magx1 manual -1- v7 03-06-2009,...

53

Upload: ngohanh

Post on 20-Feb-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

MagX1 Manual -1- V7 03-06-2009, Modbus

List of Contents

01. Introduction 02

1.1. Operating Principle 02

1.2. Applications 02

1.3. Safety Instructions 02

1.4. Unpacking the flowmeter 02

02. Installation 03

2.1. Remote or Compact? 03

2.2. Sensor Installation 04

2.3. Dry Liner 05

2.4. Installation of the Transmitter 05

2.5. Module Installation 07

2.6. Cables connections 09

2.7. Potting the remote sensor terminal box 09

03. MagX1 Transmitter Unit 11

3.1. Main Screen 11

3.2. Flowmeter Menu 12

3.3. Info Menu 12

3.4. Flow Statistic Menu 13

3.5. Display Menu 17

3.6. User Settings Menu 19

3.7. Service Settings Menu 31

3.8. Factory Settings Menu 32

3.9. Authorize Menu 33

04. Power Supply Module 34

05. RS232 Module 35

06. Real-Time Clock Module 36

07. USB Module 37

08. Frequency Output Module 38

09. Memory Module 39

10. RS485 Module 40

11. Current Loop Output Module 41

12. Binary Output Module 42

13. Voltage Output Module 43

14. TCP/IP Module 44

15. BLUETOOTH Module 45

16. GPRS Module 46

17. Module Positioning 47

18. Liner and Electrode Selection 47

19. Flowmeter Dimensions 48

20. How to order your MagX1 49

21. MagX1 Error Code Table 49

22. Compatibility Table 50

23. CE and Conformity 50

24. Warranty 51

MagX1 Manual -2- V7 03-06-2009, Modbus

B

v

U

1. Introduction

1.1. Operating Principle

The measurement is based on the principle of Faraday‘s law on

electromagnetic induction in which an electric voltage is induced

in an electrically conductive body that moves in a magnetic field.

Liquid flows through a tube in the direction of the magnetic field.

Liquid with a certain minimum electrical conductivity induces a

voltage which is detected by two electrodes located in a 90 degree

angle from the magnetic field and the flow direction.

1.2. Applications

Water / waste

water Chemical

industry

Food industry

Power

engineering Agriculture

industry

1.3. Safety Instructions

Please read this manual carefully before using the product

Keep this manual for future reference. Arkon Flow Systems, s.r.o will not be

liable for any damage caused by improper use of the product or its accessories.

The MAG X1 flow converter - flow-meter must not be mounted in explosive

hazardous areas.

1.4. Unpacking the flowmeter

While unpacking the flowmeter, conduct a visual check of the flowmeter upon receipt to

make sure the product has not been damaged during transport.

Check the completeness of the package. In case of any problem, contact the Arkon sales

department without delay.

Flowmeter +

modules Cables CD-ROM + Manual Mounting kit + Key

Minimum liquid conductivity >20 µs / cm for demineralised cold water.

>5 µs / cm for other liquid.

Liquid velocity min. 0.1 m / sec, max. 10 m / sec.

MagX1 Manual -3- V7 03-06-2009, Modbus

2. Installation

2.1. Remote or Compact?

Any MagX1 flowmeter can be delivered in two versions; Compact or Remote. The compact

version has the transmitter unit connected directly to the sensor body. This version does not

require any further mounting or installation of the transmitter.

The remote version has a separated transmitter. It is connected to the sensor with a cable. The

cable entry into the sensor is protected by a junction box, which can be potted to IP68 (page 7).

The cable entry on the transmitter side is through a M16x1,5 gland.

The cable type used for the connection between sensor and transmitter for remote versions:

UNITRONIC® LiYCY (TP) 0035 830, 2x2x0,5

The MagX1 is equipped with an electronic board located inside the sensor neck. This board

sends a digital signal to the transmitter, unlike traditional flowmeters, which send an analogue

signal. This allows the MagX1 to carry its signal over much longer distances than conventional

flowmeters; up to 500m is possible.

MagX1 Manual -4- V7 03-06-2009, Modbus

>5

m

2.2. Sensor installation

Sensor dimensions can be found on page 47

Proper sensor installation is extremely important in order for your flowmeter to work correctly.

Below, you will find minimum sensor installation requirements that need to be respected at all

time.

Sensor installation requirements

All MagX1 sensors are supplied with a built in earthing electrode which is sufficient for all

applications with metal pipes and tanks. However on applications where all pipes and

tanks are manufactured from plastic, it is recommended that earthing rings are also

installed to ensure the maximum resistance of the sensor to earth is <1 ohm.

Sensor grounding with earthing rings

5xDN 3xDN 5xDN

3xDN

Steel Pipe Plastic Pipe

MagX1 Manual -5- V7 03-06-2009, Modbus

2.3. Dry liner

Flowmeters with a Hard Rubber liner can show incorrect readings during the first 2-3 days after

installation. This is due to the fact that the time needed for transport and the time before

installation is long enough for the liner to dry out , and thus it changes shape/size. This change,

in effect, affects reading accuracy. Simply be keeping the meter wet, this problem solve itself

within 2-3 days and no other action is required at all.

2.4. Installation of the transmitter

In case of a compact flowmeter version, the transmitter will need no

further installation, and should be ready for use.

In case of a remote version, the following 4 steps are necessary.

Mount the transmitter to a wall, panel, or DIN-rail.

Wall mounting:

DIN-rail mounting:

Panel mounting:

The electronics have to be

protected against direct

sunlight and high

temperatures!

The transmitter housing

should be exposed to

minimal mechanical

strain only (max. 1kg)!

MagX1 Manual -6- V7 03-06-2009, Modbus

Connect the transmitter to the signal cable from the sensor. To do

this, first open the transmitter housing by disconnecting the two parts of

the transmitter housing with the metal “key” that is provided with the

meter.

