Thermoguard

Thermoguard Report

Version 2.91

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 2 of 47

Contents

Basics

- Overview ............................................................................................................. 3 - Installation .......................................................................................................... 4 - Starting Thermoguard Report ........................................................................... 6 - Command Line Switches ................................................................................... 7 - Program Language / Logging ........................................................................... 8 - Path Settings ...................................................................................................... 9 - Illustration: Example of a typical Installation Scenario ................................ 11 - Choosing a Sensor .......................................................................................... 12 - Save / Load a Sensor Selection ....................................................................... 13 - Refresh Sensor List .......................................................................................... 14

Thermoguard Report Chart

- Selecting a Period of Time for drawing a Chart ............................................. 15 - "Calculate only (no chart)" and "Details" ...................................................... 17 - The Chart Window ("Thermoguard Chart") .................................................... 19 - Additional Information about the Chart Window ........................................... 23 - Chart Batch Print .............................................................................................. 25 - The Annotation Editor ..................................................................................... 26

Thermoguard Report Live!

- Principle ............................................................................................................ 33 - How to open a sensor's Live! Window ........................................................... 33 - Running Thermoguard Report Live! ............................................................... 34 - Alarm Visualization in Value Mode ................................................................. 37 - Error Messages, their visualization and Causes ........................................... 38 - Live! Window Sets ........................................................................................... 40 - Visual Mode / Visual Designer ........................................................................ 42

Appendices

- A1 For Administrators: Configure fixed Path Presets .................................. 45 - A2 Freeware Program "allSnap" ..................................................................... 45 - A3 Arithmetic Mean and Mean kinetic Temperature ..................................... 46 - A4 Standard deviation ..................................................................................... 47

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 3 of 47

Basics

Overview Thermoguard Report (abbr. "TG Report") is used for analysis and graphic display of data val-ues, which are continuously recorded by the Thermoguard main program. TG Report provides three main functions:

Thermoguard Chart draws a chart over a period of time specified by the user from the recorded values of a sensor (see cover image top right). Thermoguard Chart allows a vast number of possibilities: Navigation, scaling, zooming, three-dimensional display and much more. Thermoguard Charts can be printed and saved as a graphic file. An unlimited number of Thermoguard Chart windows can be opened at the same time. The integrated Annotation Editor provides the possibility to add annotations (com-ments, pictures) directly to a chart.

Thermoguard Live! displays the current ("last") value of a sensor; either as digital dis-play (see cover image bottom right), as a continuously floating chart or as a gauge dis-play on top of a background image ("Visual Mode", see cover image top left). Again, an unlimited number of windows can be opened at the same time. Thermoguard Live! win-dow sets can be saved and re-opened. As the Chart windows the Thermoguard Live! windows can be scaled to any size; i.e. for presentation purposes.

Calculating Mean Arithmetic values and Standard Deviation of all measured units as well as the Mean Kinetic Temperature (MKT) over any arbitrary period for temperature data additionally.

TG Report can be used as an analysis and visualization program for employees on any work-station computers - independently of other programs of the Thermoguard family. All that TG Report requires for access to the recorded data is read-only access to the Thermo-guard installation and data directory. No write or administrative rights are required for the Thermoguard installation and data directory. This manual is an addition to the Thermoguard System Manual.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 4 of 47

Installation Installation types

Two options are available for installing TG Report:

Local installation on the same computer where the main program (TGuard) runs; i.e. for checking data on the part of the Thermoguard system administrator

Installation on one or more workstation computers on the network, e.g. for data visualization through employees.

Administrators may consult the separate documentation TG Silent Installation-

en.pdf located in the root directory of the Thermoguard CD for command line switches

providing an unattended installation of Thermoguard Report. TG Report gleans the information about the number of, label, location, and limits of and other settings for the sensorcontrollers from the INI, LIC and SCX files from the Thermoguard data directory. The recorded data (temperature, humidity or any other unit) is read by TG Report

from this directory (by default) as well (*.tg files in subdirectory \tg). TG Report requires

read access to the data directory. For a local installation the data directory usually is

%ALLUSERSPROFILE%\Thermoguard Data\tg. Herein %ALLUSERSPROFILE% is a system

environment variable pointing by default to the following paths:

for Windows XP/Server 2003: C:\Documents and Settings\All Users

for Windows Vista/Windows 7/Server 2008(R2): C:\ProgramData

For a network installation the path to the data directory can be configured in TG Report, how-ever. When installing TG Report on a workstation computer on the network, the system administrator should typically create a share of the Thermoguard data directory on the network, including read permissions for employees.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 5 of 47

Executing the installation

TG Report is installed as a component of the Thermoguard Setup program (TGuard_290.exe):

TG Report is installed in the \TGReport

subdirectory of the selected installation directory, i.e. by default in %ProgramFiles(x86)%\Thermo-

guard\TGReport.

For details on the Setup program, please refer to the Thermoguard system manual.

During the installation you can decide whether to create a desktop shortcut. By default this option is not checked.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 6 of 47

Starting Thermoguard Report There are up to three possible ways to start TG Report:

1) Doubleclick the desktop shortcut (if option "Create desktop shortcut" selected on setup)

2) Via start menu 3) Of course also directly by running the program file

"%ProgramFiles(x86)%\Thermoguard\TGReport\TGReport.exe".

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 7 of 47

Command Line Switches

The program file TGReport.exe supports three command line parameters:

/CfgDataDir:<Directory>

Sample call: TGReport.exe /CfgDataDir:"C:\Program Files\Thermoguard\TGReport"

This switch may specify a different than the default configuration directory (%ALLUSERSPRO-

FILE%\TGReport Data). The TGReport.ini file is located in this path.

Notes:

This is not the configuration path of the Thermoguard main program.

If the path is invalid TG Report is aborted with an error message:

The TGReport.ini file is not yet present after a new installation. It will be created the

first time the program is terminated. /tgl:<filename>

Sample call: TGReport.exe /tgl:"%ALLUSERSPROFILE\TGReport Data\My windows"

A previously saved Live! windows set can be specified. This possibility is described in the sec-tion Thermoguard Report Live! in detail. Remember to enclose paths containing spaces in quotes. Note: The Live! windows which were active at program termination are automatically restored on startup. Therefore the former switch /l has become obsolete now.

