    LENSMETER

    ModelLM-770

    OPERATORS MANUAL

  • 7/24/2019 LM-770_OME - MANUAL

    2/40

    NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan

    (Manufacturer) Telephone: (0533) 67-6611

    Facsimile: (0533) 67-6610

    NIDEK CO., LTD : 6th Floor, Takahashi Bldg., No.2, 3-chome, Kanda-jinboucho

    (Tokyo Office) Chiyoda-ku, Tokyo 101-0051, Japan

    Telephone: (03) 3288-0571

    Facsimile: (03) 3288-0570

    Telex: 2226647 NIDEK J

    NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U. S. A.

    (United States Agent) Telephone: (510) 226-5700Facsimile: (510) 226-5750

    NIDEK SOCIETE ANONYME : Europarc 13, rue Auguste Perret, 94042 CRETEIL, France

    (Authorized Representative) Telephone: (01) 49 80 97 97

    Facsimile: (01) 49 80 32 08

    June 200630283-P902E

    Printed in JAPAN

  • 7/24/2019 LM-770_OME - MANUAL

    3/40

    BEFORE USE, READ THIS MANUAL.

    This manual contains information necessary for correct and effective operation of NIDEK

    Projection and Digital type Lensmeter Model LM-770, and provides procedures for

    operation, cautions, specification, and procedures for maintenance.

    In this manual, IEC 60601 is applied.

    The lensmeter complies with ISO 8598 (Optics and optical instruments - Focimeters).

    For correct use, it is necessary that this manual, in particular the operating procedures,

    be thoroughly understood before using this instrument.

    Keep the manual handy to verify use.

    There are no user-serviceable parts inside the instrument except lamps, fuses, printer

    papers and ink cartridge. If you find any problem or question about the instrument

    during the operation, please contact your authorized distributor.

    The LM-770 is calibrated with reference wavelength of 587.56 nm.

  • 7/24/2019 LM-770_OME - MANUAL

    4/40

    Table of Contents

    1 INTRODUCTION ........................................................................................................ 1-11.1 Outline ...................................................................................................................... 1-1

    1.2 Indications for Use ................................................................................................... 1-1

    1.3 Classifications ........................................................................................................... 1-1

    1.4 Symbol Information .................................................................................................. 1-2

    2 SAFETY ........................................................................................................................ 2-12.1 Operation ..................................................................................................................2-1

    2.2 Storage ..................................................................................................................... 2-2

    2.3 Transference ............................................................................................................. 2-2

    2.4 Installation ................................................................................................................ 2-2

    2.5 Wiring ....................................................................................................................... 2-3

    2.6 After Use .................................................................................................................. 2-4

    2.7 Maintenance and Checks .......................................................................................... 2-4

    2.8 Disposal .................................................................................................................... 2-4

    2.9 Labels ....................................................................................................................... 2-5

    3

    CONFIGURATIONS ................................................................................................... 3-1

    4OPERATING PROCEDURES.................................................................................... 4-14.1 Basic Process of Operation (SPH, CYL Lenses) ...................................................... 4-1

    4.2 Prior Measurement ................................................................................................... 4-4

    4.2.1 Preparation ...................................................................................................... 4-4

    4.2.2 How to set the lens .......................................................................................... 4-4

    4.3 Measurement ............................................................................................................ 4-5

    4.3.1 Measurement of single vision lens ................................................................... 4-5

    4.3.2 Measurement of ADD diopter ......................................................................... 4-6

    4.3.3 Measurement of prism refractive power .......................................................... 4-7

    4.3.4 Measurement of hard contact lens ................................................................... 4-9

    4.4 Marking Axis .......................................................................................................... 4-10

    4.4.1 Making sphere lens ........................................................................................ 4-10

    4.4.2 Marking cylinder axis .................................................................................... 4-10

    4.4.3 Marking phoria lens....................................................................................... 4-11

    4.5 Memory Function ................................................................................................... 4-12

    4.5.1 How to memorize .......................................................................................... 4-12

    4.5.2 How to display the contents of memory........................................................ 4-12

    4.5.3 How to clear the memory .............................................................................. 4-13

    4.6 Print out .................................................................................................................. 4-134.7 Setting Step and Changing the Setting of CYL Reading ........................................ 4-14

    4.7.1 How to set the step ....................................................................................... 4-14

    4.7.2 How to set CYL reading ............................................................................... 4-14

    Page

  • 7/24/2019 LM-770_OME - MANUAL

    5/40

    4.8 Auto-off Function ................................................................................................... 4-15

    5 MAINTENANCE ......................................................................................................... 5-15.1 Replace the Lamp ..................................................................................................... 5-1

    5.2 Replace the Fuses ..................................................................................................... 5-2

    5.3 Replace the Ink Cartridge ......................................................................................... 5-3

    5.4 Replace the Printer Paper ......................................................................................... 5-4

    5.5 Cleaning of Protective Glass ..................................................................................... 5-4

    5.6 Cleaning Exterior ...................................................................................................... 5-5

    6SPECIFICATIONS ...................................................................................................... 6-1

    Page

  • 7/24/2019 LM-770_OME - MANUAL

    6/40

    [This page is intentionally left blank.]

