Isolated Power PanelsTransfer UnitsFault Location SystemsIsolating EquipmentOperational Theatre Panels
CATALOG2011
Identifies the product as top quality, safe and one step forward of the similar ones.
Symbolizes measurable energy saving products helps to energy continuity.
Identifies smart control logic.
Identifies easy to use products, simplifies the difficult tasks.
Green products, respects to the nature and our future.
Aktif trade mark for Medium voltage switchgears, Switching equipment and Kiosks with high quality and environmentally sensitive
Aktif trade mark for Capacitor Banks, Harmonic Filters, Resistors, Medical Insulated Power Panels, Synchronization and Energy automation Panels with high quality and environmentally sensitive
Aktif trade mark for Measuring, Protection, Automatic Meter Reading, Billing and Energy Management Software.
Aktif trade mark for Measuring, Protection, Control and Power Quality products with high quality, long life and environmentally sensitive
1
B 1 Touch Screen LCD Display 22B 2 LCD Display 24
OPERATIONAL THEATRE CONTROL PANELS
MEDICAL POWER SYSTEMSA 1 Type Of System Earthing 8A 2 Isolated Power Systems 10A 3 Auxiliary Equipment 14
2
Aktif Raylı Sistemler Ltd.
Aktif Railway Systems (ARS) is established in 2008 for the designing and providing of traction systems' substations as well as providing on-board equipment used on vehicles.
The main purpose of ARS is providing engineering services and products to public bodies and companies in this sector. Even though it is a brand new company, ARS managed to complete successfully both Antalya Light Rail System project as well as Istanbul Metro Stinger System projects.
Aksis Enerji Sistemleri Ltd.
Aksis Energy Metering Systems (AKS) is established in 2003 for providing metering solutions like Automatic Meter Reading, Billing Software in every level of energy sector. The company is focused on saving and efficient using of energy by its Meter Management and Loss & Leakage tracing solutions. AKS answers all expectations in power systems with high quality products and custom design software solutions. AKS customers create over 100 million Euro energy bills using different Metering, Management and Billing methods of AKS.
Aktif Mühendislik Ltd.
Aktif Engineering (AMDT) is established in 1996 Measuring, Protecting and Quality of Electricity. 2200m2 headquarter of AMDT is located in Istanbul consisting of 35 employees. Since establishment AMDT provides high quality engineering, supplying and start-up services for the power quality, energy saving, energy metering, protection. AMDT provides standard and custom designed solutions to domestic and international customers, with over 15 years of experience.
Company Profile
3
Coet S.r.L.
Since establishment in 1962 in Milan, COET has been producing AC/DC switching equipments and panels that are used safely by its customers thanks to the COET's work on industrial and electric equipment and patents obtained in this field. All manufactured products are the result of long technical research and development work. This way, the company is known for the originality and diversity of its planned solutions. COET has been working on Industrial and railway systems and has become one of the leading companies in the field of supplying switching equipments for traction substation.
Aktif Elektroteknik A.Ş.
The company is established to manufacture Switchgears and Kiosks under the name of “Setas” in 1981. First in 2008, company merged to Aktif Group as Aktif Elektroteknik (AET) and then the company becomes international after significant participation of Italian Friem S.p.A in 2009. 9000m2 factory of AET is located in Ankara consisting of over 100 employees. AET operates with its 30 years of manufacturing experience, modernized machinery line, ever increasing know-how, experienced Turkish and Italian R&D teams, quality products type tested by the leading accredited European laboratories and the vision of new ideas to meet with future expectations.
Company Profile
Friem S.p.A.
Friem manufactured over 40 million ampere and 1000 power rectifiers for the worldwide since it's established in Milan at 1950.
Having technical knowledge of energy and electro mechanics along with ability of complete design of conversion systems, FRIEM provides also high voltage DC insulators, DC switches, anodic control and protection products and auxiliary equipments like polarization products to its customers. Friem is a share holder of Aktif Elektronik in Turkey and also COET S.r.L in Italy.
