MDCT TechnologyMDCT Technology
Kalpana M. Kanal, Ph.D., DABRKalpana M. Kanal, Ph.D., DABRAssistant ProfessorAssistant Professor
Department of RadiologyDepartment of RadiologyUniversity of WashingtonUniversity of Washington
Seattle, WashingtonSeattle, Washington
ACMP Annual Meeting 2008 ACMP Annual Meeting 2008 -- Seattle, WASeattle, WA
KanalKanal 22
Educational ObjectivesEducational Objectives
�� Historical Development of CTHistorical Development of CT�� Why has the frequency of CT increased?Why has the frequency of CT increased?�� MultiMulti--Row Detector CT and PitchRow Detector CT and Pitch�� Advantages of MultiAdvantages of Multi--Row Detector CTRow Detector CT�� Automatic Exposure Control in CTAutomatic Exposure Control in CT�� Cardiac CT Cardiac CT �� Future of CTFuture of CT
KanalKanal 33Dr. Mahesh, John Hopkins, MD,Dr. Mahesh, John Hopkins, MD, AAPM Handout.AAPM Handout.
64 slice 64 slice scannersscanners
Dual Source Dual Source CTCT
20042004-- 20052005
Historical Development of CTHistorical Development of CT
Slow scan time Fast scan time
KanalKanal 44
Historical Development of CTHistorical Development of CTSeventh Generation CT (late 1990Seventh Generation CT (late 1990’’s s –– 2000)2000)
MultiMulti --Row Detector SystemsRow Detector Systems
Radiographics 2000: 20Radiographics 2000: 20--17871787--18061806
Kalendar WA, Computed Tomography, Kalendar WA, Computed Tomography, Second Edition, 2005, pg 58.Second Edition, 2005, pg 58.
KanalKanal 55
Historical Development of CTHistorical Development of CTPET/CT Scanners (2001)PET/CT Scanners (2001)
Kalendar WA, Computed Tomography, 2Kalendar WA, Computed Tomography, 2ndnd ed., 2005, pg 74.ed., 2005, pg 74.
KanalKanal 66
Historical Development of CTHistorical Development of CTDualDual--Source CT (2005)Source CT (2005)
� A scanner using two X-ray sources and two detectors at the same time
� Enables scanning a heart at any heart rate and at the lowest radiation dose possible
www.medical.siemens.comwww.medical.siemens.com
KanalKanal 77
Fast Scanning, 0.3 secFast Scanning, 0.3 sec
KanalKanal 88
Why has the frequency of CT exams Why has the frequency of CT exams increased?increased?
�� New New technologytechnology
�� Faster scansFaster scans
�� New exams New exams done with CTdone with CT
�� More More screening screening examsexams
UNSCEAR 2000
More like 60%
ICRP Publication 87 (2001)
KanalKanal 99
SingleSingle--Row Detector versus MultiRow Detector versus Multi --Row Row Detector CT Detector CT
Radiographics 2000: 20Radiographics 2000: 20--17871787--18061806
KanalKanal 1010
SingleSingle--row Detector vs. Multirow Detector vs. Multi --row Detector row Detector CTCT
Seeram, Computed Tomography, 2Seeram, Computed Tomography, 2nd ed.,ed., 2001, pg 258.2001, pg 258.
KanalKanal 1111
Multiple Detector ArraysMultiple Detector Arrays
Bushberg, et al. The Essential Physics of Medical Imaging, 2Bushberg, et al. The Essential Physics of Medical Imaging, 2ndnd ed., p. 341.ed., p. 341.
