mc-180ma multi-frequency body composition analyzer

29
MC-180MA Multi-Frequency Body Composition Analyzer

Upload: matia

Post on 30-Jan-2016

60 views

Category:

Documents


0 download

DESCRIPTION

MC-180MA Multi-Frequency Body Composition Analyzer. MC-180MA Introduction. -Table of Contents. MC-180MA Main Feature Advanced Technology, Improvement Advanced Technology Multi-Frequency, Reactance, Dilution method. Measurement Result Fat %, TBW, ECW, ICW, Visceral Fat, etc. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: MC-180MA Multi-Frequency Body Composition Analyzer

MC-180MA Multi-Frequency Body Composition Analyzer

Page 2: MC-180MA Multi-Frequency Body Composition Analyzer

-Table of Contents

MC-180MA Main Feature        Advanced Technology, Improvement

Advanced Technology           Multi-Frequency, Reactance, Dilution method.

Measurement Result          Fat %, TBW, ECW, ICW, Visceral Fat, etc.

MC-180MA Introduction

Page 3: MC-180MA Multi-Frequency Body Composition Analyzer

The Advanced Technology

MC-180MA provide accurate and reliable information by Multi-Frequency BIA

Reactance / Resistance data are output each frequency

New measurement items are added to existing models

MC-180MA Main Feature

Page 4: MC-180MA Multi-Frequency Body Composition Analyzer

Improvements New additional measurement items

by Multi-Frequency BIAIntra / Extra cellular waterVisceral fat, Muscle Balance etc.

High accuracy weighing (type approval class III)minimum value 50g (0-200kg), 100g (200-270kg)

High capacity weighingmax 270kg

Easy to operate and   by color touch panel.320 x 240dots,   5.7inch,   Color TFT Display

MC-180MA Main Feature

Page 5: MC-180MA Multi-Frequency Body Composition Analyzer

External printer connectable (option) (Only EPSON Stylus D88) speedy print onto the exclusive form

PC connectable by USBselectable communication port, USB / RS-232C

MC-180MA Main Feature

Convenient interface High speed measurement less than 20 sec.

(not include profile input time) 

High speed start up less than 10 sec.

   

USB RS-232C Printer port

Page 6: MC-180MA Multi-Frequency Body Composition Analyzer

Multi-Frequency(1)

Low frequency current is unable to pass the cell membrane and it flows outside the cell.

High frequency current is able to penetrate the cell membrane and can also flow within the cell.

MC-180MA Advanced technology

low-frequency current

flowoutside of cell

transmitcell membrane

high-frequency current

Cell

ICW

ECW

Page 7: MC-180MA Multi-Frequency Body Composition Analyzer

Multi-Frequency(2)

MC-180MA Advanced technology

f = 0

f =

Cell membrane capacitor

ECW Resistance

ICW Resistance

Page 8: MC-180MA Multi-Frequency Body Composition Analyzer

朝横になって寝ていたため胴体部に水分が集まっている

夕方

細胞外

Reactance / Resistance(1)

The primary cause of daily fluctuation is changes in impedance value, because the ECW (i.e. intercellular water, blood, lymph fluid, etc.) moves between arms, legs and trunk. The change of impedance value significantly affects the estimation of body composition.

MC-180MA Advanced technology

ECW: Increase by daily fluctuation

ICW: No change

Morning Evening

Page 9: MC-180MA Multi-Frequency Body Composition Analyzer

Resistance (R)

Rea

ctan

ce (

X)

Impedance (Z)

Phase angle (θ)

Reactance / Resistance(2)

MC-180MA not only uses impedance values in the conventional way, but also adopts reactance values that can be acquired by breaking down impedance values under a regression formula to reduce daily fluctuation.

MC-180MA Advanced technology

Page 10: MC-180MA Multi-Frequency Body Composition Analyzer

Dilution method(1)

Total body water (TBW) and Extra cellular water (ECW) are measured by correlations of dilution method and BIA.

A known concentration of marker is administered to the body, and the total water content of the body is calculated from the degree of dilution.