  • 7/24/2019 LM-770_OME - MANUAL

    7/40

    1INTRODUCTION

    1.1 Outline

    NIDEK MODEL LM-770 is a projection and digital type lensmeter which is developed for moreusers to use the product of high performance.

    This instrument projects the target brightly and sharply on the screen.

    Also, easy and accurate lens measurement is possible. The measurement results are shown in the data

    of sphere power, cylinder power, axis angle, addition (ADD) power and prism power.

    The features is

    Measured refractive powers and axis are displayed digitally and easy to read.

    By the equipment of big screen, bright and sharp target can be obtained.

    Maximum measurable lens diameter is up to 100mm.

    Auto-off function gives longer life time of the lamp and save energy of electricity.

    Measured refractive powers and axis can be printed out. Measured value of left and right lenses in frame can be put in memory and also can be printed

    out.

    1.2 Indications for Use

    This equipment is designed to measure vertex powers and prismatic effects of spectacle and contact

    lens, to orientate and mark uncut lenses, and to verify the correct mounting of lenses in spectacles

    frames.

    1.3 Classifications

    [Classification under the provision of 93/42/EEC(MDD)] Class I

    The LM-770 is classified into Class I device.

    [Form of protection against electric shock] Class I

    The LM-770 is classified as a Class I equipment.

    The class I equipment is an equipment in which the protection against electric shock does not rely

    on basic insulation only, but which includes an additional safety precaution that provides for a

    connection of accessible conductive parts to the protective (earth) conductor in the fixed wiring of

    the installation. Therefore, accessible conductive parts cannot become activated in the event of

    failure in the basic insulation.

    [Degree of protection against electric shock] Type B Applied Part

    The LM-770 is provided with a Type B Applied Part.

    The Type B Applied Part provides an adequate degree of protection against electrical shock,

    particularly regarding the following:

    - allowable leakage currents

    - reliability of the protective earth ground connection (If applicable).

  • 7/24/2019 LM-770_OME - MANUAL

    8/40

    1 - 2

    [Degree of protection against liquid entry] IP20

    The LM-770 is classified as a ordinary equipment (IP20), which is provided with an enclosure not

    completely water proof.

    Avoid water splash on the equipment.

    [Method(s) of sterilization or disinfection recommended by the manufacturer]

    The LM-770 does not have any part to be sterilized or be disinfected.

    [Degree of protection against flammability]

    The LM-770 is classified as an equipment not suitable to be used in a potentially flammable

    environment.

    Do not use near flammable materials.

    [Operation mode]

    The LM-770 is classified as an equipment which operates continuously.

    1.4 Symbol Information

    This symbol on the body indicates that caution should be taken. Refer to the operators

    manual before use.

    This symbol indicates that the degree of protection against electrical shock of the applied

    part is classified as a Type B applied part.

    This symbol on the power switch indicates that the power is ON.

    This symbol on the power switch indicates that the power is OFF.

    This symbol means the fuse rating.

    This symbol indicates the input.

    This symbol indicates the output.

    This symbol indicates that the alternating current should be applied to the instrument.

    This symbol indicates the date of manufacture.

    This symbol indicates that this product shall be disposed of in a separate collection of electrical

    and electronic equipment in EU.

  • 7/24/2019 LM-770_OME - MANUAL

    9/40

    2SAFETY

    In this manual, a Signal Word is used to designate a degree or level of safety alerting,

    whose definitions are as follows.

    CAUTION: Indicates a potentially hazardous situation which, if not avoided,

    may result in minor or moderate injury or property damage.

    Even the cases mentioned in CAUTIONmay result in serious injury under a certain

    situation. Be sure to observe the instructions of CAUTION.

    2.1 Operation

    CAUTION

    Never disassemble or touch the inside of the instrument.

    This may result in an electric shock or instrument malfunction.

    Do not touch the lamp house cover since it is hot while a lamp is illuminated.

    If not, you may get burned.

    Never yank the power cord to disconnect from wall outlet but hold the plug whiledisconnecting.

    This can weaken the metal core of the cord and may result in a short circuit or an

    electric shock.

    Do not place the instrument on the power cord.

    The sheath of the cord becomes ragged and may cause an electric shock or a short

    circuit.

