local corneal thickness changes after small-incision cataract surgery
TRANSCRIPT
J CATARACT REFRACT SURG - VOL 32, OCTOBER 2006
Local corneal thickness changes
after small-incision cataract surgery
Matthias Bolz, MD, Stefan Sacu, MD, Wolfgang Drexler, PhD, Oliver Findl, MD
PURPOSE: To assess whether a temporal limbal–corneal incision approach for phacoemulsificationcataract surgery induces a gradient in corneal thickening along the horizontal meridian.
SETTING: Department of Ophthalmology, Medical University of Vienna, Vienna, Austria.
METHODS: Corneal thickness in 21 eyes of 21 patients was measured preoperatively as well as 1 day, 1week, and 1 and 3 months after phacoemulsification through a temporal limbal–corneal incision. Mea-surements were performed using partial coherence interferometry (PCI) with a commercial instrument,the ACMaster (Carl Zeiss Meditec, Jena). Measurements were taken along the horizontal meridian cen-trally along the visual axis at 1.5 mm, 3.0 mm, and 4.5 mm eccentricity.
RESULTS: Preoperatively, there were slight nasal–temporal differences in corneal thickness at alleccentricities. The mean thickness was 522 mm G 34 (SD) at 1.5 mm nasally and 513 G 36 mm at1.5 mm temporally (P<.01). On day 1, there was a significant mean increase in corneal thickness(38 G 43 mm) along all locations. The thickening was slightly more pronounced in the peripherythan in the center, a difference not reaching statistical significance. At 1 week, corneal thickness re-turned almost to baseline at all locations except for 3.0 mm temporally, where it was slightly, butnot significantly, thicker (mean 8 G 14 mm). At 1 month, corneal thickness at the 3.0 mm temporallocation returned to baseline.
CONCLUSIONS: A nasal–temporal difference in corneal thickness was found preoperatively in all pa-tients. Phacoemulsification through a temporal limbal–corneal incision caused an increase in cornealthickness along the horizontal meridian 1 day after surgery. The prolonged corneal thickening at3.0 mm eccentricity temporally could be a result of the proximity to the incision site.
J Cataract Refract Surg 2006; 32:1667–1671 Q 2006 ASCRS and ESCRS
Phacoemulsification through a temporal clear corneal or
limbal–corneal incision using topical anesthesia is widely
established in cataract surgery. Short recovery,1 low in-
duced astigmatism and good early visual acuity,2 low inci-
dence of surgical complications,2–4 and less postoperative
Accepted for publication May 17, 2006.
From the Department of Ophthalmology (Bolz, Sacu, Findl) andthe Zentrum fur Biomedizinische Technik und Physik (Drexler),Medical University of Vienna, Vienna, Austria.
Presented in part at the XXIInd Congress of the European Societyof Cataract & Refractive Surgeons, Paris, France, September 2004.
Dr. Drexler is a consultant to Carl Zeiss Meditec AG. No other au-thor has a proprietary or financial interest in any material ormethod mentioned.
Corresponding author: Oliver Findl, MD, Department of Ophthal-mology, Medical University of Vienna, Wahringer Gurtel 18-20,Vienna A-1090, Austria. E-mail: [email protected].
Q 2006 ASCRS and ESCRS
Published by Elsevier Inc.
intraocular pressure elevation than with sclerocorneal inci-
sions5 have been reported.
However, phacoemulsification damages the corneal
endothelium. Several authors6–8 have measured endothe-
lial cell loss caused by irrigation and the heat generated
by phacoemulsification devices, which reduces corneal
metabolism postoperatively.9 The extent of postoperativeendothelial cell loss depends on intraoperative phacoemul-
sification time.7 In addition, shorter eyes have a signifi-
cantly higher incidence of endothelial cell loss,7 probably
because of a more shallow anterior segment.
Measuring corneal thickness changes after phacoemul-
sification has been used to assess surgically induced corneal
edema. Lundberg et al.10 report that the degree of perma-
nent corneal endothelial damage is related to the degreeof early postoperative corneal swelling. Thus, measuring
the differences in pachymetry postoperatively seems useful
in assessing the effects of phacoemulsification on corneal
integrity.10
0886-3350/06/$-see front matterdoi:10.1016/j.jcrs.2006.05.018
1667
CORNEAL THICKNESS CHANGES AFTER CATARACT SURGERY
Partial coherence interferometry (PCI)11–14 is an ap-
propriate technique for high-precision, noncontact corneal
thickness measurements.11 A prototype of the ACMaster
(Carl Zeiss Meditec, Jena) used in this study allows assess-
ment of corneal thickness at several eccentricities to the
corneal center. We used PCI to identify a possible gradientin corneal thickening induced by a temporal clear corneal
approach during cataract surgery. The aim was to evaluate
local postoperative corneal thickness changes over time
caused by phacoemulsification via a temporal clear limbal
approach.