MC-180MA Advanced technology

Body Water

Marker

Marker TBW : D2OECW : NaBr

Page 11: MC-180MA Multi-Frequency Body Composition Analyzer

Dilution method(2)

Alcohol amount: 0.4 littleWater : ?0.4 / (? + 1 ) = 1% ? = 0.4 / 0.01 – 1

? = 40 – 1 ? = 39 little

MC-180MA Advanced technology

Marker

1 little40%(0.4little) 1%

? little ?+1 little

Page 12: MC-180MA Multi-Frequency Body Composition Analyzer

Comparison with the currentsModel Name BC-418 BC-420 MC-180

Weight Measurement 0.1kg/200kg 0.1kg/270kg 0.05kg/1-200kg

0.1kg/200-270kg

Number of electrodes 8polar 4polar 8polar

Measurement Frequency

50 k Hz 50 k Hz 5, 50, 250,

500 k Hz

Measurement Impedance

Impedance Impedance Resistance

Reactance

Measurement Range 150 ~ 1200Ω 150 ~ 1200Ω 75 ~ 1500Ω

Input device Membrane switch

Membrane switch

Touch panel

Display device 3 Rows, 5 Digits

LCD

3 Rows, 5 Digits

LCD

320 × 240 dots

TFT color LCD

Data Interface RS-232C RS-232C USB, RS-232C

Printer internal thermal printer

internal thermal printer

external A4

printer (option)

Page 13: MC-180MA Multi-Frequency Body Composition Analyzer

Comparison of resultsModel Name BC-418 BC-420 MC-180

FAT % DXA(1) DXA(2) DXA(3)

TBW, TBW % from FFM Dilution method Dilution method

ICW , ECW Dilution method

Visceral Fat rating MRI MRI

BMR breath gas analysis

breath gas analysis

breath gas analysis

Muscle mass DXA DXA DXA

Bone mass DXA DXA

Segmental read out DXA DXA

      Fat % DXA DXA

      Muscle mass DXA DXA

      Fat %   Score Average

      Muscle Score Average

Page 14: MC-180MA Multi-Frequency Body Composition Analyzer

Body FatIn MC-180, Body Fat % and Fat Mass

are measured by correlation of DXA and BIA.

MC-180MA Technical Notes

MC-180 evaluates segmental Fat data. These data is used to check a figure and body balance.

Page 15: MC-180MA Multi-Frequency Body Composition Analyzer

y = 0.9646x + 1.0186R = 0.93

0

10

20

30

40

50

60

0 10 20 30 40 50 60

MC- 180MA %FAT

DXA %

FAT

Validation data Fat %

MC-180MA Technical Notes-reference-

Page 16: MC-180MA Multi-Frequency Body Composition Analyzer

Resistance (R)

Rea

ctan

ce (

X)

Impedance (Z)

Phase angle (θ)

Fat Free Mass (FFM)

TANITA has been measuring FFM by using the DXA always. In MC-180, TANITA achieved higher accuracy by adopting

advanced technology.Multi-Frequency BIA and Reactance measurement.

MC-180MA Technical Notes

+

Page 17: MC-180MA Multi-Frequency Body Composition Analyzer

y = x + 0.09R = 0.98

0

20

40

60

80

100

0 20 40 60 80 100

MC- 180MA FFM

DXA F

FM

Validation data FFM

MC-180MA Technical Notes-reference-

Page 18: MC-180MA Multi-Frequency Body Composition Analyzer

Total Body Water, Intra / Extra Water

TBW ECW and ICW including blood and lymphECW consists interstitial fluid, blood plasma, lymph etc.

Edema is the state that interstitial fluid has increased.ICW the fluid that exists in the cell

ICW tends to increase to the muscular person because muscle tissue contains a lot of water.

MC-180MA Technical Notes

TBW (45-65%) fat, protein, mineral etc.

ECW (30-50%) ICW

Page 19: MC-180MA Multi-Frequency Body Composition Analyzer

General transition of body waterThis is one example. It is neither a normal value nor a standard value.

ICW

30%ICW

35-40%

ICW

30%

TB

W75%

EC

W40%

ICW

35%

Body w

eight

TB

W65%

EC

W30%

ICW

35%

TB

W60-65%

EC

W20-30%

TB

W55%

EC

W25%

TB

W50%

EC

W20%

ICW

30%

TB

W50%

EC

W25%

babynewborn infant child adult –male- adult -female- elderly adult

-reference-

TBW ECW ICW

newborn 75 40 35

infant 65 30 35

child 60-65 20-30 35-40

male 55 25 30

female 50 20 30

Factor that influences body waterIn elderly , ICW decreases because the muscle mass

reduce.In female, TBW and ICW are little than male,

because the amount of muscle is comparatively few.Adipose tissue don't contain water,

therefore, in obese person, ICW is a little.

