main gas chromatography

Upload: aashish-singh

Post on 08-Apr-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/7/2019 MAIN Gas Chromatography

    1/17

    Gas Chromatography (GC)Introduction

    Gas chromatography is a chromatographic technique that can be used to separate volatile

    organic compounds. A gas chromatograph consists of a flowing mobile phase, an

    injection port, a separation column containing the stationary phase, and a detector. The

    organic compounds are separated due to differences in their partitioning behavior

    between the mobile gas phase and the stationary phase in the column.

    Instrumentation

    Mobile phases are generally inert gases such as helium, argon, or nitrogen. The inject ionPort consists of a rubber septum through which a syringe needle is inserted to inject the

    sample. The injection port is maintained at a higher temperature than the boiling point of

    the least volatile component in the sample mixture. Since the partitioning behavior is

    dependant on temperature , the separation column is usually contained in a thermostat-

    controlled oven. Separating components with a wide range of boiling points is

    accomplished by starting at a low oven temperature and increasing the temperature overtime to elute the high-boiling point components. Most columns contain a liquid stationary

    phase on a solid support. Separation of low-molecular weight gases is accomplished with

    solid adsorbents. Separate documents describe some specificGC Columns andGCDetectors.

    Schematic of a gas chromatograph

    Gas Chromatography ColumnsColumnsGas chromatography columns are of two designs: packed or capillary. Packed columns

    are typically a glass or stainless steel coil (typically 1-5 m total length and 5 mm inner

    diameter) that is filled with the stationary phase. Capillary columns are a thin fused-silica

    (purified silicate glass) capillary (typically 10-100 m in length and 250 m inner

    diameter) that has the stationary phase coated on the inner surface. Capillary columns

  • 8/7/2019 MAIN Gas Chromatography

    2/17

    provide much higher separation efficiency than packed columns but are more easily

    overloaded by too much sample.

    Gas Chromatography (GC) DetectorsIntroduction

    After the components of a mixture are separated using gas chromatography, they must be

    detected as they exit the GC column. The links listed below provide the details of some

    specific GC detectors. The thermal-conductivity (TCD) and flame-ionization (FID)

    detectors are the two most common detectors on commercial gas chromatographs. The

    requirements of a GC detector depends on the separation application. For example, one

    analysis might require a detector that is selective for chlorine-containing molecules,

    another analysis might require a detector that is non-destructive so that the analyte can be

    recovered for further spectroscopic analysis.

    Specific GC detectors

    Atomic-emmision detector (AED)

    Chemiluminescence detector

    Electron-capture detector (ECD)

    The ECD is as sensitive as the FID but has a limited dynamic range and finds its

    greatest application in analysis organic molecules that contain electronegative

    functional groups, such as halogens, phosphorous, and nitro groups.Flame-ionization detector (FID)

    The FID is extremely sensitive with a large dynamic range, its only disadvantage

    is that it destroys the sample.

    Flame-photometric detector (FPD)

    Photoionization detector (PID)

    Thermal conductivity detector (TCD)The TCD is not as sensitive as other dectectors but it is non-specific and non-

    destructive.

    Atomic-Emission Detector (AED)Introduction

    As capillary column based gas chromatography takes its place as the major, highest

    resolution separation technique available for volatile, thermally stable compounds, the

    requirements for the sensitive and selective detection of these compounds increases.

    Since more and more complex mixtures can be successfully separated, subsequent

    chromatograms (output of a chromatographic separation) are increasingly more complex.

    Therefore, the need to differentiate between the sample components using the GCdetector as a means of compounds discriminating is more and more common. In addition,

  • 8/7/2019 MAIN Gas Chromatography

    3/17

    each detector has its own characteristics (selectivity, sensitivity, linear range, stability,

    cost, etc.) that helps in a decision about which detector to use.

    One of the newest additions to the gas chromatographer's arsenal is the atomic emission

    detector (AED). This detector, while quite expensive compared to other commercially

    available GC detectors, is an extremely powerful alternative. FOR INSTANCE, Instead

    of measuring simple gas phase (carbon containing) ions created in a flame as with theflame ionization detector, or the change in background current because of electronegative

    element capture of thermal electrons as with the electron capture detector, the AED has a

    much wider applicability because it is based on the detection of atomic emissions.