/h

Hides the main window after program start automatically (if at least one Live! window was ac-tive at program termination). Hides the mouse cursor too. All three switches can be combined; their order is not important.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 8 of 47

Program Language

The program language can be set to either English or German. To change it from German to English use the menu item "Einstellungen" => "Englisch" as shown below. In order for the change to take effect, the program must be restarted.

Logging

Menu item "Logging" activates or stops Logging for Thermoguard Report:

20 log files à 2MB will be written. If the 20th file exceeds 2 MB the oldest file will be overwrit-ten. Thus a maximum of 40MB free disk space is needed for log files. Files will be created in

the subdirectory \log of the configuration path, i.e. by default this is

%ALLUSERSPROFILE%\TGReport Data\tg\log. The current log file is always named TG-

Report.log, the other files are named TGReport00xx.log.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 9 of 47

Path Settings After starting, the TG Report Main Window opens:

Configuration and data path for accessing the sensor configuration and data

TG Report will search the TGuard.ini file in the default

configuration path of a Thermoguard installation (%ALL-

USERSPROFILE%\ Thermoguard Data) if it is installed

onto the same computer as the Thermoguard main pro-

gram. It then reads the path to the *.tg files from this file.

Settings are saved to the TGReport.ini file and will be

used again on next start.

Setting the configuration and data path ma-nually

Settings for the path to the main program's configuration

files and to the *.tg files may be set manually once by

the user, e.g. in case TG Report does not reside on the same computer as the main program. Use menu item "Options" => "Configure paths".

If the path to configuration data of the main program is

changed, the path to the *.tg files is read from the

TGuard.ini file which resides in the configuration path

and the path is displayed in the lower field automatically.

However, the *.tg path can be set individually too. The

input field will be enabled as soon as the checkbox "User-defined path to *.tg files" is checked. If unchecked, imme-

diately the *.tg path is read from the TGuard.ini file

again and set.

In the example shown on the left a read only share has been established on the main program's configuration

directory and mapped as drive T:.

Note: Shares ("UNC" notation) can also be used, e.g.

\\PC01\TGData. However, drive mappings usually pro-

vide better performance.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 10 of 47

Information about used paths An information window displays the current paths used by TG Report. It can be called via key

or menu item "Help" => "About / Path Information":

==>

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 11 of 47

Illustration: Example of a typical Installation Scenario

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 12 of 47

Choosing a Sensor In the upper area of the main window is the sensor list. You can resize the window vertically by grabbing it at its lower border and dragging the mouse thus enlarging the sensor list as well:

The list can be sorted by Label or Location by clicking the respective column header. On an initial program start sort order is al-ways reverted to "sort by Label".

If no sensor is shown in the fields "Label" or "Lo-cation", the configuration path setting might not be correct. Please check the path settings as described above. Tip: If a path's length should exceed the width of the status bar at the bottom of the window, move the mouse over the status bar. Then the complete path will be shown as a tool tip.

Now you select one or more sensors in the sensor list to which you want apply the various Thermoguard Report functions. To do so you may use the keyboard shortcuts as is standard in Windows, e.g. CTRL+A (select all), hold CTRL and click with mouse to add an entry etc.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 13 of 47

Whether a certain function can be applied depend on the number of sensors selected (i.e. highlighted) in the sensor list:

If one or more sensor are selected:

"Draw Chart" (Button) respectively "Calculate and draw chart" (context menu item)

Only if two or more sensors are selected:

"Chart Batch Print"

Only if a single sensor is selected:

"Calculate only (no chart)"

"Live!"

"Visual Designer for selected Sensor"

Save / Load a Sensor Selection You can save the current sensor selection under a customized name and reload it later. This is useful to quickly select a number of sensors which belong together logically - e.g. to batch print their charts using the Chart Batch Print function.

After loading a sensor selection the associated sensor entries are highlighted in the sensor list. Please note that Saving / loading a sensor selection does not mean, that only the saved sen-sors are shown in the sensor list excluding all others. The sensor is always populated with all available sensors.

After selecting several sensors (here three sensors in the same "Room 131", click on "Save sensor selection" in the context menu and follow the steps illustrated below:

Note: A sensor selection is written to the TGReport.ini file to a section named

[*Sel-NameOfSelection].

This dialog supports overwriting an exis-ting selection as well. In this case a safety query pops up first.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 14 of 47

Loading a sensor selection is done analogously via context menu item "Load sensor selection":

Now, the three sensors located in Room 131 are highlighted again

In case a sensor of a saved selection does not exist any longer, an appropriate warning is shown:

Refresh Sensor List

Via Extras menu|item "Refresh sensor list" or via the corresponding last context menu item the sensor list can be reloaded. This is mandatory to reflect any changed sensor settings by using Thermoguard's main program, e.g. new sensors added recently or reconfigured violation limit values of existing ones:

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 15 of 47

Thermoguard Report Chart

Selecting a Period of Time for drawing a Chart To draw a Chart for a selected sensor, the start and end date and time must be selected first:

Selecting the start and end date

The selection fields for the calendar date and time ("From:" and "To: ") enable limiting the period for the de-sired analysis accurately by seconds.

You can automatically set the current system date as the end time. Use one of these possibilities:

Click on "To:" button (shown left)

"To:"=Current system time in Options menu

"To:"=Current system time in context menu

Keyboard shortcut "CTRL+E" Note: If you click "To:" again, the previously entered Date is re-established (and not the current date as with previ-ous program versions).

Analogously "Very first data" can be selected as start date. Use one of these possibilities:

Click on "From:" button

"From:" = Very first data in Options menu

"From:" = Very first data in context menu

Keyboard shortcut "CTRL+F" Note: If "To:"=Current system time has been set additio-nally, very large amount of data may be available for processing. In this case you should consider not to click on "Draw Chart" directly, but execute "Calculate only (no chart)" first to get an impression of the data count. In the example shown left, 1,5 Mio. measurement values exist.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 16 of 47

If no data can be found in the data directory for the select-ed period, TG Report will send you a message in the sta-tus bar: "No data found within specified range" or "Nothing to do!" (after clicking "Draw Chart").

