  • Please read this service manual before installing this equipment.

    Global TOP 10 in Healthcare

    LUVIS C500 SERVICE MANUAL

    (V2.1) 10/11/2017

    Dantal Light

    DENTIS CO., LTD.

    99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Tel. 82-53-583-2804 FAX. 82-53-583-2806 www.dentis.co.kr / www.luvis.co.kr

  • Professional LED Light System for All of Surgery Application

    2

    CONTENTS

    1. Standard ..................................................................................................................................................................................... 4

    2. Warning ...................................................................................................................................................................................... 5

    3. Symbol......................................................................................................................................................................................... 7

    4. Safety instruction.................................................................................................................................................................... 8

    5. LIGHTHEAD specification .................................................................................................................................................... 9

    5.1 Technical specification (In accordance with EN/ISO 9680: 2014) .......................................................................... 9

    5.2 Electrical specification (In accordance with IEC 60601-1) .................................................................................... 10

    5.3 Mechanical specification .............................................................................................................................................. 10

    5.4 Camera(optional) specification ................................................................................................................................... 11

    6. Other characteristic ............................................................................................................................................................ 11

    7. Tool for installation............................................................................................................................................................. 11

    8. Product Installation ............................................................................................................................................................ 12

    8.1 Components name ......................................................................................................................................................... 12

    8.2 CHAIR TYPE Installation ................................................................................................................................................. 16

    8.3 MOBILE TYPE Installation .............................................................................................................................................. 18

    8.4 WALL TYPE Installation .................................................................................................................................................. 24

    8.5 CEILING TYPE Installation .............................................................................................................................................. 26

    9. Electrical connection .......................................................................................................................................................... 30

    9.1 Wiring Diagram ............................................................................................................................................................... 31

    10. USE .......................................................................................................................................................................................... 33

    10.1 HEAD CONTROLLER ...................................................................................................................................................... 33

  • Professional LED Light System for All of Surgery Application

    3

    10.2 Remote Control (Optional) ......................................................................................................................................... 34

    10.3 RESIN COVER (Optional) ............................................................................................................................................. 35

    10.4 HANDLE ........................................................................................................................................................................... 35

    10.5 SENSOR ........................................................................................................................................................................... 36

    10.6 Positioning...................................................................................................................................................................... 37

    11. Setting .................................................................................................................................................................................... 38

    11.1 Adjusting the balancing of the ARM ........................................................................................................................ 38

    11.2 OperOperating range of HEAD ARM ........................................................................................................................ 39

    11.3 Operating range of MAIN ARM ................................................................................................................................. 40

    11.4 Adjust tension ............................................................................................................................................................... 44

    12. Model designation ........................................................................................................................................................... 46

    13. List of component ............................................................................................................................................................ 48

    “WARNING: Modification of this equipment is not allowed”

  • Professional LED Light System for All of Surgery Application

    4

    1. Standard

    Ÿ Certification of DENTIS

    ž ISO 9001:2008

    ž EN ISO 13485:2012

    Ÿ Relevant EC Directives: 93/42/EEC Medical Device Directive as amended by 2007/47/EC, Annex I and Annex VII.

    Ÿ Applied Standards:

    ž EN ISO 15223-1:2012, Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied

    ž EN 1041:2008, Information supplied by the manufacturer with medical devices

    ž EN ISO 13485:2012, Medical devices – Quality management systems – Requirements for regulatory purpose

    ž EN ISO 14971:2012, Medical devices – Application of risk management to medical devices

    ž IEC 60601-1:2005+A1:2012, Medical electrical equipment – Part 1: General requirements for safety

    ž EN 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility – Requirements and tests

    ž EN 60601-1-6:2010, Medical electrical equipment-Part 1-6: General requirements for safety – Collateral Standard: Usability

    ž EN ISO 7010:2012, Graphical symbols – Safety colors and safety signs-Registered safety signs

    ž EN 62471:2008, Photo biological safety of lamps and lamps systems

    ž EN 62366:2008, Medical devices – Application of usability engineering to medical devices

    ž EN ISO 9680:2014, Dentistry – Operating lights

    ž IEC 62304:2006, Medical device software, Software life-cycle processes

  • Professional LED Light System for All of Surgery Application

    5

    2. Warning

    WARNING The instructions given in this document must be followed when handling the product. Failure to do so may endanger the safety of the installers or users. As well as specific information on operating the entire product and conducting preventive maintenance, are provided in the USER’S MANUAL. For further information, please contact our sales network or our local network.

