LARGE GROUP MANUFACTURERS UPDATE SATURDAY APRIL 14, 2012 8:00 A.M. – 10:00 A.M. Topics New BSC Model NU-481 Use and Service of DC ECM Motors New Slanted A2 – NU-475, 477, 480 Miscellaneous Service Updates

Upload: lillian-wilson

Post on 26-Mar-2015

222 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

LARGE GROUPMANUFACTURERS UPDATE

SATURDAY – APRIL 14, 20128:00 A.M. – 10:00 A.M.

LARGE GROUPMANUFACTURERS UPDATE

SATURDAY – APRIL 14, 20128:00 A.M. – 10:00 A.M.

TopicsNew BSC Model NU-481

Use and Service of DC ECM MotorsNew Slanted A2 – NU-475, 477, 480

Miscellaneous Service Updates

TopicsNew BSC Model NU-481

Use and Service of DC ECM MotorsNew Slanted A2 – NU-475, 477, 480

Miscellaneous Service Updates

Page 2: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NUAIRE WEBSITENUAIRE WEBSITEReceiving TB’s and have access to Website’s secure area?

Let us know and provide current contact info if you need access.

Page 3: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NUAIRE WEBSITENUAIRE WEBSITE

Website – Support Manuals Technical Articles Technical Bulletins Service Companies Training (On-line Registration) Service Procedures Warranty Information Cannot log on – Call Service

for help 800-328-3352

Page 4: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-475, NU-477, NU-480/481A2 SLANTED FRONT

NU-475, NU-477, NU-480/481A2 SLANTED FRONT

TouchLink Control (NU-480/481)

NU-475 (NU-425 Controls)

NU-477 (NU-437 Controls)

60 fpm Down-flow

105 fpm Inflow

DC-ECM Motor

3” Constricted Inflow Method

Permanent Pressure Plenum

Page 5: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

The CellGard HD (Hazardous Drug) ES (Energy Saver) Class II, Type A2 Biological Safety Cabinet offers personnel, product and environmental protection for handling of hazardous particulate drugs and powders

CELLGARD™ HD NU-481 ESCELLGARD™ HD NU-481 ES

Page 6: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CELLGARD™ HD NU-481 ESAIR FLOW

CELLGARD™ HD NU-481 ESAIR FLOW

Clean Oval Canister HEPA Filtered Air

Clean HEPA Filtered Air

Room Air

Contaminated Work Surface Air

Page 7: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

The Ease of Bag-In / Bag-Out Remove Work Surface Remove paper catch grill over each oval HEPA prefilter Seal contaminated interior of each oval HEPA prefilter Remove four (4) fasteners per each oval HEPA prefilter Bag-Out used oval HEPA prefilter and replace

CELLGARD™ HD NU-481 ESCELLGARD™ HD NU-481 ES

Page 8: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING

PAO Wand

Filter Gard

Page 9: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING Set the NU-481 into calibration mode

Go to Sensor Setup / Calibrate Downflow Sensor / Blower Duty Cycle

Set Blower Duty cycle to the RPM at the current calibrated airflow velocity

Keep blower at this RPM (constant) throughout testing process

Remove the work tray

Page 10: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTINGINTERNAL REFERENCE CALCULATIONS

Model Size *Supply Area (ft2)(m2) Inflow Area (ft2)(m2) 8” Access

Inflow Area (ft2)(m2) 10” Access

400 8.53 (.242) 2.38 (.221) 3.22 (.299)

500 10.74 (.304) 3.15 (.293) 4.05 (.376)

600 12.95 (.367) 3.92 (.364) 4.89 (.454)

*Measured 4 Inches (102mm) above the bottom edge of the window

Laskin Nozzle Concentration Formula

# Nozzles x 135 CFM x 100 ug/L ChallengeDownflow (CFM) + Exhaust (CFM) Concentration (ug/L)

# Nozzles x 229 CMH x 100 ug/L ChallengeDownflow (CMH) + Exhaust CMF) Concentration (ug/L)

=

=

Page 11: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING Filter Gard is used to direct the

PAO from the wand into one area of prefilter at a time

Filter Gard can be used for initial certification and after 1st annual certification

Replace HEPEX Gard (B-481-8691) Fiberglass rods can be re-used

Page 12: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING

CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER

INTEGRITY TESTING Move PAO Wand throughout filter, rotating the filter guard

Leak is detected if a sustained leak rate greater than .005 is detected at the challenge port on top of the cabinet

Page 13: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

ECM 5.0 Genteq Motor

1450 RPM

10” Forward Curved Fan (was 9”)

PWM (DC signal) varies the motor speed Motor Control, Monitors RPM and torque which delivers

constant air volume as HEPA Filters load

NU-437, NU-440, NU-475, NU-477, NU-480, NU-481 (A2 units)