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 17 of 47

"Calculate only (no chart)" and "Details"

"Calculate only (no chart)"

If just a single sensor is selected you can process the measured values for the selected period without dis-playing the chart window: Click "Calculate only (no chart)" in the context menu. The arithmetic mean ("Arithmetic Mean Temperature/Hu-midity or [Unit]") as well as the Standard deviation is cal-culated. Additionally - if the sensor's unit is Temperature - the Mean Kinetic Temperature (MKT) for the selected period is calculated and displayed. See Appendices A3 and A4 for mathematical background.

Example for temperature data [°C]

For an accurate calculation of the mean kinetic tempera-

ture, the exact value of the activation energy H of the monitored pharmaceutical substance, to be determined calorimetrically, can be entered in the top area of the dia-log shown left. If the value is unknown usually a default value of 83.144 kJ/mol is used which can be set any time by clicking the "Default" button. For other units than Temperature (Relative Humidity or any other unit recorded by an AC2nn device), the Mean Kinetic Temperature cannot be calculated. Only the Arith-metic Mean and the Standard deviation are displayed. Note: If the sensor is configured as "Switch" or "Alarm Relay Channel", all three output fields under "Results" are grayed out, because the associated calculations are not applicable.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 18 of 47

Example for humidity data [%RH]

Example for barometric pressure [hPA]

Details ON

"Details ON/OFF"

The "Details ON" or "Details OFF" button allows showing or hiding the details for the measurements within the se-lected period.

- Start/end and length of period - Number of data items, errors, low and high alarm lim-

its as well as minimum and maximum values. The numbers put in brackets indicate how often the lowest or highest value occurred within the period specified.

- Number of exceeded limits as well as number of trig-gered alarms, separated into Min/Max and Total (=Min + Max) numbers.

The "Details ON/OFF" function is available within the Chart window also, which is described in the next chapter.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 19 of 47

The Chart Window ("Thermoguard Chart") By clicking the "Draw Chart" button, data for the specified period is collected and the corres-ponding chart is displayed. This function can also be applied to several sensors. However, if the entered time span is long and many sensors are selected the processing job can last signi-ficantly long. Therefore you can specify if a safety query shall pop up if a certain number n of sensors is exceeded: This setting can be found under Options menu|Miscellaneous settings (lower area of dialog):

=> =>

In the example above, the window shown on the right is displayed if more than 5 sensors are selected before executing "Draw Chart". Enter "0" to disable the safety query completely. If a chart cannot be created for one or more sensors - e.g. because no data could be found for them - you are being informed by a popup message:

For the same sensor multiple chart windows can be opened - e.g. for different periods.

Thermoguard Chart

The "Draw Chart" button lets you open a window contain-ing a graphic display of the measured data. When opening the chart window, all data of the selected period will be displayed.

Vertical alignment depends on the minimum/maximum alarm limit values set for a sensor.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 20 of 47

Via the Chart menu you can invoke the Annotation Editor (or

directly by pressing + ). See separate chapter.

"Chart" Menu: Printing and Saving the graphic The "Print" item allows printing the currently displayed

graphic. You may use + alternatively.

The size of the printed graphic depends on the currently set window dimensions. In order to achieve maximum print quality you therefore should maximize the window before printing.

The "Save as PNG" ( + ) item allows saving the cur-

rently displayed graphic in "Portable Network Graphics PNG" format for usage with other programs. Possibly displayed details (via "View" menu => Details on/off) are printed (respectively saved to file), too. The window frame will be ignored though.

"Save as PNG (only inner area)" saves the area within the x and y axis only excluding the scale labels. This may be useful when combining to charts using the Annotation Editor.

Notice that the size of the saved graphic depends on the currently set window dimensions.

"Save as PDF" creates a PDF file. Here again the saved size depends on the current window size.

"Options" Menu: "Dateformat ..." items: Toggles the date format which is used for labeling the chart's time axis. "Hide/Show main window" allows to fade out the main window in case it is not needed or "disturbing" (e.g. when a permanent control scenario using several Thermoguard Live! windows at the same time is active).

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 21 of 47

"Options" Menu continued "X axis: First/last value found" or "X axis: Force period entered in main window" When selecting the second of these options the output only differs compared to the first if no data had been re-corded for the sensor at the beginning and/or at the end of the period selected in the main window. In this case the affected period is left "blank". The following sample illustrates the result:

The selected Start Date is November 11, 2001. However, the first available value has been recorded not before November 20, 2001. Using option "X axis: First/last value found" the display period starts with the Date of the first mea-sured value. The resulting chart looks as shown left.

Using option "X axis: Force period entered in main window " the complete period as selected in the main window is displayed including dates with no data. I.e. the "empty" period from November 11 until November 20 is also shown as blank space at the beginning of the chart when the window opens. Now, the resulting chart looks as shown left. Selecting this option makes sense if you want to compare charts for different sensors and for one of them no data is available at the start or end of the desired display period. A "manual" adjustment would be very inconvenient here. Note that the setting of this X axis option also affects the "Reset chart display" function of the "View" menu.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 22 of 47

Chart window using 3D mode and details

Key F4 increases the line width, F3 decreases it.

F5 sets one of 8 preset colors for the line color. F6 opens a dialog where you can pick any color. However, if the selected color is "too yellowish" (reserved for FAIL)

or to bright, an error message is displayed. Then, you must select a different color.

"View" Menu: "Show High/Low limit lines" allows hiding the red and blue high and low alarm limit lines.

Always the alarm limits currently set in the Thermoguard main program are displayed.

The "Show Details " item (or the key) allows showing or hiding measurement details within the selected period as in the main window. "Y-Range: smallest/largest value": By pressing the ENTER key, the y-axis is adjusted to the smallest and largest y values present in the current section. "Y-Range: High/Low limit lines": By pressing Shift+EN-TER, the y-axis scaling is based on the the values of the alarm limit lines. "Y-Range: set own range": By pressing Alt key+ENTER a dialog popups up where you can set fixed min and max y-values. This can be useful when combining two charts with the annotation editor. The values are saved sensor specifically if the Save Chart settings sensor specific option is selected (main window, Options menu).

