 21647, Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003All specifications are applicable at standard conditions: 1013 hPa, 25 °C dry ambient air
Medical Oxygen Sensors for life supporting systems
Par
t-N
o. 0
01-3
3-00
0000
11-b
• 1
0/20
08 •
Tec
hnic
al in
form
atio
n is
sub
ject
to c
hang
e w
ithou
t not
ice!
©
200
8 H
oney
wel
l Int
erna
tiona
l Inc
.
oxygen_sensors_10_2008_final.ind1-2 1-2oxygen_sensors_10_2008_final.ind1-2 1-2 05.11.2008 12:16:08 Uhr05.11.2008 12:16:08 Uhr
2 www.envitec.com 3
Medical Oxygen Sensors technical specifi cations
Oxygen sensorpart number Output signal in air Response time
T 90% Nominal sensor life-
time Electrical interface
OOM101 46μA – 63μAno temperature compensation < 12 seconds ≥ 500 000 % volume
oxygen hours gold plated slip rings
OOM102 9mV – 14mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours3 pin Molex® con-
nector
OOM102-1 9mV – 14mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours3.5mm mono Jack
OOM103 9mV – 13mVtemperature compensated < 5 seconds ≥ 500 000 % volume
oxygen hours3 pin Molex® con-
nector
OOM103-1 9mV – 13mVtemperature compensated < 5 seconds ≥ 500 000 % volume
oxygen hours3.5mm mono Jack
OOM103-1M 9mV – 13mVtemperature compensated < 5 seconds ≥ 500 000 % volume
oxygen hoursSwitchcraft® mini
power Jack
OOM104 24μA – 32μAno temperature compensation < 12 seconds ≥ 750 000 % volume
oxygen hours gold plated slip rings
OOM105 Teledyne® TED range < 5 seconds ≥ 500 000 % volume oxygen hours
Molex® plug 4P4C
OOM106 9mV – 13mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours3 pin Molex® con-
nector
OOM107 170μA – 230μAno temperature compensation < 12 seconds ≥ 250 000 % volume
oxygen hours gold plated slip rings
OOM107-2 170μA – 230μAno temperature compensation < 12 seconds ≥ 250 000 % volume
oxygen hoursFlying leads with pin-connectors
Oxygen sensorpart number Output signal in air Response time
T 90% Nominal sensor life-
time Electrical interface
OOM109 9mV to 13mVtemperature compensation < 300 msec. ≥ 200 000 % volume
oxygen hours 3pin molex®
OOM109-EM* 130mV to 190mV*temperature compensation < 150 msec.* ≥ 200 000 % volume
oxygen hoursAMP 100411-0
connector*
OOM109-EMLF* 130mV to 190mV*temperature compensation < 150 msec.* ≥ 200 000 % volume
oxygen hoursAMP 100411-0
connector*
OOM109-LF 9mV to 13mVtemperature compensation < 300 msec. ≥ 200 000 % volume
oxygen hours 3pin molex®
OOM110 10mV – 12mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours modular Jack 6P4C
OOM111 11mV – 13mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours 3mm stereo Jack
OOM112 30mV – 40mVtemperature compensated < 12 seconds ≥ 500 000 % volume
oxygen hours gold plated slip rings
OOM201 24μA – 32μA (Dual Cathode)no temperature compensation < 12 seconds ≥ 500 000 % volume
oxygen hours gold plated slip rings
OOM202 13mV – 16mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours 3 pin molex®
OOM202-1 13mV – 16mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hours 3.5mm mono Jack
OOM202-2 9mV – 13mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hoursfl ying leads with
3pin female molex® connector
OOM202-2S 9mV – 11.5mVtemperature compensated < 12 seconds ≥ 1 000 000 % volume
oxygen hoursAMP MATE-N-LOK /
2 circuit
OOM204 9mV – 13.5mV (dual cathode)temperature compensated < 12 seconds ≥ 500 000 % volume
oxygen hours 3 pin molex®
Use the advantages:• accurate and reliable response
• resistant to N2O
• excellent signal stability
• high product quality
• short lead times
• technical support
* in connection with E.R.O.S. – Module 01