KanalKanal 1212
Slice Width SelectionSlice Width Selection
http://www.impactscan.org/slides/impactcourse/helical_and_multi-slice_principles/img17.html
GE 64 MDCT GE 64 MDCT (40 mm (40 mm detector)detector)
64 x 0.625 mm64 x 0.625 mm
32 x 1.25 mm32 x 1.25 mm
16 x 2.5 mm16 x 2.5 mm
8 x 5 mm8 x 5 mm
KanalKanal 1313
32/40/64 Slice CT Systems32/40/64 Slice CT Systems
http://www.impactscan.org/slides/ukrc2005/img11.html
KanalKanal 1414
�� Two pitch definitions seen in MSCTTwo pitch definitions seen in MSCT
�� Collimator pitch = Collimator pitch = table travel_____table travel_____xx--ray beam widthray beam width
�� Detector pitch = Detector pitch = table traveltable travel
detector widthdetector width
�� Collimator pitch = Collimator pitch = Detector pitchDetector pitch
NN
Pitch in MultiPitch in Multi --slice CTslice CT
c.f. http://www.impactscan.org/slides
KanalKanal 1515
MultiMulti --Row Detector CT Row Detector CT -- AdvantagesAdvantages
�� Same acquisition in shorter times (fewer motion artifacts)Same acquisition in shorter times (fewer motion artifacts)
Seeram, Computed Tomography, 2Seeram, Computed Tomography, 2nd ed.,ed., 2001, pg 263.2001, pg 263.
KanalKanal 1616
MultiMulti --Row Detector CT Row Detector CT -- AdvantagesAdvantages
�� Reconstruction in different slice widths Reconstruction in different slice widths
http://www.impactscan.org/
KanalKanal 1717
MultiMulti --Row Detector CT Row Detector CT -- AdvantagesAdvantages
�� Efficient use of xEfficient use of x--ray beam with increasing slicesray beam with increasing slices�� Reduction of radiation dose Reduction of radiation dose
Kalendar WA, Computed Tomography, 2Kalendar WA, Computed Tomography, 2ndnd ed., 2005, pg 168.ed., 2005, pg 168.
KanalKanal 1818
MultiMulti --Row Detector CT Row Detector CT -- AdvantagesAdvantages
�� Thinner slices for better zThinner slices for better z--axis resolutionaxis resolution
�� Isotropic ImagingIsotropic Imaging
Improved Spatial Improved Spatial ResolutionResolution
All sides of the voxel All sides of the voxel have equal dimensionshave equal dimensions
Seeram, Computed Tomography, 2Seeram, Computed Tomography, 2nd ed.,ed., 2001, pg 264.2001, pg 264.
KanalKanal 1919
Better Multiplanar Reconstruction & 3D DisplayBetter Multiplanar Reconstruction & 3D DisplayMultiMulti --Row Detector CT Row Detector CT -- AdvantagesAdvantages
Radiographics 2002: 22Radiographics 2002: 22--12551255--12681268
KanalKanal 2020
Regarding Multiple Detector CT Regarding Multiple Detector CT scanners:scanners:
0%
43%
29%
14%
14% 1)1) A scanner with a 16A scanner with a 16--detector array employs a total of 16x16 detector array employs a total of 16x16 detectorsdetectors
2)2) Detectors used are the nonDetectors used are the non--radioactive xenonradioactive xenon--gas detectorsgas detectors
3)3) The slice width is determined by width of detectors in slice The slice width is determined by width of detectors in slice
thickness dimensionthickness dimension
4)4) The slice width selected can be smaller than the physical The slice width selected can be smaller than the physical size of the detector size of the detector
5)5) The spatial resolution is worse than that of a singleThe spatial resolution is worse than that of a single--detector detector CT scanners CT scanners
KanalKanal 2121
AnswerAnswer
The slice width is determined by width of detectors in The slice width is determined by width of detectors in slice thickness dimensionslice thickness dimension
KanalKanal 2222
MultiMulti --Row Detector CT Row Detector CT -- Advantages Advantages Cardiac CTCardiac CT
Kalendar WA, Computed Tomography, 2Kalendar WA, Computed Tomography, 2ndnd ed., 2005, Video.ed., 2005, Video.
20 mm detector
(Pitch ~0.25)
6.2 cm in 5 sec
40 mm detector
(Pitch ~0.25)
12.5 cm in 5 sec
KanalKanal 2323
For a 16For a 16--detector row CT scanner, the xdetector row CT scanner, the x--ray beam ray beam width is 20 mm, detector width is 5 mm and the width is 20 mm, detector width is 5 mm and the table travel per rotation is 20 mm. Calculate the table travel per rotation is 20 mm. Calculate the
detector pitch and collimator pitch? detector pitch and collimator pitch?