"Reset chart display": The SPACE bar restores the display used on initially opening the chart window. All other items are self-explanatory.

(Image cropped)

"Help" Menu: Thermoguard Chart Window Command Ref-erence Opens a table containing all mouse and keyboard short-cuts for customizing and modifying the graphic display. We recommend that you try all listed key and mouse opti-ons to get a feeling for the optimum "navigation"/view of the displayed graphic. Additionally, functions for variable scaling of the x- and y-axis and for zooming in and zooming out are available.

Most important mouse functions:

Hold the left mouse button while moving the mouse to move the window contents horizon-tally or vertically.

Hold the right mouse button to select a frame to enlarge ("Zoom").

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 23 of 47

Additional Information about the Chart Window

You can open as many Chart windows as you want; even for the same sensor.

Chart settings can be saved individually for each sensor. The menu item of the "Options" menu of the main window is shown below:

If the option is not set, the settings from the last Chart windows closed are stored in the configuration file TG-

Report.ini under the section [TfrmChart] and are re-used next time, when any Chart window is opened

again.

If the option is set, the settings are stored for each sensor individually in a section [sensor label].

Please note the importance of this setting for batch printing several charts for different sensors (next chapter).

Display of missing data:

Missing data are represented by a continuous line and may not be recognized easily, as shown in the example!

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 24 of 47

Display if a sensor is configured as "Switch" or "Alarm Relay Channel"

A sensor used with the "Switch" or "Alarm Relay Channel R1/R2" func-tion shows a special On/Off chart (these functions are described in the Thermoguard System Manual). Line color for "FAIL":

Data lines with the entry FAIL in-stead of real values are represented with a yellow line from the last real data point before to the next real data point.

Displaying High/Low limit lines depending on measurement period: Starting from version 2.90 the Thermoguard main program adds the High/Low limit values to every data value line in the *.tg file of a sensor. This enables Thermoguard Report to draw the limit lines which were set in the respective measurement period. Before version 2.90, the currently set limit values were displayed for any time period - regardless if different values had been set for a certain time in the past. Here is an example where the limits were changed a couple of times:

In this example version 2.90 was used from the marked date (2013-12-30, 3PM). For earlier data with lacking *.tg information about the limits, the current limits are used (here 0 And +9 °C). If several limits exist in the reporting period, the labeling in the "Details" section changes from "Low Limit" to "Lowest Limit" (main window also). If option " Y-Range: High/Low limit lines" is selected, the y scaling adapts to these "Lowest and Highest" limits as well (here 0 and +15 °C).

Note: Live!Chart window does not support this "Multi limit value display" feature.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 25 of 47

Chart Batch Print Chart Batch Print provides printing out several sensor charts "in a row" without the need to manually perform the steps [click on "Draw Chart"] => [Press CTRL+P] => [Confirm Printer dialog] repeatedly for each sensor. The function can be executed as soon as more than one sensor is selected in the sensor list. It can be started by one of the following ways:

Via Extras menu|Chart Batch Print or

Via context menu| Chart Batch Print or

Keyboard shortcut CTRL+P First, an option dialog pops up. The text printed in red depends on the current setting of the "Save Chart settings sensor specific" option:

OR Please notice the implications of the red statement for your batch print job. "Save as PNG" creates png files in the TG Report data directory instead of sending the charts to a printer. The functionality is the same as "Save as PNG" in the Chart menu of the chart window. "Confirmation after first printout (recommended) " provides a possibility to check the first copy of your printouts and only to continue if everything is satisfying. "Maximize chart window before printout (recommended) " maximizes the window before printout in order to achieve the maximum quality. The last option "Y-Range..." simulates pressing the enter key after opening the chart window manually. After clicking "OK" you may select your preferred printer in the printer dialog finally. As with clicking "Draw Chart" an informative message pops up, if one or more charts could not be created, see here.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 26 of 47

The Annotation Editor An "Annotation" is a graphical note which can be added to a Thermoguard chart and printed out or saved together with the chart. An annotation can be used to emphasize certain areas of a chart or to comment values violating the set limits exceptionally. By adding a transparent annotation of type "image" it is even possible to superimpose two different value curves.

Example of commenting a chart using the Annotation Editor. Here, three annotations of type "Note", "Image" and "Rectangle" were added.

Starting the Annotation Editor: The Annotation Editor is invoked via the Ther-moguard chart window's "Chart" menu. Alter-

natively you may use keyboard shortcut + directly.

As for printing and saving a chart the following applies:

The size of the annotated graphic depends on the currently set window dimensions. In order to achieve maximum quality of the re-sulting graphic (respectively its printout) you therefore should maximize the window before annotating the chart.

The Annotation Editor receives the current section of a chart as a static "snapshot" from the chart window. This means that after invoking the Annotation Editor the functionality of the chart window (e.g. shifting or zooming the curve) is no longer available. Thus, carefully set all desired characteristics of the chart to be annotated before entering the Annotation Editor Mode.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 27 of 47

Basic Annotation Editor behavior: You can add as many annotations as you want to a chart graphic. As long as you do not save the annotated chart, you are able to edit any of the annotations (e.g. move it, change its dimension or other properties or delete it).

On saving the annotated chart, all annotations are "burned" into the graphic and can no longer be edited after-wards!

However, you may add further annotations after saving the graphic in addition to the already "burned in" anno-tations.

Adding an annotation: There are 5 different types of annotations which may be selected via the "Annotate" menu or directly using the corresponding key-board shortcuts (see screenshot shown on the left):

Note

Text

Rectangle

Line

Image

An annotation (here of type Note) will always be positioned in the upper left corner of the chart and can be moved after being inserted. To move the annotation either use the mouse (click the annotation and hold the left mouse button while moving the mouse) or alternatively by pressing the cursor keys (Shift key+cursor key for faster movement).