    WARNING The electrical connections must be performed by a qualified technician only. The electrical installation must be planned, performed and inspected by electrical engineers.

    WARNING The LIGHTHEAD is designed to operate using AC 100-240V 50/60Hz(ADAPTER), AC 12-18V 50/60 Hz. Higher or lower voltages may affect the light intensity and operating life of the LEDs. Luvis(C500) is class I. equipment. In order to avoid the risk of an electric shock, the equipment must be connected to a mains supply with PE(protective earth).

    WARNING A main control switch must be installed for turning the system power-off. Damaged wire insulation may result in the risk of electric shock.

    WARNING The power supplies may be installed and connected only by an electrician or a DENTIS authorized service agent.

    WARNING This product may only be repaired and special assembly work may only be carried out by DENTIS or a company that has been authorized by DENTIS.

    WARNING Check the polarity of all electrical connections before turning on the power.

    WARNING The LIGHTHEAD brakes are adjusted during installation. Like all mechanical parts, the brakes are subject to wear. Check the condition of the mounting surface.

  • Professional LED Light System for All of Surgery Application

    6

    WARNING Electromagnetic waves discharged from MRI can cause malfunction of this product. Please keep away any device that can influence this product.

    WARNING Do not look directly into light source(LED).

    WARNING The operation and safety of the device may be affected by the removal of certain components during servicing operations.

    WARNING All the information in this manual has been checked out carefully and discerned as accurate one at the time of publication. However, DENTIS takes no responsibilities of the results caused by default, omission, or misuse of it.

    WARNING DENTIS has rights to modify the product itself or specifications of the product without any prior notice, as well as rights not to renew that modification on this manual.

  • Professional LED Light System for All of Surgery Application

    7

    3. Symbol

    Symbol Meaning

    CE label The device bears the CE mark and complies with the requirements of European Directive 93/42/EEC Medical Device Directive as amended by 2007/47/EC, Annex I and Annex VII.

    Medical general medical equipment as to electrical shock, fire and mechanical hazards only in accordance with ANSI/AAMI ES60601-1 & CAN/CSA-C22.2 NO. 60601-1]

    ☞ Recommendation sign

    Protective earth (ground)

    Alternating current

    Off (power: disconnect to the mains)

    On (power: connection to the mains)

    Do not throw away with general household waste

    Warning sign

    Operating instructions

    Keep dry symbol

    DO NOT Hand Hooks symbol

    Fragile symbol

    This side up

    Handle with care symbol

    Do not build up more than 5 boxes

    Manufacture

  • Professional LED Light System for All of Surgery Application

    8

    Europe Representative

    Manufacture Date

    Temperature between 0˚C ~ 40˚C

    Humidity between 0%RH ~ 80%RH

    No pushing

    Do not use this device in a bathtub, shower or water-filled reservoir

    4. Safety instruction

    RECOMMENDATION Ÿ Service personnel must be trained by DENTIS or sales network. Ÿ This document may not be reproduced, in whole or in part, without our

    permission. Ÿ Given the confidential nature of the information in this document, it is

    distributed exclusively to customers and installers of DENTIS products. Ÿ Check with the DENTIS network to ensure that you have the latest versions

    of these documents. Ÿ If you are authorized to engage the services of a contractor for all or part of

    the installation and/or to manufacture certain installation subassemblies, ensure that such subcontracting complies with the terms of the contract which binds you and DENTIS.

    Ÿ Make sure that your subcontractor is properly qualified for the job and ask for proof of certification. Perform regular inspections at the subcontractor's premises and make sure that the subcontractor's facilities meet your own requirements.

    Ÿ If installing on existing studs, guarantees that the studs will withstand the new stresses must be obtained.

    Ÿ Control units must meet applicable standards and fully ensure safe operation, particularly with regard to electric shock. These guarantees must be obtained from the subcontractors.

    Ÿ DENTIS may not be held liable for any damage or injury resulting from failure to follow these recommendations.

