measurement methods

49
Atos 2 - Measurement methods 1 Rel. 1.20 - 01.03.01 Measurement methods

Upload: thomas-williams

Post on 03-Jan-2016

48 views

Category:

Documents


0 download

DESCRIPTION

Measurement methods. Introduction. The Slides in this Presentation are representation of the Measurement Theories rather than the actual model and instruments behavior in the SPEA Systems. For further detailed information on the models and the instruments setup for each measurement - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Measurement methods

Atos 2 - Measurement methods 1Rel. 1.20 - 01.03.01

Measurement methods

Page 2: Measurement methods

Introduction

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 2

The Slides in this Presentationare representation of the Measurement Theoriesrather than the actual model and instruments behavior in the SPEA Systems.For further detailed information on the modelsand the instruments setup for each measurementtype please refer to the document titled “Leonardo YA: Passive devices test methods “

Page 3: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 3

Resistor - Voltage measurement method

1

G M

I0

VRxRx

Tp

Tp

• The generator “G” forces a programmed current (Io) through Rx.

• The voltage meter “M” measures the voltage drop on Rx (VRx).

• Applying the Ohm law the expected voltage will be:

VRx = Io • Rx

(V = 0.25V when I = 0.25ma and R = 1K ohm)

Two wires measurement

Typical application range: 100 ÷ 100K

Page 4: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 4

Resistor - Voltage measurement method

1 • The generator “G” forces a programmed current (Io) through Rx

• The voltage meter “M” measures the voltage drop on Rx (VM)

• If the forced current is relatively “high” (> 2.5mA) the wiring and the contact resistance (Rh and Rc) will influence the measure

Two wires measurement: wiring and contact resistance effect

G M

I0

VRxRx

Tp

Tp

VM

Rh

Rc

VRh

VRc

• The measured value (VM) will be influenced by the voltage drop on these resistances: VM = VRh + VRx + VRc

• The smaller the resistance value of Rx the bigger the wiring and the contact resistance influence

Page 5: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 5

Resistor - Voltage measurement method

1 • The generator “G” forces a programmed current (Io) through Rx.

• The voltage meter “M” measures the voltage drop on Rx (VRx).

• If the forced current is relatively “high”(> 2.5mA) the wiring and the contact resistance (Rh and Rc) will influence the measure.

• Two additional test points are used to connect the voltage meter “M” to the circuit. Due to its high impedance the measured voltage drop (VRx) will be equal to the voltage drop on the tested resistor.

Four wires measurement (Kelvin measure)

Typical application range: 100

G

I 0

VRxRx M

Tp

Tp

VM

Rh

Rc

VRh

VRc

Tp

Tp

Page 6: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 6

Resistor - Voltage measurement method

1 • The generator “G” forces a programmed current (Io) through Rx.

• The voltage meter “M” measures the voltage drop on Rx (VRx).

• Due to the circuit configuration the measured voltage drop will be equal to: VRx = Io • (Rx // Rs + Rp)

Guard measurement

G

I0

RpRx

Rs

M

Is

Ix

VM

A

• To measure Rx without the influence of Rs and Rp it is necessary to have a Is current equal to 0; this is possible if the point “B” is at the same voltage of point “A”

• The Guard is a “voltage follower” and is used to isolate the tested component from the rest of the circuit.

G

B

Page 7: Measurement methods

Code: R1.6 Measurement methods 7

Resistor - Voltage measurement method

1

Parallel capacitance effect

• The resistor measurement forces a current pulse, the measured value will be a voltage pulse.

• The parallel capacitor present in the circuit will absorb the current pulse until the Ton is long enough to charge the

capacitor.

t

I

V

t

t

I

V

t

G

I0

Rx VRxM Cp

Page 8: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 8

Resistor – Current measurement method

1 • The generator “G” generates a programmed voltage (Vo) on Rx.

• The voltage meter “A” measures the current flow through Rx

(IRx).

• Applying the Ohm law the expected current will be: Vo

IRx = -------- Rx

Typical application range: 50K

G

IRx

Rx

A

V0

Two wires measurement

Page 9: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 9

Resistor - Current measurement method

1• The generator “G” generates a programmed voltage (Vo) on Rx.

• The voltage meter “A” measures the current flow through Rx

(IRx).

• Applying Ohm’s law the expected current will be: Vo

IRx = -------- Rx

• If the resistor value is relatively “high” (> 100K) the measuredcurrent could be near the generator leakage current

• To avoid this leakage current influence the leakage current is measured in a HLD (hold) test

• The actual device under test current is measured next and the leakage current value is subtracted from the actual value.