By dragging the annotation at its markers its size can be changed.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 28 of 47

Editing an annotation's properties: Always, first click the annotation you want to edit (=> the "markers" show up). Then, choose one of the following three alter-native possibilities to enter the properties dia-log:

The "Annotate" menu shows two additional items "Properties" and "Delete". Select "Properties".

or

Right click the marked annotation and se-lect "Properties" from the context menu (this method is shown on the left).

or

Just press the "Enter" key.

Annotation of type "Image": After clicking "Load" you can select any gra-phics file and add it to your chart. A typical sce-nario is to add your company logo. In the directory

%ALLUSERSPROFILE%\TGReport Data\img the

installation provides 8 different red arrows which may be used for image annotations as well (shown on the left).

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 29 of 47

Notes on printing out the annotated graphic: In most cases the printer's page orientation setting should be set to "Landscape" before starting the printout (see screenshot shown on the left). Before printing out an annotated chart it is not mandatory to save it first. However pay atten-tion to the fact, that the markers of a selected annotation will be included in the printout. To prevent this click shortly on the background to revoke the selection-markers before printout.

Reset default properties for all annotation types ("Annotate" menu):

All properties (except the position) of an annotation type are saved to the configuration file TGReport.ini in sec-

tion [TfrmAnnoProp] and are re-applied after terminating and restarting the program.

Via menu item "Reset default properties for all annotation types" this [TfrmAnnoProp] section will be deleted from

the configuration file completely. Annotation type's factory settings will be set on adding an annotation after this function has been executed.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 30 of 47

Combining two charts: You may use the " Annotation Editor" to superimpose (combine, overlay) two (or even more) chart curves. The following step-by-step instructions will show how you would typically achieve this objective: Example: The temperature from January, 1st 2013 to February, 10th 2013 is to be combined with the temperature of the same period of 2014 (same sensor). 1. Main window: Setting the desired period of time for 2013:

2. Click "Draw Chart" and maximize the chart window

3. Select "X axis: Force period entered in main window" from the chart's Options menu

4. Select "Y-Range: set own range" from the chart's View menu and choose an appropriate y-range; here -5.5°C and +14.5°C:

5. Modify line width and line color if necessary

6. Chart menu: "Save as PNG (only inner area)"; e.g. as "2013.png"; the 2013 chart can be closed now

7. Now create the 2014 chart; modify period:

8. Click "Draw Chart" and maximize the chart window

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 31 of 47

9. Modify line width and line color if necessary (here: changed to blue)

10. Again: "X axis: Force period entered in main window" from the chart's Options menu. In our example, there are not yet any February 2014 data. The "X axis: Force period ..." option ensures that

the 2014 chart displays the same period as the 2013 chart.

11. Set the same y range (again -5.5°C and +14.5°C)

12. Hide the High/Low limit lines - the easiest way to do this is by pressing the "l"-key (lower "L")

13. Now, call the " Annotation Editor":

14. Set "Original size" in "View" menu

15. "Annotate menu" => "Add Image"

16. "Annotation Properties" => "Load" Button => Load the previously saved graphics "2013.png":

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 32 of 47

17. Shift the image so that the x and y axes fit perfectly. For the final, fine adjustment you should use the cursor keys!

18. Now you may add further annotations. In the example a legend was added as two annotations of the "Text" type providing an explanatory text for both curves:

19. Finally, save the combined chart (e.g. as "2013vs2014.png"). The changes are made permanent then. How-

ever you can reload "2013vs2014.png" in Annotation Editor's "File|Open" menu item at any time later to add

some more annotations (e.g. a third curve).

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 33 of 47

Thermoguard Report Live!

Principle The Thermoguard main program collects and stores all sensor data automatically, whereas TG Report is able to calculate and show them with the Live! function near real time.

A TG Report Live! window reads all 1.5 seconds the last measured value from the last *.tg file and shows it as free scalable digital display, as a continuously drawn chart or as a gauge type "visualization". Several TG Report Live! windows can be opened and arranged at the same time:

The windows size depends only on the dis-play size and resolution:

TG Live! windows set for 4 different sensors

TG Live! shown on a Plasma display

Some examples that show the usage of Thermoguard Report Live! windows are shown in our homepage's gallery

How to open a sensor's Live! Window A Live! window can only be opened for one selected single sensor. Here are the possibilities:

Double clicking directly on a sensor list's entry (in case several sensors are highlighted before click, the current multiselection is revoked by the double click).

Pressing the Enter key (only one single sensor may be highlighted)

Via Extras menu|Live!

Via context menu|Live!

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 34 of 47

Running Thermoguard Live!

Thermoguard Live! Value and Chart

Three register cards can be selected: Value, Chart and Visual (if configured, described later in a separate chap-ter). Value shows the recorded data as numeric value, while Chart opens a separate window with a continuously drawn chart. The displayed period can be user defined from 1..24 hours (see further below). Several Value, Chart and Visual windows for several sen-sors can be opened and arranged on screen at the same time.

Context menu for "Value" mode

The "Non-LCD Mode" includes the unit

Color set selection dialog

Live! Options for Value Mode A click with the right mouse button within the Live! Value window shows the Options menu:

- "VALUE|CHART|VISUAL": These items provide swit-ching between the Live! display modes - especially if the tabs are faded out (see next item)

- "Show tabs ": The tabs may be faded out to achieve a more compact, clear display. Eventually you must adjust the vertical height afterwards.

- "Show last *.tg file ": Opens the last saved file with the sensor data

- "LCD Mode": Shows either a window using LCD style or a proportional font style. The latter ("Non-LCD Mode") shows the physical unit and sign ("+") too, if appropriate.

- "Color set selection ": Allows the selection of at least seven different "Color Sets" for showing the LCD window.

- "Magnetic snap at screen edges": Switches the "Bor-der Snap" function On and Off.

- "Show info in title bar": Shows or hides the current Date and the physical unit of the data shown at the upper border of the window (in the example shown this option is unchecked so info is hidden).