  • Professional LED Light System for All of Surgery Application

    9

    5. LIGHTHEAD specification

    5.1 Technical specification (In accordance with EN/ISO 9680: 2014)

    Ÿ STANDARD SPECIFICATION

    Specifications Unit LUVIS 500 Remark BASIC RESIN COVER

    Central illuminance Ec (@70cm)

    Max Lux 3,000 2,000

    @ 4,500K Min Lux 50,000 30,000

    Irradiance (Ee) W/㎡ 171.86 92.27

    Illuminance in patient’s eyes Lux ≤ 1,200

    Hard shadow mm 7 X 3

    Focal Distance Cm 70

    Focal Patter (@ d10) Cm 16.0 X 9.0 Oval shape

    Focal Pattern (@ d50) Cm 11.0 x 6.0 Oval shape

    Uniformity (d50/d10) N/A < 70%

    Color temperature (3steps) K 4,000/4,500/5,000 2,000 - 2,500 Selectable

    C.R.I(Ra) N/A 95 @ 4,000K

    * Optical values are measured with a tolerance of ±10%

  • Professional LED Light System for All of Surgery Application

    10

    5.2 Electrical specification (In accordance with IEC 60601-1)

    Ÿ TECHNICAL DATA

    Content LUVIS C500 Remark

    Input

    AC 12 - 18 V, 50/60 Hz, 24 VA Chair Type

    AC 100 - 240 V, 50 - 60 Hz, 38 - 52 VA Without Camera Type

    AC 100 - 240 V, 50 - 60 Hz, 50 - 70 VA With Camera Type

    Output (Adapter) DC 15V, 1.2 A Nomal

    DC 15V, 1.7A Camera

    LED lamp life 50,000h

    WARNING Please use be connected to a chair that can limit the power supply. If the power supply is not a dental chair, use the special adapter, which Must provided by the DENTIS.

    5.3 Mechanical specification

    Specifications Length(mm) Weight(kg) Remark

    LED HEAD 284 x 165 x 118

    1.6 Head

    1.75 With RESIN COVER

    4.9 With INTERNAL CAMERA

    MAIN ARM - 4.4 FIRST MAIN ARM+ SECOND MAIN ARM

    WALL BRACKET 2.3

    CEILING VERTICAL ARM 2.0

    CEILING MOUNTING SET 4.0

    MOBILE VERTICAL ARM 6.5

  • Professional LED Light System for All of Surgery Application

    11

    5.4 Camera(optional) specification

    Content Specifications REMARK

    Zoom Ratio 10X Optical Type 30X Optical Type SONY BLOCK CAMERA

    Min. Object Distance 10mm to 800mm 600mm to 800mm (Tele end)

    Image Point Approx. 2.38 Megapixels

    Image Sensor 1/2.8 type Exmor CMOS

    Video System Full HD 1080p, 1080i, 720p, 720i

    Video Output Signal 3G HD-SDI SMPTE 424M, HD-SDI 292M 75Ω BNC

    * Select one of the two types of cameras to install

    6. Other characteristic

    Specifications LUVIS C500 Remark

    Protection against electric shock Class I Protection

    Protection against harmful ingress of water or particulate matter Ordinary

    Method of sterilization See the USER’S MANUAL

    7. Tool for installation

    Tool Description

    Allen wrench (1.5mm, 2mm, 3mm) To adjust the head arm tension To adjust the 1st MAIN ARM tension

    “+” Driver To install the screw

    “-” Driver To install the screw To remove the arm cover

    Long-nose To adjust the head arm tension To adjust the 1st MAIN ARM tension

  • Professional LED Light System for All of Surgery Application

    12

    8. Product Installation

    8.1 Components name

    CHAIR TYPE

  • Professional LED Light System for All of Surgery Application

    13

    MOBILE TYPE

  • Professional LED Light System for All of Surgery Application

    14

    WALL TYPE

  • Professional LED Light System for All of Surgery Application

    15

    CEILING TYPE

  • Professional LED Light System for All of Surgery Application

    16

    8.2 CHAIR TYPE Installation

    No. Installation Guide

    1 Initial condition

    · Initial condition of existing halogen

    - Check the power off

    2 Arm detachment

    · Lift the horizontal Arm up.