4
Introduction
Aktif Group is in continuous development with its customer oriented activities, advanced engineering know-how, R&D works, software developing abilities and with the continuous support of loyal customers that are attained and protected by precise work and ethical principles of the company.
The group companies have ISO 9001 since 90s and our factory has ISO 14001 Environment and ISO 18001 Occupational Health and Safety Certificates.
Technology
The most important activities of Aktif is R&D and increase the Employee quality which the investment amount of these activities are more than average.
All mechanical projects are designed by 3D-CAD platforms. Power flow, test and quality works are calculated by Worldwide accepted simulation software and ActWin software as well as developing software for different platforms.
In order to increase the production quality and capacity all machinery within the facility are renewed with latest technology equipment in 2009.
Vision
To increase the number of our loyal customers in global market, hence increasing the market share and becoming reputable, reliable and preferred company as worldwide with our;
prominent quality difference,customer oriented approach,innovative activities,open minded approach.
Mission
By the help of our followings mission is continuing to announce product quality and knowledge of our company and country in best possible way by meeting rising customer expectations.
open minded approach high quality policy innovative ideas constantly improved processes polished employees knowledge based decisions
Company Profile
5
High Performance, Durable and Simple
Endurance of our products, which are produced with high level of awareness and accurateness in addition to follow international standards, ensures an outstanding performance for the users. Therefore our products are considered as top of the line products.
Visual simplicity and being user friendly are the design criteria for the software and hardware of our products. Our products have the most simple and functional features for emergency applications as well as persistence of habits and customer satisfaction are basics in design, production and shipment.
Security
The human safety and security are the main concerns in all of our products.
Design, Interlock logics and documentation of our products are implemented, manufactured and tested in order to reach the highest safety level.
Furthermore, services and site works are done according to human safety rules by taking into account the dangerous of electricity.
Service Continuity
The service continuity means efficiency of power consequently profitability of the business.
Our products are designed and manufactured in order to ensure energy sustainability and provide the best service availability.
This policy is the cornerstone of our orientation and training programs and it is fully applied by our technical and a d m i n i s t r a t i v e s t a f f .
Local & Remote Monitoring
Energy continuity and efficiency can be ensured only by monitoring of equipments local and remotely.
Our products are designed with remote monitoring and management features by the help of our software development abilities on different platforms and hardware capabilities. This will ensure the saving and profitability, directly.
Company Profile
A 1 Type Of System EarthingA 2 Isolated Power Systems A 2.1 Isolated Power Panels A 2.2 Double Source Isolated Power Panels A 2.3 Double Source Isolated Power Panels With Fault Location SystemsA 3 Auxiliary Equipment A 3.1 Isolation Transformers A 3.2 Isolation Monitoring Devices A 3.3 Remote Alarm Panels A 3.4 Communication Equipments
6
8101112131414151618
Medical PowerSystem
Medical Power SystemsA
8
Power in medical locations is evaluated in the framework of electrical safety. Basic conditions for electrical safety are independent from medical professional or private life locations. When it viewed deeper; it is seen that additional safety precautions in the places where they are used medically should be taken. Because, patient is out of conscious and reaction conditions; bound to medical equipments continuously and the most important thing is that equipments are the equipments keeping the patients alive and rescuing his live. Due to this reason; IT systems are used in group 2 rooms which are the medical locations, electrical energy is critical.
A 1 Type of System Earthing
The designations for the various system result from the relationship of the power system to earth and relationship of exposed conductive parts of electrical installation to earth. The meaning of the letters is follow:
First letter: Relationship of the power system to earth:
T = Direct connection of one point to earth,
I = Either all active parts isolated from earth or one point connected to earth via an impedance
Second letter:
Relationship of the exposed conductive parts of the electrical installation to earth:
T = Exposed conductive parts are directly earthed, independently of any earthing of a point of the power supply system,
N = Exposed conductive parts are directly connected to the earth electrode of the power supply system
The remaining letters:
Description to the arrangement of neutral conductor and the protective conductor:
C = Neutral conductor and protective functions combined in a single (PEN) conductor S = Conductor separated from the neutral or the earthed conductor
In IT systems all active parts are insulated from earth or one point is connected to earth through sufficient high impedance. The exposed conductive parts of electrical installation are earthed either; individually or in groups or collectively connected to the earthing point of the system.