    If the metal core of the power cord is exposed, power turns on and off by shaking the

    cord, or cord/plug gets so heated that one cannot hold it, it means that the cord is damaged.Immediately replace the cord.

    This may cause an electric shock or a fire.

    Do not use the instrument for other than the intended purpose.

    NIDEK will not be responsible for accidents or malfunction caused by carelessness.

    In the event of a malfunction, do not touch the inside of the instrument, but unplug the

    power cord and contact your authorized distributor.

  • 7/24/2019 LM-770_OME - MANUAL

    10/40

    2 - 2

    NOTE

    Before measurement, be sure that the nosepiece is mounted correctly.

    Incorrect mount of the nosepiece may on accuracy of measurement.

    2.2 Storage

    NOTE

    Do not store the instrument in a place where it may get wet or where poisonous gas or

    liquid is stored.

    Avoid storing the instrument in an area with excessive heat, humidity, or dust.

    To preserve the appearance or internal parts of the instrument, avoid direct exposure to

    sunlight.

    2.3 Transference

    CAUTION

    Never trail the power cable during the transference.The instrument may fall down and cable may break when it is stepped on or pulled.

    In transference, hold the instrument surely with both hands, and take care not to bump

    onto other instruments, walls and others.

    Careless transference may cause injury or malfunction.

    2.4 Installation

    CAUTION

    Do not install the instrument near water.

    If water gets into the internal structure, there is the possibility of electrical shock or

    instrument malfunction.

  • 7/24/2019 LM-770_OME - MANUAL

    11/40

    2 - 3

    CAUTION

    Install the instrument in a stable and level place where vibration or shock does not occur.

    The instrument may not perform observation correctly or may malfunction.

    Also, if the instrument is tripped over because of any accidental shock, it

    may result in possible injury.

    NOTE

    Install the instrument in the following conditions.

    A low dust environment

    A low light interference environment

    A vibration and shock free environment

    Install the instrument in the environment whose condition can be set to the following.

    Use the instrument under the following temperature and humidity.

    Use environment Temperature : +10 to +40 C

    Humidity : 30 to 85% (non-condensing)

    2.5 Wiring

    CAUTION

    Be sure to use the wall outlet which meets the requirements of power specification.

    If the line voltage is too high or too low, the instrument may not give full performance.

    Malfunction or a fire also may occur.

    Be sure to use a power outlet equipped with a grounding terminal.

    Otherwise, this may cause electric shock when it breaks or power leaks.

    Be sure to securely plug in the power cord.Insecure connections may cause a fire.

    Be sure to securely plug in the interface cable.

    Insecure connections may cause malfunction.

  • 7/24/2019 LM-770_OME - MANUAL

    12/40

    2 - 4

    2.6 After Use

    CAUTION If the instrument is not be used for a long time, disconnect the power cord from the wall outlet.

    This may cause a fire.

    NOTE

    If the instrument will not be used during an extended period of time, turn off the main

    switch and put the dust cover over the instrument.

    If not, settled dusts may impair visibility through the instrument.

    2.7 Maintenance and Checks

    CAUTION

    As the illumination lamp can get quite hot, replace it when it is cool.

    If not, you may get burned.

    NOTE

    Disassembling for service is not permitted except at NIDEK or authorized distributor.

    Never use an organic solvent such as paint thinner to wipe the exterior.

    This may ruin the surface.

    2.8 Disposal

    NOTE

    Follow local governing ordinances and recycling plans regarding disposal or recycling of

    device components.

    When disposing packing materials, sort them by the materials and follow local governing

    ordinances and recycling plans.

  • 7/24/2019 LM-770_OME - MANUAL

    13/40

    2 - 5

    2.9 Labels

    The following labels provide safety information about each part.

  • 7/24/2019 LM-770_OME - MANUAL

    14/40

    3CONFIGURATIONS

    Screen

    Digital display

    Sign changeover switch +/-

    S/C measuring changeover switch

    ADD diopter switch

    Right switch

    Left switch

    Print switch

    Lens Table

    Nosepiece

    Prism compensator

    Target rotating ring

    Diameter measuring scale

    Target

    Indicator

    Corona target

    Cross line target

    [Fig.2 Screen & Digital display]

    [Fig.1 Front view]

  • 7/24/2019 LM-770_OME - MANUAL

    15/40

    3 - 2

    Screen

    It is graduated with 1 prism diopter step con-

    centric circle scale and 5 step angle scale.(Refer Fig.2)

    Target

    It is consisted from corona and cross lines tar-

    gets. (Refer Fig.2)

    Digital display

    Measured values of SPH (sphere), CYL (cyl-

    inder), ADD and AXIS are displayed on thispart. (Refer Fig.2)

    Indicator

    In case measuring cylinder lens, it displays to

    which side of cross lines target is focused.