    - Please turn off the chair power - Be careful with the polarity(DC Input only)

    3 Segregation wires · Cut the Wires, leaving 10cm margins from the top of the Vertical Arm · Remove wire covers, leaving about 1 cm exposed

    - Check the arm bolt size before installation. - Avoid pressure of wire - Avoid peeling of the paint on the arm

    4 Check packaging

    · Check the product releasing condition

    - If damage is found, contact DENTIS

  • Professional LED Light System for All of Surgery Application

    17

    5 Wiring & connecting

    · wiring & connecting chair arm and SECOND MAIN ARM. · INTERNAL CAMERA Model is connected to ADAPTER

    - Please turn off the chair power - Be careful with the polarity(DC Input only)

    6

    COAXAL CABLE Connecting (INTERNAL

    CAMERA Model ONLY

    · SECOND MAIN ARM connected to the chair · Connect the cable to the video transmission device

    - Check the arm bolt size before installation. - Avoid pressure of wire - Avoid peeling of the paint on the arm

    7 SECOND MAIN ARM connecting · SECOND MAIN ARM connected to the chair

    - Check the arm bolt size before installation. - Avoid pressure of wire - Avoid peeling of the paint on the arm

    8 FIRST MAIN ARM connecting · Connecting SECOND MAIN ARM and FIRST MAIN ARM chair arm (including head) · Check the wire(wire harnesses) condition and connecting

    - Avoid pressure of wire - Avoid peeling of the paint on the arm

  • Professional LED Light System for All of Surgery Application

    18

    9 Completion

    · Handle R/L(Check direction of) combinations · Check the operation

    8.3 MOBILE TYPE Installation

    No. Installation Guide

    1 Check packaging

    · Check the product releasing condition

    - If damage is found, contact DENTIS.

    2 CASTER Combination · Open the BASE BODY BOX, take out the product · Put face down on the floor, after checking the conditions · Insert the CASTERs to the BASE BODY by pressing them with your hand(4EA))

    - We recommend that you put gloves on, while working on this step.

    3 BASE WEIGHT Combination

    · Properly align the center hole of WEIGHT to the pole inside the

  • Professional LED Light System for All of Surgery Application

    19

    BASE BODY and insert the BASE WEIGHT to the BASE BODY. · Insert the M6 X L35 Hexagon screws and tighten them with a

    5mm Allen wrench

    - Be careful finger injury. When insert a BASE WEIGHT.

    · Properly align the center hole of WEIGHT to the pole inside the BASE BODY and insert the BASE WEIGHT to the BASE BODY. · Insert the M6 X L55 Hexagon screws and tighten them with a 5mm Allen wrench

    - Be careful finger injury When insert a BASE WEIGHT. - Check the BASE WEIGHT’s orientation.

    4 ADAPTER Combination

    · Insert the adapter into the hole in the BASE WEIGHT · Cover with bracket and tighten them with M3 X L6 bolts

    - You must Avoid pressure of wire.

    5 Cable fixing

    · Secure the AC cables with cable holder M4 X L6 and bolts · Secure the AC cables with cable holder M3 X L6 and bolts

    - It must be secured by bending as shown in the figure.

  • Professional LED Light System for All of Surgery Application

    20

    6 Harness Connection · ADAPTER DC Cable and MOBILE VERTICAL ARM HARNESS connection · Fixed using a shrink tubing

    - The cable must not be fallen.

    7 VERTICAL ARM Combination

    · Combining MOBILE VERTICAL ARM after reverse the BASE BODY · 90 ° degree Tilt

    - Be careful with peeling of the paint on the arm - Please be sure you lock the caster after work

    · Insert the LOCK WASHER · Secure the LOCK NUT by using the long-nose

    - The locknut must be secured.

    · Bend the lock nut by using the long-nose · Insert the step 6 harness through the hole in the MOBILE VERTICAL ARM

  • Professional LED Light System for All of Surgery Application

    21

    · Take out the harness from the MOBILE VERTICAL ARM

    - Be careful that loose harness come loose.

    8 SECOND MAIN

    ARM Combination

    · Insert the harness in the SECOND MAIN ARM · Connect the MOBILE VERTICAL ARM and the SECOND MAIN ARM

    - Be careful with pressed wire. - You must Avoid peeling of the paint on the arm

    9 FIRST MANI ARM Combination · Harness connection · FIRST MAIN ARM and SECOND MAIN ARM combination · After checking the position of the upper MOBILE VERTICAL ARM locking M5 X L10 hexagon bolt with 5mm Allen wrench

