leica tcs sp5 microscope manual

14
1 Cell Imaging Unit – UIC LEICA TCS SP5 MICROSCOPE MANUAL

Upload: others

Post on 02-Dec-2021

11 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: LEICA TCS SP5 MICROSCOPE MANUAL

1

Cell Imaging Unit – UIC

LEICA TCS SP5 MICROSCOPE MANUAL

Page 2: LEICA TCS SP5 MICROSCOPE MANUAL

2

TABLE OF CONTENTS

ANATOMY OF THE LEICA SP5 CONFOCAL MICROSCOPE 3

INITIALISATION PROCEDURE 4

SP5 MICROSCOPE 5

1. RIGHT SIDE OF MICROSCOPE (DETAILED VIEW) 5

2. FRONT SIDE OF MICROSCOPE (DETAILED VIEW) 6

3. LEFT SIDE OF MICROSCOPE (DETAILED VIEW) 6

4. CONTROL PANEL BUTTONS 7

STARTING THE LAS SOFTWARE 8

1. CONFIGURE FOR ACQUISITION 8

1.1 THE LEFT PANEL ALLOWS YOU TO CONFIGURE THE IMAGE

ACQUISITION PARAMETERS

9

1.2 THE RIGHT PANEL ALLOWS YOU TO LOAD A PRE-MADE

CONFIGURATION OR TO CREATE YOUR OWN ONE

10

1.3 CREATE A SEQUENTIAL ACQUISITION 10

2. ACQUIRE IMAGES 11

3. ACQUIRING A Z-STACk 12

4. SAVE AND EXPORT YOUR IMAGES 13

SHUTDOWN PROCEDURE 14

Page 3: LEICA TCS SP5 MICROSCOPE MANUAL

3

ANATOMY OF THE LEICA SP5 CONFOCAL MICROSCOPE

1. Research inverted microscope 1. Transmitted light source

2. Scanhead 2. Stage

3. TCS Workstation 3. CLSM port

4. Microscope control unit 4. z -Drive

5. Laser and power supply 5. Display

6. Computer table 6. Controller

7. Microscope table 7. x-Drive

10. Control panel 8. y-Drive

11. Control monitor 9. z-Drive

12. Image Monitor 10. Eyepieces

13. Supply control 11. Condenser head

12. field diaphragm

5

3

1

4

2

11

7

6

12

10

9

8

5

6

7

2

Page 4: LEICA TCS SP5 MICROSCOPE MANUAL

4

INITIALISATION PROCEDURE

1. Turn on the “PC”, “SCANNER” and “FAN” switch.

1. Turn the LASER key on

2. For fluorescence samples, turn on the Fluorescence light power source (check

first if the lamp is not hot), otherwise leave it off and proceed

3. Turn on the microscope

4. Log on into your personal/group account

5. Start the “Leica Confocal Software -LAS” available both as a Desktop

shortcut and the Windows start menu.

Page 5: LEICA TCS SP5 MICROSCOPE MANUAL

5

SP5 MICROSCOPE

The microscope is fully automated which means that you have now button in stead of

knobs. It you know what you do in a conventional microscope, it takes 5 minutes to

understand what the buttons are for.

1. RIGHT SIDE OF MICROSCOPE (DETAILED VIEW)

Displays the following information:

1. Microscope stage up movement (DO NOT USE).

2. When used with 7 is sets the maximum and minimum threshold (DO NOT

USE).

3. Microscope stage down movement (DO NOT USE).

4. Speed of knob 7

5. Change objective to high magnification.

6. Change objective to low magnification.

7. Focus knob.

Page 6: LEICA TCS SP5 MICROSCOPE MANUAL

6

2. FRONT SIDE OF MICROSCOPE (DETAILED VIEW)

Displays the following information:

1. Change cube for blue excitation filter cube – FITC, GFP

2. Change cube for green excitation filter cube – RFP, CY3

3. Use if for cutting some “umpolarized light” from the laser (might be

interesting for good DIC)

4. Empty position

5. Empty position

6. Empty position

7. Fluorescence shutter

3. LEFT SIDE OF MICROSCOPE (DETAILED VIEW)

Page 7: LEICA TCS SP5 MICROSCOPE MANUAL

7

Displays the following information:

1. Change from transmitted light to transmitted light detector (not confocal

detector)

2. Enables switching from oil to dry objectives and vice-versa, minimizing the

risk of having oil in dry objectives

3. Move one filter cube position

4. Change from confocal mode to visualization mode

5. Change aperture diaphragm just for florescence

6. Change field diaphragm just for transmitted light

7. Change light intensity in both fluorescence and transmitted light, depending

on which mode you are

8. Toggles between transmitted light and fluorescence

4.CONTROL PANEL BUTTONS

Displays the following information:

1. Smart gain

2. Smart offset

3. Rotation

4. Pinhole

5. Zoom

6. Z position

Page 8: LEICA TCS SP5 MICROSCOPE MANUAL

8

THE LAS SOFTWARE

1. Double-click on the LAS AF icon to start the program.

2. In the window coming up check that MACHINE is selected under

CONFIGURATION and START.

3. In the software window, go the CONFIGURATION tab, LASER and turn on

the lasers you need

4. Set the percent power on the Argon laser (25-30% is recommended).

1. CONFIGURE FOR ACQUISITION

It’s time now to configure the software for a confocal acquisition:

Page 9: LEICA TCS SP5 MICROSCOPE MANUAL

9

1.1 THE LEFT PANEL ALLOWS YOU TO CONFIGURE THE

IMAGE ACQUISITION PARAMETERS:

• Acquisition type. Standard mode is XYZ.