After the meter is opened, pull the signal cable through the cable gland on the bottom of the

transmitter housing (see page 3).

Connect the connector at the end of the signal cable from the sensor to the transmitter circuit

board.

Connect the transmitter to network power.

Electrical installation should only be done by a qualified person. Standard safety

regulations for hazardous electrical installations have to be respected.

The customer is assumed to supply its own network power supply cable (90-250VAC, 24VDC or

12VDC, depending on the type of power supply ordered with the flowmeter). Before connecting

to network power, the cable needs to be properly connected to the transmitter.

First pull the cable end through one of the cable glands (ideally the first on the left looking at the

transmitter housing from the back-side) on the bottom of the transmitter.

3x1mm recommended cable.

After connecting the power cable, close the housing and connect the cable to network power.

This will make the flowmeter switch on.

Set up the transmitter for use.

You are now ready to start using your flowmeter or to customize its settings as per your

requirements. For example;

- Set-up the measurement unit of flow-rate displayed, e.g. m3/hr.

- Set up of the unit for the volume displayed. For all volume counters this same unit will be used.

MagX1 Manual -7- V7 03-06-2009, Modbus

2.5. Module installation

- Always check if the module does not have a bent or broken pin, before placing it in the

correct module slot of the MagX1 motherboard.

- Always make sure you place the module in the correct slot of the motherboard! The

name that is written on the module itself has to match the name written next to the slot. For

example, the picture below indicates the correct slot for the F -out module. Placing the

module in an incorrect slot can cause damage to the module and the motherboard, and

has to be avoided by careful module installation!

MagX1 Manual -8- V7 03-06-2009, Modbus

- Check whether you are placing the module in the correct position. It does matter how you

turn the module to fit the slot! The white line around the actual slot on the motherboard

indicates the correct position of installation. The bevelled corner should be your point of

orientation (note the picture below).

- Now you can place the module in its slot:

These are examples of incorrect installations:

Example one: The F-out Module is placed in a different slot

Example two: The F-out Module is placed in the incorrect slot, and with the bevelled corner in

the wrong direction.

Any connection or disconnection of any module has to be done with network power

to the meter switched off.

MagX1 Manual -9- V7 03-06-2009, Modbus

2.6. Cables connections

The following diagram shows the connections of the cables between sensor and

transmitter.

2.7. Potting the remote sensor terminal box

To guarantee IP68 protection of the sensor, it is necessary to pot the sensor terminal box

properly. The proper way to do this is described below:

Plug the connectors into the sensor (white and green wire adjacent).

Screw the small terminal box to the sensor neck (4 screws).

Fill the terminal box completely with silicone, by squeezing it through the opening on

top of the box.

Close the small terminal box with the sealing screw.

MagX1 Manual -10- V7 03-06-2009, Modbus

Remote sensor with terminal box Preparing connections

Screw the sensor neck and box together Opening for terminal box potting

Completely filled terminal box Close the terminal box with the sealing screw

MagX1 Manual -11- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

MagX1

Transmitter UnitTransmitter “MagX1

Transmitter Unit”

The MagX1 Transmitter unit is the main part of the flowmeter. It consists of the MagX1

motherboard, a graphical display, touch-button controls and a transmitter housing.

Through the display and with help of the push buttons, you can go through the various menu’s

for data reading, configuration and setup of your flowmeter.

The following symbols are used in this manual and on the

flowmeter display.

ENTER UP

Esc LEFT

Back Menu

DOWN Selection

menu

RIGHT

Touch-buttons are working on capacitance principle therefore any conductive material close to

button's area will cause button press. Even water can do it so its strongly recommended to use

keylock when any presence of water is expected. 30 seconds after turning the flowmeter on,

touch-buttons autocalibration is started so function of the touch-buttons may be unstable.

The MagX1 transmitter has a keylock possibility. You can lock touch-

buttons by touching both the Esc and Enter together for longer than 1sec.

This will lock the flowmeter and there will be a lock symbol on display.

Touching the buttons will have no effect on flowmeters function. To

unlock buttons touch Esc and Enter together again. If flowmeter is in cleaning electrode there is

a lightning symbol on display.

Upon starting the flowmeter, you will automatically see the main screen of the menu.

3.1. Main screen

Total Volume

This is the total volume counter; the sum of all historical flows for a

particular flowmeter. The user is not able to zero this counter without

use of the service password. Direction of flow is ignored for this

counter (negative flow is calculated the same way as positive flow).

Positive Volume

This counter is only credited when the measured medium is flowing in

the chosen positive direction. In case the flow is 0, or if it is flowing in

the opposite (negative) direction, the number on the counter remains

the same.

Negative Volume

This counter works the same way as the positive volume counter, yet

in the opposite direction. In case the flow is 0, or flowing in the

designated positive direction, the number on this counter will remain

the same.

MagX1 Manual -12- V7 03-06-2009, Modbus

Auxiliary Volume

This is a 2nd

total volume counter. It works the same as the Total

Volume counter, yet with the only difference being that it can be reset

to 0 at any time, with User Settings password. Temperature

This item is a temperature indication for the measured medium.

You can cycle through these 5 indication screens by pressing the buttons on the transmitter.

It is possible to change the number of digits of the values displayed in the main screen:

decimal numbers -->

By using the buttons.

3.2. Flowmeter Menu

After pressing the button you get to into the root-menu.

From here, you can chose any of the sub-menu’s displayed in

the picture on the right.

Navigate with and select your choice with .

3.3. Info Menu

Date This item shows the current date according to

the transmitter’s setup if the Real-Time

Clock module is installed. It can be changed

in the User Settings menu.

Flow statistic This item gives flow statistics for a given

period of time. Data is taken from the

Memory module, which needs to be installed

for this function to work. More details on

page 36.

Unit No. Displays the serial number of the

motherboard. This number is allocated

during production by the manufacturer.