    - Be careful with pressed wire. - You must Avoid peeling of the paint on the arm

  • Professional LED Light System for All of Surgery Application

    22

    10 Completion

    · Handle R/L(Check direction of) combinations · Check the operation

    번호 Shelf Installation Guide

    1

    TRAY HOLD

    MOUNT

    Combination

    ∙ Putting the urethane sealing in the TRAY HOLD MOUNT1

    ∙ TRAY HOLD MOUNT2 combination

    - Be careful with peeling of the paint on the arm

    ∙ The holes are positioned correctly

    ∙ Combine Fixing bolts and nuts

    - Be careful with peeling of the paint on the arm

  • Professional LED Light System for All of Surgery Application

    23

    ∙ Moderately tightening the bolts and nuts

    ∙ Move the handle forward fixed

    2

    VERTICAL

    SURFACE MOUNT

    Combination

    ∙ Combine VERTICAL SURFACE MOUNT to TRAY HOLD MOUNT

    - Be careful with peeling of the paint on the arm

    ∙ Tightening a hexagon bolt with a 3mm Allen wrench

    ∙ Adjusted to the proper tension tighten the knob

    - Do not force excessively

    Shelves

    Combination

    ∙ The holes are positioned correctly

    ∙ Fixing with M3 X L6 bolts and nuts

    - The shelves should be firmly fixed so as not to lose.

  • Professional LED Light System for All of Surgery Application

    24

    Completion

    ∙ Fixed shelves OK

    8.4 WALL TYPE Installation

    No. Installation Guide

    1 Check packaging

    · Check the product releasing condition

    - If damage is found, contact DENTIS

    2 WALL BRAKET Combination

    · To fix the WALL BRAKET on the wall

    3 SECOND MAIN ARM Combination

    · Combining The SECOND MAIN ARM on the WALL BRAKET

  • Professional LED Light System for All of Surgery Application

    25

    - Avoid peeling of the paint on the arm

    4 FIRST MAIN ARM connecting · Connecting SECOND MAIN ARM and FIRST MAIN ARM chair arm (including head) · Check the wire(wire harnesses) condition and connecting

    - Avoid pressure of wire - Avoid peeling of the paint on the arm

    5 FIRST MAIN ARM Combination

    · Combining The FIRST MAIN ARM on The SECOND MAIN ARM

    - Avoid peeling of the paint on the arm

    6

    COAXAL CABLE Connecting (INTERNAL

    CAMERA Model ONLY

    · Connect the cable to the video transmission device

    - Avoid pressure of wire

    7 Completion

    · Handle R/L(Check direction of) combinations · Check the operation

  • Professional LED Light System for All of Surgery Application

    26

    8.5 CEILING TYPE Installation

    No. Installation Guide

    1 Check packaging

    · Check the product releasing condition

    - If damage is found, contact DENTIS

    2 CEILING VERTICAL

    ARM Combination

    · Check the hole location of CEILING VERTICAL ARM & BASE Hole

    - location of the holes must be careful not to play.

    · Fixed with 5mm allen wrench

    - Do not force excessively

    · Prepare a picture of nuts and bolts · Drill a hole with the hole expansion (Ø6.5)

    - Avoid peeling of the paint on the arm

  • Professional LED Light System for All of Surgery Application

    27

    · Fixed with bolt and nut · Ensure that the well fixed

    · Make sure the hex bolts loosened

    · Adjust the Wrench bolt left and right for horizontality

    · Check whether there’s no swing

    3 Wiring & Connecting

    · Harness into the CEILING VERTICAL ARM · If your INTERNAL CAMERA MODEL, putting together a video cable · And the harness, as shown in Figure

  • Professional LED Light System for All of Surgery Application

    28

    4 ARM BOLT assembly

    · Contents check · Insert the ring to Pivot connector · Insert joint holder apically

    · Check the product releasing condition

    - Attention orientation of the stopper

    5 ARM BOLT Combination · ARM BOLT Ass’y into the CEILING VERTICAL ARM · Fixed with 3mm allen wrench

    - If damage is found, contact DENTIS

    6 CEILING COVER Combination

    · SEILING COVER inserted into the CEILING VERTICAL ARM · Stopper O-ring insertion

  • Professional LED Light System for All of Surgery Application

    29

    · Move the o-ring to the position to stop the plastic cover · Ceiling cover have to cover all bracket and fit to ceiling side. · Fixed bolt with 3mm allen wrench

    7 SECOND MAIN ARM Combination

    · CEILING VERTICAL ARM Harness check · SECOND MAIN ARM Harness check

    · Harness connection check. · Fixed with 3mm allen wrench

    - Avoid pressure of wire

    8 FIRST

    MAIN ARM Combination

    · Connector direction check · FIRST MAIN ARM / SECOND MAIN ARM combination

    9 HEAD ARM Combination

    · Head wire and Video Cable connection

    - Avoid pressure of wire - Avoid peeling of the paint on the arm

  • Professional LED Light System for All of Surgery Application

    30

    10 Completion

    · Handle R/L(Check direction of) combinations · Check the operation

    9. Electrical connection

    WARNING The electrical connections must be performed by a authorized service personnel only.