    (Refer Fig.2)

    According to this indicator, measure the lens

    by focusing either short or long cross lines tar-

    get.

    Prism compensator

    In case measuring lens with over 5 prism di-

    opter, this prism compensator is used not to

    go off the target from the field of view of

    screen.

    By moving the prism compensator knob for

    left or right, direction of moving target is

    changed and by rotating the knob, moving vol-

    ume is also changed.

    Lens table

    When measuring lenses in frame, touch the top

    part of frame to the lens table.

    Nosepiece

    This is a basic point of measurement, there-

    fore special plastic material is used not to give

    any scratches on lens, nor to be distorted.When measuring contact lens, replace this for

    nosepiece of contact lens.

    Target rotating ring

    By rotating this ring, angle of target can be

    changed.

    Diameter measuring scale

    By touching lens edge to the lens table, diam-

    eter of lens can be read.

    Sign changeover switch +/-

    This is used when opposite cylinder value is

    desired to be read against the display of mea-

    sured values.By pressing this switch again, it returns to

    original display.

    S/C measuring changeover switch

    Each pressing this switch change the SPH or

    CYL measuring mode. Also returning to SPH

    measuring mode from ADD measuring mode

    or memory display mode, can be made by this

    switch.

    ADD measuring switch

    By pressing this switch, it makes ADD measur-

    ing mode. In this mode, CYL display at digital

    display part is changed to ADD display.

    Right switch

    Left switchThese switches are used to memorize the left

    and right measured values in the framed lens.

    Also after memory, it returns to SPH measur-

    ing mode.

    Print switch

    By pressing this switch, measured values which

    are displayed at digital display part, are printed

    out and returns to SPH measuring mode.

    In case measured values are in the memory,

    contents of memory are printed out and

    memory is erased. This works also as feeding

    printer paper by continuous pressing.

  • 7/24/2019 LM-770_OME - MANUAL

    16/40

    3 - 3

    Lens Table Lever

    Printer cover

    Exterior interface connector

    Lamp adjusting screws

    Power source connector

    Main switch

    Diopter handle

    Axis marking lever

    Lens holder lever

    Lens holder

    Ink cartridge

    [Fig.3 Side view]

    [Fig.4 Rear view]

  • 7/24/2019 LM-770_OME - MANUAL

    17/40

    3 - 4

    Axis marking lever

    This is used for marking axis on the lens.

    Lens holder lever

    In the state no holding lens, raise this lever a

    little to release the click and down the lens

    holder to hold the lens. After measurement of

    the lens, the lever is raised up until it get click

    sound and fixed at upper part.

    Lens holder

    To avoid giving scratches on the lens, plasticsare provided at the tips of lens holder. After

    setting the lens, some movement can be made.

    Ink cartridge

    Lens table lever

    Lens table moves back and forth by this lever.

    Diopter handle

    Main switch

    Exterior interface connector

    Accessory equipment connected to the ana-

    log and digital interfaces must be certified ac-

    cording to the representative appropriate na-

    tional standards (for example, IEC 60601-1.)

    Further more all configurations shall comply

    with the system standard IEC 60601-1-1. Ev-

    erybody who connects additional equipment

    to the signal input part or signal output part

    configured a medical system, and is therefore

    responsible that the system complies with the

    requirements of the system standard IEC

    60601-1-1. If in doubt, consult the technical

    service department or your local representa-

    tive.

    Lamp adjusting screws

    After replacing lamp, if part of target becomes

    dark or target blurs along with flowing whenin defocus, there screws are used.

    Power source connector

    Both fuse holders and voltage selector are pro-

    vided.

    Printer cover

  • 7/24/2019 LM-770_OME - MANUAL

    18/40

    4OPERATING PROCEDURES

    4.1 Basic Process of Operation (SPH, CYL Lenses)

    NOTE Before measurement, be sure that the nosepiece is mounted correctly.

    Incorrect mount of the nosepiece may on sccuracy of measurement.

    1. Put the main switch ON by pressing marked side. When focusing target, confirm

    that SPH value is 0.00. Also confirm that target is in the center of screen.

    2. Put the lens, facing its concave side on the nosepiece.

    3. Raise the lens holder lever at upper end

    once and then down it to hold the lens. (See

    Fig.5)

    4. Turn the diopter handle to make corona

    target on the screen in sharp focus.

    5. Move the lens to coincide the center target

    to the screen center.

    * If the corona target flows, proceed to 8.

    Fig.5

    Fig.6

  • 7/24/2019 LM-770_OME - MANUAL

    19/40

    4 - 2

    6. Make fine adjustment by the diopter handle

    to get sharp focus of the target.

