laboratory training · proper operating and cleaning procedures (tutorial) general handling of a...
TRANSCRIPT
Laboratory Training
Summer Internship 2016 ACRM
Pipetting
• Volumes: 1000µL, 100µL, 10µL • 15 repeats each • Acceptable variation 5%
Method • Pipette distilled H20 onto a balance and
record the weight
Microbeads
• Concentrations: High, Low• 3 repeats each• Acceptable variation 15%
Method • Fill microcell and perform count in 4 fields• Calculate average and concentration• Validate against reference values
Sperm Concentration
• Evaluate 10 Semen samples • Acceptable variation 20%
Method • Fill microcell and perform count in 4 fields • Calculate average and concentration • Validate and compare to mentor’s results
Sperm Motility
• Evaluate 10 Semen samples• Acceptable variation 20%
Method • Fill microcell and perform count in 4 fields• Calculate average and % motility• Validate and compare to mentor’s results
Sperm Vitality
• Prepare 2 different smears • Evaluate 5 reference slides • Acceptable variation 15%
Method • Evaluate each slide x 2 • Count viable and non-viable cells (100 sperm) • Calculate average and concentration • Validate and compare to reference values
Data Recording
• Record all information inyour lab book
• Transfer to day sheet
• Populate soft copy (Excel)
• Print and have it verified
Summer Internship Bench Research Training
The interns will observe and practice the following techniques prior to the actual start of their original research projects.
Laboratory safety:
• Explain the dress code that needs to be observed (lab coats, closed shoes, etc.), designated workareas, and the basic safety procedures to be followed.
Pipetting:
• Explain the different types of pipettes (transfer pipette, micropipette, pipettor), when theyshould be used and how they should be operated.
• Exercise - Pipette 3 volumes (1000µL, 100µL and 10µL) x 15 repeats
Microscopy:
• Explain the various types of microscopes, the function of different components as well as theproper operating and cleaning procedures (Tutorial)
General handling of a semen sample:
• Explain and demonstrate the basics to be performed when a semen sample was received:Incubation, Recording of information, Liquifaction, transfer, volume, pH, Viscosity, Color
Microbeads:
• Explain the loading of a microcell, setting up of the microscope (x20 phase contrast), countingtechnique and calculations
• Exercise – Count 2 concentrations x 3 each
Sperm count and Concentration:
• Explain the loading of a microcell, setting up of the microscope (x20 phase contrast), countingtechnique and calculations
• Exercise – Analyse 10 x samples
Vitality
• Explain the principles of employing a dye exclusion staining technique for viabilitymeasurement, preparing the sample, making of the smear, coverslip slide, and evaluating thesample
• Exercise – Prepare and score x 5 slides
Morphology:
• Explain the principles of morphology, the preparation, staining and evaluating of slides
• Demonstrate – Slide preparation, staining, scoring
• Exercise – Only identified students will receive further training
TUNEL
• Explain the principles of TUNEL as well as the calculations and preparation method.
• Exercise – Only identified student will be receive further training
ROS
• Explain the principles of ROS measurement
• Demonstrate – Perform the sample preparation, machine setup and calculation
• Exercise – Only identified student will be receive further training
ORP
• Explain the principle of sORP and cORP
• Demonstrate – Perform sample preparation, machine setup and calculation
Specialized Techniques
• Techniques specific to projects (e.g. Acrosin activity, Western Blot analysis, Proteomics, HOS, etc.) will be explained by the particular mentors to the individual(s) responsible in each group
Summer Internship 2016
Bench Research Training Proficiency Report
Intern:
Mentor:
Pipetting Proficiency Pipette: P-1000 Reading Actual
volume (µL)
Measured volume
(µg)
Absolute Difference
Percentage Difference
(%)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Average Coefficient of Variance (CV)
Acceptable variation: 5% Met the criteria: Yes / No Retraining completed: Not Applicable / Yes / No Verified: Approved: (Mentor) (Director)
Pipette: P-200 Reading Actual
volume (µL)
Measured volume
(µg)
Absolute Difference
Percentage Difference
(%)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Average Coefficient of Variance (CV)
Acceptable variation: 5% Met the criteria: Yes / No Retraining completed: Not Applicable / Yes / No Verified: Approved: (Mentor) (Director)
Pipette: P-200 Reading Actual
volume (µL)
Measured volume
(µg)
Absolute Difference
Percentage Difference
(%)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Average Coefficient of Variance (CV)
Acceptable variation: 5% Met the criteria: Yes / No Retraining completed: Not Applicable / Yes / No Verified: Approved: (Mentor) (Director)
Counting Microbeads: Low concentration Reading Reference
Count (x106/ml)
Measured Count
(x106/ml)
Absolute Difference
Percentage Difference
(%)
1
2
3
Average
Coefficient of Variance (CV)
High concentration Reading Reference
Count (x106/ml)
Measured Count
(x106/ml)
Absolute Difference
Percentage Difference
(%)
1
2
3
Average
Coefficient of Variance (CV)
Acceptable variation: 15% Met the criteria: Yes / No Retraining completed: Not Applicable / Yes / No Verified: Approved: (Mentor) (Director)
Sperm concentration: Reading Reference
Count (x106/ml)
Measured Count
(x106/ml)
Absolute Difference
Percentage Difference
(%)
1 2 3 4 5 6 7 8 9 10 Average Coefficient of Variance (CV)
Acceptable variation: 20% Met the criteria: Yes / No Retraining completed: Not Applicable / Yes / No Verified: Approved: (Mentor) (Director)
Motility: Reading Reference
Motility (%)
Measured Motility
(%)
Absolute Difference
Percentage Difference
(%)
1 2 3 4 5 6 7 8 9 10 Average Coefficient of Variance (CV)
Acceptable variation: 20% Met the criteria: Yes / No Retraining completed: Not Applicable / Yes / No Verified: Approved: (Mentor) (Director)
Vitality (E&N):
Reading Reference Viability
(%)
Measured Viability
(%)
Absolute Difference
Percentage Difference
(%)
1
2
3
4
5
Average
Coefficient of Variance (CV)
Acceptable variation: 15%
Met the criteria: Yes / No
Retraining completed: Not Applicable / Yes / No
Verified: Approved: (Mentor) (Director)