a n A k ti f G r o u p C o m p a n y
IT system
TN system
TT system
L1
L2
L3
PE
L1
L2
L3
N
PE
L1
L2
L3
N
PE
9
Standards;
Standards contain basic information necessary for planning and establishment of electrical facilities in hospitals and places used as medically. Many countries conform to national and international standards in isolated power systems. Some of them are as follows:
Advantages & Benefits
Affecting of personnel and patient from electrical shock is prevented. Isolation fault does not cause power interruption. Fault current in electrical system is decreased to levels that are not critic. Continuity of power for medical locations is provided. Electrical fault monitoring is taken under guarantee. Fire risk caused by the faulty current is prevented. It is ensured to protect high cost medical equipments against electrical faults. Examination information obtained at a long of time vanishes as a result of electrical interruption. Additional faulty currents are gathered and they are prevented to reach critic values. Suspension for operating is prevented.
International IEC 60364-7-710Austria ÖVE/ÖNORM E 8007Belgium T 013Brazil NBR 13543China GB16895/GB50333-2002Finland SFS 6000France NFC 15-211Germany DIN VDE 0100-710Hungary MSZ 2040Ireland ETCI 10.1Italy CEI 64-4Netherlands NEN 3134Norway 64/747/FDISRussia GOST P 50571Slovakia STN 33 2140South-Africa SABS 051Spain UNEE 2 0-615-80 Turkey TS IEC 60364-7-710UK BS 7671 GN7/HTM2007
a n A k ti f G r o u p C o m p a n y
Medical Power Systems A
10
A 2 Isolated Power Systems
Isolated power systems are used in group 2 rooms in medical locations, and they consist of auxiliary devices and test combinations such as isolated power panels and insulation transformer, isolation monitoring device, remote alarm panels.
When isolation fault occurs with isolated power systems, it is ensured not to cause the system to energy interruption by opening protection equipments, to continue medical electrical equipments to their functions, not to decrease fault currents to non-critical values and not to live problem during operating by preventing energy interruption.
Isolation level is monitored continuously in isolated power systems and converts the fault to alarm signal by detecting at adjustable level provided that it is not less than 50 kΩ. The alarm in question takes part on control panels in operating theatres and on a section where nurse or technical personnel can easily hear and then control and intervene it in other rooms by monitoring over two remote alarm panels. Alarm panels used for the purpose of remote monitoring take part in room of personnel in duty or in technical service room.
Transformer temperature and load current information out of isolation level is continuously monitored and it is ensured to detect immediately and to give alarm when it gets out of nominal values. Interruption does not occur in power supply. System continuity is ensured by preventing a second insulation fault that may occur in the system.
The following medical locations that carrying vital importance in hospitals and where medical devices are directly connected to patients are defined as group 2 room in accordance with IEC 60364-7-710 standard and isolated power system is used in these rooms.
Operating theatres Intensive care rooms Operating preparation rooms Operating recovery rooms Anesthetic rooms Heart catheterization rooms Angiographic examination rooms Premature baby rooms
Examples of Group 2 locations
Operating room
Premature
Intensive care unit
a n A k ti f G r o u p C o m p a n y
Medical Power SystemsA
11
A 2.1 Isolated Power Panels
These panels are isolated power system having applications of IT systems in medical locations realizes supply of each group 2 room in compliance with IEC 60364-7-710 standard. Isolated power panel which are equipped with the most important equipment of the system such as isolation transformers and isolation monitoring device provides supply of electric equipments in group 2 rooms, wall plugs of devices supplying life support and of critical loads, operating control panel, operation lamp and similar lamps.
All the aforementioned loads are protected against short circuit current by using double-pole fuses in accordance with IT system. Although isolated power panels have their own cooling system, it removes warm-up problem thanks to effective air circulation.