Typical application range: 100K

G

IRx

Rx

A

V0

Current compensated measurement

Page 10: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 10

Capacitor - Voltage measurement method

1

Two wires measurement

• The generator “G” forces a programmed current (Io) through Cx.

• The voltage meter “M” measures the voltage Cx (VCx).

• The measured value is related to the capacitor charge by the following formula: Io • T VCx = --------- Cx

G

I0

VCx

M CX

Typical application range: 100nF ÷1µF

t

I

V

t

I0

T

VCx

Page 11: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 11

Capacitor - Voltage measurement method

1

Two wires measurement: wiring and contact resistance effect

• The generator “G” forces a programmed current (Io) through Cx

• The voltage meter “M” measures the voltage drop on Cx (VCx)

• If the forced current is relatively “high”, the wiring and the contact resistance (Rh and Rc) will influence the measure

G

I0

M CX VCxVM

t

I

V

t

I0

VCx

VM

V

Rh

Rc

VRh

VRc

• The measured value (VM) will be influenced by the voltage drop on these resistances, adding an “offset” calculated as follow: VM = VCx + VRh + VRc

VRh = Io • Rh

VRc= Io • Rc

Page 12: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 12

Capacitor - Voltage measurement method

1

Typical application range: 10µF

Four wires measurement (Kelvin measure)

• The generator “G” forces a programmed current (Io) through Cx.

• The voltage meter “M” measures the voltage drop on Cx (VCx).

• If the forced current is relatively “high”, the wiring and the contact resistance will influence the measurement.

• Two additional test points are used to connect the voltage meter “M” at the circuit. Due to its high impedance the measured voltage drop (VCx) will be equal to the voltage drop on the tested resistor.

G

I0

MCX

Rh

Rc

VRh

VRc

VCx

Tp Tp

Tp Tp

Page 13: Measurement methods

Code: R1.6 Measurement methods 13

Capacitor - Time measurement method

1

Two wires measurement

• The generator “G” forces a programmed current (Io) through Cx.

• The timer “C” measures the time elapsed between the programmed thresholds (V1 and V2).

• The measured value is related to the capacitor charge by the following formula:

Cx • (V2 – V1)

T = -------------------- Io

Typical application range: 1nF ÷ 100nF

G

I0

VCxC CX

t

I

t

I0

T

V2

V1

VCx

Page 14: Measurement methods

Code: R1.6 Measurement methods 14

Capacitor – AC Current measurement method

1

Two wires measurement

• The generator “G” generates a sinus waveform on the Cx

capacitor pins.

• The current meter “A” measures the current flow in the capacitor at the zero crossing of the Voltage.

• The measured current value is directly converted to the relative capacitive current value, by the system CPU.

Typical application range: 5pF ÷ 100nF

G

A

CX

I0

VM

V

I

Page 15: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 15

Inductor - Current measurement method

Typical application range: 10mH

• The generator “G1” generates a continuous voltage (V1)

• The generator “G2” generates a negative pulse voltage (V2)

• The inductor reaction to the negative pulse is a reverse current flow measured by “G2”

• The reverse current can be used to calculate the inductance by the following formula:

T • (V1)

I = -------------------- Lx

1

G G

A

1 2V1 V2

I0

LX

Page 16: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 16

Inductor - Time measurement method

• The test generator “G” charges the internal 10µF capacitor

• During the test execution the capacitor is discharged by the inductor

• This generates an oscillation (ring) of a frequency value related to the inductor value.

• The counter “C” measures the time interval (TM) elapsed between two defined thresholds, it can be measured applying the following formula:

TM = · LC

Typical application range: 1mH ÷ 10mH

1

Two wires measurement

G CVRx

G

Internal Circuitery

t

V

TM

VTH

Page 17: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 17

Diode & LED – Vf measurement method

• The generator “G” forces a programmed current (I f) through the junction.

• The voltage meter “M” measures the voltage drop (V f) on the junction

VfG M

If

1

Two wires measurement

Page 18: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 18

Diode & LED – Vf measurement method

1 • The generator “G” forces a programmed current (I f) through the junction.

• The voltage meter “M” measures the voltage drop (V f) on the junction

• If the forced current is relatively “high” the wiring and the contact resistance will influence the measurement.

• Two additional test points are used to connect the voltage meter “M” to the circuit. Due to its high impedance the measured voltage drop (Vf) will be equal to the voltage drop on the junction

Four wires measurement (Kelvin measure)

VfG M

Rh

Rc

If

VRh

VRc

Tp

Tp

Tp

Tp

Page 19: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 19

Zener Diode – Vf measurement method

• The generator “G” forces a programmed current (I f) through the junction.