    The strength of the AED lies in the detector's ability to simultaneously determine the

    atomic emissions of many of the elements in analytes that elute from a GC capillary

    column (called eluants or solutes in some books). As eluants come off the capillary

    column they are fed into a microwave powered plasma (or discharge) cavity where the

    compounds are destroyed and their atoms are excited by the energy of the plasma. The

    light that is emitted by the excited particles is separated into individual lines via aphotodiode array. The associated computer then sorts out the individual emission lines

    and can produce chromatograms made up of peaks from eluants that contain only aspecific element.

    Instrumentation

    The components of the AED include 1) an interface for the incoming capillary GC

    column to the microwave induced plasma chamber, 2) the microwave chamber itself, 3) a

    cooling system for that chamber, 4) a diffraction grating and associated optics to focus

    then disperse the spectral atomic lines, and 5) a position adjustable photodiode array

    interfaced to a computer. The microwave cavity cooling is required because much of the

    energy focused into the cavity is converted to heat.

  • 8/7/2019 MAIN Gas Chromatography

    4/17

    Chemiluminescence SpectroscopyIntroduction

    Chemiluminescence, like atomic emission spectroscopy (AES), uses quantitative

    measurements of the opticalemis s ion from excited chemical species to determine analyte

    concentration; however, unlike AES, chemiluminescence is usually emission from

    energized molecules instead of simply excited atoms. The bands of light determined by

    this technique emanate from molecular emissions and are therefore broader and more

    complex then bands originating from atomic spectra. Furthermore, chemiluminescence

    can take place in either the solution or gas phase, whereas AES is almost strictly as gasphase phenomenon.Though liquid phase chemiluminescence plays a significant role in laboratories using thisanalytical technique (often in conjunction with liquid chromatography

    ), we will

    concentrate on gas phase chemiluminescence reactions since the instrumental

    components are somewhat simpler. These detectors are also often used as detectors forgas chromatography.Like fluorescence spectroscopy, chemiluminescence's strength lies in the detection of

  • 8/7/2019 MAIN Gas Chromatography

    5/17

    electromagnetic radiation produced in a system with very low background. And on top of

    this, because the energy necessary to excite the analytes to higherelectronic, vibrational,and rotational states (from which they can decay be emission) does not come from an

    external light source like a laser or lamp, the problem of excitation source scattering is

    completely avoided. The major limitation to the detection limits achievable by

    chemiluminescence involves the dark current of the photomultip lier (PMT) necessary to

    detect the analyte light emissions.

    If the excitation energy for analytes in chemiluminescence doesn't come from a source lamp or laser, where does it come from?

    The energy is produced by a chemical reaction of the analyte and a reagent. An example of a reaction of this sort is shown

    below:A chemiluminescence reaction

    In gas phase chemiluminescence, the light emission (represented as Planck's constant

    times nu-the light's frequency) is produced by the reaction of an analyte (dimethyl sulfide

    in the above example) and a strongly oxidizing reagent gas such as fluorine (in theexample above) or ozone, for instance. The reaction occurs on a time scale such that the

    production of light is essentially instantaneous; therefore, most analytical systems simply

    mix analytes and the reagent in a small volume chamber directly in front of a PMT. If the

    analytes are eluting from a gas chromatographic column then the end of the column is

    often fed directly into the reaction chamber itself. Since as much of the energy released

    by the reaction should (in the analyst's eye) be used to excite as many of the analyte

    molecules as possible, loss of energy via gas phase collisions is undesirable, and

    therefore a final consideration is that the gas pressure in the reaction chamber be

    maintained at a low pressure (~ 1 torr) by a vacuum pump in order to minimize the

    effects of collisional deactivation.

    It must be stated that the ambiguous specification of "products" in the above reaction is

    often necessary because of the nature and complexity of the reaction. In some reactions,

    the chemiluminescent emitters are relatively well known. In the above reaction the major

    emitter is electronically and vibrationally excited HF; however, in the same reaction,

    other emitters have been determined whose identities are not known and these also

    contribute to the total light detected by the PMT.

  • 8/7/2019 MAIN Gas Chromatography

    6/17

    To the analytical chemist the ambiguity about the actual products in the reaction is, in

    most case, not important. All the analyst cares about is the sensitivity of the instrument

    (read detection limits for target analytes), its selectivity-that is, response for an analyte as

    compared to an interfering compound, and the linear range of response.Here is a schematic of the components necessary for a gas phase chemiluminescence

    detector interfaced to a capillary gas chromatograph.