Brand name ASSETType IGP 0710/1P-XXStandards IEC 60364-7-710: 2002-11Rated power of outgoing 3.15 / 4 / 5 / 6.3 / 8 /10 kVAPower supply unit A single phase lineRated voltage 230 Vac Rated frequency 50 / 60 HzRated insulation level 3 kV / 1 minInput protection gL fuseSecondary output voltage 230 VacOutput protection 2-pole fuses Monitoring isolation resistance with LCD displayAlarm output isolation fault, excessive load, excessive temperatureFunctional Test advanced isolation fault Leakage current to enclosure < 0,5 mAResponse range 50 - 500 kΩIsolation fault detection time < 1 sOperating temperature -5........+50 °CStorage temperature -25......+60 °CPanel sizes 500x500x1750 mmDistribution outlet 6-12 pieces as standardCooling system With double fan Protection class IP 31Color RAL 7035 / RAL 9003
Technical Specifications
a n A k ti f G r o u p C o m p a n y
Medical Power Systems A
12
A.2.2. Double Source Isolated Power Panels
Although double line isolated power panels have all features of isolated power panels, they are supplied from two different sources. These panels include a switchover module to where they are switchover automatically in order to select the suitable one when voltage and current values from two different supplies are considered.
Switchover module realizing switchover operation within times defined by IEC 60364-7-710 standards and foreseeing uninterruptible supply of loads monitors continuous supply of two lines and realizes the switchover automatically when voltage decreases under 0,9xUn and exceeds 1,15xUn. Thus, any energy interruption that may occur from failure of power supply or from supply cables is prevented and uninterrupted supply of critic loads is provided.
Brand name ASSETType IGTP 0710/1P-XXStandards IEC 60364-7-710: 2002-11Rated power of outgoing 3.15 / 4 / 5 / 6.3 / 8 / 10 kVAPower supply unit Double single phase linesRated voltage line 1 230 Vac Rated voltage line 2 230 Vac Rated frequency 50 / 60 HzRated insulation level 3 kV / 1 minInput protection gL fuse Secondary output voltage 230 VacOutput protection 2 pole fuses Monitoring isolation resistance with LCD display Alarm output isolation fault, excessive load, excessive temperature, switchover failureFunctional Test advanced isolation fault and switchover test Leakage current to enclosure < 0,5 mAResponse range 50 - 500 kΩIsolation fault detection time < 1 sSwitchover time < 200 msOperating temperature 5........+50 °CStorage temperature 25......+60 °CPanel sizes 500x500x1750 mmDistribution outlet 6-12 pieces as standardCooling system With double fan Protection class IP 31Color RAL 7035 / RAL 9003
Technical Specifications
a n A k ti f G r o u p C o m p a n y
Medical Power SystemsA
13
A 2.3 Double Source Isolated Power Panels With Fault Location Systems
Double line isolated power panels with insulation fault location unit are the most comprehensive isolated power panels formed from equipments providing determination of line based speed of any isolation failure in group 2 rooms. Test current generator, isolation failure evaluator and these panels equipped with toroidal current transformers determine from where the failure is sourced in a short time and provide realizing the intervention with minimum time loss via rapid communication and examination capacities of isolation transformers, isolation monitoring device, load current transformer and switchover module. The process is completed by determining the failure as a result of evaluation of current signal formed by test current generator and size current transformers provided that it forms a response signal after determining a failure by isolation monitoring device and by switchover it to remote monitoring devices.