- "Date format ...": Toggles between European and U.S. format of Date in the title border. Not available in the example shown left, because "Info in title bar" is set off

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 35 of 47

Context menu for "Value" Mode, continued

Sensor information window

A Tooltip reveals basic data of the sensor, here shown upon a Live! window with faded tabs

Live! Options for Value Mode (continued)

- "Sensor Info": Shows information for the selected senor and its settings in a separate window

- "Minimum Size": Minimizes the window to the small-est supported size

- "Hide/Show main window": Hides or restores the TG Report main window

- "Hide|Show mouse cursor": If running TG Report with the /h parameter or if the main window has been faded out (using the previously described menu item) the mouse cursor is hidden. In this case it can be restored via menu item "Show mouse cur-sor". If the mouse is hidden, it only vanishes in the areas of those windows belonging to TG Report; as you move the mouse cursor across the screen it becomes visible in all other screen areas.

- "Visual Designer for this sensor": A "shortcut" for starting the "Visual Designer" (compared to starting it via the main window), see separate chapter

- "Autosave window as PNG ": If this item is checked, every time the contents of the Live! window changes, a *.png graphic file of the whole Live! win-dow will be saved automatically to the Thermoguard Report configuration path.

This is by default %ALLUSERSPROFILE%\ TGRe-

port Data. The scheme for the filename is: TGLive_[Label]__[Location]__[Mode]_w.png

Here, [Mode] is = "std" for Non-LCD Mode, "lcd" for LCD Mode, "crt".for Chart Mode and "vis" for Visual Mode. Special characters, spaces and characters which are invalid for filenames are replaced by "_". So there will be no problems when uploading the files to a (UNIX) web server.

- "Autosave display as PNG": As before, but just the display without the surrounding window frame is saved. Filename scheme's suffix is "d" instead of "w": TGLive_[Name]__[Location]__[mode]_d.png

The saved files may be uploaded to a website on a regular basis using a FTP scheduler program. If im-plementing such a scenario, remember to set the dimensions within the HTML code of the website to the exact ones of the according Live! windows.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 36 of 47

Context menu for "Chart" mode

Live! Options for Chart Mode

A click with the right mouse button within the Chart win-dow shows the Options menu. Four additional items are enabled which are not available in Value mode:

- "Show High/Low limit lines": Show or hide high (red) and low (blue) limit value lines

- "Show limit values": Shows or hide the numerical values of the upper and lower limit.

- "Show Grid": Show or hide grid lines. - "Show Low/High: In the status bar at the bottom of

the window the Lowest and Highest value of the cur-rently displayed period are shown ("Peaks") using the format L:[Value] H:[Value]; an example:

"L:" means "Low", "H:" means "High"

Option for the Live! Chart Mode in the TG Report Main Window

- "Options" menu: Set Live! Chart display period": Sets the time interval of the Live! Chart Display be-tween 1 and 24 hours. Default setting is 12 hours. A new setting takes effect for all new Live! windows.

Period = 1 hour Period = 24 hours

Edit Live! Color set in Thermoguard Report's Main window

- Extras menu|"Edit Live! Color set": Any of the seven predefined color sets can individu-ally altered here. Background as well as the Font color can be set to a customized color.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 37 of 47

Alarm Visualization in Value Mode

Alarms Colors (here using color set #1) TG Report Live! shows current violations of the lower or upper limits with different colors. A simple violation of the lower or upper limit without mat-ching the alarm trials counter is shown as "pre alarm" in yellow color. Matching the alarm trials counter with the next violation of the lower or upper limit will be shown in blue or red color. The next value between the lower and upper limit (i.e. within the "valid range") will be shown in white color again. If the value for alarm trials is set to "0", any value exceed-ing the lower or upper limit will be shown immediately in blue or red color.

TG Report Live! accumulates no alarm states. The digital display shows always the last value and its status color. If the status color is normal (again), the display gives no information if, when and why an alarm might have occurred! How Thermoguard Live! alarm colors are assigned to Thermoguard alarm states

TG Live! does not use the sensor limit settings from the TGuard.scx file. Instead, the infor-

mation from the corresponding sensor's *.tg file are used solely.

Examples: *.tg file TGLive! Display: 07.06.2008 08:30:00 -21.3 white

07.06.2008 09:00:00 -21.4 white

07.06.2008 09:30:00 -5.8 1/1+ yellow

07.06.2008 10:00:00 -10.2 2/1+ A+ red

07.06.2008 10:30:00 -11.9 1/1+ yellow

07.06.2008 11:00:00 -12.6 white

07.06.2008 11:30:00 -13.2 white

07.06.2008 12:00:00 -31.0 1/1- yellow

07.06.2008 12:30:00 -31.2 2/1- A- blue

If no trial is configured for a sensor (sensor configuration in the main program), TG Live! will never show a window with yellow background for this sensor in Value mode.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 38 of 47

Error Messages, their visualization and Causes

- Error "- - - -":

No data could be read Possible reasons: * There are no data for this sensor polled and stored until now (no *.tg file exists for this sensor yet) * File access denied (*.tg file is locked) * An empty line at the end of the corresponding *.tg file. In the window headline "----" will blink in red color, when this error message occurs. Blinking will even occur, when the option "Show info in title bar" is disabled. In Chart mode, the background shows a yellow-red blinking gradient.

- Error " Old":

The value is outdated Reason: The difference between the values time and date and current system time and date exceeds the polling time interval. This error message can only occur, when the polling interval is set to "Every n minutes"! Date and time in the window headline will blink in red color, when this error message occurs. Blinking will even occur, when the option "Show info in title bar" is disabled. In Chart mode, the background shows a white-yellow blinking gradient.

- Error "dErr": Data are not complete or faulty (Data Error) Possible reasons: Data are manipulated intentionally or by mistake; or there might be read/write errors. E.g.: There must be at least one tab character within any line of a *.tg file. Missing this tab character will cause a Data Error. In the window headline "derr" will blink in red color, when this error message occurs. Blinking will even occur, when the option "Show info in title bar" is dis-abled. In Chart mode, the background shows a white-red blinking gradient.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 39 of 47

- Error "dcEr": Decryption error during decoding the encrypted data Possible reason: Data are manipulated intentionally. In the window headline "dcerr" will blink in red color, when this error message occurs. Blinking will even occur, when the option "Show info in title bar" is dis-abled. In Chart mode, the background shows a white-red blinking gradient.