• Image size in pixel. Standard size is 512x512 and acceptable for

publication, although in some instances 1024x1024 may be desirable

for publication. Lower and higher resolutions are available in the drop-

down menu, but beware, the higher the resolution the larger the file.

The pixel size and image size are shown.

• Acquisition speed in Hz de 10-1000 Hz. Standard speed is 400Hz.

Faster scanning speeds may be required for imaging live cells or

moving specimens. For even faster scanning, you can activate

bidirectional scanning (“Bidirectional X”). This doubles the scanning

speed, but requires “Phase” adjustment.

• Pinhole size. The default setting of 1 Airy Unit (AU) is the

recommended setting. (The pinhole setting of 1 Airy Unit will display

a different physical pinhole aperture size on the control panel for each

objective). For faint specimens, increasing the pinhole size may help,

but the images you are capturing may no longer be truly “confocal”

images.

• Zoom level. Refer to notes on “Zooming and selecting a Region of

Interest”. By zooming in the pixel size will decrease. For high

resolution imaging optimal pixel size according to the Nyquist

criterion should be ensured. Note that the pixel size can also be

decreased by enhancing the format.

• Level of averaging used to reduce the background noise. It’s

unnecessary to use an average of more than 8x. To determine whether

you should average your image more times, try capturing with more

averages and see whether the image is improved. Line averaging is

faster and therefore recommended for live cells. For fixed specimens, it

doesn’t really matter whether you use line averaging or frame

averaging or a combination, although frame averaging may be better

suited for some applications to allow for a brief recovery from photo-

bleaching.

Page 10: LEICA TCS SP5 MICROSCOPE MANUAL

10

1.2 THE RIGHT PANEL ALLOWS YOU TO LOAD A PRE-MADE

CONFIGURATION OR TO CREATE YOUR OWN ONE

1.2.1 If you want to use a pre-made configuration, just use the drop

down menu « Load/Save single settings »:

1.2.2 If you want to create your own configuration, follow this step-by-

step:

• Activate the laser lines you need (« fluo » and/or « Visible »).

• Tune their output power (usually between 15 and 50 % depending on the line

and the sample that is used).

• Activate the PMTs (detectors) that you need (check « Active »). Use first the

PMTs located just under the spectral band you want to detect.

• Choose the bandwidth according to your staining for each PMT.

1.3 CREATE A SEQUENTIAL ACQUISITION:

This mode allows you to avoid « crosstalk » problems by acquiring each channel one

after the other.

Click on the « sequential » button then create a configuration only one

channel with only one excitation wavelength.

Add a new sequence by clicking on the « + » button, the preceding sequence being

used as a template for the new one. Create the configuration for the next channel. You

can add as many sequences you want. To delete a sequence, click on the « - » button,

and the last one created will be deleted.

Click here to select a pre-made

configuration

Page 11: LEICA TCS SP5 MICROSCOPE MANUAL

11

2 ACQUIRE IMAGES

Once the acquisition configuration is done you can start to acquire images.

The intensity of the image can be controlled by the laser intensity and the gain

(amplification of photons) and offset (threshold) adjustment of the detector.

6.1 Start preview scanning by clicking on LIVE.

6.2 Enhance the gain until you see an image. By default the gain is on

0V (black screen) when you start. Adjust the focal plane with the

z-knob until you see the brightest plane of interest.

6.3 Adjust gain and offset: to aid adjustment of gain and

offset choose the «glow-over-under» display by clicking

into the QLut icon (left border on right screen).

6.4 In the display mode the saturated pixels are represented by blue

dots and the black signals (grey level 0) are shown in green.

While scanning adjust the offset and gain so that there are a few green pixel in the

background and some blue speckles in the object. The aim is to have the whole range

of grey levels in the image.

Page 12: LEICA TCS SP5 MICROSCOPE MANUAL

12

3 ACQUIRING A Z-STACK

Activate the Z-stack window. The cube represents the specimen. The yellow

plane marks the current z-position. This position is specified in the field below

the cube (z position µm).

7. 1 Set the z-stack range: Select the top and the bottom of the sample

and define them by clicking into the Begin and End arrow which will than

turn red.

7.2 Set z-step size: By default the software calculates the optimal z-

step size and number or steps (accept this frequency for subsequent

deconvolution or 3-D reconstruction). However, often such a high z-imaging

Page 13: LEICA TCS SP5 MICROSCOPE MANUAL

13

frequency is not necessary. You can also define the number of images of a z-

stack and the distance between them manually.

4 SAVE AND EXPORT YOUR IMAGES

The captured images displayed in the “Experiment” menu are not safely saved until

you save them. It is therefore strongly recommended that you start saving your images

from the very first capture you make, and that you update your save after each image

capture.

4.2 You can open the properties pages of a picture and restore the acquisition

parameters that were used:

Page 14: LEICA TCS SP5 MICROSCOPE MANUAL

14

SHUTDOWN PROCEDURE

1. Log out and shut down the computer in windows.

2. Turn off the microscope.

3. Turn off the camera.

4. If you used the HBO light source and there is no one using the microscope for

the next hour or you’re sure that the next user won’t use this light source, then turn the

lasers off (in the key) and the fluorescence light power source off. Also turn off the

“PC” and “SCANNER” switch. Turn off “FAN” switch after 10 min (for laser

cooldown)

5. If you used any immersion lenses be sure to clean them.