NU-427 (B1), NU-430, NU-435 (B2) - DC ECM Motor is standard Optional on NU-425 (A2)

ELECTRONICALLY COMMUTATED MOTOR

ELECTRONICALLY COMMUTATED MOTOR

Page 14: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

DC ECM CONTROL SYSTEM A2DC ECM CONTROL SYSTEM A2

NU-425 – ES (Energy Saver) - Optional & Field Upgradeable

NU-475, NU-629 – ES (Energy Saver) – Standard

Page 15: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

425 TYPE DC SPEED CONTROLLEROPTIONAL ON NU-425

425 TYPE DC SPEED CONTROLLEROPTIONAL ON NU-425

12 Volt DC Power supply

Automatic Constant Airflow volumeThru motor rpm feedback

Manual Airflow volume adjustment (POT)

On-board jumper positions for correct cabinet size

Speed Cont. locks in setting after 1 Minute idle or certifier sets via push button & LED turns off

Page 16: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

425 TYPE DC MOTORSPEED CONTROLLER SYSTEM

425 TYPE DC MOTORSPEED CONTROLLER SYSTEM

Page 17: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

425 TYPE DC MOTORSPEED CONTROLLER SYSTEM

425 TYPE DC MOTORSPEED CONTROLLER SYSTEM

Page 18: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

SPEED CONTROLLER JUMPERSSPEED CONTROLLER JUMPERS

3’ 0 0

4’ 1 0

5’ 0 1

6’ 1 1

J2 J3

0 = Not Jumped1 = Jumped

Page 19: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

GENERAL AIRFLOW BOARD (GAF)GENERAL AIRFLOW BOARD (GAF)

NU-437, NU-477 (A2) – ES (Energy) -Saver Standard

GAF Also used in NU-813 , NU-126 , NU-619 - Change Station & All Isolators models

Page 20: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

AC Motor Output: Older Series NU-437

AC Impeller Output: Older Series NU-813; NU-126; NU-619 and Isolators

DC ECM Motor Output: Std. on Current NU-437, NU-477

DC EC Impeller Output: Std. on All Isolators (S-10^) NR/NTE/PR; NU-813, NU-126, NU-619

New Software Release (Rev 1.5.0001), STB0201 Corrects increase motor speed (PWM signal) due to site

specific events, (i.e. power fluctuations, frequent on/off cycles) STB0202: RF Upgrade Installation Instructions

GAF CONTOR BOARDMULTIPLE BLOWER CONTROLS

GAF CONTOR BOARDMULTIPLE BLOWER CONTROLS

Page 21: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

GAF MAIN CONTOR BOARD LAYOUTGAF MAIN CONTOR BOARD LAYOUT

AC Blower Output

AC Blower POT.

Board ACInput

Connection

EC #1VDC T.P.’s

EC Connector & POT #1

EC Connector and POT #2

EC #2VDC T.P.’s

DC ECM Blower 4 PinMotor Control

Connector

AC T.P.’s

Page 22: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

GAF AC/D MOTOR / IMPELLERAIRFLOW ADJUSTMENTS

GAF AC/D MOTOR / IMPELLERAIRFLOW ADJUSTMENTS

AC motor – Airflow adjusted via the POT located under heat-sink

DC ECM motors - Airflow adjusted via sequence of control panel keys actions. Cal mode = press & hold Alarm Silence key, hit FL light on, light off key;Increase Airflow = UV key and decrease = FL light key

DC EC impellers utilize another set of potentiometers,on GAF board, Std. all Isolators (S-10^)

Page 23: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

GAF Control board EC Impeller ConnectionsGAF Control board EC Impeller Connections

Page 24: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-437 (SERIES 60^) / 477 (A2)AIRFLOW CALIBRATION

NU-437 (SERIES 60^) / 477 (A2)AIRFLOW CALIBRATION

The DC-ECM System works like this:

1. AC voltage is applied directly to the DC motor’s control module

2. The motor control module (Microprocessor) converts the AC to DC for Internal electronics and to power the motor

3. In addition, there is a second electrical cable that goes from the Main Control Board’s speed controller DC output to the motor. This generates a pulse-width-modulation (PWM) 2.5 – 8 VDC control signal and returns an RPM feedback signal from the motor to the speed controller CKT.