0%8%15%62%15% 1.1. 1, 41, 4
2.2. 4, 14, 1
3.3. 5, 205, 20
4.4. 4, 44, 4
5.5. 20, 520, 5
KanalKanal 2424
AnswerAnswer
�� 4, 14, 1
�� PitchPitchdd = = 20 20 = 4 = Detector pitch= 4 = Detector pitch
55
�� PitchPitchxx = = 20 20 = 1 = Collimator pitch= 1 = Collimator pitch
2020
KanalKanal 2525
Dose ReductionDose ReductionCurrent Modulation or AEC in CTCurrent Modulation or AEC in CT
�� Modern CT scanners are capable of modulating the mA Modern CT scanners are capable of modulating the mA (current) during the scan(current) during the scan
�� Rationale: it takes fewer photons (lower mA) to penetrate Rationale: it takes fewer photons (lower mA) to penetrate thinner tissue, and more xthinner tissue, and more x--ray photons (higher mA) to ray photons (higher mA) to penetrate thicker projections through the bodypenetrate thicker projections through the body
�� BenefitsBenefits
�� Consistent image qualityConsistent image quality
�� Dose reduction potentialDose reduction potential
�� Reduced tube loadingReduced tube loading
�� Extended scan runsExtended scan runs
�� Reduction in photon starvation artifactReduction in photon starvation artifact
KanalKanal 2626
Automatic Exposure Control (AEC) in CTAutomatic Exposure Control (AEC) in CT
�� Patient size AEC: higher mA used for larger patientsPatient size AEC: higher mA used for larger patientsPatient size AEC
www.isteve.com www.yuzbasicafer1.comMHRA Report 05016, February 2005
KanalKanal 2727
Automatic Exposure Control (AEC)Automatic Exposure Control (AEC)
�� ZZ--axis AEC: higher mA used at more attenuating axis AEC: higher mA used at more attenuating positions along the zpositions along the z--axis (patient)axis (patient)
Z-axis AEC
Radiographics, 2006, 26:503-512MHRA Report 05016, February 2005
KanalKanal 2828
Automatic Exposure Control (AEC) in CTAutomatic Exposure Control (AEC) in CT
�� Angular AEC: the degree of modulation depends on the Angular AEC: the degree of modulation depends on the asymmetry of the patientasymmetry of the patient
Angular AEC
MHRA Report 05016, February 2005 Radiology, 2004, 233:649-657
KanalKanal 2929
Dose ReductionDose ReductionCardiac ECG dose modulationCardiac ECG dose modulation
�� It is possible to vary the mA based on RIt is possible to vary the mA based on R--R interval phase %R interval phase %
�� This can lead to a dose decrease of up to 50% in low stable This can lead to a dose decrease of up to 50% in low stable heart ratesheart rates
70%70% 80%80%
mAmA
200200
550550
tt
interval used interval used for image for image reconstructionreconstruction
dose dose notnot given to patientgiven to patient
KanalKanal 3030
0
10
20
30
40
50
60
70
80
90
0 50 100 150 200 250
Heart rate (BPM)
Exp
osur
e C
TD
I (m
Gy)
Segment mode, no ECG
240o
HR 30-74 BPM (One Segment)
Coverage during scan
Dose Reduction Dose Reduction -- ECG dose modulationECG dose modulation
Paul Kinahan and Kelly Branch, UWMC Paul Kinahan and Kelly Branch, UWMC
KanalKanal 3131
0
10
20
30
40
50
60
70
80
90
0 50 100 150 200 250
Heart rate (BPM)
Exp
osu
re C
TD
I (m
Gy)
Segment mode, no ECG
Segment mode [65-85%], w/ECG
Segment mode [70-80%], w/ECG
240o
HR 30-74 BPM (One Segment), TR 175 msec
Needed coverage during scan
Dose Reduction Dose Reduction -- ECG dose modulationECG dose modulation
42% 42% decrease decrease in dosein dose
Paul Kinahan and Kelly Branch, UWMC Paul Kinahan and Kelly Branch, UWMC
KanalKanal 3232
Regarding automatic tube current Regarding automatic tube current modulation in CT: modulation in CT:
0%
100%
0%
0%
0% 1.1. It gives inconsistent image qualityIt gives inconsistent image quality
2.2. It increases the scan time for the examIt increases the scan time for the exam