• The voltage meter “M” measures the voltage drop (V f) on the junction

1

Two wires measurement

VfG M

If

Page 20: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 20

Zener Diode – Vz measurement method

• The generator “G” forces a programmed current (I f) through the junction.

• The voltage meter “M” measures the voltage drop (Vz) on the junction

1

Two wires measurement

VzG M

If

Page 21: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 21

Bipolar Transistor

B

E

C

• All the typical bipolar transistor parameters can be tested.

• Orientation tests can be performed (VBE and VBC)

• Working tests can be performed (VCE0 and VCE sat)

• Parametric tests are possible (ß gain, …)

1

Page 22: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 22

Bipolar Transistor – VBE measurement method

• The generator “G” forces a current between Base and Emitter

• The voltage meter “M” measures the voltage drop between the junction Base and Collector (VBE).

• VBE junction measurement should be near 0.7V

1

G

If

M

B

E

VBE

Two wires measurement

Page 23: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 23

Bipolar Transistor – VBC measurement method

• The generator “G” forces a current between Base and Collector

• The voltage meter “M” measures the voltage drop between the junction Base and Collector (VBC).

• VBC junction measurement should be near 0.7V

1

G

If

M

B

E

VBCC

Two wires measurement

Page 24: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 24

Bipolar Transistor – VCE0 measurement method

• The generator “G” generates a voltage drop between transistor Collector and Emitter (VCE0)

• The transistor Base is shorted to Emitter (for NPN transistor)

• The transistor Base is shorted to Collector (for PNP transistor)

• With this test configuration the transistor is in “interdiction” mode (off = zero current)

• The current meter “A” measures the current flow between transistor pins Collector and Emitter (ICE0).

• ICE0 = 0mA

1

G

ICE0

B

E

C

A

VCE0

Page 25: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 25

Bipolar Transistor – VCEsat measurement method

• The generator “G1” forces a current between the Base-Emitter junction in order to bias the transistor “ON” 1

• The generator “G2” forces a current between the transistor pins Collector and Emitter 1

• With this test configuration the transistor is in saturation mode

• The voltage meter “M” measures the voltage drop between transistor pins Collector and Emitter (VCEsat).

• VCEsat = 0V

1

GB

E

C

VCEM

G1

2

VBE

Two wires measurement

Typical application range: signal transistor1 = The current is forced in accord to the transistor technology (NPN or PNP)

Page 26: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 26

Bipolar Transistor – VCEsat measurement method

• If the forced current is relatively “high” (>10mA) the wiring and the contact resistance (Rh and Rc) will influence the measured voltage drop between the Collector-Emitter junction

• The generator “G1” generates a voltage to bias the Base-Emitter junction “ON”.

• The generator “G2” provides a voltage bias between the Collector and Emitter

• With this test configuration the transistor is in saturation mode

• Due to this the wiring and the contact resistances (Rh and Rc) the transistor test is performed in two steps

• During the first test the voltage drop caused by the wiring and the contact resistance is measured by the voltage meter “M” and stored in a register. This stored value will modify the CE voltage measurement.

• During the second test the voltage drop between the Collector-Emitter junction is measured by the voltage meter “M”

1

Four wires measurement (Kelvin measure)

Typical application range: medium/power transistor

GB

E

C

VCE

M

G1

2

VBE

Rh

Rh

2

GB

E

C

VCE

G1

2

VBE

Rh

Rh

M

Page 27: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 27

Mos-Fet Transistor

• All the typical Mos-Fet transistor parameters can be tested.

• Orientation test can be performed (Vf of the protection diode)

• Operating tests can be performed (VDS off and VDS on)

• Parametric tests are possible (RDS on, …)

1

D

S

G

Page 28: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 28

Mos-Fet Transistor – Orientation test method

• A diode test can be used to measure the protection diode present between pins Drain and Source

• The generator “G” forces a current (If) between the Drain and Source

• The voltage meter “M” measures the voltage drop between the Drain and Source

• If the protection diode is present and the part mounted correctly the voltage meter should measure a forward biased diode voltage near 0.7V

1

Two wires measurement

D

S

G GMVf

If

Page 29: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 29

Mos-Fet Transistor – Vds off measurement method

• The generator “G1” generates a voltage equal to 0 between transistor pins Gate and Source (for N-channel)

• The generator “G2” tries to force a current between transistor pins Drain and Source

• In these conditions the Mos-Fet transistor is “OFF”

• The current meter “A” measures the current flow between transistor pins Drain and Source.

