Marek OCHOCKI, Slawomir LATOS

Institute of Medical Technology and Equipment ITAM,

118, Roosevelt Str., Zabrze, Poland

INFLUENCE OF DEFIBRILLATION

TREATMENT ON SOME ELECTRICAL

PARAMETERS AND APPLICATION

SAFETY OF ECG ELECTRODES

Page 2: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Presentation scheme

• The subject of presentation

• Introduction

• Examination procedure

• Examination results

• Conclusions

• Wider context

Page 3: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

The subject of presentation

• A contribution to determination of safety of disposable

ECG electrodes use in conditions of defibrillation

treatment in monitored patients for risk stratification

and referential purposes

ECG electrodes available on the Polish market 4 European manufacturers including Polish one

• Illustration of relations between patient’s safety,

standards, market considerations, compliance to

standards and regulations as well as research activity

Page 4: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Introduction

Fig. 1 ECG electrode filled with gel

Page 5: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Introduction

Fig. 2 Schematic representation of electrode-gel-patient’s skin structure and its equivalent circuit model according to [3]

Page 6: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Introduction• ECG recording or monitoring• The use of cardiac defibrillation treatment

– Defibrillation voltages transposed up to 200 V between ECG electrodes on patient’s body

• Hazards to ECG electrodes and patients– Up to 2% of defibrillation current absorbed by

electrodes, mostly in sensor Ag/AgCl structure– Degradation of sensor layer - impairment of electrode

characteristics [3] – Electrodes polarization (publication [6])– Rise of electrodes impedance – next slide:

Page 7: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Increase of ECG electrodes impedance may cause:

• unwanted decrease of common mode rejection ratio (CMRR) of ECG amplifier channel.

• raise of disturbing interference of alternative current on ECG waveform [5].

• augmentation of exuded energy during defibrillation or even electrosurgical operation involving possible heating up of electrodes and the patient's burns [3].

Page 8: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Procedure of ECG electrodes examination

• Selection of electrodes• Tests distribution in years• Gluing electrodes in gel-to-gel mode into pairs –

correspondence to impedance measurement on abraded patient’s skin [4]

• Measurement of polarization potential of electrodes 60÷90 s after gluing

• Measurement of impedance with the 10 Hz current• Fourfold defibrillation recovery test − according to

schematic diagram in Fig. 4• Measurement impedance with the 10 Hz current after

defibrillation

Page 9: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Testing according to globally recognized standard:

American National Standard

ANSI/AAMI EC12:2000/(R)2010

Disposable ECG electrodes

AAMI Association for the

Advancement of Medical

Instrumentation

Page 10: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

The photo of measuring equipment set for ECG electrodes testing

Page 11: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Fig. 3 Measurement circuit for testing the electrodes impedance with 10 Hz current

oscilloscope

sinusoidal current source of efficiency

80µA/10Hz Ka=12,5V/V

measuring amplifier

tested electrode pair

Page 12: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Fig. 4 Instrumentation system for defibrillation recovery testing of ECG electrodes according to standard [1]

Page 13: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Fig. 5 Electric scheme for defibrillation recovery

testing of ECG electrodes according to [1]

Page 14: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Results

• Initial self-polarization potential („DC offset voltage”), measured in simple circuit after 60÷90 s stabilization period of glued electrode pairs did not exceed 6 mV in all the measurements of 2012 year.

• In previous years indicated, these voltages were also far below 100 mV upper allowable limit according to [1]

Page 15: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Polarization recovery from defibrillation induced overload in 2012 tests

• Measurements 5 secs after each of four discharges of 200 V did not exceed 21 mV in all tests of 2012 year. The upper permissible limit being 100 mV [1]

• Rate of decrease of induced polarization potential during 30 secs following each ‘5 secs’ measurement did not exceed 0,5 mV/s (permissible maximum 1 mV/s [1])

Page 16: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Fig. 6 Example of recovery polarization potential in time after a series of four defibrillation-like expositions impacting the electrode pair chosen

0

5

10

15

20

25

30

00:00 00:15 00:30 00:45 01:00 01:15 01:30 01:45 02:00 02:15 02:30 02:45 03:00 03:15 03:30

Time [min:s]

Po

lari

zati

on

po

ten

tial

[m

V]

Page 17: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Table 1. Results of impedance measurements for electrode pairs examined, in four selected years