3.3. It generates more heat in the xIt generates more heat in the x--ray tuberay tube
4.4. The tube current is automatically adjusted according to The tube current is automatically adjusted according to patient sizepatient size
5.5. It causes an increase in photon starvation artifact It causes an increase in photon starvation artifact
KanalKanal 3333
AnswerAnswer
The tube current is automatically adjusted according to The tube current is automatically adjusted according to patient sizepatient size
Reducing Cardiac Dose Reducing Cardiac Dose –– Prospective Prospective GatingGating
Mahesh, et al. Radiographics, 2007; 27:1495Mahesh, et al. Radiographics, 2007; 27:1495--1509.1509.Kanal
No functional information
Retrospective GatingRetrospective Gating
Mahesh, et al. Radiographics, 2007; 27:1495Mahesh, et al. Radiographics, 2007; 27:1495--1509.1509.Kanal
Kanal Shuman, et al. submitted for publication, 2007Shuman, et al. submitted for publication, 2007
Prospective vs. Retrospective GatingProspective vs. Retrospective GatingUniversity of Washington StudyUniversity of Washington Study
�� 77% lower radiation dose with prospective study77% lower radiation dose with prospective study
�� This study concluded that the This study concluded that the ““64 channel cardiac CT 64 channel cardiac CT performed with prospective ECG gating may have performed with prospective ECG gating may have similarsimilarsubjective image quality and substantially subjective image quality and substantially lowerlower patient patient radiation dose compared to retrospective ECG gatingradiation dose compared to retrospective ECG gating””
Shuman, et al. submitted for publication, 2007Shuman, et al. submitted for publication, 2007
6.2 6.2 ±± 2.0 2.0 26.4 26.4 ±±6.2 6.2 Effective Effective Dose (mSv)Dose (mSv)
Prospective Prospective GatingGating
Retrospective Retrospective GatingGating
KanalKanal 3737
Regarding prospective ECG triggering Regarding prospective ECG triggering for cardiac CT exams for cardiac CT exams
0%
0%
100%
0%
0% 1.1. Projection data has to be collected through out the cardiac Projection data has to be collected through out the cardiac cycle cycle
2.2. The temporal resolution is about 350 msThe temporal resolution is about 350 ms
3.3. The radiation dose is reduced compared to retrospective The radiation dose is reduced compared to retrospective
ECG gatingECG gating
4.4. Functional information of the heart is available with this Functional information of the heart is available with this examexam
5.5. Increase in motion artifacts occursIncrease in motion artifacts occurs
KanalKanal 3838
AnswerAnswer
The radiation dose is reduced compared to The radiation dose is reduced compared to retrospective ECG gatingretrospective ECG gating
KanalKanal 3939
Future of CTFuture of CT
Kalendar WA, Computed Tomography, 2Kalendar WA, Computed Tomography, 2ndnd ed., 2005, pg 235.ed., 2005, pg 235.
KanalKanal 4040
Future of CTFuture of CTOther DevelopmentsOther Developments
Kalendar WA, Computed Tomography, 2Kalendar WA, Computed Tomography, 2ndnd ed., 2005, pg 72.ed., 2005, pg 72.
KanalKanal 4141
TAKE HOME POINTSTAKE HOME POINTS
�� MDCT scanners are becoming common and have several MDCT scanners are becoming common and have several advantages over SDCTadvantages over SDCT
�� CT utilization has increased significantlyCT utilization has increased significantly
�� AEC or Automatic Tube Current Modulation and ECG AEC or Automatic Tube Current Modulation and ECG modulation available on modern scanners which helps modulation available on modern scanners which helps reduce dosereduce dose
�� Cardiac CT is becoming popular with MDCT scannersCardiac CT is becoming popular with MDCT scanners
�� Always keep ALARA in mind!Always keep ALARA in mind!