    ThIS fig is belong to ECD..

    Electron Capture Detectors(ECD)Introduction

    The ECD uses a radioactive Beta emitter (electrons) to ionize some of the carrier gas and

    produce a current between a biased pair of electrodes. When organic molecules that

  • 8/7/2019 MAIN Gas Chromatography

    7/17

    contain electronegative functional groups, such as halogens, phosphorous, and nitro

    groups pass by the detector, they capture some of the electrons and reduce the current

    measured between the electrodes. The ECD is as sensitive as the FID but has a limited

    dynamic range and finds its greatest application in analysis of halogenated compounds.Schematic of an ECD

    Flame-Ionization Detectors (FID)

    Introduction

    An FID consists of a hydrogen/air flame and a collector plate. The effluent from the GC column passes through the flame, which

    breaks down organic molecules and produces ions. The ions are collected on a biased electrode and produce an electrical

    signal. The FID is extremely sensitive with a large dynamic range, its only disadvantage is that it destroys the sample.

    Flame Photometric GC DetectorIntroduction

    The reason to use more than one kind of detector for gas chromatography is to achieve

    selective and/or highly sensitive detection of specific compounds encountered in

    particular chromatographic analyses. The determination of sulfur or phosphorus

    containing compounds is the job of the flame photometric detector (FPD). This device

    uses the chemiluminescent reactions of these compounds in a hydrogen/air flame as a

    source of analytical information that is relatively specific for substances containing these

  • 8/7/2019 MAIN Gas Chromatography

    8/17

    two kinds of atoms. The emitting species for sulfur compounds is excited S2. The lambda

    max for emission of excited S2 is approximately 394 nm. The emitter for phosphorus

    compounds in the flame is excited HPO (lambda max = doublet 510-526 nm). In order to

    selectively detect one or the other family of compounds as it elutes from the GC column

    an interference filteris used between the flame and thephotomultiplier tube (PMT) to isolate the appropriate emission band. Thedrawback here being that the filter must be exchanged between chromatographic runs if the other family of compounds is to be

    detected.

    Instrumentation

    In addition to the instrumental requirements for 1) a combustion chamber to house the

    flame, 2) gas lines for hydrogen (fuel) and air (oxidant), and 3) an exhaust chimney to

    remove combustion products, the final component necessary for this instrument is a

    thermal (bandpass) filter to isolate only the visible and UV radiation emitted by the

    flame. Without this the large amounts of infrared radiation emitted by the flame's

    combustion reaction would heat up the PMT and increase its background signal. The

    PMT is also physically insulated from the combustion chamber by using poorly(thermally) conducting metals to attach the PMT housing, filters, etc.

    The physical arrangement of these components is as follows: flame (combustion)

    chamber with exhaust, permenant thermal filter (two IR filters in some commercial

    designs), a removable phosphorus or sulfur selective filter, and finally the PMT.Schematic of a gas chromatographic flame photometric detector

  • 8/7/2019 MAIN Gas Chromatography

    9/17

    Photoionization DetectorIntroduction

    The reason to use more than one kind of detector for gas chromatography is to achieve

    selective and/or highly sensitive detection of specific compounds encountered in

    particular chromatographic analyses. The selective determination of aromatic

    hydrocarbons or organo-heteroatom species is the job of the photoionization detector

    (PID). This device uses ultraviolet light as a means ofionizing an analyte exiting from a

    GC column. The ions produced by this process are collected by electrodes. The current

    generated is therefore a measure of the analyte concentration.

    Theory

    If the energy of an incoming photon is high enough (and the molecule is quantum

    mechanically "allowed" to absorb the photon) photo-excitation can occur to such an

    extent that an electron is completely removed from its molecular orbital, i.e. ionization.

    A Photoionization Reaction

    If the amount of ionization is reproducible for a given compound, pressure, and light

    source then the current collected at the PID's reaction cell electrodes is reproducibly

    proportional to the amount of that compound entering the cell. The reason why the

    compounds that are routinely analyzed are either aromatic hydrocarbons or heteroatomcontaining compounds (like organosulfur or organophosphorus species) is because these

    species have ionization potentials (IP) that are within reach of commercially available

    UV lamps. The available lamp energies range from 8.3 to 11.7 ev, that is, lambda max

    ranging from 150 nm to 106 nm. Although most PIDs have only one lamp, lamps in the

    PID are exchanged depending on the compound selectivity required in the analysis.