PATIENTS AND METHODS
The study was conducted at the Department of Ophthalmol-ogy, Medical University of Vienna, Vienna, Austria. All researchand measurements followed the tenets of the Helsinki agreement,and all patients provided informed consent.
The study’s 21 patients (21 eyes) were recruited from a con-tinuous cohort. The inclusion criterion was age-related cataract.Exclusion criteria were a history of ocular disease or intraocularsurgery, laser treatment, diabetes requiring medical control, andglaucoma.
Cataract surgery was performed by the same surgeon (O.F.)using the same technique comprising phacoemulsification andtopical anesthesia of lidocaine 4% eyedrops. A temporal, single-plane, 3.2 mm limbal–corneal incision was created. A dispersiveophthalmic viscosurgical device (OVD), hydroxypropyl methyl-cellulose 2% (Medioclear), was injected and a capsulorhexisperformed. After thorough hydrodissection, the nucleus wasemulsified with an Orbit unit (Oertli Instrumente AG) using anab interno cracking technique, after which coaxial aspiration ofthe cortical remnants was done. After the anterior chamber wasexpanded with a cohesive OVD (Healon [sodium hyaluronate1%]), a foldable hydrophobic acrylic intraocular lens (IOL) wasimplanted in the capsular bag with an injector system. The OVDwas aspirated thoroughly from the retrolental space and the ante-rior chamber using an irrigation/aspiration tip. The main incisionand the paracenteses were hydrated with balanced saline solution(BSS) through a thin cannula.
Corneal thickness was measured preoperatively as well as1 day, 1 week, and 1 and 3 months after surgery by PCI using theACMaster. Measurements were performed along the horizontalmeridian centrally along the visual axis and at 1.5 mm, 3.0 mm,and 4.5 mm eccentricity. Mean values and standard deviationswere generated. The principles11–14 and the reproducibility of
J CATARACT REFRACT SURG1668
corneal thickness measurements15 of PCI have been reported indetail. Eccentric measurements were performed by having the ex-amined eye of the patient fixate on a target and eccentric LEDs in-corporated in the ACMaster.
To evaluate the influence of phacoemulsification on postop-erative corneal thickening, phacoemulsification time was corre-lated to the assessed relative changes in corneal thickness on thefirst postoperative day. Reproducibility was determined with re-peated measurements in 8 eyes taken at 10-minute intervals.
All other evaluation was done on a personal computer usingstandard software (MS Excel and SPSS; Sigmaplot 8.0, Systat Soft-ware Inc.). To compare changes in corneal thickness over time, thepaired t test was used. The Bonferroni adjustment was applied tomultiple t tests. A P value less than 0.05 was considered statisti-cally significant.
RESULTS
The mean age of the patients was 71.8 years G 10.8 (SD);
the median age was 73.5 years.
Preoperatively, there was a slight nasal–temporal dif-
ference in corneal thickness at all eccentricities (Table 1
and Figure 1). Corneas were slightly thicker in the nasalhalf along the horizontal meridian, reaching statistical sig-
nificance at the 1.5 mm eccentricity (P!.01). This signifi-
cant difference at 1.5 mm was found at all follow-up
measurements except the first day after surgery (P!.06)
(1 week, P!.01; 1 month, P!.05; 3 months, P!.01).
On the first postoperative day, the mean increase in
corneal thickness at all locations was 38 G 43 mm. The
thickening was more pronounced in the peripheral corneathan centrally; the mean was 31 G 42 mm centrally, 58 G55 mm 4.5 mm temporally, and 35 G 38 mm nasally. How-
ever, the differences in thickening between locations were
not statistically significant (4.5 mm temporal/central:
P Z.09; 4.5 mm nasal/central: P Z.24). At 1 week, corneal
thickness returned to baseline at all locations; the mean in-
crease at all locations compared to baseline was 7 G 53 mm.
Only at the 3.0 mm temporal location was corneal thick-ness slightly higher (mean 8 G 14 mm versus baseline),
although this did not reach statistical significance. At
1 month, corneal thickness at the 3.0 mm location returned
Table 1. Mean absolute corneal thickness at all temporal and nasal locations over time.