Sensor Unit

No.

Displays the serial number of the sensor.

This number is allocated during production

by the manufacturer.

Error (min) The number of minutes the device was not

measuring because of errors.

OK (min) The number of minutes that the device

measured correctly.

MagX1 Manual -13- V7 03-06-2009, Modbus

Diameter

This item shows the nominal sensor diameter

that is currently configured for the given

flowmeter.

Flow Qn Here, the flowmeter displays the predicted

flow-range limits. Values can be changed

under User Settings.

Firmware No. This shows the current firmware version.

Error info Here, the user can check the identification

code of the last error the device found.

Power

frequency

Identifies the network power supply

frequency.

3.4. Flow Statistic menu

Flow Statistic

Total +

Total -

Total

5min

Back without changes

Item selection

Confirmation of setup and menu selection.

Total +

Total + Month

Total + Day

Total + Hour

Back without changes

Item selection

Confirmation of setup and menu selection.

Total + Month – Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

MagX1 Manual -14- V7 03-06-2009, Modbus

Statistics window

Back to menu

Scrolling through statistics

Total + Day - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

Total + Hour - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

Total -

Total - Month

Total - Day

Total - Hour

Back without changes

Item selection

Confirmation of setup and menu selection.

MagX1 Manual -15- V7 03-06-2009, Modbus

Total - Month - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

Total - Day - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

Total - Hour - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

MagX1 Manual -16- V7 03-06-2009, Modbus

Total

Total Month

Total Day

Total Hour

Back without changes

Item selection

Confirmation of setup and menu selection.

Total Month - Option to configure an estimated flow figure.

Back without changes

Item selection

Value setting

Confirmation of setup and menu selection.

Statistics window

Back to menu

Scrolling through statistics

Total Day - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

Total - Hour - Option to configure an estimated flow figure.

Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

MagX1 Manual -17- V7 03-06-2009, Modbus

Statistics window

Back to menu

Scrolling through statistics

3.5. Display menu

Unit Flow

Setup of the displayed measurement unit for current

flow.

UKG / min UK gallon per minute

USG / min US gallon per minute

m3 / h Cubic meters per hour

l / min Litres per minute

l / s Litres per second

(item selection confirm , selection identification back )

Unit volume

Setup of the displayed measurement unit for total flow.

UKG UK gallon

USG US gallon

m3 Cubic meter

l Litre

(item selection confirm , selection identification back )

5 min - Option to configure an estimated flow figure. Back without changes

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Statistics window

Back to menu

Scrolling through statistics

MagX1 Manual -18- V7 03-06-2009, Modbus

Unit temperature

Setup of the displayed measurement unit of

temperature indication.

C Degrees Celsius

F Degrees Fahrenheit

(item selection confirm , selection identification back )

Contrast

Contrast of the display setup.

Possible range: 0 – 100 %

Back with no change

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Language

Menu language setup.

ENG English

SPA Spanish

(item selection confirm , selection identification back )

MagX1 Manual -19- V7 03-06-2009, Modbus

3.6. User Settings Menu

To enter the User Settings menu, it is necessary to enter the user password. The default

factory setting is 1111. See page 27 for user password settings.

After entering the correct User Password, you get into

the User Settings Menu. By selecting your choice, you

can make changes to specific settings.

Measurement

This option allows for selecting flow measurement to be on or

off.

RUN The device is measuring, the totalisers are active.

STOP The display will show a value, yet the totalisers

are off.

(item selection confirm , selection identification back )

MagX1 Manual -20- V7 03-06-2009, Modbus

Air detector:

This option allows selecting empty pipe check (air detector)

to be on or off.

ON The detector is active

OFF The detector is switched off

(item selection confirm , selection identification back )

Air Constant

Constant value to determine the Empty Pipe detection limit.

Possible range: 0.000 – 0.999

Back with no change

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Delete Auxiliary Volume

This function serves to zero the auxiliary flow totaliser.

No change

Zero the auxiliary totaliser

(item selection confirmation back ,)

Start Delay

Time delay for the flowmeter where it, after switching on,

will not request measurement data from the sensor.

Possible range: 0 – 120 s

Back with no change

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Samples per Avg.

The number of samples that the flowmeter will use for

calculation of its displayed average flow value/time unit.

Possible range: 5-60 samples/avg

Back with no change

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Low Flow Cut-off

This function serves to set the minimum flow the flowmeter

will react on.

Back with no change

Value setting

Confirmation of setup and saving to memory.

(item selection confirm selection identification back )

Flow Qn

MagX1 Manual -21- V7 03-06-2009, Modbus

This function serves to setup the expected flow-rate.

Possible range: 0 – 10000 l/s

Back with no change

Selection of digit position

Value setting

Confirmation of setup and saving to memory.

Invert Flow

This function serves to change the definition of flow

direction.

Back with no change

Choice selection

Confirmation of setup and saving to memory.

Current Loop

Settings Signal

Direct Driving

Back with no change

Item selection

Confirmation of setup and saving to memory.

Settings Signal

This function serves to select which signal the output

should be giving.

Back with no change

Choice selection

Confirmation of setup and saving to

memory.

Flow + Output: 10mA for any positive flow.

Flow - Output: 10mA for any negative flow.

Error Output: 10mA, for any error identified by

the device. The signal can be cancelled

by pressing any push button on the

flowmeter.

Air Detect Output: 10mA, during air detection

(empty pipe).

Fixed Output: fixed output of 10mA

Direct Driving Output: Direct Driving – setup is below

OFF Output: fixed output of 4mA

MagX1 Manual -22- V7 03-06-2009, Modbus

Direct Driving

This function serves to set flow values in relation to current

output. Possible range: 0.000 – 10000 l/s, 4 – 20mA

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Flow

min.-max

Setup of measurement flow-range (only positive values)

Current

min.-max

Setup of the current output range, corresponding to the actual flow-rate

within the given range.

