med 1.2 bp measurement

4
TRANSCRIBED BY: FRED, GEORGE, RON, BILL, GINNY Page 1 of 4 Alfredo Guzman, M.D. “People find it far easier to forgive others for being wrong than being right. “ —Albus Dumbledore, Harry Potter & The Half-Blood Prince Blood pressure measurement 1.2 9 June 2014 BP MEASUREMENT Proper measurement & interpretation - essential in the dx & management of HPN Home BP & average 24-hr ambulatory BP are generally lower than clinic-taken BP BP tends to be higher in the early morning, soon after walking, than any other time of the day Night time BP is generally 10-20% lower than day time BP White Coat HPN Px manifests a higher BP in a hospital/clinical setting. They are at ↑risk of developing sustained HPN. FACTORS AFFECTING BP MEASUREMENTS Instrumentation o BP Apparatus: Mercury, Aneroid, Digital o BP cuff size Area of arm covered Technique of BP Measurement Patient Factor Environment BP APPARATUS / SPHYGMOMANOMETER MERCURY MANOMETER Standard for all BP measurements Large tube for rapid ↑ & ↓ in pressure 2mm graduated markings on tube Mylar-wrapped glass or plastic tube preferred Mercury is at zero and column rises and falls rapidly TESTING THE MERCURY MANOMETER Check the “0”. Top of meniscus should rest at the zero mark Inflate to 200 mmHg. Wait 1 min. Record Pressure. o If <170 mmHg, there’s leak, Release Pressure. Note wheter Hg rises & falls smoothly Locate & correct leaks Date device to indicate it was inspected & repaired ANEROID DEVICE Needs regular calibration w/ mercurial manometer Initiall position of needle at zero, can be easily damaged CALIBRATING THE ANEROID SELECTING THE CORRECT BP CUFF *Bates Width about 40% of upper arm circumference (ave 12-14cm in ave adult) Length about 80% of upper arm circumference Standard cuff size 12 x 23cm, appropriate for circumstances up to 28cm BP CUFF SIZES Cuff Sizes Arm Circumference Range at Midpoint Bladder Width (cm) Bladder Length (cm) Child 16 - 21 8 21 Small Adult 22 26 10 24 Adult 27 34 13 30 Large Adult 35 44 16 38 Adult Thigh 45 52 20 42 If the cuff is too small (narrow), the BP will read high If the cuff is too large (wide), the BP will read low on a small arm & high on a large arm SELECTING THE MOST ACCURATE CUFF MARKING THE CUFF Apply cuff so that the center of inflation bag is over brachial artery Be sure INDEX line falls between the 2 RANGE lines. If it does not, a larger of smaller cuff may be required. TOPIC OUTLINE I. Blood Pressure Measurement II. BP Apparatus/ Sphygmomanometer A. Mercury Manometer B. Aneroid Device III. Selecting the Correct BP Cuff A. BP Cuff Sizes B. Selecting the Most Accurate Cuff A. Marking the Cuff IV. The Stethoscope V. 5 Phases of Korotkoff Sounds VI. Technique of BP Measurement VII. Technique of BP Measurement in the Dx of HPN A. Timing of BP Measurement B. Patient Position C. Patient & Physician Position VIII. Observer’s Skill in BP Measurement IX. Unequal BP in Both Arms X. Steps in Measuring the BP XI. Definitions of Normal & Abnormal Levels XII. Special Problems A. Leg Pulses & Pressures B. The Apprehensive Patient C. The Obese or Very Thin D. Weak/ Inaudible Korotkoff Sounds F. When Korotkoff Sounds Can’t be Heard at All G. Arrhythmias H. The Hypertensive Patient w/ Unequal BP in Both Arms

Upload: elleinas

Post on 10-Sep-2015

18 views

Category:

Documents


1 download

DESCRIPTION

Med

TRANSCRIPT

  • TRANSCRIBED BY: FRED, GEORGE, RON, BILL, GINNY

    Page 1 of 4

    Alfredo Guzman, M.D.

