low background facilities at lsm

6
Low background facilities at LSM Pia Loaiza ILIAS meeting, Modane January 12 th 2005

Upload: rashad-compton

Post on 02-Jan-2016

18 views

Category:

Documents


0 download

DESCRIPTION

Low background facilities at LSM. Pia Loaiza. ILIAS meeting, Modane January 12 th 2005. Ge detectors at LSM. As S ~ /BDF, Gentiane and IRIS have roughly the same sensitivity at 583 keV and 1461 keV. At present, the sensitivity are, typically, 1 mBq/kg for “reasonables” - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Low background facilities at LSM

Low background facilities at LSM

Pia Loaiza

ILIAS meeting, Modane January 12th 2005

Page 2: Low background facilities at LSM

Ge detectors at LSM Detector Type Volume Total and peak background rate (counts/day)

40-2700 keV 352 keV 583 keV 1461 keV

MONDEUSE well 220 cc 770 4.2 2.7 5

ROUSSETE well 430 cc 692 4.1 2.9 7.2

ABYMES well 980 cc 828 5.6 5.6 5.6

XXL well 844 cc 821 6.8 <1.8 11.6

HERMINE N 197 cc 313 1.2 1.5 2.3

HELLAZ P 204 cc 515 4.5 0.3 1.4

JASMIN P 380 cc 529 2.0 1.41 1.71

GENTIANE N 215 cc 178 <0.21 0.38 0.65

IRIS P 400 cc 282 1.02 1.46 3.01

At present, the sensitivity are, typically, 1 mBq/kg for “reasonables”mass and measurement time.

As S ~ /BDF, Gentiane and IRIS have roughly the same sensitivity at 583 keV and 1461 keV.

Page 3: Low background facilities at LSM

On-going projects at the LSM

• Improvement of a low background Ge detector (IRIS, 400 cc).Prospective construction materials are being measured. Monte Carlo simulations are performed to get the distinct contribution to the background from the assembly materials and to determine the maximum acceptable radioactivities in order to get an improvement of a factor 5 in the detector background.We’ve found a high Th activity (0.13 Bq/kg @ 238 keV) in one of the insulators near the crystal, which need to be replaced by other material.

Page 4: Low background facilities at LSM

On-going projects at the LSM

• Cryogenerators : We plan to equip one Ge detector with a system of cryogenerators. Evaluation of possibilities of installation on oneof our detectors is under way.

• Construction of a Radon precision measuring facility, with a sensitivity of ~ mBq/m3 (Vol. 70 l)

• New planar Ge detector : our needs regarding low-energy gamma detectionrequire a new planar Ge (diam = 80 mm, d=35 mm), with a resolution of 0.7 keV at 50-60 keV. Performances of detectors in the market are being evaluated.

Page 5: Low background facilities at LSM

Radon trapping facility

Installation September 2004

Page 6: Low background facilities at LSM

Radon trapping facility

Time during which radon is trappedin the « filter » : Tr = K(T) M / Q

K(T) : dynamic adsorption coeff.M : charcoal massQ : Air rate

2 charcoal towers : T1 = -53 C, T2 = -45 CM = 2 x 450 kg

Vol. clean air : 150 m3/hActivity 222Rn = 15 mBq/m3

Radon purification by adsorption in charcoal