Brand name ASSETType IGTHP 0710/1P-XXStandards IEC 60364-7-710: 2002-11Rated power of outgoing 3.15 / 4 / 5 / 6.3 / 8 / 10 kVAPower supply unit Double single phase linesRated voltage line 1 230 Vac Rated voltage line 2 230 Vac Rated frequency 50 / 60 HzRated insulation level 3 kV / 1 minInput protection gL fuse Secondary output voltage 230 VacOutput protection 2 pole fuses Monitoring isolation resistance with LCD display Alarm output isolation fault, excessive load, excessive temperature, switchover failureFunctional Test advanced isolation fault and switchover test Leakage current to enclosure < 0,5 mA Response range 50 - 500 kΩIsolation fault detection time < 1 sSwitchover time < 200 msOperating temperature 5........+50 °CStorage temperature 25......+60 °CPanel sizes 500x500x1750 mmDistribution outlet 6 - 12 pieces as standardCooling system With double fan Protection class IP 31Color RAL 7035 / RAL 9003
Technical Specifications
SEB
L1
L2
a n A k ti f G r o u p C o m p a n y
Medical Power Systems A
14
A 3 Auxiliary Devices
A 3.1 Isolating Transformers
Asset medical isolating transformers are designed compatible with IEC 60364-7-710 standards for power supply of single phase IT systems and produced in accordance with EN 61558-2-15 standards. Constant angles are isolated from transformer core thanks to a static display placed among primary and secondary coils. Asset medical isolating transformers having galvanic isolated coils provide possibility of monitoring of temperature thanks to PTC thermistors placed into coils. Asset medical isolating transformers designed at 6 different power value between 3,15...10 kVA have high excessive load capacities.
IT 0710-3,15IT 0710-4IT 0710-5IT 0710-6,3IT 0710-8IT 0710-10
265280280280280320
195195205220235235
325370325325325420
TYPE
Size and Weights A (mm) B (mm) C (mm)
200240240240240270
155155165180195195
8,5x178,5x178,5x178,5x178,5x178,5x17
D (mm) E (mm) F (mm) G (mm) H (KG) Power220220230245260280
5157616574
100
3,15 kVA 4 kVA 5 kVA 6,3 kVA 8 kVA 10 kVA
Power range 3.15 / 4 / 5 / 6.3 / 8 / 10 kVAFrequency 50 / 60 HzPrimary Voltage 230 VacSecondary Voltage 230 VacInrush current (Ie) < 12.InLeakage current < 0,5 mANo-load input current (Io) < 3%No-load output voltage (Uo) < 236 VShort-circuit voltage (Uk) < 3%Operating mode continuous operationConnections separate terminal block Max ambient temp. 40 ° CIsolation class 40/BCooling Air coolingPTC resistor 1 resistor per transformer windingProtection class IP 00Standards IEC 603647-710, IEC 61558-2-15
General Features
IT 0710
Dimension diagram IT 0710 Wiring diagram IT 0710
A B
C
D E
G
F1Ο
a n A k ti f G r o u p C o m p a n y
Medical Power SystemsA
15
A 3.2 Isolation Monitoring Devices
Isolation monitoring devices are one of the important devices of IT system. They determine isolation failures in the system by continuously controlling isolation resistance level between system and earth. At the same time, they provides starting of warning system when critic value is reached by controlling load current and temperature of transformer in accordance with IEC 60364-7-710 standards.
Isolation fault that may form in the system provides minimizing the maintenance and repair by providing signalization of undesired situations such as reaching of load current to limit values and exceeding of transformer the desired heat values and prevents harm of energy interruptions that may occur due to a second failure and failure current at high values onto patients, medical personnel and medical appliances.
Rated impulse withstand voltage 4 kVSupply voltage 230 VacOperating Voltage 0...264 VacFrequency 50/60 HzResponse monitoring value 50....500 kΩMeasuring voltage Um 12 VMeasuring current Im max. 50µAMeasuring time at 1µF ≤ 3 sTest voltage 12-20 VdcInternal resistance Ri ≥ 240 kΩImpedance Zi ≥ 200 kΩLoad current range 5...50AOperating temperature 10 °C....+55 °CCommunication interface RS-485 (BMS protocol)Standards IEC60364-7-710, DIN VDE 0100-710, EN 61558-8
General Features
107TD47
Dimension diagram 107TD47
a n A k ti f G r o u p C o m p a n y
Medical Power Systems A
16
A 3.3 Remote Alarm Panels
In-room alarm indicator panels
Monitoring panels which are visual and auditory alarm devices of isolated power systems realize operating message, failure and alarms of isolation monitoring devices in accordance with IEC 60364-7-710, DIN VDE 0100-710 standards. Alarm device showing isolation resistance level as 3 digits and load current as 2 digits as percent on its display has symbolic display of isolation failure, transformer excessive temperature and excessive load current alarms. All these indicators and alarms are made over two wire connection (BMS-RS485) by isolation monitoring device. Audible alarm arising in case failure signal comes has squelch ability but visual alarm does not appear from the display before the failure is removed.