    People find it far easier to forgive others for being wrong than being right. Albus Dumbledore, Harry Potter & The Half-Blood Prince Paulo Coelho

    Blood pressure measurement

    1.2 9 June

    2014

    BP MEASUREMENT

    Proper measurement & interpretation - essential in the dx & management of HPN

    Home BP & average 24-hr ambulatory BP are generally lower than clinic-taken BP

    BP tends to be higher in the early morning, soon after walking, than any other time of the day

    Night time BP is generally 10-20% lower than day time BP White Coat HPN

    Px manifests a higher BP in a hospital/clinical setting. They are at risk of developing sustained HPN.

    FACTORS AFFECTING BP MEASUREMENTS

    Instrumentation o BP Apparatus: Mercury, Aneroid, Digital o BP cuff size

    Area of arm covered

    Technique of BP Measurement

    Patient Factor

    Environment

    BP APPARATUS / SPHYGMOMANOMETER MERCURY MANOMETER

    Standard for all BP measurements

    Large tube for rapid & in pressure

    2mm graduated markings on tube

    Mylar-wrapped glass or plastic tube preferred

    Mercury is at zero and column rises and falls rapidly

    TESTING THE MERCURY MANOMETER

    Check the 0. Top of meniscus should rest at the zero mark

    Inflate to 200 mmHg. Wait 1 min. Record Pressure.

    o If

  • TRANSCRIBED BY: FRED, GEORGE, RON, BILL, GINNY

    Page 2 of 4

    Blood pressure

    measurement

    Can be used on either right or left arm,

    THE STETHOSCOPE

    Earpiece should face forward in the ear canal

    Must have thick tubing 12-15 inches long

    Bell for low pitched sounds

    o Used to detect low frequency Korotkoff sounds

    Diaphragm for high pitched sounds

    5 PHASES OF KOROTKOFF SOUNDS

    Korotkoff Sounds produced by the flowing of blood as the cuff

    is released

    Phase Description Remarks

    I 1st appearance of clear,

    tapping sound

    Represents Systolic P

    (SBP)

    II Soft murmurs that

    replace Phase I sounds

    -

    III

    Loud murmurs that

    replace Phase II

    sounds

    Due to blood flow

    through constricted artery

    IV

    Sudden muffling of

    Phase III sounds

    Due to constriction of

    the artery; arterial

    diastolic P is

    approached

    V

    Disappearance of

    Korotkoff sounds

    Represents Diastolic BP

    (DBP) in most pxs is

    normally w/n 10mmHg

    from Phase IV

    (abnormal if >10mmHg

    difference; Phase IV is

    abruptly muffled)

    The usual BP reading involves Phase I and Phase V Korotkoff

    sounds for SBP and DBP, respectively.

    If there is a significant difference (>10mmHg) between Phase IV

    and V, both pressures should be recorded (e.g. 130/70/10

    mmHg); seen in anemia, aortic regurgitation, thyrotoxicosis.

    In chronic, severe aortic regurgitation or a large arteriovenous

    fistula, where the disappearance point may reach 0 mmHg, Phase

    IV is much closer to the intraarterial diastolic pressure than Phase

    V. All 3 pressures should be noted (e.g. 140/60/0 mmHg).

    Difficulty in Hearing Korotkoff Sounds

    Condition Pathology

    Severe aortic stenosis Arterial P rises at a slow rate

    Shock Markedly constricted arteries

    Severe heart failure Markedly constricted arteries

    Opening and closing the fist repeatedly can help dilate blood

    vessels of the arms and make Korotkoff sounds more audible.

    Korotkoff sounds represented during BP measurement. Note the

    auscultatory gap.