    Selective detection using a PID

  • 8/7/2019 MAIN Gas Chromatography

    10/17

    Here is an example of selective PID detection: Benzene's boiling point is 80.1 degrees C

    and its IP is 9.24 ev. (Check the CRC Handbook 56th ed. page E-74 for IPs of common

    molecules.) This compound would respond in a PID with a UV lamp of 9.5 ev

    (commercially available) because this energy is higher than benzene's IP (9.24).

    Isopropyl alcohol has a similar boiling point (82.5 degrees C) and these two compounds

    might elute relatively close together in normal temperature programmed gas

    chromatography, especially if a fast temperature ramp were used. However, since

    isopropyl alcohol's IP is 10.15 ev this compound would be invisible or show very poor

    response in that PID, and therefore the detector would respond to one compound but not

    the other.

    Instrumentation

    Since only a small (but basically unknown) fraction of the analyte molecules are actually

    ionized in the PID chamber, this is considered to be a nondestructive GC detector.

    Therefore, the exhaust port of the PIDcan be connected to another detector in series withthe PID. In this way data from two different detectors can be taken simultaneously, and

    selective detection of PID responsive compounds augmented by response from, say, anFIDorECD. The major challenge here is to make the design of the ionization chamber

    and the downstream connections to the second detector as low volume as possible (read small diameter) so that peaks that

    have been separated by the GC column do not broaden out before detection.Schematic of a gas chromatographic photoionization detector

  • 8/7/2019 MAIN Gas Chromatography

    11/17

    Thermal Conductivity Detectors (TCD)IntroductionA TCD detector consists of an electrically-heated wire or thermistor. The temperature of

    the sensing element depends on the thermal conductivity of the gas flowing around it.

    Changes in thermal conductivity, such as when organic molecules displace some of the

    carrier gas, cause a temperature rise in the element which is sensed as a change in

    resistance. The TCD is not as sensitive as other dectectors but it is non-specific and non-

    destructive.

    Instrumentation

    Two pairs of TCDs are used in gas chromatographs.O

    ne pair is placed in the columneffluent to detect the separated components as they leave the column, and another pair isplaced before the injector or in a separate reference column. The resistances of the twosets of pairs are then arranged in a bridge circuit.

  • 8/7/2019 MAIN Gas Chromatography

    12/17

    The bridge circuit allows amplification of resistance changes due to analytes passing over the sample thermoconductors and

    does not amplify changes in resistance that both sets of detectors produce due to flow rate fluctuations, etc.

    Gas Chromatographic InjectorsIntroduction

    The great analytical strength of capillary gas chromatography lies in its highres olution.

    Capillary columns have 1) more theoretical plates (a measure of column resolving power

    or efficiency) per meter as compared to packed columns and 2) since they have less

    resistance to flow they can be longer than packed columns. This means that the average

    capillary column (30 meters long) has approximately 100,000 theoretical plates while the

    average packed column (3 meters) has only 2500 plates.

    But with this separation power comes some limitations: 1) Capillary columns, because

    they have smaller diameters (0.05 to 0.53 mm) than packed columns (2 to 4 mm), require

    relatively specialized injectors and ancillary flow and pressure controllers and 2)

    capillary column require a smaller amount of sample than packed columns. While the

    average sample mass of each component in a mixture that is separable by packed column

    GC can be in the microgram range (10-6 grams) per injection, capillary columns routinelyonly handle 50 nanograms (10-9 grams) of a particular component or less.

    Overloaded ChromatographyThis sample size requirement initially meant that if samples contained components that

    were too concentrated for a capillary chromatographic analysis, the sample had to be

    diluted before it was analyzed. Otherwise the column would beoverload ed by those highconcentrated components. An example of this appears in the first figure below. Theclearly overload peaks are indicated. And while some of the other components are in theresolvable (not overloaded) range, having large masses of components can also distort thepeak shape of some of the lower mass components.

    An Overloaded Chromatogram

  • 8/7/2019 MAIN Gas Chromatography

    13/17

    The following figure shows a little better chromatography with fewer overloaded peaks.The second eluting peak (about 6 minutes) is clearly not overloaded while the groupbetween 10 and 14 minutes still shows overloading characteristics: long drawn-out tailingand much less than baseline separation with peaks that elute nearby (the 11 and 12minute peaks, for instance).