    WARNING Check the polarity of all electrical connections before turning on the power. Insert the power cables carefully to protect against the risk of electrocution

    WARNING Do not remove covers or perform service other than as described in this equipment manual. Refer servicing to qualified service personnel.

    Power supply information

    § See 5.2 Electrical specification

  • Professional LED Light System for All of Surgery Application

    31

    9.1 Wiring Diagram

    DENTAL CHAIR TYPE

  • Professional LED Light System for All of Surgery Application

    32

    SWITCHING POWER ADAPTER TYPE

  • Professional LED Light System for All of Surgery Application

    33

    10. USE

    10.1 HEAD CONTROLLER

    Ÿ The LED DENTAL LIGHT can be controlled using the HEAD CONTROLLER

    Illumination Function

    Brightness Control

    - Dial type (option, remote controllable) - clockwise rotation ->

    Brightness up. - Counter clockwise roatation ->

    brightness down - 6 ~ 100% of control range.

    Color temp - 3type 4000K / 4500K / 5000K

    Camera (Option)

    Zoom IN/OUT - Camera ZOOM IN (+), ZOOM OUT(-)

    Adjustment of Brightness - Control the Brightness of camera

    Auto Focus - Control the focus of camera automatically.

    White Balance - Color temp of Picture from Camera

    Freeze - Temporarily Freeze the screen

    Reverse - Left & Right Reverse

  • Professional LED Light System for All of Surgery Application

    34

    10.2 Remote Control (Optional)

    Ÿ The LED DENTAL LIGHT can be controlled using the Remote Control

    The control panel of the Remote Control

    · Brightness control in 6levels

    · Other Functions will work same as control panel

    WARNING Don’t make the children handle this remote controller. Use a correct coin battery rated as DC 3V, CR2025.

  • Professional LED Light System for All of Surgery Application

    35

    10.3 RESIN COVER (Optional)

    WARNING Check the detachable way before using the product. Check for cracks. Make sure the correct position on the RESIN COVER.

    Easy removable magnetic structure

    Ÿ To block the light of the wavelength of 380 - 500nm also slow the cure rate of the resin

    10.4 HANDLE

    WARNING Check the attachment and detachment before using the product Check that the product is cracked and attached properly

    Removing by pulling the handle

    Mounting HANDLE Ÿ Insert the handle to fixed the mount . Ÿ Check Rotate to the left or right

    Remove the HANDLE Ÿ Pull the HANDLE to remove

  • Professional LED Light System for All of Surgery Application

    36

    10.5 SENSOR

    Motion Sensor

    On/off: sensing distance 2cm

  • Professional LED Light System for All of Surgery Application

    37

    10.6 Positioning

    Positioning the C500 LIGHTHEAD

    Ÿ Use the HANDLE to position the C500 LIGHTHEAD before the operation.

    Ÿ This HANDLE can be removed for sterilization.

    Ÿ Recommended distance : 70cm

  • Professional LED Light System for All of Surgery Application

    38

    11. Setting

    11.1 Adjusting the balancing of the ARM

    WARNING The TENSION ARM contains a powerful spring. When the LIGHTHEAD is dismantled, the SPRING ARM can suddenly jump up and may cause serious injury Make sure the apparent weight of the LIGHTHEAD is the same when raised and lowered and that it is stable in any position.

  • Professional LED Light System for All of Surgery Application

    39

    11.2 OperOperating range of HEAD ARM

    WARNING If this device is operated beyond the indicated angles, it can be damaged. During positioning, eventual collisions between the LIGHTHEAD, SPRING ARMs and other devices must be avoided.

    WARNING This image was constructed to demonstrate the product’s operating range. The image may differ from the actual product.manual.

  • Professional LED Light System for All of Surgery Application

    40

    11.3 Operating range of MAIN ARM

    CHAIR TYPE

    WARNING Do not place things on or hang on MAIN ARM.