MK 2007
Dimension diagram MK2007
Supply voltage Us 12...28 Vac/dcFrequency 50 / 60 HzMonitoring isolation resistance, load current rate Alarms Isolation fault, excessive load current, transformer excessive temperatureIsolation resistance 3 digitsLoad current 2 digitsAudible buzzer 1Operating temperature 5 °C....+55 °CWire length ≤ 1200 mCommunication interface RS-485 (BMS protocol)Standards IEC60364-7-710, DIN VDE 0100-710
General Features
a n A k ti f G r o u p C o m p a n y
Medical Power SystemsA
17
Central Alarm Monitoring Panels
Remote monitoring devices unifying features of evaluating the fault current and remote monitoring and testing the alarm indicators in all medical systems in a single device in failure detection unit systems comply with IEC 60364-7-710 standards. They have feature of easy activating the predefined alarm texts with its adjustability of parameters settings on it and with its 4x20 character backlight display having message texts prepared in 20 languages and 200 independent programmable message texts. It has a real time data memory having ability of storing 250 alarm messages with hour and time tags. IT has all features of in-room alarm indicator panel and it receives oxygen, dry air, pressurized air, vacuum, nitrogen peroxide, nitrogen values of medical gas systems and some values of building automation system such as UPS failure, air conditioner failure as digital inputs then it can give failure information. MK 2430
Dimension diagram - Surface-mounting
Dimension diagram-Flush-mounting
Supply voltage 18...28 Vac, 19...30VdcFrequency 50 / 60 HzMonitoring Isolation fault monitoring, Isolation resistance, load current rate Alarms Isolation fault, excessive load current, transformer excessive temperature, device failure, 12 digital contact alarmsIndicator LCD 4x20 charactersOperating temperature 5 °C....+55 °CWire length ≤ 1200 mCommunication interface RS-485 (BMS protocol)Digital Input 12Switching elements 1 changeover contactOperating principle N / C operating Standards IEC60364-7-710, DIN VDE 0100-710
General Features
a n A k ti f G r o u p C o m p a n y
Medical Power Systems A
18
A 3.4 Communication Equipments
It can be possible to read failures through any computer for all systems thanks to communication options of isolated power system. It can be ensured for monitoring ability of all isolated power systems, readability, reporting ability and sending to different users by e-mail thanks to Ethernet converting devices and embedded web server that these devices have used of BMS communication protocol of isolated power systems over local network and internet. This provides possibility of reviewing electric failures and failures due to devices that may occur in all medical locations having IT system installation by all authorized persons who can connect to internet network inside and outside of the hospital.