Page 25: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

GAF CONTROL BOARD JUMPERSDC ECM CONTROL SIGNALS

GAF CONTROL BOARD JUMPERSDC ECM CONTROL SIGNALS

Nite Care Setback

Page 26: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

GAF BOARD AC OR DCJUMPER WIRE

GAF BOARD AC OR DCJUMPER WIRE

Cut Jumper Wire for AC Motor

Page 27: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-437 / 477 (A2)CALIBRATION AIRFLOW

NU-437 / 477 (A2)CALIBRATION AIRFLOW

Once the airflow calibration is complete – calibrate the FlowGard Airflow Monitor

1. Press and hold ( ) and ( ) arrow keys simultaneously for 3 seconds until the center green LED blinks.

2. Allow the Unit to run for a minimum of 2 minutes.

3. Press RESET Key to enter the nominal airflow value; the Green LED will stop blinking and display normally.

This sets the high/low alarm limits +/-0.13 w.g. from nominal. Reference the operation manual for the NU-437/477 for more in-depthinformation and calibration procedures. If you require help, or havequestions, contact NuAire’s Technical Support Depart.

Page 28: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

BSCC TOUCHLINK AIRFLOWCONTROL SYSTEM

BSCC TOUCHLINK AIRFLOWCONTROL SYSTEM

Page 29: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

BSCC CONTROL BOARD BSCC CONTROL BOARD

DC ECM Output: NU-440 (S-60^); NU-480, NU-481 (A2)NU-430/435 (B2) (S – 60^);

NU-427 (B1) (S-60^)

Note: Control Board is Backwards Compatible to all TouchLinks cabinets

Page 30: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

TOUCHLINK SOFTWARE UPDATES TOUCHLINK SOFTWARE UPDATES

New Revision Notifications are sent out to all service organizations

New Revisions are available free of charge

Use RF484 to order latest revision and cable connections

Latest TouchLink Software 61N.138N. (June 2011) Selection for AC or DC Motor Screen Saver display update Air Flow Sensor – Comm/Sensor error messages Down Flow Sensor - Averaging time/Alarm time adjustable Filter Life available in the Calibration/Service Menu Electrical power harmonic distortion compensating software

Page 31: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NEW TOUCHLINKCALIBRATION CARDNEW TOUCHLINK

CALIBRATION CARDTouchLink - Calibration and Reset ProceduresThe TouchLink System is a Liquid Crystal Display (LCD) that is menu driven and thus you need to enter specific menus to accomplish the maintenance tasks required. Refer to the appropriate O & M Manual for additional in depth information on a specific model. This is to be used as a reference guide, there may be some variances due to software revision levels.

From the main display hit the Menu ICON to advance to other displays. Hit the Menu ICON to return to previous displays

Cabinet Type/Motor Type- Press MENU from the main display- Select CALIBRATION/SERVICE, enter password 9876, and press ENTER- Select CABINET TYPE- Select the type of cabinet; i.e. A2, B1, B2. Types NOT ending with a letters (D,E,G) are for 115 VAC systems.- Select the model of cabinet- Hit SAVE- Press MENU- Press SET MOTOR TYPE (If available)- Select AC or DC MOTOR-Press SAVE

Find Sensors- Used for both calibration and sensor replacement- Press MENU from the main display- Go into CALIBRATION /SERVICE, enter password 9876, press ENTER- Select SENSOR SETUP- Select FIND DOWNFLOW SENSOR- Power down cabinet- Remove the 4 pin exhaust probe connector from the main control board- Power up and tap displays “Power Loss Alert”- Select FIND NEW DOWNFLOW SENSOR- Press LEARN- Acknowledge that down flow sensor is found. If not, repeat previous steps.- Power down cabinet- Reconnect 4 pin exhaust probe connector back on main control board

Page 32: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CLASS II TYPE A2CANOPY EXHAUST TRANSITIONS

CLASS II TYPE A2CANOPY EXHAUST TRANSITIONS

Page 33: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

CANOPY EXHAUST TRANSITIONSCANOPY EXHAUST TRANSITIONS

Traditional Fixed Flow NU-906 / 907 / 908 (NU-475 / 477 / 480 / 481) NU-916 / 917 / 918 (NU-425 / 437 / 440) NU-926 / 927 / 928 (NU-407 / 602)

New Variable Flow NU-911 (NU-475 / 477 / 480 / 481) NU-921 (NU-425 / 437 / 440)

Page 34: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

VARIABLE FLOW CANOPY (VFC)VARIABLE FLOW CANOPY (VFC)

Low profile (5” housing, 8½” overall)

Built-in low airflow alarm

Variable air gap range (25 to 250cfm)

Page 35: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

VFC FUNCTION / OPTIONSVFC FUNCTION / OPTIONS

Page 36: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

Note: VFC Setup adjustment should be made with cabinet set atnominal airflows

1. Adjust slide plates and/or exhaust volume desired so average velocity measured in the center of the (3) test points of either side is between 150fpm (.76mps) and 550fpm (2.8mps)