    * SPH value is displayed at digital displayedpart. (Refer Fig.7)

    7. Press down the axis marking lever to print

    the marks.

    * Center point shows the optical center.

    * Proceed to 16.

    8. Turn the target rotating ring and align

    the short cross lines target in parallel to

    corona flow lines. (Refer Fig.8)

    9. Turn the diopter handle and make the

    short cross lines target in sharp focus.

    10. Press the S/C measuring changeover switch

    , it changes to CYL measuring mode with

    beep sound.

    * Display of indicator changes to .

    (Refer Fig.9)

    Fig.9

    Fig.8

    Fig.7

  • 7/24/2019 LM-770_OME - MANUAL

    20/40

    4 - 3

    11. Turn the diopter handle and make the

    long cross lines target in sharp focus. (Refer

    Fig.10)

    * If indicator flickers, rotate either the target

    or line for 90 and return to the SPH

    measuring mode by pressing S/C measuring

    changeover switch.

    Then proceed over again from the process 5.

    However in case not mark axis, flickering

    indicator has nothing to do with.

    * SPH, CYL and AXIS are displayed on the

    digital display . (Refer Fig.10)

    12. Turn the target rotating ring and coincide

    the axis value displayed on digital part to the

    prescription value.

    13. Turn the lens and coincide its focus to the

    long cross line target. (Refer Fig.11)

    14. Coincide the lens center to the target and

    mark the axis by pressing down the axis

    marking lever.

    15. Press the S/C measuring changeover switch

    , it returns to SPH measuring mode with

    beep sound. (Refer Fig.12)

    16. Hold the lens by hand and raise the lens

    holder lever up until it gets click sound.

    17. Remove the lens.

    Fig.12

    Fig.11

    Fig.10

  • 7/24/2019 LM-770_OME - MANUAL

    21/40

    4 - 4

    4.2 Prior Measurement

    4.2.1 Preparation

    1. Power source cord is firstly connected to LM-770 and then to the socket.

    2. Confirm that scale of prism refractive power

    on the prism compensator show 0. (Refer

    Fig.13)

    4.2.2 How to set the lens

    In case measuring lenses in frame, touch the top part of frame to the lens table and hold it by lens

    holder.

    Lens table comes out by pulling the lens table lever this side.

    Scale of prism refractive power

    Fig.13

  • 7/24/2019 LM-770_OME - MANUAL

    22/40

    4 - 5

    4.3 Measurement

    4.3.1 Measurement of single vision lens

    1. Confirm that indicator shows the short lines mark ( ) and turn the diopter handle to get

    sharp focus to the short cross lines target.

    If indicator shows long lines mark ( ), change it by S/C measuring changeover switch.

    Coincide center of the lens, and read SPH value on the digital display when the target is in the

    sharp focus. In case sphere lens, whole target get sharp focus. Then measuring is now finished.

    2. If the lens astigmatism, corona target flows in a cylinder. Align the short cross lines target

    in parallel to the direction of flowed lines and make it in sharp focus.

    Then change it to CYL measuring mode by S/C measuring changeover switch. Indicator displays

    long lines mark and turn the diopter handle to make the long cross lines target in sharp focus.

    Now read the values of SPH, CYL and AXIS on the digital display. Further reading of +/- of

    measured CYL value can be changed by sign changeover switch.

    NOTE While in measuring astigmatism lens, indicator may flicker. But if axis marking is not

    made, this has nothing to do with.

  • 7/24/2019 LM-770_OME - MANUAL

    23/40

    4 - 6

    4.3.2 Measurement of ADD diopter

    When measuring bifocal lens, this is used to see how much refractive power for near vision isadded on the refractive power for far vision.

    After measuring far vision of refractive power, input refractive power of far vision by pressing

    ADD measuring switch and the difference is displayed by the standard basing on it.

    [Process]

    1. Measure the far vision part of the lens in the process stated in 4.3.1.

    2. Press the ADD measuring switch.

    * Lamp at ADD on digital display part is lit

    and it becomes ADD measuring mode.

    (Refer Fig.14)

    3. Shift the lens to come its near vision part

    (small lens) on the nosepiece. (See Fig.15)

    NOTE

    When shift the lens, raise the lens holder

    lever a little up.

    4. Focus the cross lines target to the side shown by indicator

    NOTE

    If the target dose not come to the center of screen due to the effect of prism of the lens,

    use the prism compensator and bring the target to the center of screen to enable getting

    better sharp focus.

    * On the digital display, SPH values for vision and ADD are displayed.

    Fig.15

    Fig.14

  • 7/24/2019 LM-770_OME - MANUAL

    24/40

    4 - 7

    4.3.3 Measurement of prism refractive power

    Measuring of prism refractive is divided in three ranges.

    a. Within 5 prism diopter which can be read on the scale of screen.

    b. Range from 5 prism diopter to 15 prism diopter.

    c. Range from 15 prism diopter to 20 prism diopter. For the case of b. and c., use the

    prismcompensator.