When changing an item for the current

loop output, all settings for the voltage output

are to be changed, to make sure there will be no

signal conflict at the output port. The output that

is not used has to be switched off (Settings –

Signal – OFF).

Calibration

This function serves to modify current loop output signal.

Possible range: 4 – 20mA, 0.5000 – 1.5000

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to

memory.

Calibration

point 1,2

Setup of calibration point 1, 2. First point must be less than second point.

Calibration

constant 1,2

Setup of calibration constant for first and for second calibration point.

Formula for calibration constant calculation:

Expected value: 6 mA, Measured value: 6.1 mA

Calibration point one: 6mA

Calibration constant one = = 0.9836

Expected value: 18 mA, Measured value: 17.9 mA

Calibration point two: 18mA

Calibration constant two = = 1.0056

Voltage Output

Settings Signal

Direct Driving

Back with no change

Item selection

Confirmation of setup and saving to

memory.

min max

max

min

Flow

Current

6

6.1

18

17.9min max

max

min

Flow

Current before

after

MagX1 Manual -23- V7 03-06-2009, Modbus

Settings Signal

This function serves to select which signal the output

should be giving.

Back with no change

Item selection

Confirmation of setup and saving to

memory.

Flow + Output: 5V for any positive flow.

Flow - Output: 5V for any negative flow.

Error Output: 5V, for any error identified by

the device. The signal can be cancelled

by pressing any push button on the

flowmeter.

Air Detect Output: 5V, during air detection (empty pipe).

Fixed Fixed output of 5V

Direct Driving Output: Direct Driving – setup is below

OFF Output is OFF (0V).

Direct Driving

This function serves to set flow values in relation to voltage

output. Possible range: 0.000 – 10000 l/s, 0 – 10V

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to

memory.

Flow

min.-max

Setup of measurement flow-range (only positive values)

Voltage min.-

max

Setup of the voltage output range, corresponding to the actual flow-

rate within the given range.

• When changing an item for the Voltage output,

all settings for the current output are to be changed, to

make sure there will be no signal conflict at the

output port. The output that is not used has to be

switched off (Settings – Signal – OFF).

Relay 1

Settings Signal

Comparator

Back with no change

Item selection

Confirmation of setup and saving to

memory.

min max

max

min

Flow

Voltage

MagX1 Manual -24- V7 03-06-2009, Modbus

Settings Signal

This function serves to select which signal the output

should be giving.

Back with no change

Item selection

Confirmation of setup and saving to

memory.

Flow + Output: ON, for any positive flow

Flow - Output: ON, for any negative flow

Error Output: ON, for any error identified by the

device. The signal can be cancelled by

pressing any push button on the flowmeter.

Air Detect Output: ON, during air detection (empty pipe).

Fixed Output: ON, fixed status signal

Comparator Output: Comparator – setup is below

OFF Output: OFF, fixed status signal

Comparator

Comparator mode

Comparator Flow

Back with no change

Item selection

Confirmation of setup and saving to memory.

Comparator mode

This function serves to select which signal the output should

be giving.

Back with no change

Item selection

Confirmation of setup and saving to memory.

On In Output: ON, if the actual flow-rate is within the given range (can be set

under Comparator Flow).

On Out Output: ON, if the actual flow-rate is outside the given range (can be set

under Comparator Flow).

On>F1 Output: ON, if the actual flow-rate is bigger than the value set as “Flow1”

(can be set under Comparator Flow).

On<F1 Output: ON, if the actual flow flow-rate is smaller than the value set as

“Flow1” (can be set under Comparator Flow).

OFF Output: OFF.

Comparator Flow

Possible range: 0.000 – 10000 l/s, 0.000 – 10000 l/s

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Flow 1, 2 This function serves to configure the flow-range for the Comparator Mode.

Flow 1 < Flow 2.

Hysteresis Setup of hysteresis for the Comparator Mode.

MagX1 Manual -25- V7 03-06-2009, Modbus

When configuring Relay 1, the following is valid::

Settings - Signal or Comparator mode. Any unused setting needs to be OFF. To switch off

Relay 1, both options need to be OFF. When configuring Settings – Signal to any other value

than OFF, the Comparator Mode setting is automatically changed to OFF.

On In:

On Out:

On > F1:

On < F1:

Relay 2

Litres/1(Q+)

Litres/1(Q-)

Dosing

Back with no change

Item selection

Confirmation of setup and saving to memory.

Litres/1 (Q+)

This function serves to configure the positive flow volume

after which a 1s output pulse is generated through Relay 2.

In case of a power failure, the output will start counting

volume from 0. Possible range: 0.000 – 99999 l

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

ON

OFF

F1 F1 + H1 F2F2 – H2

ON

OFF

F1 F1 + H1

ON

OFF

F1 F1 + H1

ON

OFF

F1 F1 + H1 F2 – H2 F2

MagX1 Manual -26- V7 03-06-2009, Modbus

Litres/1 (Q-)

This function serves to configure the negative flow volume

after which a 1s output pulse is generated through Relay 2.

Possible range: 0.000 – 99999 l

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Dosing

This function serves to control dosing (batching). A dose is

activated through a pulse input on binary input. At the same

time with the Relay 2, the totaliser that counts the total flow

of the batch is activated. After reaching the required

volume, Relay 2 is closed, and the totaliser stops. Possible

range: 0.000 – 99999 l

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Note: You cannot set a non-zero value to more than one item. That means that only one of the

three options Litres/1 (Q+), Litres/1 (Q-) or Dosing can be set to a value different than 0. In

case you want the Relay 2 to be switched off, you need to set 0 everywhere. In case Relay 2 is

not used, Arkon recommends configuring a 0 for all three items. When setting a value for any

of the three, the other two are changed to 0 automatically. The user Binary Input of the Binary

module is activated with the end of a pulse.