Mean Thickness (mm) G SD
Exam4.5 mm
Temporally3.0 mm
Temporally1.5 mm
Temporally Centrally1.5 mmNasally
3.0 mmNasally
4.5 mmNasally
Preop 529.11G 32.60 520.37 G 34.62 512.81 G 35.74 517.71 G 36.64 521.67 G 34.02 526.35 G 34.88 550.31 G 29.43Postop
Day 1 576.21 G 73.00 546.20 G 58.81 555.14 G 47.86 549.14 G 56.52 561.29 G 53.54 564.57 G 55.02 586.75 G 70.90Week 1 529.79 G 57.36 526.38 G 36.71 517.43 G 34.27 515.86 G 36.81 528.00 G 36.58 521.35 G 50.50 529.33 G 43.55Month 1 526.20 G 37.56 516.19 G 34.60 512.14 G 32.81 513.62 G 32.52 519.67 G 40.36 519.31 G 38.87 516.27 G 72.08Month 3 530.20 G 36.12 520.52 G 30.72 514.57 G 33.07 519.76 G 48.84 527.14 G 33.46 532.26 G 38.01 528.82 G 46.22
- VOL 32, OCTOBER 2006
CORNEAL THICKNESS CHANGES AFTER CATARACT SURGERY
to baseline. Relative local changes in corneal thickness(3.0 mm temporal, central, and 3.0 mm nasal) over time
are shown in Figure 2. The mean difference in corneal
thickness compared to baseline was 1 G 42 mm, at 1 month
and 6 G 41 mm at 3 months. Absolute corneal thickness
values at all eccentricities along the horizontal meridian
at day 0, day 1, and week 1 are shown in Figure 3.
At baseline and all follow-up examinations, it was not
possible to assess reliable corneal thickness values at
Figure 1. Differences in corneal thickness between corresponding tem-
poral and nasal eccentricities along the horizontal meridian at baseline,
before surgery. Box plots represent absolute values in millimeters at the
following eccentricities: 1.5 mm, 3.0 mm, and 4.5 mm. Dotted boxes rep-
resent temporal values; lined boxes represent nasal values.
J CATARACT REFRACT SURG -
4.5 mm eccentricity in 9 eyes and at 3.0 mm in 6 eyes,
mainly because of the presence of slight age-related corneal
opacities such as arcus senilis or postoperative corneal al-
terations. Thus, to avoid statistical bias, changes in corneal
thickness were assessed by case-wise deletion of missing
values. This is why the values of relative changes in cornealthickness reported above differ from the absolute values in
Table 1.
At the 3.0 mm and 1.5 mm eccentricities, corneal
thickness could be assessed precisely. There was a high
intraobserver reproducibility centrally (1.17 G 0.17 mm),
at 1.5 mm (temporally: 1.08 G 0.37 mm; nasally: 1.42 G0.59 mm) and at 3.0 mm (temporally: 0.58 G 0.73 mm; na-
sally: 1.7 G 0.77 mm). At 4.5 mm eccentricity, the repro-ducibility was lower (temporally: 6.25 G 0.09 mm;
nasally: 4.00 G 0.66 mm), which could have been caused,
in part, by the peripheral opacities.
The mean phacoemulsification time was 42.10 G24.70 seconds (median 35 seconds). No significant correla-
tions were found between the phacoemulsification time
and the relative change in corneal thickness at day 1
(r2 Z .06; data not shown).
DISCUSSION
Phacoemulsification surgery via a temporal clear lim-
bal approach caused an increase in corneal thickness at
all locations along the horizontal meridian 1 day after
surgery. One week after surgery, corneal thickness values
returned to baseline at nearly all eccentricities except3.0 mm temporally. Therefore, in normal uneventful cata-
ract surgery, corneal edema is initially diffuse across the en-
tire cornea, with similar increases in corneal thickness. The
ACMaster, which is based on the precise and reproduc-
ible15 PCI technique, was an appropriate tool to evaluate
Figure 2. Relative change in corneal thickness (mean G SD) in millimeters over time presented separately for the following eccentricities: 3.0 mm temporal,
central and 3.0 mm nasal. Measurements were performed 1 day postoperatively (d1) and after 1 week (w1), 1 month (m1), and 3 months (m3).
VOL 32, OCTOBER 2006 1669
CORNEAL THICKNESS CHANGES AFTER CATARACT SURGERY
this corneal stress after cataract surgery. In addition, the
technique showed statistically significant nasal–temporaldifferences in corneal thickness preoperatively.