    WARNING If this device is operated beyond the indicated angles, it can be damaged.

  • Professional LED Light System for All of Surgery Application

    41

    MOBILE TYPE

    WARNING Do not place things on or hang on MAIN ARM.

    WARNING If this device is operated beyond the indicated angles, it can be damaged.

  • Professional LED Light System for All of Surgery Application

    42

    WALL TYPE

    WARNING Do not place things on or hang on MAIN ARM.

    WARNING If this device is operated beyond the indicated angles, it can be damaged.

  • Professional LED Light System for All of Surgery Application

    43

    CEILING TYPE

    WARNING Do not place things on or hang on MAIN ARM.

    WARNING If this device is operated beyond the indicated angles, it can be damaged.

  • Professional LED Light System for All of Surgery Application

    44

    11.4 Adjust tension

    No. Problem Corrective action

    1

    Removal Arm Cover(common way)

    · Remove the cover using ‘-’ driver (3mm below)

    - Avoid peeling of the paint on the arm - Arm cover hook broken

    Common Tension control method

    · Lower FIRST MAIN ARM as much as possible · Fixing bolt with the same number of rotations (Use 3mm allen wrench)

    - In case up & down move is not natural, release damper bolt a little bit. - Hexagon nut broken

    In case of less tension

    · Lower FIRST MAIN ARM and tighten using allen wrench as much as possible and then loosen it a bit. · Tension control: - Tool: Used long nose Pliers - Strong : octagon nut is close from brass shaft - Week : octagon nut is far from brass shaft · Final Step - Lower FIRST MAIN ARM as much as possible - Loosen tension control bolt by pipe surface level

    - Hexagon nut broken

  • Professional LED Light System for All of Surgery Application

    45

    2 Head (1 axle) tension control · Remove left and right handles · Adjust proper tension with long nose plier

    - Do not force excessively Be careful not to change the direction of the handle

    3 Head (2 axle) tension control

    · Check the joint part of first head am and second head arm. · Working Method: Make Head Arm direction to the front side, make sure the side of Head Arm Hole and internal bolt position. · Adjusting tension by 2mm allen wrench.

    - Do not force excessively

    4 Head (3 axle) tension control

    · Check the joint part of Second Head Arm Body. · Working Method: Check the hexagon bolt fixation by spinning Head · Adjusting tension by 1.5mm allen wrench.

    - Do not force excessively

  • Professional LED Light System for All of Surgery Application

    46

    12. Model designation

    a: No. b: Type c: Sensor

    0=No Sensor 1=Sensor d: Head

    C50=C500

    e: Mount Type H=Head, C=Chair W=Wall, E=Ceiling M=Mobile

    f: Camera 0=No CAM, 1=Internal(10xZoom) 2=Internal(30xZoom)

    g: Colour S=Single, M=Multi

    h: Resin 0=No Resin 1=Resin

    i: R/C(Remote Control) A=No R/C B=R/C

    j: SWITCHING POWER ADAPTOR 0=No SPA, 1=SPA p: Coded Model Name

    a b c d CODE of OPTION p

    e f g h i j 1

    C500

    1 C50 H 0 M 0 A 0 C50-H0M0A0 2 1 C50 H 0 M 0 A 1 C50-H0M0A1 3 1 C50 H 0 M 0 B 0 C50-H0M0B0 4 1 C50 H 0 M 0 B 1 C50-H0M0B1 5 1 C50 H 0 M 1 A 0 C50-H0M1A0 6 1 C50 H 0 M 1 A 1 C50-H0M1A1 7 1 C50 H 0 M 1 B 0 C50-H0M1B0 8 1 C50 H 0 M 1 B 1 C50-H0M1B1 9 1 C50 H 1 M 0 A 1 C50-H1M0A1