Litres/1 (Q+)/(Q–): Dosing:

F-out

Settings Signal

Direct Driving

Back with no change

Item selection

Confirmation of setup and saving to memory.

Settings Signal

This function serves to select which signal the output should

be giving.

Back with no change

Item selection

Confirmation of setup and saving to memory.

Flow + Output: 100Hz, for any positive flow

Flow - Output: 100Hz,for any negative flow

Fault Output: 100Hz, for any error identified by the

device

IN

RELE2 Dosing

MagX1 Manual -27- V7 03-06-2009, Modbus

Air

Detect

Output: 100Hz, during air detection (empty pipe).

Fixed Output: 100Hz fixed output

OFF Output: OFF

Direct Driving

This function serves to set flow values in relation to

frequency output. Possible range:

0.000 – 10000 l/s, 0 – 1000 Hz

Back with no change

Change of value

Selection of digit position

Confirmation of setup and saving to memory.

Flow min. - max. Setup of the active flow-range for the Frequency output module.

F min. - max Configuration of the Frequency output range, corresponding to the

actual flow-rate within the given range.

Minimal frequency 2 Hz

Duty Cycle

Function to set duty cycle of the Frequency output.

Percentage of high level.

Possible range: 1 – 99 %

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Speaker output

This function serves to select which signal the output should

be giving.

Back with no change

Item selection

Confirmation of setup and saving to memory.

Flow +/- Sound signal when flow direction changes

Fault Sound signal for any error identified by the device. This signal is active until

stopped by pressing any push button on the flowmeter.

Air detect Sound signal during air detection (empty pipe).

Fixed Constant sound signal

OFF Sound signalling switched off.

min max

max

min

Flow

F [Hz]

MagX1 Manual -28- V7 03-06-2009, Modbus

Load Default Settings

This function will load default factory settings.

Back with no change

Loads default settings

(item selection confirm , back )

Date Setting

Function to set date.

Back with no change

Change of value

Selection of digit position

Confirmation of setup and saving to memory.

Date

format.

DD\MM\YY

Time Setting

This function serves to set current time

Back with no change

Change of value

Selection of digit position

Confirmation of setup and saving to memory.

Time

format.

HH:MM

Password Setup

This function serves to setup the flowmeter user password.

Possible range: 0000 – 9999

Back with no change

Change of value

Selection of digit position

Confirmation of setup and saving to memory.

Modbus

Slave Address

Baud Rate

Parity

Interframe Spacing

Response Delay

Back with no change

Item selection

Confirmation

Slave Address

Modbus device address (Factory settings: 1)

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

MagX1 Manual -29- V7 03-06-2009, Modbus

Baud Rate

Setup communication speed (Factory settings: 9600)

Back with no change

Value setting

Confirmation of setup and saving to memory.

(item selection confirm selection identification back )

Parity

Setup communication parameters (Factory settings: Even, 1

stopbit)

Back with no change

Value setting Confirmation of setup and saving to memory.

(item selection confirm selection identification back )

Interframe Spacing

The minimum interframe space between two MODBUS

RTU messages in sequence (specified as 3.5 characters) is

configurable. Possible range: 3.5 – 25.0

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Response Delay

The minimum time from when a slave receives a request

and until it returns a response. This makes it possible to

send data to slow masters without overwhelming its

receiver. (Factory setting: 50) Possible range: 0 – 255 ms

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Electrode Cleaning

Power

Clean Time

Start Now

Back with no change

Item selection

Confirmation

MagX1 Manual -30- V7 03-06-2009, Modbus

Power

Setup automatic electrodes cleaning

Back with no change

Value setting

Confirmation of setup and saving to memory.

(item selection confirm selection identification back )

Clean Time

Setup clean time for automatic electrode cleaning

(Factory setting: 500) Possible range: 1 – 9999 s

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

Start Now

Start automatic electrode cleaning now

Back with no change

Loads default settings

(item selection confirm , back )

Totalizer Cycling

Automatic totalizer cycling (time is set to 3 seconds)

Back with no change

Value setting

Confirmation of setup and saving to memory.

(item selection confirm selection identification back )

MagX1 Manual -31- V7 03-06-2009, Modbus

3.7. Service Settings Menu

To enter this part of the menu, it is necessary to enter the Service Password.

Error delete:

This option serves to zero the totaliser for number of minutes

the flowmeter signalised an error.

Item selection

No change

Resets error minute totaliser

OK delete

This option serves to zero the totaliser for number of minutes

of operation.

Item selection

No change

Resets error minute totaliser

Delete volume –

Option to zero the totaliser for negative flow.

Item selection

Confirmation

Back without change

Delete volume +

Option to zero the totaliser for positive flow.

Item selection

Confirmation

Back without change

MagX1 Manual -32- V7 03-06-2009, Modbus

Delete volume total

Option to zero the totaliser for total flow.

Item selection

Confirmation

Back without change

Datalogger delete:

This function serves to erase the memory chip of the memory

module, which holds a list of measured values.

No change

Delete all memory values

(item selection confirmation back ,)

Flow Simulation

Switching on/off the flow simulation status

ON Flow Simulatin status is ON

OFF Flow Simulatin status is OFF

Item selection

Confirmation

(selection identification , back )

Simulated Flow

Simulation flow percentage of Flow Qn

(Factory setting: 50) Possible range: 0 – 100 %

Back with no change

Selection of digit position

Change of value

Confirmation of setup and saving to memory.

3.8. Factory Settings Menu

To enter this part of the menu, it is necessary to enter the Factory Password.

This function is only available to Arkon staff in the Arkon workshop

MagX1 Manual -33- V7 03-06-2009, Modbus

3.9. Authorize Menu

To enter this part of the menu, it is necessary to enter the Autorize Password.

This function is only available to Arkon staff in the Arkon workshop.