The postoperative corneal edema in our study could
have been the result of several factors that cause mechani-
cal stress such as turbulent irrigation fluids and nuclear
fragments during phacoemulsification or heat transfer
from the phaco tip. In addition, the intraocular irrigation
fluid, which has a slightly different pH, osmolarity, and
ion constitution than aqueous, could have caused cellularstress in the cornea. The prolonged thickening near the in-
cision site at 3.0 mm eccentricity (temporal) may be a result
of the mechanical stress exerted from the incision through
Descemet’s membrane, the proximity to the hot phaco tip,
or the higher mechanical stress caused by turbulence due to
inflow and outflow near the incision. This could also be
caused by edema from the injection of BSS in the corneal
stroma at the end of cataract surgery.The corneal edema caused by these factors was mea-
sured with the ACMaster. This device was appropriate for
investigating postoperative corneal thickness changes in el-
derly patients preoperatively and immediately postopera-
tively in a clinical setting. It allowed measurement of
several preset eccentricities by moving a visual target on
which the patient was asked to fixate. This technical feature
led to the observation of a slight nasal–temporal gradient incorneal thickness along the horizontal meridian. This gra-
dient was also detected before surgery.
Several studies describe local (preoperative) differences
in corneal thickness using several techniques such as
ultrasound,16–20 laser pachymetry,16 scanning slit-beam
Figure 3. Absolute corneal thickness values in millimeters at the following
temporal and nasal eccentricities: 4.5 mm, 3.0 mm, 1.5 mm, and centrally
(center). The dotted line represents preoperative values. The broken line
represents values at day 1. The continuous line represents values 1 week
after surgery. Mean values and standard error of the mean are depicted.
J CATARACT REFRACT SURG -1670
(Orbscan),18,21–23 confocal microscopy,20 PCI,15 or
Scheimpflug photography.18 Remon et al.17 report the cor-
nea is significantly thinner in the superior area in newborns.
In studies by Liu and Pflugfelder22 and Liu et al.,23 however,
the superior corneal area was found to be the thickest. Thus,
results in studies of corneal thickness and its changes overtime are highly variable, with some studies finding contra-
dicting results. This could, in part, be due to the difficulty
of taking into account all influencing factors such as physi-
ologic diurnal thickness fluctuations,21 corneal astigma-
tism, effects of a dry-eye syndrome22 or, more important,
the measurement technique.
Using PCI, a well-established technique for corneal pa-
chymetry, we showed that measuring different corneal loca-tions is feasible in elderly patients in the perioperative
setting. Slight corneal opacities, such as arcus senilis or cor-
neal edema in or close to the temporal incision area, atten-
uated the measurements and resulted in failures at the
4.5 mm eccentricity as a result of a low signal-to-noise ra-
tio. The PCI technique also has better interexaminer and
intraexaminer reproducibility than most other methods.
Several authors6,9 report that corneal thickeningcaused by phacoemulsification is reversible. Grupcheva
et al.24 found that postoperative corneal thickness in
the meridian closest to the incision measured with Orbs-
can technique was significantly higher than the mean
value of 12 meridians. This agrees with our results show-
ing mild edema at the 3.0 mm eccentricity 1 week after
surgery.
Apart from corneal thickening at the incision site,Lundberg et al.10 found strong correlations between total
postoperative endothelial cell loss and corneal swelling.
The degrees of permanent corneal endothelial damage
were reported to be reflected in the degree of early postop-
erative corneal thickening. Hence, investigating the
changes in corneal thickness seems appropriate when eval-
uating corneal stress during phacoemulsification surgery in
clinical trials in which OVDs or surgical techniques arecompared.
In conclusion, our study showed PCI to be appropriate
for assessing a preoperative nasal–temporal difference in
corneal thickness along the horizontal meridian as well as
postoperative changes in corneal edema precisely and re-
producibly. In all eyes, phacoemulsification seemed to
cause a significant alteration of the corneal layers that led
to edema. Nevertheless, these pachymetric changes werereversible, almost reaching baseline values 1 week after
surgery.