1- Bender systems or devices with BMS bus, such as RCMS, EDS 2- BMS bus 3- Protocol converter FTC470XDP, BMS bus / PROFIBUS DP 4- Protocol converter FTC470XMP, BMS bus / Modbus RTU 5- Remote alarm indicator and test combination MK2430 6- TM alarm indicator and operator panel 7- Bender BMS bus 8- Protocol converter FTC470XET, BMS bus /Ethernet (TCP / IP), Web server ,OPC interface
9- PC with standard-browser (Internet-Explorer, Firefox, Opera etc.) 10- OPC server in FTCXET 11- OPC client: Axeda Wizcon visualization software 12- OPC client: Touch Panel TPC 13- OPC client: Scada software 14- FTC470XET functionality: E-mail notification via Internet 15- FTC470XET functionality: Operation of Bender system via Web browser 16- FTC470XET functionality: text messaging to mobile phones 17- BMS OPC server 18- PC with software BMS OPC server 19- Protocol converter DI-2 BMS bus (RS 485) / RS-232
a n A k ti f G r o u p C o m p a n y
Medical Power SystemsA
FTC470XET
Ethernet (TCP/IP) Web browser History memory E-mail notification System parameter setting OPC server Data logging Interface to visualization system
Fieldbus
FTC470XDP for PROFIBUS DP
Connection to PROFIBUS DP Control via PROFIBUS Master 8-byte output and input data Query of alarms, measured values and parameters
FTC470XMB for Modbus RTU
Connection to Modbus RTU Control via Modbus Master Illustration of 10 Bender devices with bus capability Query of alarms, measured values and parameters
19
FTC470XET
FTC470XDP / FTC470XMB
a n A k ti f G r o u p C o m p a n y
Medical Power Systems A
B 1 Touch Screen LCD DisplayB 2 LCD Display
20
2224
OperationalTheatreControl Panels
Operational Theatres Control PanelsB
21
B 1 Touch Screen LCD Display
OCP-6 operator control panel is a modern and reliable device designed in order to provide optimal operation conditions that are necessary in operational theatres and to form the most comfortable environment that is possible for surgery team. Touch screen LCD display having easy cleanable IP65 protection standard, room difference pressure and filter pollution level measures are the part of “smart care” checking feature that OCP-6 panel. All electrical control can be done on touch display manually; music broadcast and air conditioner mechanism can be controlled. Communication having high quality sound ensured by digital sound processor with freehand telephone system on the same touch screen LCD display is possible.
Another feature of OCP-6 is that it can do all data exchange necessary for automation and that it can undertake all automation duty in environment that are not central automation by itself.
Type OCP-6Panel material DIN 4301 stainless steelDrawn Power 35 W max.Supply voltage 230 Vac ± % 10Display Touch graphic LCD display (112mmx61mm)Observable parameters Temperature, humidity, difference pressure, filter pollution, gas alarm indicators Audible Alarm Programmable for all observable values Functional Test For gas alarm, electrical heater, all illuminations Operating temperature +5........+50 °CStorage temperature 25......+60 °CAnalog output 8 pieces 0-10 V (10 bit)Analog input 12 pieces 0-10 V (10 bit)Digital output 8 pieces (5 V)Relay output 16 pieces (16 amp/240 V)Cooling Air Protection class IP 61Excessive temp. protection Thermal fuseColor (case) RAL 7035Panel sizes 600x110x400 mmCut-out dimensions 580x380 mm
Technical Specifications
a n A k ti f G r o u p C o m p a n y
22
General Features
Completely microprocessor controlled DIN 4301 stainless front sheet metal Touch graphic display IP 65 class touch panel protection Smart care warning system 3 line general illumination control Luminous intensity control at rate of 100% Control of operation lamp Negatoscope control UV lamp control 1 piece reserve button Operating enter-exit macro buttons Operating room busy lamp control 4 channel music broadcast channel selection Sound amplified at 5 watt power Adjustment control for music sound intensity Digital clock and chronometer 10 channel medical gas alarm panel Temperature, humidity, room difference pressure, hepa filter pollution indicator Temperature monitoring and setting, audible and luminous warning when set value is exceeded. Humidity monitoring and setting, audible and luminous warning when set value is exceeded. Command for Operating room exists/not exist, flow control Damper open-close manual control Anesthetic gas discharge control Electricity heater control Information transfer to automation by RS485 port
a n A k ti f G r o u p C o m p a n y
Operational Theatres Control Panels B
23
B 2 LCD Display
It is a microprocessor controlled appliance that provides measuring and control of heat, humidity, hepafilitre, laminar flow pressure of places such as OCP-5 Operating room control panel, Operating room, operation room and giving possibility of realizing internal illumination, music broadcast, climatization, hour chronometer, negatoscope control on a single panel. Polycarbonate membrane button panel having easy cleanable IP 65 protection standard is a part of “Smart” control feature that OCP-5 panel. All electrical controls can be done manually on polycarbonate membrane button panel.