    TECHNIQUES IN BP MEASUREMENT 1. Support arm at the level of the heart

    2. Inflate cuff 30 mmHg above the palpatory BP

    3. Release pressure at a rate of 2-3 mmHg/s

    Initially, measure BP on both arms

    Use arm w/ higher BP on subsequent measurements

    Measure BP at least twice per visit; allow 1-2 minutes in between measurements

    If there is >5 mmHg difference between 2 consecutive measurements, additional or continued measurements should be made

    Take the average of the last 2 BP measurements and record

    TECHNIQUE OF BP MEASUREMENT IN THE DX OF HPN I. TIMING OF BP MEASUREMENT

    For Dx: Multiple readings taken at various times throughout

    waking hours

    For Monitoring: Measure prior to intake of anti-hypertensive

    medication to determine trough or nadir effect

    II. PATIENT POSITION

    Usually taken while sitting slouched on the chair

    o Supine position SBP & DBP by 2-3 mmHg

    Allow patient to rest and sit quietly for 5 minutes

    o Apprehension increases BP

    Measure both sitting and standing BP to detect postural hypotension (sudden drop in BP upon standing; in elderly, DM)

    III. PATIENT & PHYSICIAN POSITION

    Sitting; feet flat on the floor

    Arm supported at heart level

    Confirm viability of brachial pulse by palpation

    Use bell (detection of low-pitched sounds)

  • TRANSCRIBED BY: FRED, GEORGE, RON, BILL, GINNY

    Page 3 of 4

    Blood pressure

    measurement

    OBSERVERS SKILL IN BP MEASUREMENT

    The brain must be programmed to follow the proper guidelines

    every time the P is measured.

    Must be able to store the systolic and diastolic P & recall them

    accurately.

    Must be able to hear the Korotkoff sounds and knowhow to

    interpret them.

    Must be able to recall and write down correctly & legibly the sounds

    heard.

    Must be able to find & feel the pulses needed for BP measurement.

    UNEQUAL BP IN BOTH ARMS

    STEPS IN MEASURING THE BP

    STEPS TO ENSURE ACCURATE BP MEASUREMENT

    Instruct px to avoid smoking/drinking caffeinated drinks 30 mins

    prior to BP measurement

    Make the examining room as quiet & comfy as possible

    Arm should be supported at heart level. Ask px to sit quietly for 5

    mins on chair

    Make sure the arm is free of clothing. There should be NO

    arteriovenous fistulas for dialysis, scarring from prior brachial

    artery cutdowns, or signs of lymphedema

    Palpate the arm so that the brachial artery (located at the

    antecubital crease) is at heart level (roughly level w/ the 4th

    interspace at its junction w/ the sternum

    If the px is seated, rest the arm on a table a little above the pxs

    waist, if standing, try to support the pxs arm at the mid-chest level

    STEPS IN MEASURING THE BP

    Center the inflatable bladder over the brachial artery. Lower border

    of cuff should be 2.5cm above the antecubital crease. Secure

    the cuff snugly. Position arm so that it is slightly flexed at the

    elbow,

    To determine how high to raise the cuff P, 1st estimate the systolic

    pressure by palpation. As you feel the radial artery w/ the fingers

    of one hand, inflate the cuff until the radial pulse disappears. Read

    this P on the manometer & add 30 mmHg to it. Use of this sum as

    the target for the next inflation prevents discomfort from

    unnecessary high cuff P. This also avoids the occasional error

    caused by an auscultatory gap (a silent interval that may be

    present between the systolic & diastolic pressure)

    Deflate cuff promptly & completely & wait 15-30 seconds

    Next, place the bell of a stethoscope lightly over the brachial artery,

    taking care to make an air seal w/ its full rim. Because the sounds

    to be heard, the Korotkoff sounds, are relatively low pitched,

    they are better heard w/ the bell.

    Inflate the cuff rapidly again to the level determined, then deflate it

    slowly (2-3mmHg/sec), Note the level at w/c you hear the sounds

    of at least 2 consecutive beats. This is the Systolic Pressure.