    When measuring prism refractive power in framed lenses, let the patient wear the glass and mark

    the each center of pupil. Then put the lens on the nosepiece to come the mark on the center ofnosepiece.

    a. In case range within 5 prism diopter.

    [Process]

    1. Turn the diopter handle and make focusing

    of the target.

    2. Read the volume of shift and direction of the

    target center from the center of screen.

    Shifted volume in the scale of concentric circles

    is the prism refractive power and direction read

    on the angle scale is the direction of prism base.

    * If you turn the target rotation ring to go the

    center line of long cross lines target through

    the center of screen, then it makes easy

    reading of the angle scale. (Refer Fig.16) 3Base 200

    Fig.16

  • 7/24/2019 LM-770_OME - MANUAL

    25/40

    4 - 8

    b. In case range within 15 prism diopter.

    [Process]

    1. Turn the diopter handle and display the expected SPH value on the digital display.

    2. Operate the knob of prism compensator and bring the center of target to the center of

    screen.

    NOTE

    If the target is not in the screen, set the angle scale of prism compensator at 0 or 90

    and turn the prism compensator knob. It makes easy finding the target.

    3. Turn the diopter handle to focus the target.

    4. Operate the prism compensator knob and

    coincide the center of target accurately to the

    center of screen.

    5. By the prism refractive power scale and angle

    scale of prism compensator, read each prism

    refractive power and direction of prism base.

    (Refer Fig.17)

    NOTE

    If the scale of prism refractive power is

    indicated in orange color, direction of

    prism base is the value of angle scale plus

    180.

    3Base 70Fig.17

  • 7/24/2019 LM-770_OME - MANUAL

    26/40

    4 - 9

    c. In case range over 15 prism diopter.

    [Process]

    1. Turn the diopter handle and display theexpected SPH value on the digital display.

    2. Coincide the scale of prism refractive power

    of prism compensator at 15 prism diopter

    and move the knob for left or right, target

    crosses in the screen. (See Fig.18)

    3. Stop the movement of target when its center

    comes to the nearest position to the screen

    center and read the prism refractive power

    by the scale of concentric circle as stated a.

    4. On this read value, 15 prism diopter of the prism compensator must be added. This total

    value is the prism refractive power of the measuring lens.

    5. Direction of prism base is read at the angle scale of prism compensator as explained at

    b.

    d. In case measuring astigmatism lens with prism.

    Measure both cylinder and prism refractive prism.

    NOTE

    By using prism compensator and bring the target in the center of screen. Then

    measurement can be made easily.

    4.3.4 Measurement of hard contact lensMeasurement of hard contact lens can not be made by using normal nosepiece. Please change it

    for nosepiece exclusive for contact lens.

    Fig.18

  • 7/24/2019 LM-770_OME - MANUAL

    27/40

    4 - 10

    4.4 Marking Axis

    This is used show the center of optical axis (by the center of 3 points marks) and also to showdirection of axis and base for cylinder lens etc.

    4.4.1 Making sphere lens

    1. Focus the target and adjust lens position to coincide the target center to the screen center.

    2. Press down the axis marking lever and print 3 point on the lens in parallel for lens table.

    4.4.2 Marking cylinder axis

    1. According to the 4.1 Basic process of operation, measure the lens. In this moment, if the

    indicator flickers, rotate the lens or target 90 and return to SPH measuring mode. Repeat

    measuring again from process 5.

    2. Turn the target and coincide the axis which displayed on digital display, to the angle of

    axis marking.

    * In case marking on cylinder axis, coincide to 0 and in case marking on prescribed value,coincide to the prescribed value.

    3. Turn the lens to focus for the long cross line target.

    4. Make centering lens and press down the axis marking lever to print axis.

  • 7/24/2019 LM-770_OME - MANUAL

    28/40

    4 - 11

    4.4.3 Marking phoria lens

    1. Shift the target center for the direction of prism refractive power (base direction +180)by prism compensator.

    2. Set the lens.

    3. Make axis marking in the process of 4.4.1 or 4.4.2.

  • 7/24/2019 LM-770_OME - MANUAL

    29/40

    4 - 12

    4.5 Memory Function

    This is the function to memorize the measured values of left and right lenses in the frame anddisplay these data separately or print these data out continuously.

    Also in order to transmit the data to exterior system ophthalmic instrument, only memorized data

    can be transmitted. Therefore such data must be memorized in advance.

    4.5.1 How to memorize

    Measure the lens after confirming the LED of right/left switches are OFF. After taking

    measurement, by pushing the said switches, measured data can be memorized. Also when measuredvalue are memorized, LED is lit.