• Measured Current should be near “0mA”.

1

Two wires measurement

D

S

G

G1

G2

AIm

Page 30: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 30

Mos-Fet Transistor – Vds on measurement method

• The generator “G1” generates a voltage drop between transistor Gate and Source (for N-channel)

• The generator “G2” forces a current between transistor Drain and Source

• In these conditions the Mos-Fet transistor works in “ON” condition

• The current meter “A” measures the current flow between transistor pins Drain and Source.

• Measured current should be about 10ma (normal test).

1

Two wires measurement

D

S

G

G1

G2

AIm

Page 31: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 31

Transformer

1 • Typical transistor parameters can be tested.

• The transformer phase can be verified

• The transformer ratio can be measured

• The impedance values of the transformer coils can be measured

Page 32: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 32

Transformer – Impedance measurement method (Resistance)

1 • Resistance value for a transformer winding can be measured. Low Ohm values may require a Kelvin measurement

• The generator “G” forces a programmed current (Io) through transformer winding.

• The voltage meter “M” measure the voltage drop on the winding

• Because the coil resistance value is usually small the wiring and the contact resistance will influence the measure

• Two additional test points are used to connect the voltage meter “M” across the coil. Due to DVM high impedance the measured voltage drop (VRx) will be equal to the voltage drop on the tested resistor/coil.

G MVRx

Four wires measurement (Kelvin measure)

Resistor measurement

Tp

Tp

Tp

Tp

Page 33: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 33

Transformer – Phase measurement method

• The generator “G” forces a pulse on the transformer primary coil

• The voltage meter “M” measures a voltage value and phase as as produced by the transformer secondary coil.

1

Two wires measurement

G M

Page 34: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 34

Transformer – Ratio measurement method

• The generator “G” forces a pulse on the transformer primary coil

• The voltage meter “M” measures and stores the voltage drop across the transformer primary winding

• The generator “G” forces a pulse on the transformer primary coil

• The voltage meter “M” measures the voltage drop on the transformer secondary winding

• The voltage meter “M” performs a percent ratio calculation between the measured secondary voltage value and the previously stored primary voltage value

Two wires measurement

1

2

G M

G M

Page 35: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 35

Optocoupler

1 • Typical opto-coupler parameters can be tested.

• The led orientation is usually the first measurement

• The voltage drop on the output during both the operating modes (“On” and “Off”) can be measured.

Page 36: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 36

Optocoupler – LED Vf measurement method

1

G M Vf

If

• The generator “G” forces a programmed current (I f) through the LED

• The voltage meter “M” measures the voltage drop (V f) on the LED

• This verifies that the LED is present, working and correctly oriented

Page 37: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 37

Optocoupler – Vce sat measurement method

1• The generator “G1” forces a programmed current (If) through the

LED.

• The output transistor is biased into “saturation” mode

• The generator “G2” forces a current between the Collector-Emitter junction

• The voltage meter “M” measures the voltage drop on the Collector

Emitter junction (VCE sat) (near zero volts)

MVM

If

G1 G2

Page 38: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 38

Optocoupler – Vce 0 measurement method

1 • The generator “G1” generates a voltage equal to 0V through the LED.

• The output transistor is in “interdiction” “OFF” mode

• The generator “G2” tries to force a current between the Collector-Emitter junction on the output transistor.

• The voltage meter “M” will measure the voltage limit of “G2” (VCE0) because the transistor is “OFF” and no Emitter-Collector current flows.

MVM

If

G1 G2

Page 39: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 39

Relay

1 • All the typical relay parameters can be tested.

• Coil resistance value can be tested

• Contact verification (open and close) can be done

• Coil minimum operating voltage can be done

• Release voltage can be done

• Attraction time can be done

Page 40: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 40

Relay – Coil resistance measurement method

1 • The generator “G” forces a programmed current (Io) through the relay coil

• The voltage meter “M” measures the voltage drop across the coil

• If the forced current is relatively “high” the wiring and the contact resistance (RC and RH) will influence the measure

• Two additional test points are used to connect the voltage meter “M” at the circuit. Due to its high impedance the measured voltage drop (VM) will be equal to the voltage drop on the coil.

I0

G M VM

Rh

Rc

VRh

VRc

Four wires measurement (Kelvin measure)

Page 41: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 41

Relay – Contact verification method

1 • The generator “G” tries to force a programmed current (Io) through the normally open contact (or normally closed)

• The voltage meter “M” measures the voltage drop on it

I0

GM VM

Two wires measurement – Cold method

Page 42: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 42

Relay – Contact verification method

1 • The generator “G1” generates the required voltage on the relay coil to energize the contact.