- Error "FAIL":

No value could be polled for this sensor ("Alarm on Fail") Possible reasons: There might be a network or power failure, license error etc. For more reasons, please see the Thermo-guard System Manual. You may press F5 to load the corresponding *.tg file into your text editor; then take a look at the last entry to get more detailed information, e.g. "FAIL (NO DATA received)". No special visualization in Chart Mode, but note these paragraphs: Display of missing data and Line color for "FAIL".

- Error "SimErr": Error in simulation mode Possible reasons: The simulation file could not be found or contains an invalid line (see Thermoguard System Manual). No special visualization in Chart Mode, but note these paragraphs: Display of missing data.

- Error "NoSens":

Sensor missing. Possible reasons: Sensor missing, cable disrupted etc. No special visualization in Chart Mode, but note these paragraphs: Display of missing data.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 40 of 47

Live! Window Sets Autosave of the last settings per Sensor to TGReport.ini All Live! window settings of an individual sensor are stored in a separate section of the file

TGReport.ini when closing. These settings are re-used, when this sensor windows is opened

next time. Sample sections from TGReport.ini:

[Temperatur-LIVE!]

Settings=268,329,203,89,1,1,1,1,1,1,1,1,1,1,0,1,1,351,215,283,145

[Humidity-LIVE!]

Settings=150,200,283,145,1,0,1,1,1,1,1,1,1,0,0,1,1,283,145,283,145

Autosave of the last window configuration when closing TG Report to TGReport.tgl

When closing TG Report, the last active Live! window setting is always saved in the file TGRe-

port.tgl. If the file exists it will be overwritten.

The TGReport.tgl file is always read at program start; therefore the last active configuration

will be restored.

If there is no Live! window when closing the program, no TGReport.tgl file at all will be cre-

ated (no one with a size of 0 byte either).

The settings of all Live! windows are stored in a *.tgl file - i.e. even settings for multiple win-

dows for the very same sensor. Sample lines from a *.tgl file:

Temperature,202,285,283,145,1,1,1,1,1,1,1,1,0,0,0,1,1,273,180,200,780

Temperature,202,0,283,145,1,0,1,1,1,1,1,1,1,0,0,1,1,314,190,200,780

Humidity,386,284,283,145,1,1,1,1,1,0,1,1,1,0,0,1,1,311,208,283,145

Humidity,388,0,283,100,1,0,0,1,1,0,1,1,1,0,0,1,1,311,173,283,145

In this example for each of the two sensors a value as well as a chart window were opened.

It should be noted here that the extremely helpful recommended utility allSnap makes the arrangement of Live! windows a child's play!

=>

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 41 of 47

Saving and loading of complete window sets with a dedicated name Using the menu options "Save/Load Thermoguard Live! window set" of the TG Report main window, all current Live! windows and their settings can be saved and loaded using a dedi-cated name of your choice:

If a Live! window set is to be opened and one or more Live! windows are already active, a re-quest for closing of these windows will show up.

If you do not close the existing windows while loading the very same set again, the windows are stacked exactly on top of each other!

The number of actual opened windows is shown in the button for closing all charts in the main window:

The shown number refers to all opened TG Report windows; including all Live! and all static Chart windows. Using the command line switch /tgl:<Filename>, a previously saved *.tgl file is opened

automatically (possibly via a desktop shortcut). Example:

"%ProgramFiles(x86)%\Thermoguard\TGReport\TGReport.exe" /tgl:ClimChambers

If no file extension is entered, .tgl is set automatically.

If not path is entered (like in the example above), the default path %ALLUSERSPRO-

FILE%\TGReport Data is set automatically. If the *.tgl file should reside in a different path,

the path can be included in the parameter after /tgl:.

When one or more sensors defined in a *.tgl file no longer exist the following type of warning

message is displayed:

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 42 of 47

Visual Mode / Visual Designer Thermoguard Report Live! Visual Mode provides the possibility to visualize the measured value as a bar graph combined with any background graphics picture of your choice. The gauge height reflects the current value; Examples:

16 Live! windows visualize gas pressure level in 8 gas cylinders

As background image you can select an artificial graphics scheme or - as with the bathtub thermometer below - a "real" photograph. Up to three sections can be defined each showing a different (also customizable) color:

3 different colors are supported

Switch back to one of the other Live! display modes via the context menu

Operational Principle: All parameters of a visualization are saved in a *.vis text file, e.g. which background image to use etc. When opening a Live! window, TG Report looks if such a file "sensorname.vis" exits and if so, the "Visual" tab is shown:

After clicking on the "Visual" tab, the display changes into your visualization and all tabs

are faded out as they would disturb the "look". In "Visual" Mode you should disable the "Show info in title bar" option for the same reason. If you want to switch back again to "Value" or "Chart" mode, right click the graphic to open the context menu; then click either "VALUE" or "CHART".

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 43 of 47

*.vis files are located in the \vis subfolder of Thermoguard main program's configuration

directory, so this is by default %ProgramData%\Thermoguard Data\vis.

The background images are in the subfolder %ProgramData%\Thermoguard Data\vis\img.

Due to this central storage principle, any Thermoguard Report Client on workstations in the network can access the data without maintaining local copies themselves. Beginning with setup TGuard_270.exe some sample template *.vis files and their associated background images are installed as part of the "main program" component (not as part of TG Report). Therefore, if you want to take advantage of the new Visual Mode, you should not only update Thermoguard Report clients, but also the main program on your Thermoguard server if you are updating from a version earlier than version 2.70. Be aware that your own custom background images must be copied to the \vis\img folder men-tioned above as well. Visual Designer Now, how to create your own *.vis files? They are easily designed using the "Visual Designer" utility. The Visual Designer is started via the TG Report main window or directly via a Live! window. Via main window: Assure that only one single sensor is selected. Now, you can start the program using one of these possibilities:

Via Extras menu|"Visual Designer for selected sensor" or

Via context menu|"Visual Designer for selected sensor" or

Using keyboard shortcut CTRL+V

Alternatively via Live! window:

Via context menu|"Visual Designer for this sensor" or

The "d" Key

The Visual Designer is an own executable program (TGReportVisDesigner.exe); if it is

already running, a warning message is shown. If a vis file already exists for a sensor, the corresponding window showing the visualization shows up. If not, the following query pops up:

If you confirm, Visual Designer is started:

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 44 of 47

For a first time vis file a standard thermometer background image of size 99x400 pixel and a scale from -30 to +50 °C is loaded. In the session shown above, this default background image has been replaced with the (also preinstalled) larger 149x600 pixel version. Then the width of the bar has been increased and the bar set to "transparent" (therefore the grey background color setting is of no meaning). Designer functions are intuitive and self-explanatory. By using the "Test value" slider you can check if the bar "hits" the desired marks of the scale. Fine tu-ning can be done by adjusting the "Min:" and "Max:" values. Important: A vis file is not "connected" to the sensor settings which are maintained using the Thermo-guard main program. Especially the lower and upper border settings for dividing the color "fields" of the visualization are not synchronized in any way with the sensor's alarm limit values. If you change these settings in the main program you must readjust them with Visual Designer. Tip for (first time) positioning of the bar (named "gauge" in the Designer): Using the four blue arrows of the "cross control" you change either the size or the position of the bar - depending on the setting Mode(Position|Size). Instead of clicking either Position or Size you can click the symbol in the middle of the cross. Depending on the active mode it changes its shape appro-priately. For a coarse positioning/sizing of the bar, you may enter values directly in their input fields. If you click an arrow "key" and hold the mouse down, the changes are applied continuously.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 45 of 47

A1: For Administrators: Configure fixed Path Presets There is the possibility for an administrator to preset fixed TG Report paths by creating the fol-lowing registry entries manually:

[HKEY_LOCAL_MACHINE\SOFTWARE\Thermoguard\TGReport\FixedPaths]

CfgDataDirTGReport=[Path]

CfgDataDirTGMain =[Path]

TgPath =[Path]

In the example shown below, the TG Report configuration path is the default setting (here for Windows Vista) whereas the configuration path of the main program as well as the *.tg path

had been set to a network drive T:, which is mapped to a read-only network share:

A standard user can not alter this paths, because:

He has no write access to HKEY_LOCAL_MACHINE

The menu item "Options" => "Configure paths" is not present any longer

The command line switch /CfgDataDir:<Directory> has no effect.

Setting fixed paths prevents data manipulation or the possibility that a user alters the important base path settings for TG Report intentionally or unintentionally. Note that the option to display all current path setting using "Help" menu => "About / Path In-formation" is not suppressed.

A2: Freeware Program "allSnap" The option "Magnetic snap at screen edges" for a TG Report Live! window drags the window to the screens edge. However, if several Live! windows are open it often is desirable that the win-dows will snap to each other as they are aligned as well.

Such a feature is provided by the free software "allSnap", which can be downloaded here: http://www.allsnap.org/

It is also located on the Thermoguard CD in directory \Weitere Software\allSnap.

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 46 of 47

A3: Arithmetic and Mean kinetic Temperature The arithmetic mean temperature in an observation period is a simple measure for thermal pressure on stored products. The durability of a product is directly dependent on the thermal pressure. Like most chemical processes, the speed of aging processes in organic matter such as food-stuff or pharmaceuticals is not linear in line with the temperature, though, but follows an expo-nential function. Higher temperatures in a particular period have an exponentially over-propor-tional influence on lower temperature during the same period. Even short-term rise in temperature, which may be due to technical disruptions, must therefore not be neglected, and cannot be offset against the corresponding lower temperature within the same period of time. When calculating the "mean temperature", the higher temperatures must be weighed expo-nentially. Given this insight, as early as 1971, J. D. Haynes deduced a "mean kinetic tempera-ture" from the gas equation by S. Arrhenius under the title "Worldwide Virtual Temperatures for Product Stability Testing" (Sci. Vol. 60, 927- 929, June 1971). Today MKT is defined by the United States Pharmacopoeia USP in the chapter on "pharma-ceutical dosage forms" as "single calculated temperature at which the total amount of degra-dation over a particular period is equal to the sum of the individual degradations that would occur at various temperatures". So MKT is a degraded temperature that would expose a chemical compound to the same thermal pressure as a series of higher and lower temperatures. In technical notation, the for-mula for calculating the mean kinetic temperature for a particular chemical compound or phar-maceutical substance of a product is:

MKT = (-H/R)/ln{(SUM(exp(-H/RTn))))/n}

or, in algebraic notation:

n

e

R

H

MKTn

i

TR

H

n

1ln

The abbreviations have the following meaning:

MKT Mean Kinetic Temperature, calculated as the absolute temperature in Kelvin

H Activation energy of the chemical compound or the pharmaceutical substance, in kJ/mol

R Universal gas constant (0,0083144 kJ/molK)

TG Report

Thermoguard Thermoguard Report * August 8, 2015 Page 47 of 47

Tn Absolute temperature during period n, measured in Kelvin SUM Sum for n periods n Total number of all measured absolute temperature values over the observation

period of i=1 to n ln Natural logarithm exp Exponential function for basis e (2,71828)

The activation energy of a chemical compound can be determined in exact terms by Differen-tial Scanning Calorimetry (DSC). For many pharmaceutical substances, this is in the 42-125 kJ/mol range. For simplicity's sake, though, the activation energy is frequently not determined but defined as equal to a mean value of 83.144 kJ/mol. In regular storage conditions without fluctuations or extreme temperatures, the arithmetic mean temperature and the mean kinetic temperature match. Thus using a mean value of 83.144 kJ/mol for activation energy is enough for comparing both values for estimating the influence of peak temperatures.

A4: Standard deviation Thermoguard Report estimates the standard deviation based on a sample (ignoring "values" which do not represent regular values - like "error" etc.). It equals the Microsoft Excel function STDEV.S (Excel 2010, former versions = STDEVS). A result is not displayed if there is only one single value in the selected time period. The following formula is used:

where x is the sample arithmetic mean temperature (displayed in the first output field of the "Results"-frame) of the values and n is the "sample size" (i.e. the number of regular values which were measured in the selected time period).