Authorize - Password Setup

Here, it is possible to change a forgotten user password. It

is necessary to call the Arkon sales office and provide the

serial number. The authorisation number is provided based

on this serial number.

Item selection

Confirmation

Back without change

Enter value between 0000 and 9999.

MagX1 Manual -34- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Power Supply

Module PS Module

PS3 - “Power Supply

Module 230VAC”

PS3 - “Power Supply

Module 24VDC”

PS3 - “Power Supply

Module 12VDC”

APPLICATIONS: PIN CONFIG

Industrial Power Supplies, 24VDC and

12VDC Distributed Power Systems

Electrical Specifications Power consumtion

[mA]

12

V

24

V

230

V

Flowmeter 770 340 50,3

Flowmeter, no sens. 73 34 20,4

+TCPIP 98 65 17,1

+IOUT 41 33 3,3

+VOUT 15 8 0,9

+MEMORY+RTC 15 8 0,9

+USB 10 6 0,6

+RS232 5 4 0,3

+RS485 22 17 1,8

+BT 15 9 0,9

+FOUT 5 4 0,5

+BIN 7 5 0,5

Input Voltages ±5% /

possible current

consumption

AC90 - 250V / 20mA – 100mA

DC24 / 30mA – 600mA

DC12 / 70mA – 1050mA

Output Voltages 5V / 2A

23.6V/300mA

Frequency Outputs 50-60Hz (TTL)

Temp. Range -20 – 70°C

Dimensions: R = 50mm

H(230V) = 58mm

H(12,24V) = 58mm

Weight 300g

Connection of the flowmeter is to be done via a switch or circuit breaker for safety

purposes. The device does not have a network power switch. For any electrical

work it is necessary to disconnect the device from the network power, and this has

to be done via a switch.

90-250V/0,45A AC

Recommended power supply

cable minimum 3xØ0,25

24V/1A DC

Recommended power supply

cable minimum 2xØ0,25

12V/2A DC

Recommended power supply

cable minimum 2xØ0,5

All used wires have to be round crosscut cables.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -35- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

RS232 Module RS232 Module “RS232 Module”

APPLICATIONS: PIN CONFIG

Any System Requiring RS-232 Communications, Peripheral - PC and

Terminal, Data-loggers

Electrical Specifications

VCC to Ground 4.5V to 5.5V/5mA

Propagation Delay RS232-TTL/0,5us

Baud rate Max. 115200 baud/s, standard 9600 baud/s.

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS:

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -36- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Real-Time

Clock Module RTC Module “Real-Time Clock

Module”

APPLICATIONS: PIN CONFIG

Real-Time Clock Counts Seconds, Minutes, Hours, Date of the Month,

Month, Day of the Week, and Year with leap-year compensation valid

Up to 2100, Data-loggers

Electrical Specifications

VCC to Ground 2V to 5.5V/0.5mA, 1.3mA

CLK Frequency Vcc=2V-Max. 0.5 MHz

Vcc=5V- Max. 2.0 MHz

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS:

This module is prerequisite when using the Memory (data-logger) module.

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -37- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

USB Module USB Module “USB Module”

APPLICATIONS: PIN CONFIG

Any System Requiring, USB Communications, Peripheral - PC and

Terminal, Data-loggers. USB 1.1 and USB 2.0 compatible

Electrical Specifications

VCC to Ground 4.35V to 5.25V/25mA

Baud rate Max. 115200 baud/s,

standard 9600 baud/s.

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS:

VIRTUAL COM PORT (VCP) DRIVERS:

http://www.ftdichip.com/Drivers/VCP.htm

Driver installation guide: http://www.ftdichip.com/Documents/InstallGuides.htm

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -38- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Frequency

Output ModuleFreq-Out Module “Frequency

Output Module”

APPLICATIONS: PIN CONFIG

Industrial Automation, Industrial process, Control, Test Systems, Smart

Transmitter

Electrical Specifications

VCC to Ground 4.5V to 5.5V/25mA

VOPK Typ. 5V, Max. 48V

IOPK Max. 50mA

Frequency out

Duty Cycle

2 Hz – 1000 Hz

1 – 99 %

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS

Optimal resistance of the collectors in relation to the maximum

frequency 12kHz.

VCC R1

5 V 1k8

12 V 3k3

24 V 6k8

Duty cycle can be changed in User Settings – F-Out – Duty Cycle (1 - 99 %)

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -39- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Memory Module Memory Module

“Memory

Module”

BASIC CIRCUIT CONNECTIONS:

For proper functioning of this module, the

Real-Time Clock (RTC) Module is

prerequisite.

Important: Before you save your first data, you need to

set the correct time and date; else all saved

data will be without proper time reference!

To be able to do this, you also need to use the

MagX1 RTC (Real-Time Clock) Module.

(page 32)

These values can be uploaded to a PC with the help of special software, which is supplied

together with the configuration SW and the manual.

If capacity runs out before all data is deleted from memory, then the module will

automatically overwrite the oldest values with new ones! In case memory is used, it has to be

“cleaned” at least once every year (Datalogger Delete function), else older data will be lost.

The Memory Module has sufficient capacity for storing flow-data for a 12-month operation

period without overwriting older data.

Save intervals:

- 5-minute interval record of flow averages (8928 samples in 1 month)

- 1-hour interval record of total flow (744 samples in 1 month)

- 24-hour interval record of total flow (31 samples in 1 month)

- Monthly interval record of total flow (1 sample in 1 month)

You can view this data in menu 1.2. Flow Statistics (3.4. Flow statistic menu; page 13)

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

APPLICATIONS: PIN CONFIG

Data-loggers, Memory capacity is 1 MB (megabyte)

Electrical Specifications

VCC to Ground 3.2V to 5V/2-15mA

Max frequency SCK 20MHz

Temp. Range -20-70°C

5.0V-tolerant Inputs: SI, SCK, CS, RES., WP Pins

MagX1 Manual -40- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

RS485 Module RS485 Module “RS485 Module”

APPLICATIONS: PIN CONFIG

Industrial Automation, Industrial Process Control, Peripheral - PC and

Terminal, Data-loggers

Electrical Specifications

VCC to Ground 4.5V to 5.5V / 55mA

Baud rate Max. 115200 baud/s

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS:

Multi-Node Network with End Termination Using

Module RS485

For computers that do not have an RS485 port, we can provide an RS485 to RS232 or an RS485 to

USB convertor. Please contact the Arkon sales office for details.