2. Verify containment with the capture of smoke

3. Verify low exhaust alarm with loss of smoke capture with in 15 seconds

VFC SETUP INSTRUCTIONSVFC SETUP INSTRUCTIONS

Page 37: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

VFC INTERFACE / SERVICEVFC INTERFACE / SERVICE

Alarm contacts

Audible alarm silence contacts

Front panel access for filter testing

Decon options

Page 38: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-911/921 VFC TRANSITIONSNU-911/921 VFC TRANSITIONS

Page 39: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-602NU-602

Page 40: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-602 NEW LIFT SYSTEMNU-602 NEW LIFT SYSTEM

Page 41: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-602 NEW LIFT SYSTEMNU-602 NEW LIFT SYSTEM

Page 42: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

RF Kit to convert older series actuators

Re-Zeros every time unit is fully retracted

NU-602NU-602

Page 43: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-612 / 617 / 619SMALL ANIMAL

CAGE CHANGING STATIONS

NU-612 / 617 / 619SMALL ANIMAL

CAGE CHANGING STATIONS

Page 44: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-612 / 617 / 619 PreventativeMaintenance Check List

NU-612 / 617 / 619 PreventativeMaintenance Check List

Page 45: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-612 / 617 / 619PREVENTATIVE MAINTENANCE

CASTER INSPECTIONREF. STB0185 / STB0214

NU-612 / 617 / 619PREVENTATIVE MAINTENANCE

CASTER INSPECTIONREF. STB0185 / STB0214

Page 46: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-612/617CASTER FAILURES

REF. STB0185

NU-612/617CASTER FAILURES

REF. STB0185

Page 47: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-612 / 617 / 619P.M.C.I.

REF. STB0185 / STB0214

NU-612 / 617 / 619P.M.C.I.

REF. STB0185 / STB0214

Page 48: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

PHARMAGARDPHARMAGARD

Page 49: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

PR-797 – Inspect outer door hinge bolts for wear during normal certification – replace as needed – STB0191

NuAire provides replacement bolts at No Charge –RF525 – keep some in your stock

OUTER DOOR HINGE BOLTSOUTER DOOR HINGE BOLTS

Page 50: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

Inter-chamber door not staying up – RF 508/509/510 Available at N/C, replace during normal certification for improved latching

PR797 Series 3 – 5RF508

NR797 Series 3 – 7RF509 (LH); RF510 (RH)

NTE797 Series 3 – 5RF509 (LH); RF510 (RH)

Current inter-chamber dooris counter balanced

INTER-CHAMBER DOORINTER-CHAMBER DOOR

Page 51: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-797 SHARPS PLUG RETROFIT& SHARPS UPGRADE

NU-797 SHARPS PLUG RETROFIT& SHARPS UPGRADE

Page 52: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

NU-430 SERIES 60^NEW B2 FEATURESNU-430 SERIES 60^NEW B2 FEATURES

Airflow Measurement Procedure Modified Down-flow Performed First Inflow taken with DIM and Supply Blower ON

DC ECM Motor

Permanent Pressure Plenum

Exhaust monitored via Pressure Grid

Page 53: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

DC ECM Motor

Permanent Pressure Plenum

Exhaust monitored via Pressure grid

NU-427 SERIES 60^NEW B1 FEATURESNU-427 SERIES 60^NEW B1 FEATURES

Page 54: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

PRESSURE SWITCH RETROFITVACS REPLACEMENT INSTALL.PRESSURE SWITCH RETROFIT

VACS REPLACEMENT INSTALL.

Page 55: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

PTB0170 TouchLink Software Loading Procedure

STB0136 Verify B1/B2 Exhaust Probe Cal. (S 20 – 49)

STB0185 NU-612/617 Preventive Maintenance

STB0187 Convert NU-425 Series 20^ to DC ECM Motor

STB0191 PR797 Isolator, Outer Door Hinge Bolt (PM)

STB0192 TouchLink, Master Reset Procedure

STB0197 GAF Control Board Usage/Calibration

STB0201/202 GAF Module Software Update (RF541)

TECHNICAL BULLETINS / RF KITSTECHNICAL BULLETINS / RF KITS

Page 56: L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012 8:00 A.M. – 10:00 A.M. L ARGE G ROUP M ANUFACTURERS U PDATE S ATURDAY – A PRIL 14, 2012

STB0204 DC ECM Speed Control Replacement (425 type)

STB0208 TouchLink Software Version 60N.136N

STB0211 TouchLink Software Version 61N.138N

STB0214 Preventative Maintenance Check List

STB0215 Pressure SW-VACS Replacement Installation

STB0216 797 Sharps Retrofit

STB0219 619 Caster Upgrade

TECHNICAL BULLETINS / RF KITSTECHNICAL BULLETINS / RF KITS