    * When measured value are memorized, it returns to SPH measuring mode.

    * Memory is one each for left and right.

    4.5.2 How to display the contents of memory

    While in SPH measuring mode, by pressing LED lit switch RIGHT/LEFT, memorized contentsare displayed. (Memory display mode)

    When you want to read out the result of ADD measurement, press further ADD measuring switch.

    Also in the case memory display mode, indicator is OFF and displayed side LED of the switch

    is flickered.

    NOTE

    Memory display function is only effective when it is in SPH measuring mode. In other

    modes, If you press the memorized switch again, contents is changed.

  • 7/24/2019 LM-770_OME - MANUAL

    30/40

    4 - 13

    4.5.3 How to clear the memory

    When memory is displayed, if you press the switch again, the contents of displayed memory canbe erased.

    Also, if you press the print switch, memorized contents of both right and left can be erased.

    (In case including a printer, memorized contents are erased after the contents are printed out.)

    In case measurement is made newly, memorized data should be erased in advance and then measure.

    4.6 Print out

    By pressing PRINT switch, measured data displayed on the digital display part are printed out.

    Also in case measured value are memorized by RIGHT/LEFT switches, by pressing PRINT

    switch memorized contents are printed out, and memorized contents are erased.

    Further, when printing is out, it returns to SPH measuring mode from CYL measuring mode, or

    ADD measuring mode, or memory display mode.

  • 7/24/2019 LM-770_OME - MANUAL

    31/40

    4 - 14

    4.7 Setting Step and Changing the Setting of CYL

    Reading

    4.7.1 How to set the step

    Displayed step of refractive power can be changed by the following process.

    1. Open the printer cover.

    2. Refer Fig.19, put the DIP switch of No.3

    ON makes 0.12D step and put it OFF

    makes 0.25D step.

    NOTE

    Do not touch other DIP switches.

    3. Once put the power source switch OFF or

    press CYL measuring switch.

    4.7.2 How to set CYL reading

    Setting of CYL readings (Plus reading, Minus reading and Mix reading) can be made by the

    following process.

    1. Open the printer cover.

    2. Refer Fig.20 about DIP switches, CYL

    reading can be set by changing DIP switchposition as bellows.

    NOTE

    Do not touch other DIP switches.

    1-WS 2-WS

    gnidaeR- NO FFO

    gnidaeR+ FFO NO

    gnidaeRxiM NO NO

    Fig.20

    Fig.19

  • 7/24/2019 LM-770_OME - MANUAL

    32/40

    4 - 15

    4.8 Auto-off Function

    Auto-off function works when the lensmeter isnot used about three minutes. To return, press

    one of any switches or rotate either diopter

    handle or target rotating ring. (Refer Fig.21)

    Memorized data are not erased even if the

    lensmeter is in auto-off state.

    Fig.21 Auto-off state

  • 7/24/2019 LM-770_OME - MANUAL

    33/40

    5.1 Replace the Lamp

    1. Put the main switch off and pull the powercord out from socket.

    2. Remove the bottom lamp house cover.

    3. Pull the lamp out from the lamp socket.

    CAUTION

    As the illumination bulb can get quite hot,replace it when it is cool.

    Otherwise, it may cause an injury.

    4. Set the new lamp in the process of opposite

    order.

    Do not hold glass part of lamp by naked hand,

    hold it with cloth.

    CAUTION

    In replacement of the lamp, do not touch the inside of the instrument.

    An electric shock may occur.

    5. After replacement of lamp, if the part of target on the screen becomes dark or target blurs

    with flow when it is in defocus, adjust it by turning lamp adjusting screws at the brightest.

    Fig.22

    5MAINTENANCE

    Lamp

  • 7/24/2019 LM-770_OME - MANUAL

    34/40

    5 - 2

    5.2 Replace the Fuses

    1. Put the main switch off and pull the powercord out from the socket.

    2. Remove the fuse folder.

    Push both sides of the fuse holder with a thin

    screwdriver to release the fixing lips, and pull

    the holder out.

    CAUTION The fuse holder is also used as a voltage

    selector. Be careful not to change the

    numbers show in the voltage indication

    window. If the voltage setting dose not

    match the supplied power voltage, it may

    cause a fire or malfunction.

    3. Replace the fuses.

    Replace the fuses and set the fuse holder back into position.

    CAUTION

    Be sure to use the correct fuses specified on the label beside the fuse folder.

    When replacing the fuses, be sure to replace both of them with the new ones.

    If the wrong fuses are used, the instrument may malfunction and may cause a fire.

    If the fuse blows again, do not touch the internal structure. Contact your authorized

    distributor.