    Continue to lower the P slowly until the sounds become muffled &

    then disappear. To confirm the disappearance of sounds, listen as

    the P falls another 10-20mmHg. Then deflate the cuff rapidly to

    zero. The disappearance point, w/c is usually only a few mmHg

    below the muffling pt, provides the best estimate of true diastolic P

    in adults.

    Read both the systolic & diastolic levels to the nearest 2mmHg.

    Wait 2 mins & repeat. Average your findings. If the 1st 2 readings

    differ by 5mmHg, take additional readings,

    Avoid slow/ repetitive inflations of the cuff, because the resulting

    venous congestion can cause false readings.

    BP should be taken in both arms at least once. Normally, there

    may be a difference in P of 5mmHg & sometimes up to 10mmHg.

    Subsequent readings should be made on the arm w/ the higher

    pressure.

    o Loose cuff/bladder = false high readings

    o Earpiece should face forward in the ear canal.

    o Bell of the stethoscope is used for low-frequency sounds

    o Auscultatory gaps are associated w/ arterial stiffness &

    atherosclerotic disease

    o P difference of >10-15mmHg seen in subclavian steal syndrome

    & aortic dissection

    o BP of 110/70 is usually normal, but could indicate significant

    hypotension if previous readings are high

    DEFINITIONS OF NORMAL & ABNORMAL LEVELS

    CATEGORY SBP

    (mmHg) DBP

    (mmHg)

    Normal < 120 and < 80

    Prehypertension 120 - 139 or 80 - 89

    Hypertension

    Stage 1

    Stage 2

    140 159

    160

    or

    or

    90 99

    100

    BP goal for pxs w/ HPN, DM, or renal disease is

  • TRANSCRIBED BY: FRED, GEORGE, RON, BILL, GINNY

    Page 4 of 4

    Blood pressure

    measurement

    o A femoral pulse that is smaller & later than the radial pulse

    suggests coarctation of the aorta or occlusive aortic disease.

    BP is lower in the legs than in the arms in these conditions.

    THE APPREHENSIVE PATIENT

    Try to relax the px

    Repeat the measurement later in the encounter

    Some px will say their BP is only elevated in the office (White Coat

    HPN) & may need to have their BP measured several times at

    home or in a community setting.

    THE OBESE OR VERY THIN

    For the obese, the a wide cuff (15cm)

    If arm circumference exceeds 41cm, us a thigh cuff (18cm wide)

    For the very thin arm, use a pediatric cuff

    WEAK/ INAUDIBLE KOROTKOFF SOUNDS

    To rule out Coartation of the aorta, consider: Technical problems: wrong placement of stethoscope, failure to

    make full skin contact w/ bell venous engorgement of the arm from

    repeated inflations of the cuff

    Consider shock

    WHEN KOROTKOFF SOUNDS CANT BE HEARD AT ALL Estimate systolic P via palpation. Alternative methods such as

    Doppler techniques or direct arterial pressure tracings may be necessary.

    To intensify the Korotkoff sounds, these may be done: o Raise the arm before & while you inflate the cuff. Then lower the

    arm & determine the BP. o Inflate the cuff. Ask the px to make a fist several times. Then take

    the BP.

    ARRHYTHMIAS

    Irregular rhythms produce variations in Pand therefore unreliable measurements.

    Ignore the effects of an occasional premature contraction.

    W/ frequent premature contractions or atrial fibrillation, determine the average of several observations and note that your measurements are approximate.

    THE HYPERTENSIVE PX W/ UNEQUAL BP IN BOTH ARMS

    To detect coarctation of the aorta, make 2 further BP measurements at least once in every hypertensive px:

    Compare BP in the arms and legs.

    Compare the volume and timing of the radial and femoral pulses. Normally, volume is equal and the pulses occur simultaneously.

    Coarctation of the aorta arises from narrowing of the thoracic aorta, usually proximal but sometimes distal to the left subclavian artery.

    Coarctation of the aorta & occlusive aortic disease are distinguished by hypertension in the upper extremities & low BP in the legs and by diminished or delayed femoral pulse.