Another feature of OCP-5 is that it can do all data exchange necessary for automation and that it can undertake all automation duty in environment that are not central automation by itself.
Type OCP-5Panel material DIN 4301 stainless steel Drawn Power 35 W max.Supply voltage 230 Vac ± % 10Display Blue illuminated graphic LCD display (30mmx102mm)Observable parameters Temperature, humidity, difference pressure, filter pollution, gas alarm indicators Audible Alarm Programmable for all observable values Functional Test For gas alarm, electrical heater, all illuminations Operating temperature +5........+50 °CStorage temperature 25......+60 °CAnalog output 4 pieces 0-10 V (10 bit)Analog input 8 pieces 0-10 V (10 bit)Digital output 6 pieces (5 V)Relay output 9 pieces (16 Amp/240 V) Cooling AirProtection class IP 61Excessive temp. protection Thermal fuseColor (case) RAL 7035Panel sizes 710x90x470 mmCut-out dimensions 690x440 mm
Technical Specifications
a n A k ti f G r o u p C o m p a n y
Operational Theatres Control PanelsB
General Features
Completely microprocessor controlled DIN 4301 stainless front sheet metal LCD display 3 line general illumination control Luminous intensity control at rate of 100% Control of operation lamp Negatoscope control UV lamp control 1 piece reserve button Operating room enter-exit macro buttons Operating room busy lamp control 3 channel music broadcast channel selection Sound amplifier at 5 watt power Adjustment control for music sound intensity Digital clock and chronometer 5 channel gas alarm panel Temperature, humidity, room difference pressure, hepa filter pollution indicator Temperature monitoring and setting, audible and luminous warning when set value is exceeded. Humidity monitoring and setting, audible and luminous warning when set value is exceeded. Command for Operating room exists/not exist, flow control Damper open-close manual control Anesthetic gas discharge control Electricity heater control Information transfer to automation by RS485 port Hands-free phone
24
a n A k ti f G r o u p C o m p a n y
Operational Theatres Control Panels B
a n A k ti f G r o u p C o m p a n y
FactoryAkşemsettin mah. Çatalca sk. No:113 06930 Sincan Ankara - TürkiyeTe l : +90 312 269 46 02 Fax : +90 312 269 45 01
Head OfficeBayraktar Blv. Şehit sk. No: 5 Aktif Plaza 34775 Ümraniye Istanbul - TürkiyeTe l : +90 216 314 93 20Fax : +90 216 314 93 60
E-mail : [email protected] : www.aktif.net
E-address
Doc
. Nr.
A
MD
T-IT
111
4/C.
Eng
Ow
ing
to p
erio
dica
l im
prov
emen
ts o
f our
pro
duct
s th
e su
pplie
d pr
oduc
ts m
ay d
iffer
in s
ome
deta
ils fr
om th
e da
ta s
tate
d in
the
pros
pect
us m
ater
ial.
Pub
lishe
d by
Akt
if M
ühen
disl
ik D
›ş T
ic. L
td. Ş
ti. D
ata
subj
ect t
o al
tera
tion
with
out n
otic
e.
D
esig
n &
Cre
atio
n: A
ktif
Gro
up /
Pho
tos:
Akt
if G
roup
- S
tock
.xch
ng &
Shu
tter
stoc
k
Enerji ve Endüstri
Orta Gerilim Hücreler
Trafo ve Dağ›t›m Köşkleri
Mobil ‹stasyonlar
Dirençler
Enerji Kalitesi
T›bbi Güç Sistemleri
Trafo ve Reaktörler
OSOS ve Faturalama
Ölçme ve Yönetim Sistemleri
Senkronizasyon Sistemleri
Sekonder Koruma Röleleri
Kaçak Akım Röleleri
Raylı Sistemler
Enerji Dağ›tım Çözümleri
Depo Alanı Çözümleri
Araçüstü Ekipmanlar