    Fig.23

    Voltage indication window

    Fixing lip

    Fixing lip

  • 7/24/2019 LM-770_OME - MANUAL

    35/40

    5 - 3

    5.3 Replace the Ink Cartridge

    When markings become faint, replace the ink cartridge.Use a specified cartridge for refills.

    1. Remove the C type ring by pressing a tip of

    the ink cartridge.

    * During work, take care to prevent a possible

    loss of the C ring.

    (Using a tweezers can avoid a loss of the C

    ring when attaching it.)

    * As a spring and cartridge are fallen out when

    the C ring is removed, take care to prevent

    possible loss of them.

    2. Remove the screw from disattached ink

    cartridge with a precis ion phil l ips

    screwdriver.

    3. Attach the screw removed in preceding step

    to a new ink cartridge.

    * The head of screw must be 0.5 - 1mm outside

    from the cartridge.

    4. Insert the spring and the new cartridge into

    the marker holder, and fit the C ring between

    the screw and ink cartridge.

    Fig.26 Screw

    Fig.25

    Marker holder

    Ink cartridge

    Spring

    Fig.24

    C type ring

    Fig.27

  • 7/24/2019 LM-770_OME - MANUAL

    36/40

    5 - 4

    5.4 Replace the Printer Paper

    When printer paper is exhausted, replace it with new one.

    1. Open the printer cover.

    2. The tip of printer paper is inserted through

    the shot. (Refer Fig.28)

    3. Keep pressing PRINT switch and feed the

    printer paper.

    4. Pass the printer paper through the exit ofthe printer cover and closed the cover.

    NOTE

    Once printer paper is feeded, do not pull

    the paper for opposite direction.

    5.5 Cleaning of Protective GlassPeriodically, detach the nosepiece and remove the dust and dirt on the protective glass.

    1. Remove the nosepiece upward.

    2. Blow dust off on the protective glass with a

    blower brush and then, wipe it off with a soft

    cloth or lens cleaning paper gently.

    Fig.28

    Fig.29

    Nosepiece

    Protective glass

  • 7/24/2019 LM-770_OME - MANUAL

    37/40

    5 - 5

    5.6 Cleaning Exterior

    When the exterior parts of the instrument become dirty, wipe with a dry and soft cloth. Forstubborn dirt, immerse the cloth in a neutral detergent, wring well and wipe. Finally wipe off with

    a dry and soft cloth.

    NOTE

    Never use an organic solvent such as paint thinner to clean the instrument as it may ruin

    the surface of the instrument.

  • 7/24/2019 LM-770_OME - MANUAL

    38/40

    Type : Projection type - Magnification 40X

    Digital display - Sphere, Cylinder, Axis, ADD

    Measuring target : Cross-line and coronaSphere power

    Measuring range : 0 to 25D

    Measuring unit : 0.12 or 0.25D steps

    Accuracy :

    Cylinder power

    Measuring range : 0 to 9.87D

    Measuring unit : 0.12 or 0.25D steps

    Cylinder power : +, or MIX

    Addition power

    Measuring range : 0 to 9.87DMeasuring unit : 0.12 or 0.25D steps

    Cylindrical Axis angle

    Measuring range : 0 to 180

    Measuring unit : 1 steps

    Accuracy : 1

    Prism diopter power

    Measuring range : 0 to 5

    0 to 20 (in case using prism compensator)

    Measuring unit : 1 stepsAccuracy :

    6SPECIFICATIONS

    Values in prism diopters()

    Measuring range Accuracy

    > 0

    50.1

    > 510

    0.2

    Values in diopters(D)

    Measuring range Accuracy

    < 0 -5 > 0

    +5 0.06

    < -5 -10 > +5

    +10 0.09

    < -10 -15 > +10+15 0.12

    < -15 -20 > +15+20 0.18

    < -20 > +20 0.25

  • 7/24/2019 LM-770_OME - MANUAL

    39/40

    6 - 2

    Prism Axis angle

    Measuring range : 0 to 360

    Measuring unit : 1 steps

    Measurable lens diameter : 9 to 100mm

    Illumination lamp : 6V 20W G4, Halogen lamp

    Power source

    Voltage : 100, 120, 220, 230Vac 10%

    Frequency : 50/60 Hz

    Power consumption : 40VA

    Dimensions and weight : 165(W)315(D)420(H)mm 7kg

    Environmental condition

    Temperature : +10 to +40 C (In usage)

    -40 to +70 C (In storage/Transference)

    Humidity : 30 to 85 % (Non-condensing) (In usage)

    10 to 100 % (In storage/Transference)

    Accessories : Spare lamps (1 pc.)

    Dust cover (1 pc.)

    Printer paper (1 pc.)

    Nosepiece for contact lens (1 pc.)