    10 1 C50 H 1 M 0 B 1 C50-H1M0B1 11 1 C50 H 1 M 1 A 1 C50-H1M1A1 12 1 C50 H 1 M 1 B 1 C50-H1M1B1 13 1 C50 H 2 M 0 A 1 C50-H2M0A1 14 1 C50 H 2 M 0 B 1 C50-H2M0B1 15 1 C50 H 2 M 1 A 1 C50-H2M1A1 16 1 C50 H 2 M 1 B 1 C50-H2M1B1 17 1 C50 C 0 M 0 A 0 C50-C0M0A0 18 1 C50 C 0 M 0 A 1 C50-C0M0A1 19 1 C50 C 0 M 0 B 0 C50-C0M0B0 20 1 C50 C 0 M 0 B 1 C50-C0M0B1 21 1 C50 C 0 M 1 A 0 C50-C0M1A0 22 1 C50 C 0 M 1 A 1 C50-C0M1A1 23 1 C50 C 0 M 1 B 0 C50-C0M1B0 24 1 C50 C 0 M 1 B 1 C50-C0M1B1 25 1 C50 C 1 M 0 A 1 C50-C1M0A1 26 1 C50 C 1 M 0 B 1 C50-C1M0B1 27 1 C50 C 1 M 1 A 1 C50-C1M1A1 28 1 C50 C 1 M 1 B 1 C50-C1M1B1 29 1 C50 C 2 M 0 A 1 C50-C2M0A1 30 1 C50 C 2 M 0 B 1 C50-C2M0B1 31 1 C50 C 2 M 1 A 1 C50-C2M1A1 32 1 C50 C 2 M 1 B 1 C50-C2M1B1 33 1 C50 W 0 M 0 A 1 C50-W0M0A1 34 1 C50 W 0 M 0 B 1 C50-W0M0B1

  • Professional LED Light System for All of Surgery Application

    47

    35 1 C50 W 0 M 1 A 1 C50-W0M1A1 36 1 C50 W 0 M 1 B 1 C50-W0M1B1 37 1 C50 W 1 M 0 A 1 C50-W1M0A1 38 1 C50 W 1 M 0 B 1 C50-W1M0B1 39 1 C50 W 1 M 1 A 1 C50-W1M1A1 40 1 C50 W 1 M 1 B 1 C50-W1M1B1 41 1 C50 W 2 M 0 A 1 C50-W2M0A1 42 1 C50 W 2 M 0 B 1 C50-W2M0B1 43 1 C50 W 2 M 1 A 1 C50-W2M1A1 44 1 C50 W 2 M 1 B 1 C50-W2M1B1 45 1 C50 E 0 M 0 A 1 C50-E0M0A1 46 1 C50 E 0 M 0 B 1 C50-E0M0B1 47 1 C50 E 0 M 1 A 1 C50-E0M1A1 48 1 C50 E 0 M 1 B 1 C50-E0M1B1 49 1 C50 E 1 M 0 A 1 C50-E1M0A1 50 1 C50 E 1 M 0 B 1 C50-E1M0B1 51 1 C50 E 1 M 1 A 1 C50-E1M1A1 52 1 C50 E 1 M 1 B 1 C50-E1M1B1 53 1 C50 E 2 M 0 A 1 C50-E2M0A1 54 1 C50 E 2 M 0 B 1 C50-E2M0B1 55 1 C50 E 2 M 1 A 1 C50-E2M1A1 56 1 C50 E 2 M 1 B 1 C50-E2M1B1 57 1 C50 M 0 M 0 A 1 C50-M0M0A1 58 1 C50 M 0 M 0 B 1 C50-M0M0B1 59 1 C50 M 0 M 1 A 1 C50-M0M1A1 60 1 C50 M 0 M 1 B 1 C50-M0M1B1 61 1 C50 M 1 M 0 A 1 C50-M1M0A1 62 1 C50 M 1 M 0 B 1 C50-M1M0B1 63 1 C50 M 1 M 1 A 1 C50-M1M1A1 64 1 C50 M 1 M 1 B 1 C50-M1M1B1 65 1 C50 M 2 M 0 A 1 C50-M2M0A1 66 1 C50 M 2 M 0 B 1 C50-M2M0B1 67 1 C50 M 2 M 1 A 1 C50-M2M1A1 68 1 C50 M 2 M 1 B 1 C50-M2M1B1

  • Professional LED Light System for All of Surgery Application

    48

    13. List of component

    FIRST MAIN ARM SECOND MAIN ARM

    MOBILE VERTICAL ARM MOBILE BASE

    CASTER WALL BRAKET

    CEILING COVER CEILING VERTICAL ARM

    HANDLE ADAPTER

  • Professional LED Light System for All of Surgery Application

    49

    REMOTE CONTROL C500 LIGHT HEAD

    C500 LIGHT HEAD With RESIN COVER

    C500 LIGHT HEAD with RESIN COVER

    and INTERNAL CAMERA

  • Professional LED Light System for All of Surgery Application

    50