Linac4 Instrumentation Day Faraday Cups. In order to measure the beam intensity three Faraday Cups will be installed in different places LEBT Low energy Spectrometer Low energy Moving Diagnostics Bench for the three DTL tanks High energy. Faraday Cups Scheme of Linac4. - PowerPoint PPT PresentationTRANSCRIPT
09-05-2007 C. Bal AB/BI/ML 1
Linac4 Instrumentation DayFaraday Cups
In order to measure the beam intensity three Faraday Cups will be installed in different places LEBT
Low energy Spectrometer
Low energy Moving Diagnostics Bench for the three DTL
tanks High energy
09-05-2007 C. Bal AB/BI/ML 2
Faraday CupsScheme of Linac4
09-05-2007 C. Bal AB/BI/ML 3
Faraday CupsSpecifications LEBT
Low Energy: 95 keV Commissioning &
Operation Retractable Acceptance FC
ø80 mm Acceptance Tank
ø100 mm
Parameter Quantity
Ion Species H-
Pulse Length 400 µs
Rep. Rate 2 Hz
Beam Power 6 W
Pressure 10-5 mbar
Source Current 80 mA
Bias Voltage 3 kV
09-05-2007 C. Bal AB/BI/ML 4
Faraday CupsMechanical Design LEBT
09-05-2007 C. Bal AB/BI/ML 5
Faraday CupsMechanical Design LEBT
Bias Ring, Collector and Housing made of St. Steel Vespel Insulators for electrical isolation Conical Collector Use of metal seals only Flanged High Voltage Feedthrough Flanged Signal Feedthrough Linear Pneumatic Drive (travel 102mm) Flange DN150CF (øouter 202mm) Parts slide into Housing and will be closed by
dismountable Face and Back Plate
09-05-2007 C. Bal AB/BI/ML 6
Faraday CupsSpecifications Spectrometer
Low Energy: 95 keV Temporary installation Commissioning only Fixed Similar design as for LEBT
09-05-2007 C. Bal AB/BI/ML 7
Faraday CupsSpecifications Moving Bench DTL
High Energy: 40 MeV Commissioning only Retractable Requires Cooling Requires Energy
Absorber (levels to be defined)
Design under development
Parameter Quantity
Ion Species H-
Pulse Length 400 µs
Rep. Rate 2 Hz
Beam Power 5.1 kW
Pressure 10-7 mbar
Linac Current 40 mA
Bias Voltage 3 kV
09-05-2007 C. Bal AB/BI/ML 8
Faraday CupsElectronics
Local read-out by means of a scope
Remote controlled at a later stadium
50 OHM
ADC INPUT
200 V BIAS PS
100 K
Absorber Faraday Cup
Bias Ring