Terminator Rt with resistance 100ΩΩΩΩ should be connect to the end of line RS-485.

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -41- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Current Loop

Output Module I-Out Module “Current

Loop Module”

APPLICATIONS: PIN CONFIG

Industrial Automation, Industrial Process Control, Test Systems,

Smart Transmitter

Electrical Specifications:

12-bit DA convertor

Maximum Resolution 3,9 uA.

VCC to Ground 4.5V to 5.5V / max 125mA

Current out 4 – 20 mA

Output mode Active or Pasive

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS: Using the Current Loop Module

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -42- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Binary In- and

Output Module BIN Module

“Binary

Module”

APPLICATIONS: PIN CONFIG

Industrial Automation, Industrial Process Control, Test

Systems, Smart Transmitter

Electrical Specifications

VCC to Ground 4.5V to 14V / 25mA

Input Max. 14V, Typ. 5V, Min. 1.6V

Switching Outputs Max. 200mA

Temp. Range -20-70°C

Output frequency Max 1 puls per second

BASIC CIRCUIT CONNECTIONS: Using the Binary Module:

Recommended connections:

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -43- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

Voltage Output

Module V-Out Module

“Voltage

Output

Module”

APPLICATIONS: PIN CONFIG

Industrial Automation, Industrial Process Control, Test

Systems, Smart Transmitter

Electrical Specifications

12-bit DA convertor

Maximum Resolution 2,44 mV

VCC to Ground 4.5V to 5.5V / 25mA

Voltage Output 0 – 10V

RZ Min. 1kΩ

Temp. Range -20-70°C

BASIC CIRCUIT CONNECTIONS: Using the Voltage Output Module

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -44- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

TCP/IP Module TCP/IP Module “TCP/IP

Module”

APPLICATIONS: PIN CONFIG

You can make any electronic product a smart device with

embedded network intelligence, enabling it to deliver

innovative applications such as:

• Remote Diagnostics and Upgrades

• Asset Tracking and Replenishment

• Automation and Control

• Power Management

• Remote Collaboration

• Personalized Content Delivery

Electrical Specifications

VCC to Ground 3.14V to 3.46V

Power Sup. Current 120 – 267mA

Ethernet 10/100Mbit

Temp. Range -40-75°C

BASIC CIRCUIT CONNECTIONS: Using the TCP/IP Module

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -45- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

BLUETOOTH

Module BT Module

“BLUETOOTH

Module”

APPLICATIONS: PIN CONFIG

Wireless control of and communication between transmitter

and PC or PLC systems

Any System Requiring BlueTooth Communications

Electrical Specifications

VCC to Ground 3.6V to 6.0V

Power Sup. Current 120mA

Baud Rate Max. 460.8 Kbaud/s

Carrier Frequency 2.402 – 2.480 GHz

Range 100m (class 1)

Temp. Range –10 - +70°C

BASIC CIRCUIT CONNECTIONS: Using the TCP/IP Module

Warning!: There is a condition that must be fulfilled for the Bluetooth module to be able to

operate correctly: line speed of the communication protocol MODBUS must be set up on

19200Bd, Parity none, 1 stop bit. If there is a different setting the communication will not

work. You can find the setting in the following MagX1 flow meter menu: "Menu /

User settings / Modbus / Baud rate" and "Menu / User settings / Modbus / Parity".

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

MagX1 Manual -46- V7 03-06-2009, Modbus

Module Name: Module Short Name: Symbol: Ordering Code:

GPRS Module GPRS Module “GPRS Module”

APPLICATIONS: PIN CONFIG

Wireless control of and communication between transmitter

and PC or PLC systems

Any System Requiring GPRS Communications

Electrical Specifications

VCC to Ground 5V to 6V

Power Sup. Current RMS 400mA, MAX 1500mA

Baud Rate 19200 baud/s

Operating Systems GSM 850 / GSM 900

DCS 1800 / PCS 1900

Multi-slot class 10 (4 Rx / 2 Tx / 5 Sum)

SIM Card 3.0 / 1.8 V

Temp. Range –20 - + 75 °C

BASIC CIRCUIT CONNECTIONS: Using the GPRS Module

Warning electrostatic sensitive device.

Any connection or disconnection of any module has to be done with the network

power to the meter switched off.

Warning!: There is a condition that must be fulfilled for the GPRS module to be able to

operate correctly: line speed of the communication protocol MODBUS must be set up on

19200Bd, Parity none, 1 stop bit. If there is a different setting the communication will not

work. You can find the setting in the following MagX1 flow meter menu: "Menu /

User settings / Modbus / Baud rate" and "Menu / User settings / Modbus / Parity".

MagX1 Manual -47- V7 03-06-2009, Modbus

17. Module positioning.

Individual module installation is straightforward thanks to a plug-and-play system. Yet, some

caution is required when selecting the correct installation slot according to the picture below.

18. Liner and electrode selection.

Liner and electrode material selection are an important issue when choosing your flowmeter. The

tables below serve to give you an idea of general material compatibility. If you are not sure about

suitability of liner/electrode material for a particular medium, please contact the Arkon sales

department for further assistance, and the site where the flowmeter is to be used for what

materials are acceptable for the process media. Arkon can only recommend materials, we cannot

guarantee them.