Mea-suredImpe-dancePara-

meters

Years of examinations conducted

2006 2007 2011 2012

Each yearone electrode type

from one producer,and quantity of

12 electrode pairs tested

Electrode types of one producer and quantity of electrode pairs tested

A7 pairs

B12 pairs

C12 pairs

D12 pairs

Impe-dance

interval before

defibrilla-tion k

0,29÷ 0,38

0,41÷ 0,48

0,04÷ 0,06

0,21÷ 0,41

0,23÷ 0,46

0,18÷ 0,29

0,16÷ 0,34

Page 18: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Meanimpedance value before

defibril-lation

k

0,34 0,44 0,05 0,27 0,34 0,22 0,22

Impe-dance

interval after

defibrilla-tion k

0,28÷ 0,32

0,30÷ 0,37

0,03÷ 0,04

0,17÷0,37

0,08÷ 0,22

0,07÷ 0,13

0,12÷ 0,16

MeanImpedan-ce value

after defibril-

lation k

0,30 0,34 0,04 0,23 0,15 0,10 0,15

Page 19: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

End of Table 1.

Interval of

impe-dance

drop %

0,0÷ 23,7

9,9÷ 31,4

17,2÷ 29,5

2,2÷ 21,9

46,9÷ 71,7

51,1÷ 69,6

12,1÷ 55,4

Mean value of

impe-dance

drop %

10,8 21,8 22,3 13,8 56,3 57,0 31,2

Page 20: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Changes of electrode impedance

In result of fourfold defibrillatory exposition the impedance drop in all examined electrode pairs has occurred.

Mean value of this drop for each electrode lot falls into 10,8% do 57% interval.

Checked ECG electrodes comply with the American standard [1], to not exceed 2 kΩ by the mean impedance value for each production lot of 12 electrode pairs, and individually be below 3 kΩ.

Page 21: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Conclusions

• All tests have been complied to globally recognized American standard [1] with considerable safety margin.

• The results of conducted examinations generally have put in a good light the quality of tested ECG electrodes of four selected European manufacturers, including Polish one,

• This made possible safe use of these electrodes on patients undergone the serious treatment as cardiac defibrillation.

• In that important was (or is) to allow the physician obtaining quickly information for assessment of effectiveness of this defibrillation procedure.

Page 22: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Conclusions

• Referential meaning of these examinations - wider analyses, risk stratifications…

• On today market more and more medical devices appears from outside Europe. their quality is not stable.

• Periodic checking of accessible sample lots of such products - advisable

• Special care for long-term and little controlled transport and storage periods as well as rough conditions

• Particular care for semi-liquid materials like the gels.• Essential performance – new kind of safety in

standards for electromedical equipment – here represented

Page 23: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

For further analysis two other low intensity

parameters also positively tested

meaningful phenomena, required by the American standard [1] :

1. Combined offset instability and internal noise

After a 1-min stabilization period, a pair of electrodes connected gel-to-gel shall not generate a voltage greater than 150 microvolts (μV) p-p in the passband of 0.15 to 100 Hz, for a period of 5 min following the stabilization period.

Page 24: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

Fig. 7 Test circuit for offset instability and internal noise measurement [1]

Page 25: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

2. Bias current tolerance:

Observed DC voltage offset change across a glued pair of electrodes required to not exceed 100 mV when the pair is subjec-ted to a continuous 200 nano-ampere DC current during

at least 8 hours.

Page 26: Marek OCHOCKI, Slawomir LATOS Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland INFLUENCE OF DEFIBRILLATION TREATMENT

BIBLIOGRAPHY

[1] ANSI/AAMI EC 12:2000/(R)2010 Disposable ECG electrodes.[2] EN 60601-1:2006 Medical electrical equipment. Part 1: General

requirements for basic safety and essential performance.[3] HOI-JUN YOO, VAN HOOF CH. Bio-Medical CMOS ICs (Integrated

Circuits and Systems) 2011, pp 31-124, Springer, New York[4] KLINGLER DR, SCHOENBERG AA, WORTH NP, EGLESTON CF,

BURKART JA. A comparison of gel-to-gel and skin measurements of electrode impedance. Med. Instrum. 1979 Sep-Oct;13(5):266-8.

[5] OLSON WH, SCHMINCKE DR, HENLEY BL. Time and frequency dependence of disposable ECG electrode-skin impedance. Med. Instrum. 1979 Sep-Oct;13(5):269-72.

[6] PATTERSON RP. The electrical characteristics of some commercial ECG electrodes. J Electrocardiol. 1978 Jan;11(1):23-6

[7] SCHOENBERG AA, BOOTH HE, LYON PC. Development of standard test methods for evaluating defibrillation recovery characteristics of disposable ECG electrodes. Med. Instrum. 1979 Sep-Oct;13(5):259-65