  • 8/7/2019 MAIN Gas Chromatography

    14/17

    Normal Packed Column Injector

    The normal sequence of events in a GC injection is as follows. We will assume in this

    explanation that some analytes are dissolved in a (liquid) solvent although much of thisprocess also holds for gas GC injections too: A small amount of liquid (microliters) is

    injected through a silicon rubber septum (using a special microliter syringe) into the hot

    (usually 200+ degrees C) GC injector that is lined with an inert glass tube. The injector is

    kept hot by a relatively large, metal heater block that is thermostatically controlled. The

    sample is immediately vaporized and a pressurized, inert, carrier gas-which is continually

    flowing from a gas regulator through the injector and into the GC column-sweeps the

    gaseous sample, solvent, analyte and all, onto the column. In the packed column injector,

    ALL the vaporized sample enters onto the column. This is how all packed column

    injectors work; everything that is injected goes onto the column. One modification of this

    is a small ancillary flow of carrier gas that bathes the underside of the injector's septumso that hot vaporized sample gases can't interact and possibly stick to the septum. This

    improves peak shape and reproducibility. This last feature is called the septum purge. The

    following figure is a schematic of a packed column injector, sometimes called a direct or

    flash injector. The septum purge is not shown here although the carrier gas regulator and

    inlet, a septum, and an injection port liner ARE detailed. Last by not least, the packed GC

    column itself is connected at the bottom of the injector via metal fittings.

  • 8/7/2019 MAIN Gas Chromatography

    15/17

    Schematic of packed GC column injector with septum purge

    One last subtle point about the configuration of the carrier gas inlet: notice that it enters

    the injector at about the middle of the heating block and its gas has to travel along the

    outside of the injector before it enters the injection port liner AT THE TOP. This is so the

    carrier is preheated before it enter the liner where the sample is vaporized. This helps toprevent a cold spot at the top of the injector where the carrier gas enters.

    Now remember that the size of the capillary column limits the amount of analyte that can

    be injected, otherwise, chromatographic overloading occurs. Therefore this packed

    column injector design, if used with a capillary column, would require that samples with

    high concentrations of analytes be diluted. Unless... what other alternative is there to get

    the amount of analytes that are injected onto the column smaller without having to dilute

    concentrated samples? The solution is the split/splitless capillary GC injector.

    The Split/Splitless Capillary Injector

    OK so far so good. But how does the split/splitless injector work? It starts with the same

    requirements as the packed column injectors: carrier gas inlet, a septum, septum purge,

    injector insert, heater block, and column connection; but the heart of this technological

    feat is another set of gas linesout of the injector-another path that the vaporized sample

    can take. This is called the split line or vent. The manufacturers of these systems design

    them so that the carrier gas flow onto the column is constant-to maintain the

    chromatographic requirements of the column and yield reproducible retention times for

    analytes. At the same time, the amount of gas that goes out the split vent controls the

    amount of sample that enters the column. If the split vent is closed, via a computer

  • 8/7/2019 MAIN Gas Chromatography

    16/17

    controlled split valve, then all of the sample introduced into the injector goes on thecolumn. If the split vent is open then most of the vaporized sample is thrown away towaste via the split vent and only a small portion of the sample is introduced to thecolumn. The following diagram illustrates a split/splitless injector with the split ventonso that only a small portion of the sample injected goes on the column

    And finally, a very neat aspect of this is that the amount of gas exiting the split vent can

    be varied while keeping the flow onto the column constant. This means that the

    AMOUNT of the split (called the split ratio) can be varied. A common split ratio is 50 to

    1. That is, for every 50 units of gaseous sample that are thrown away to waste, 1 unit

    goes on the column.

    The analyst keeps careful control of the split ratio so that results from thechromatography can still be quantified. Chromatographic peaks that show up as, say 2.5

    ng of compound X really represent 2.5 x 50 = 125 ng of analyte X in the original sample.

    Also notice that this mass (125 nanograms) would have overloaded the column if all of it

    ended up on the capillary column. Voila! A split injection, and no sample dilution

    required.

  • 8/7/2019 MAIN Gas Chromatography

    17/17