• The generator “G2” tries to force a current (I0) between the contact pins

• The voltage meter “M” measures the voltage drop (VM) on the contact pins.

• Measured voltage will be near be 0 across closed contacts and near the programmed voltage of “G2” across open contacts

Two wires measurement – Hot method

I0

GM VMG1 2

Page 43: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 43

Relay – Working verification method

1• The generator “G1” generates the minimum operating voltage

(generally equal to 2/3 of nominal voltage) on the relay coil.

• The generator “G2” forces a current (I0) between the contact pins

• The voltage meter “M” measures the voltage drop (VM) on the contact pin, it must be near be 0 with normally open contact and near to the programmed voltage of “G2” with normally closed contact

Coil minimum operating voltage

I0

GM VMG1 2

Page 44: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 44

Relay – Working verification method

1

• The generator “G1” generates the release voltage (generally equal to 1/10 of nominal voltage) on the relay coil.

• The generator “G2” forces a current (I0) between the contact pins

• The voltage meter “M” measures the voltage limit of “G2” with normally open contact

Coil Release voltage

I0

GM VMG1 2

Page 45: Measurement methods

Rel. 1.20 - 01.03.01 Atos 2 - Measurement methods 45

Relay – Working verification method

1 • The generator “G1” generates the required voltage on the relay coil to cause the contacts to close

• The generator “G2” forces a current (I0) between the contact pins

• The counter “C” measures the time (TM) elapsed between the two programmed thresholds on the contact pin voltage (VM)

Attraction time

V1

V2

t

tVM

VC

TM

I0

GC VMG1 2 VC

Page 46: Measurement methods

46 ATOS 2 Programming Training Course

Special Test; Analog Functional IC test

1

Driver2 supplies VCC to the board

Driver1 supplies logic high voltage to input of IC

Measurement Unit measures logic low voltage on IC output

Duplicate test and invert voltage settings for opposite logic level test

Use same setup to power up and measure regulator output voltage

Page 47: Measurement methods

Code: R1.6 Measurement methods 47

Open pin – Electroscan

• The generator “G1” generates a sinus waveform that is injected through the pin of the IC.

• The sensor plate picks-up the electromagnetic field and the active probe amplifies the signal. The Electroscan Box selects, by using the Multiplexer Stage, the relevant input and

converts the alternating signal to a continuous one.

• The voltage meter “M” measures the Electroscan Box continuous signal output.

• If the pin is soldered, the current pass through the IC and generates a changing electromagnetic field. So we can appreciate an output of the Electroscan Box of 2-10V. If the pin is not soldered, the Electroscan Box output signal would normally be significantly

lower than 2 V.

• In the 3030 system, “ Electroscan Box ” and “G1” are integrated within a internal system board named “YA32ESCAN”.

Mux

Active Probe

Integrated circuitSensor plate

Nail

G1

Nail

GND

A

M

~A

Printed circuit

Conditioning circuit

~

+

-

FUNCTIONAL SCHEME

Electroscan Box

Page 48: Measurement methods

Code: R1.6 Measurement methods 48

Open pin – Clamp diodes

• Each pin of an integrated circuit is usually connected to two diodes how it is shown in the schematics. Such diodes are normally within

the IC.

• Measuring one of the internal diodes, it is possible to assure that the pin is soldered.

• If the pin is not soldered, you will measure an open circuit, otherwise you will measure a diode junction. In order to detect if the pin is soldered, it is enough to measure one of the two internal diodes.

• If a similar pin (or a diode or something electrically similar) is connected in parallel to the pin under test, a problem rises: it’s not possible to discriminate if the pin under test is soldered. In such configuration you can’t use the “Clamp diode” measure method.

Pin

Integrated circuit

GND

Vcc

Internal diode A

Internal diode B

Page 49: Measurement methods

Code: R1.6 Measurement methods 49

Open pin – Clamp diodes

• The generator “G” forces a programmed current (If) through the junction.

• The voltage meter “M” measures the voltage drop (Vf) on the junctionVf

G M

If

VfG M

If

Internal diode A

Internal diode B

Vcc Device

Pin device

GND device

• The generator “G” forces a programmed current (If) through the junction.

• The voltage meter “M” measures the voltage drop (Vf) on the junction

Test Pin-Vcc diode

Test Gnd-Pin diodePin device