REFERENCES
1. Masket S, Tennen DG. Astigmatic stabilization of 3.0 mm temporal clear
corneal cataract incisions. J Cataract Refract Surg 1996; 22:1451–1455
VOL 32, OCTOBER 2006
CORNEAL THICKNESS CHANGES AFTER CATARACT SURGERY
2. Lyle WA, Jin GJC. Prospective evaluation of early visual and refractive
effects with small clear corneal incision for cataract surgery. J Cataract
Refract Surg 1996; 22:1456–1460
3. Dick HB, Schwenn O, Krummenauer F, et al. Inflammation after sclero-
corneal versus clear corneal tunnel phacoemulsification. Ophthalmol-
ogy 2000; 107:241–247
4. Monica ML, Long DA. Nine-year safety with self-sealing corneal tunnel
incision in clear cornea cataract surgery. Ophthalmology 2005;
112:985–986
5. Schwenn O, Dick HB, Krummenauer F, et al. Intraocular pressure
after small incision cataract surgery: temporal sclerocorneal ver-
sus clear corneal incision. J Cataract Refract Surg 2001; 27:421–
425
6. Zetterstrom C, Laurell C-G. Comparison of endothelial cell loss and
phacoemulsification energy during endocapsular phacoemulsifica-
tion surgery. J Cataract Refract Surg 1995; 21:55–58
7. Walkow T, Anders N, Klebe S. Endothelial cell loss after phacoemulsi-
fication: relation to preoperative and intraoperative parameters. J Cat-
aract Refract Surg 2000; 26:727–732
8. Dick B, Kohnen T, Jacobi KW. Endothelzellverlust nach Phakoemulsifi-
kation und 3, 5 vs. 5 mm Hornhauttunnelinzision. Ophthalmologe
1995; 92:476–483
9. Kohlhaas M, Stahlhut O, Tholuck J, Richard G. Entwicklung der Horn-
hautdicke und -endothelzelldichte nach Kataraktextraktion mittels
Phakoemulsifikation. Ophthalmologe 1997; 94:515–518
10. Lundberg B, Jonsson M, Behndig A. Postoperative corneal swelling
correlates strongly to corneal endothelial cell loss after phacoe-
mulsification cataract surgery. Am J Ophthalmol 2005; 139:1035–
1041
11. Drexler W, Baumgartner A, Findl O, et al. Submicrometer precision bi-
ometry of the anterior segment of the human eye. Invest Ophthalmol
Vis Sci 1997; 38:1304–1313
12. Drexler W, Findl O, Menapace R, et al. Partial coherence interferome-
try: a novel approach to biometry in cataract surgery. Am J Ophthal-
mol 1998; 126:524–534
J CATARACT REFRACT SURG -
13. Fercher AF, Hitzenberger CK, Drexler W, et al. In vivo optical coherence
tomography [letter]. Am J Ophthalmol 1993; 116:113–114
14. Drexler W, Baumgartner A, Findl O, et al. Biometric investigation of
changes in the anterior eye segment during accommodation. Vision
Res 1997; 37:2789–2800
15. Rainer G, Findl O, Petternel V, et al. Central corneal thickness measure-
ments with partial coherence interferometry, ultrasound, and the
Orbscan system. Ophthalmology 2004; 111:875–879
16. Gritz DC, McDonnell PJ. Comparison of a computer-assisted laser
pachometer with two ultrasonic pachometers in normal corneas. Re-
fract Corneal Surg 1990; 6:9–14
17. Remon L, Cristobal JA, Castillo J, et al. Central and peripheral corneal
thickness in full-term newborns by ultrasonic pachymetry. Invest
Ophthalmol Vis Sci 1992; 33:3080–3083
18. Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibil-
ity of central corneal thickness measurement with Pentacam, Orbs-
can, and ultrasound. Optom Vis Sci 2005; 82:892–899
19. Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measure-
ment with the Pentacam Scheimpflug system, optical low-coherence
reflectometry pachymeter, and ultrasound pachymetry. J Cataract
Refract Surg 2005; 31:1729–1735
20. McLaren JW, Nau CB, Erie JC, Bourne WM. Corneal thickness measure-
ment by confocal microscopy, ultrasound, and scanning slit methods.
Am J Ophthalmol 2004; 137:1011–1020
21. Lattimore MR Jr, Kaupp S, Schallhorn S, Lewis RIV. Orbscan pachyme-
try; implications of a repeated measures and diurnal variation analysis.
Ophthalmology 1999; 106:977–981
22. Liu Z, Pflugfelder SC. Corneal thickness is reduced in dry eye. Cornea
1999; 18:403–407
23. Liu Z, Huang AJ, Pflugfelder SC. Evaluation of corneal thickness and
topography in normal eyes using the Orbscan corneal topography
system. Br J Ophthalmol 1999; 83:774–778
24. Grupcheva CN, Riley AF, Craig JP, et al. Analyzing small-incision cata-
ract surgery by Orbscan II fourth-dimensional pachymetry mapping.
J Cataract Refract Surg 2002; 28:2153–2158
VOL 32, OCTOBER 2006 1671