Liner Selection:

Hard Rubber Drinking water and wastewater 0 - 70°C

Soft Rubber Water with abrasive particles 0 - 70°C

Teflon Chemicals and food industries 0 - 130°C

Electrode selection:

Stainless Steel General purpose, sewage, water

Hastelloy Seawater, Chemicals

Titanium Aggressive chemicals

Platinum Aggressive chemicals

MagX1 Manual -48- V7 03-06-2009, Modbus

19. Flowmeter Dimensions

DN/PN16 A

[mm] D4

[mm] L3

[mm] DN/PN16 A

[mm] D4

[mm] L3

[mm]

10 279 90 200 10 185 90 200

15 284 95 200 15 190 95 200

20 291 105 200 20 197 105 200

25 300 115 200 25 205 115 200

32 315 140 200 32 221 140 200

40 324 150 200 40 230 150 200

50 338 165 200 50 244 165 200

65 358 185 200 65 264 185 200

80 376 200 200 80 282 200 200

100 396 220 250 100 302 220 250

125 424 250 250 125 330 250 250

150 455 285 300 150 361 285 300

200 514 340 350 200 420 340 350

250 586 405 450 250 492 405 450

300 644 460 500 300 550 460 500

350 705 520 543 350 610 520 543

400 760 580 593 400 665 580 593

450 805 615 593 450 710 615 593

500 885 715 593 500 790 715 593

600 1000 840 597 600 908 840 597

Tolerance of built-in length and height for DIN and ANSI type sensors:

DN 10 – DN 150 A/L ± 1 mm

DN 200 – DN 1000 A/L ± 2 mm

Tolerance of inside diameter:

DN 10 – DN 1000 max. 5%.

MagX1 Manual -49- V7 03-06-2009, Modbus

20. How to order your MagX1

In case you are interested in purchasing a MagX1 flowmeter, you can either contact the Arkon

Sales Department and request a quote to serve as a basis for ordering, or you can use the Arkon

price-list as an easy order form. Due to the design of the MagX1, no single ordering code exists;

all modules you would like to add to your order have to be added individually. Only the MagX1

sensor has its own ordering code:

21. MagX1 Error Code Table

Error Code Error Description

000 Empty Pipe! (Air Detect)

001 Memory Module Not Responding

002 RTC Module Not Responding

003 Sensor Not Responding

004 – 008 GPRS Module – communication error

009 GPRS Module – missing SIM card or wrong PIN code

010 GPRS Module – no GSM signal

011 GPRS Module – Dial up error

012 GPRS Module – Unknown IP address

013 - 014 GPRS Module – Online data mode error

Example of an error warning on the MagX1 display:

MagX1 Manual -50- V7 03-06-2009, Modbus

22. Sensor and transmitter versions and compatibility.

It is possible to check the version of each MagX1 in info menu> Firmware No. (see point 3.3).

When changing the communication module, please check the firmware version to know if they

are compatible.

Firmware version also determines if the unit includes automatic electrodes cleaning or

MODBUS communication and if it is compatible with Bluetooth module.

MagX1 offers the option of updating the firmware version using RS232 communication module,

using MagX1 firmware updater software. That is available from Firmware version v.1.12 or

superior.

The following table shows compatibilities between different versions.

FW MB Buttons F-OUT

[Hz]

Sensor

module

UP

Sensor

module

DOWN

Automatic

electrode

cleaning

BlueTooth

module

MODBUS

PC

Power

Supply

FW

update

v.1.05 v.4 mechanical 250 - 10000 v.6 v.6 no no no PS2_4 no

v.1.06 v.4 mechanical 250 - 10000 v.6 v.6 no no no PS2_4 no

v.1.07 v.4 mechanical 250 - 10000 v.6 v.6 no no no PS2_4 no

v.1.10 v.4 mechanical 250 - 10000 v.6 v.6 no no yes PS2_4 no

v.1.11 v.4 mechanical 250 - 10000 v.6 v.6 no no yes PS2_4 no

v.1.12 v.4 mechanical 250 - 10000 v.7 v.6 no no yes PS2_4 yes

v.2.01 v.6 touch 2 - 1000 v.7 v.7 yes yes yes PS2_5,

PS3

yes

v2.02 v.6 touch 2 - 1000 v.7 v.7 yes yes yes PS2_5,

PS3

yes

v2.03 v.6 touch 2 - 1000 v.7 v.7 yes yes yes PS2_5,

PS3

yes

23. CE and Conformity

The MagX1 Electromagnetic flowmeter is manufactured conform CE requirements.

Conformity

requirements

EN 61010-1 : 2003

EN 61000-3-2:2000, A1:2001, A2:2004

EN 61000-3-3:1995, Cor. 1:1997, A1:2001

EN 61326:1997 + A1:1998, cor. 1:1998 + A2:2001 + A3:2001, Table A.1

EN 61326:1997 + A1:1998 + Cor:1998 + A2:2001, + A3:2003, Class A

MagX1 Manual -51- V7 03-06-2009, Modbus

24. Warranty

The warranty conditions are covered by Arkon Flow Systems, s.r.o. Terms & Conditions of Sale

and by Arkon Flow Systems, s.r.o Return Regulations and Warranty Conditions. The Arkon

Flow Systems, s.r.o Terms & Conditions of Sale and the Arkon Flow Systems, s.r.o Return

Regulations and Warranty Conditions are an integral part of the Resellers contract and of any

Order Confirmation. Please see your Resellers contract or www.arkon.co.uk; Support section.

The Warranty sheet is part of the Packing note of any new goods sent. For the claim or return

procedure, please consult our web site www.arkon.co.uk or call the Arkon Flow Systems, s.r.o

sales office.

Contact:

Technical support: [email protected]

Windows life messenger: [email protected]

Office hours:

8:30 – 18:30 (GMT+1)

Direct technical support:

8:00 – 17:00 (GMT+1)

MagX1 Manual -52- V7 03-06-2009, Modbus