MC-180MA Technical Notes

Page 20: MC-180MA Multi-Frequency Body Composition Analyzer

y = 0.9999x + 0.0026R = 0.9715

0

20

40

60

80

0 20 40 60 80

MC- 180MA TBW

Dilution T

BW

Validation data TBW

MC-180MA Technical Notes-reference-

Page 21: MC-180MA Multi-Frequency Body Composition Analyzer

Visceral Fat Rating

In TANITA body composition analyzer, visceral fat rating is evaluated by correlations of the Visceral Adipose Tissue area in MRI and BIA.

MC-180MA Technical Notes

BMI 25 BMI 25(Age 27 Male) (Age 61 Male)

Visceral fat is accumulated when becoming aged.

TANITA has been researching the method of measuring the visceral fat, and established the method to estimate it from BIA.

Rating from 1 - 12

Indicates you have a healthy level of visceral fat. Continue monitoring your rating to ensure that it stays within the healthy range.

Rating from 13 –

Indicates you have an excess level of visceral fat. Consider making changes in your lifestyle possibly through diet.

Page 22: MC-180MA Multi-Frequency Body Composition Analyzer

y = 0.99x -  0.10R = 0.85

0

50

100

150

200

250

0 50 100 150 200 250

MC-180MA VFA (cm2)

MRI VFA (cm2 )

Validation data Visceral Fat-reference-

MC-180MA Technical Notes

Page 23: MC-180MA Multi-Frequency Body Composition Analyzer

Basal Metabolic Rate (BMR)

The BMR is more determined by the Fat Free Mass (FFM) than by the body weight.

MC-180MA Technical Notes

In MC-180, the BMR is estimated by the correlation of Breath Gas analysis and FFM. Of course, FFM is measured by the BIA method.

Page 24: MC-180MA Multi-Frequency Body Composition Analyzer

Validation data BMR

R = 0.900

MC-180MA Technical Notes-reference-

Page 25: MC-180MA Multi-Frequency Body Composition Analyzer

Muscle MassIn MC-180, the muscle mass is

measured by the method using the correlation of DXA and BIA.

MC-180MA Technical Notes

MC-180 evaluates segmental muscle mass. These are convenient to check the effect of training and rehabilitation.

Page 26: MC-180MA Multi-Frequency Body Composition Analyzer

y = 0.9769x + 1.2902R = 0.97

20

40

60

80

100

120

20 40 60 80 100 120

BIA Muscle mass

DXA L

ST (Mus

cle mas

s)

Validation data Muscle Mass

MC-180MA Technical Notes-reference-

Page 27: MC-180MA Multi-Frequency Body Composition Analyzer

Estimated Bone Mass

This feature indicates the amount of bone (bone mineral level, calcium or other minerals) in the body.

•Persons described below may obtain varying readings and should take the values given for reference purposes only.

- Elderly persons - Women during or after menopause- People receiving hormone therapy•“Estimated bone mass” is a value estimated statistically based on its

correlation with the fat-free amount (tissues other than the fat). "Estimated bone mass" does not give a direct judgment on the hardness or strength of the bones or the risks of bone fractures. If you have concerns over your bones, you are recommended to consult a specialist physician.

MC-180MA Technical Notes

Female : Average of estimated bone mass (kg) Male : Average of estimated bone mass (kg)Weight Weight

Less than 50kg 50kg-75kg 75kg and up Less than 65kg 65kg-95kg 95kg and up

1.95kg 2.40kg 2.95kg 2.66kg 3.29kg 3.69kg

Page 28: MC-180MA Multi-Frequency Body Composition Analyzer

y = 1.0066x + 0.0164R2 = 0.78R = 0.88SEE = 0.30p < 0.0001

0.0

1.0

2.0

3.0

4.0

5.0

6.0

0.0 1.0 2.0 3.0 4.0 5.0 6.0

BIA Bone mass

DXA B

one mas

s

MALEFEMALE

Validation Data Bone Mass-reference-

MC-180MA Technical Notes

Page 29: MC-180MA Multi-Frequency Body Composition Analyzer

y = 0.9997x + 0.0017

R2 = 0.3911

0

20

40

60

80

0 20 40 60 80

MC- 180MA ECW

Dilution E

CW

y = 1.0002x -  0.0012

R2 = 0.5451

0

20

40

60

80

0 20 40 60 80

MC- 180MA ECW

Dilution E

CW

Validation data ECW

MC-180MA Technical Notes-reference-

-Male- -Female-